Vous êtes sur la page 1sur 5

Excerpt from the Proceedings of the COMSOL Users Conference 2006 Birmingham

Simulation of Blood Flow through the Mitral Valve of the Heart: A


Fluid Structure Interaction Model.
Daniel M. Espino*1, Michael A. Watkins1, Duncan E.T. Shepherd1, David W.L. Hukins1 and
Keith G. Buchan2
1
Department of Mechanical and Manufacturing Engineering, University of Birmingham,
Birmingham, UK, 2Aberdeen Royal Infirmary, Aberdeen, UK
*Corresponding author: Department of Mechanical and Manufacturing Engineering, University of
Birmingham, Birmingham, UK, d.m.espino@bham.ac.uk

Abstract: Mitral valve failure causes death if it Several Finite Element (FE) models of the
is not corrected surgically. Such surgical repair mitral valve have been developed; however,
can be improved by understanding mitral there are fewer that account for the effect of fluid
biomechanics. A two-dimensional FSI model of flow through the valve (Einstein et al., 2005a,b;
the mitral valve was generated, using an ALE Van Loon et al., 2006; Watton et al., 2006). Of
mesh. A simple approximation of the heart the developed FSI models of the mitral valve, all
geometry was used, the valve dimensions were have simulated the flow of blood through the
based on measurements made. The viscosity of valve in a conduit and ignored the effect of the
blood and the elastic properties of the valve left ventricle of the heart on blood flow.
leaflets were obtained from the literature. Valve Furthermore, one model simulated a
closure was simulated using contact equations. polyurethane replacement mitral valve (Watton
The simulation predicted a large vortex behind et al., 2006), and another a single leaflet in a tube
the anterior leaflet during inflow of blood into (Van Loon et al., 2006). Only one FSI model of
the left ventricle, in agreement with MRI scans an actual mitral valve, with two leaflets has been
available in the literature. Leaflet deformations developed (Einstein et al., 2005a,b).
agreed with results from experiments in the The FSI model of the mitral valve we are
literature and with our previous experimental developing includes the walls of the left ventricle
results. In conclusion, a two-dimensional model of the heart, both anterior and posterior leaflets
of the mitral valve has been developed, and will of the valve, and the outflow tract of the aorta.
be further developed to investigate heart valve
failure and subsequent surgical repair. 2. Methods
Keywords: Mitral valve, Fluid-Structure 2.1 Model overview
Interaction, Biomechanics.
A two-dimensional FSI model of the mitral
1. Introduction valve within the heart was generated using
Comsol Multiphysics (Version 3.2). Lagrange
Failure of the mitral valve of the heart can be multipliers have been used to apply the forces
fatal if it is not corrected surgically (Espino et exerted on the deforming structure due to the
al., 2005, 2006a). Improvement in the flow of fluid, as has been done for other heart
understanding of the biomechanics of the mitral valve FSI simulations (De Hart et al., 2003).
valve would aid current knowledge of mitral Velocity constraints were applied to the fluid on
valve function, its failure and possible surgical the fluid-structure boundary, to ensure fluid at
correction. Therefore, a Fluid-Structure that location moved at the same velocity as the
Interaction (FSI) model of the mitral valve has moving structure, as in other simulations (De
been generated. Hart et al., 2003). Structural deformation and
The mitral valve is present in the left side of fluid dynamics are determined simultaneously.
the heart, and functions normally to allow blood An Arbitrary Lagrange Euler (ALE) mesh was
to flow into the left ventricle of the heart when it used, to allow FSI simulations to be performed.
is filling. This valve then closes when blood is
pumped out from the heart towards the body. In
doing so, it prevents the regurgitation of fluid.
Excerpt from the Proceedings of the COMSOL Users Conference 2006 Birmingham

2.2 Geometry Table 1. Values of parameters used to describe blood


and heart valves during simulations.
Two geometries of the left ventricle of the
heart have been generated. Initially a simple Parameter Value
approximation of the heart was made, as being Blood density (kg/m3) 1.06 × 103
elliptical in shape (i.e. half an ellipse), the major Blood viscosity (Pa.s) 2.70 × 10-3
and minor axis were 70 mm and 24 mm Leaflet density (kg/m3) 1.06 × 103
respectively (based on measurements on heart Anterior leaflet Young’s
specimens, figure 1). The dimensions of the modulus (MPa) 2.0 × 106
valve leaflets were based on measurements made Posterior leaflet Young’s
(approximately 20 mm length and 0.7 mm thick; modulus (MPa) 1.0 × 106
figure 1b).
Leaflet Poisson’s ratio 0.49

2.4 Boundary conditions

Inflow boundary conditions were applied to


simulate valve opening during diastole and valve
closure during systole. During diastole an inflow
velocity was applied at the atrium and a pressure
condition was applied at the apex of the heart, to
apply the relevant ventricular pressures. During
systole, ventricular pressure was applied from
the apex of the heart, and outflow boundary
conditions were applied at the aortic outlet.
During systole only the stage of outflow of blood
a) b)
was simulated.
Figure 1. Model of the left ventricle geometry used.
a) Elliptical model of the left ventricle; b) dissected 2.5 Contact simulation
specimen of the mitral valve, with part of the heart
muscle dissected, showing the mitral valve. Valve closure was simulated using contact
equations between the boundaries of the two
A geometrically more accurate model has mitral valve leaflets that come into contact.
also been generated that matches a study in Integration coupling variables were used to allow
which measurements of pressure, flow and flow information regarding the position of the
patterns were used to quantitatively validate the contacting boundaries available to each other all
prediction of the model, for inflow during valve along the boundary of the leaflets.
opening (diastole; Reul et al., 1980). The two contacting boundaries were
The simplified elliptical model has been used separated into small segments, with nodes at the
to simulate systolic ventricular pressures applied end of each small segment. The distance between
to generate blood flow within the heart chamber, nodes of one boundary on one leaflet and the
causing valve closure, and blood flow through whole boundary on the other leaflet was
the aortic outlet. determined, and applied to equation 1.

2.3 Parameters  Pe − gk / P g > 0


F = 1
The viscosity of blood and the elastic  P − gk g ≤ 0
properties of the valve leaflets were obtained
from the literature (Table 1). In equation 1, P is the applied fluid pressure,
k is the stiffness of the contact, and g the
distance between the two contacting points. F is
the resulting contact force at each node.
Excerpt from the Proceedings of the COMSOL Users Conference 2006 Birmingham

These contact forces were calculated at each


node for both leaflet boundaries that came into
contact. The force acting along the whole
boundary was then determined by interpolation
between the two nodes at the end of each
boundary (equation 2). It was assumed that the
force in between two nodes would be
continuous.

Fb = (r1/r2 × (F2 –F2)) + F1 2

In equation 2, F1 is the contact force at node-


1 and F2 the contact force at node-2 (where node-
1 and node-2 define the start and end points of a
boundary, respectively). Fb defines the contact Figure 3. Predicted flow through a model of the left
force at any point on the boundary. While. r1/r2 ventricle with a geometry that corresponds directly to
defines the ratio of the distance from node-1 to a a heart from which measurements of flow and
given point (r1), to the whole length of the subsequent flow patterns were made.
boundary (r2).
Leaflet deformations, during opening and
3. Results closure, agreed with results from experiments in
the literature (figure 4). During systole, outflow
The flow of blood caused the valve leaflets to of blood occurred through the aortic outlet (as
deform, in turn altering the flow of blood. The reported in the medical literature), and mitral
simulation predicted a large vortex behind the valve leaflet closure agreed with our previous
anterior leaflet during inflow of blood into the experimental results (Espino et al., 2006b).
left ventricle (figure 2), in agreement with MRI
scans available in the literature (Bronzino, 2000).

Figure 4. Simulated closure of the mitral valve of the


heart.

Results of simulating the closure of the mitral


Figure 2. Large vortex behind anterior leaflet valve leaflets using the contact equations, are
generated during diastole.
shown in figure 5. Little overlap occurred
between the two leaflets, which had different
Measurements were taken from specific
material properties (table 1).
points (figure 3), for quantitative comparison
with published experimental results (Reul et al.,
1980). Reasonable agreement was obtained.
Excerpt from the Proceedings of the COMSOL Users Conference 2006 Birmingham

results (Espino et al., 2005, 2006a,b), and


general literature (Bronzino et al., 2000; Reul et
al., 1980). Overall, our predicted results show
good agreement with observations available in
the literature.
This model will be used in future, to
determine stresses experienced by mitral valves
undergoing in vitro testing in our Laboratory to
understand the stresses the valve experienced.
a)
Furthermore, it is to be developed into a three-
dimensional model.

5. Conclusions

A two-dimensional model of the mitral valve of


the heart has been developed. The results from
simulations agree with observations in the
medical literature. The model will be used to
b)
investigate heart valve failure and subsequent
surgical repair.

6. References

Bronzino, Biomedical Engineering Handbook.


CRC Press LLC, Boca Raton (2000).
De Hart et al., A three-dimensional
computational analysis of fluid–structure
interaction in the aortic valve, Journal of
c) Biomechanics, 36, 103-112 (2003).
Espino et al., The role of chordae tendineae in
Figure 5. Simulated mitral valve leaflet closure. a) mitral valve competence, Journal of Heart Valve
Leaflets before application of pressure; b) leaflet Disease, 14, 603-609 (2005).
closure position after application of pressure; c) close Espino et al., Mitral valve repair: an in vitro
up of leaflet section in contact.
comparison of the effect of surgical repair on the
pressure required to cause mitral valve
4. Discussion regurgitation, Journal of Heart Valve Disease,
15, 375-381 (2006a).
A two dimensional FSI model of the mitral Espino et al., Determination of the pressure
valve within the heart has been generated, and required to cause mitral valve failure, Medical
transient simulations have been performed. Engineering and Physics, 28, 36-41 (2006b).
Simulations performed include the inflow of Einstein et al., The relationship of normal and
blood into the left ventricle of the heart, and abnormal microstructural proliferation to the
outflow of blood through the aorta. The latter mitral valve closure sound, Journal of
simulation has required the use of contact Biomechanical Engineering, 127, 134-147
equations to simulate the closure. (2005a).
One of the models simulated was developed Einstein et al., Non-linear fluid-coupled
to match the geometry of a heart investigated in a computational model of the mitral valve, Journal
study where flow patterns and measurements of of Heart Valve Disease, 14, 376-385 (2005b).
flow and pressure were made. This has allowed Reul et al., Fluid mechanics of the natural mitral
direct quantification of some results. A more valve , Journal of Biomechanics, 14, 361-372
generalized model has been used more (1980).
extensively, but validation is more qualitative, Van Loon et al., A fluid-structure interaction
e.g. comparison to our previous experimental method with solid-rigid contact for heart valve
Excerpt from the Proceedings of the COMSOL Users Conference 2006 Birmingham

dynamics, Journal of Computational Physics,


217, 806-823 (2006).
Watton et al., Dynamic modelling of prosthetic
chorded mitral valves using the immersed
boundary method, Journal of Biomechanics, In
Press (2006).

7. Acknowledgements

The authors thank: The British Heart


Foundation for a Junior Fellowship awarded to
DME (FS/05/033), and the Nuffield Foundation
for an Undergraduate Research Bursary awarded
to MAW.

Vous aimerez peut-être aussi