Vous êtes sur la page 1sur 16

SPECIAL ISSUE PAPER 127

A tutorial on ultrasonic physics and imaging techniques


M Halliwell
Medical Physics and Bioengineering, Bristol General Hospital, Bristol, UK

The manuscript was received on 7 May 2009 and was accepted after revision for publication on 5 October 2009.

DOI: 10.1243/09544119JEIM656

Abstract: Ultrasound is a widely used modality for both therapy and diagnosis in medicine
and biology. Currently, in the field of medical diagnosis, ultrasound is responsible for about
one in five of all diagnostic images. The physical characteristics of medical ultrasound, along
with its behaviour as it interacts with biological tissues, are described in this tutorial. The role of
ultrasound in therapeutic and diagnostic applications is briefly described. In view of the
importance of ultrasound as a medical imaging modality, the basic technological building
blocks utilized in diagnostic ultrasound scanners are also described. Many of these topics are
the subjects of other papers in this special issue where they are dealt with in more detail.

Keywords: acoustic oscillation

1 INTRODUCTION ‘infra-sound’ has a frequency less than 20 Hz, the


lower limit of human hearing. Unsurprisingly,
The purpose of this medical ultrasound tutorial is to ‘sound’ occupies the in-between frequency range
introduce in a fairly uncomplicated manner: the (from 20 Hz to 20 kHz). The frequencies commonly
basic physics of ultrasound; the generation and used in biological and imaging applications range
detection of high-frequency mechanical waves: the from about 500 kHz to more than 50 MHz.
interactions of ultrasound with biological tissues,
and some of the more frequently used techniques for
medical imaging. Specific consideration of the use of 3 PROPAGATION
ultrasound in therapeutic and diagnostic roles is
described elsewhere in this special issue; many of A useful simple model of tissues shows them as a
the more complex details will be found there. three-dimensional (3D) matrix of balls connected to
their neighbours by springs (Fig. 1). In this model, a
movement of any one of the balls is transmitted to
2 ULTRASOUND its neighbours by the altered tensions in the
compressed or expanded springs.
Ultrasound is an example of an ‘acoustic oscillation’. Propagation of ultrasound, the oscillatory move-
Acoustic oscillation is defined as: movement of ment of particles about a mean position, occurs by
particles in an elastic medium about an equilibrium the action of intra-particle forces (springs) on each
position. Other instances of acoustic oscillation are particle (ball) in the elastic medium (human tissue).
sound and infra sound. The only difference between The way in which displacements propagate through
these is in the frequency of oscillation. For the the medium is described by the wave equation. In its
acoustic oscillation to be called ‘ultrasound’, its
simplest form, the wave equation for a one-dimen-
frequency must be greater than 20 kHz, approxi-
sional (1D) wave describes the time change of
mately the upper limit of human hearing. Similarly,
pressure in terms of its change with distance as
*Corresponding author: Medical Physics and Bioengineering,
L2 DP L2 DP
Bristol General Hospital, Guinea Street, Bristol BS1 6SY, UK. ~c2 ð1Þ
email: mike.halliwell@bris.ac.uk Lt 2 Lx2

JEIM656 Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


128 M Halliwell

Fig. 2 Propagation of acoustic oscillations is achieved


through movement and changes in the relative
spacing of particles in response to high and low
pressure parts of the acoustic wave. (a) The
Fig. 1 Simple ‘balls on springs’ model of matter. The
upper graph represents the pressure fluctuation
balls represent the particles of the material and
with distance at an instant in time. The distance
the springs represent the forces between each
occupied by one cycle of pressure is the
particle and all its neighbours. The movement
wavelength (it is the distance travelled by the
of a single particle will change each force; this is
wave in one period, T). (b) The bottom graph of
represented by a compression or expansion of
the acoustic wave indicates the changing
each spring
pressure with time at a fixed position in front
of the transducer. T is the period of a single
In three dimensions, the wave equation becomes oscillation. The frequency is 1/T

 2 
L2 DP 2 L DP L2 DP L2 DP
~c z z ð2Þ
Lt 2 Lx2 Ly 2 Lz2
In Fig. 2, the period, T, of the wave is the time
interval between sequential occurrences of the same
where DP is the change in local pressure in the
phase and amplitude of the varying pressure. The
medium and c, the speed of the wave, is given by
frequency of the wave, f, is the reciprocal of the
  period
k
c~H ð3Þ
r
f ~1=T ð4Þ
where k is the bulk modulus of the tissue and r is the
density. For a wave travelling at speed c, the wavelength, l, is
given by
The bulk modulus (k) of a substance describes
behaviour under uniform compression. It is defined
l~c=f ð5Þ
as the pressure increase needed to cause a given
relative decrease in volume. The units are Pa.
The relationship can be expressed as
The density (r) of a material is its mass per unit
volume. The units are kg per cubic metre. c~lf ð6Þ
It is worth noting that, generally, the bulk modulus
will be a function of DP, and in turn DP is a function Note that c, the speed of sound, is dependent on the
of x, y, z, and t. Consequently, in real tissues, the mechanical characteristics of the material (Fig. 3).
speed of sound will not be absolutely constant For low-amplitude waves, the speed of sound is
during the passage of the ultrasound wave. constant for a particular material, with a small
dependence on temperature and frequency. How-
ever, the mechanical characteristics of the material
3.1 Period, frequency, and wavelength
also depend on the local pressure; consequently,
A simple depiction of medical ultrasound is as a high-amplitude waves can cause local changes in
sinusoidal variation of pressure with time. This is sound speed leading to non-linear propagation. This
represented graphically in Fig. 2. phenomenon is described later.

Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine JEIM656

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


A tutorial on ultrasonic physics and imaging techniques 129

A practical consequence of non-linear propagation is


the generation of harmonic frequencies during wave
propagation. The high-pressure portions of the wave
travel at higher speed than the low-pressure ones,
leading to the initial sinusoidal wave shape becom-
ing a more saw-tooth shape (Fig. 4).
Mathematically, the pressure front will become a
Fig. 3 Speed of sound differences in different material shock wave with a zero rise time. In reality the high-
types, using the balls and springs model. (a)
frequency components are sufficiently attenuated
Light balls (low density) and stiff springs (high
bulk modulus) gives a high speed of sound. The that a fully developed shock does not occur in soft
low-mass ball reaches its final particle displa- tissues at diagnostic power levels. However, the
cement in a short time as the stiff spring rapidly higher harmonics are sufficient in amplitude for
transmits enough force to accelerate the light them to be used to advantage in diagnostic imaging
particle. (b) Heavy balls (high density) and systems. Under the appellation ‘tissue harmonic
weak springs (low bulk modulus) give low imaging’, the harmonic frequency components are
speed of sound. A result of the combination of
processed and contribute to diagnostic images. Two
reduced force and greater mass is that the more
massive ball takes a longer time to reach the practical examples of the clinical advantages of
final particle displacement tissue harmonic imaging are that deep structures
can be visualized at higher resolution and that some
reverberation artefacts can be reduced.
3.2 Non-linear propagation
3.2.1 Speed of sound dependent on frequency
4 WAVES
For most materials, the propagation speed depends
on the frequency. In a more realistic form of the Acoustic waves in general are described as propagat-
wave equation the speed, c, must be replaced by the ing energy in 3D space, as described by the 3D wave
phase velocity. equation. For purposes of analysis, this propagation
The phase velocity (or phase speed), vp, of a wave
is the rate at which the phase of the wave propagates
in space. This is the speed at which the phase of any
one frequency component of the wave travels. For
such a component, any given phase of the wave (for
example, the crest) will appear to travel at the phase
velocity. The phase speed is given in terms of the
wavelength l (lambda) and period T

vp ~l=T ð7Þ

This phenomenon is ‘dispersion’. In soft tissues the


effect is small and the speed of sound is essentially
constant for the range of frequencies normally used
for medical ultrasound. Frequency has a much more
significant effect in relation to the dependence of Fig. 4 Progression of a sinusoidal pressure wave to a
attenuation with frequency. saw-tooth wave in a material where the speed
of sound increases with pressure. The high-
pressure parts of the wave travel more quickly
than the low-pressure parts. After a sufficient
3.2.2 Speed of sound dependent on pressure
travel time, the high-pressure parts will catch
In most materials, propagation speed depends on up with the low-pressure parts to form a ‘shock’
the amplitude of the wave. The non-linear wave wave. In soft tissue it is not possible for the
peak to overtake the trough, but in water waves
equation is
this phenomenon leads to the waves breaking
 2  into surf. In soft tissues the shock wave does
L2 DP 2 L DP L2 DP L2 DP not travel, as the very high frequencies asso-
~c ðP Þ z z ð8Þ
Lt 2 Lx2 Ly 2 Lz2 ciated with it are massively attenuated

JEIM656 Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


130 M Halliwell

is categorized into two fundamental modes: long-


itudinal and transverse.

1. Longitudinal waves: the direction of particle


displacement is the same as the direction of
energy flow.
2. Transverse waves: the direction of particle dis- Fig. 5 Reflection of acoustic waves at a mismatch
placement is perpendicular to that of energy flow. between acoustic impedances in the medium.
These waves are also called shear waves. For a wave incident from the left, the small balls
would bounce off the large ones; for a wave
incident from the right, the large balls would
4.1 Waves in biological tissues overshoot the small ones and the extra energy
The anisotropic nature of the forces between in the increased extension of the spring is
reflected towards the right
particles in human tissues determines the supported
wave types.

N Bone: in bony structures vibrations in all three reflections. Whenever a change in acoustic impe-
directions are supported and both longitudinal dance occurs, reflection will take place. Acoustic
and transverse (shear) waves propagate. impedance, Z, is of two types: specific or character-
N Soft tissue: in soft tissues, and liquid components istic acoustic impedance and acoustic impedance.
of the body, the coupling forces between particles The difference is that specific (characteristic) im-
in the direction of energy flow are large and pedance refers to the impedance at a point whereas
longitudinal waves with velocities of thousands of acoustic impedance refers to impedance at a surface.
metres per second are supported. The coupling
Specific acoustic impedance is the complex ratio of
forces perpendicular to this direction are weak;
the effective sound pressure at a point to the
consequently transverse particle movements are
effective particle velocity at that point. Impedance
rapidly attenuated. Shear waves are also sup-
can be expressed in either its constituent units
ported in soft tissues. However, they have a
(pressure per velocity per area) or in rayls. The rayl
relatively low velocity, in the order of hundreds
equals 1 pascal-second per metre (Pa s/m), or
of metres per second and high attenuation
equivalently, 1 Newton-second per cubic metre
coefficients, which render them unsuitable for
(N s/m3). In SI base units, that is kg/s m2
imaging purposes. Recent developments in elas-
tography make use of shear-wave propagation.  
Z~p=v~I v2 ~p2 I ð9Þ
N Mode conversion: in the situation of a longitudinal
wave incident at a non-normal angle to a
where p is the sound pressure (N/m2 or Pa), v is the
boundary, some of the energy of longitudinal
particle velocity (m/s), and I is the sound intensity
particle movement can cause particle movement
(W/m2).
in the transverse direction. This process is known
as mode conversion. Acoustic impedance is the total reaction of a
medium to acoustic transmission through it, repre-
4.2 Reflection sented by the complex ratio of the pressure to the
effective flux, that is particle velocity multiplied by
At a discontinuity of the mechanical characteristics surface area, through the medium. In other words,
in the path of the ultrasound wave, reflection of the sound pressure p divided by the particle velocity
energy occurs. In the simple example illustrated in v and the surface area S, through which the acoustic
Fig. 5, this discontinuity is represented by a change wave propagates. For example, in plane, single-
in the mass of the particles along the path and a frequency travelling waves the acoustic impedance
change in intra-particular forces. For a wave incident is equal to the specific (characteristic) impedance
from the left, say, reflection can be thought of in divided by the surface area, S
purely mechanical terms as the smaller particle
‘bouncing off’ the more massive particle – in much Z~p=vS ð10Þ
the same way that a ping-pong ball would bounce if
run against a snooker ball. Sometimes vS is referred to as the volume velocity.
Acoustic impedance is used to describe the me- For practical purposes, acoustic impedance, Z, for
chanical properties of tissues relevant to such any medium is calculated as the square root of the

Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine JEIM656

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


A tutorial on ultrasonic physics and imaging techniques 131

product of the density, r, and the bulk modulus, k The practical significance of these reflectivities in
pffiffiffiffiffiffi diagnosis is that ultrasound is limited to use
Z~ rk ð11Þ primarily in the assessment of soft tissues. Bone
presents a significant barrier and air is essentially
In the simple tissue model, the density relates impenetrable.
approximately to the mass of the particles and the Coupling/Matching layer: the acoustic impedance
bulk modulus to the forces between them. of transducer material is significantly greater than
N small masses, weak springs 5 low acoustic im- that of soft tissue. To reduce reflection, and conse-
pedance; quently improve transmission, at the probe/skin
N large masses, strong springs 5 high acoustic im- boundary, matching layers of solid materials of
intermediate acoustic impedance are incorporated
pedance.
pffiffiffiffiffiffiffiffi over the face of the transducer. This is an issue for
Using equation (3) (c 5 k=r), the computation equipment manufacturers. A related impedance
most often used to calculate Z is the product of the issue for the end users is the presence of air between
density and the speed of sound the probe face and the patient’s skin. The acoustic
impedance mismatch here is overcome by the use of
Z~rc ð12Þ a coupling gel with an impedance very close to that
of skin. A by-product of using gel as a coupling agent
Typical acoustic impedances for biological materi- is that the gel also acts as a lubricant. This facilitates
als, in rayls, are: easy manipulation of the probe over the skin surface.

Air 0.0004 6106


Lung 0.18 6106 4.2.2 Real reflections
Fat 1.34 6106
Water 1.48 6106 In real systems, boundaries between media are not
Blood 1.65 6106 flat and infinite. In the body, reflection processes
Muscle 1.71 6106 depend critically on the actual physical dimensions
Skull Bone 7.80 6106 in relation to the wavelength of the ultrasound being
used. Real reflections are considered to be some
combination of the two extreme types of interaction:
4.2.1 Reflection coefficients
non-specular (or scattering) and specular reflection.
In specific geometrical situations, reflection can be Scattering is the reflection of ultrasonic waves by
quantified. The amplitude reflection coefficient, R, is discontinuities or surface irregularities, small, rela-
given by the equation tive to the wavelength of the waves.
Pr ðZ2 {Z1 Þ For objects in tissue which have a maximum
R~ ~ ð13Þ dimension less than about a tenth of the wavelength
Pi ðZ2 zZ1 Þ
of the ultrasound energy, the vibrating object then
acts as a point radiator creating a spherical field, so
where Pr and Pi are the reflected and incident
the energy is re-radiated or scattered equally in all
pressure amplitudes, Z1 and Z2 are the acoustic
directions. Real objects, with dimensions closer in
impedance coefficients for the two media.
size to the wavelength, re-radiate ultrasound in a
The intensity reflection coefficient is
non-uniform manner, with a directivity pattern
  depending on the size–wavelength relationship and
Ir ðZ2 {Z1 Þ 2
~ ð14Þ the surface morphology. As the object dimensions
Ii ðZ2 zZ1 Þ increase to the limiting situation of a large flat
surface, the directivity pattern is identical to that of a
These equations hold for a plane-wave incident on
an infinite plane boundary at normal incidence. specular reflection (see Fig. 6)
Making that assumption, and using values from Specular reflection is the perfect, mirror-like re-
the table of acoustic impedances, it is possible to flection of ultrasound from a surface. Ultrasound
show that at a blood fat boundary approximately 10 from a single incident direction is reflected into a
per cent of the incident pulse amplitude is reflected. single reflected direction. The law of reflection, states
For soft tissue to bone, the value is 66 per cent, and that the angle of incidence equals the angle of
for soft tissue to air (lung), it is 99 per cent. reflection.

JEIM656 Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


132 M Halliwell

Fig. 7 Specular (mirror-like) reflection. At boundaries


large and flat relative to the wavelength, the
ultrasound beam is reflected without any
change in its characteristics (beam width,
convergence, or divergence) and in a direction
from the surface which is at an angle equal to
that of the incident beam. The angle of the
beam is measured with respect to a line which
is perpendicular to the surface (the ‘normal’).
For specular reflection, the angle of incidence is
equal to the angle of reflection. Tissue bound-
aries and blood vessel walls are targets which
Fig. 6 Scatter of ultrasound at boundaries smaller exhibit specular reflection and are best seen
than the wavelength. The targets can be when the ultrasound beams fall perpendicu-
considered as very small re-radiators of the larly on them
ultrasound. If such sources of ultrasound are
much smaller than a wavelength, the energy is tissue. The speckle pattern is entirely artefactual and
radiated uniformly in all directions. An extreme is characteristic of the imaging system.
example would be oscillations of a single ball in Interference occurs when the spatial separation of
the ‘balls on springs’ tissue model. All of its
the individual scatterers is less than the wavelength.
springs would be equally disturbed and the
displacements would propagate equally in all This results in areas of alternating high- and low-
directions. Real boundaries are larger than this level echoes. Speckle patterns are strongly wave-
and show some directivity length and therefore frequency dependent. They also
change as the angle of insonation of the scattering
Specular reflection can be considered to be a region is altered. Changing this angle alters the
special case of scattering. For specular reflection to relative separation in the beam direction and hence
occur, the interface between tissues must be rela- the pattern of interference. For randomly distributed
tively large (more than 10 wavelengths in diameter) targets the speckle pattern is random noise and
and relatively flat (less than one wavelength-high entirely artefactual. In real tissues the targets are not
irregularities). The boundary between liver and purely randomly distributed. The underlying tissue
kidney is an example of this kind of boundary for components impose a structure upon which the
ultrasound with a wavelength of less than 1 mm random speckle pattern is superimposed. The
(frequency more than 3 MHz). Pulses are reflected, textural pattern of the ultrasound images from tissue
as replicas of themselves, at an angle of reflection combines a random, fine echo texture, component
equal to the angle of incidence. This kind of with structural information at a coarser dimension.
reflection is often called ‘mirror-like reflection’ as it This is the reason that the textural pattern from the
is entirely analogous to optical reflection in flat glass liver is different from that from the kidney.
surfaces (Fig. 7). In spite of the fact that the echo patterns are
largely artefacts, many end users confuse the fine
Speckle is another special case of scattering. A
echo texture of the speckle component with true
speckle pattern is formed by the constructive and
resolution (Fig. 8).
destructive interference of ultrasound re-radiated
uniformly from randomly positioned closely packed Speckle reduction. Speckle can be reduced in
small objects (less than a tenth of the wavelength) in images by techniques such as spatial and frequency

Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine JEIM656

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


A tutorial on ultrasonic physics and imaging techniques 133

with differing speeds of sound. Refraction is de-


scribed by Snell’s Law

sin hi c1
~ ð15Þ
sin ht c2

where hi and ht are the angles of incidence and


transmission respectively, and c1 and c2 are the
speeds of sound in the two media (Fig. 9).
For angles of incidence other than 90u, the
direction of the main axis of the ultrasound beam
will be deviated across boundaries between tissues
with different sound speeds. For example, a beam
travelling from muscle to fat at an incidence angle of
60u will be deviated by 8u. This effect is significant in
imaging applications and causes considerable image
degradation in many patients. Real targets are
incorrectly located on the image leading to blurring
and loss of resolution.
Fig. 8 Speckle resulting from constructive and de-
structive interference of waves scattered from
clusters of small targets. Each small target re- 4.4 Attenuation
radiates the ultrasound as in scattering, but
because there are several scatterers in close As ultrasound propagates, energy is lost either by
proximity, the ultrasound wave can either absorption (as heat) or by reflection. Both specular
reinforce or cancel out in the tissues around reflection and scattering contribute to this loss of
the scatterers. If the waves happen to be in step energy from the beam. The intensity reduces
(pressure peaks in line) they reinforce (con-
exponentially with distance
structive interference). If waves are out of step
so that the pressure peak of one corresponds
with the pressure trough of another, then in Ix ~I0 expð{axÞ ð16Þ
that direction the waves cancel; in other words,
the interference between waves is destructive.
The pattern of the textural pattern is both
direction and frequency dependent. This allows
speckle reduction to be implemented in two
ways: by using multiple angles of view (steering
the beam–spatial compounding) and by trans-
mitting pulses with differing centre frequencies
(frequency compounding)

compounding. These take advantage of the spatial


and frequency dependence of speckle patterns.
Multiple images of the same region are created with
beams progressively steered so as to insonate at a
variety of different angles with pulses at a series of
different centre frequencies. Real targets are un-
affected by changes in beam angle or frequency. The
speckle patterns will be affected by these changes.
Following frame-averaging, speckle in the displayed
image will blur to a mid-grey background, leaving Fig. 9 Refraction due to the speed of sound difference
real targets conspicuous against it. across an inclined boundary. The ultrasound
image is constructed with an assumption that
4.3 Refraction the beam travels in a straight line; a deviation of
the beam away from this direction can lead to
As with all wave phenomena, ultrasound suffers misregistration and subsequent blurring of the
refraction as it passes a boundary between materials targets in the images

JEIM656 Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


134 M Halliwell

where I is the intensity, x is the distance, and a is the are operated at their resonance; consequently, their
attenuation coefficient. thickness is determined. For fundamental reso-
The description of attenuation is simplified with nance, the thickness must equal one half-wave-
the introduction of a logarithmic representation of length, remembering that it is the wavelength in the
the intensity variation. Taking logarithms of equa- transducer material. In a commonly used piezo-
tion (16) electric material, lead zirconate titanate (PZT), the
speed of sound is approximately 3300 m/s. The wave-
logðIx =I0 Þ~{ax ð17Þ length of sound in PZT at a frequency of 5 MHz is
approximately 0.7 mm. Consequently, 5 MHz reso-
The logarithm of the ratio of intensities is called the nant elements would have to be 0.35 mm thick. The
Bel. In view of the magnitudes involved in medical other dimensions, height, width, and shape, depend
applications, the more frequent notation is the upon the application. Typically, they can be single flat
decibel (dB). The usage is the term ‘intensity level’ or bowl-shaped elements of circular or rectangular
where shape up to 100 wavelengths in dimension or they can
be in the form of rectangular arrays of side-by-side
intensity level~10 log10 I1 =I0 dB ð18Þ elements about 0.2 wavelengths wide and 5–10
wavelengths long, with either flat or convex front
For example, in a situation where the intensity at the faces. Whatever its particular construction, the trans-
location of interest (I1) is 100 times less, than the ducer arrangement and its housing is usually referred
initial intensity I0, the intensity level would be to as the ‘probe’.
described as being ‘minus 20 dB’.
The attenuation coefficient, a, is frequency de-
pendent and in the order of 0.5 to 0.7 dB per 4.6 Continuous wave ultrasound
centimetre per MHz. The loss of signal strength A continuous wave is an acoustic wave of constant
due to attenuation is a limiting factor for diagnostic amplitude and frequency and, strictly speaking, of
purposes and typically restricts the depth of imaging infinite duration. In practice, ultrasound is consid-
to about 15 cm at 5 MHz. ered to be a continuous wave if is transmitted without
interruption while the equipment is functioning. Its
duration is never infinite but is long enough for its
4.5 Wave production and detection amplitude and frequency to be considered constant.
Acoustic oscillations can be generated in many ways. Continuous wave ultrasound propagating in a
Simple sound-wave generators include plucked medium of low attenuation between parallel reflect-
guitar strings and beaten drum skins. In both of ing surfaces can create a standing wave field. The
these examples, the mechanical vibrations of the standing wave field is one in which there are
surfaces cause the air molecules with which they are stationary regions of high and low pressure. Stand-
in contact to vibrate in sympathy. As it is a higher ing wave fields have been shown to cause blood flow
frequency, ultrasound is usually generated using stasis in chick embryos in vivo. They are also used as
mechanical vibrations which result from the elec- a tool for contactless separation, concentration and
trical excitation of a transducer. manipulation of microparticles and biological cells.
These devices convert or ‘transduce’ the electrical The basis is the acoustic radiation force, a non-linear
exciting signal into a mechanical movement via a effect which causes particles to be attracted to either
variety of mechanisms. The inverse piezoelectric the nodes or antinodes of the standing wave. In
effect is the most frequently employed but capacitive therapy, continuous wave ultrasound is used in
and solid-state devices also exist. applications such as physiotherapy and surgery.
Ultrasound detection is often undertaken by the Diagnostically, continuous wave ultrasound is used
same transducer, now operated in the inverse mode solely in Doppler applications.
(for example, utilizing the direct piezoelectric effect).
Changes in local pressure around the transducer (the
4.7 Pulsed ultrasound
mechanical disturbance) are converted to electrical
signals. A pulse wave is a non-sinusoidal waveform. The
The thickness of the transducer governs its shape of the wave is determined by the character-
resonant frequency. This is the frequency of max- istics of the exciting electrical pulse and the proper-
imum efficiency and sensitivity. Usually, transducers ties of the transducer. Typically, pulsed ultrasound is

Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine JEIM656

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


A tutorial on ultrasonic physics and imaging techniques 135

generated as a sequence of regularly occurring Manufacturers make use of the increased sensi-
pulses. The pulse duration and the time interval tivity and bandwidth of transducers by labelling
between pulses depend on the application. them as having a range of usable frequencies equal
In the frequency domain, the primary character- to the upper and lower limits of the bandwidth; for
istics of an ultrasound pulse are its centre frequency example, a typical diagnostic probe for abdominal
and its bandwidth. For typical excitation with a use may be labelled as a ‘3 to 7 MHz probe’. This
short, high-amplitude voltage spike, the centre means that the ultrasound pulse has a centre
frequency of the pulse is the mechanical resonant frequency of 5 MHz but a bandwidth sufficient for
frequency of the transducer elements. The resonant images to be created at filtered, received frequencies
frequency depends on the thickness of the elements from 3 to 7 MHz.
and cannot be altered for a particular transducer.
Changing the centre frequency of the pulse requires
4.7.1 Beam shape
changing the thickness of the elements; in other
words, it requires the probe to be changed. Typical beam shapes and characteristics of ultra-
The bandwidth of the pulse is a measure of the sound emitted from the transducers used in medical
range of frequencies it contains. Usually, the width is and biological systems are detailed in Fig. 11.
measured between the 23 dB points and is referred
to as the full-width, half maximum (FWHM) value
(Fig. 10).
The upper and lower frequencies are taken as the
usable limits of the pulse. Above and below these
frequencies, the signal amplitude is less than half of
its maximum value and is considered to be insuffi-
cient for most purposes. Similar considerations
apply to the overall sensitivity characteristic of the
transducer. Its response is taken to be the square of
that of the pulse it launches.

Fig. 10 An ultrasound pulse typically contains four or


five cycles. The number of cycles determines
the length and frequency content of the pulse.
The fewer the cycles, the shorter the pulse and Fig. 11 Beam from a plane transducer is fairly well
the smaller (better) the resolution along the collimated but diverges after the near-field/
beam (axial). The frequency content of the far-field boundary. Focusing the beam is
pulse is described by its bandwidth. A wide important and is frequently achieved by
bandwidth comes from a pulse with few curving the emitted wavefront, either with a
cycles. Very long pulses are narrow band- lens for a single-element transducer, or by
width, the limit is an infinitely long pulse (a timing delays for the pulses exciting an array
continuous wave) which has a single fre- transducer. The curved wave front converges
quency component to the focal point and diverges beyond it

JEIM656 Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


136 M Halliwell

4.7.2 Therapy ultrasound


Pulsed ultrasound is used for some physiotherapy
applications. The rationale is that the time interval
between pulses enables induced tissue heating to
dissipate, allowing non-thermal effects of ultrasound
to predominate in the patient’s treatment.
Surgical and high-intensity focused ultrasound
(HIFU) applications also use pulses. The reason here
is to enable delivery of a specified dose of energy to
heat and to destroy the target volume. Energy Fig. 12 Pulse echo technique. A pulse launched
deposition is determined by a combination of towards the target is partially reflected at the
acoustic power, pulse number, and pulse duration. target, causing an echo to return to the
transducer. The time elapsed, t (measured at
the transducer), is equal to twice the time
4.7.3 Diagnostic ultrasound taken for the pulse to travel to the target. The
range of the target is computed assuming a
The majority of diagnostic imaging techniques also value, c, for the speed of sound in the medium.
use pulsed ultrasound. Pulse duration and ampli- Range, d 5 K t6c
tude control the characteristics of equipment per-
formance and are tailored for specific applications. horizontal distance between two echoes represents
High-resolution imaging requires pulses to be as the time interval between them. The vertical excur-
short as possible; these pulses are created by exciting sion, in a linear system, is directly proportional to
the transducer with a single voltage pulse of less the amplitude of the signal. In many practical
than a half-period duration. Imaging of blood flow systems the vertical excursion is logarithmically
uses longer pulses; these may be generated by proportional to the received echo amplitude. This
excitation with a gated sinusoidal voltage. takes account of the significance of the low-level
scattered echoes from tissues.

4.8 The pulse echo technique


5.1.1 Radiofrequency display
Similar to its application as the echo location
method used by bats and dolphins, this method In their raw, unprocessed form, echo data are
forms the basis of the majority of diagnostic systems. radiofrequency (RF) electrical signals. The display
Ultrasound is launched in the form of pulses; of these RF signals incorporates more than simply
reflections of those pulses at tissue boundaries information about the amplitude of the signals and
return to the probe as echoes (Fig. 12). their time of arrival back at the probe face. There is
Imaging relies on those reflections to yield also detail concerning the phase changes that have
information concerning the range and direction of occurred during the reflection and interference
interfaces. The range is deduced from the time processes (Fig. 13(a)). Historically, this additional
interval between launching the ultrasound pulse and information has been largely unused. Development
receiving its echo. This time is converted into a of high-speed, wide dynamic range analogue-to-
measure of target range by making an assumption digital conversion hardware and increases in the
about the speed of propagation of the pulse in the processing power of computers has enabled the
body tissues. The directional location of the reflect- construction of scanners which take advantage of
ing target is assumed to be identical with the this information. Colour Doppler imaging, synthetic
direction in which the ultrasound pulse was initially aperture, and tissue characterization techniques are
launched from the face of the probe. available in some commercial equipment and
developments of these processes are to be expected.

5 DISPLAY AND IMAGING TECHNOLOGIES


5.1.2 Unipolar, filtered display
5.1 A-mode
A more frequently used display involves processing
A display method in which echo amplitude is plotted the RF signals, usually by full wave rectification and
along the y-axis versus time on the x-axis. The filtering, to produce a unipolar display. Target

Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine JEIM656

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


A tutorial on ultrasonic physics and imaging techniques 137

Fig. 13 (a) Upper trace is the radio frequency (RF)


scan, the raw ultrasound echo train detected
and amplified. The ultrasound echo train is a Fig. 14 B-scan is formed by changing the amplitude
series of pressure fluctuations at a frequency fluctuations of the A-scan to a brightness
approximately the same as that of the trans- modulated time-base line. The peaks on the
ducer. The transducer converts the pressure A-scan corresponding to individual targets are
changes to an electrical signal. This voltage now shown as bright dots on the B-scan. The
trace is very similar to the signals used in radio brightness depends on the amplitude of the A
transmitters and receivers and so was given scan peak
the name ‘radio frequency signal’ in the early
days of ultrasound development. (b) Lower The brightness of the dot is related to the
trace is the A-scan, the result of rectifying and amplitude of that particular received echo; the
envelope-detecting the RF trace. The radio position of the dot along the time-base line is
frequency fluctuations are too rapid and proportional to the time of arrival of the echo at
irregular for easy discrimination by eye.
the probe face. The position of the time-base line on
Changing the echo train to a lower frequency
representation (the A scan) makes it easier to the screen is linked to the spatial location of the
appreciate the size and presence of targets ultrasound beam. The direction is governed by the
along the beam direction at which the ultrasound pulse is launched
from the probe face (Fig. 15).
boundaries are displayed as short, fast rise time, Frequently, the display shows a representation of
spikes positioned along the display at locations the face of the probe, either a horizontal straight line,
proportional to their time of arrival. for a linear array flat faced probe, or a curved line,
The initial use of such displays was for flaw when the probe used has a curved face (a curvilinear
detection in metal constructions, typically in the probe). Time-base lines are then drawn as though
shipbuilding and aircraft industries. For medical emanating from this face in the direction deter-
purposes, the 1D A-mode displays were used in the mined by the scanner.
differentiation of solid and cystic lesions and in the
measurement of structures such as the fetal head.
Medically, the displays have very largely been super-
seded by the two-dimensional (2D) images produced
by B-mode scanners. An example of a clinical use
which still utilizes the A-scan is the accurate
determination of tissue thickness. For example, prior
to surgical eye correction procedures, the thickness of
the cornea is measured using A-mode ultrasound to
an accuracy of about 100 microns.
Fig. 15 Sequential pulses from the face of the trans-
ducer are depicted as vertical time-base lines
on the display screen. The horizontal starting
5.2 B-mode point of each vertical scan line corresponds
with the site on the probe face from which the
B-mode (‘brightness modulation’) scanning is a corresponding ultrasound beam emerges.
major component of by far the most frequently used Echoes are drawn at the appropriate depth
technique in diagnostic ultrasound imaging. The on each line. The final scan shows the cross-
principle is that echoes are displayed as bright dots sectional view of targets beneath the transdu-
along a time-base line. cer

JEIM656 Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


138 M Halliwell

5.3 C-scan: constant depth scanning


The C-scan yields a 2D plan view of the object. Early
systems involved mechanical movement of the
transducer over the surface in an x–y raster. The
display was a single dot whose position was linked
to that of the transducer. The brightness of the dot
was modulated by the echo strength at each x–y
location and from a depth along the z-axis deter-
mined by a range-gated receiver. The constant
depth scans were time-consuming to acquire and
found limited use in medical applications. Devel-
opments of 3D scanning have revitalized the C-
scan, which can now be created by resampling in
planes orthogonal to the original 2D slices within
the data volume [1]. Fig. 16 M-mode trace takes echoes from a single
direction within the scan plane and traces
out the movement of targets with time.
Stationary targets, such as the chest wall, will
5.4 M-mode
appear as parallel horizontal lines. Moving
The motion mode display is a variant of the single targets, such as the leaflets of a heart valve,
line B-mode scan. The vertical time-base line will trace their movement over time. From the
trace, values such as total excursion and rate
contains the echoes deep to the stationary trans-
of movement can be measured. The M-mode
ducer and along one of the ultrasound beams. The is used most frequently in examining the
horizontal time-base is a constant speed shift of motion of cardiac structures
those echoes. Stationary echoes are displayed as
horizontal lines. Echoes from moving structures, 5.5.2 Beamformer
such as heart valves, draw out distance–time
Each of these pulses is channelled via the beamfor-
graphs of their movements. From these traces,
mer to transducer array. The beam former deter-
the displacement, speed and acceleration of mov-
mines which group of elements in the array are
ing structures may be computed. Those character-
excited. An ultrasound pulse is launched from that
istics are of clinical significance in cardiac dis-
group, propagates through the patient, and echoes
orders (Fig. 16).
from boundaries along the path return to the probe
face to be converted to electrical signals. The
number of elements in the transmission group is
5.5 Real-time ultrasound B-mode scanner
usually between 10 and 15. On reception, the
The simplified construction layout and sequence of number of elements in the receive group may well
events in a basic scanner is illustrated in Fig. 17. be increased as deeper and deeper echoes return.
The reason for this is to optimize the focusing and
sensitivity performance for distant targets. Essen-
5.5.1 Pulser tially, this process increases the aperture and focal
The pulser generates a repetitive sequence of depth of the receiver with time.
voltage pulses. The rate of pulsing depends on the
type of scan and, in particular, the maximum depth
5.5.3 Receiver
of display required. The time interval between
pulses is usually sufficiently long to allow echoes This amplifies the raw electrical signals. Echoes from
from the deepest targets to reach the probe face close targets produce signal voltages in the order of
before the next pulse is launched. In soft tissue with millivolts. However, pulse and echo attenuation
a speed of sound of 1540 m/s (1.54 mm/ms) the time introduced by soft-tissue interactions during propa-
interval for a depth of 15 cm (a typical liver gation dictates that mean signal levels fall exponen-
dimension) will be 200 ms (100 ms for the pulse to tially with time of arrival (target range). The fall in
reach the deepest part and 100 ms for the echo to signal strength is compensated by an increase in the
return). receiver gain with time, logarithmically. Such com-

Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine JEIM656

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


A tutorial on ultrasonic physics and imaging techniques 139

Fig. 17 Block diagram of a typical real-time scanner. The pulser or clock creates a regular
sequence of electrical signals which are processed by the beamformer to be sent to the
appropriate group of elements in the probe. The pulse created by the electrical pulse
travels into the patient and a train of echoes returns. These echoes are converted to an
electrical signal in the probe. The receiver amplifies these to make them large enough to
process. The processing black box cleans up the echoes and conditions them before they
are stored in the computer memory. The display of the memory is the ultrasound image,
which is annotated with details of machine settings, patient details, and date. Image
storage is usually digital and often in a picture archiving and communications system
(PACS) system

pensation is termed depth compensation, time gain


compensation or swept gain compensation. It is
implemented at the receiver by user operated
controls (Fig. 18).

5.6 Image processing


Just as the RF signals were processed to make an
easily intelligible A-mode display, so the echo signals
need processing for the B-mode image. The main
difference is that the signal amplitude is used to
modulate the brightness of the time-base line and
the appreciation of image detail now relies much
more on the physiology of human perception. A
large range of processing techniques is employed to Fig. 18 Exponentially increasing amplification is ap-
achieve optimum image performance. They will be plied to the received echoes to compensate for
used in a variety of ways which will depend on the attenuation within tissues. This is called
clinical context. For example, liver scanning requires variously, depth gain, swept gain, or time gain
display characteristics which emphasize low signal compensation. The aim is to ensure that
similar targets appear with similar brightness,
discrimination to allow the visualization of liver
irrespective of their depth. The limit to
metastases. However, cardiac applications need amplification, subject to the electronic noise
characteristics which suppress low-level artefactual levels in the receiver, is reached at a particular
echoes and noise to enable the cardiac chambers to depth, after which the echo brightness reduces
be delineated clearly. and structural detail is lost

JEIM656 Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


140 M Halliwell

The different mixtures of processing techniques


are under operator control, but more usually, they
are stored in the machine as application-specific
presets which the operator selects prior to each
investigation.

5.6.1 Display
The display is invariably a computer monitor which
shows the image accumulating in the computer
memory. Each time-base line is digitized and echo
signals are written into the appropriate location for
subsequent display. In normal operation, memory
locations are refreshed as new echo information is
received and processed. Typically, complete scans of
100 ultrasound lines can be performed at a rate of
about 25/s.

5.7 Resolution
5.7.1 Spatial resolution
The spatial resolution (Fig. 19) of any imaging
system relates to the physical dimensions of the
energy pulse travelling/propagating through the
tissue and is critically dependent on the wavelength.
In microscopy, for example, electron microscopes
have significantly greater resolution than optical
microscopes in direct proportion to the ratio of the
relative wavelengths associated with electrons and
light photons. Similarly, in diagnostic ultrasound,
frequencies of 10 MHz have about twice the resolu-
tion of 5 MHz systems, the relative wavelengths in
soft tissue are 0.15 mm and 0.3 mm. The spatial
resolution of a pulse-echo ultrasound imager is
frequently assessed by linear measurement of the
Fig. 19 Spatial resolution is defined in terms of how
dimensions of the image produced by scanning a well the system can visualize small targets.
‘point’ target. The image is essentially the point- The three directions in which it is usually
spread function of the system – the point-spread determined are axial (along the beam direc-
function relates to the physical dimensions of the tion), lateral (across the beam in the scan
ultrasound pulse and thus to the ‘clarity’ or spatial plane), and slice thickness or elevation (across
resolution of the system. This measure of resolution the beam perpendicular to the scan plane).
is difficult to relate to the clinical situation; for The measurements are made by scanning a
phantom that contains wire targets embedded
example, the length of the point-spread function of a in a tissue-like substance and measuring the
nylon wire target in a gel is not directly applicable to width and length of the resulting bright dots
the accuracy of measurement of the diameter of an
intra-hepatic vessel in vivo. pulses are generated at a rate which is limited by the
requirement that the returning echoes can be
processed unambiguously and do not overlap with
5.7.2 Temporal resolution
echoes from a preceding pulse. The time interval
During the observation of moving targets, the rate at between pulses, for M-mode and Doppler, or
which new images are made, the sample rate, between frames, for imaging, is a measure of
determines the fidelity of measurement. Ultrasound temporal resolution.

Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine JEIM656

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


A tutorial on ultrasonic physics and imaging techniques 141

ACKNOWLEDGEMENTS

I thank numerous colleagues in the Department of


Medical Physics at University Hospital Bristol NHS
Trust for their help and encouragement. In particu-
lar, thanks go to Ms T. Robinson, Dr S. Curtis and Dr
H. Liang. I must also thank Professor P. N. T. Wells.
for the kind invitation to participate in this special
issue.

F Author 2010

Fig. 20 Contrast resolution is the ability to display REFERENCES


regions which have small differences in echo
size. As well as machine signal-processing
aspects, the physical size of the beam is a 1 Prager, R. W., Ijaz, U. Z., Gee, A. H., and Treece, G.
limiting factor. Beams which are wider than M. Three-dimensional ultrasound imaging. Proc.
the target will override the true echoes from IMechE, Part H: J. Engineering in Medicine, 2010,
the target with echoes from surrounding 224(H2), 193–224. DOI: 10.1243/09544119JEIM586.
tissue. In this example, the cyst becomes
invisible when the beam is too wide
BIBLIOGRAPHY

Allan, P. L., McDicken, W. N., Pozniak, M. A., and


5.7.3 Contrast resolution
Dubbins, P. A. Clinical Doppler ultrasound, 2000
Contrast resolution is the ability to depict the (Churchill Livingstone, London).
difference between regions of differing reflectivity. Bushong, S. C. and Archer, B. Diagnostic ultrasound:
A clinical example is the display of metastatic cancer physics, biology and instrumentation, 1991 (Mosby-
Year Books Inc., Philadelphia).
deposits against the background of the liver par-
Evans, D. H. and McDicken, W. N. Doppler ultrasound:
enchyma. This ability depends on several factors to physics, instrumentation and signal processing, 2nd
do with signal processing but also with the physical edn, 2000 (Wiley, Chichester).
dimensions of the beam. Narrow beams are more Fish, P. Physics and instrumentation of diagnostic
likely to return signals which relate to the echogeni- medical ultrasound, 1990 (John Wiley & Sons,
city of the lesion; wide beams will add echoes from Chichester).
tissues surrounding the lesion and will reduce its Hedrick, W. R., Hykes, D. L., and Starchman, D. E.
visibility. In Fig. 20 the extreme situation of a cyst Ultrasound physics and instrumentation, 1995 3rd
edn (St. Louis, MO: Mosby).
(which should return no echoes) is demonstrated for
Hill, C. R., Bamber, J. C., and ter Haar, G. R. Physical
two beam sizes. Contrast resolution is difficult to principles of medical ultrasonics, 2004 (John Wiley &
measure but is one of the major performance Sons, Chichester).
differences between scanners. Hofer, M., Hofer, M., and Telger, T. C. Teaching
manual of color duplex sonography, 2000 (Thieme
Publishing Group, Stuttgart).
Hoskins, P., Thrush, A., Martin, K., and Whittingham,
T. A. Diagnostic ultrasound: physics and equipment,
6 CONCLUSIONS 2003 (Greenwich Medical Media, London).
Kremkau, F. W. Diagnostic ultrasound: principles and
Ultrasound is a very widely used modality both in instruments, 5th edn, 1998 (Philadelphia, W. B.
therapy and diagnosis and it is estimated that Saunders Company).
approximately 20 per cent of all medical images are McDicken, W. N. Diagnostic ultrasonics: principles and
made with ultrasound. There is no sign that its use is use of instruments, 3rd rev. edn, 1990 (Churchill,
Livingstone).
diminishing and new applications and new technol-
Szabo, T. L. Diagnostic ultrasound imaging: inside out,
ogies are regularly introduced. This paper has 2004 (Elsevier, Academic Press).
illustrated the basic physical properties of ultrasound. Taylor, K. J. W., Burns, P. N., and Wells, P. N. T.
For more depth on the topics touched on here, the Clinical applications of Doppler ultrasound, 1995
other articles in this special issue may be consulted. (Raven, New York).

JEIM656 Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016


142 M Halliwell

Wells, P. N. T. Physical principles of ultrasonic Wells, P. N. T. Advances in ultrasound techniques and


Diagnosis, 1969 (Academic Press, Inc., London). instrumentation, 1993 (Churchill Livingstone, New
Wells, P. N. T. Biomedical ultrasonics, 1977 (Academic York).
Press, Inc., London). Zagzebski, J. A. Essentials of ultrasound physics, 1996
Wells, P. N. T. Ultrasonics in clinical diagnosis, 1977 (Mosby, St Louis).
(Churchill Livingstone, Edinburgh).

Proc. IMechE Vol. 224 Part H: J. Engineering in Medicine JEIM656

Downloaded from pih.sagepub.com at PENNSYLVANIA STATE UNIV on September 18, 2016

Vous aimerez peut-être aussi