Académique Documents
Professionnel Documents
Culture Documents
The manuscript was received on 7 May 2009 and was accepted after revision for publication on 5 October 2009.
DOI: 10.1243/09544119JEIM656
Abstract: Ultrasound is a widely used modality for both therapy and diagnosis in medicine
and biology. Currently, in the field of medical diagnosis, ultrasound is responsible for about
one in five of all diagnostic images. The physical characteristics of medical ultrasound, along
with its behaviour as it interacts with biological tissues, are described in this tutorial. The role of
ultrasound in therapeutic and diagnostic applications is briefly described. In view of the
importance of ultrasound as a medical imaging modality, the basic technological building
blocks utilized in diagnostic ultrasound scanners are also described. Many of these topics are
the subjects of other papers in this special issue where they are dealt with in more detail.
2
L2 DP 2 L DP L2 DP L2 DP
~c z z ð2Þ
Lt 2 Lx2 Ly 2 Lz2
In Fig. 2, the period, T, of the wave is the time
interval between sequential occurrences of the same
where DP is the change in local pressure in the
phase and amplitude of the varying pressure. The
medium and c, the speed of the wave, is given by
frequency of the wave, f, is the reciprocal of the
period
k
c~H ð3Þ
r
f ~1=T ð4Þ
where k is the bulk modulus of the tissue and r is the
density. For a wave travelling at speed c, the wavelength, l, is
given by
The bulk modulus (k) of a substance describes
behaviour under uniform compression. It is defined
l~c=f ð5Þ
as the pressure increase needed to cause a given
relative decrease in volume. The units are Pa.
The relationship can be expressed as
The density (r) of a material is its mass per unit
volume. The units are kg per cubic metre. c~lf ð6Þ
It is worth noting that, generally, the bulk modulus
will be a function of DP, and in turn DP is a function Note that c, the speed of sound, is dependent on the
of x, y, z, and t. Consequently, in real tissues, the mechanical characteristics of the material (Fig. 3).
speed of sound will not be absolutely constant For low-amplitude waves, the speed of sound is
during the passage of the ultrasound wave. constant for a particular material, with a small
dependence on temperature and frequency. How-
ever, the mechanical characteristics of the material
3.1 Period, frequency, and wavelength
also depend on the local pressure; consequently,
A simple depiction of medical ultrasound is as a high-amplitude waves can cause local changes in
sinusoidal variation of pressure with time. This is sound speed leading to non-linear propagation. This
represented graphically in Fig. 2. phenomenon is described later.
vp ~l=T ð7Þ
N Bone: in bony structures vibrations in all three reflections. Whenever a change in acoustic impe-
directions are supported and both longitudinal dance occurs, reflection will take place. Acoustic
and transverse (shear) waves propagate. impedance, Z, is of two types: specific or character-
N Soft tissue: in soft tissues, and liquid components istic acoustic impedance and acoustic impedance.
of the body, the coupling forces between particles The difference is that specific (characteristic) im-
in the direction of energy flow are large and pedance refers to the impedance at a point whereas
longitudinal waves with velocities of thousands of acoustic impedance refers to impedance at a surface.
metres per second are supported. The coupling
Specific acoustic impedance is the complex ratio of
forces perpendicular to this direction are weak;
the effective sound pressure at a point to the
consequently transverse particle movements are
effective particle velocity at that point. Impedance
rapidly attenuated. Shear waves are also sup-
can be expressed in either its constituent units
ported in soft tissues. However, they have a
(pressure per velocity per area) or in rayls. The rayl
relatively low velocity, in the order of hundreds
equals 1 pascal-second per metre (Pa s/m), or
of metres per second and high attenuation
equivalently, 1 Newton-second per cubic metre
coefficients, which render them unsuitable for
(N s/m3). In SI base units, that is kg/s m2
imaging purposes. Recent developments in elas-
tography make use of shear-wave propagation.
Z~p=v~I v2 ~p2 I ð9Þ
N Mode conversion: in the situation of a longitudinal
wave incident at a non-normal angle to a
where p is the sound pressure (N/m2 or Pa), v is the
boundary, some of the energy of longitudinal
particle velocity (m/s), and I is the sound intensity
particle movement can cause particle movement
(W/m2).
in the transverse direction. This process is known
as mode conversion. Acoustic impedance is the total reaction of a
medium to acoustic transmission through it, repre-
4.2 Reflection sented by the complex ratio of the pressure to the
effective flux, that is particle velocity multiplied by
At a discontinuity of the mechanical characteristics surface area, through the medium. In other words,
in the path of the ultrasound wave, reflection of the sound pressure p divided by the particle velocity
energy occurs. In the simple example illustrated in v and the surface area S, through which the acoustic
Fig. 5, this discontinuity is represented by a change wave propagates. For example, in plane, single-
in the mass of the particles along the path and a frequency travelling waves the acoustic impedance
change in intra-particular forces. For a wave incident is equal to the specific (characteristic) impedance
from the left, say, reflection can be thought of in divided by the surface area, S
purely mechanical terms as the smaller particle
‘bouncing off’ the more massive particle – in much Z~p=vS ð10Þ
the same way that a ping-pong ball would bounce if
run against a snooker ball. Sometimes vS is referred to as the volume velocity.
Acoustic impedance is used to describe the me- For practical purposes, acoustic impedance, Z, for
chanical properties of tissues relevant to such any medium is calculated as the square root of the
product of the density, r, and the bulk modulus, k The practical significance of these reflectivities in
pffiffiffiffiffiffi diagnosis is that ultrasound is limited to use
Z~ rk ð11Þ primarily in the assessment of soft tissues. Bone
presents a significant barrier and air is essentially
In the simple tissue model, the density relates impenetrable.
approximately to the mass of the particles and the Coupling/Matching layer: the acoustic impedance
bulk modulus to the forces between them. of transducer material is significantly greater than
N small masses, weak springs 5 low acoustic im- that of soft tissue. To reduce reflection, and conse-
pedance; quently improve transmission, at the probe/skin
N large masses, strong springs 5 high acoustic im- boundary, matching layers of solid materials of
intermediate acoustic impedance are incorporated
pedance.
pffiffiffiffiffiffiffiffi over the face of the transducer. This is an issue for
Using equation (3) (c 5 k=r), the computation equipment manufacturers. A related impedance
most often used to calculate Z is the product of the issue for the end users is the presence of air between
density and the speed of sound the probe face and the patient’s skin. The acoustic
impedance mismatch here is overcome by the use of
Z~rc ð12Þ a coupling gel with an impedance very close to that
of skin. A by-product of using gel as a coupling agent
Typical acoustic impedances for biological materi- is that the gel also acts as a lubricant. This facilitates
als, in rayls, are: easy manipulation of the probe over the skin surface.
sin hi c1
~ ð15Þ
sin ht c2
where I is the intensity, x is the distance, and a is the are operated at their resonance; consequently, their
attenuation coefficient. thickness is determined. For fundamental reso-
The description of attenuation is simplified with nance, the thickness must equal one half-wave-
the introduction of a logarithmic representation of length, remembering that it is the wavelength in the
the intensity variation. Taking logarithms of equa- transducer material. In a commonly used piezo-
tion (16) electric material, lead zirconate titanate (PZT), the
speed of sound is approximately 3300 m/s. The wave-
logðIx =I0 Þ~{ax ð17Þ length of sound in PZT at a frequency of 5 MHz is
approximately 0.7 mm. Consequently, 5 MHz reso-
The logarithm of the ratio of intensities is called the nant elements would have to be 0.35 mm thick. The
Bel. In view of the magnitudes involved in medical other dimensions, height, width, and shape, depend
applications, the more frequent notation is the upon the application. Typically, they can be single flat
decibel (dB). The usage is the term ‘intensity level’ or bowl-shaped elements of circular or rectangular
where shape up to 100 wavelengths in dimension or they can
be in the form of rectangular arrays of side-by-side
intensity level~10 log10 I1 =I0 dB ð18Þ elements about 0.2 wavelengths wide and 5–10
wavelengths long, with either flat or convex front
For example, in a situation where the intensity at the faces. Whatever its particular construction, the trans-
location of interest (I1) is 100 times less, than the ducer arrangement and its housing is usually referred
initial intensity I0, the intensity level would be to as the ‘probe’.
described as being ‘minus 20 dB’.
The attenuation coefficient, a, is frequency de-
pendent and in the order of 0.5 to 0.7 dB per 4.6 Continuous wave ultrasound
centimetre per MHz. The loss of signal strength A continuous wave is an acoustic wave of constant
due to attenuation is a limiting factor for diagnostic amplitude and frequency and, strictly speaking, of
purposes and typically restricts the depth of imaging infinite duration. In practice, ultrasound is consid-
to about 15 cm at 5 MHz. ered to be a continuous wave if is transmitted without
interruption while the equipment is functioning. Its
duration is never infinite but is long enough for its
4.5 Wave production and detection amplitude and frequency to be considered constant.
Acoustic oscillations can be generated in many ways. Continuous wave ultrasound propagating in a
Simple sound-wave generators include plucked medium of low attenuation between parallel reflect-
guitar strings and beaten drum skins. In both of ing surfaces can create a standing wave field. The
these examples, the mechanical vibrations of the standing wave field is one in which there are
surfaces cause the air molecules with which they are stationary regions of high and low pressure. Stand-
in contact to vibrate in sympathy. As it is a higher ing wave fields have been shown to cause blood flow
frequency, ultrasound is usually generated using stasis in chick embryos in vivo. They are also used as
mechanical vibrations which result from the elec- a tool for contactless separation, concentration and
trical excitation of a transducer. manipulation of microparticles and biological cells.
These devices convert or ‘transduce’ the electrical The basis is the acoustic radiation force, a non-linear
exciting signal into a mechanical movement via a effect which causes particles to be attracted to either
variety of mechanisms. The inverse piezoelectric the nodes or antinodes of the standing wave. In
effect is the most frequently employed but capacitive therapy, continuous wave ultrasound is used in
and solid-state devices also exist. applications such as physiotherapy and surgery.
Ultrasound detection is often undertaken by the Diagnostically, continuous wave ultrasound is used
same transducer, now operated in the inverse mode solely in Doppler applications.
(for example, utilizing the direct piezoelectric effect).
Changes in local pressure around the transducer (the
4.7 Pulsed ultrasound
mechanical disturbance) are converted to electrical
signals. A pulse wave is a non-sinusoidal waveform. The
The thickness of the transducer governs its shape of the wave is determined by the character-
resonant frequency. This is the frequency of max- istics of the exciting electrical pulse and the proper-
imum efficiency and sensitivity. Usually, transducers ties of the transducer. Typically, pulsed ultrasound is
generated as a sequence of regularly occurring Manufacturers make use of the increased sensi-
pulses. The pulse duration and the time interval tivity and bandwidth of transducers by labelling
between pulses depend on the application. them as having a range of usable frequencies equal
In the frequency domain, the primary character- to the upper and lower limits of the bandwidth; for
istics of an ultrasound pulse are its centre frequency example, a typical diagnostic probe for abdominal
and its bandwidth. For typical excitation with a use may be labelled as a ‘3 to 7 MHz probe’. This
short, high-amplitude voltage spike, the centre means that the ultrasound pulse has a centre
frequency of the pulse is the mechanical resonant frequency of 5 MHz but a bandwidth sufficient for
frequency of the transducer elements. The resonant images to be created at filtered, received frequencies
frequency depends on the thickness of the elements from 3 to 7 MHz.
and cannot be altered for a particular transducer.
Changing the centre frequency of the pulse requires
4.7.1 Beam shape
changing the thickness of the elements; in other
words, it requires the probe to be changed. Typical beam shapes and characteristics of ultra-
The bandwidth of the pulse is a measure of the sound emitted from the transducers used in medical
range of frequencies it contains. Usually, the width is and biological systems are detailed in Fig. 11.
measured between the 23 dB points and is referred
to as the full-width, half maximum (FWHM) value
(Fig. 10).
The upper and lower frequencies are taken as the
usable limits of the pulse. Above and below these
frequencies, the signal amplitude is less than half of
its maximum value and is considered to be insuffi-
cient for most purposes. Similar considerations
apply to the overall sensitivity characteristic of the
transducer. Its response is taken to be the square of
that of the pulse it launches.
Fig. 17 Block diagram of a typical real-time scanner. The pulser or clock creates a regular
sequence of electrical signals which are processed by the beamformer to be sent to the
appropriate group of elements in the probe. The pulse created by the electrical pulse
travels into the patient and a train of echoes returns. These echoes are converted to an
electrical signal in the probe. The receiver amplifies these to make them large enough to
process. The processing black box cleans up the echoes and conditions them before they
are stored in the computer memory. The display of the memory is the ultrasound image,
which is annotated with details of machine settings, patient details, and date. Image
storage is usually digital and often in a picture archiving and communications system
(PACS) system
5.6.1 Display
The display is invariably a computer monitor which
shows the image accumulating in the computer
memory. Each time-base line is digitized and echo
signals are written into the appropriate location for
subsequent display. In normal operation, memory
locations are refreshed as new echo information is
received and processed. Typically, complete scans of
100 ultrasound lines can be performed at a rate of
about 25/s.
5.7 Resolution
5.7.1 Spatial resolution
The spatial resolution (Fig. 19) of any imaging
system relates to the physical dimensions of the
energy pulse travelling/propagating through the
tissue and is critically dependent on the wavelength.
In microscopy, for example, electron microscopes
have significantly greater resolution than optical
microscopes in direct proportion to the ratio of the
relative wavelengths associated with electrons and
light photons. Similarly, in diagnostic ultrasound,
frequencies of 10 MHz have about twice the resolu-
tion of 5 MHz systems, the relative wavelengths in
soft tissue are 0.15 mm and 0.3 mm. The spatial
resolution of a pulse-echo ultrasound imager is
frequently assessed by linear measurement of the
Fig. 19 Spatial resolution is defined in terms of how
dimensions of the image produced by scanning a well the system can visualize small targets.
‘point’ target. The image is essentially the point- The three directions in which it is usually
spread function of the system – the point-spread determined are axial (along the beam direc-
function relates to the physical dimensions of the tion), lateral (across the beam in the scan
ultrasound pulse and thus to the ‘clarity’ or spatial plane), and slice thickness or elevation (across
resolution of the system. This measure of resolution the beam perpendicular to the scan plane).
is difficult to relate to the clinical situation; for The measurements are made by scanning a
phantom that contains wire targets embedded
example, the length of the point-spread function of a in a tissue-like substance and measuring the
nylon wire target in a gel is not directly applicable to width and length of the resulting bright dots
the accuracy of measurement of the diameter of an
intra-hepatic vessel in vivo. pulses are generated at a rate which is limited by the
requirement that the returning echoes can be
processed unambiguously and do not overlap with
5.7.2 Temporal resolution
echoes from a preceding pulse. The time interval
During the observation of moving targets, the rate at between pulses, for M-mode and Doppler, or
which new images are made, the sample rate, between frames, for imaging, is a measure of
determines the fidelity of measurement. Ultrasound temporal resolution.
ACKNOWLEDGEMENTS
F Author 2010