Vous êtes sur la page 1sur 1

UJI VALIDASI METODE TES

ALAT HEMATOLOGI SYSMEX XNL350

Data running bahan kontrol selama 15 kali sebagai berikut.

Run WBC RBC HGB HCT PLT


1 3.09 2.38 6.1 18.20 83
2 3.00 2.38 6.1 18.20 88
3 3.09 2.37 6.2 17.90 83
4 2.93 2.41 6.2 18.20 91
5 2.95 2.36 6.1 17.80 88
6 3.00 2.38 6.2 18.00 87
7 2.93 2.37 6.2 18.00 95
8 3.16 2.37 6.1 17.90 91
9 2.87 2.33 6.1 17.70 88
10 3.02 2.37 6.2 17.90 89
11 3.02 2.36 6.1 18.00 94
12 3.02 2.37 6.2 18.10 89
13 2.91 2.37 6.1 17.70 104
14 3.00 2.37 6.1 18.10 104
15 3.04 2.34 6.1 17.80 113

Mean 3.00 2.37 6.14 17.97 92.47


SD 0.08 0.02 0.05 0.17 8.43
Coefficient
of Variation
(CV) 0.025 0.008 0.008 0.010 0.091
Laboratory Improvement Amendments (CLIA) 1988
Kesimpulan Guideline
Laboratory Improvement Amendments (CLIA) 1988
Guideline

Balikpapan, 21 Januari 2019


Yang melakukan uji, Kepala Instalasi Laboratorium,

Sri Puji Astuti, A.Md.AK Oktien Harsari P, dr., SpPK


NIP. 197803311997032001 NIP. 19691010200002008