Académique Documents
Professionnel Documents
Culture Documents
com
ScienceDirect
Publishing Services by Elsevier
International Journal of Naval Architecture and Ocean Engineering 9 (2017) 589e597
http://www.journals.elsevier.com/international-journal-of-naval-architecture-and-ocean-engineering/
Abstract
This study aims to verify experimentally the specifications of the data acquisition system required for the precise measurement of signals in
an underwater explosion (UNDEX) experiment. The three data acquisition systems with different specifications are applied to compare their
precision relatively on maximum shock pressures from UNDEX. In addition, a method of assessing the acquired signals is suggested by
introducing the concept of measurement uncertainty. The underwater explosion experiments are repeated five times under same conditions, and
assessment is conducted on maximum quantities acquired from underwater pressure sensors. It is confirmed that the concept of measurement
uncertainty is very useful method in accrediting the measurement results of UNDEX experiments.
Copyright © 2017 Society of Naval Architects of Korea. Production and hosting by Elsevier B.V. This is an open access article under the
CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
Keywords: Naval ship; Underwater explosion (UNDEX) experiment; Shock pressure; Data acquisition system; Uncertainty of measurement
http://dx.doi.org/10.1016/j.ijnaoe.2017.04.002
p2092-6782 e2092-6790/Copyright © 2017 Society of Naval Architects of Korea. Production and hosting by Elsevier B.V. This is an open access article under the
CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
590 S.-J. Moon et al. / International Journal of Naval Architecture and Ocean Engineering 9 (2017) 589e597
pressure sensor #3
pressure sensor #2
6m
pressure sensor #1
2m
explosive (TNT)
4m
11 m
depth of 15 m. A 1 kg explosive (TNT) was installed in the The underwater shock wave generated by the explosion is
center of the water tank, and three underwater pressure sensors superimposed on the hydrostatic pressure. The pressure history
were installed at a radius of 2 m, 4 m, and 6 m, respectively. of the shock wave at a fixed location starts with an instanta-
The installation depth of the explosive and the sensors is 5 m. neous pressure increase to a peak followed by a decline. That,
The three aforementioned DAQ systems were used for col- initially, is usually approximated by an exponential function.
lecting the experiment data. The peak pressure in the shock front depends on the size of the
Among the experiment results, Fig. 2 shows the signals explosive charge and the stand-off distance from this charge at
measured by pressure sensor #1 and DAQ system No. 2. It can which the pressure is measured. Peak pressure pm is expressed
be checked that shock wave pressure due to underwater ex- by the empirical equation of Cole (1948) as the following
plosion, reflected wave pressure from the wall of the water empirical formula:
tank, and pressure from bubbles were measured.
In this study, only the shock wave pressure part is consid- 1=3 A
ered. The signals collected from each sensor and DAQ system W
pm ¼ K$ ðunit : MPaÞ ð2Þ
are analyzed and only the shock wave part is shown in Fig. 3, R
and a similar pattern can be seen.
where K and A are constants that depend on the explosive should have a certain value. That value can be calculated from
charge type. For example, according to the value presented in the sound speed in water and the distance theoretically. The
Bradley (1991) for TNT, K ¼ 52:4, A ¼ 1:13. W is the weight sound speed in water can be calculated from the following
of the explosive charge in kilograms. R is the distance between empirical equation (Wilson, 1960):
the explosive charge and target (sensor) in meters.
The maximum pressure of shock wave was extracted from Table 2
Fig. 3 and summarized in Table 2. Relative difference was also Maximum pressure and relative difference.
calculated based on the value predicted from Eq. (2) for each Distance System No. 1 System No. 2 System No. 3 Equation (2)
measured pressure. It can be checked that, while DAQ system
2m 19.52 MPa 23.09 MPa 24.06 MPa 23.94 MPa
No. 1 has a relative difference of about 20%, DAQ systems 18.5% 3.6% 0.5%
No. 2 and No. 3 show relative difference about 5%. 4m 8.68 MPa 11.20 MPa 10.38 MPa 10.94 MPa
From Fig. 3, the time interval between peaks from values at 20.7% 2.4% 5.1%
each peak was summarized in Table 3. Since each sensor was 6m 6.27 MPa 6.85 MPa 6.79 MPa 6.92 MPa
9.4% 1.0% 1.9%
installed at interval of 2 m, the time interval between peaks
S.-J. Moon et al. / International Journal of Naval Architecture and Ocean Engineering 9 (2017) 589e597 593
Table 6 9:766 mV pffiffiffi.
Calibration information of pressure sensor and DAQ system No. 3. uðdDAQ;2 Þ ¼ 3 ð0:073 mV=kPaÞ ¼ 38:62 kPa
2
Corrective Uncertainty
value (ab. 95% level of
ð9Þ
confidence, k ¼ 2) The estimated uncertainty in Table 7 should be used to
Pressure Sensor 0 3% calculated the uncertainty of maximum pressure (pm ), which is
DAQ System No. 3 0 1 mV called combined uncertainty (uc ). If combined uncertainty
inputs x1 ; x2 ; /xN do not have any correlations e in other
words, if they can be obtained independently e combined
uncertainty can be obtained using Eq. (4) and Eq. (10) below.
as shown in Table 7. The standard uncertainty from repeated
measurements uðpm;0 Þ is calculated using the following Eq. XN
2 X
vf N
uc ðyÞ ¼ u2 ðxi Þ ¼ ½ci uðxi Þ ð10Þ
2 2
(6).
vxi
pffiffiffi i¼1 i¼1
u pm;0 ¼ std pm;0 n ð6Þ
uðxi Þ is the standard uncertainty of xi , and ci is called sensitive
Here, std and n are the standard deviation and number of coefficient. Using Eq. (10), the sensitive coefficient for each
repetitions, respectively. The degree of freedom is calculated measurement in Eq. (5) is calculated to be 1. On the other
from n number of independent measurements, and it is ðn 1Þ hand, the calculation of the D.O.F. for combined uncertainty is
for simple cases. As shown in Table 7, the uncertainty esti- known to be very complicated. It is derived by calculating the
mated through repetitive experiments is Type A uncertainty. effective degree of freedom (veff ) using the Welch-
Table 6 shows that the uncertainty for the pressure sensor is Satterthwaite formula (Satterthwaite, 1946) below.
expressed as the relative uncertainty, and that it is Type A
uncertainty evaluation. Using the average value of the u4c ðyÞ
veff ¼ ð11Þ
measured maximum pressured, its conversion into absolute X
N
½ci uðxi Þ
4
Table 7
Standard uncertainty of maximum pressure. Table 8
Distance Average Standard Standard D.O.F. Standard uncertainty of pressure sensor and DAQ system No. 3.
(m) (MPa) deviation uncertainty Distance (m) Average uðdsensor Þ uðdDAQ;1 Þ uðdDAQ;2 Þ
(MPa) (MPa) (MPa) (MPa) (MPa) (MPa)
2 23.30 0.553 0.247 4 2 23.30 0.350 0.007 0.039
4 10.19 0.370 0.166 4 4 10.19 0.153 0.007 0.039
6 7.01 1.042 0.466 4 6 7.01 0.105 0.007 0.039
8 4.72 0.261 0.117 4 8 4.72 0.071 0.007 0.039
S.-J. Moon et al. / International Journal of Naval Architecture and Ocean Engineering 9 (2017) 589e597 595
Table 9 Table 13
Measurement uncertainty for distance of 2 m. Measurement uncertainty for distance of 2 m.
pm;0 dsensor uðdDAQ Þ W R
Uncertainty 0.247 0.350 0.007 (MPa) 0.039 (MPa) Uncertainty 0.015 (kg) 0.030 (m)
(MPa) (MPa) Sensitive coefficient 9.0 13.5
Sensitive coefficient 1 1 1 1 D.O.F. ∞ ∞
D.O.F. 4 ∞ ∞ ∞ Combined uncertainty 0.428 (MPa)
Combined uncertainty 0.430 (MPa) Effective D.O.F. ∞
Effective D.O.F. 36 Expanded uncertainty 0.856 (MPa)
Expanded uncertainty 0.860 (MPa)
Table 10 Table 14
Expression of measurement uncertainty. Expression of measurement uncertainty on Eq. (2).
Distance (m) Average Uncertainty (MPa) Equation (2) Distance (m) Average (MPa) Uncertainty (MPa) (ab. 95%
(MPa) (ab. 95% level of (MPa) level of confidence, k ¼ 2)
confidence, k ¼ 2) 2 23.94 0.86
2 23.30 0.86 23.94 4 10.94 0.40
4 10.19 0.46 10.94 6 6.92 0.25
6 7.01 0.96 6.92 8 5.00 0.18
8 4.72 0.29 5.00
Table 11 Using Eq. (4), the model equation for maximum pressure
Calibration information of balancer and tape scale. estimation equation can be expressed as Eqs. (12)e(14) below.
Corrective Uncertainty (ab. 95% level of 1=3 A
Value confidence, k ¼ 2) W
pm ¼ K$ ð12Þ
Electronic Balancer (Weight) 0 3% R
Steel Scale (Distance) 0 3%
W ¼ W0 þ delectronic balancer ð13Þ
Table 12
Measurement uncertainty for explosive weight and distance of 2 m.
W0 delectronic balancer R0 dsteel scale
National Instruments Corporation, 2016a. Data Sheet NI 9234. http://www.ni. PCB Piezotronics Inc., 2016. Model W138A10/ICP Pressure Sensor Installa-
com (Accessed 10 December 2016). tion and Operation Manual. Document Number 21354, Revision B. http://
National Instruments Corporation, 2016b. Device Specification NI PXIe-5160. www.pcb.com (Accessed 10 December 2016).
http://www.ni.com (Accessed 10 December 2016). Satterthwaite, F.E., 1946. An approximate distribution of estimates of variance
National Instruments Corporation, 2016c. NI 449x Specifications. http://www. components. Biom. Bull. 2, 110e114.
ni.com (Accessed 10 December 2016). Wang, H., Zhu, X., Cheng, Y.S., Liu, J., 2014. Experimental and numerical
Park, I.K., Cho, D.S., Kim, J.C., 2003a. Development of measurement system investigation of ship structure subjected to close-in underwater shock wave
for the underwater explosion shock test of naval ships. J. Soc. Nav. Archit. and following gas bubble pulse. Mar. Struct. 39, 90e117.
Korea 40, 66e74 (in Korean). Wilson, W., 1960. Speed of sound in sea water as a function of T, P, and
Park, I.K., Kim, J.C., Ahn, C.W., Cho, D.S., 2003b. Measurement of naval ship salinity. J. Acoust. Soc. Am. 38, 641.
responses to underwater explosion shock loadings. Shock Vib. 10,
365e377.