Vous êtes sur la page 1sur 3

Encore“ 500 and Eclipse“ Electronic Calibration

(Pump Software version 1.5.40 or higher)


Review safety information in MDE 3804 Encore/Eclipse Start ON
OFF
up/Service manual first.
The following steps MUST be performed in the order given or the
calibration will not be properly completed.

1 Turn on the Weights & Measures switch (located on side 1) Prover Can Size
Note! This step has
”no default value,
2 Press F1 you must enter a
prover can size.

3 Press 0 1 2 8 ENTER

4 Press 7 . 8 ENTER

Note! To calibrate, a conversion


factor must be set and is indicated
5 Enter prover can size (e.g. '5' for five gallons) and press ENTER
in the left digit in the PPU display.

6 Enter the Side number (1 or 2) and press


ENTER

7 Remove nozzle and select grade or lift lever Side 1 or 2


NOTE: only pure products can be calibrated

8 Dispense product into prover can as close to the zero as possible


and hang up nozzle
ENTER
The volume display shows the actual pulse count - press

9 Enter the offset (number of cubic inches or centimeters away from Cubic inches of
zero) and press ENTER
deviation from Zero
10 Enter 0 if offset is above zero or 9 if offset is below zero and
press ENTER

11
The volume display will be flashing the adjusted volume (e.g. 5.008
gallons should equal + 2 cubic Inches on the prover scale) You now
have the option of pressing CLEAR to return to Step 10 to re-enter the
offset data.

12 Press to return to step 5 and if no further calibration is required


ENTER

continue to the next step

NOTE: If is not pressed before the Weights & Measures switch


ENTER

is turned off, calibration will be lost. 0 = at zero or above


9 = below zero
13 Turn Weights & Measures switch off
14 Press F2 to exit programming
Notes:
Electronic Calibration must always be manually verified before returning the
dispenser to service

For additional assistance contact your local Gilbarco Veeder-Root Distributor or Authorized Service
Contractor.

Version 1.1 February 14, 2003


Encore“ 500 and Eclipse“ Electronic Calibration
(Prior to Pump Software version 1.5.40)
OFF ON
Review safety information in MDE 3804 Encore/Eclipse Start
up/Service manual first.
The following steps MUST be performed in the order given or the
calibration will not be properly completed. Prover Can Size
Note! This step has
1 Turn on Weights & Measures switch (located on side 1) ”no default value,
you must enter a
2 Press F1 prover can size
3 Press 0 1 2 8 ENTER

4 Press 7 . 8 ENTER

Side 1 or 2
5 Enter prover can size (e.g. '5' for five gallons) and press ENTER

6 Enter the Side number (1 or 2) and press ENTER

7 Remove nozzle and select grade or lift lever


NOTE: only pure products can be calibrated

8 Dispense product into prover can as close to the zero as possible


and hang up nozzle, the volume display will show flashing zeros.

9 Enter the offset (number of cubic inches or centimeters away from


zero) press ENTER
Cubic inches of
10 Enter 0 if offset is above zero or 9 if offset is below zero and deviation from Zero
press ENTER

11 Press to return to step 5 and if no further calibration is required


ENTER

continue to the next step


ENTER
NOTE: If is not pressed before the Weights & Measures switch
is turned off, calibration will be lost.

12 Turn Weights & Measures switch off


13 Press F2 to exit programming 0 = at zero or above
9 = below zero

Notes:
Electronic Calibration must always be manually verified before returning the
dispenser to service

For additional assistance contact your local Gilbarco Veeder-Root Distributor or Authorized Service
Contractor.

Version 1.1 February 14, 2003


Encore“ 300 Electronic Calibration NOTE! The ECAL
switch is always
Review safety information in MDE 3804 Encore/Eclipse Start on side 1.
up/Service manual first.

Note: Electronic Calibration requires a Conversion factor (CC 17)


PRIOR to any calibration.

Prior to entering Command Code 90 for side 1 or Command Code


91 for side 2, make sure the Weights and Measures switch is turned
on

The following steps MUST be performed in the order given or the OFF ON
calibration will not be properly completed

Press Prover Can Size


F1
Enter command code 90 or 91and press
ENTER
, you will be prompted
005 90 Note! This step has
”no default value,
you must enter a
to enter the prover can size in Gallons (or whatever units the
conversion factor is set for) and press prover can size
ENTER

Remove the nozzle and raise the lever or select the grade.

Dispense fuel into the prover can as close as possible to the ”zero
mark and hang up the nozzle, - Any re-activation of the handle
switch prior to the completion of the calibration procedure will
result in a lost calibration
005 90
Enter the number of cubic inches the fuel is away from zero in the
prover can and press ENTER Cubic inches of
deviation from
Enter either a ” 0 if the fuel level is at or above zero, or a ” 9 if Zero (up to 99)
the fuel level in the prover is below zero
ENTER
Press to return to the main display to enter a new prover size,
side, or grade. The volume display will display the adjusted volume
amount (eg 5.008 should equal +2 cu. Inches in the prover).
005 90
OR OR
If you are done calibrating turn OFF the Weights and Measures
switch to exit programming mode. Note: if this switch is turned off
prior to the completion of calibration programming - calibration
will be lost. 0 = at zero or above
9 = below zero

Notes:
Electronic Calibration should always be manually verified before returning the dispenser to service

Calibration on blenders can only be checked by testing a pure High and Low programmed grade, any
testing of blended grades will be inaccurate.

For additional assistance contact your local Gilbarco Distributor or Authorized Service Contractor.

Vous aimerez peut-être aussi