Vous êtes sur la page 1sur 2

PRODUCT SPECIFICATIONS

Omni FTIR Multi-Gas CEMS


Continuous emissions monitoring system
for the measurement of 10 or more gases
The Thermo Scientific™ Omni FTIR
Multi-Gas CEMS is designed to
meet 40 CFR Part 60 standards by
using the state-of-the-art Thermo
Scientific™ Antaris™ IGS gas analyzer
to simultaneously monitor more
than ten gases with high precision,
sensitivity, speed and reliability within
a tough industrial environment.

Features
• Meets 40 CFR Part 60/63 standards
and HCl Performance Specifi cations The system is capable of simultaneously The Omni FTIR Multi Gas CEMS has the
monitoring 10 or more gases, using an unique ability to switch to a Flowback
• Leading-edge Antaris IGS analyzer
advanced software platform to calibrate mode that purges the sample stream all
for monitoring multiple gases, and analyze spectra with high sensitivity the way from the analyzer through the
including HCl and precision. The system supports sample line to the tip of the probe. This
automated dynamic spiking that can be feature enables clean and continuous
• Capable of serving as a Reference
remotely triggered through the plant operation of the system and eliminates
Method 321 for HCl
installed data acquisition system (DAS) or the likelihood of sample line contamination.
• Automated dynamic spiking triggered data control system (DCS). In addition, the
by plant’s DAS or DCS system is easily integrated with the DAS or The Omni FTIR CEMS uses a modular
DCS through MODBUS or 4-20 mA analog design for reliable trouble-free operation
• Unique flowback mode through outputs with digital I/O. in demanding conditions, requiring minimal
sampling system for continued method maintenance. The Antaris IGS
clean operation analyzer is a primary component of the
system and uses a dynamically aligned
Introduction interferometer, high precision mirrors, and
The Thermo Scientific Omni FTIR Multi-Gas a thermo-electrically cooled detector to
Continuous Emissions Monitoring System offer superior minimum detection limit
(CEMS) uses the fi eld-proven Thermo (MDL), selectivity, accuracy and speed.
Scientific Antaris Industrial Gas System This combination of speed and perfor-
(IGS) analyzer to deliver research-grade mance is ideal for detailed analysis of
continuous monitoring of multiple gases in rapidly changing complex gas mixtures
a robust package designed specifi cally to
in the stack.
meet 40 CFR Part 60 standards and related
performance specifi cations.

Designed to help users meet performance


specifications for gases such as Hydrogen
Chloride (HCl), the Omni FTIR CEMS can
alsobe employed as a Reference Method
321 for HCl.
Thermo Scientific Omni FTIR Multi Gas CEMS System components
Specifications The main components of the system are the
Linearity <2% of minimum measuring range
Antaris IGS analyzer, Sample Handling System,
Zero drift <3% of minimum measuring range
and an advanced software platform.
Span drift <2% of minimum measuring range Sample handling system
Influence of ambient <5% of minimum measuring range for Consists of a direct extractive probe, heated
temperature zero and span with 20°C as reference sample line (not included with the system), pump
Influence of sample assembly, air dryer, and purge gas generator.
None (pressure compensation)
gas pressure The probe assembly consists of a stainless steel/
Flow ~5 l/min hastelloy probe barrel and a heated filter assembly.
<3 minutes for all non-reactive gases; The entire sample stream, from the probe stinger
Response time T95 <5 minutes for NH3, HCl and HF spiking; to the analysis bench, is maintained at or above
<10 minutes for NH3, HCl and HF dry a temperature of 185°C to prevent sample loss or
Availability >98% condensation.
Instrument air 80-120 PSI, 5 SCFM Advanced software platform
Ambient temperature +60 to +95 °F/+15 to +35°C The PC gathers information from the FTIR and
Physical dimensions
22" (W) × 70" (H) × 36" (D); temperature controller and reports readings over
31" (W) × 70" (H) × 36" (D) (built system) analog and digital I/O lines, Ethernet, and Serial.
Weight ~ 440 lbs (200kg)
Sample line length Up to 100 feet
Applications
The Omni FTIR CEMS uses the hot-wet extractive
Power requirements 208v, 240 VAC, 60 AMPS
measurement technique and can be used in
Analog signals 4-20 mA for each measured concentration
various applications including coal-fi red power
FTIR data valid, temperature alarm for sample probe, heated lines, sample plants, cement kilns, waste incinerators, and to
conditioning assembly, and FTIR cell; Status measurement/maintenance; study industrial processes with multiple effluents.
Calibration Gas mode inputs and outputs: check span, check spike, check zero,
background and flowback, blowback, and sample; MODBUS over TCP
and RS232
Compound MDL 1 min.
Compound
(ppm) avg (ppm)
Thermo Scientific Antaris IGS Analyzer
™ ™
Carbon monoxide CO 0–2000 0.15
The Thermo Scientific Antaris IGS analyzer was specifically Nitric oxide NO (gas turbine) 0-500 0.40
developed to meet the needs of demanding gas applications.
Sulfur dioxide SO2 0– 1000 0.40
Developed as an industrial FTIR system that can be deployed
in either a rack-mount manufacturing environment or a table-top Hydrogen chloride HCl 0–100 0.10
quality control area, the Antaris IGS analyzer provides the Ammonia NH3 0–100 0.10
industry’s highest possible performance in calibration and Water H2O 0–40.0 vol% 0.1 vol%
stability, method transferability and high speed data acquisition.
Nitrogen dioxide NO2 0–50.0 0.10
Nitrous oxide N2O 0–50.0 0.10
Hydrogen fluoride HF 0–14.0 0.15
To maintain optimal product performance, you need immediate access to experts worldwide,
as well as priority status when your air quality equipment needs repair or replacement. Carbon dioxide CO2 0–45.0 vol% 0.2 vol%

We offer comprehensive, flexible support solutions for all phases of the product life cycle. Sulfur hexafluoride SF6 0–6.0 0.003
Through predictable, fixed-cost pricing, our services help protect the return on investment
and total cost of ownership of your Thermo Scientific products.

USA India China Europe


27 Forge Parkway C/327, TTC Industrial Area +Units 702-715, 7th Floor Ion Path, Road Three,
Franklin, MA 02038 MIDC Pawane Tower West, Yonghe Winsford, Cheshire CW73GA
Ph: (508) 520-0430 New Mumbai 400 705, India Beijing, China 100007 UK Ph: +44 1606 548700
Fax: (508) 520-2800 Ph: +91 22 4157 8800 Ph: +86 10 84193588 Fax: +44 1606 548711
orders.aqi@thermofisher.com india@thermofisher.com info.eid.china@thermofisher.com sales.epm.uk@thermofisher.com

Find out more at thermofisher.com/air

© 2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific
and its subsidiaries unless otherwise specified. EPM_OmniCEMS_DS_1117

Vous aimerez peut-être aussi