Vous êtes sur la page 1sur 9

Indian J. Anaesth.

2004; 48 (4) :
278 INDIAN JOURNAL OF ANAESTHESIA, AUGUST 2004
278
ORIGINAL ARTICLE

POINT-OF-CARE TESTING IN ANAESTHESIOLOGY AND


INTENSIVE CARE – AN OVERVIEW
Dr. Subhash Gupta1 Dr. Abhijit Bhattacharya2

SUMMARY
The term point-of-care (POCT) testing, also called near patient testing or bedside testing, refers to the performance of diagnostic
tests at or near the patient. These tests are performed outside the confines of clinical laboratories in health care facilities. Many
descriptive titles have been used for POCT, including Point-of-care diagnostics, Alternate site testing, Bedside testing, Ancillary
laboratory testing, near and patient side testing, and Decentralized and Distributed testing.The application of point-of-care testing in
the operating room or intensive care unit is reviewed and its importance emphasized.
Keywords : Point-of-care testing, Anaesthesia, Critical care.

Introduction analysis. After the analysis, reports are communicated to


Traditionally, most of the laboratory tests ordered the treating physician. (figure 1)
from acute areas like operating room (OR) or Intensive This multistep and multiperson central laboratory
Care Unit (ICU) were and still are performed in the central program may lead to long delays between the time a
or stat laboratory and decisions regarding the change in laboratory test is ordered and the time the result is received
therapy or initiation of appropriate therapy were taken only by the clinical staff and may not keep pace with the
after the reports were made available. Considerable amount dynamically evolving clinical needs of the unstable, critically
of time was consumed, depending upon the type of test ill patient.2 The intensive care unit (ICU)/operating room
ordered or the response time of the system involved in (OR) diagnostic and therapeutic processes are rendered
testing.1 These tests often had to be repeated because of inefficient and ineffective if the ICU/OR staff finds that
error in sampling, labeling, or inadequate sampling, causing test results are classified as “pending” and are not available
further delay. The laboratory procedure begins with the at the time they are required for decision making. Even
physician’s order for laboratory studies, using testing when laboratory results arrive at the ICU/OR, they reflect
profiles and prioritization levels (standard, urgent, and stat) the past and not the patient’s present condition.
pre configured by the laboratory administration. The samples
are drawn and transported to the central laboratory for A few of the problems arising out of the delays
involved with traditional laboratory testing are listed in
table 1. These can definitely change the outcome of the
critically sick patients with rapidly changing internal milieu
and large volume shifts, in the operating room setting and
the ICU. Analysis of serum electrolytes is an example,
reports of which, if not available immediately, will be
useless in terms of replacement therapy.

Table - 1 : Problems with traditional laboratory tests.

Increased therapeutic turn around time


Fig. 1 : Traditional steps involving centralized laboratory testing Delayed data availability
Delayed decision making
1. M.D., Asso. Consultant, Increased blood loss
2. D.A., M.D., D.Acp., Senior Consultant,
Department of Anaesthesiology, Increased pre-analytic error
Pain and Perioperative Medicine, Sir Ganga Ram Hospital, More redundant blood tests
New Delhi – 110060. India
Correspond to :
Dr. S. Gupta With the advent of transplant surgery, especially
E-mail : s_grh99@yahoo.com liver transplant, and major cardiac surgery the need for
(Accepted for publication on 23-06-2004) immediate availability of laboratory test results became
GUPTA, BHATTACHARYA : POINT OF CARE TESTING IN ANAESTH. AND ICU 279

overwhelming. Tableside laboratory tests which could be Types of POC testing 7-9
carried out in accordance with the requirement of surgery Initially, the critical care profile included whole
and postoperative intensive care, prompted anaesthesiologists blood ionized calcium (Ca++) and electrolytes with blood
to search for a viable and efficient alternative and Point- gases and hematocrit. Now biosensor based whole blood
of-care technology for the OR and the ICU. measurements in critical care clusters include glucose,10
ionized magnesium,11 lactate,12 urea nitrogen, creatinine,
Point-of-care testing
and CO2 content.13 Manufacturers are designing and
The term point-of-care (POCT) testing, also called introducing unique new in vitro, ex vivo, and in vivo
near patient testing or bedside testing, refers to the systems14-16 - future problem solvers in the Point-of-care
performance of diagnostic tests at or near the patient. These paradigm. POCT has focused on various combinations of
tests are performed outside the confines of clinical the following testing profiles: blood gas analyses, critical
laboratories in health care facilities. Many descriptive titles care electrolytes (sodium, potassium, chloride, glucose,
have been used for POCT, including Point-of-care diagnostics, lactate, ionized calcium, ionized magnesium, blood urea
Alternate site testing, bedside testing, ancillary laboratory nitrogen, and creatinine), hematology variables (hemoglobin
testing, near and patient side testing, and decentralized and and hematocrit), coagulation parameters (activated
distributed testing. In the Point-of-care framework, testing clotting time [ACT], prothrombin time [PT], activated
is performed on whole blood immediately on request with partial thromboplastin time [aPTT]), and myocardial
user friendly devices located within the vicinity of the markers (creatine kinase MB isoenzyme [CK-MB],
patient either directly at the patient’s bedside or in a ‘mini’ myoglobin, Troponin I and T). Additional analytes and
or ‘stat’ laboratory within the ICU/OR setting. testing profile configurations are continuously under
development.
Historical perspective - Evolution of POC testing
Laboratory testing started at the bedside or ward
level in 19th century London hospitals when ward laboratories
were introduced to chemically analyze urine and fecal
specimens. Through the years this scenario has changed
because of advancements in laboratory medicine and testing
profiles. The site of testing kept on going farther and farther
away from the patient, in the form of a central laboratory
catering to the entire hospital. With the development of
reliable methods of measuring blood gases and pH in the
early 1960s, these measurements became an essential guide
for managing critically ill patients. At the beginning, these
tests were limited to only cardiorespiratory laboratories in
Fig. 2 : POC technology in ICU/ OR
conjunction with respiratory functions and cardiac
catheterization. Later, these gained popularity with critical
Types of monitors for POCT
care centers and operation rooms because of faster and
easy availability of results. Point-of-care testing devices can be used in vitro, ex
vivo, or in vivo and include many options. The analyzer
Now technologic advancement characterised by traditionally has been defined by laboratory medicine as an
microchemistry, microcomputerization, miniaturization and in vitro measurement device that requires the permanent
non invasive testing has revolutionized the concept of present removal of blood, fluids, or tissue from the patient. A
day laboratories. Point-of-care testing first became a monitor is defined as a measurement device that does not
necessity when rapid changes during liver transplant protocols require permanent removal of blood, fluid, or tissue from
demanded immediate knowledge of levels of ionized calcium.3 the patient and can be invasive or non invasive. Figure 2
Along with these, the economically driven need to shorten illustrates the POC modalities that are used in the OR and
length of hospital stay, reduce hospital costs, and manage the ICU.
scarce resources has prompted health care providers to
implement POC testing, without compromising patient A blood monitor can be used ex vivo or in vivo. Ex
outcome. Biosensors and biosensor based methods, and vivo technologies use specimens that are drawn from the
whole blood analysis have enabled a shift to immediate patient and analyzed just outside of the body, with some
diagnostic testing at the point-of-care.3-6 of them allowing the return of the specimens to the
280 INDIAN JOURNAL OF ANAESTHESIA, AUGUST 2004

patient (i.e., intra arterial catheter attached to an Table - 2 : Potential advantages and disadvantages.
analyzer17). In vivo technologies provide the opportunity to
measure pH, blood gases, electrolytes, lactate, and glucose Potential advantages Potential disadvantages
levels directly, using invasive sensors. In 1994 Inomata16
Decreased therapeutic turn around time Lack of adequate documentation
demonstrated the use of fibreoptic pulmonary catheters
to measure the SVO2 continuously and correlated this with Rapid data availability Problems of training and competency

the cardiac output (CO) and in the same year Haller18 Increased real time patient management Poor analytic performance
demonstrated the use of an intra arterial fibreoptic
Augmented clinical decision making Sample handling error
sensor for continuous monitoring of PO2, PCO2 and pH.
Subcutaneous monitoring of glucose concentrations is Decreased preanalytic error Postanalytic error (e.g., transcription
error or communication failure)
another example of in vivo biosensors whose clinical
relevance was shown by Fischer19 in 1994. Further, the use Decreased iatrogenic blood loss Increased costs
of gastric intramucosal pH monitoring was studied by
Fewer redundant blood tests Unauthorized testing
Doglio20 as a prognostic index in critically ill patients.
Other examples of non invasive monitors include testing Test clustering Limited test menu
exhaled air, urine and transcutaneous sampling. Some of Integration with performance maps, No critical values notification system
this technology is already available or is under development algorithms, and care paths. and/or documentation
(Figure 3).
Decreased therapeutic Turn Around Time (TTAT)
Therapeutic Turn-Around Time (TTAT) is the time
between ordering the tests to the implementation of the
actions generated by the results. Salem21 and Fletcher,22
stated that there is considerable saving of time by the use
of bedside microchemistry instruments. POC testing can
facilitate patient care management by allowing for the
implementation of treatment decisions right away rather
than having to wait, sometimes for hours, for the results.
By the time the results are available, the condition of the
patient may be completely different from the situation that
prompted the ordering of the particular test.
With application of POC testing, order writing is
Fig. 3 : Types of POC Testing
abbreviated or not even required, specimen transport is
minimized, and processing steps, TTAT is diminished.
Potential advantages and disadvantages of POC testing Manual delivery of specimens and reporting intermediaries
As evaluation of the risk benefit ratio of a particular are not required (Figure 4).
medication needs rigorous randomized placebo controlled
trials, it is very difficult to determine advantages vs
disadvantages of POC testing in such a trial. Inspite of
this, POC testing has been used in numerous studies and
certain advantages and disadvantages of POC testing
have been suggested. Reduction of Turn-Around Time (TAT)
is one the advantages that have been well documented.21
Often POC testing is performed by persons who may
not have the type of training that ensures quality control,
proper documentation of test results and proficiency
testing.
Various potential advantages and disadvantages are Fig. 4 : Flow chart showing decreased steps of POC testing in
listed in Table 2. comparison to traditional laboratory.
GUPTA, BHATTACHARYA : POINT OF CARE TESTING IN ANAESTH. AND ICU 281

Kost13 also observed that POC testing reduces the and delays in analyzing samples (i.e., preanalytic error)
TAT drastically and makes the reports available instantly. that are associated with traditional laboratory testing.26
In one of the studies, Fleisher23 showed that TAT was
comparable between Point-of-care testing and urgent care Personnel
centralized laboratory (UCL) equipped with fully automated The concept of POCT generally does away with
sample delivery system and reporting interface. But in other requirement of specially trained staff to run the equipment.
laboratories that are not fully automated, there was no Although the anaesthesiologists acknowledge the advantages
difference in analytic TAT. But, the time delay was mainly of POC testing, they believe that having to perform these
in transport of the sample. The decrease in the turn-around- tests themselves detracts from their ability to render direct
time (TAT) allows the treating anaesthesiologist to act patient care and that this is an added responsibility that
promptly by augmenting clinical decisions and altering the they sometimes are reluctant to accept and would prefer be
treatment according to real time test results. left to laboratory personnel. On the other hand, laboratory
personnel and respiratory therapists (in the case of arterial
Blood conservation and fewer redundant tests blood gases) may perceive POC testing as an invasion of
In the OR and the ICU, where frequent blood draws their duties and as a threat to their job security.27 Untrained
are required for laboratory analysis, blood conservation is personnel as opposed to trained technicians operating in
an important priority.24,25 Physicians are familiar with the central laboratories, operate POC devices present at bedside.
dreaded call from the laboratory requesting more blood This can give rise to performance failure mainly due to
because the amount sent was insufficient. Moreover, some sample handling error.
of the equipment (analyzers) used in central laboratory has
fixed menus for which minimum fixed sample size is Documentation
required. Present day microsample POC devices require The results from POC devices usually are displayed
smaller blood samples for the same test menu. Usually the on a screen, with temporary printouts facility. The actions
plasma (after centrifugal separation) is analyzed in the generated from such results follow in an efficient way, but
central laboratory in contrast to POC devices that use whole how the results make their way into the permanent chart
blood. With ever decreasing blood sample requirements for is a concern. The results may get lost and never reach the
analysis, microsample POC devices can have a great impact chart. This possibility not only poses a medicolegal issue
on blood conservation.21 but also can affect reimbursement.
In a central clinical laboratory, usually the test menu Data recording and transcription (Post analytic error)
is fixed for that apparatus and it will analyze a blood
Most of the time when reports are in printed format,
sample for all the tests available under that menu whether
the results are not documented in the data management
those were ordered or not. For example, for the haematocrit
system. And there can be post analytic errors during
the entire range of the CBC is processed by the machine.
transcription of data. The reports are usually not filed
Unfortunately, this advantage of using small sample properly and get displaced, making the data unavailable at
sizes does not always equal blood conservation. In practice a later time for reference.
with the use of POC testing for blood gas and electrolyte
analysis, some individuals persist in the practice of collecting Unauthorized testing
sufficient blood for two or three samples “just in case the This happens quite often with POC testing. Many of
cartridge fails or error on part of operator.” the unauthorized tests are performed by physicians, for
personal use or for learning purposes. The best example of
Portability this type testing is with the hand held glucometer.
Portability of the POC analyzers is another advantage.
This technology is especially useful in several situations, Critical Care profiling (CCP)
including ambulance patient transfers, air transport (even A haemodynamically unstable patient loses
outer space), in remote areas, and rural areas, where the consciousness within ten seconds of O2 deprivation and
cost even with the low volume of tests would prohibit a within six minutes permanent brain damage can result due
traditional laboratory setting. to drained creatine phosphate and adenosine triphosphate
(ATP) energy reserves. With the availability of bedside
Reduction in preanalytic error multianalyzers, critical care profile (CCP) should suggest
An advantage of POC testing is that it reduces the whether tissues are receiving adequate oxygen or not, and
risk of errors associated with handling, transport, labeling, why the availability has been compromised. Figure 5
282 INDIAN JOURNAL OF ANAESTHESIA, AUGUST 2004

identifies a proposed CCP to assess sufficient fuel and end stage renal disease patients. Infusion of large volumes
oxygen supply to support vital organs. of fluid can change plasma matrices. Transfusion of large
amounts of citrated blood and blood components leads to
calcium binding and a decrease in ionic calcium in liver
transplantation.
In the ICU setting, hypocalcaemia can occur in
patients of sepsis, renal or cardiac disease, pancreatitis or
after major surgery.38 Since such patients are often transfused
citrated blood components, monitoring of ionized calcium
Fig. 5 : Figure identifies a proposed CCP to becomes mandatory. Measurements done at the bedside will
assess sufficient fuel and oxygen to support vital organs
ensure early intervention.
POC tests can be protocolized for specific clinical Hypoxia
need. These points can be better understood in the following Lactate is the only blood marker whose presence
clinical situations. generally means that the tissues are receiving inadequate
Diabetes mellitus oxygen. Lactate is the end product of the anaerobic
metabolism of glucose and is clinically sensitive to global
We take the example of diabetes mellitus which is and to some extent regional hypoxia. In addition to hypoxia,
the most common entity requiring multiple blood glucose the degree of lactate elevation may reflect any of the
level estimations in a single day. Stress hyperglycaemia following: inadequate lactate metabolism as a result of a
and diabetes mellitus are commonly found among critically severely compromised liver, the concentration of glucose
ill adults.28,29,30 during hypoxia,39 the cause of hypoxia (sepsis, exercise,
Ellison31 reported a rise in blood glucose levels from seizure, acute myocardial infarction, and so forth),40
144 to 774 mgdl-1 in 95 minutes during surgery. Much treatment with lactated Ringer’s solution, or sudden release
milder elevations of blood glucose levels threatens vital of pent up lactate as perfusion is restored. Most of the
tissues if oxygenation is inadequate.32 Maintaining adequate blood gas measuring equipments currently available
but not excessive glucose levels in the critically ill patients incorporate serum lactate estimation facility.
makes blood glucose levels an important candidate for critical
care profile (CCP). Blood gas analysis and haematocrit
Arterial blood gas analysis and haematocrit are the
Malmberg30 reported that glycaemic control among most common tests ordered in ICU and operating room
such individuals may reduce long term morbidity and settings. Immediate results are of prime importance in such
mortality. Frequent venous sampling or arterial blood glucose situations because correction of pH, PO2, and electrolyte
monitoring may add to the growing problem of iatrogenic derangements in an early phase of disease can alter the
blood loss and anaemia among ICU patients, as well as outcome of the patient.41
delay in clinical decision making because of a long turnaround
time.24 One solution to this problem is adoption of bed side Monitoring gas exchange at the lung level can
glucometry, an accepted method for estimating blood glucose help predict adequacy of ventilation and perfusion. PaO2
levels in ambulatory and hospital ward patients.33,34 appears to be the most sensitive parameter of the adequacy
of diffusion of oxygen across the lung.41 Haematocrit or
Electrolyte imbalance haemoglobin concentration provides a measure of oxygen
Critically ill patients and those undergoing major transport capability.
surgical procedures have a tendency towards abnormality So a bedside monitor providing the above tests will
in the concentration of potassium and calcium, pre-existing be able to produce stat results with a small blood sample
or iatrogenic e.g. decrease in potassium level in a and corrective steps can be taken immediately.
hyperventilating patient, where a normal value for potassium
may be low for that pH. Coagulation disorder
In kidney transplantation and liver transplantation, The number of surgical patients receiving
there can be major shifts of potassium and calcium levels anticoagulant therapy is on the rise. Patients undergoing
during the course of surgery. Electrolyte imbalance may major surgery receive large volumes of fluid and blood
also be present preoperatively,35-37 e.g. hyperkalaemia in leading to problems of coagulation. The use of nonspecific
GUPTA, BHATTACHARYA : POINT OF CARE TESTING IN ANAESTH. AND ICU 283

haemostasis monitoring may not clarify whether or not Future developments in poct analyzers and monitors
patients should receive blood versus blood products. Cardiac Point-of-care testing (POCT) is a major force in the
anaesthesiologists were quick to recognize the problem and future evolution of hospital care, with prospects for even
instituted point-of-care testing facilities in the OR for greater expansion of accessibility, speed, and also, hopefully,
haemostasis monitoring.42 Success of broad intra-operative accuracy of results. POCT test menus will continue to expand,
coagulation assessment requires techniques combining
with more coagulation testing, chemistries, and infectious
accuracy and consistency of measurement comparable to
screening, but also on site drug screening, intra-operative
laboratory systems with device characteristics that maintain
hormone levels, and microchip DNA diagnostics. Non
feasibility for bedside use.
invasive POCT will expand beyond the GlucoWatch glucose
Traditional methods of monitoring coagulation monitor and the Bilichek noninvasive bilirubin monitor to
like the bleeding time, activated clotting time (ACT), noninvasive CBCs and Pap smears.44
prothrombin time (PT), activated partial thromboplastin
time (aPTT), platelet count, fibrinogen, and fibrinogen Technology improvements
degradation products, have no or minimal predictive power POCT analyzers already have been miniaturized from
for perioperative bleeding. And, moreover, these tests are large, bulky, bench top units to very small lightweight,
time consuming and require expertise in obtaining samples disposable, hand held devices. Microchip, biochip, and
to get accurate results. nanochip technology45,46 may allow bedside testing for
Thromboelastography (TEG), a POC device, does analytes, microorganisms, or genetic analysis. Techniques
appear to accurately predict peroperative/postoperative are being developed for the insertion and monitoring of
bleeding and, therefore, may have value in targeting microchips within human tissues and organs.
coagulation treatment.42 Currently there are two viscoelastic
Advances in compact disc (CD-ROM) technology
coagulation monitoring devices, TEG and Sonoclot (SCT).43
also are progressing to “labs on a CD”. The CDs could be
These monitoring devices provide information regarding
read at the ICU bedside using standard or customized optical
the entire clotting system including platelet dysfunction
readers.46
and fibrinolysis.
The promptness of viscoelastic results are because Modular design
of use of whole blood instead of separation of cellular and During the last decade, there has been increasing
acellular components as is the case with traditional stress on modular design of monitors so as to incorporate
coagulation tests. The rapid growth in TEG and SCT stand alone bed side devices into single platform. Devices
investigations provides evidence that these devices are filling like pulse oximeters, non invasive blood pressure,
a unique need for POC testing. capnograph, and cardiac output machines have been converted
Figure 6 illustrates a proposed critical care profile into modules that are inserted into the modular racks of the
in a case of end stage renal disease (ESRD) for kidney bedside physiologic monitor. This modular conversion
transplantation. conserves space at the bedside and consolidates all devices
into one networked platform that provides data display and
management, and facilitates biomedical maintenance and
general product support.48
A generic POCT modular rack with a universal
interface that connects to all physiologic monitoring systems/
networks could be constructed. Such a rack would house a
variety of POCT modules and the user could select the
modules most needed to address specific clinical situations.
POCT modular analyzers could function with single or
multiuse cartridges which would be disposable. Finally,
stand alone POCT devices may interface with the monitoring
system through a generic POCT data recipient module or
Fig. 6 : Typical critical care profile (CCP) for patient undergoing kidney
transplantation (End-stage renal disease ESRD, kidney transplantation KTP, directly through a port to the bedside monitor or to the
Random blood sugar RBS, haematocrit Hct.) monitoring system network.
284 INDIAN JOURNAL OF ANAESTHESIA, AUGUST 2004

Sample procurement invasive Connectivity


Research programs in diabetes management are POCT connectivity developments have focused on
developing innovative methods of procuring and measuring three levels of interface and information transfer, starting
samples for glucose analysis that may become applicable with interfacing POCT devices with the users, connectivity
to critical care POCT. The first technique approaches the expands into consolidating all devices on to one platform
intermittent or continuous acquisition of capillary bed or and finally networking it to a DMS, L/HIS.
interstitial samples with laser microporation. Reverse
Summary
iontophoresis, a second sampling technology, relies on the
application of a minielectric current to the intact skin to Interest in the paradigm of Point-of-care testing has
cause glucose from the interstitial fluid to raise though the increased in recent years. By providing faster turnaround
epidermis. A third approach being studied is the insertion time, the clinician can rapidly respond to changes in the patient’s
of an indwelling subcutaneous glucose sensor. status, theoretically resulting in optimized treatment and shorter,
less expensive hospitalizations.52 Analysis in point-of-care
Analyte and vasculature assessment - non invasive testing is no longer performed exclusively by skilled medical
Pulse oximetry is the standard at every bedside ICU technologists but also by multiskilled personnel including
for continuous determination of oxygen saturation. Other anaesthesiologists, nursing staff, respiratory therapist,
analytes possibly could be assessed through “imetry.” emergency medical staff, physicians, and other medical staff.27,53
Bilirubin is currently evaluated by neonatal spectroscopy. With rapid advances in technology, POC testing is
In addition to non invasive spectroscopic assessment of a dynamic work in progress; the analyzers are becoming
analytes, the microcirculation may be non invasively smaller, faster, and more user friendly and have been
evaluated using orthogonal polarization spectral imaging.48 demonstrated to achieve accuracy with increasingly smaller
blood samples, aiding in blood conservation.
POCT informatics : Data management and
connectivity POCT has focused on various combinations of the
Connectivity for POCT devices has evolved from following testing profiles: from blood gas analyses, critical
care electrolytes to hematology variables, coagulation
point-of-service workstations to standardized POCT data
parameters, and myocardial markers. The anaesthesiologist
transmission protocols to remote roaming wireless connectivity
and every critical care therapist must decide which bedside
with automatic data capture. Patient results must be available
test or cluster of tests are indicated for their given patient
to health care providers using the laboratory/hospital
population. To determine this, other factors must be
information system (L/HIS) and transmitted to the electronic
considered, including advantages and disadvantages, analysis
medical record. In the central laboratory the interface and
of test accuracy, clinical impact, and cost benefit ratio. A
management of data have been nearly perfected. Ideally, recent editorial by Jahn,54 put forward recommendations for
laboratory devices wherever they are based or used, the implementation of near patient testing. He suggested a
including POCT devices, should be managed and interfaced strong multidisciplinary and interdepartmental team approach
in a manner similar to those of the central laboratory.49,50,51 is necessary to ensure cost effectiveness, better analytic
Data management performance, reliability, comparability, and quality control.
The dispersion of POCT devices throughout the The transfer of blood gas and other blood
hospital and critical care environment creates unique measurements germane to the critically ill from the
challenges for data management. These problems stem laboratory to the ICU/OR should have profound effect on
from the use of POCT devices of various designs and critical care as did the introduction of laboratory based
technologies in multiple sites by a host of personnel with blood gas analyzers more than 40 years ago.
minimal training and perspective in standard laboratory
Acknowledgements
practice.49 To address these issues Data Management
We gratefully acknowledge Dr. Chand Sahai,
System (DMS) would require “smart” software programmed
associate consultant of our department, for providing help
with artificial intelligence and algorithms that are designed
in preparation of this manuscript.
to identify and expediently communicate device and
data deviations to the local device user, the local DMS References
manager, or preferentially to centrally based laboratory. 1. Fiallos MR. Point-of-care testing. Pediatr Clin North Am
Such communications would require that the next 2001; 48: 589-599.
generation of POCT data management systems include 2. Kost GJ, Ihrmeyer SS, Chernow B et al. The laboratory-
real time bidirectional connectivity and broad based, robust clinical interface: Point-of-care testing. Chest 1999; 115:
communication support. 1140-1154.
GUPTA, BHATTACHARYA : POINT OF CARE TESTING IN ANAESTH. AND ICU 285

3. Kost GJ, Jammal MA, Safwat AM. Monitoring of ionized 19. Fischer U, Rebrin K, Woedtke TV et al. Clinical usefulness of
calcium during human hepatic transplantation: Critical values the glucose concentration in the subcutaneous tissue:
and their relevance to cardiac and hemodynamic management. Properties and pitfalls of electrochemical biosensors. Horm
Am J Clin Pathol 1986; 86: 61-70. Metab Res 1994; 26: 515-522.
4. Kost GJ, Shirey TL. New whole blood testing for laboratory 20. Doglio GR, Pusajo JF, Egurrola MA et al. Gastric mucosal
support of critical care at cardiac transplant centers and US pH as a prognostic index of motility in critically ill patients.
hospitals. Arch Pathol Lab Med 1990; 114: 865-868. Crit Care Med 1991; 19: 1037-1040.
5. Kost GJ. New whole blood analyzers and their impact on 21. Salem M, Chernow B, Burke R et al. The bedside diagnostic
cardiac and critical care. Crit Rev Clin Lab Sci 1993; 30: blood testing: Its accuracy, rapidity, and utility in blood
153-202. conservation. JAMA 1991; 266: 382-389.
6. Kost GJ. Point-of-care testing in intensive care. In: Tobin M, 22. Fletcher AB. The essential role of the laboratory in the optimal
ed. Principles and Practice of Intensive Care Monitoring. care of the sick neonate. J Int Fed Clin Chem 1990; 2: 166-172.
New York; McGraw-Hill 1998; 1267-1296. 23. Fleisher M, Schwartz MK. Automated Approach to Rapid
7. Kost GJ. Planning and implementing pointof-care testing Response Testing – A comparative Evaluation of Point-of-
systems In: Tobin M, ed. Principles and Practice of Intensive Care and Centralized Laboratory Testing. Am J Clin Pathol
Care Monitoring. New York; McGraw-Hill 1998; 1297-1328. 1995; 104: 18-25.
8. Kost GJ, Hague C. In vitro, ex vivo, and in vivo biosensor 24. Zimmerman JE, Seveff MG, Sun X et al. Evaluating laboratory
systems. In: Kost GJ, ed. Handbook of clinical automation, usage in the intensive care unit: Patient and institutional
robotics, and optimization. New York; John Wiley and Sons characteristics that influence frequency of blood sampling.
1996; 648-753. Crit Care Med 1997; 25: 737-748.
9. Zaloga GP. Evaluation of bedside testing options for the 25. Chernow B, Jackson E, Miller J. Blood conservation in acute
critical care unit. Chest 1990; 97: 185S-190S. care and critical care. AACN Clin Issues 1996; 7: 191-197.
10. Kost GJ, Wiese DA, Bowen TP. New whole blood methods 26. Steffes MW, Gillen JL, Fuhnnan SA. Delivering clinical
and instruments: Glucose measurement and test menus for laboratory services to intensive care units. Clin Chem 1996;
critical care. J Int Fed Clin Chem 1991; 3: 160-172. 42: 387-391.
11. Altura BT, Dell Orfano K, Yeb Q et al. A new ion selective 27. Lamb LS Jr, Parrish RS, Goran SF et al. Current nursing
electrode (ISE) for ionized magnesium (Mg++) in whole blood, practice of point-of-care laboratory diagnostic testing in critical
plasma and serum. Clin Chem 1991; 37: 948-949. care units. Am J Crit Care 1995; 4: 429-434.
12. Aduen J, Bernstein WK, Khastgir T et al. The use and clinical 28. Young B, Ott L, Dempsey R et al. Relationship between
importance of a substrate specific electrode for rapid admission hyperglycemia and neurologic outcome of severly
determination of blood lactate concentrations. JAMA 1994; brain-injured patients. Ann Surg 1989; 210: 466-472.
272: 1678-1685. 29. Capes SE, Hunt D, Malmbeg K et al. Stress hyperglycemia
13. Kost GJ, Trent JKT, Saaed D. Indications for measurement and increased risk of death after myocardial infarction in
of total carbon dioxide in arterial blood. Clin Chem 1988; 34: patients with and without diabetes mellitus: a systemic
1650-1652. overview. Lancet 2000; 355: 773-778.
14. Arnold MA. New developments and clinical impact of 30. Malmberg K. Prospective randomized study of intensive
noninvasive monitoring. In: Kost GJ, ed. Handbook of insulin treatment in long term servival after acute myocardial
Automation, Robotics, and Optimization. New York: John infarction in patients with diabetes mellitus. BMJ 1997; 314:
Wiley and Sons, 1995. 1512-1515.
15. Kost GJ, Hague C. In: Kost GJ, ed. Handbook of Automation, 31. Ellison DA, Forman DT. Transient hyperglycemia during
Robotics, and Optimization. New York: John Wiley and Sons abdominal aortic surgery. Clin Chem 1990; 36: 815-817.
1995. 32. Inamura K, Smith ML, Olsson Y, Siesjoe BK. Pathogenesis of
16. Inomata S, Nishikawa T, Taguchi M. Continuous monitoring substantia nigra lesion following hyperglycemic ischemia:
of mixed venous oxygen saturation for detecting alterations in Changes in energy metabolites, cerebral blood flow, and
cardiac output after discontinuation of cardiopulmonary by morphology of pars reticulata in a rat model of ischemia. J
pass. Br J Anaesth 1994; 72: 11-16. Cereb Blood Flow Metab 1988; 8: 375-384.
17. Widness JA, Kulhavy JC, Johnson KJ. Clinical performance of 33. Harding K. A comparison of four glucose monitors in a hospital
an In-Line Point-of-Care monitor in neonates. Pediatrics 2000; medical surgical setting. Clin Nurse Spec 1993; 7: 13-16.
106: 497-504. 34. Ray JG, Hamielec C, Mastracci T. Pilot study of the accuracy
18. Haller M, Kilger E, Briegel J et al. Continuous intra-arterial of bedside glucometry in the intensive care unit. Crit Care
blood gas and pH monitoring in critically ill patients with Med 2001; 29: 24-28
severe respiratory failure. A prospective criterion standard 35. Prough DS. Physiologic acid-base and electrolytes changes in
study. Crit Care Med 1994; 22: 580-587. acute and chronic renal failure patients. Anesthesiol Clin North
Am 2000; 18: 809-833.
286 INDIAN JOURNAL OF ANAESTHESIA, AUGUST 2004

36. Sladen RN. Anesthetic considerations for the patient with 46. Wilding P, Kricka LJ. Micro-microchips. Just how small can
renal failure. Anesthesiol Clin North Am 2000; 18: 863-882. we go? Trends Biotechnol 1999; 17: 465-468.
37. Joseph AJ. Perioperative care of the patient with renal failure. 47. Halpern NA. ICU bedside technology: Point-of-Care
Med Clin North Am 2003; 87: 193-210. diagnostics and networks. Crit Care Clin 2000; 16: 623-639.
38. Lepage R, Legare G, Racicot C et al. Hypocalcemia induced 48. Groner W, Winkelman JW, Harris AG et al. Orthogonal
during major and minor abdominal surgery in humans. J Clin polarization spectral imaging: A new method for study of the
Endocrinol Metab 1999; 84: 2654-2658. microcirculation. Nat Med 1999; 5: 1209-1213.
39. Siesjo BK. Mechanism of ischemic brain damage. Crit Care 49. Halpern NA, Brentjens T. Point-of-care testing informatics:
Med 1988; 16: 954-963. The critical care hospital interface. Crit Care Clin 1999; 15:
40. Schuster HP. Prognostic value of blood lactate in critically ill 577-591.
patients. Resuscitation 1984; 11: 141-146. 50. Kane B. Point-of-care testing: Instant gratification? Ann Intern
41. Shirey TL. Critical care profiling for informed treatment of Med 1999; 130: 870-872.
severely ill patients. Am J Clin Pathol 1995; 104: S79-S87. 51. Parker JD. LIS vendor, POCT manufacturer, laboratorian:
42. Shore Lesserson L, Manspeizer HE, De Perio M et al. Who’s responsible for POCT data management? Med Lab
Thromboelastography-guided transfusion algorithm reduces Obser 1995; 27: 8-11.
transfusions in complex cardiac surgery. Anesth Analg 1999; 52. Burke MD. Turnaround time, point-of-care, and a future role
88: 312-319. for the pathologist. Am J Clin Pathol 1993; 100: 89-90.
43. Hett DA, Walker D, Pilkington SN et al. Sonoclot analysis. 53. Herr DM, Newton NC, Santrach PJ et al. Airborne and rescue
Br J Anaesth 1995 ; 75: 771-776. point-of-care testing. Am J Clin Pathol 1995; 104: S54-S58.
44. Bissell M. Point-of-care testing at the millennium. Crit Care 54. Jahn UR, Van Aken H. Near-patient testing – Point-of-care
Nurs Q 2001; 24: 39-43. or points of costs and convenience? Br J Anaesth 2003;
45. Kricka LJ. Revolution on a square centimeter. Nat Biotechnol 90(4): 425-427.
1998; 16: 513-514.

SPECIAL ATTENTION!
Dear Members,
Are you ae one of those who are not getting our esteemed Journal regularly?
Any of the following are possible :
• May be our Journal is going to your old address? Please check
• You may have some dues to be paid towards ISA? Please verify
Kindly take the trouble of writing to the ISA Secretariat (Dr. J. Rangananthan Secretary, ISA (National,
IMA Building Campus, No. 12, Saradha College Road, Near Five Roads, Salem - 636004, T.N.) to ascertain
your membership status and inclusion in the mailing list of IJA.
- Editor

ATTENTION! ISA MEMBERS

Kindly complete the following form, and send it to ISA Secretariat (Dr. J. Rangananthan Secretary, ISA
(National), IMA Building Campus, No. 12, Saradha College Road, Near Five Roads, Salem - 636 004 T.N.)
The ‘Life Members List’ (Journal Mailing List) is being updated. Your co-operation is highly appreciated.
Name :
Qualification and Year of Passing :
ISA No. : Year of Membership :
Address :
State : Pin :

Vous aimerez peut-être aussi