Monthly Maintenance
PEP to Go Effective Date: 10/29/2018 | HOOD05162002956766
Unrestricted © Siemens Healthcare GmbH, 2018
Atellica® IM Analyzer
Monthly Maintenance
Monthly Maintenance
Clean the Exterior of the Aspirate Probes
Clean the IM Fan Filters
2
Effective Date: 10/29/2018 | HOOD05162002956766
Unrestricted © Siemens Healthcare GmbH, 2018
Atellica® IM Analyzer
Monthly Maintenance
Cleaning the Exterior of the Aspirate Probes
A
1. Unlock the IM analyzer front cover: B
A. On the Command bar, select System
C
B. Select the Status tab
C. Select the IM analyzer
D. Select to pause or stop the analyzer
E. Select Front Cover to unlock the front cover
2. Lift open the IM front cover D
E
* This procedure is continued on the next slide.
3
Effective Date: 10/29/2018 | HOOD05162002956766
Unrestricted © Siemens Healthcare GmbH, 2018
Atellica® IM Analyzer
Monthly Maintenance
Cleaning the Exterior of the Aspirate Probes (continued)
3. Carefully lift the aspirate probe arms to the highest position
Note: Do not remove the aspirate probes. Do not touch the probe tip
when working with the probes. Do not bend the probe.
1. Grasp the left side of the fan filter and pull it out of the slot
4. Tap the filter to remove excess water and let it air dry
Note: The customer parts kit includes spare fan filters. Use a spare filter
while the cleaned filter dries. Once dry, place the cleaned filter in the
customer parts kit for future use.
5. Reinstall a spare fan filter with the plastic grid facing away from the
panel
5
Effective Date: 10/29/2018 | HOOD05162002956766
Unrestricted © Siemens Healthcare GmbH, 2018
Please note that the learning material is for training purposes only!
For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers.
This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not
necessarily reflect the latest version of the software and hardware available at the time of the training.
The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions.
Note: Some functions shown in this material are optional and might not be part of your system.
Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the
future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information.
The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages.
All names and data of patients, parameters and configuration dependent designations are fictional and examples only.
All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners.
Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability.
6
Effective Date: 10/29/2018 | HOOD05162002956766
Unrestricted © Siemens Healthcare GmbH, 2018