Académique Documents
Professionnel Documents
Culture Documents
2 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon
Femoral
Arteries within the descending thoracic aorta rather than the
(Left and Right Leg) abdominal aorta, which generally has more calcified
plaque. Other sources of perforations include abra-
Figure 1. Pumping action of a single-chambered sions and punctures that occur during handling and
intra-aortic balloon; arrows indicate the direction of introduction, as well as manufacturing defects. Manu-
blood flow facturers continue to improve the quality of balloon
materials and catheters. Careful review of the instruc-
pack that allows over two hours of operation. Some tions for insertion and contraindications for use, along
manufacturers offer additional battery packs for ex- with prompt removal when blood is observed in the
tended operation. Some units have handles and casters catheter, are the most prudent approaches to prevent-
for easier mobility and securing brackets for air or ing injuries from perforated balloons. See “Intra-aortic
ambulance transport. Balloon Perforations” in ECRI’s Health Devices (cited
below) for more information.
Reported problems
Balloons of one manufacturer can be interconnected
R R
to the pump of another manufacturer only after the Arterial
Pressure
following criteria are met:
T T
• The diffusion and purity of the shuttle gas are P P
ECG
compatible with the inserted balloon.
Q Q
• The connections are leak-free. S
S
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 3
Healthcare Product Comparison System
Given that IABP patients are typically critically ill equipment versus purchasing the equipment outright.
and sometimes even balloon dependent, malfunction Because it examines the cash-flow impact of initial
of an IABP without a backup unit available can be acquisition costs and operating costs over a period of
disastrous. While IABPs are generally very reliable, time, LCC analysis is most useful for comparing alter-
they are complex devices, and malfunctions have been natives with different cash flows and for revealing the
known to occur in all brands. Therefore, ECRI recom- total costs of equipment ownership. One LCC tech-
mends that during in-hospital use, at least one backup nique — present value (PV) analysis — is especially
IABP be on hand at all times in case of malfunction. useful because it accounts for inflation and for the time
value of money (i.e., money received today is worth
more than money received at a later date). Conducting
Purchase considerations a PV/LCC analysis often demonstrates that the cost of
The IABP is not appropriate for use in every hospital. ownership includes more than just the initial acquisi-
Patients require constant, intensive, highly skilled care tion cost and that a small increase in initial acquisition
that is beyond the capability of many facilities. Compe- cost may produce significant savings in long-term op-
tency involves regular use of the technology, which is a erating costs. The PV is calculated using the annual
problem if IABPs are in small hospitals that have infre- cash outflow, the dollar discount factor (the cost of
quent opportunities to treat patients requiring an IABP. capital), and the lifetime of the equipment (in years)
Before purchasing an IABP, the hospital must assess its in a mathematical equation.
ability to effectively use and support this technology.
The following represents a sample seven-year
IABPs are life-support systems; therefore, alarms PV/LCC analysis for an IABP.
should be distinct and easily identified. Alarms asso-
ciated with critical parameters (e.g., trigger loss, vac- Present Value/Life-Cycle Cost Analysis
uum loss, trigger change, balloon disconnect, balloon
Assumptions
leak, balloon overinflation) should activate both audi-
ble and visual alarms and deflate the balloon. If the • Operating costs are considered for years 1 through 7
alarm is silenced, a visual display should clearly indi-
• Dollar discount factor is 6%
cate which alarm is disabled. It should also be possible
for the operator to silence certain alarms while leaving • The facility uses the IABP for 150 procedures/year
others active.
Capital Costs
The unit’s design should facilitate transport, includ- • Balloon pump = $50,000
ing operation and maneuverability in an elevator. The
pump should also protect against disconnection and Total Capital Costs = $50,000
catheter dislodgment during transport and should be Operating Costs
rugged enough to withstand the mechanical shocks
that it may encounter during transport. In addition, • Balloon catheters for 150 procedures =
moving the pumps should not cause undue physical $112,500/year
strain to the clinical staff. Total Operating Costs = $112,500/year
Cost containment
PV = ($780,257)
Because IABPs require the use of disposable balloon
Other costs not included in the above analysis that
catheters for each procedure, the pump’s initial acquisi-
should be considered for budgetary planning include
tion cost does not accurately reflect the total cost of
those associated with the following:
ownership. Therefore, a purchase decision should be
based on issues such as life-cycle cost (LCC), local service • Utilities
support, discount rates and non-price-related benefits
• Other disposables and accessories, including he-
offered by the supplier, and standardization with exist-
lium
ing equipment in the department or hospital (i.e., pur-
chasing all balloon catheters from one supplier). • Contributions to overhead
An LCC analysis can be used to compare high-cost As illustrated by the above sample PV/LCC analy-
alternatives and/or to determine the positive or nega- sis, the initial acquisition cost is only a fraction of the
tive economic value of a single alternative. For exam- total cost of operation over seven years. Therefore,
ple, hospitals can use LCC analysis techniques to rather than making a purchase decision based solely
examine the cost-effectiveness of leasing or renting on the acquisition cost of a balloon pump, buyers
4 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon
should consider operating costs over the lifetime of the interfaces for pulse oximeters, and at least one model
equipment. can transmit data via modem.
For further information on PV/LCC analysis, cus-
tomized analyses, and purchase decision support, Bibliography
readers should contact ECRI’s SELECT™ Group. Aksnes J, Abdelnoor M, Berge V, et al. Risk factors of
septicemia and perioperative myocardial infarction
Because many hospitals do not purchase service
in a cohort of patients supported with intra-aortic
contracts or cannot afford to have a pump out of service
balloon pump (IABP) in the course of open heart
until a service representative responds, the ease of
surgery. Eur J Cardiothorac Surg 1993;7(3):153-7.
servicing is often an important consideration. The
service and operator’s manuals should be comprehen- Armstrong B, Zidar JP, Ohman EM. The use of in-
sive, clearly written, and well organized. The pumps traaortic balloon counterpulsation in acute myocar-
should be designed to allow easy service and mainte- dial infarction and high risk coronary angioplasty.
nance. The printed circuit boards should be easily J Interv Cardiol 1995 Apr;8(2):185-91.
accessed, and it should be easy to identify defective
boards. Some suppliers offer software to help the facil- Benn A, Feldman T. The technique of inserting an
ity’s biomedical engineering staff diagnose equipment intra-aortic balloon pump. J Crit Illn 1992
problems and perform routine maintenance. Mar;7(3):435-45.
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 5
Healthcare Product Comparison System
Kantrowitz A, Cardona RR, Freed PS. Percutaneous Medical electrical equipment — part 1: general re-
intra-aortic balloon counterpulsation. Crit Care quirements for safety. Amendment 2 [standard].
Clin 1992 Oct;8(4):819-37. IEC 60601-1-am2 (1995-03). 1995.
Nishida H, Koyanagi H, Abe T, et al. Comparative Medical electrical equipment — part 1: general re-
study of five types of IABP balloons in terms of quirements for safety. Section 1. Collateral standard:
incidence of balloon rupture and other complica- safety requirements for medical electrical systems.
tions: a multi-institutional study. Artif Organs 1994 IEC 60601-1-1 (1992-06). 1992.
Oct;18(10):746-51. Medical electrical equipment — part 1: general re-
quirements for safety. Section 1. Collateral standard:
Ohman EM, George BS, White CJ, et al. Use of aortic
safety requirements for medical electrical systems.
counterpulsation to improve sustained coronary ar-
Amendment 1 [standard]. IEC 60601-1-1-am1 (1995-
tery patency during acute myocardial infarction:
11). 1995.
results of a randomized trial. Circulation 1994
Aug;90(2):792-9. Medical electrical equipment — part 1: general re-
quirements for safety. Section 2. Collateral standard:
Sakamoto T, Arai H, Maruyama T, et al. New algo- electromagnetic compatibility — requirements and
rithm of intra aortic balloon pumping in patients tests. IEC 60601-1-2 (2001-09). 2001.
with atrial fibrillation. ASAIO J 1995 Jan-
Mar;41(1):79-83.
Underwood MJ, Firmin RK, Graham TR. Current con- Citations from other ECRI publications
cepts in the use of intra-aortic balloon counterpul- Health Devices
sation. Br J Hosp Med 1993 Oct;50(7):391-7.
Intra-aortic balloon perforations [hazard]. 1989
Webster JG, ed. Encyclopedia of medical devices and
Dec;18(12):440-1.
instrumentation. Vol. 3. New York: John Wiley &
Sons; 1988:1661. Undetected pacemaker spike signals on IABP patients
[hazard]. 1989 Dec;18(12):441-2.
Standards and guidelines Low utilization of intra-aortic balloon pump [User
Note: Although every effort is made to ensure that the Experience Network™]. 1992 Sep;21(9):335.
following list is comprehensive, please note that other
applicable standards may exist. Intra-aortic balloon pumps: inspection and infrequent
use [User Experience Network™]. 1993 Feb;22(2):96.
American College of Cardiology/American Heart Asso-
ciation. ACC/AHA guidelines for the management Blood residue in Datascope System 97 IABP can cause
of patients with acute myocardial infarction. G03. malfunction [hazard]. 1997 May;26(5):219.
J Am Coll Cardiol 1999 Sep;34(3):890-911.
Intra-aortic balloon perforations [hazard update].
Guidelines for the evaluation and management of 1997 May;26(5):217-8.
heart failure. Committee on Evaluation and Man-
agement of Heart Failure. G013. Circulation 1995 Intra-aortic balloon pumps [evaluation]. 1997 May;
Nov;92(9):2764-84 (revised 2001). 26(5):184-215.
American National Standards Institute/Association Premature depletion of Datascope 97 intra-aortic bal-
for the Advancement of Medical Instrumentation. loon pump helium cylinders. 1998 Mar;27(3):117-8.
Safe current limits for electromedical apparatus
[standard]. 3rd ed. ANSI/AAMI ES1-1993. 1985 (re- Health Devices Inspection and Preventive Maintenance
vised 1993). System
International Electrotechnical Commission. Medical
Intra-aortic balloon pumps. Procedure no. 432.
electrical equipment — part 1: general requirements
for safety [standard]. IEC 60601-1 (1988-12). 1988.
Healthcare Risk Control
Medical electrical equipment — part 1: general re-
quirements for safety. Amendment 1 [standard]. Intra-aortic balloon pumps [risk analysis].
IEC 60601-1-am1 (1991-11). 1991. 1996;4:Critical care:2.
6 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 7
Healthcare Product Comparison System
About ECRI . . .
ECRI is a nonprofit health services research agency and a Collaborating Center of the World Health
Organization, providing information and technical assistance to the healthcare community to support
safe and cost-effective patient care for more than 25 years. The results of ECRI’s research and
experience are available through its publications, information systems, databases, technical assis-
tance program, laboratory services, seminars, and fellowships.
Our full-time staff includes a wide range of specialists in healthcare technology, hospital admini-
stration, financial analysis, risk management, and information and computer science, as well as
hospital planners, attorneys, physicists; biomedical, electrical, electronic, chemical, mechanical, and
registered engineers; physicians; basic medical scientists; epidemiologists and biostatisticians; and
writers, editors, and communications specialists.
Underlying ECRI’s knowledge base in healthcare technology are its integrity and objectivity. ECRI
accepts no financial support from medical product manufacturers, and no employee may own stock
in or consult for a medical equipment or pharmaceutical company.
The scope of ECRI’s resources extends far beyond technology. ECRI keeps healthcare professionals,
manufacturers, legal professionals, information specialists, and others aware of the changing trends
in healthcare, healthcare standards and regulations, and the best ways to handle environmental and
occupational health and safety issues. ECRI also advises on management issues related to healthcare
cost containment, accreditation, risk management, human resources, quality of care, and other
complex topics.
ECRI has more than 35 publications, databases, software, and services to fulfill the growing need
for healthcare information and decision support. They focus on three primary areas: healthcare
technology, healthcare risk and quality management, and healthcare environmental management.
8 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon
SIGNAL INPUTS ECG, pressure ECG, pressure ECG, pressure ECG, pressure
AUTO CONDENSATION
REMOVAL Not specified Without interruption Without interruption 2 hr purge
DISPLAY LCD LCD High-resolution Flat vacuum
electroluminescent fluorescent dot type
Type/size 640 x 480 pixels 3 trace (4 color) 3 trace 2 trace/3.6 x 4.9"
Parameters ECG; aortic blood ECG, arterial pres- ECG, arterial pres- ECG, arterial
pressure; trigger sure, balloon pres- sure, balloon pres- pressure, balloon-
inflate/deflate sure, HR, freeze sure, HR, freeze gas pressure, HR,
status waveforms; mode, balloon mode, balloon volume arterial pressure
digital values for volume, helium and set, balloon volume (digital values),
systolic, diastolic, battery reserve, pumped, helium and ECG lead, assist
augmented, and mean timing reference battery reserve, frequency, battery
blood pressure; HR; display, diagnos- timing reference gauge
battery voltage; tics, trigger high- display, diagnos-
He tank pressure; lights on ECG wave- tics, cursor for AP,
volume-weaning form, timing refer- numerical values of
status; error ence with numerical inflation and
messages; time/date settings, new help, deflation pressures,
cursor for AP & BPW AC or DC operation **
Heart rate alarm No No (trigger loss) No No (trigger loss)
QRS indicator Yes Yes Yes Yes
Pressure alarm Yes MAP/PDP MAP No (trigger loss)
Lead select Not specified 3- or 12-lead 3- or 12-lead 3-lead
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 9
Healthcare Product Comparison System
ALARMS
Trigger loss/change Yes/yes Yes/no Yes/no Yes/no
Vac/pressure loss Not specified No/yes Yes/yes Yes/yes
Balloon disconnect Yes Yes Yes Yes
Balloon leak Yes Yes Yes Yes
Balloon overinflate Yes Yes Yes Yes
Systolic inflate No Yes Yes No *
Loading failure No Yes Yes Yes
High pressure Yes Yes Yes Yes
Power loss Yes Yes Yes Yes
BALLOON SIZES, cc 32, 40 30, 40, 50 30, 40, 50 Any 20-50 volume
(percutaneous DL), (percutaneous DL), controlled in 1 cc
sheathless sheathless increments
Compatibility All suppliers Datascope Datascope Datascope
PURCHASE INFORMATION
Price, unit Not specified $48,975 $53,975 $45,750
Balloons Not specified $715-750 $715-750 $750
Warranty 1 year 1 year, negotiable 1 year, negotiable 1 year
Year first sold 2001 1997 2000 1993
Fiscal year April to March October to September October to September October to September
OTHER SPECIFICATIONS Built-in 5 cm wide, No external or No external or Designed to hang
2-channel strip- replaceable parts; replaceable parts; from the rails or
chart recorder; auto purge and auto auto purge and auto footboard of an ICU
automatic documenta- helium refill w/o helium refill w/o bed; 3-position cart
tion of alarm pumping interrup- pumping interrup- for OR use; auto
conditions with tion; continuously tion; continuously purge (2 beats)
waveforms, alarm displayed calibrat- optimized AP scale every 2 hr; ECG
information, and ed/scaled BPW indi- for display; contin- baseline noise
date/time stamp. cates IABP or cathe- uously displayed filter suppresses
ter malfunction; calibrated/scaled shift; diagnostic
timing-error alert BPW indicates IABP mode displays all
for >100% deflation; or catheter malfunc- operating parameters
auto print on alarm tion; auto print (15 including inflate/
(2 min to 4 hr); min to 4 hr) on deflate times;
auto weaning; alarm selectable by separate high-fill-
annotation includes user; ECG w/continu- pressure and kinked-
trigger mode, alarm ous gain; annotation line alarms;
status, assisted and includes date, time, approved for use in
unassisted values, *** mode of operation, aircraft; has added
date, and time. † 2-channel waveform, aircraft cart. †
ECG lead select. ††
10 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Circulatory Assist Units, Intra-Aortic Balloon
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 11
Healthcare Product Comparison System
12 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.