Vous êtes sur la page 1sur 11

ARTICLE IN PRESS

Water Research 39 (2005) 3098–3108


www.elsevier.com/locate/watres

Comparison of electrocoagulation and chemical


coagulation pretreatment for enhanced virus removal using
microfiltration membranes
Bintuan Zhua, Dennis A. Clifforda,, Shankararaman Chellama,b
a
Department of Civil and Environmental Engineering, University of Houston, 4800 Calhoun Road, Houston, TX 77204-4003, USA
b
Department of Chemical Engineering, University of Houston, Houston, TX 77204-4004, USA
Received 20 December 2004; received in revised form 17 May 2005; accepted 18 May 2005

Abstract

This research studied virus removal by iron electrocoagulation (EC) followed by microfiltration (MF) in water
treatment using the MS2 bacteriophage as a tracer virus. In the absence of EC, MF alone achieved less than a 0.5-log
removal of MS2 virus, but, as the iron-coagulant dosage increased, the log virus removal increased dramatically. More
than 4-log virus removal, as required by the Surface Water Treatment Rule, was achieved with 6–9 mg/L Fe3+. The
experimental data indicated that at lower iron dosages and pH (o8 mg Fe/L and pH 6.3 and 7.3) negatively charged
MS2 viruses first adsorbed onto the positively charged iron hydroxide floc particles before being removed by MF. At
higher iron dosages and pH (49 mg Fe/L and pH 8.3), virus removal was attributed predominantly to enmeshment
and subsequent removal by MF. Additionally, the experimental data showed no obvious influence of ionic strength in
the natural water range of 107–102 M on MS2 virus removal by EC-MF. Finally, EC pretreatment significantly
outperformed chemical coagulation pretreatment for virus removal. The proposed mechanism for this improved
performance by EC is that locally higher iron and virus concentrations and locally lower pH near the anode improved
MS2 enmeshment by iron flocs as well as adsorption of MS2 viruses onto the iron floc particles.
r 2005 Elsevier Ltd. All rights reserved.

Keywords: Microfiltration; Coagulation; Electrocoagulation; Virus; Bacteriophage; Water treatment; Membrane filtration

1. Introduction Several published studies have reported partial (0.2- to


3-log) virus removal by MF (Coffey et al., 1993;
Microfiltration (MF) has been used to completely or Madaeni et al., 1995; Urase et al., 1996; Roberts,
significantly remove turbidity, bacteria, and protozoa 1997), which is significantly less than the 4-log virus
from water and wastewater (Jacangelo et al., 1991; removal mandated by the Surface Water Treatment
Mallevialle et al., 1996; Madaeni, 1999). However, MF Rule (SWTR). Therefore, MF by itself cannot meet the
alone is not an efficient barrier for virus removal, SWTR virus removal requirement even though the
because viruses are typically smaller than its pores. turbidity and protozoa regulations can be met easily.
The addition of a coagulant prior to membrane
Corresponding author. Tel.: +713 743 4266; filtration has been suggested to generally improve
fax: +713 743 4260. product water quality and reduce membrane fouling
E-mail address: daclifford@uh.edu (D.A. Clifford). by coagulating the dissolved organic matter ahead of the

0043-1354/$ - see front matter r 2005 Elsevier Ltd. All rights reserved.
doi:10.1016/j.watres.2005.05.020
ARTICLE IN PRESS
B. Zhu et al. / Water Research 39 (2005) 3098–3108 3099

filter (Lahoussine-Turcaud et al., 1990; Wiesner et al., 2.3. Test water


1992). Usually, chemical coagulation (CC) with iron or
aluminum salt is used, however electrocoagulation (EC) Three synthetic fresh waters, roughly characterized as
is another possibility. To our knowledge, this is the first low (I ¼ 107 M), medium (I ¼ 6  103 M), and high
report of EC pretreatment for MF. ionic strength (I ¼ 1:8  102 M), were used in this
EC has been widely studied in water and wastewater research. These were prepared from deionized (DI)
treatment to remove heavy metals, organics, bacteria, water and reagent-grade chemicals to simulate fresh
hardness, turbidity, and other contaminants (Horner water containing monovalent and divalent cations, and
and Duffey, 1983; Tsouris et al., 2001; Can et al., 2003; alkalinity. The low-ionic-strength water was DI water,
Mills, 2000). In the EC process, the electrodes are and the high-ionic-strength water was DI water spiked
consumed as the coagulant is generated and precipi- with 3.0 mM NaHCO3, and 10 mM CaCl2. The medium-
tated; no liquid chemical is added; alkalinity is not ionic-strength water used in most of the experiments
consumed; and pH adjustment is not needed. Addition- consisted of DI water spiked with 3.0 mM NaHCO3, and
ally, compared with CC, the EC process reportedly 1.0 mM CaCl2. Its pH was 8.3 without any adjustments.
requires less coagulant and produces less sludge (Horner
and Duffey, 1983; Mills, 2000). According to one 2.4. Membrane microfilter
estimate, the space required for EC is less than CC
because EC does not require chemical storage, dilution, All filtration experiments were carried out using a
and rapid mixing (Mills, 2000). Because EC systems hydrophilic, 0.22-mm pore size, modified polyvinylidene
typically use solid iron or aluminum anodes rather than fluoride (PVDF) membrane filter (Durapore, Millipore
corrosive iron or aluminum salt solutions, EC units can Corp. Bedford, MA). A fresh membrane with an
be more easily incorporated into ‘‘packaged’’ plants and effective filtration area of 4.1 cm2 was used for each
transportable water treatment plants for use in remote experiment. The manufacturer confirmed that its surface
areas or in emergency water supply treatment, which was negatively charged in the pH range of our
was one of the driving forces to undertake this research. experiments, 6.3–8.3.
The objectives of this study were to (a) evaluate the
efficacy of iron EC-MF for virus removal, (b) investigate 2.5. EC unit
the effect of EC-MF operating parameters on virus
removal, (c) compare virus removal by EC-MF and CC- A bench-scale EC unit with three anode–cathode pairs
MF, and (d) suggest possible mechanisms for virus was designed and built for this research (Fig. 1). It
removal by EC-MF. consisted of a 200-mL active-volume, flow-through
electrode chamber with 22-cm long rod-shaped iron
anodes and porous cylindrical stainless steel cathodes.
The total anode surface area was 100 cm2 and the
2. Materials and methods current density was typically 0.25 mA/cm2. By adjusting
the operating current and flow rate of source water, the
2.1. Virus desired iron concentration was obtained.
As can be seen in Fig. 1, the recycle pump discharged
The indicator-organism virus used throughout the into the annular area between the cathode and anode of
study was the bacterial virus, MS2. It is 0.025 mm in size, each of the three cathode-anode pairs in order to flush
icosahedral in shape, and contains a single strand of the anode where iron coagulant was continuously
ribonucleic acid with 3569 nucleotides (Valegård et al.,
1990). The virus stock was bought from American Type
Culture Collection (ATCC # 15597-B1). Because bac-
Source Water
terial viruses are also called bacteriophages (Brock and
PlanView of Anode and Cathode
Madigan, 1991), we use the terms ‘‘virus’’ and ‘‘bacter-
iophage’’ interchangeably to refer to MS2. Internal Diameter: 0.85 cm

EC Unit
Stainless Steel Cathode
2.2. MS2 virus propagation and MS2 virus assay
Iron Anode Diameter: 0.48 cm
The procedure of MS2 virus propagation and MS2
virus assay was described in our previous paper (Zhu et
al., 2004). MS2 samples were assayed using a modified
Recycle Pump
agar-overlay technique, which counts viable viruses
using Escherichia coli as the host bacterium. Fig. 1. Diagram of batch bench-scale EC system.
ARTICLE IN PRESS
3100 B. Zhu et al. / Water Research 39 (2005) 3098–3108

generated. During the continuous flow experiments (see spiked with MS2 virus was fed into the top of empty EC
Section 2.8), raw water entered the unit through an inlet unit and when the level reached the overflow port, the
tube that terminated near the center of the unit at about water was recycled at about 500 mL/min. Prior to
one half the water depth. reaching steady state, about 2 L of effluent water was
wasted. After reaching steady state as predetermined by
2.6. Calculation of virus removal the volume required reach a constant iron concentration
in the effluent, the effluent water was directed to the feed
As is commonly practiced, virus removal in these bottle of the MF system. Viruses in the EC unit feed
experiments is expressed as a log reduction value (LRV) tank and in the MF effluent were counted for the
calculated using the permeate (CP) and feed concentra- determination of log removal using Eq. (1).
tions (CF) of the MS2 virus (EPA, 2001):
2.9. Bench-scale CC-MF system
LRV ¼ log removal value ¼ logðC F =C P Þ. (1)
Because the target virus concentration in the feed In order to compare EC-MF with chemical coagula-
water was 106 pfu/mL, the maximum achievable LRV in tion (CC)-MF, CC-MF experiments were carried out
our experiments was 6.0. (Experimental LRVs were using the procedure described in a previous paper (Zhu
lower than 6.0 in all cases.) et al., 2004). Source water spiked with MS2 virus was
coagulated using FeCl3 (1 min rapid mixing at 100 rpm),
2.7. Fe (III) measurement flocculated (30-min slow mixing at 30 rpm), and then
filtered in the dead-end mode using the PVDF flat-sheet
The concentrations of Fe3+ generated by the EC unit membrane.
were analyzed in triplicate using flame atomic absorp-
tion spectroscopy (Flame AA-AAnalyst 300, Perkin- 2.10. MS2 viability during the course of the experiments
Elmer Corporation, CT, USA) according to Method
3111 in Standard Methods (Clesceri et al., 1998). If virus inactivation occurred during EC pretreatment,
it would result in an overestimation of intrinsic log virus
2.8. Bench-scale EC-MF system removal by combined EC and membrane filtration. To
determine if virus inactivation occurred as a result of
The EC unit treated source water containing viruses EC, pressure change, or mixing during MF, two
before it was microfiltered. Source water entered from separate control experiments were conducted. The first
the top of the EC unit. Once inside the unit, it was involved EC only and the second involved pressure and
recycled at about 500 mL/min to ensure that the mixing during filtration. In the first experiment, viruses
electrodes remained free of surface deposits and that were electrocoagulated, and then samples were taken
the Fe3+ generated was quickly released from the from the EC unit for viral counting without performing
surface and hydrolyzed. Following the practice of other any filtration.
EC researchers (Tsouris et al., 2001; Han et al., 2002; In the second experiment, which determined the effect
Holt et al., 2002), all virus-removal data from the EC of pressure and mixing on virus viability, the initial
unit was collected in a batch mode, i.e., the EC unit was sample (Sample 1) for virus enumeration was taken from
filled with virus-spiked water, and turned on for 40 s to the source water before EC; the final sample (Sample 2)
generate the desired target amount of iron. After EC, the was taken from the last 20 mL of coagulated feed water
solution was directly filtered by transferring it to the feed present in the pressurized (69 kPa, 10 p) feed bottle. This
bottle where it was gently and continuously mixed sample (Sample 2) had been electrocoagulated, pressur-
during MF. The bench-scale membrane system was ized and mixed for about 1 h (no filtration was
operated in the dead-end filtration mode at a constant performed). The difference in viral counts between
transmembrane pressure of 69 kPa (10 psig) maintained Samples 1 and 2 was considered to be a measure of
by compressed air. Finally, samples for analysis of the effects of pressure and mixing on viability.
viruses were taken before the EC unit, after the EC unit,
and after the membrane filter. 2.11. Bench-scale CC-MF system using the EC unit as a
Although batch EC experiments were carried out hydraulic mixer only
under non-steady-state conditions as was the practice of
other researchers, a limited number of continuous flow, During the EC experiments, water in the EC unit was
steady-state EC-MF experiments for virus removal were re-circulated at a high rate (500 mL/min) to keep it well-
also carried out at pH 7.3 with varying iron dosages mixed. Such hydraulic mixing differed from the mechan-
using the same source water as used in the batch ical mixing provided by a rectangular paddle mixer
experiments. The procedure used for the continuous during chemical coagulation jar tests, and could possibly
experiments was as follows: 300 mL/min of source water have changed virus removal by MF. Therefore, to insure
ARTICLE IN PRESS
B. Zhu et al. / Water Research 39 (2005) 3098–3108 3101

that any differences between EC and CC could be (55.85 g mol1), and F is Faraday’s constant (96,486 C
attributed solely to the electrochemical generation of eq1).
Fe3+, chemical coagulation was conducted in the EC When operated at steady state, which was reached
unit with the power to the electrodes turned off and by after approximately 8 EC unit volumes (1600 mL) of
using recirculation for 40s rapid mixing during chemical throughput, the Fe3+ concentration (mg/L) in the EC
coagulation. effluent water was a function of the current and the
The source water in these CC control experiments detention time in the EC unit. By adjusting the operating
using the EC unit for mixing had the same composition current and flow rate of source water to be treated, a
as that used in the EC-MF experiments. Four coagulant desired iron concentration was obtained. For example,
dosages—0, 2, 5, and 10 mg Fe/L—were targeted. The when the feed flow rate was 300 mL/min and the
procedure for chemical coagulation using the EC unit operating current was 0.25 A, the steady-state iron
solely to provide hydraulic mixing was as follows: (a) concentration was 9.6 mg/L. Based on the observed
400 mL of source water containing MS2 viruses was steady state Fe(III) concentration and operating current,
added from the top of the EC unit using a glass beaker; the EC unit operating curve shown in Fig. 2 was
(b) simultaneously, a predetermined dose of FeCl3 stock obtained. The calculated current efficiency was 93%
solution (2000 mg-Fe/L) was quickly added to the EC based on the slope of the linear best-fit curve with zero
unit from the top using a syringe; and (c) the recycle intercept.
pump (500 mL/min) was turned on while the EC power In all cases, no iron was detected (method detection
supply remained off. The detention time of the source limit 0.03 mg Fe/L) in the microfiltered water indicating
recirculating water in the un-powered EC unit was that the EC unit predominantly generated particulate
maintained at 40 s, which was the same as all other EC- iron.
MF experiments. After 40 s of recycle mixing in the un-
powered EC unit, the chemically coagulated solution
3.2. Possibility of MS2 virus removal or inactivation
was immediately transferred to the feed bottle where it
within the EC unit
was gently and continuously mixed during filtration.
The operation of the bench-scale membrane system
Our previous research demonstrated that chemical
following chemical coagulation was the same as that
coagulation with ferric chloride did not inactivate MS2
described in the bench-scale EC-MF system. Finally,
viruses (Zhu et al., 2004). However, because EC uses
samples for virus analysis were taken before coagulation
voltage and current to generate iron, chlorine produc-
and after MF.
tion is also possible, which would inactivate viruses (Liu
et al., 1997; Drees et al., 2003).
Experiments were carried out to investigate the
3. Results and discussion possibility of inadvertent virus inactivation by chemical
disinfection or virus removal by accumulation on the
3.1. Operation of EC unit anode. Virus samples were taken from the spiked raw
water (Sample 1) and the electrocoagulated water
After designing and fabricating the EC unit, it was immediately after operating the EC unit (Sample 2). If
tested for its ability to generate iron in proportion to the
current by operating it continuously at different current
values, while source water was pumped into the top at a
12
constant flow rate of 300 mL/min. Water was recycled in
Fe(III) Concentration(mg/L)

the unit at 500 mL/min and exited through the overflow 10


port at 300 mL/min. Samples for analysis of Fe(OH)3(s)
measured as Fe, generated by EC unit, were taken from 8
the effluent of the EC unit at each different operating-
current condition. 6
To determine the current efficiency, the amount of
iron generated was calculated using Faraday’s Law (Eq. 4
(2)): Experimental Data
2 Theoretical Value
I  t  MW
m¼ , (2)
ZF 0
0 0.05 0.1 0.15 0.2 0.25 0.3
where m is the mass in grams of Fe generated at a
EC Operating Current (A)
specific current (I, amperes) over a time interval
(t, seconds), Z is the number of electrons transferred Fig. 2. Fe(III) generated at 300 mL/min flow rate as a function
per Fe atom, MW is the molecular weight of EC operating current.
ARTICLE IN PRESS
3102 B. Zhu et al. / Water Research 39 (2005) 3098–3108

chlorine was generated or viruses were accumulated on 2.0


the EC anode, a reduction in the number of viable
viruses would be observed. The results of triplicate

Normalized MS2 Virus


Concentration (N/N0)
experiments conducted at different iron doses and pH 1.5
values are shown in Fig. 3. The virus concentrations of
Samples 1 and 2 were compared using the t-test based on
1.0
n2 ¼ 4 degrees of freedom and a 0.05 level of
significance. All test statistics were less than the critical
value (t0.05(4) ¼ 2.776). Hence, the virus concentrations 0.5
before and after EC treatment were essentially the same pH = 6.3
pH = 7.3
demonstrating that viruses did not accumulate on the pH = 8.3
anode nor were viruses rendered less viable by EC. 0.0
0 2 4 6 8 10 12 14
Fe(III) Concentration (mg/L)
3.3. Effect of mixing- and pressurization on MS2 virus
viability during filtration Fig. 4. Determination of mixing and pressurization on MS2
viability during filtration. Error bars represent one standard
The results of additional control experiments con- deviation.
ducted to establish that virus concentrations in the
electrocoagulated mixture in the pressurized feed bottle
did not change significantly during MF are shown in
Fig. 4. Sample 1 (N 0 ) was taken before the EC unit, and
Sample 2 (N) was taken from the feed bottle near the 3.4. Effect of Fe (III) concentration and cake thickness
end of filtration. The virus concentrations of Sample 1 on MS2 virus removal by EC-MF
(average of 3 replicates) and Sample 2 (average of 3
replicates) were also compared using the t-test. The results of EC-MF experiments conducted to
Samples 1 and 2 were not significantly different based determine the effects of iron dosage and pH on virus
on the t-test statistics, all of which were less than critical removal in medium ionic-strength background water
value of t0.05(4) ¼ 2.776. Therefore, operating pressure with 106 pfu/mL virus concentration are shown Fig. 5.
(69 kPa, 10 psig) or stirring time (1 h) had no effect on In the absence of EC, MF alone achieved less than 0.5-
virus viability demonstrating that any virus removal log virus removal. With increasing iron dosage during
observed during EC-MF experiments was due to the EC pretreatment, virus removal was dramatically
combination of EC and filtration and not inactivation improved. The 4-log virus removal required by the
during extended mixing at 69 kPa (10 psig). These results SWTR (EPA, 2001) was achieved at Fe(III) concentra-
are consistent with those in our previous CC-MF study tions in the range of 6–9 mg/L as pH increased from 6.3
(Zhu et al., 2004). to 8.3.
The membrane alone (220 nm nominal pore size) was
not expected to substantially remove the 25 nm MS2
bacteriophage directly by sieve action. Moreover,
2.0 because the MS2 bacteriophage (isoelectric pH, 3.9)
and the surface of the membrane are both negatively
charged in the studied pH range (6.3–8.3), substantial
Normalized MS2 Virus
Concentration (N/N0)

1.5 virus adsorption on the membrane was not expected.


The significant virus reductions shown in Fig. 5 suggests
two mechanistic possibilities: (1) enmeshment of viruses
1.0
by the Fe(OH)3 precipitates, and/or (2) a two-step
process where initially Fe(OH)3 flocs are formed, which
0.5 are positively charged in the pH range 6.3 and 7.3 and
pH =6.3 near neutral at pH 8.3 (O’Melia, 1972). In the second
pH =7.3
pH =8.3 step, the negatively charged viruses adsorb onto the
0.0 positively charged Fe(OH)3(s) flocs that were subse-
0 3 6 9 12 15 quently well removed by MF. A similar mechanism
Fe(III) Concentration (mg/L) applies for adsorption of soluble anions onto hydrous
Fig. 3. Determination of virus inactivation by electrocoagula- metal oxides such as goethite (FeOOH) and gibbsite
tion. (N is MS2 concentration after EC treatment and N 0 is (AlOOH) (Hingston, 1981; Schindler, 1981). This is also
MS2 concentration before EC treatment.) Error bars represent analogous to the viruses concentration procedure
one standard deviation. recommended in ‘‘Standard Method 9510D’’ (Clesceri
ARTICLE IN PRESS
B. Zhu et al. / Water Research 39 (2005) 3098–3108 3103

Log Removal of MS2 Bacteriophage


6 6
Adsorption Enmeshment
5 dominant dominant
5
4
pH = 8.3 SWTR virus requirement
3 4
Fe(III) = 12.6 mg/L
Fe(III) = 9.8 mg/L
2 Fe(III) = 7.4 mg/L 3
Fe(III) = 4.5 mg/L
1 Fe(III) = 2.3 mg/L
Log Removal of MS2 Bacteriophage

Fe(III) = 0.0 mg/L 2


0 pH=6.3
1 pH=7.3
5 pH=8.3
0
4
0 2 4 6 8 10 12 14
pH = 7.3
3 Fe(III) Concentration (mg/L)
Fe(III) = 13.1mg/L
Fe(III) = 10.1mg/L
2 Fe(III) = 7.3 mg/L
Fig. 6. The effect of iron dose and pH on MS2 log removal by
Fe(III) = 6.4 mg/L EC-MF.
1 Fe(III) = 2.8 mg/L
Fe(III) = 0.0 mg/L
0
3.5. Effect of pH on MS2 virus removal by EC-MF
5
As shown in Fig. 6, a bulk solution pH of 6.3 and
4 pH = 6.3
lower Fe(III) dosages (o9 mg/L) generally achieved
3 Fe(III) = 10.0 mg/L slightly higher MS2 removal compared with 7.3 and
Fe(III) = 7.6 mg/L
Fe(III) = 6.0 mg/L even 8.3, which is approximately the pH of least
2 Fe(III) = 4.9 mg/L
Fe(III) = 3.0 mg/L
solubility for iron (O’Melia, 1972). However, pH did
1 Fe(III) = 1.7 mg/L not have a substantial effect on virus removal at larger
Fe(III) = 0.0 mg/L
0 Fe(III) dosages. At pH 6.3, Fe(OH)+ 2 and Fe(OH)
2+
are
0 20 40 60 80 100 120 the dominant dissolved iron species and are thought to
Volume Filtered/Membrane Area (mL/cm2)
be the species present on the surfaces of the positively
charged Fe(III) flocs (O’Melia, 1972). At pH 7.3, Fe(III)
Fig. 5. The effect of coagulant dose and volume filtered (cake flocs are less positively charged compared with pH 6.3
thickness) on MS2 removal by EC-MF. because Fe(OH)+ 2 and Fe(OH)03 are the dominant
dissolved and surface bound iron species. Fe(III) flocs
are near neutral at pH 8.3 because Fe(OH)03 is the
dominant dissolved iron species. Since pKa of the MS2
et al., 1998) wherein they are first adsorbed onto bacteriophage is 3.9, it is substantially negatively
positively charged flocs and later eluted in alkaline charged at all pH conditions investigated in this study.
solution where flocs become negatively charged and/or Hence, greater MS2 removal at lower Fe(III) dosages
more soluble. and at pH 6.3 by EC-MF is consistent with MS2
Log virus removal increased with iron-coagulant dose adsorption onto flocs and subsequent removal by the
(Fig. 5) because as dose increased, (1) the coagulant membrane. Similar results were obtained in CC-MF
surface area and number of adsorptions sites available experiments conducted by us (Zhu et al., 2004).
increased and (2) more Fe(OH)3 precipitated that During iron EC, the following electrochemical reac-
encapsulated additional viruses. The above mechanisms tions take place:
of virus removal are consistent with the negligible 0 3+ 
Anode: Fe (s) ¼ Fe +3e and
impact of membrane cake layer thickness on virus 3+ +
Fe +3H2O ¼ Fe(OH)3(s)+3H
removal, which can be deduced from Fig. 5, where cake  

layer thickness increased with increasing volume filtered.


Cathode: 3H2O+3e ¼ 3OH +32H2
0

As observed, virus removal remained approximately Overall: Fe +3H2O ¼ Fe(OH)3(s)+32H2


constant at any fixed iron dose even as thicker cake
layers were deposited as filtration progressed. These
results are similar to those observed in our previous iron Thus, ferric ions produced at the anode hydrolyze to
coagulation-microfiltration research (Zhu et al., 2004). produce protons, thereby reducing the pH in the area
To further investigate virus removal mechanisms, near the anode. The significant advantage of EC over
experimental data from different pH conditions were CC is that the hydroxide ions generated at the cathode
examined. neutralize these protons in the bulk solution eliminating
ARTICLE IN PRESS
3104 B. Zhu et al. / Water Research 39 (2005) 3098–3108

the consumption of alkalinity during metal-ion hydro- of other researchers (Tsouris et al., 2001; Han et al.,
lysis, which occurs during CC when ferric or aluminum 2002; Holt et al., 2002) have been obtained by unsteady-
salts are used. The bulk-solution-pH effect on virus state operation of the EC unit in batch mode. In order to
removal (Fig. 6) would have been decreased by the fact establish the equivalence of batch unsteady-state experi-
that virus adsorption onto the coagulant primarily mental results described thus far and continuous-flow
occurred in the low-pH region near the anode. Here, EC pretreatment, limited steady-state experiments were
the iron (III) hydroxide complexes would be even more also conducted at pH 7.3 corresponding to water
positively charged at all studied bulk solution pH treatment EC practice. Negligible differences between
conditions (6.3–8.3). Thus, the pH near the anode had the two operating modes at pH 7.3 can be deduced from
a major effect on MS2 virus adsorption onto iron floc, Fig. 8. Based on these tests, which were run in triplicate
whereas, the bulk solution pH had a less noticeable at five different iron dosages, it can be inferred that
effect. operating an EC unit in batch mode, as is typically
At high iron dosages (49 mg/L in our system), employed in lab-scale studies, provides meaningful
Fe(OH)3 precipitates encapsulate MS2 viruses regardless results that are also directly applicable to continuous
of pH and remove them by enmeshment resulting in operation. (It should be noted that this work did not
similar removals at pH 6.3, 7.3 and 8.3 (Fig. 6). explicitly consider scaling-up the microfiltration pro-
cess.)
3.6. Effect of ionic strength on MS2 removal by EC-MF

MS2 removal by EC-MF was investigated in the range 3.8. Comparison of chemical- and electrocoagulation
of ionic strength expected in fresh waters: low pretreatment
(I ¼ 107 M), medium (I ¼ 6  103 M), and high
(I ¼ 1:8  102 M). Results of these experiments, which A comparison of the chemical coagulation and EC
were carried out at pH 6.3 are depicted in Fig. 7. As pretreatment methodologies for MF at different pHs
observed, ionic strength in the 104 to 18 mM range did and iron dosages is presented in Fig. 9. Fe(III)
not significantly influence virus removal. Regardless of coagulation either by EC or CC pretreatment dramati-
ionic strength, iron generated by EC dramatically cally improved virus removal. Furthermore, regardless
improved the MS2 virus removal from o0.5 log in the of pH, the Fe(III) dosage required to achieve a given
absence of pretreatment to 44 logs at 7 mg Fe/L. MS2 virus LRV was always significantly less for EC
compared with CC pretreatment. For example, 4-log
(99.99%) reduction of viruses was readily achieved by
3.7. Comparison of virus removal by EC-MF between
EC with 6–9 mg/L Fe(III) dosage, whereas only 2-log
batch unsteady-state and continuous flow steady-state
virus reduction was achieved with CC pretreatment in
operation conditions
this same Fe(III) dosage range. The fact that EC
significantly outperformed chemical coagulation de-
In actual practice, EC systems are operated continu-
serves explanation.
ously ultimately reaching steady-state. However all
results presented thus far in this paper as well as those
Log Removal of MS2 Bacteriophage

Log Removal of MS2 Bacteriophage

6 6

5 5

4 4

3 3

2 2
pH =6.3

1 DI Water + 3 mM NaHCO3 + 10 mM CaCl2


1 Batch mode
DI Water + 3 mM NaHCO3 + 1 mMCaCl2
DI Water Continuous flow mode
0 0
0 2 4 6 8 10 12 0 2 4 6 8 10 12 14
Fe(III) Concentration (mg/L) Fe(III) Concentration (mg/L)

Fig. 7. The effect of Fe(III) dosage and ionic strength on MS2 Fig. 8. MS2 virus removal by EC-MF at batch non steady-state
virus removal by EC-MF. Feed MS2 virus concentra- and continuous steady-state operation conditions at pH ¼ 7.3,
tion ¼ 106 pfu/mL. feed MS2 virus concentration ¼ 106 pfu/mL.
ARTICLE IN PRESS
B. Zhu et al. / Water Research 39 (2005) 3098–3108 3105

6
Log Removal of MS2 Bacteriophage

pH =8.3 pH =7.3 pH = 6.3


5

4
EC
EC EC
3

2 CC

CC CC
1

0
0 2 4 6 8 10 12 0 2 4 6 8 10 12 0 2 4 6 8 10 12
Fe(III) Concentration (mg/L)

Fig. 9. Comparison of virus removal by EC-MF and CC-MF as a function of iron dose.

Log removal of MS2 bacteriophage


Fig. 9 demonstrates that iron was used more
effectively during EC compared with CC as performed 3 CC-MF using EC as a mixer
CC-MF using a square-jar mixer
in this study. Iron is generated continuously over the
extended area of the anode in EC compared with the
point addition of iron in CC. Hence, freshly precipitated 2
flocs are generated and more effectively dispersed in EC,
which will increase adsorptive removal of viruses as
reported in ‘‘Standard Method 9510D’’ (Clesceri et al.,
1998). As demonstrated in Section 3.2, there was no 1
virus inactivation or accumulation within the EC unit.
However, differences in mixing or materials of construc-
tion between EC-MF and CC-MF processes could have 0
caused EC to perform better than CC. Results from 0 2 4 6 8 10 12
control experiments designed to evaluate these experi- Fe(III) Concentration (mg/L)
mental artifacts are presented next.
Fig. 10. The effect of possible difference in hydraulic mixing on
virus removal by CC-MF at pH ¼ 6.3, and MS2 feed
3.9. CC-MF for MS2 virus removal using the EC Unit as concentration ¼ 106 pfu/mL.
a mixer

The effect of hydraulic mixing in the EC unit on virus


removal was investigated using the EC unit solely as a CC resulting in EC outperforming CC as MF pretreat-
hydraulic mixer for CC with the power turned off (no ment for virus removal.
electrochemical iron generation). Batch CC-MF tests
were conducted with 400 mL of MS2 virus-spiked test
water. This test water and FeCl3 (at doses of 0, 2, 5, and 3.10. Proposed explanation for why EC-MF
10 mg Fe/L) were simultaneously input to the EC unit outperformed CC-MF
Fig. 10 depicts very similar trends for virus removal
by CC-MF using the standard square ‘‘gator’’ jar for During chemical coagulation, the pH, viruses, and
mixing and CC-MF using the EC unit as a mixer with iron hydroxides were uniformly dispersed in the jar.
the recirculation pump on and the EC power off. Log However, during EC, iron coagulant was continuously
removal of MS2 virus was essentially the same at a fixed generated and released from the anode surface to water.
iron dose under both mixing conditions, demonstrating Hence, the iron concentration decreased as distance
that hydraulic mixing in the unit did not enhance virus from the anode increased. Also, the pH in the area near
removal by EC pretreatment compared with CC. Hence, the anode decreased due to the local hydrolysis of Fe3+
virus adsorption onto iron flocs or enmeshment by iron ions to produce protons. This has been verified by Chen
flocs was significantly better during EC compared with et al. (2002) in an EC system that used an aluminum
ARTICLE IN PRESS
3106 B. Zhu et al. / Water Research 39 (2005) 3098–3108

anode. Additionally, since the MS2 virus was negatively 3.10.2. Experimental verification for the improvement of
charged under our experimental conditions (pH 6.3, 7.3, virus removal at locally high iron and virus concentrations
and 8.3), and the anode was positively charged after near the anode
turning on the EC power, negatively charged MS2 virus A high-iron, -virus concentration CC-MF process was
could electrophoretically move resulting in a higher designed to simulate the EC-MF process conditions of
virus concentration near the anode. Further, in the locally high iron and MS2 virus concentrations in the
immediate region around the anode, there was a locally area near the anodes.
higher concentration of Fe(OH)3(s) flocs compared with The first experiment simulated a local doubling of the
the average concentration of Fe(OH)3(s) flocs during iron and virus concentrations at pH 6.3. During
chemical coagulation in the jar mixer. Thus, locally adsorption and mixing, the doubled iron and virus
higher coagulant and virus concentrations might have concentrations were 2  106 pfu/mL and 10 mg/L, re-
been the reasons for better performance of EC spectively. After adsorption, mixing, and dilution, the
compared with CC. These possibilities are examined in final iron and virus concentrations were 106 pfu/mL and
the following two sections by using a simple model of 5 mg-Fe/L, respectively. The experimental procedure
virus transport and by performing experiments to was as follows:
determine the effects of virus and iron coagulant
concentrations on virus removal.
(1) Add source virus into 100 mL source water (DI
water spiked with 1 mM CaCl2, and 3 mM NaH-
3.10.1. Simple model of virus movement in the EC unit CO3). The target virus concentration was
Two-dimensional virus transport in the EC unit; from 2  106 pfu/mL.
the inlet to the outlet (convection dominated) and from (2) Take samples after 4 s mixing (Sample 1).
the cathode to the anode (electrophoresis) was consid- (3) Add iron coagulant solution to produce a target
ered. A simple expression for the virus electrophoretic Fe(OH)3(s) concentration of 10 mg/L as Fe.
velocity (v) was developed by equating the Stokes drag (4) Take samples after 24 s mixing (Sample 2).
force and the force induced by the imposed electric field (5) Dilute the mixture back to the desire average 5 mg-
(Zhu, 2004): Fe/L and 106 pfu/mL virus concentrations by adding
QE   100 mL source water without MS2 viruses present.
v¼ 1  eð6pZrt=mÞ , (3) (6) Mix the diluted sample for 10 s, and take samples for
6pZr
iron and virus analysis (Sample 3).
where Q (Coulombs) is the net surface charge of the (7) Transfer the water to the feed bottle of the MF
MS2 virus (647, 680, and 685 e at pH 6.3, 7.3, and 8.3, system.
respectively (Penrod et al., 1996)); E is the electrical field (8) Filter the water using MF in the dead-end mode at
strength (0.57 V/mm); t is the time; m is the mass of a constant pressure (69 kPa, 10 psig).
single virus (2,484,700 Dalton or 4.1  1018 g as re- (9) Take samples from the effluent and analyze for virus
ported by Tito et al. (2000)); Z is the water viscosity concentration (Sample 4).
(8.9  104 N s/m2 at 23 1C); and r is the virus radius
(12.5  109 m).
This simple model revealed that under our experi- These experiments were performed at pH 6.3 and in
mental conditions, the maximum virus electrophoretic triplicate as were similar experiments for 4- and 20-times
velocity was reached in much less than one second for all higher iron and MS2 virus concentrations. The final iron
studied pHs. As depicted in Fig. 1, the annular distance and virus concentrations (Sample 3) for all of these
between the anode and the cathode was 1.85 mm and the experiments were 5 mg-Fe/L, and 106 pfu/mL, respec-
hydraulic detention time in the EC unit was 24 s tively. There was no statistically significant difference
(excluding recycle). Thus, a virus particle entering at between Samples 1 and 2 for any experiment, indicating
the maximum radial distance from the anode (1.85 mm) that the increased levels of Fe(III) did not affect the
would have to move towards the anode at a minimum virus viability.
velocity of 0.077 mm/s in order to reach the anode Fig. 11 summarizes the LRVs of all the high-iron,
before exiting the unit. The electrophoretic velocities high-virus concentration experiments based on the
calculated from Eq. (3) were 5.07, 5.33, and 5.37 mm/s at difference between Samples 3 and 4 for each experiment.
pHs 6.3, 7.3, and 8.3, respectively. Because the electro- Virus removal obtained during the control experiment
phoretic velocities were all greater than 0.077 mm/s, was 1.5 log, whereas, greater removals, 2.6-, 3.8-, and 5-
MS2 viruses could readily travel the maximum distance log, were achieved at 2-, 4-, and 20-times higher iron and
to the anode during their residence time in the EC unit, virus concentrations, respectively. Thus, locally high
which would result in a higher virus concentration in the iron and virus concentrations significantly improved
proximity of the anode. virus removal, which helps explain why EC with its
ARTICLE IN PRESS
Log Removal of MS2 Bacteriophage B. Zhu et al. / Water Research 39 (2005) 3098–3108 3107

6.0 virus concentrations within the EC unit improved virus


removal.
5.0 These bench-scale results, which demonstrate the
significant advantage of iron EC compared with CC
4.0 for virus removal, should be verified at the pilot scale.
Furthermore, the results here suggest that EC might also
3.0 be superior to CC for natural organic matter removal
during membrane pretreatment.
2.0

1.0
Acknowledgments
0.0
0 4 8 12 16 20 We thank Prof. Michael Benedik (Department of
Iron and Virus Concentration Factor Biology and Biochemistry, University of Houston) and
Dr. Vidya Gopalakrishnan (Department of Molecular
Fig. 11. Enhancements in virus removal resulting from 2 to 20-
times higher iron and virus concentrations. Final iron
Genetics, The University of Texas, M.D. Anderson
concentration ¼ 5 mg/L, pH ¼ 6.3, final target MS2 virus Cancer Center) who instructed us in the methods of
concentration ¼ 106 pfu/mL. enumerating coliphages. Wendong Xu is acknowledged
for his help in settling up the experiments. This research
was funded by grants from the Texas Higher Education
Coordinating Board (project No. 003652-0412-1999),
the National Science Foundation (BES-0134301), and
locally higher iron and virus concentrations near the the University of Houston. The contents do not
anode was superior to CC for virus removal. necessarily reflect the views and policies of the sponsors
nor does the mention of trade names or commercial
products constitute endorsement or recommendation for
use.
4. Conclusions

Iron electrocoagulation proved to be an excellent


References
pretreatment for virus removal by MF. The LRV for
MS2 virus suspended in simulated natural water was 4.5 Brock, T.D., Madigan, M.T., 1991. Biology of Microorgan-
(99.997% removal) when generating 10 mg/L Fe by EC. isms. Prentice-Hall, Englewood Cliffs, NJ.
As expected, virus removal improved with increasing Can, O.T., Bayramoglu, M., Kobya, M., 2003. Decolorization
iron concentration. Without EC pretreatment, the LRV of reactive dye solutions by electrocoagulation using
for MF was only 0.5 (32% removal). aluminum electrodes. Ind. Eng. Chem. Res. 42 (14),
At low iron dosages and lower pH values (6.3 and 3391–3396.
7.3), virus removal by EC-MF was predominantly due Chen, X., Chen, G., Yue, P.L., 2002. Investigation on the
to adsorption of the negatively charged viruses onto the electrolysis voltage of electrocoagulation. Chem. Eng. Sci.
57 (13), 2449–2455.
positively charged iron flocs and the subsequent removal
Clesceri, L.S., Greenberg, A.E., Eaton, A.D., 1998. Standard
of the flocs by MF. However, virus removal at high
Methods for the Examination of Water and Wastewater,
Fe(III) dosages was attributed to enmeshment prior to 20th ed. American Public Health Association/American
removal by MF. These results suggest that other Water Works Association/Water Environment Federation,
clarification technologies capable of removing flocs Washington, DC.
(e.g. media filtration) would also be effective for virus Coffey, B.M., Stewart, M.H., Wattier, K.L., Wale, R.T., 1993.
removal, as was MF in this study. Evaluation of microfiltration for metropolitan’s small
In a comparison of the processes, EC significantly domestic water systems. AWWA Membrane Technology
outperformed chemical coagulation (CC) with ferric Conference Proceedings, Baltimore, MD.
chloride as a pretreatment for MF. For example, at pH Drees, K.P., Abbaszadegan, M., Maier, R.M., 2003. Compara-
tive electrochemical inactivation of bacteria and bacter-
7.3 and an iron dose of 10 mg/L, the LRVs for MS2
iophage. Water Res. 37 (10), 2291–2300.
virus removal were 4.5 and 2.0 for EC and CC,
EPA, 2001. Low-pressure membrane filtration for pathogen
respectively. The dramatic improvement in virus re- removal: application, implementation, and regulatory
moval by EC compared with CC held true for all iron issues. EPA 815-C-01-001, Office of Water (4601).
dosages (2–10 mg/L) and pHs tested (6.3, 7.3, and 8.3). Han, M., Song, J., Kwon, A., 2002. Preliminary investigation of
It was proposed and subsequently verified by modeling electrocoagulation as a substitute for chemical coagulation.
and additional experiments that locally higher iron and Water Supply 2 (5-6), 73–76.
ARTICLE IN PRESS
3108 B. Zhu et al. / Water Research 39 (2005) 3098–3108

Hingston, F.J., 1981. A review of anion adsorption. In: Penrod, S.L., Olson, T.M., Grant, S.B., 1996. Deposition
Anderson, M.A., Rubin, A.J. (Eds.), Adsorption of kinetics of two viruses in packed beds of quartz granular
Inorganics at Solid–Liquid Interfaces. Ann Arbor Science, media. Langmuir 12 (23), 5576–5587.
Ann Arbor, MI. Roberts, P., 1997. Efficient removal of viruses by a novel
Holt, P.K., Barton, G.W., Wark, M., Mitchell, C.A., 2002. A polyvinylidene fluoride membrane filter. J. Virol. Meth. 65
quantitative comparison between chemical dosing and (1), 27–31.
electrocoagulation. Colloids Surf. A: Physicochem. Eng. Schindler, P.W., 1981. Surface complexes at oxide-water
Aspects 211, 233–248. interfaces. In: Anderson, M.A., Rubin, A.J. (Eds.), Adsorp-
Horner, G., Duffey, J.G., 1983. Electrochemical Removal of tion of Inorganics at Solid–Liquid Interfaces. Ann Arbor
Heavy Metals from Waste Water. The American Electro- Science, Ann Arbor, M.I (Chapter 1).
platers Society Annual Meeting, Denver, CO. Tito, M.A., Tars, K., Valegard, K., Hajdu, J., Robinson, C.V.,
Jacangelo, J.G., Laine, J.M., Carns, K.E., Cummings, E.W., 2000. Electrospray time-of-flight mass spectrometry of the
Mallevialle, J., 1991. Low-pressure membrane filtration for intact MS2 virus capsid. J. Am. Chem. Soc. 122 (14),
removing Giardia and microbial indicators. J. Am. Water 3550–3551.
Works Assoc. 83 (9), 97–106. Tsouris, C., DePaoli, D.W., Shor, J.T., Hu, M., Ying, T.-Y.,
Lahoussine-Turcaud, V., Wiesner, M.R., Bottero, J.Y., Mal- 2001. Electrocoagulation for magnetic seeding of colloidal
levialle, J., 1990. Coagulation pretreatment for ultrafiltra- particles. Colloids Surf. A: Physicochem. Eng. Aspects 177
tion of a surface water. J. Am. Water Works Assoc. 82 (12), (3), 223–233.
76–81. Urase, T., Yamamoto, K., Ohgaki, S., 1996. Effect of pore
Liu, W.K., Brown, M., Elliott, T.S., 1997. Mechanisms of the structure of membranes and module configuration on virus
bactericidal activity of low amperage electric current (DC). retention. J. Memb. Sci. 115 (1), 21–29.
J. Antimicrob. Chemother. 39 (6), 687–695. Valegård, K., Liljas, L., Fridborg, K., Unge, T., 1990. The
Madaeni, S.S., 1999. The application of membrane technology three-dimensional structure of the bacterial virus MS2.
for water disinfection. Water Res. 33 (2), 301–308. Nature 345, 36–40.
Madaeni, S.S., Fane, A.G., Grohmann, G.S., 1995. Virus Wiesner, M.R., Veerapaneni, S., Brejchova, D., 1992. Improve-
removal from water and wastewater using membranes. J. ments in membrane microfiltration using coagulation
Memb. Sci. 102 (1-3), 65–75. pretreatment. Proceedings of the Fifth Gothenburg Sympo-
Mallevialle, J., Odendaal, P.E., Wiesner, M.R., 1996. Water sium, Nice, France.
Treatment Membrane Processes. McGraw-Hill, New York. Zhu, B., 2004. The development of an electrocoagulation-
Mills, D., 2000. A new process for electrocoagulation. J. Am. microfiltration process for virus removal from drinking
Water Works Assoc. 92 (6), 34–43. water. Doctoral Dissertation, University of Houston,
O’Melia, C.R., 1972. Coagulation and flocculation. In: Weber, Houston, TX.
W.J. (Ed.), Physicochemical Processes for Water Quality Zhu, B., Clifford, D., Chellam, S., 2004. Virus removal by iron
Control. Wiley, New York (Chapter 2). coagulation-microfiltration. Water Research, under revision.

Vous aimerez peut-être aussi