Vous êtes sur la page 1sur 12

November 2002

Circulatory Assist Units, Intra-Aortic Balloon


additional support to pharmacologic treatments to par-
tially restore cardiac function, as well as act as an aid
Scope of this Product Comparison
to the heart before, during, and after cardiac surgery.
This Product Comparison covers intra-aortic bal-
loon pump (IABP) consoles and intra-aortic balloon IAB pumping is often used to treat patients with
(IAB) catheters. IAB catheters are disposable and unstable angina, patients who have suffered a myocar-
are often supplied by the IABP manufacturer. Ac- dial infarction (augmenting coronary perfusion helps
cessories such as balloon interconnects (for use develop collateral circulation to aid the healing of
when one manufacturer’s balloon is used on an- ischemic areas of the myocardium) or other pump
other manufacturer’s IABP), training simulators, failure, and patients undergoing percutaneous trans-
and cardiac-output computers are not included. For luminal coronary angioplasty or myocardial revascu-
more information, see the Product Comparison ti- larization. It is also used for weaning patients with
tled CARDIAC OUTPUT UNITS, THERMAL DILUTION. inadequate cardiac output from cardiopulmonary by-
pass surgery.
UMDNS information
This Product Comparison covers the following Principles of operation
device term and product code as listed in ECRI’s The coronary arteries, which originate at the base
Universal Medical Device Nomenclature System™ of the ascending aorta, supply blood and oxygen to the
(UMDNS™): heart. As the heart contracts (systole), these vessels
• Circulatory Assist Units, Intra-Aortic Balloon are compressed, reducing coronary perfusion. As a
[10-846] result, 70% to 90% of coronary perfusion occurs during
the heart’s filling stage (diastole). To assist in the
perfusion of the coronary arteries, the counterpulsa-
Purpose tion technique — inflation and deflation in synchrony
with the cardiac cycle — was developed.
The IABP is a mechanical counterpulsation device
that benefits the patient by reducing afterload as well In preparation for counterpulsation, the balloon
as increasing coronary and systemic blood flow. A catheter, which is mounted on a flexible radiopaque
balloon is placed in the aorta, and its inflation and catheter, is inserted, typically percutaneously with a
deflation results in a more favorable myocardial sup- guide wire through the femoral artery into the aorta.
ply-and-demand balance. This balance reduces after- Less commonly, the balloon catheter can also be surgi-
load and augments the heart’s diastolic pressure, cally placed into the aorta using a prosthetic graft. The
leading to an increased cardiac output (Christenson et balloon catheter is attached to the IABP console, which
al. 1997). Because IABP therapy reduces the workload displays heart rate, arterial pressure, balloon pressure
and oxygen demand of the heart while increasing oxy- and volume, alarm conditions, setup instructions, and
gen supply and cardiac output, it can be used to treat electrocardiogram (ECG) readings. The console uses
a number of cardiovascular conditions. IABPs provide the R wave of the ECG for automatic timing control for

214897 5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA


424-008 Telephone +1 (610) 825-6000 ● Fax +1 (610) 834-1275 ● E-mail hpcs@ecri.org
Healthcare Product Comparison System

the inflation and deflation of the balloon. Balloons are


available in several volumes; 30 to 50 cc balloons are
most commonly used for adult patients. Pediatric sizes
from 2.5 to 20.0 cc are also available, although not all
IABPs can accommodate them.
The IAB forces the blood toward the heart during
each diastolic phase of the cardiac cycle. During this
phase, the aortic valve is closed, and the blood is
channeled into the coronary and carotid arteries, in-
creasing coronary and cerebral perfusion (see Fig. 1).
Blood is also displaced systemically as the balloon is
inflated, improving systemic perfusion. Catheters are
typically double lumen and have an outer gas lumen
and a central lumen through which aortic pressures
can be monitored and dye can be injected to determine
the position of the balloon catheter.
The physician maneuvers the balloon catheter until
it occupies the descending aorta; fluoroscopy, chest
x-ray, or transesophageal echocardiography verifies
positioning. Balloon size is selected so that 75% to 90%
of the aortic lumen is occluded when the balloon is
inflated. The balloon is made of a thrombosis-resistant
polymer and has either one or two chambers. Dual-
chambered balloons are designed for unidirectional
flow (toward the heart); the more widely used single-
chambered balloons encourage bidirectional flow.
The balloon is cyclically inflated and deflated using
helium shuttle gas kept in a storage compartment in
the console. A pneumatic system automatically moves
a preset volume of gas to and from the balloon by
alternately applying pressure and vacuum. Helium is
a low-density gas that is better suited for pumping
(particularly with narrow catheters for percutaneous
insertion), and it is now used instead of carbon dioxide.
However, because of its small molecular size, helium
tends to diffuse through the balloon wall more rapidly
than carbon dioxide, requiring more frequent replen-
ishing; therefore, most pumps have automatic purging
and refilling at fixed intervals.
Inflation and deflation must be synchronized with
the cardiac cycle to maximize effectiveness. Timing
markers are displayed on the ECG or on the aortic
pressure waveform on the IABP monitor to indicate pressure or vacuum loss, balloon leak or disconnect,
inflation and deflation periods. The duration of each is power loss, heart rate change, or inflation during sys-
set to be a fraction of the diastolic and systolic intervals tole (late inflation). Inflate/deflate controls allow the
(see Fig. 2). Most units feature automatic timing adjust- adjustment of timing logic according to the patient’s
ment using beat-to-beat analysis of the ECG waveform heart rate. All the listed units have built-in electrosur-
to correct for rate changes and arrhythmias. Weaning gical interference suppression (ESIS) and pacemaker-
the patient from an IABP is accomplished by gradually signal rejection.
decreasing inflation volume or pumping frequency.
IAB pumping is sometimes performed during trans-
Audible and visual alarms may occur under the port, either within the hospital or in transit (e.g.,
following conditions: loss or change of trigger stimulus, between hospitals). Many IABPs include a battery

2 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon

Systole Diastole A number of clinical problems have been reported


in the literature. Perforation of the arterial wall during
To Carotid
Arteries insertion is an obvious danger, especially when nego-
tiating circuitous vessels. Total occlusion of the aorta
Aortic can result in aortic wall trauma and damage to red
Valve
blood cells and platelets; therefore, balloons should be
selected to be nonocclusive, according to appropriate
guidelines. The balloon should be periodically inflated
Inflated and deflated, regardless of the therapy, to prevent
Balloon thrombus formation in its folds when deflated. Other
problems include kinking, insertion into false chan-
Descending nels, and plaque dislodgment. Limb and kidney is-
Aorta chemia caused by poor circulation, trauma, or catheter
Coronary obstruction have also been reported.
Arteries
IABs are fragile and must be handled with care to
help prevent leaks. Although balloon perforations are
From ECG relatively rare, they are generally the problem most
Balloon Electrodes
commonly reported to ECRI, and they can have very
Catheter
serious consequences, such as gas embolism. The first
From Pressure indication of perforation is often blood in the catheter.
Transducer
Most reported perforations have been caused by calci-
Intra-aortic
fied plaques. Studies (Cohen et al. 1995; Cox et al.
Balloon Pump 1995) have shown that using smaller IABs in shorter
Control
patients may reduce the incidence of IAB perforations
because it enables the tip of the IAB to be positioned
C289UN3A

Femoral
Arteries within the descending thoracic aorta rather than the
(Left and Right Leg) abdominal aorta, which generally has more calcified
plaque. Other sources of perforations include abra-
Figure 1. Pumping action of a single-chambered sions and punctures that occur during handling and
intra-aortic balloon; arrows indicate the direction of introduction, as well as manufacturing defects. Manu-
blood flow facturers continue to improve the quality of balloon
materials and catheters. Careful review of the instruc-
pack that allows over two hours of operation. Some tions for insertion and contraindications for use, along
manufacturers offer additional battery packs for ex- with prompt removal when blood is observed in the
tended operation. Some units have handles and casters catheter, are the most prudent approaches to prevent-
for easier mobility and securing brackets for air or ing injuries from perforated balloons. See “Intra-aortic
ambulance transport. Balloon Perforations” in ECRI’s Health Devices (cited
below) for more information.
Reported problems
Balloons of one manufacturer can be interconnected
R R
to the pump of another manufacturer only after the Arterial
Pressure
following criteria are met:
T T
• The diffusion and purity of the shuttle gas are P P
ECG
compatible with the inserted balloon.
Q Q
• The connections are leak-free. S
S

• The correct volume of shuttle gas is placed in the


C289UN3B

balloon. Balloon Deflated Inflated Deflated


State (Systole) (Diastole) (Systole)
• The inflation and deflation times are not compro-
mised by inadequate gas volumes or increased re-
sistance of interconnect components. Figure 2. Relationship among the ECG, arterial
• All alarm systems remain operative and unaffected pressure, and balloon state
by the interconnection.

©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 3
Healthcare Product Comparison System

Given that IABP patients are typically critically ill equipment versus purchasing the equipment outright.
and sometimes even balloon dependent, malfunction Because it examines the cash-flow impact of initial
of an IABP without a backup unit available can be acquisition costs and operating costs over a period of
disastrous. While IABPs are generally very reliable, time, LCC analysis is most useful for comparing alter-
they are complex devices, and malfunctions have been natives with different cash flows and for revealing the
known to occur in all brands. Therefore, ECRI recom- total costs of equipment ownership. One LCC tech-
mends that during in-hospital use, at least one backup nique — present value (PV) analysis — is especially
IABP be on hand at all times in case of malfunction. useful because it accounts for inflation and for the time
value of money (i.e., money received today is worth
more than money received at a later date). Conducting
Purchase considerations a PV/LCC analysis often demonstrates that the cost of
The IABP is not appropriate for use in every hospital. ownership includes more than just the initial acquisi-
Patients require constant, intensive, highly skilled care tion cost and that a small increase in initial acquisition
that is beyond the capability of many facilities. Compe- cost may produce significant savings in long-term op-
tency involves regular use of the technology, which is a erating costs. The PV is calculated using the annual
problem if IABPs are in small hospitals that have infre- cash outflow, the dollar discount factor (the cost of
quent opportunities to treat patients requiring an IABP. capital), and the lifetime of the equipment (in years)
Before purchasing an IABP, the hospital must assess its in a mathematical equation.
ability to effectively use and support this technology.
The following represents a sample seven-year
IABPs are life-support systems; therefore, alarms PV/LCC analysis for an IABP.
should be distinct and easily identified. Alarms asso-
ciated with critical parameters (e.g., trigger loss, vac- Present Value/Life-Cycle Cost Analysis
uum loss, trigger change, balloon disconnect, balloon
Assumptions
leak, balloon overinflation) should activate both audi-
ble and visual alarms and deflate the balloon. If the • Operating costs are considered for years 1 through 7
alarm is silenced, a visual display should clearly indi-
• Dollar discount factor is 6%
cate which alarm is disabled. It should also be possible
for the operator to silence certain alarms while leaving • The facility uses the IABP for 150 procedures/year
others active.
Capital Costs
The unit’s design should facilitate transport, includ- • Balloon pump = $50,000
ing operation and maneuverability in an elevator. The
pump should also protect against disconnection and Total Capital Costs = $50,000
catheter dislodgment during transport and should be Operating Costs
rugged enough to withstand the mechanical shocks
that it may encounter during transport. In addition, • Balloon catheters for 150 procedures =
moving the pumps should not cause undue physical $112,500/year
strain to the clinical staff. Total Operating Costs = $112,500/year
Cost containment
PV = ($780,257)
Because IABPs require the use of disposable balloon
Other costs not included in the above analysis that
catheters for each procedure, the pump’s initial acquisi-
should be considered for budgetary planning include
tion cost does not accurately reflect the total cost of
those associated with the following:
ownership. Therefore, a purchase decision should be
based on issues such as life-cycle cost (LCC), local service • Utilities
support, discount rates and non-price-related benefits
• Other disposables and accessories, including he-
offered by the supplier, and standardization with exist-
lium
ing equipment in the department or hospital (i.e., pur-
chasing all balloon catheters from one supplier). • Contributions to overhead
An LCC analysis can be used to compare high-cost As illustrated by the above sample PV/LCC analy-
alternatives and/or to determine the positive or nega- sis, the initial acquisition cost is only a fraction of the
tive economic value of a single alternative. For exam- total cost of operation over seven years. Therefore,
ple, hospitals can use LCC analysis techniques to rather than making a purchase decision based solely
examine the cost-effectiveness of leasing or renting on the acquisition cost of a balloon pump, buyers

4 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon

should consider operating costs over the lifetime of the interfaces for pulse oximeters, and at least one model
equipment. can transmit data via modem.
For further information on PV/LCC analysis, cus-
tomized analyses, and purchase decision support, Bibliography
readers should contact ECRI’s SELECT™ Group. Aksnes J, Abdelnoor M, Berge V, et al. Risk factors of
septicemia and perioperative myocardial infarction
Because many hospitals do not purchase service
in a cohort of patients supported with intra-aortic
contracts or cannot afford to have a pump out of service
balloon pump (IABP) in the course of open heart
until a service representative responds, the ease of
surgery. Eur J Cardiothorac Surg 1993;7(3):153-7.
servicing is often an important consideration. The
service and operator’s manuals should be comprehen- Armstrong B, Zidar JP, Ohman EM. The use of in-
sive, clearly written, and well organized. The pumps traaortic balloon counterpulsation in acute myocar-
should be designed to allow easy service and mainte- dial infarction and high risk coronary angioplasty.
nance. The printed circuit boards should be easily J Interv Cardiol 1995 Apr;8(2):185-91.
accessed, and it should be easy to identify defective
boards. Some suppliers offer software to help the facil- Benn A, Feldman T. The technique of inserting an
ity’s biomedical engineering staff diagnose equipment intra-aortic balloon pump. J Crit Illn 1992
problems and perform routine maintenance. Mar;7(3):435-45.

Buchanan SA, Langenburg SE, Mauney MC, et al.


Stage of development Ambulatory intraaortic balloon counterpulsation.
Many changes have occurred since IABPs were first Ann Thorac Surg 1994 Nov;58(5):1547-9.
clinically used in 1968. Improvements in IABP tech-
nology — in particular, simplification of controls and Christenson JT, Badel P, Simonet F, et al. Preopera-
greater use of automation — have eased the initiation tive intraaortic balloon pump enhances cardiac per-
and management of treatment. Manufacturers are formance and improves the outcomes of redo CABG.
making the units smaller and lighter to facilitate Ann Thorac Surg 1997 Nov;64(5):1237-44.
transport; reduce catheter diameter; and change bal-
Christenson JT, Simonet F, Badel P, et al. Evaluation
loon size and shape to improve the hemodynamic effect
of preoperative intra-aortic balloon pump support in
of pumping. Multilumen catheters with ECG, pres-
high risk coronary patients. Eur J Cardiothorac
sure, and cardiac output capabilities are more preva-
Surg 1997 Jun;11(6):1097-103.
lent. The percutaneous method of insertion is more
widely used because it allows balloon insertion to be Cohen M, Patel JJ, Dohad S, et al. Pilot prospective
performed outside the operating room (e.g., in the evaluation of counterpulsation with different intra-
catheterization lab). ESIS and timing technologies are aortic balloon volumes on cardiac performance in hu-
being improved through microprocessor algorithms mans. Cathet Cardiovasc Diagn 1995 Sep;36(1):82-7.
that can track many cardiac arrhythmias. IABPs may
become further automated by the development of mi- Cox PM Jr, Kellett M, Goran SF, et al. Plaque abrasion
croprocessor-based closed-loop feedback systems, re- and intra-aortic balloon leak. Chest 1995
quiring less direct clinical intervention. Arterial Dec;108(6):1495-8.
pressure wave triggering yielding direct timing infor-
Darovic GO. Hemodynamic monitoring: invasive and
mation is being refined to prevent interference from
noninvasive clinical application. 3rd ed. Philadel-
the inflated balloon.
phia: WB Saunders; 2001.
The applications of IABPs are also changing. For
example, technology advances have allowed their use Frederiksen JW, Smith J, Brown P, et al. Arterial
in pediatric cases. A major trend is starting IABP use helium embolism from a ruptured intraaortic bal-
earlier in the course of therapy, increasing its role in loon. Ann Thorac Surg 1988 Dec;46(6):690-2.
prevention. IABPs may also see more frequent use in Göl MK, Bayazit M, Emir M, et al. Vascular complica-
cases requiring increased cerebral blood flow. Double- tions related to percutaneous insertion of intraaor-
balloon pumping (in the pulmonary artery and aorta) tic balloon pumps. Ann Thorac Surg 1994 Nov;
is being investigated, and dual pneumatic pumps may 58(5):1476-80.
be designed for this purpose. Left-ventricular assist
devices provide a partial bypass of the left ventricle Gott JP. Intraaortic balloon pump counterpulsation.
(from the left atrium to the aorta) and can be used In: Mora CT, ed. Cardiopulmonary bypass. New
in conjunction with an IABP. Some models have York: Springer-Verlag; 1997.

©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 5
Healthcare Product Comparison System

Kantrowitz A, Cardona RR, Freed PS. Percutaneous Medical electrical equipment — part 1: general re-
intra-aortic balloon counterpulsation. Crit Care quirements for safety. Amendment 2 [standard].
Clin 1992 Oct;8(4):819-37. IEC 60601-1-am2 (1995-03). 1995.

Nishida H, Koyanagi H, Abe T, et al. Comparative Medical electrical equipment — part 1: general re-
study of five types of IABP balloons in terms of quirements for safety. Section 1. Collateral standard:
incidence of balloon rupture and other complica- safety requirements for medical electrical systems.
tions: a multi-institutional study. Artif Organs 1994 IEC 60601-1-1 (1992-06). 1992.
Oct;18(10):746-51. Medical electrical equipment — part 1: general re-
quirements for safety. Section 1. Collateral standard:
Ohman EM, George BS, White CJ, et al. Use of aortic
safety requirements for medical electrical systems.
counterpulsation to improve sustained coronary ar-
Amendment 1 [standard]. IEC 60601-1-1-am1 (1995-
tery patency during acute myocardial infarction:
11). 1995.
results of a randomized trial. Circulation 1994
Aug;90(2):792-9. Medical electrical equipment — part 1: general re-
quirements for safety. Section 2. Collateral standard:
Sakamoto T, Arai H, Maruyama T, et al. New algo- electromagnetic compatibility — requirements and
rithm of intra aortic balloon pumping in patients tests. IEC 60601-1-2 (2001-09). 2001.
with atrial fibrillation. ASAIO J 1995 Jan-
Mar;41(1):79-83.

Underwood MJ, Firmin RK, Graham TR. Current con- Citations from other ECRI publications
cepts in the use of intra-aortic balloon counterpul- Health Devices
sation. Br J Hosp Med 1993 Oct;50(7):391-7.
Intra-aortic balloon perforations [hazard]. 1989
Webster JG, ed. Encyclopedia of medical devices and
Dec;18(12):440-1.
instrumentation. Vol. 3. New York: John Wiley &
Sons; 1988:1661. Undetected pacemaker spike signals on IABP patients
[hazard]. 1989 Dec;18(12):441-2.
Standards and guidelines Low utilization of intra-aortic balloon pump [User
Note: Although every effort is made to ensure that the Experience Network™]. 1992 Sep;21(9):335.
following list is comprehensive, please note that other
applicable standards may exist. Intra-aortic balloon pumps: inspection and infrequent
use [User Experience Network™]. 1993 Feb;22(2):96.
American College of Cardiology/American Heart Asso-
ciation. ACC/AHA guidelines for the management Blood residue in Datascope System 97 IABP can cause
of patients with acute myocardial infarction. G03. malfunction [hazard]. 1997 May;26(5):219.
J Am Coll Cardiol 1999 Sep;34(3):890-911.
Intra-aortic balloon perforations [hazard update].
Guidelines for the evaluation and management of 1997 May;26(5):217-8.
heart failure. Committee on Evaluation and Man-
agement of Heart Failure. G013. Circulation 1995 Intra-aortic balloon pumps [evaluation]. 1997 May;
Nov;92(9):2764-84 (revised 2001). 26(5):184-215.
American National Standards Institute/Association Premature depletion of Datascope 97 intra-aortic bal-
for the Advancement of Medical Instrumentation. loon pump helium cylinders. 1998 Mar;27(3):117-8.
Safe current limits for electromedical apparatus
[standard]. 3rd ed. ANSI/AAMI ES1-1993. 1985 (re- Health Devices Inspection and Preventive Maintenance
vised 1993). System
International Electrotechnical Commission. Medical
Intra-aortic balloon pumps. Procedure no. 432.
electrical equipment — part 1: general requirements
for safety [standard]. IEC 60601-1 (1988-12). 1988.
Healthcare Risk Control
Medical electrical equipment — part 1: general re-
quirements for safety. Amendment 1 [standard]. Intra-aortic balloon pumps [risk analysis].
IEC 60601-1-am1 (1991-11). 1991. 1996;4:Critical care:2.

6 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon

Supplier information start of inflation to the start of deflation. The defla-


tion interval is the period from the start of deflation
Aisin Human Systems to the start of inflation.
Aisin Human Systems Co Ltd [304377] Triggering modes: Methods used for triggering infla-
2-3 Showa-cho tion of the balloon. Units can be triggered from the
Kariya-shi, Aichi Pref 448-8650 R wave of the QRS complex. Inflation can also be
Japan triggered from the arterial pressure waveform,
Phone: 81 (566) 248222 pacemaker spike (atrial, ventricular, or atrial-ven-
Fax: 81 (566) 249377 tricular sequential), external monitor input signal,
E-mail: omura@ai-aisin.co.jp or internal IABP trigger.
Internet: http://www.aisin.co.jp
Max pump rate, bpm: The maximum rate, in beats per
Arrow International minute, at which full inflation and deflation occur
for given inflation and deflation times.
Arrow International CR as [384439]
Prazska 209 Frequency weaning: A ratio that is set to gradually
CZ-500 04 Hradec Kralove decrease pumping frequency at the end of therapy.
Czech Republic Volume weaning by decreasing the balloon inflation
Phone: 420 (49) 5759111 volume is also possible.
Fax: 420 (49) 5759222 Balloon sizes, cc: The balloon volume, measured in
E-mail: jitka_vodstrcilova@arrow.cz cubic centimeters (DL — double lumen; SL — single
Internet: http://www.arrowintl.com lumen).
Arrow International Inc [101060] Compatibility: Some systems are compatible for use
2400 Bernville Rd with other manufacturers’ balloons.
Reading PA 19605
Operating time, hr: The portable operating time of the
Phone: (610) 378-0131, (800) 233-3187
unit. This depends on either the battery or the gas
Fax: (610) 478-3199
supply.
E-mail: webmaster@arrowintl.com
Internet: http://www.arrowintl.com Abbreviations:
CE mark — Conformite Europeene mark
Datascope
Datascope Corp CLA — City of Los Angeles
Cardiac Assist Div [107876] DL — Deciliter
15 Law Dr
Fairfield NJ 07004-0011 ECG — Electrocardiogram
Phone: (973) 244-6100, (800) 777-4222
ESU — Electrosurgical unit
Fax: (973) 244-6279
E-mail: monitoring-dtld@datascope.com ETL — ETL Testing Laboratories
Internet: http://www.datascope.com
FDA — U.S. Food and Drug Administration
Datascope GmbH [153264]
Fabrikstrasse 35 HR — Heart rate
D-64625 Bensheim
Germany IABP — Intra-aortic balloon pump
Phone: 49 (6251) 17050 ICU — Intensive care unit
Fax: 49 (6251) 67877
E-mail: monitoring-dtld@datascope.com IEC — International Electrotechnical Commission
Internet: http://www.datascope.com
LCD — Liquid crystal display

About the chart specifications MDD — Medical Device Directives


The following terms are used in the chart: OR — Operating room
Timing logic: The timing of inflation and deflation psi — Pounds per square inch
points in the cardiac cycle is operator adjustable.
The inflation and deflation intervals are also adjust- QRS — Wave complex representing the beginning
able. The inflation interval is the period from the of ventricular contraction

©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 7
Healthcare Product Comparison System

UL — Underwriters Laboratories features and characteristics are standard and which


are not, some may be optional, at additional cost.
Note: The data in the charts derive from suppli-
ers’ specifications and have not been verified through For those models whose prices were supplied to us
independent testing by ECRI or any other agency. in currencies other than U.S. dollars, we have also
Because test methods vary, different products’ specifi- listed the conversion to U.S. dollars to facilitate com-
cations are not always comparable. Moreover, prod- parison among models. However, keep in mind that
ucts and specifications are subject to frequent changes. exchange rates change often.
ECRI is not responsible for the quality or validity of
the information presented or for any adverse conse- Need to know more?
quences of acting on such information.
For further information about the contents of this
When reading the charts, keep in mind that, unless Product Comparison, contact the HPCS Hotline at +1
otherwise noted, the list price does not reflect supplier (610) 825-6000, ext. 5265; +1 (610) 834-1275 (fax); or
discounts. And although we try to indicate which hpcs@ecri.org (e-mail).

About ECRI . . .
ECRI is a nonprofit health services research agency and a Collaborating Center of the World Health
Organization, providing information and technical assistance to the healthcare community to support
safe and cost-effective patient care for more than 25 years. The results of ECRI’s research and
experience are available through its publications, information systems, databases, technical assis-
tance program, laboratory services, seminars, and fellowships.

Our full-time staff includes a wide range of specialists in healthcare technology, hospital admini-
stration, financial analysis, risk management, and information and computer science, as well as
hospital planners, attorneys, physicists; biomedical, electrical, electronic, chemical, mechanical, and
registered engineers; physicians; basic medical scientists; epidemiologists and biostatisticians; and
writers, editors, and communications specialists.

Underlying ECRI’s knowledge base in healthcare technology are its integrity and objectivity. ECRI
accepts no financial support from medical product manufacturers, and no employee may own stock
in or consult for a medical equipment or pharmaceutical company.

The scope of ECRI’s resources extends far beyond technology. ECRI keeps healthcare professionals,
manufacturers, legal professionals, information specialists, and others aware of the changing trends
in healthcare, healthcare standards and regulations, and the best ways to handle environmental and
occupational health and safety issues. ECRI also advises on management issues related to healthcare
cost containment, accreditation, risk management, human resources, quality of care, and other
complex topics.

ECRI has more than 35 publications, databases, software, and services to fulfill the growing need
for healthcare information and decision support. They focus on three primary areas: healthcare
technology, healthcare risk and quality management, and healthcare environmental management.

8 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon

Product Comparison Chart


MODEL AISIN HUMAN SYSTEMS ARROW INTERNATIONAL ARROW INTERNATIONAL ARROW INTERNATIONAL

Corart BP21 ACAT 1 PLUS AutoCAT H-8000 TransAct IABP

WHERE MARKETED Japan Worldwide Worldwide Worldwide

FDA CLEARANCE Not specified Yes Yes Yes


CE MARK (MDD) Not specified Yes Yes Yes
TIMING LOGIC Auto timing adjust Auto timing adjust Auto timing adjust Auto timing adjust
for rate changes and for rate changes and based on analysis of for rate changes and
arrhythmias based on arrhythmia based on ECG, AP (if arrhythmia based on
beat-beat analysis; beat-beat analysis, available), and BPW real-time beat-beat
in manual mode, auto timing on to optimize assist, analysis, inflate
inflation point set setup, predictive automatically point auto adjusts
at dicrotic notch, and real-time detects irregular HR changes (Weissler
deflation point set modes selectable rhythms, implements Formula), deflate at
before systolic by clinician; ECG specialized timing R wave to 160 msec
upstroke at R wave trigger: 20-80% R-R algorithms, before R wave, hold
if R wave occurs (inflation), 30-120% compensates for deflate to R wave in
first R-R (deflation); premature beats atrial fib (ECG mode
pressure trigger: only); pressure-
0-35% (inflation), trigger timing com-
35-75% (deflation) pensation, time set-
tings between modes
require min adjust
INFLATION VOL ADJUST
Increments, cc NA, automatic 0.5 1 1
TRIGGERING MODES R wave (pattern, R wave, QRS, R wave, pressure, R wave with pacer
peak, pacer V), pressure, atrial pacer AV, internal, spike rejection,
aortic pressure (10, fib, pacer A, pacer pacer V peaks, pacer AV,
15, 20, 25 mm Hg) * V, internal pressure, internal

MAX PUMP RATE, bpm 200 200 200 200


FREQUENCY WEANING 1:1, 1:2, 1:3, 1:4 1:1, 1:2, 1:4, 1:8 1:1, 1:2, 1:3, 1:8 1:2, 1:3, 1:4, 1:8

SIGNAL INPUTS ECG, pressure ECG, pressure ECG, pressure ECG, pressure

AUTO CONDENSATION
REMOVAL Not specified Without interruption Without interruption 2 hr purge
DISPLAY LCD LCD High-resolution Flat vacuum
electroluminescent fluorescent dot type
Type/size 640 x 480 pixels 3 trace (4 color) 3 trace 2 trace/3.6 x 4.9"
Parameters ECG; aortic blood ECG, arterial pres- ECG, arterial pres- ECG, arterial
pressure; trigger sure, balloon pres- sure, balloon pres- pressure, balloon-
inflate/deflate sure, HR, freeze sure, HR, freeze gas pressure, HR,
status waveforms; mode, balloon mode, balloon volume arterial pressure
digital values for volume, helium and set, balloon volume (digital values),
systolic, diastolic, battery reserve, pumped, helium and ECG lead, assist
augmented, and mean timing reference battery reserve, frequency, battery
blood pressure; HR; display, diagnos- timing reference gauge
battery voltage; tics, trigger high- display, diagnos-
He tank pressure; lights on ECG wave- tics, cursor for AP,
volume-weaning form, timing refer- numerical values of
status; error ence with numerical inflation and
messages; time/date settings, new help, deflation pressures,
cursor for AP & BPW AC or DC operation **
Heart rate alarm No No (trigger loss) No No (trigger loss)
QRS indicator Yes Yes Yes Yes
Pressure alarm Yes MAP/PDP MAP No (trigger loss)
Lead select Not specified 3- or 12-lead 3- or 12-lead 3-lead

Colons separate data on similar models of a device. This is the first of


* Also fixed rate (60, 90, 120 bpm). two pages covering
** Also mode of operation (operator or AutoPilot). the above model(s).
These specifications
continue onto the
next page.

©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 9
Healthcare Product Comparison System

Product Comparison Chart


MODEL AISIN HUMAN SYSTEMS ARROW INTERNATIONAL ARROW INTERNATIONAL ARROW INTERNATIONAL

Corart BP21 ACAT 1 PLUS AutoCAT H-8000 TransAct IABP

ALARMS
Trigger loss/change Yes/yes Yes/no Yes/no Yes/no
Vac/pressure loss Not specified No/yes Yes/yes Yes/yes
Balloon disconnect Yes Yes Yes Yes
Balloon leak Yes Yes Yes Yes
Balloon overinflate Yes Yes Yes Yes
Systolic inflate No Yes Yes No *
Loading failure No Yes Yes Yes
High pressure Yes Yes Yes Yes
Power loss Yes Yes Yes Yes
BALLOON SIZES, cc 32, 40 30, 40, 50 30, 40, 50 Any 20-50 volume
(percutaneous DL), (percutaneous DL), controlled in 1 cc
sheathless sheathless increments
Compatibility All suppliers Datascope Datascope Datascope

HELIUM CYLINDER SIZE


Volume, L 0.49 See footnote ** See footnote ** 95 cc
Pressure 14.7 mPa See footnote ** See footnote ** 2,150 psi
OPERATING TIME, hr 2 2/4 (1 batt/2 batt) 2 ≥2 @ 120 bpm
CHARGING TIME, hr 8 8 (90%), 4 (80%) 4 (80%), 6 (100%) 6-8
LINE POWER, VAC 100 90-264, 47-63 Hz 100/120, 220/240; 90/132, 184/264
47-63 Hz
H x W x D, cm (in) 84 x 34 x 61 71 x 30.5 x 50.8 71 x 30.5 x 50.8 61 x 48.3 x 22.9
(33.1 x 13.4 x 24) (28 x 12 x 20) (28 x 12 x 20) (24 x 19 x 9)
WEIGHT, kg (lb) 39.5 (87.1) 40.9 (90) 44.4 (98) 27.3 (60)

PURCHASE INFORMATION
Price, unit Not specified $48,975 $53,975 $45,750
Balloons Not specified $715-750 $715-750 $750
Warranty 1 year 1 year, negotiable 1 year, negotiable 1 year
Year first sold 2001 1997 2000 1993
Fiscal year April to March October to September October to September October to September
OTHER SPECIFICATIONS Built-in 5 cm wide, No external or No external or Designed to hang
2-channel strip- replaceable parts; replaceable parts; from the rails or
chart recorder; auto purge and auto auto purge and auto footboard of an ICU
automatic documenta- helium refill w/o helium refill w/o bed; 3-position cart
tion of alarm pumping interrup- pumping interrup- for OR use; auto
conditions with tion; continuously tion; continuously purge (2 beats)
waveforms, alarm displayed calibrat- optimized AP scale every 2 hr; ECG
information, and ed/scaled BPW indi- for display; contin- baseline noise
date/time stamp. cates IABP or cathe- uously displayed filter suppresses
ter malfunction; calibrated/scaled shift; diagnostic
timing-error alert BPW indicates IABP mode displays all
for >100% deflation; or catheter malfunc- operating parameters
auto print on alarm tion; auto print (15 including inflate/
(2 min to 4 hr); min to 4 hr) on deflate times;
auto weaning; alarm selectable by separate high-fill-
annotation includes user; ECG w/continu- pressure and kinked-
trigger mode, alarm ous gain; annotation line alarms;
status, assisted and includes date, time, approved for use in
unassisted values, *** mode of operation, aircraft; has added
date, and time. † 2-channel waveform, aircraft cart. †
ECG lead select. ††

Colons separate data on similar models of a device.


* Automatic timing adjustment maintains proper timing as heart rate changes.
** 33 L @ 500 psi, disposable; optional 105 L @ 2,216 psi refillable D tank, 100 L @ 2,900 psi refillable D tank.
*** Quality assurance log and electronic key recognition of volume.

Prints select 2-channel waveform, patient hemodynamics, assist ratio, balloon volume, and ECG lead. The H-8000 TransAct IABP
is CLA and ETL approved and meets requirements of IEC 601-1 and UL 544.
††
Annotation also includes trigger, alarm status, assisted and unassisted numerics, 360-degree rotatable screen (removable for
remote mounting), and real-time modem.

10 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon

Product Comparison Chart


MODEL ARROW INTERNATIONAL DATASCOPE

KAAT II PLUS System 98XT with


CardioSync2 Software *
WHERE MARKETED Worldwide Worldwide

FDA CLEARANCE Yes Yes


CE MARK (MDD) Yes Yes
TIMING LOGIC Auto timing adjust Adapts ECG defla-
for rate changes and tion, ECG timing and
arrhythmia based on triggering to pro-
beat-beat analysis, vide better augmen-
auto timing on tation, unloading,
setup, predictive and tracking of iso-
and real-time lated PVCs, bi-
modes selectable geminy, couplets,
by clinician; ECG and atrial fibrilla-
trigger: 20-80% R-R tion; pressure
(inflation), 30-120% trigger detects
R-R (deflation); sustained arrhyth-
pressure trigger: mias; adjusts and
0-35% (inflation), optimizes deflation
35-75% (deflation) and assisted beats

INFLATION VOL ADJUST


Increments, cc 0.5 NA
TRIGGERING MODES R wave, QRS, R wave w/pacer spike
pressure, atrial rejection; pressure
fib, pacer A, pacer with auto threshold
V, internal adjustment; pacer
V/A-V; pacer A **
MAX PUMP RATE, bpm 200 200
FREQUENCY WEANING 1:1, 1:2, 1:4, 1:8 1:1, 1:2, 1:3

SIGNAL INPUTS ECG, pressure ECG, arterial


pressure
AUTO CONDENSATION
REMOVAL Without interruption Fully automatic
DISPLAY LCD Electroluminescent

Type/size 3 trace (4 color) 8.3 x 6.2"


Parameters ECG, arterial pres- ECG, arterial pres-
sure, balloon pres- sure (values
sure, HR, freeze for syst, diast,
mode, balloon mean, augmented),
volume, helium and dR, HR trend, ECG
battery reserve, lead, ECG size,
timing reference pressure scale and
display, diagnos- trigger threshold,
tics, trigger high- balloon dynamics,
lights on ECG augmentation alarm
waveform set, opt balloon
pressure waveform,
trigger source,
timing and fill mode
Heart rate alarm No (trigger loss) Yes
QRS indicator Yes Yes
Pressure alarm No (trigger loss) Yes
Lead select 3-lead 12-lead

Colons separate data on similar models of a device. This is the first of


* Model listed is currently marketed; specifications current as of June 2001. two pages covering
** Variable internal rate. the above model(s).
These specifications
continue onto the
next page.

©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 11
Healthcare Product Comparison System

Product Comparison Chart


MODEL ARROW INTERNATIONAL DATASCOPE

KAAT II PLUS System 98XT with


CardioSync2 Software *
ALARMS
Trigger loss/change Yes/no Yes/yes (standby)
Vac/pressure loss No/yes Yes/no (aug alarm)
Balloon disconnect Yes Yes
Balloon leak Yes Yes
Balloon overinflate Yes NA, autofill control
Systolic inflate Yes NA, auto control
Loading failure Yes Autofill failure
High pressure Yes Yes
Power loss Yes Yes
BALLOON SIZES, cc 30, 40, 50 34, 40 (8 Fr);
(percutaneous DL), 25, 34, 40 (9.5 Fr);
sheathless 40, 50 (10.5 Fr) **
Compatibility Datascope Arrow International,
Xemex, TMP

HELIUM CYLINDER SIZE


Volume, L See footnote *** 0.5 L; 0.69 L
Pressure See footnote *** 2,900; 2,200 psi
OPERATING TIME, hr 3 2.25
CHARGING TIME, hr 8 (90%), 4 (80%) 18
LINE POWER, VAC 90/132, 180/250 100-120, 220-240

H x W x D, cm (in) 88.9 x 30.5 x 63.5 109 x 42.7 x 56.6


(35 x 12 x 25) (43.3 x 16.8 x 22.3)
WEIGHT, kg (lb) 56.1 (124), 62.3 27.7 (61), cart;
(137) with optional 15.4 (34), battery;
D tank 39.1 (86.3), console
PURCHASE INFORMATION
Price, unit $48,975 $53,790
Balloons $715-750 $745-895
Warranty 1 year 1 year
Year first sold 1992 2000
Fiscal year October to September July to June
OTHER SPECIFICATIONS No external or 1-button start-up;
replaceable parts; print strip mode;
auto purge and auto manual fill mode; 2
helium refill w/o hr auto purge; ESU
pumping interrup- noise suppression w/
tion; continuously ECG amplifier; blood
displayed calibrat- detect sensor;
ed/scaled BPW indi- defibrillator over-
cates IABP or cathe- load protection; RS-
ter malfunction; 232 connector; diag-
timing-error alert nostic output;
for >100% deflation; service diagnostics
auto print on alarm software; universal
(2 min to 4 hr); transport system;
annotation includes R-Trac automatically
trigger mode, alarm matches the most
status, date, and appropriate
time; flash card for deflation timing
user-defined setup method to the
and storage, modem patient's rhythm.
for remote communi- †
cations. ††

Colons separate data on similar models of a device.


* Model listed is currently marketed; specifications current as of June 2001.
** 8 Fr is co-lumen, 9.5 Fr is dual lumen. Pediatric sizes: 2.5, 5, 7, 12, and 20 cc.
*** 33 L @ 500 psi, disposable; optional 105 L @ 2,216 psi refillable D tank, 100 L @ 2,900 psi refillable D tank.

Quality assurance log and electronic key recognition of volume.
††
Prints select 2-channel waveform, patient hemodynamics, assist ratio, balloon volume, and ECG lead. The H-8000 TransAct IABP
is CLA and ETL approved and meets requirements of IEC 601-1 and UL 544.

12 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.

Vous aimerez peut-être aussi