Kinetic™ QRS ECG Algorithm


The Kinetic™ QRS ECG Algorithm is designed to be embedded into a
device or software suite to provide precise identification of the QRS
complex and provide RR interval measurements. Any device that
requires HR, RR variability, triggers off of the R peak, or uses any
component of the QRS complex can take advantage of the precision
and accuracy of Kinetic™ QRS.

In any algorithm, QRS detection is the most difficult and most important
task. Kinetic™ QRS excels at identifying the QRS complex based on
its ability to recognize its intrinsic properties. It does not matter if the
amplitude is low or high, if it is lower or higher than the following T
wave, or if there is noise present, Kinetic™ QRS correctly identifies it.

Below are the results of the Kinetic™ QRS ECG Algorithm when tested
against the AHA and MIT databases. This data was used in Monebo's
FDA 510K submission.

Summary Results of AHA and MIT Testing


Database QRS Se QRS +P VEB Se VEB +P
AHA 99.56 99.90 82.49 95.65
MIH-BIH 99.45 99.45 87.03 87.76
NST 91.56 85.66 81.79 53.19

AHA - The American Heart Association Database for Evaluation of Ventricular Arrhythmia Detectors
MIT-BIH - The Massachusetts Institute of Technology-Beth Israel Arrhythmia Database
NST - The Noise Stress Database
Se - Sensitivity: True Positive/(True Positive + False Negative)
+P - Positive Predictivity: True Positive/(True Positive + False Positive)
VEB - Ventricular Ectopic Beat