DST-i Oscilloscope Software

Instruction Manual
IMPORTANT Product Name: DST-i Oscilloscope Software
Number of License: One
NOTICE TO USER:
Please read the following License Agreement carefully. The License Agreement is a contract. This software product of “DST-i Oscilloscope
Software”, including any document related thereto, is protected by copyright laws and international treaties. You must agree to all of the terms and
conditions of the License Agreement before installing or using wholly or partially the product.
If you do not agree to any of the terms and conditions of the License Agreement, do not use this product.

License Agreement
This software product of ” DST-i Oscilloscope Software”, including any document related thereto (the “Software”) is the copyrighted work of DENSO
CORPORATION (“DENSO”) and its licensor. Under this Agreement, the Software is licensed to you by DENSO subject to the terms and conditions
below, not sold nor assigned to you by DENSO.
If you use the Software as a natural person, “you” herein means such natural person, and if you use the Software as an officer, employee, agent or other
member (including a partner of a partnership) of any legal entity, “you” herein means such legal entity.
This Agreement shall take effect as from the day you agree to all of the terms and conditions of the License Agreement.

1. Limited License
DENSO grants you a non-exclusive license to install the Software into one certain DST-i hardware in your possession or under your control
(“Hardware”), and to use the Software for the sole purpose of diagnosing the trouble of vehicle electronic control systems (“Purpose”).

2. Restrictions
(i) You shall not install the Software for any purpose other than Purpose, and shall not use the Software for any purpose other than Purpose.
(ii) You shall not use the Software with any hardware other than Hardware.
(iii) You shall not rent, lease, sell, sublicense, assign, or otherwise transfer the Software or this Agreement to any third party; provided, however, that
you may assign all of the Software and this Agreement, not part thereof, to your assignee perpetually subject to agreement of such assignee to all
of the terms and conditions of this Agreement.
(iv) You shall not copy, reverse engineering, decompile, disassemble, merge, modify or translate the Software. However, you may make one copy of
the Software for the sole purpose to make a backup of the Software.
(v) You shall not remove nor obscure DENSO’s copyrights, trademarks or other proprietary notices or legends from any of the materials of the
Software.
(vi) You shall not use the Software and/or technology provided by DENSO, or any other products, software and/or technology manufactured or
developed by using them for the purposes of disturbing international peace and security, including (1) the design, development, production,
stockpiling or use of weapons of mass destruction such as nuclear, chemical or biological weapons or missiles, (2) the other military activities, or
(3) any use supporting these activities.
(vii) You shall not sell, export, dispose of, license, rent, transfer, disclose or otherwise provide the Software to any third party, whether directly or
indirectly, with knowledge or reason to know that the third party or any other party will engage in the activities described above.
(viii) You shall not directly or indirectly, export, re-export, transship or otherwise transfer the Software in violation of any applicable export control
laws or regulations promulgated and administered by the governments of the countries asserting jurisdiction over the parties or their transactions.
(ix) You shall not illegally download the Software from official Webpage of DENSO.

3. Violation of this Agreement


Unauthorized install, copy or use of the Software may result in severe civil and criminal penalties, and will be prosecuted to the maximum extent
possible.
All your rights granted hereunder shall automatically terminate if you fail to comply with any terms and conditions of this Agreement. You shall
destroy or erase all of the Software from all media in your possession or under your control if you fail to comply with any terms and conditions of this
Agreement. The termination of this Agreement shall not prejudice any rights and remedies that DENSO may have against you.

4. Disclaimer of Warranty; Limitation of Liability


THE SOFTWARE IS PROVIDED “AS IS” WITHOUT ANY KIND OF WARRANTY OF ANY PURPOSE. THE SUPPLY OF THE SOFTWARE
OR THE GRANT OF RIGHTS HEREUNDER BY DENSO SHALL NOT IMPLY ANY WARRANTY AGAINST INFRINGEMENT OF
INTELLECTUAL PROPERTY RIGHTS HELD BY A THIRD PARTY. DENSO MAKES NO REPRESENTATIONS NOR WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
DENSO ASSUMES NO RESPONSIBILITY FOR THE APPLICATION OF, ERRORS NOR OMISSION IN THE SOFTWARE. IN NO EVENT
SHALL DENSO BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL, PUNITIVE OR OTHER
DAMAGES ARISING OUT OF OR RESULTING FROM ANY PART OF THE SOFTWARE, THE INSTALLATION OR COPY OF THE
SOFTWARE, THE USE OF THE SOFTWARE, OR INABILITY TO USE THE SOFTWARE. FURTHERMORE, DENSO SHALL NOT BE
LIABLE FOR ANY LOSS, DAMAGES OR COSTS ARISING OUT OF LOST PROFITS OR REVENUE, LOSS OF USE OF THE SOFTWARE,
LOSS OF DATA OR EQUIPMENT, COST OF RECOVERING THE SOFTWARE, DATA OR EQUIPMENT, THE COSTS OF SUBSTITUTE THE
SOFTWARE, MEDIA, DATA OR EQUIPMENT OR OTHER SIMILAR COSTS, AND SHALL BE FULLY INDEMNIFIED FROM ANY CLAIM
ASSERTED BY YOU OR ANY THIRD PARTY.

5. Governing Law
This Agreement shall be governed as to all matters including validity, construction and performance, by and under the laws of Japan, without reference
to its conflicts of law principles.

6. Arbitration
Any disputes arising out of this Agreement shall be finally settled by arbitration in accordance with the Rules of Conciliation and Arbitration of the
International Chamber of Commerce. The award of arbitration shall be final and binding upon the parties. Arbitration shall be held in Tokyo, Japan.

T02145Z
Preface

Preface
Thank you for purchasing "set with an oscilloscope".
Before using this product, read these instructions carefully so that you can use it correctly and safely.
The DST-i oscilloscope software should be the software for (hereinafter referred to as DST-i)
model with an oscilloscope.
DST-i model without an oscilloscope and hardware other than DST-i cannot be used.
An SD memory card (optional) is required to use the DST-i oscilloscope software.

-i-
Safety Instructions

Safety Instructions
This product is intended for the use by a properly trained automotive technician. The following safety
instructions should be adhered to prevent harm to the user and/or the vehicle.
Since a variety of work procedures, technologies, tools, parts, etc. as well as technician’s skill are used and
involved in the diagnosis and maintenance/service of vehicle, various results are predicted and it is
impossible to provide a comprehensive set of advice and safety messages that can cover all cases.
Therefore, it is the responsibility for an automotive technician to have an adequate knowledge of the
diagnosis system. It is also important to perform work appropriately and to ensure your safety and the safety
of those around you, as well as to maintain the safety of the vehicle to be diagnosed and the relevant
devices to perform appropriate diagnosis and maintenance.
The precondition for use of the DST-i is that a user has sufficient knowledge of vehicle diagnostics and the
operation of the DST-i to safely use this product.

This document uses symbols for warnings, cautions, and prohibitions. These symbols and their meanings are as follows.
Please make sure you fully understand the meanings of these symbols before reading the rest of this document.

Indicates an item for which incorrect handling can result in a major


WARNING accident involving death or serious injury.

Indicates an item for which incorrect handling can lead to injury or


CAUTION damage to property. Under certain conditions, more serious
consequences may result.

PROHIBITED Indicates a prohibited method of handling.

MANDATORY Indicates a mandatory method of handling.

- ii -
Safety Instructions

WARNING and CAUTION for handling this product

WARNING
Perform diagnostics and repair work according to the precautions included in "For safe
diagnostics".
Reference: Page v For safe diagnostics (Safety Instructions)
Refer to safety instructions and diagnosis procedures for the vehicle to be diagnosed and to those
provided by the manufacturer of devices before using this product, and adhere to those
instructions.
Failure to obey the precautions could result in an accident.
Do not perform any work while the vehicle is being driven.
Doing so could result in an accident.
Route the cable so it does not tangle with the technician or the operation control unit.
Failure to do so could result in an accident.
Always observe the following rules. Failure to do so can result in abnormal heat generation, fire,
explosion, or electric shock.
- Do not disassemble or alter this product.

- Do not connect this product to anything with voltage exceeding the ratings of this product.

- Do not connect the probe to any parts with voltage exceeding the ratings of this product.

- iii -
Safety Instructions

CAUTION
Use wheel chocks to block the wheels prior to working on the vehicle.
Failure to do so could result in an accident.
When working in low visibility areas, for example under the vehicle, always make the vehicle
ignition switch OFF to ensure that the vehicle cannot be moved.
Failure to do so could result in an accident.
When you start the engine or move the vehicle, make sure everyone is clear of the vehicle.

More than 100 V is generated in the ECU and injector. Be careful not to be shocked by this high
voltage.
Before removing any parts, disconnect the negative battery cable.

Turn the vehicle ignition switch OFF before connecting or disconnecting the connector or the
electric terminal of the vehicle unless otherwise specified.
Failure to do so may cause damage on the vehicle side electric circuit.
Wear safety glasses and protective clothing during work near any rotating parts.
Parts can be thrown out by the rotating engine, resulting in an accident.
Do not remove the radiator cap unless the engine is cool.
Doing so will result in being splashed with high temperature and pressurized engine coolant.
Do not touch high temperature areas such as exhaust system, manifold, engine, radiator, etc. with
bare hands.
Failure to do so may cause a burn injury.
Wear gloves when touching and handling the high-temperature engine and parts.

Do not route the cable of this product over the engine compartment, especially if the engine is
running.
Failure to follow this instruction could result in an accident where the cable and your clothes are caught in
the belt and pulley.
Do not work anywhere that water could come in contact with the equipment.

Do not drop this tool or cause it to have a strong impact.

Do not lift or drag the DST-i by holding the cable connected to the DST-i. Hold the body of the
DST-i when moving or installing it.
Do not apply any load causing stress to the connector connected to this product.
Failure to do so could result in an accident.
Before use, inspect for foreign material or pin breakage in each connector of this product.

- iv -
Safety Instructions

For safe diagnostics

Do not place any metal tools on the battery.

Do not create any sparks near the battery.


Failure to do so may cause ignition of battery gas.
Keep a lit cigarettes, sparks, open flames and other ignition sources away from the vehicle and
the battery.
Be sure to turn the ignition switch OFF, turn the headlight and other accessories OFF before
disconnecting the negative battery cable.
Remove personal accessories such as rings, bracelets and watches prior to working near the
battery.
Failure to do so could result in an accident.
Technicians handling the battery and technicians near the battery must wear safety glasses and
protective gloves.
If splashed with battery fluid, wash the effected area with soapy water for 10 minutes, then
proceed to a medical facility for further treatment.
Do not rub or touch your eye directly while work near the battery.
Failure to do so may cause a burn injury on your eye and skin from the battery fluid.
Do not allow connection between the "+" and "-" of the battery terminals with jumper cable/wire or
a tool.
Do not ground an electric terminal that is or may be a part of the electric system.
Failure to do so may cause the damage of vehicle side electric circuit.
Make sure the working area has exhaust gas ventilation or is well ventilated to prevent exhaust
gas poisoning.
The engine exhaust gas contains odorless lethal gas, which may cause death or severe illness due to
breathing the exhaust.
Do not use this product in an environment that accumulates explosive vapor, such as an
underground location or a closed room.
Do not smoke, strike a match or use a lighter during work.
Failure to do so may cause ignition of battery gas or explosive gas.
Do not use a flammable spray or a flushing spray during usage of the DST-i.

Have a fire extinguisher capable of extinguishing a gasoline, chemical or electrical fir in case of an
accidental fire.
All technicians must wear safety glasses and protective clothing.
Fuel, oil vapor, high temperature vapor, poisonous emission gas, acid, refrigerant and other foreign
substances may be emitted due to failure and malfunction of the vehicle system.
When connecting or disconnecting the oscilloscope probe or ground cable to this product, ensure
correct orientation of the connector and gently insert it.
If the connector is connected in the wrong orientation, inserted or pulled out at the wrong angle, there is a
risk of damaging the connector terminal and this may cause the DST-i to malfunction.
Do not pull on the cable, hold on the connector section to disconnect the cable.
Pulling on the cord section can break the wires in the cable.

-v-
Table of contents

Table of contents
1 Before Use
1-1 Basic operation ..................................................................................................................................... 1
1-2 Starting.................................................................................................................................................. 1
1-3 Ending................................................................................................................................................... 7

2 Functions of the DST-i Oscilloscope Software


2-1 Functions .............................................................................................................................................. 8
Waveform observation functions ............................................................................................................. 9
Long duration data measurement functions ............................................................................................ 9

3 Operation of the DST-i Oscilloscope Software


3-1 Waveform observation ........................................................................................................................ 10
Screen configuration ............................................................................................................................. 10
Stopping/Starting ................................................................................................................................... 11
Time range setting................................................................................................................................. 12
Setting the channel (waveform display/non-display switching, ground position)................................... 13
Voltage range setting............................................................................................................................. 14
Trigger setting ....................................................................................................................................... 15
3-2 Long duration data measurement ....................................................................................................... 18
Long duration data measurement flow .................................................................................................. 18
Screen configuration ............................................................................................................................. 18
Trigger mode setting ............................................................................................................................. 19
Time range setting................................................................................................................................. 20
Setting the channel (waveform display/non-display switching, ground position)................................... 21
Voltage range setting............................................................................................................................. 22
Setting conditions for recording long duration data measurement ........................................................ 23
Start recording ....................................................................................................................................... 25
<To record without a start trigger>..................................................................................................... 25
<To record with a start trigger>.......................................................................................................... 26
Stop recording ....................................................................................................................................... 29
3-3 Saving data ......................................................................................................................................... 30
3-4 Retrieving data.................................................................................................................................... 34
3-5 Deleting data....................................................................................................................................... 40
3-6 Zero point correction ........................................................................................................................... 45
3-7 Save confirmation message display/non-display setting .................................................................... 47
3-8 Termination of the oscilloscope........................................................................................................... 49

4 Troubleshooting
4-1 If the screen freezes while using......................................................................................................... 50
4-2 If the system message is displayed .................................................................................................... 50

- vi -
1 Before Use

1 Before Use
1-1 Basic operation
The DST-i has six operation buttons.

Up button
Left button
Right button

Down button

Guidance area B button A button


T02153E

The function of each button varies depending on the screen.


In the lower portion of the screen is a guidance area explaining how to operate the system.
Operate using the information displayed in the guidance area.

1-2 Starting
1. Insert an SD memory card installed with DST-i oscilloscope software to the DST-i unit.

NOTE
Refer to the "Instruction manual for DST-i hardware with oscilloscope" for the details of the SD
memory card.
Refer to the following download site for how to install DST-i oscilloscope software.
http://www.ds3.denso.co.jp/dst-i/setup/

Compatible cards
Notch to be on SD SDHC
the right side

Confirm the logo mark of


the card.
T02585E

-1-
1 Before Use

2. Connect cables used for oscilloscope functions to DST-i.

NOTE
Refer to the "Instruction manual for DST-i hardware with oscilloscope" for how to connect
cables.
When using the DST-i oscilloscope software, do not use the AC/DC adapter.
Otherwise, the correct value is not measured.

Cap
The tip of the probe can be removed.
Tip of the probe

Ch1 Ch2
Oscilloscope probe Do not use the AC/DC adapter.

Connect the oscilloscope probe to Ch1 or Ch2.


The probe can be connected to both Ch1 and Ch2.
T02719E

Connect to vehicle ground

Connect when the waveform is unstable due to electronic noise.


T02235E

<Optional products>

Ch1 Ch2
About the guides

<DST-i side> <4-channel adapter side>

4-channel adapter

Guide (male) Guide (female)

Line up the male and female portions of the guide


when connecting the connector.
Connect the 4-channel adapter to Ch2. *It cannot be connected to Ch1.
The connector may only be connected to Ch2. T02059E

-2-
1 Before Use

3. Connect DST-i and the vehicle side diagnosis connector with a datalink cable.

Connected to vehicle side diagnosis connector

Notch (Center)

Tighten with screws. T02005E

CAUTION
Do not use a datalink cable other than those intended for DST-i.
Check the connector for adhesion of foreign substances or breakage of connector pins before
connecting it.
When connecting the datalink cable to DST-i and the vehicle side diagnosis connector, pay
attention to the connector direction, and connect it straight and gently.
Connecting the cable in an incorrect direction or inserting it at an angle may damage the connector
terminal and cause a malfunction of the vehicle or DST-i.
Do not apply a load to the connector connected to DST-i.
Applying a load may damage the connector terminal and cause a malfunction of the vehicle or DST-i.

-3-
1 Before Use

4. Turn the DST-i mode switch ON.


The opening screen is displayed.

Mode switch

T01897E

T02419E
Opening screen

NOTE
Turning the mode switch ON makes the power indicator of DST-i turned to green.

-4-
1 Before Use

5. Press any button while the opening screen ON.


The main menu screen is displayed.

㪤㪸㫀㫅㩷㪤㪼㫅㫌 

1UEKNNQ 2%1UEKNNQ 5GVVKPIU

㧦5GNGEV‫ޓ‬$$CEM‫ޓ‬#1-㧔‫ޓ‬㧕
T02587E
Main menu screen

NOTE
The main menu screen is not displayed when an SD memory card with the oscilloscope
software installed is not inserted.
When pressing any button shows the error screen, refer to the following page.
Reference: Page 6 When the error screen is shown (Chapter 1 Before Use/Starting)

The following table shows the functions that can be selected in the main menu screen.
Function Description
Observes waveform of signals inputted to each channel on the DST-i screen, when
Oscillo
an oscilloscope probe is connected to DST-i.
Observes waveform of signals inputted to each channel on the PC screen, when an
PCOscillo
oscilloscope probe is connected to DST-i.
Settings Changes and customizes DST-i settings.

6. Select "Oscillo" in the main menu screen, and press the "A" button.
The waveform observation screen is displayed.

㪤㪸㫀㫅㩷㪤㪼㫅㫌 

1UEKNNQ 2%1UEKNNQ 5GVVKPIU

㧦5GNGEV‫ޓ‬$$CEM‫ޓ‬#1-㧔‫ޓ‬㧕
T02587E T02322E
Main menu screen Waveform observation screen

-5-
1 Before Use

When the error screen is shown


If the error screen is displayed when one of the buttons is pressed in the opening screen, or when "Oscillo" is
selected and the "A" button is pressed in the main menu screen, these are the possible causes for this.
Error code

Error screen (In opening screen) Error screen (In main menu screen)
T02590E

NOTE
Each error code in the lower right of the screen is assigned to the following causes.

Error code: 00000002


Possible cause Check Item
The SD memory card does not have DST-i Install DST-i oscilloscope software to the SD memory
oscilloscope software installed. card.
Format the SD memory card and re-install DST-i
oscilloscope software.
The SD memory card is defective.
If the error screen is still shown, use another SD
memory card.
Error code: 00000006
Possible cause Check Item
Use the DST-i that was used for installing DST-i
The combination of DST-i and SD memory oscilloscope software.
card is incorrect. Or re-install DST-i oscilloscope software to the SD
memory card.

NOTE
If any other error codes are displayed, please contact your distributor.

-6-
1 Before Use

1-3 Ending
1. Close the DST-i oscilloscope software.

2. Turn the DST-i mode switch OFF.

3. Remove the datalink cable from DST-i and the vehicle side diagnosis connector.

CAUTION
When removing the datalink cable from DST-i and the vehicle side diagnosis connector, pull it
out of the connector straight and gently.
Pulling out the cable at an angle may damage the connector terminal and cause a malfunction of the
vehicle or DST-i.
When pulling out the cable, always pull it out holding the connector, not the cord.
Pulling out the cable holding the cord may disconnect the cable.

4. Remove the cables used for the oscilloscope functions from DST-i.

NOTE
Refer to the "Instruction manual for DST-i hardware with oscilloscope" for how to disconnect
the cables.

-7-
2 Functions of the DST-i Oscilloscope Software

2 Functions of the DST-i Oscilloscope Software


2-1 Functions
The DST-i oscilloscope software has two functions: "Waveform observation" and "Long duration data measurement".
Function Waveform observation Long duration data measurement
Allows observation of signal waveform as Long duration waveform measurement
Details of
an oscilloscope and saving the observation data can be recorded and the recorded
function
data. data can be saved.
Trigger mode (Time range):
Setting • AUTO(500 ns/div to 5 ms/div) Trigger mode (Time range):
conditions • NORMAL • AUTO(10 ms/div to 10 s/div)
• SINGLE
Long duration displaying (as per recording
Screen 1-screen display (real time displaying) time setting)
Recording time: 1 min. to 40 hrs.

Operation flow
Set a trigger mode.

"NORMAL" "SINGLE" "AUTO"

A trigger is set. Time range is set.

"500 ns/div to 5 ms/div" "10 ms/div to 10 s/div"

Waveform Long duration


observation data measurement

Waveform observation data is saved. Recording conditions are set.


Recording time setting
Pretime setting
If above parameters are not set, they will be
automatically set depending on the time range.

A trigger is set.

A trigger is set. A trigger is not set.

Recording long duration data measurement starts.

Long duration measurement data is saved. T02417E

-8-
2 Functions of the DST-i Oscilloscope Software

Waveform observation functions


The following table shows operation and setting of waveform observation functions.
Function Operation and setting Refer to:
Stop/Start Page 11
Time range setting Page 12
Waveform display/non-display switching Page 13
Waveform
Ground position setting Page 13
observation
Voltage range setting Page 14
Trigger setting Page 15
Saving data Page 30

Long duration data measurement functions


The following table shows operation and setting of long duration data measurement functions.
Function Operation and setting Refer to:
Recording time setting Page 23
Long duration data Pretime setting Page 24
measurement Trigger setting Page 26
Saving data Page 30

-9-
3 Operation of the DST-i Oscilloscope Software

3 Operation of the DST-i Oscilloscope Software

NOTE
The following explanation uses a screen that observes 4 channel waveforms using a 4-channel adapter
set.
The default condition of this product allows waveforms for 2 channels only.

3-1 Waveform observation


This screen is used to observe the inputted waveforms.

Screen configuration
The following figures show screen examples and the parts of the screen.

Trigger position: T mark

Channel 4 ground position


Trigger position display
Channel 3 ground position Trigger channel display

Channel 2 ground position Trigger level display


Trigger level: Triangular mark
Channel 1 ground position Trigger slope: Arrow mark
Trigger slope display
Trigger mode display
Time range display display Menu

Selection items
Channel 4 voltage range display
Channel 1 display
Channel 4 display
Channel 1 voltage range display
Channel 3 voltage range display
Channel 2 display
Channel 2 voltage range display Channel 3 display
T02323E

- 10 -
3 Operation of the DST-i Oscilloscope Software

Stopping/Starting
1. Press the "B" button in the waveform observation screen.
Waveform observation stops and the data replay screen is displayed.

T02322E T02324E
Waveform observation screen Data replay screen

2. Press the "B" button in the data replay screen.


The waveform observation screen in the suspended observation is displayed.

T02325E
Waveform observation screen
(Observation suspended)

NOTE
Wave displayed in the data replay screen can be saved.
Refer to the following for the data saving.
Reference: Page 30 Saving data (Chapter 3 Operation of the DST-i Oscilloscope Software)

3. Press the "B" button in the waveform observation screen while observation is suspended.
Waveform observation resumes.

T02514E
Waveform observation screen
(Resume observation)

- 11 -
3 Operation of the DST-i Oscilloscope Software

Time range setting

1. Select the time range display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Select a time range value with the right/left button, and press the "A" button.
The value you have set is displayed in the time range display.

T02326E
Waveform observation screen
(Time range display)

Operation buttons
Right button : Increases the time range value in increments of one step.
Left button : Decreases the time range value in increments of one step.

NOTE
By pressing the up/down button, you can change the time range value in increments of three
steps.
The value increases with the up button, and decreases with the down button.

- 12 -
3 Operation of the DST-i Oscilloscope Software

Setting the channel (waveform display/non-display switching, ground position)

1. Select the channel display from the lower selection items of the waveform observation screen, and
press the "A" button.

2. Select display/non-display of waveforms with the right/left button, and press the "A" button.
When you select OFF, OFF is displayed in the channel display.

NOTE
Set "ON" to show waveforms and "OFF" to hide them.

T02328E
Waveform observation screen
(Channel display)

Operation buttons
Right button : Sets the waveform display/non-display switching to OFF.
Left button : Sets the waveform display/non-display switching to ON.

3. When you set the waveform display/non-display switching to ON, specify the ground position (0 V
position) with the up/down button, and press the "A" button.

T02440E
Waveform observation screen
(Ground position)

Operation buttons
Up button : Increases the ground position. (Only when the waveform display/non-display
switching is set to ON)
Down button : Decreases the ground position. (Only when the waveform display/non-display
switching is set to ON)

- 13 -
3 Operation of the DST-i Oscilloscope Software

Voltage range setting

1. Select the voltage range display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Select a voltage range value with the up/down button, and press the "A" button.
The value you have set is displayed in the voltage range display.

T02329E
Waveform observation screen
(Voltage range display)

Operation buttons
Up button : Increases the voltage range value in increments of one step. ("OFF" is added to
the end.)
Down button : Decreases the voltage range value in increments of one step. ("OFF" is added to
the end.)

NOTE
By pressing the right/left button, you can change the voltage range value in increments of
three steps.
The value increases with the right button, and decreases with the left button. ("OFF" is added to the
end.)
When you set the voltage range to "OFF", the signal of the channel is not displayed.

- 14 -
3 Operation of the DST-i Oscilloscope Software

Trigger setting
Various settings related to trigger can be made.
• Trigger mode setting
• Trigger channel setting
• Trigger level setting
• Trigger slope setting
• Trigger position setting

Trigger mode setting


Trigger mode can be set.
Setting items
Item Description
Renews the waveform after the designated trigger conditions are detected or after a
AUTO certain time elapses. Setting the time range from 10 ms/div to 10 s/div allows long
duration data measurement.
A waveform is displayed every time the designated trigger conditions are met.
NORMAL Under the conditions that no triggers can be set, no waveform will be shown. In the
upper of the screen, "WAITING" is displayed.
Stops the trigger after displaying a waveform when designated trigger conditions are
first met.
SINGLE
Under the conditions that no triggers can be set, no waveform will be shown. In the
upper of the screen, "WAITING" is displayed.

1. Select the trigger mode display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Select a trigger mode with the right/left button, and press the "A" button.
The mode you have selected is displayed in the trigger mode display.

T02331E
Waveform observation screen
(Trigger mode display)

Operation buttons
Right button : Moves the mode selection cursor to the right.
Left button : Moves the mode selection cursor to the left.

- 15 -
3 Operation of the DST-i Oscilloscope Software

Trigger channel setting


Trigger channels are switched over.

1. Select the trigger channel display from the lower selection items of the waveform observation
screen, and press the "A" button.

2. Select a channel you want to set the trigger conditions with the right/left button, and press the "A"
button.
The channel you have selected is displayed in the trigger channel display.

T02330E
Waveform observation screen
(Trigger channel display)

Operation buttons
Right button : Moves the channel selection cursor to the right.
Left button : Moves the channel selection cursor to the left.

Trigger level setting


A trigger level for a selected trigger channel is set.

1. Select the trigger level display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Specify a trigger level with the up/down button, and press the "A" button.
The trigger level you have set is displayed in the right side of the screen with a triangular mark.

Trigger level: Triangular mark

T02332E
Waveform observation screen
(Trigger level display)

Operation buttons
Up button : Increases the trigger level.
Down button : Decreases the trigger level.

- 16 -
3 Operation of the DST-i Oscilloscope Software

Trigger slope setting


A trigger slope for selected trigger channels are switched over.

1. Select the trigger slope display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Select rise/fall of the trigger slope with the right/left button, and press the "A" button.
Marks of the rise/fall are displayed in the trigger slope display and at the right of the screen.

Trigger slope :‫ޓ‬


A rise mark or a fall mark.

T02333E
Waveform observation screen
(Trigger slope display)

Operation buttons
Right button : Moves the slope selection cursor to the right.
Left button : Moves the slope selection cursor to the left.

Trigger position setting


Trigger position for selected trigger channels is set.

1. Select the trigger position display from the lower selection items of the waveform observation
screen, and press the "A" button.

2. Specify a trigger position with the right/left button, and press the "A" button.
The trigger position you have set is displayed in the upper part of the screen with a T mark.
Trigger position: T mark

T02334J
Waveform observation screen
(Trigger position display)

Operation buttons
Right button : Moves the trigger position to the right.
Left button : Moves the trigger position to the left.

- 17 -
3 Operation of the DST-i Oscilloscope Software

3-2 Long duration data measurement


Long duration waveform measurement data can be recorded.
Setting trigger mode to AUTO and time range from 10 ms/div to 10 s/div allows measuring a long duration data
measurement.
Data obtained by pressing the recording start mark cannot be displayed as a waveform after resetting the time range
while replaying a data.

Long duration data measurement flow

1. Set the trigger mode to "AUTO". Refer to : Trigger mode setting

2. Set the time range from 10 ms/div to 10 s/div. Refer to : Time range setting

3. Set the long duration data measurement conditions. Refer to : Long duration data measurement conditions

4. Set the trigger. Refer to : Trigger setting

5. Start recording long duration data measurement. Refer to : Starting recording

6. Save the long duration measurement data. Refer to : Data saving


T02263E

Screen configuration
The following figures show screen examples and the parts of the screen.

Channel 4 ground position


Trigger position display
Channel 3 ground position Trigger channel display

Channel 2 ground position Trigger level display

Channel 1 ground position


Trigger slope display
Trigger mode display
Time range display Menu

Selection items
Channel 4 voltage range display
Channel 1 display
Channel 4 display
Recording start mark Channel 3 voltage range display
Recording time Pretime display
display Channel 3 display
Channel 1 voltage range display Channel 2 display Channel 2 voltage range display

T02355E

- 18 -
3 Operation of the DST-i Oscilloscope Software

Time range, Sampling rate, Recording time and measurement data file size
Sampling rate Maximum recording Maximum file size
Time range
(Fixed) time (per 1 recording) (per 1 recording)
10 s/div 32 S/s 40 hrs. Approx. 20 MB
5 s/div 64 S/s 40 hrs. Approx. 35 MB
2 s/div 160 S/s 40 hrs. Approx. 90 MB
1 s/div 320 S/s 40 hrs. Approx. 180 MB
500 ms/div 640 S/s 40 hrs. Approx. 360 MB
200 ms/div 1.6 kS/s 20 hrs. Approx. 440 MB
100 ms/div 3.2 kS/s 10 hrs. Approx. 440 MB
50 ms/div 6.4 kS/s 5 hrs. Approx. 440 MB
20 ms/div 16 kS/s 2 min. (ch.2: 4 min.) Approx. 8 MB
10 ms/div 32 kS/s 1 min. (ch.2: 2 min.) Approx. 8 MB

NOTE
The sampling rate and the recording time differs depending on the time range to be set.
Time can be record differs depending on the sampling rate.
The more detail that is recorded, the shorter the maximum recording time.
Use PC oscilloscope software for recording the further detailed data.

Trigger mode setting


1. Select the trigger mode display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Select "AUTO" with the right/left button, and press the "A" button.
"AUTO" is displayed in the trigger mode display.

T02441E
Waveform observation screen
(Trigger mode display)

Operation buttons
Right button : Moves the mode selection cursor to the right.
Left button : Moves the mode selection cursor to the left.

- 19 -
3 Operation of the DST-i Oscilloscope Software

Time range setting


1. Select the time range display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Select the value from 10 ms/div to 10 s/div with the right/left button, and press the "A" button.
The value you have set is displayed in the time range display.

T02442E
Waveform observation screen
(Time range display)

Operation buttons
Right button : Increases the time range value in increments of one step.
Left button : Decreases the time range value in increments of one step.

NOTE
By pressing the up/down button, you can change the time range value in increments of three
steps.
The value increases with the up button, and decreases with the down button.

- 20 -
3 Operation of the DST-i Oscilloscope Software

Setting the channel (waveform display/non-display switching, ground position)

1. Select the channel display from the lower selection items of the waveform observation screen, and
press the "A" button.

2. Select display/non-display of waveforms with the right/left button, and press the "A" button.
When you select OFF, OFF is displayed in the channel display.

NOTE
Set "ON" to display waveforms and "OFF" to hide them.

T02443E
Waveform observation screen
(Channel display)

Operation buttons
Right button : Sets the waveform display/non-display switching to OFF.
Left button : Sets the waveform display/non-display switching to ON.

3. When you set the waveform display/non-display switching to ON, specify the ground position (0 V
position) with the up/down button, and press the "A" button.

T02444E
Waveform observation screen
(Ground position)

Operation buttons
Up button : Increases the ground position. (Only when the waveform display/non-display
switching is set to ON)
Down button : Decreases the ground position. (Only when the waveform display/non-display
switching is set to ON)

- 21 -
3 Operation of the DST-i Oscilloscope Software

Voltage range setting

1. Select the voltage range display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Select a voltage range value with the up/down button, and press the "A" button.
The value you have set is displayed in the voltage range display.

T02445E
Waveform observation screen
(Voltage range display)

Operation buttons
Up button : Increases the voltage range value in increments of one step. ("OFF" is added to
the end.)
Down button : Decreases the voltage range value in increments of one step. ("OFF" is added to
the end.)

NOTE
By pressing the right/left button, you can change the voltage range value in increments of
three steps.
The value increases with the right button, and decreases with the left button. ("OFF" is added to the
end.)
When you set the voltage range to "OFF", the signal of the channel is not displayed.

- 22 -
3 Operation of the DST-i Oscilloscope Software

Setting conditions for recording long duration data measurement


This screen allows setting of various conditions for long duration data measurement.

NOTE
You can set the following when observing waveforms.

Recording time setting


The recording time can be set.

1. Select the recording time display from the lower selection items of the waveform observation
screen, and press the "A" button.

2. Select the recording time with right/left buttons, and press the "A" button.
The time you have set is displayed in the recording time display.

T02446E
Waveform observation screen
(Recording time display)

Operation buttons
Right button : Increases the recording time value in increments of one step.
Left button : Decreases the recording time value in increments of one step.

NOTE
By pressing the up/down buttons, you can change the voltage range value in increments of
three steps.
The value increases with the up button, and decreases with the down button.

- 23 -
3 Operation of the DST-i Oscilloscope Software

Pretime setting
A long duration data measurement starts when the recording start mark is pressed, or when the trigger conditions
are met as they are set.
This software is able to record data taken before the start of recording. The data located before the start of
recording is called a "pretrigger".
The recording time for a pretrigger is referred to as "pretime" and the "pretime" is a settable parameter. This is
called "pretime setting".

1. Select the pretime display from the lower selection items of the waveform observation screen, and
press the "A" button.

2. Select the pretime with the right/left button, and press the "A" button.
The value you have set is displayed in the pretime display.

T02447E
Waveform observation screen
(Pretime display)

Operation buttons
Right button : Increases the pretime in increments of one step.
Left button : Decreases the pretime in increment of one step.
("AUTO" is displayed at the end.)

NOTE
If "AUTO" is selected as pretime, the time per one screen (12 frames of the set time range)
will be pretime.
In case of the data having pretrigger less than 1 sec. in AUTO, "00h00m01" is displayed on
the upper right of the screen.
The pretime that can be set differ depending on the time range.

- 24 -
3 Operation of the DST-i Oscilloscope Software

Start recording
Long duration measurement data can be recorded to an SD memory card.

NOTE
Data saving methods
• To record without a start trigger
• To record with a start trigger

<To record without a start trigger>

1. Select "OFF" in the trigger level display from the lower selection items of the waveform observation
screen, and press the "A" button.

T02515E
Waveform observation screen

2. Select the recording start mark in the lower left part of the waveform observation screen, and press
the "A" button.
Recording of long duration data measurement is started, and "REC" is displayed in the upper part of the
screen.

T02448E
Waveform observation screen

NOTE
For data obtained by pressing the recording start mark, waveforms cannot be displayed after
resetting the time range while replaying a data.

- 25 -
3 Operation of the DST-i Oscilloscope Software

<To record with a start trigger>


Various settings related to trigger can be made.
• Trigger channel setting
• Trigger level setting
• Trigger slope setting

1. Select "ON" in the trigger level display from the lower selection items of the waveform observation
screen, and press the "A" button.

T02479E
Waveform observation screen

Trigger channel setting


Trigger channels are switched over.

1. Select the trigger channel display from the lower selection items of the waveform observation
screen, and press the "A" button.

2. Select a channel you want to set the trigger conditions for with the right/left button, and press the
"A" button.
The channel you have selected is displayed in the trigger channel display.

T02464E
Waveform observation screen
(Trigger channel display)

Operation buttons
Right button : Moves the channel selection cursor to the right.
Left button : Moves the channel selection cursor to the left.

- 26 -
3 Operation of the DST-i Oscilloscope Software

Trigger level setting


A trigger level for a selected trigger channel is set.

1. Select the trigger level display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Specify a trigger level with the up/down button, and press the "A" button.
The trigger level you have set is displayed in the right side of the screen with a triangular mark.

Trigger level: Triangular mark

T02465E
Waveform observation screen
(Trigger level display)

Operation buttons
Up button : Increase the trigger level.
Down button : Decrease the trigger level.

Trigger slope setting


A trigger slope for selected trigger channels are switched over.

1. Select the trigger slope display from the lower selection items of the waveform observation screen,
and press the "A" button.

2. Select rise/fall of the trigger slope with the right/left button, and press the "A" button.
Marks of the rise/fall are displayed in the trigger slope display and at the right of the screen.

Trigger slope :‫ޓ‬


A rise mark or a fall mark.

T02466E
Waveform observation screen
(Trigger slope display)

Operation buttons
Right button : Moves the slope selection cursor to the right.
Left button : Moves the slope selection cursor to the left.

- 27 -
3 Operation of the DST-i Oscilloscope Software

Start recording

1. Select the recording start mark in the lower left part of the waveform observation screen, and press
the "A" button.
The system is now in the state of waiting for detecting a trigger, and "WAITING" is displayed in the upper
part of the screen.

T02450E
Waveform observation screen
Upon detecting a condition of a trigger for starting measurement, recording of measurement data
starts, and "REC" is shown in the upper part of the screen.

T02451E
Waveform observation screen

- 28 -
3 Operation of the DST-i Oscilloscope Software

Stop recording

1. After the set recording time elapses, the screen turns into the data replay screen.
The recorded data is tentatively named "temp".

T02452E
Data replay screen

NOTE
Even before reaching the recording time set, pressing the "A" button can save the data
recorded up to the time the button is pressed.

T02453E

If you do not save the data, press the "B" button in the data replay screen, select "NO" in the
save confirmation message screen and press the "A" button.
The screen returns to the waveform observation screen.
If the save confirmation message screen is not set to be displayed, it will not be displayed.

T02454E

- 29 -
3 Operation of the DST-i Oscilloscope Software

3-3 Saving data


You can save a screen image and a long duration measurement data.

Saving screen images

NOTE
Waveforms of saved screen image are not scalable.

1. Select "SAVE" from the lower selection items of the data replay screen, and press the "A" button.
The save menu screen is displayed.

T02455E T02336E
Data replay screen Save menu screen

- 30 -
3 Operation of the DST-i Oscilloscope Software

2. Select "SCREEN COPY" in the save menu screen, and press the "A" button.
The data is saved and the saving completed screen is displayed.

T02337E T02338E
In-process screen Saving completed screen

NOTE
The following message is displayed when the file name you want to save already exists.
Selecting "OK" and pressing the "A" button overwrites the file.
Selecting "BACK" and pressing the "A" button returns the screen to the save menu screen.
* You cannot save data during observation.

T02339E

3. Press the "A" button in the saving completed screen.


The screen returns to the save menu screen.

NOTE
File names are automatically set with "SC + Serial numbers".
You can save up to 100 files.

- 31 -
3 Operation of the DST-i Oscilloscope Software

Saving long duration measurement data

NOTE
Waveforms of saved long duration measurement data are scalable.

1. Select "SAVE" from the lower selection items of the long duration measurement data replay screen,
and press the "A" button.
The save menu screen is displayed.

T02456J T02340E
Long duration measurement data replay screen Save menu screen

2. Select "DST-i OSCILLO DATA" in the save menu screen, and press the "A" button.
The data is saved and the saving completed screen is displayed.

T02337E T02341E
In-process screen Saving completed screen

NOTE
The following message is displayed when the file name you want to save already exists.
Selecting "OK" and pressing the "A" button overwrites the file.
Selecting "BACK" and pressing the "A" button returns the screen to the save menu screen.
* You cannot save data during observation.

T02339E

- 32 -
3 Operation of the DST-i Oscilloscope Software

3. Press the "A" button in the saving completed screen.


The screen returns to the save menu screen.

NOTE
File names are automatically set with "DO + Serial numbers".
You can save up to 100 files.

- 33 -
3 Operation of the DST-i Oscilloscope Software

3-4 Retrieving data


You can retrieve and replay a screen image and a long duration measurement data.

Retrieving screen images

NOTE
Waveforms of the screen image are not scalable.

1. Select "MENU" from the lower selection items of the waveform observation screen, and press the
"A" button.
The menu screen is displayed.

T02459E T02610E
Waveform observation screen Menu screen

- 34 -
3 Operation of the DST-i Oscilloscope Software

2. Select "OPEN" in the menu screen, and press the "A" button.
The open menu screen is displayed.

T02344E
Open menu screen

NOTE
Only when "WAVE DATA" folder or "DATA LOGGER" folder in an SD memory card storing a
data obtained in Ver.1.00 are in a lower hierarchy is "WAVE DATA" or "LOGGER DATA"
displayed on the open menu screen.
When retrieving data obtained in Ver.1.00, select "WAVE DATA" or "LOGGER DATA" in the
open menu screen and designate a file.

T02418E

3. Select "SCREEN COPY" in the open menu screen, and press the "A" button.
The screen image selection screen is displayed.

T02345E
Screen image selection screen

- 35 -
3 Operation of the DST-i Oscilloscope Software

4. Select the file name you want to replay on the screen image selection screen, and press the "A"
button.
The replay screen of the screen image you have selected is displayed.

T02346E
Screen image replay screen

NOTE
The operational guidance and the file name of the screen image which is being replayed are
displayed in the guidance area alternately.

5. Press the "A" button or the "B" button on the screen image replay screen.
The screen returns to the screen image selection screen.

- 36 -
3 Operation of the DST-i Oscilloscope Software

Retrieving long duration measurement data

1. Select "MENU" from the lower selection items of the waveform observation screen, and press the
"A" button.
The menu screen is displayed.

T02459E T02610E
Waveform observation screen Menu screen

2. Select "OPEN" in the menu screen, and press the "A" button.
The open menu screen is displayed.

T02611E
Open menu screen

NOTE
Only when "WAVE DATA" folder or "DATA LOGGER" folder in an SD memory card storing a
data obtained in Ver.1.00 are in a lower hierarchy is "WAVE DATA" or "LOGGER DATA"
displayed on the open menu screen.
When retrieving data obtained in Ver.1.00, select "WAVE DATA" or "LOGGER DATA" in the
open menu screen and designate a file.

T02418E

- 37 -
3 Operation of the DST-i Oscilloscope Software

3. Select "DST-i OSCILLO DATA" in the open menu screen, and press the "A" button.
The long duration measurement data selection screen is displayed.

T02348E
Long duration measurement data selection screen

4. Select the file name you want to open in the long duration measurement data selection screen, and
press the "A" button.
The replay screen of the selected long duration measurement data you have selected is displayed.

T02349E
Long duration measurement data replay screen

NOTE
You can reset time ranges, channels (waveform display/non-display switching, ground
position), and voltage ranges and display waveforms in the long duration measurement data
replay screen.
How to set these parameters is the same as those set in the waveform observation screen.
For data obtained by "DATA LOGGER" in Ver.1.00 and by pressing the recording start mark,
waveforms cannot be displayed after resetting the time range.

- 38 -
3 Operation of the DST-i Oscilloscope Software

5. Select a scroll bar in the long duration measurement data replay screen, and press the "A" button.

T02462E
Long duration measurement data replay screen

NOTE
Use the right/left buttons to manually move frame by frame through long duration
measurement data.
Move forward and backward through frames with the right and left buttons, respectively.
Use the up/down buttons to move through the "Left end of the scroll bar", "Recording start
position", and "Right of the scroll bar" selections.

6. Press the "B" button in the long duration measurement data replay screen.
The screen returns to the long duration measurement data selection screen.

- 39 -
3 Operation of the DST-i Oscilloscope Software

3-5 Deleting data


You can delete a screen image and a long duration measurement data.

Deleting screen images

1. Select "MENU" from the lower selection items of the waveform observation screen, and press the
"A" button.
The menu screen is displayed.

T02459E T02350E
Waveform observation screen Menu screen

- 40 -
3 Operation of the DST-i Oscilloscope Software

2. Select "DELETE" in the menu screen, and press the "A" button.
The delete menu screen is displayed.

T02351E
Delete menu screen

NOTE
Only when "WAVE DATA" folder or "DATA LOGGER" folder in an SD memory card storing a
data obtained in Ver.1.00 are in a lower hierarchy is "WAVE DATA" or "LOGGER DATA"
displayed on the delete menu screen.
When deleting data obtained in Ver.1.00, select "WAVE DATA" or "LOGGER DATA" in the
delete menu screen and designate a file.

T02460E

3. Select "SCREEN COPY" in the delete menu screen, and press the "A" button.
The screen image selection screen is displayed.

T02345E
Screen image selection screen

- 41 -
3 Operation of the DST-i Oscilloscope Software

4. Select the file name you want to delete in the screen image selection screen, and press the "A"
button.
The delete confirmation screen is displayed.

T02352E
Delete confirmation screen

5. Select "OK" in the delete confirmation screen, and press the "A" button.
The data is deleted, and the screen returns to the screen image selection screen.

CAUTION
You cannot restore deleted data. Make sure that you really want to delete the data before
deleting it.
To cancel the deletion, select "BACK" in the delete confirmation screen and press the "A"
button.

Deleting the long duration measurement data

1. Select "MENU" from the lower selection items of the waveform observation screen, and press the
"A" button.
The menu screen is displayed.

T02459E T02350E
Waveform observation screen Menu screen

- 42 -
3 Operation of the DST-i Oscilloscope Software

2. Select "DELETE" in the menu screen, and press the "A" button.
The delete menu screen is displayed.

T02353E
Delete menu screen

NOTE
Only when "WAVE DATA" folder or "DATA LOGGER" folder in an SD memory card storing a
data obtained in Ver.1.00 are in a lower hierarchy is "WAVE DATA" or "LOGGER DATA"
displayed on the delete menu screen.
When deleting data obtained in Ver.1.00, select "WAVE DATA" or "LOGGER DATA" in the
delete menu screen and designate a file.

T02460E

3. Select "DST-i OSCILLO DATA" in the delete menu screen, and press the "A" button.
The long duration measurement data selection screen is displayed.

T02461E
Long duration measurement data selection screen

- 43 -
3 Operation of the DST-i Oscilloscope Software

4. Select a file to be deleted in the long duration measurement data selection screen, and press the
"A" button.
The delete confirmation screen is displayed.

T02612E
Delete confirmation screen

5. Select "OK" in the delete confirmation screen, and press the "A" button.
Delete data and return to the long duration measurement data selection screen.

CAUTION
You cannot restore deleted data. Make sure that you really want to delete the data before
deleting it.
To cancel the deletion, select "BACK" in the delete confirmation screen and press the "A"
button.

- 44 -
3 Operation of the DST-i Oscilloscope Software

3-6 Zero point correction


You can readjust the zero point of the oscilloscope.
Perform zero point correction after removing all the cables for the oscilloscope.

NOTE
The oscilloscope cables include the oscilloscope probe, the ground cable, and the 4-channel adapter
(optional).

1. Check that the waveform of each channel is horizontal at the 0 V position.

2. Select "MENU" from the lower selection items of the waveform observation screen, and press the
"A" button.
The menu screen is displayed.

T02613E T02379E
Waveform observation screen Menu screen

3. Select "ADJUST ZERO (zero point correction)" in the menu screen, and press the "A" button.
The correction confirmation screen is displayed.

T02380E
Correction confirmation screen

- 45 -
3 Operation of the DST-i Oscilloscope Software

4. Select "START" in the correction confirmation screen, and press the "A" button.
When the zero point correction is completed, the screen returns to the menu screen.

T02337E
In-process screen

NOTE
To cancel zero point correction, select "BACK" in the correction confirmation screen, and
press the "A" button.

- 46 -
3 Operation of the DST-i Oscilloscope Software

3-7 Save confirmation message display/non-display setting


The display/non-display of the save confirmation message screen displayed when the "B" button is pressed without
first saving the long duration measurement data can be switched.

T02516E
Save confirmation message screen

NOTE
If you check the check box in the save confirmation message screen, the save confirmation message
screen will not be displayed.

1. Select "MENU" from the lower selection items of the waveform observation screen, and press the
"A" button.
The menu screen is displayed.

T02459E T02467E
Waveform observation screen Menu screen

2. Select "SETTING" in the menu screen, and press the "A" button.
The message display setting screen is displayed.

T02468E
Message display setting screen

- 47 -
3 Operation of the DST-i Oscilloscope Software

3. Select a check box in the message display setting screen, and press the "A" button.
Check mark is displayed in a check box.

NOTE
Uncheck the check box to display the save confirmation message screen again.

4. Select "BACK" in the message display setting screen, and press the "A" button.
The screen returns to the menu screen.
Subsequently, the save confirmation message screen is not displayed.

- 48 -
3 Operation of the DST-i Oscilloscope Software

3-8 Termination of the oscilloscope


1. Select "MENU" from the lower selection items of the waveform observation screen, and press the
"A" button.
The menu screen is displayed.

T02459E T02381E
Waveform observation screen Menu screen

2. Select "OSCILLOSCOPE END" in the menu screen, and press the "A" button.
DST-i is restarted, and the opening screen is displayed.

- 49 -
4 Troubleshooting

4 Troubleshooting
4-1 If the screen freezes while using
If the PC freezes while using the DST-i oscilloscope software, perform the following procedure.

1. Turn the DST-i mode switch OFF.


If the PC still freezes, please contact your distributor.

2. Turn the DST-i mode switch ON to restart the software.


If the PC still freezes, please contact your distributor.

4-2 If the system message is displayed


Message Check point
SD Card is not found. Please insert SD An SD memory card was not found while saving data.
Card. Insert an SD memory card into DST-i.
There was not enough space in the SD memory card while
saving data. (There is less than 300 KB of free space.)
SD Card is Full. Could not create new file.
Delete unnecessary data.
* You cannot save data during measurement.
A failure occurred during execution of zero point correction.
Adjusting failure.
Please contact your distributor.
The maximum number of files (100) that can be saved is
exceeded.
Error. Maximum number of files.
Delete unnecessary data.
* You cannot save data during measurement.
The SD memory card is write-protected.
Error, SD is write-protected. Cancel the write protection.
* You cannot save data during measurement.
The file was not correct during readout of the saved data.
Failed to read file.
Check the version of the DST-i oscilloscope software.

- 50 -
4 Troubleshooting

DENSO Overseas Subsidiary Company & TAS CORPORATION


name/Address Tel/Fax
DENSO AUTOMOTIVE SYSTEMS AUSTRALIA PTY. LTD. Sales:
Address: 2-46 Merrindale Drive, Croydon, Victoria 3136, Australia Tel: +61-3-8761-1400
Fax: +61-3-8761-1505
Service:
Tel: +61-3-8761-1449
Fax: +61-3-8761-1505
DENSO INTERNATIONAL ASIA PTE. LTD. Sales & Service:
Address: 51 Science Park Road, #01-19 The Aries, Science Park II, Tel: +65-6776-8268
Singapore 117586 Fax: +65-6776-8698
PT. DENSO SALES INDONESIA Sales & Service:
Address: JL.Gaya Motor I, Sunter II, Kel Sungai Bambu, Tel: +62-21-651-2279
Tanjung Priok, Jakarta Utara, Jakarta-Indonesia Fax: +62-21-651-2284
DENSO (MALAYSIA) SDN. BHD. Sales & Service:
Address: Lot 2, Jalan P/1, Section 13, 43650 Bandar Baru Bangi, Tel: +60-3-5569-9933
Selangor Darul Ehsan, Malaysia Fax: +60-3-5567-3301
PHILIPPINE AUTO COMPONENTS, INC. Sales & Service:
Address: 109 Unity Avenue, Carmelray Industrial Park 1, Tel: +63-49-549-3030
4037 Canlubang, Calamba City, Laguna, Philippines Fax: +63-49-549-3088
DENSO TAIWAN CORP. Sales & Service:
Address: 525, Sec. 2 Mei Su Road Yang Mei Town Taoyuan Hsien, Tel: +886-3-482-8001
Taiwan, R.O.C. Fax: +886-3-482-8003
DENSO SALES (THAILAND) CO.,LTD. Sales:
Address: 888 Moo 1, Bangna-Trad Rd. Km. 27.5, Tambol Bangbo, Tel: +66-2-315-9550
Amphur Bangbo, Samutprakarn 10560 Thailand Fax: +66-2-315-9556
Service:
Tel: +66-2-315-9550
Fax: +66-2-315-9557
DENSO EUROPE B.V. Sales & Service:
Address: Hogeweyselaan 165, 1382 JL Weesp, The Netherlands Tel: +31-294-493-493
Fax: +31-294-417-122
DENSO SALES MIDDLE EAST & NORTH AFRICA FZE. Service:
Address: P.O.Box 261986, Jebel Ali Free Zone South, Dubai, Tel: +971-04-880-7789
United Arab Emirates Fax: +971-04-880-7790
DENSO (CHINA) INVESTMENT CO.,LTD. Sales:
Address: Room No.518, The Beijing Fortune Building, Tel: +86-10-6590-8025
No.5 Dong San Huan Bei Lu, Chaoyang District, Beijing, Fax: +86-10-6590-8065
100004, China Service:
Tel: +86-10-6590-8021
Fax: +86-10-6590-9044
DENSO INTERNATIONAL INDIA PVT. LTD. Sales & Service:
Address: Plot No.3, Sector-3, IMT Manesar, Gurgaon, Tel: +91-11-2617-6693
Haryana-122 050, India Fax: +91-11-2618-2474
DENSO PRODUCTS AND SERVICES AMERICAS, INC. Sales:
Address: 3900 Via Oro Avenue, Long Beach, California 90810, U.S.A. Tel: +310-834-8652
Fax: +310-952-7406
Service:
Tel: +310-834-8652
Fax: +310-952-7599
DENSO SALES CANADA, INC. Sales & Service:
Address: 195 Brunel Road, Mississauga, Ontario, Canada, L4Z 1X3 Tel: +1-905-890-0890
Fax: +1-905-890-8474

DENSO SALES KOREA CORPORATION Sales:


Address: 131, Seonggogae-ro, Uiwang-si, Gyeonggi-do, Korea Tel: +82-31-340-1786
437-120 Fax:+82-31-8033-7211
TAS CORPORATION Sales & Service:
Address: 248-2 Shimoshiota Narumi-cho, Midori-ku, Nagoya, Aichi, Tel: +81-52-621-3277
458-0801, JAPAN Fax: +81-52-621-3288

- 51 -