IR-CA Series Single color radiation


thermometer [Software volume]
Model: IR-CAB, IR-CAE, IR-CAG,
IR-CAI, IR-CAM, IR-CAN,
IR-CAP, IR-CAR, IR-CAS
IR-CAT, IR-CAU, IR-CAW

Always keep this instruction manual with the thermometer.

Please be sure to deliver this instruction manual to a person


who uses this thermometer
INST No.INE-453-P3

Request and notices


This instruction manual is explaining only "soft volume" for using single color radiation
thermometer (Model IR-CAB, IR-CAE, IR-CAG, IR-CAI, IR-CAM, IR-CAN, IR-CAP, IR-CAR,
IR-CAS, IR-CAT, IR-CAU, IR-CAW).
To use single color radiation thermometer (Model IR-CAB, IR-CAE, IR-CAG, IR-CAI,
IR-CAM, IR-CAN, IR-CAP, IR-CAR, IR-CAS, IR-CAT, IR-CAU and IR-CAW) correctly and
safely, read the separate instruction manual [IR-CA series Hardware volume]

Request to the operator of the thermometer


This instruction manual describes the maintenance of the thermometer, too.
Keep this instruction manual with the thermometer.
If you have unclear points or need technical assistance, please contact your sales agent of
CHINO Corporation.
Notices
1. The information in this manual is subject to change without notice and does not
represent a commitment on the part of CHINO Corporation.
2. No part of this manual may be reproduced or transmitted in any form or by any means,
electronic or mechanical, including photocopying and recording, for any purpose other
than the purchaser's personal use without permission of CHINO Corporation.
3. CHINO Corporation shall not be liable for any operation results.

Preface !
To use single color radiation thermometer (Model IR-CAB, IR-CAE, IR-CAG, IR-CAI,
IR-CAM, IR-CAN, IR-CAP, IR-CAR, IR-CAS, IR-CAT, IR-CAU and IR-CAW) correctly
and safely, please keep the following safety measures for the operation and storage of the
thermometer.
Serial No. plate 【Serial No. Label
for thermometer】
This instruction manual is for the
thermometer with the software Software
version . . Model version No.
The software version of your
thermometer is indicated in the I R-CA BGTS3 / □.□□
label on the rear side. RANGE 20-1000 ℃
(Ref to [3.1 Overview] in separate NO. I A049B001
instruction manual [IR-CA series
Hardware volume])
Serial No. Measuring
range

♦Symbols in this instruction manual


The symbols shown below are used depending on important degrees for using the thermometer safely
and avoiding unexpected situations.
Important
Symbols Contents
degree
This symbol is attached to a title for the sentence with Warning .
1 !
Warning For avoiding dangerous accidents (may cause death or serious injury)
2 like as electrical shock, fires, or troubles/damages of the thermometer.

3 Caution For avoiding injury or in physical damage to the thermometer.

4 Remarks Information that we suggest to read carefully.

5 Reference Information that you can use as a reference.

-C1-
INST No.INE-453-P3 Software Ver.1.0□

CONTENTS Software Version Ver.1.0

1. Digital display··························································· 1 4.5 Modulation time constant unit selection·························· 14


1. Digital display······································································ 1 4.6 Contact output item selection
(Model IR-CA K only) ············································· 14
2. Operating mode and engineering mode ················ 1
4.7 Automatic emissivity calculation type selection
2.1 Parameter programming /selection
(External setting: Model IR-CAQ 5 only)············ 15
(Operating mode)···························································· 1
4.8 Measuring wavelength selection
2.2 Parameter programming /selection
(Model for IR-CAW only) ·············································· 16
(Engineering mode) ························································ 2
4.9 Zero/span adjustments ······················································ 17
3. Operating mode ························································ 3 4.9-0 Zero/span adjustment
3.1 Emissivity programming················································· 3 implementation selection········································· 17
3.1.1 Emissivity programming for model IR-CAB, 4.9-1 “Zero” side adjustments…Adjustment by furnace
IR-CAE, IR-CAG, IR-CAI, IR-CAM, IR-CAN, temperature at zero side ··········································· 17
IR-CAS, IR-CAT and IR-CAU································· 3 4.9-2 “Span” side adjustments…Adjustment by furnace
3.1.2 Emissivity programming for model IR-CAW ············ 3 temperature at span side··········································· 17
3.1.2-1 Emissivity programming for measuring 4.10 Analog output correction programming························ 18
wavelength λ = 8 to 13µm···················· 3 4.10.1 Programming of number of analog output
3.1.2-2 Emissivity programming for measuring correction datas (number of knee-point)······················· 18
wavelength λ = 1.55µm························ 4 4.10.2 Programming of number of analog output
3.1.2-3 Emissivity programming for measuring correction data No.················································· 18
wavelength λ = 0.9µm·························· 4 4.10.3 Entering of data before analog output correction . 19
3.2 Automatic emissivity calculation····································· 5 4.10.4 Entering of data after analog output correction .··· 19
3.2.1 Automatic emissivity calculation 4.10.5 Selection of analog output correction for
(Internal):Input of reference temperature······················ 5 programming item . ············································· 20
3.2.2 Automatic emissivity calculation 4.11 Analog input programming
(External…Model IR-CA 5 only): (Model IR-CA 5 only) ·········································21
Selection of external programming···························· 5 4.11.1 Selection of analog input items··································· 21
3.3 Signal modulation mode selection··································· 6 4.11.2 Emissivity scaling for remote emissivity
3.4 Modulation degree programming···································· 7 programming ························································· 22
3.4.1 Modulation time constant programming 4.11.3 Temperature scaling for automatic emissivity
(Effective when the “dELy” is selected in the signal calculation ······························································ 23
modulation mode.) ······················································ 7 4.12 Communications programming
3.4.2 Damping degree selection (Model IR-CA S only) ······································ 24
(Effective when the “PEAk” is selected in the signal 4.12.1 Address programming················································· 24
modulation mode.) ······················································ 7 4.12.2 Communication speed selection································· 24
3.5 Alarm programming 4.12.3 Communication protocol mode selection·················· 25
(Model IR-CA J only Constant output possible : 4.13 Key lock··································································· 25
As for other models “AL”,”AH” of the status marker
5. Self-diagnostic function and Overflow/underflow··26
are lighted) ·································································· 8
5.1 Self-diagnostic function································································ 26
3.5.1 Alarm type selection
5.2 Overflow/underflow ························································· 26
(High alarm, low alarm or alarm disable) ····················· 8
3.5.2 Alarm set-point (1 point only in high alarm in [6.4.5-1
6. List of starting up modes···········································27
Alarm type selection])·················································· 8
6.1 Starting up modes······························································ 27
4. Engineering mode ···················································· 9 6.2 Screens ··············································································· 27
4.1 Analog output scaling programming······························· 9
7. Parameter programming/selection ···························27
4.1.1 Analog output minimum value
7.1 Parameter programming/selection
programming…”OutL”··········································· 9
(operating mode)························································ 27
4.1.2 Analog output maximum value
7.2 Parameter programming/selection
programming…”OutH” ·········································· 9.
( engineering mode) ··················································· 28
4.2 Analog dummy output programming ····························· 10
4.3 Hold functions (Hold disable, Sample hold, Peak hold)
selection ··································································· 11
4.3.1 Hold functions type selection········································ 11
4.3.2 Peak hold reset type selection ····································· 12
4.3.3 Peak hold reset time programming······························· 13
4.4 Measurement unit selection·············································· 13

-C 2 -
INST.No.INE-453-P3 Software Ver.1.0□

1. Digital display
1 Digital display *(6): Displayed in Model IR-CAW only
After the installation, the connections and the
wirings are completed, turn all mains power on. Tb CONT MEM PEAK
Measurement starts and the measured value is AL AH
(6) (4)
displayed in the main display.
When the low alarm or the high alarm activates, the C
(5)
status marker “AL” or “AH” will light. F
(On the condition that the low alarm or the high (1)
alarm has been programmed)
(1) Main display:7-segment LCD 4 digits .. (3)
Displays measured value or
parameters ε
(2)
(2) Sub display: 7-segment LCD 4 digits ..
Displays parameter menus.
(3) Sub marker :ε … Emissivity
(4) Status marker :AL … Low alarm [Measurement mode]
: AH … High alarm The measured value is displayed
(5) Temperature unit: C …ºC (Celsius) in the main display.
F …ºF (Fahrenheit) Nothing is displayed in the sub display
*(6) Measuring wavelength display: Nothing is indicated
…8 to 13µm and in the sub marker.
1 indicated…1.55µm,
2 indicated…0.9µm
(*) is indicated at [3.1.2-2 Emissivity programming of measuring wavelength λ = 1.55µm] at
[3.1.2-3 Emissivity programming of measuring wavelength λ = 0.9µm]

2. Operating mode and Engineering mode


O p e ra tin g m o d e
2.1 Parameter programming/selection
M e a su re m e n t m o d e
(Operating mode)
In the measurement mode, program or select a SEL

parameter by using SEL key. E m issiv ity p ro g ra m m m in g

The operator mode covers programming or


selection of emissivity, automatic emissivity A u to m a tic e m issiv ity d isa b le
calculation, signal modulation mode, c a lc u la tio n e n a b le

modulation degree, and alarms. For each SEL


program menu, refer to [3 Operating mode].
Y

A u to m a tic e m issiv ity c a lc u la tio n


The right figure shows the flow chart of the
parameter menus by pressing SEL key in the SEL

measurement mode.
[Measuring wavelength M o d u la tio n

selection in Model
IR-CAW series only] D e la y
Peak

N
SEL
Y

SEL
T im e c o n sta n t D a m p in g d e g re e
Emissivity programmming 8 to 13 μm
SEL
SEL

Emissivity programmming 1.55 μm A la rm m o d e


SEL
SEL
Emissivity programmming 0.9μm
A la rm te m p e ra tu re

Refer to [3.1.2 Emissivity programming of Model IR-CAW SEL


-1-
INST.No.INE-453-P3 Software Ver.1.0□

2. Operating mode and Engineering mode


2.2 Parameter programming/selection (Engineering mode)
In the measurement mode, press SEL key and ENT key [ Measurement mode ]
simultaneously to display the engineering mode shown Tb CONT MEM PEAK
in the right figure for programming or selecting a AL AH
parameter. C
The engineering mode covers programming or selection F
of analog output scaling, analog dummy output, hold
function (without hold function, peak hold or sample
  
hold), temperature unit, contact output, automatic
emissivity calculation (internal or external), zero/span ε
adjustment, analog output correction, analog input (for
the thermometer with the option of analog input), and
[ Engineering mode ]
communications (for the thermometer with the option of
Tb CONT MEM PEAK
communications interface). For each program menu,
refer to [6.5 Engineering mode] . AL AH
The lower figure shows the flow chart of the parameter C
menus by pressing SEL key in the engineering mode. F

Engneering mode
Measurement mode In programming of parameters, the screen will move to the measurement
mode unless any key is pushed for 1 minute.
SEL + ENT

Analog output Peak hold/ Temperature Zero/Span Analog output Communications


Analog
(Low Sample hold unit adjustment correction
input enable
N address
temperature) (Lower side) (Number of data)
SEL SEL
SEL
Y

SEL
SEL SEL
Fixed Number of Analog input
number unir data=0
N usage Communications
Peak hold? N
Analog output Zero/Span speed
SEL SEL
(High adjustment
Y

(Higher side) SEL


temperature) Contact
Reset type Data number Analog input
output type
SEL scaling
SEL (Lower limit
SEL SEL SEL
value)
Analog dummy Temperature
Wavelength
output Hold time before correction N
N selection
reset? SEL

SEL SEL
SEL Analog input
Y

Automatic emissivity scaling


Temperature after
Reset time calculation enable/ (Higher limit
correction
disable(Int/Ext) value)
SEL SEL
SEL SEL
All data
Correction calculation input
orer (No,Yes,cancel) SEL

-2-
INST.No.INE-453-P3 Software Ver.1.0□

3. Operating Mode
3.1 Emissivity programming
3.1.1 Emissivity programming for model IR-CAB, IR-CAE, IR-CAG, IR-CAI,
IR-CAM, IR-CAN, IR-CAS, IR-CAT and IR-CAU.
If the emissivity of object measured is low, the temperature displayed becomes lower than the exact
temperature and the emissivity is to be compensated.
Tb CONT MEM PEAK
(1) To program the emissivity value, press SEL key in the
AL AH
measurement mode screen until “ ε ” appears in the sub C
marker as shown in the right figure.
F
(2) By pressing key, the most significant digit of the sub
display will blink.
  
(3) Use or key to program the desired value.
ε
(4) To shift to the next lower digit, press key. The next
lower digit will blink.
(5) Repeat the above procedure of (3) and (4) up to the least Sub display
significant digit for programming (4 digits). Sub marker
(6) To activate this programming, press ENT key. Main display

Remarks The programmable range is 1.999 to 0.050.


The default value is 1.000.

3.1.2 Emissivity programming for model IR-CAW


3.1.2-1 Emissivity programming of measuring wavelength λ = 8 to 13µm
(1) Confirm the wavelength selection display under the main Tb CONT MEM PEAK
marker is turning off. AL AH
To program the emissivity value of 8 to 13µm, press C
SEL key in the measurement mode screen until “ ε ” F
appears in the sub marker as shown in the right figure.
(2) By pressing key, the most significant digit of the sub   
display will blink. ε
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key. The next
Sub display
lower digit will blink.
Sub marker
(5) Repeat the above procedure of (3) and (4) up to the least
Main display
significant digit for programming (4 digits).
Nothing is indicated
(6) To activate this programming, press ENT key.

Remarks The programmable range is 1.999 to 0.050.


The default value is 1.000.

-3-
INST.No.INE-453-P3 Software Ver.1.0□

3. Operating Mode

3.1.2-2 Emissivity programming of measuring wavelength λ = 1.55µm


(1)To program the emissivity value of 1.55µm, press SEL
Tb CONT MEM PEAK
key in the measurement mode screen until “ ε ” appears AL AH
in the sub marker as shown in the right figure.
C
By pressing SEL key, indicate the 1 wavelength selection
F
display under the main marker.
(2) By pressing key, the most significant digit of the sub
  
display will blink.
(3) Use or key to program the desired value.
ε
(4) To shift to the next lower digit, press key. The next
lower digit will blink. Sub display
(5) Repeat the above procedure of (3) and (4) up to the least Sub marker
significant digit for programming (4 digits). Main display
(6) To activate this programming, press ENT key. 1 indicated

Remarks The programmable range is 1.999 to 0.050.


The default value is 1.000.

3.1.2-3 Emissivity programming of measuring wavelength λ = 0.9µm


(1) To program the emissivity value of 0.9µm, press SEL Tb CONT MEM PEAK
key in the measurement mode screen until “ ε ” appears AL AH
in the sub marker as shown in the right figure. C
By pressing SEL key 2 times, indicate the 2 wavelength F
selection display under the main marker.
(2) By pressing key, the most significant digit of the sub   
display will blink.
ε
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key. The next
lower digit will blink. Sub display
(5) Repeat the above procedure of (3) and (4) up to the least Sub marker
significant digit for programming (4 digits). Main display
(6) To activate this programming, press ENT key. 2 indicated

Remarks The programmable range is 1.999 to 0.050.


The default value is 1.000.

-4-
INST.No.INE-453-P3 Software Ver.1.0□

3. Operating Mode

3.2 Automatic emissivity calculation


The emissivity is automatically calculated when you coincident the measured value to the reference
temperature value entered by function keys or by an analog input (option).

Remarks This function is effective when “ In ” (internal) or “ Ext ” (external) is selected in


[4.7 Automatic emissivity calculation type selection.].

3.2.1 Automatic emissivity calculation(Internal ):Input of reference temperature


In the measurement mode, the emissivity is automatically calculated by entering the reference
temperature value measured by a thermocouple or other sensor in advance.
(1) To program the reference temperature value, press SEL Tb CONT MEM PEAK
key in the measurement mode screen until “Auto”        AL AH
appears in the sub display as shown in the right figure. C
(2) By pressing key, the most significant digit of the F
main display will blink.
  
(3) Use or key to program the desired value.
ε
(4) To shift to the next lower digit, press key.
The next lower digit will blink.
(5) Repeat the above procedure of (3) and (4) up to the least Sub display
significant digit for programming (4 digits).
(6) To activate this programming, press ENT key. Main display

Remarks • The programmable range is 0 to 6280ºC.


• The default is the minimum temperature value of the measuring range.

3.2.2 Automatic emissivity calculation (External …Model IR-CA 5 only)


: Selection of external programming
In the measurement mode, the emissivity is automatically calculated by entering the reference
temperature value by the analog input (option).

Reference Necessary to program the scaling of the analog input in [4.11 Analog input
programming])
(1) To select the external programming, press SEL key in the Tb CONT MEM PEAK
measurement mode screen until “Auto” appears in the sub       AL AH
display as shown in the right figure. C
(2) By pressing key, “ok” will blink in the main display. F
(3) Confirm that the analog input is connected to the
input/output terminals (9 and 10).   
(Ref. [Connections/Wirings] in separate instruction manual ε
[IR-CA Series Hardware volume])
(4) To activate this selection, press ENT key.
Sub display
Main display

Caution For the correct measurement, confirm the connections of the analog input.

-5-
INST.No.INE-453-P3 Software Ver.1.0□

3. Operating Mode
3.3 Signal modulation mode selection
Two kinds of the signal modulation mode are available, one is “dELy” (displays the measured
value smoothly for fluctuating original signal) or the other is “PEAk” (displays the measured value
by modulating the original signal after the tracing of peak value).
To display the measured value based on the original signal, select “dELy” and program the
modulation degree to “0.0”. (Ref. [6.4.4-1 Modulation time constant programming])
(1) To select the signal modulation mode, press SEL key in Tb CONT MEM PEAK
     AL AH
the measurement mode screen until “Modu” appears in
C
the sub display as shown in the right figure. F
(2) By pressing key, “dELy” or “PEAk” will blink in the
main display.  
(3) Use or key to select the desired mode. ε
(4) To activate this selection, press ENT key.
Sub display
Main display
The measured value is displayed by the
first-order lag signal based on the Tb CONT MEM PEAK
dELy modulation time constant programmed in      AL AH
[6.4.4-1 Modulation time constant C
programming]. F
When the measured temperature is rising up,
the measured value is displayed by the  
original signal. When the measured value is ε
PEAk falling down, the measured value is
displayed by the damping degree
programmed in [6.4.4-2 Damping degree Sub display
selection]. Main display

-6-
INST.No.INE-453-P3 Software Ver.1.0□

3. Operating Mode
3.4 Modulation degree programming
Caution Execute the programming of the modulation degree together with [3.3 Signal
modulation mode selection].
When the "dELy" is selected in the signal modulation mode, the first-order lag degree can be
adjusted. Further, when the “PEAk” is selected, the damping degree of signal after tracking the
value can be adjusted.

3.4.1 Modulation time constant programmming


(Effective when the “dELy” is selected in the signal modulation mode.)
When the input signal is fluctuated, the measured value is displayed smoothly.
[Decimal point 1 column]
  Tb CONT MEM PEAK
(1) To program the modulation time constant, press SEL key
        AL AH
in the measurement mode screen until “tAu” appears in the
C
sub display as shown in the right figure.
F
(2) By pressing key, the most significant digit of the main
display will blink.
   
(3) Use or key to program the desired value. ε
(4) To shift to the next lower digit, press key.
The next lower digit will blink.
Main display Sub display
(5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits). [Decimal point 2 column]
(6) To activate this programming, press ENT key.   Tb CONT MEM PEAK
        AL AH
C
Remarks ・The programmable range is 0.0 to 99.9(second)
F
or 0.00 to 9.99(second).
・ The default value is 0.0 second (original   
signal).  
ε
Reference • By selecting “dELy” and programming to
“0.0”, the measured value based on the
original signal is displayed. Sub display
Main display
3.4.2 Damping degree selection
(Effective when the "PEAk" is selected in the signal modulation mode.)
Tb CONT MEM PEAK
(1) To select the damping degree, press SEL key in the
      AL AH
measurement mode screen until “dEc” appears in the sub
C
display as shown in the right figure.
F
(2) By pressing key, “0” at the least significant digit in the
main display will blink.
  
(3) Use or key to select the desired damping degree ε
[0, 2, 5 or 10(ºC/second)] or [10, 5, 2 or 0(ºC/second)].
(4) To activate this selection, press ENT key.
Main display Sub display

Remarks • The default value is 0 (ºC/second).

-7-
INST.No.INE-453-P3 Software Ver.1.0□

3. Operating Mode
3.5 Alarm programming (Model IR-CA J only contact output possible:
As for other models ”AL”, ”AH” of the status marker are lighted.)
This programming is for the selection of the alarm type (high alarm, low alarm or alarm disable)
and the programming an alarm set-point (1 point only in the high alarm or the low alarm).
3.5.1 Alarm type selection (High alarm, low alarm [Alarm disable]
or alarm disable) Tb CONT MEM PEAK
(1) To select the alarm type, press SEL key in the AL AH
measurement mode screen until “A.Mod” appears in the C
sub display as shown in the right figure. F
(2) By pressing key, “no”, “Hi” or “Lo” will blink in the
main display.
(3) Use or key to select the desired type. ε
(4) To activate this selection, press ENT key.

[Alarm type] Main display Sub display


No: Alarm disable
Hi: High alarm (The open collector of the contact output is ON when the measured value
exceeds the alarm set-point)
Lo: Low temperature alarm (The open collector of the contact output is OFF when the
measured value exceeds the alarm set-point)
[High alarm] [Low alarm]
Tb CONT MEM PEAK Tb CONT MEM PEAK
AL AH AL AH
C C
F F

ε ε

Sub display Sub display


Main display Main display
3.5.2 Alarm set-point (1 point only in high alarm or low alarm
in [3.5.1 Alarm type selection]
(1) To program the alarm set-point, press SEL key in the Tb CONT MEM PEAK
measurement mode screen until “ALM” appears in the AL AH
sub display as shown in the right figure. C
(2) By pressing key, the most significant digit of the main F
display will blink.
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key. ε
The next lower digit will blink.
(5) Repeat the above procedure of (3) and (4) up to the least Sub display
significant digit for programming (4 digits). Main display
(6) To activate this programming, press ENT key.

Remarks • The programmable range is 0 to 6280ºC.


• The defaults are: Hi (High alarm) … Maximum value of the measuring range
Lo (Low alarm) … Minimum value of the measuring range

-8-
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4. Engineering mode
In the measurement mode, press SEL and ENT keys simultaneously to display the engineering
mode screen. Refer to the engineering mode screen in [2.2 Parameter programming/selection
(Engineering mode)]

Caution To return to the measurement mode screen from the engineering mode screen
• The screen will be automatically returned to the measurement mode screen if
any key is not pressed for 1 second in the programming of each parameter.
• The screen will return to the measurement mode screen when the
programming from [4.1 Analog output scaling programming] to [4.10 Analog
output correction programming] is completed.
Note: Up to [4.1 Analog input programming] in case of IR-CA5.
Up to [4.12 Communications programming] in case of IR-CAS.

4.1 Analog output scaling programming


4.1.1 Analog output minimum value programming…"OutL"
This programming is for the minimum value of the temperature scaling when the measured value is
outputted by analog signal. [Minimum value]
(1) To program the analog output minimum value, press SEL Tb CONT MEM PEAK
key in the engineering mode screen until “OutL” appears in AL AH
the sub display as shown in the right figure. C
(2) By pressing key, the most significant digit of the main F
display will blink.
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key. ε
The next lower digit will blink.
(5) Repeat the above procedure of (3) and (4) up to the least Sub display
significant digit for programming (4 digits). Main display
(6) To activate this programming, press ENT key.
Reference • The programmable range is 0 to 6280ºC.
• The defaults is OutL (Minimum value)… Minimum value of the measuring range

4.1.2 Analog output maximum value programming…"OutH"


This programming is for the maximum value of the temperature scaling when the measured value is
outputted by analog signal.
(1) To program the analog output maximum value, press SEL [Maximum value]
key in the engineering mode screen until “OutH” appears Tb CONT MEM PEAK
in the sub display as shown in the right figure.
AL AH
(2) By pressing key, the most significant digit of the main C
display will blink. F
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key. The next
lower digit will blink. ε
(5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
(6) To activate this programming, press ENT key. Main display Sub display

Reference • The programmable range is 0 to 6280ºC.


• The defaults is: OutH (Maximum value)… Maximum value of the measuring
-9-
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.2 Analog dummy output programming
This programming is for analog dummy output based on programmed ratio. The output of 0 to
100% is corresponding to 4 to 20mA.
Note) : This screen will not move to the measurement mode even if any key is not pushed for 1 minute.
.(1) To program the dummy analog output, press SEL key in
Tb CONT MEM PEAK
the engineering mode screen until “OutC” appears in the AL AH
sub display as shown in the right figure.
C
(2) By pressing key, the most significant digit of the main F
display will blink.
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key.
ε
The next lower digit will blink.
(5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits). Main display Sub display
(6) To activate this programming, press ENT key.
Then, the programmed value is outputted as a dummy
output.

Caution When “OutC” is displayed, “0” is displayed in the main display. However, the
dummy output corresponding to “0” is not outputted even if ENT key is pressed
on this condition.
Take the above procedure to program to “0”. If the dummy output is not required,
press SEL key with “0” displayed to move to the next programming mode.

Remarks • Programmable range is 0 to 100 (%).


Example) 0 (%) = 4mA DC
50 (%) = 12mA DC,
100 (%) = 20mA DC
• The default value is 0 (%)= 4mA DC.

- 10 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.3 Hold functions (Hold disable, sample hold, peak hold) selection
This mode is for the selection of the hold function (hold function disable, sample hold or peak hold).
When the peak hold is selected, [4.3.2 Peak hold reset type (time reset or remote contact reset)
selection] is necessary. Furthermore, when the internal time reset is selected, [4.3.3 Peak hold reset
time programming] is necessary.
For the sample hold, only the selection of the hold mode is required
4.3.1 Hold function type selection
[Hold function disable]
(1) To select the hold function type, press SEL key in the Tb CONT MEM PEAK
engineering mode screen until “HoLd” appears in the sub AL AH
display as shown in the right figure. C
F
(2) By pressing key, “no”, “PEAk” or “SAMP” will blink in
the main display.
(3) Use or key to select the desired type.
ε
(4) To activate this selection, press ENT key.
Sub display
Main display
[Peak hold]
[Hold function types] Tb CONT MEM PEAK
no : Hold function AL AH
disable C
PEAk : Peak hold F
SAMP : Sample hold

ε
【 Sample hold 】

Sub display
Main display
Input form
[Sample hold]
Tb CONT MEM PEAK
AL AH
C
Output form
F

Remote contact input reset


ε

Sub display
Main display

Caution For temperature display and analog output on the sample hold mode, the
programmed value is hold by to going down point in remote contact input.

Caution For the sample hold, the measured value is hold by the remote contact.
(The pulse width more than 55m/sec is required for the remote contact.)
Remarks • The default is no (hold function disable).

- 11 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.3.2 Peak hold reset type selection
(It is necessary to select when the peak hold is selected in [Reset disable]
Tb CONT MEM PEAK
[4.3.1 Hold function type selection].)
AL AH
(1) After the selection of “PEAk” in [4.3.1 Hold function C
type selection], press SEL key until “HrSt” (selection F
screen of peak hold reset type) appears in the sub
display as shown in the right figure.
(2) By pressing key, “no”, “in” or “Et” will blink in ε
the main display.
(3) Use or key to select the desired type. Sub display
(4) To activate this selection, press ENT key. Main display
[Internal reset] [External reset]
[Reset types] Tb CONT MEM PEAK Tb CONT MEM PEAK
no : Reset disable AL AH AL AH
in : Internal (time reset) C C
Et : External (remote contact F F
input reset)

ε ε
Remarks • The default is no
(reset disable). Sub display Sub display
Main display Main display

【Peak hold (Reset disable)】 【Peak hold internal (Time reset)】

Input Input

Output Output
TR TR

【Peak hold external (Remote contact reset)】

Caution TR: Reset time (0.0 to 99.9sec)


Input Hold time is less than 0.0 to
0.1 in 0.0 programmed.

For the remote contact reset is


Output Caution
reset by the remote contact
input.
External contact output (The pulse width is required
more than 55msec for the
remote contact input.)

- 12 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.3.3 Peak hold reset time programming
(It is necessary to program the reset time when the peak hold is selected in [4.3.1 Hold function
type selection] and then “in” (internal reset) is selected in [4.3.2 Peak hold reset type selection].)
(1) After the selection of “in” (time reset), press SEL key until Tb CONT MEM PEAK
“r.tim” (programming screen of peak hold reset time) appears in AL AH
the sub display as shown in the right figure. C
(2) By pressing key, the most significant digit of the main F
display will blink.
(3) Use or key to program the desired value.
ε
(4) To shift to the next lower digit, press key.
The next lower digit will blink.
(5) Repeat the above procedure of (3) and (4) up to the least Sub display
Main display
significant digit for programming (4 digits).
(6) To activate this programming, press ENT key.
Then, the programmed value is outputted as a dummy output.

Remarks • The programmable range is 0.0 to 99.9 (second).


• The default value is 0.0 (second).

4.4 Measurement unit selection


This mode is for the selection of the measurement unit for measured temperature.
(1) To select the measurement unit, press SEL key in the [ ºC ]
Tb CONT MEM PEAK
engineering mode screen until “unit” appears in the sub
AL AH
display as shown in the right figure.
C
(2) By pressing key, “C” or “F” will blink in the main F
display.
(3) Use or key to select the desired unit.
ε
(4) To activate this selection, press ENT key.

Sub display
Main display
[Measurement units] [ ºF ]
C : ºC (Celsius) Tb CONT MEM PEAK
F : ºF (Fahrenheit) AL AH
C
F

Sub display
Main display

Remarks The default is C (ºC)

- 13 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.5 Modulation time constant unit selection
[3.4.1 Modulation time constant programming] to the cipher below the decimal point of when that
sets up the fixed number is selected when it puts it
(1) To select the fixed number, press SEL key in the engineering Tb CONT MEM PEAK
mode screen until “tAu.d” appears in the sub display as AL AH
shown in the right figure. C
(2) By pressing key, “1” or “2” will blink in the main F
display.
(3) Use or key to select the decimal point.
ε
(4) To activate this selection, press ENT key.

Sub display
Main display
Remarks The default is 2 (decimal point 2 Column)

[No contacts output]


4.6 Contact output item selection (Model IR-CAK only) Tb CONT MEM PEAK
This mode is for the selection of the contacts output item type. AL AH
C
(1) To select the contacts output item type, press SEL key in the
F
engineering mode screen until “dout” appears in the sub
display as shown in the right figure.
(2) By pressing key, “no”, “ALM” or “Err” will blink in the ε
main display.
(3) Use or key to select the desired type.
Sub display
(4) To activate this selection, press ENT key. Main display
[High/low temperature alarm]
Tb CONT MEM PEAK
[Contact output items]
no : No contacts output AL AH
ALM : High/low temperature alarm C
Err :Self-diagnostic abnormal F

Remarks The default is ALM (High/low temperature alarm) Main display


Sub display

[Self-diagnosis abnormal]
Tb CONT MEM PEAK
AL AH
C
F

Sub display
Main display

- 14 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.7 Automatic emissivity calculation type selection
(The external setting :Model IR-CA5 only)
This mode is for the selection of the automatic emissivity calculation function (disable or enable).
In case of using the automatic emissivity calculation functions, select the internal or external setting.
[Calculation disable]
(1) To select the automatic emissivity calculation type, press
Tb CONT MEM PEAK
SEL key in the engineering mode screen until “E.Aut” AL AH
appears in the sub display as shown in the right figure. C
(2) By pressing key, “no”, “in” or “Et” will blink in the F
main display.
(3) Use or key to select the desired type. ε
(4) To activate this selection, press ENT key.
Sub display
Main display

[Internal automatic emissivity


calculation(by keys)]
Tb CONT MEM PEAK
[Automatic emissivity AL AH
calculation types] C
no: Calculation disable F
in : Internal setting (by key)
Et: External setting (by analog input)
ε

Sub display
Main display

[External automatic emissivity


calculation(by analog input)]
Tb CONT MEM PEAK
AL AH
C
F

Sub display
Main display

Remarks • The default is no (automatic emissivity calculation disable).

- 15 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.8 Measuring wavelength selection (Model for IR-CAW only)
(1) To select the measuring wavelength, press SEL key in the Tb CONT MEM PEAK
AL AH
engineering mode screen until “dEtC” appears in the sub
C
display as shown in the right figure. F
“1coL”, “tP”, “inGA” or “Si” is displayed.

•If pressing SEL key in this mode, the screen ε


Caution
moves to [4.9 Zero/Span adjustments] by
skipping “measuring wavelength selection”. Sub display
•To select the “measuring wavelength selection” Main display
mode, operate the following procedure from (2)
to (5)

(2)By pressing key, “1col”, “tP” , “inGA”or “Si” will blink in the main display.
(3) Use or key to select the measuring wavelength in following table.
(4) To activate this selection, press ENT key.
(5) Press SEL key, move to next item.

Main display Measuring Measuring wavelength Remarks


temperature range
8 to 13µm 20 to 400ºC
Measuring wavelength
1coL 20 to 3000ºC 1.55µm 400 to 1000ºC
automatic switching
0.9µm 1000 to 3000ºC
tP 20 to 500ºC 8 to 15µm 20 to 500ºC
inGA 300 to 1000ºC 1.55µm 300 to 1000ºC
Si 800 to 3000ºC 0.9µm 800 to 3000ºC

Remarks • The default is “1 coL”. (Measuring temperature range is 20 to 3000ºC)

Caution •A difference in level may happen to measurement temperature by switching of


measurement wavelength in the “measuring wavelength automatic switching”
of 1coL.
•In the range of -100ºC to 100ºC at the measuring wavelength automatic
switching of “1coL” does not use the thermometer for to control.

- 16 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.9 Zero/span adjustments
This mode is for the zero and span adjustments of your thermometer by entering the black body
temperature at the minimum temperature range (zero side) and the maximum temperature range
(span side) by using your block body furnace.
It is necessary to measure the furnace temperature at the zero side and the span side by a reference
radiation thermometer in advance.
4.9-0 Zero/span adjustment implementation selection
(1) To enter the Zero/span adjustment implementation
Tb CONT MEM PEAK
selection side temperature, press SEL key in the
engineering mode screen until “CAL” appears in the sub AL AH
display as shown in the right figure. The etc. is displayed C
to main display department “no” the present measurement F
wavelength'. When it pushes a SEL key here, the zero and
span adjustments skips and enter into the Analog output
correction programming screen. The following operation ε
to implement the zero and span adjustments (2).
(2) By pressing key, the most significant digit of the main
display will blink "no".
(3) Use or key to program the desired value.
(4) To activate this programming, press ENT key.
(5) Pushing the SEL key it enters into the setting of the next.
4.9-1 “Zero” side adjustment …Adjustment by furnace temperature at zero side
(1) To enter the zero side temperature, press SEL key in the [Zero side]
engineering mode screen until “ZEro” appears in the sub Tb CONT MEM PEAK
display as shown in the right figure. AL AH
(2) By pressing key, the most significant digit of the main C
display will blink. F
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key.
The next lower digit will blink. ε
(5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
(6) To activate this programming, press ENT key. Sub display
Main display
Remarks • The programmable range is 0 to 6280ºC.
• The default is the minimum temperature value of the measuring range.

4.9-2 “Span” side adjustment …Adjustment by furnace temperature at span side


(1) To enter the span side temperature, press SEL key in the [Span side]
engineering mode screen until “SPan” appears in the sub Tb CONT MEM PEAK
display as shown in the right figure.
AL AH
(2) By pressing key, the most significant digit of the main C
display will blink. F
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key.
The next lower digit will blink. ε
(5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
(6) To activate this programming, press ENT key. Sub display
Main display
Remarks • The programmable range is 0 to 6280ºC.
• The default is the maximum temperature value of the measuring range.

- 17 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.10 Analog output correction programming
This mode is for the correction of the analog output by knee-points (up to 8 points).
It is necessary to program from [4.10.1 Programming of number of analog output data] to [4.10.5
Selection of analog output correction for programming item] continuously.
4.10.1 Programming of number of analog output correction data (number of knee-point)
Program the number of the knee-point. When “0” is programmed, no analog output correction is
executed. [Number of analog output
(1) To program the number of the analog output correction data, correction datas( number of knee-
press SEL key in the engineering mode screen until “CP.n” point)]
appears in the sub display as shown in the right figure. Tb CONT MEM PEAK

(2) By pressing key, the least significant digit of the main AL AH


display will blink. C
F
(3) Use or key to program the desired value.
(4) To activate this programming, press ENT key.
ε

Sub display
Main display

Remarks • The numeric figure in the main display is available from 0 to 9.


However, the programmable range is 0 and 2 to 8.
• The default value is 0.

4.10.2 Programming of analog output correction data No.


Program the analog output correction data No, for the data to be corrected.
(1) After the programming of the number of the analog output [Analog output correction
correction data (number of knee-point) in [6.5.8-1 data No.]
Programming of number of analog output correction data Tb CONT MEM PEAK
(number of knee-point)], press SEL key until “no” AL AH
(programming screen of analog output correction data No.) C
appears in the sub display as shown in the right figure. F
(2) By pressing key, the least significant digit of the main
display will blink.
ε
(3) Use or key to program the desired value.
(4) To activate this programming, press ENT key.
Sub display
Main display

• The numeric figure in the main display is available from 0 to 9.


Remarks
However, the programmable range is 1 to 8.
• The default value is 1.

- 18 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode

4.10.3 Entering of data before analog output correction


Enter the data before the analog output correction.
(1) After the programming of the analog output correction data [Entering of data before analog
No. in [4.10.2 Programming of analog output correction data output correction]
No.], press SEL key until “in” (entering screen of data Tb CONT MEM PEAK
before analog output correction) appears in the sub display as AL AH
shown in the right figure. C
(2) By pressing key, the most significant digit of the main F
display will blink.
(3) Use or key to program the desired value.
ε
(4) To shift to the next lower digit, press key.
The next lower digit will blink.
(5) Repeat the above procedure of (3) and (4) up to the least Sub display
significant digit for programming (4 digits). Main display
(6) To activate this programming, press ENT key.

Remarks • The programmable range is 0 to 6280ºC.


• The default is 0ºC.

4.10.4 Entering of data after analog output correction


Enter the data after the analog output correction.
(1) After the entering of the data before the analog output [Entering of data after
correction in [4.10.3 Entering of data before analog output analog output correction]
Tb CONT MEM PEAK
correction], press SEL key until “out” (entering screen of
data after analog output correction) appears in the sub display AL AH
as shown in the right figure. C
F
(2) By pressing key, the most significant digit of the main
display will blink.
(3) Use or key to program the desired value. ε
(4) To shift to the next lower digit, press key.
The next lower digit will blink. Sub display
(5) Repeat the above procedure of (3) and (4) up to the least Main display
significant digit for programming (4 digits).
(6) To activate this programming, press ENT key.

• The programmable range is 0 to 6280ºC.


Remarks
• The default is 0ºC.

- 19 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.10.5 Selection of analog output correction for programming item
This mode is for the selection of continuous entering of analog output correction data, execution of
correction command or exiting from this programming.
(1) After the entering of the data after the analog output [Continuous data entering]
correction in [4.10.4 Entering of data after analog output Tb CONT MEM PEAK
correction], press SEL key until “End” (selection screen of AL AH
analog output correction programming item) appears in the C
sub display as shown in the right figure. F
(2) By pressing key, the most significant digit of the main
display will blink.
ε
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key.
The next lower digit will blink. Sub display
(5) Repeat the above procedure of (3) and (4) up to the least Main display
significant digit for programming (4 digits).
(6) To activate this programming, press ENT key. [Exit from programming]
Tb CONT MEM PEAK
AL AH
C
[Correction programming items] F
cont: For continuous entering.
Press SEL key to return to the Programming
screen of [4.10.2 Programming of analog ε
output correction data No.]
CAnc: For exiting from this programming.
Press SEL key to move to the next mode. Sub display
Main display
Go: For execution of correction command.
[Correction command
execution]
Tb CONT MEM PEAK
AL AH
C
F

Sub display
Main display

Remarks • The default is cont (continuous entering).

- 20 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.11 Analog input programming(Model IR-CA 5 only)
4.11.1 Selection of analog input items
This mode is for the selection of the analog input (analog input disable, for remote emissivity
programming or for automatic emissivity calculation).
[Analog input disable]
(1) To select the analog input item, press SEL key in the Tb CONT MEM PEAK
engineering mode screen until “Ain” appears in the sub AL AH
display as shown in the right figure. C
(2) By pressing key, “no”, “ErMt” or “EAut” will blink F
in the main display.
(3) Use or key to select the desired type.
(4) To activate this selection, press ENT key. ε

Sub display
Main display

[Remote emissivity
[Analog input items] programming]
no: Analog input disable Tb CONT MEM PEAK
ErMt: For remote emissivity AL AH
programming C
EEAut: For automatic emissivity calculation F

Sub display
Main display

[Automatic emissivity
calculation]
Tb CONT MEM PEAK
AL AH
C
F

Sub display
Main display

Remarks • The default is no (analog input disable).

- 21 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.11.2 Emissivity scaling for remote emissivity programming
This mode is for the scaling of the emissivity value when the remote emissivity programming is
selected.
(1) To program the minimum emissivity value for the scaling, [ Minimum value of emisivity
press SEL key in the engineering mode screen until “AinL” scaling]
appears in the sub display as shown in the right figure. Tb CONT MEM PEAK
MEAS HOLD AL AH
(2) By pressing key, the most significant digit of the main
display will blink. C
F
(3) Use or key to program the desired value.
MAX AVE εr
(4) To shift to the next lower digit, press key. MIN TC NO
The next lower digit will blink. ε
(5) Repeat the above procedure of (3) and (4) up to the least C
significant digit for programming (4 digits).
Sub display
(6) To activate this programming, press ENT key. Main display
(7) To program the maximum emissivity value for the scaling,
press SEL key in the engineering mode screen until “AinH” [ Maximum value ofemisivity
appears in the sub display as shown in the right figure. scaling]
Tb CONT MEM PEAK
(8) Repeat the above procedure from (2) to (6) .
AL AH
C
F

Sub display
Main display

Remarks • The programmable range is 1.999 to 0.001.


• The defaults are:
AinL : The default is 0.050
AinH : The default is 1.999

Caution Confirmation of setting value by “remote emissivity programming”.


In the emissivity programming screen in the operator mode, you can confirm the
value programmed by the remote emissivity programming. However, even if you
change the analog input value on the condition that the programmed value is
displayed, the displayed value is not changed. For your confirmation of the new
value, it is necessary to return to the measurement mode and then display the
emissivity programming screen again.

- 22 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.11.3 Temperature scaling for automatic emissivity calculation
This mode is for the scaling of the temperature value when the automatic emissivity calculation is
selected.
(1) To program the minimum temperature value for the scaling, [Minimum value of
press SEL key in the engineering mode screen until “AinL” temperature scaling]
appears in the sub display as shown in the right figure. Tb CONT MEM PEAK
(2) By pressing key, the most significant digit of the main AL AH
display will blink. C
F
(3) Use or key to program the desired value.
(4) To shift to the next lower digit, press key.
The next lower digit will blink.
ε
(5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
Sub display
(6) To activate this programming, press ENT key. Main display
(7) To program the maximum temperature value for the scaling,
press SEL key in the engineering mode screen until “AinH” [Maximum value of
appears in the sub display as shown in the right figure. temperature scaling]
(8) Repeat the above procedure from (2) to (6). Tb CONT MEM PEAK
AL AH
C
F

Sub display
Main display

Remarks • The programmable range is 0 to 6280ºC.


• The defaults are:
AinL (Minimum value) … Minimum value of the measuring range
AinH (Maximum value) … Maximum value of the measuring range

- 23 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.12 Communications programming(Model IR-CA S only)
This mode is for the programming of the address (effective in multi-drop connections) of this
thermometer and communications speed.
4.12.1 Address programming
Program the address within from 1 to 32.
(1) To program the address of this thermometer, press SEL key [ Address ]
in the engineering mode screen until “Adr” appears in the Tb CONT MEM PEAK
sub display as shown in the right figure. AL AH
(2) By pressing key, the least significant digit of the main C
display will blink. F

(3) Use or key to program the desired value.


(4) To activate this programming, press ENT key.
ε

Sub display
Main display

Remarks • The numeric figure in the main display is available from 1 to 99.
However, the programmable range is 1 to 32.
• The default value is 1.

4.12.2 Communication speed selection


Select the communication speed.
(1) To select the communications speed, press SEL key in the [ Communication speed ]
engineering mode screen until “SPd” appears in the sub Tb CONT MEM PEAK
display as shown in the right figure. AL AH
(2) By pressing key, “4.8”, “9.6” or “19.2” will blink in the C
main display. F

(3) Use or key to select the desired type.


(4) To activate this selection, press ENT key. ε
[Communications speed]
4.8: 4800 bps
Sub display
9.6: 9600 bps Main display
19.2: 19200 bps

Remarks • The default value is 9.6 (9600bps).

Caution To return to the measurement mode screen from the engineering mode screen
• The screen will be automatically returned to the measurement mode screen if
any key is not pressed for 1 second in the programming of each parameter.
• The screen will return to the measurement mode screen when the programming
from [4.1 Analog output scaling] to [4.10 Analog output correction] is
completed.
Note: Up to [4.10 Analog input programming] in case of model IR-CA 5 .
Up to [4.12 Communications programming] in case of model IR-CA S .

- 24 -
INST.No.INE-453-P3 Software Ver.1.0□

4. Engineering mode
4.12.3 Communication protocol mode selection
Select the communication protocol, MODBUS protocol mode or IR-CA series exclusive protocol
mode.
(1) To select the communications protocol mode, press SEL key [MODBUS RTU mode ]
in the engineering mode screen until “Prot” appears in the Tb CONT MEM PEAK
sub display as shown in the right figure. AL AH
(2) By pressing key, “rtu” or “PriV” will blink in the main C
display. F

(3) Use or key to select the desired mode.


(4) To activate this selection, press ENT key. ε

[Communications protocol mode]


rtu: MODBUS RTU mode Sub display
Main display
PriV: IR-CA series exclusive protocol mode.
[ IR-CA series exclusive
protocol mode]
Tb CONT MEM PEAK
Remarks • The default is rtu (MODBUS RTU mode). AL AH
C
F

Sub display
Main display

4.13 Key lock


This mode is for the locking of the function keys temporally when you do not want to change the
programmed parameters.
(1) To activate the key lock, press SEL, and ENT keys [ Key lock ]
simultaneously for 2 seconds in the measurement mode Tb CONT MEM PEAK
screen until “Lock” appears in the sub display. AL AH
(2) During the key lock, “Lock” is displayed in the sub C
display as shown in the right figure. F
(3) To release the key lock, press SEL, and ENT keys
simultaneously for 2 seconds in the condition that “Lock”
is displayed in the sub display. ε
(4) “Lock” will disappear and the screen will return to the
measurement mode screen. Sub display
Main display

Remarks • The default is none (Key lock not using).

- 25 -
INST.No.INE-453-P3 Software Ver.1.0□

5. Self-diagnostic function and Overflow/underflow


5.1 Self-diagnostic function
This self-diagnosis function is built-in. An error No will be displayed on abnormal conditions.
Output
Sub display Contents Countermeasure
(Note*)

The ambient temperature is Use the thermometer in the Yes


abnormal. environment from 0 to 50ºC.

*1 Motor revolution is Built-in motor is abnormal. Return to Yes


abnormal. factory to us, and order the repair.

E2PROM abnormal.
(Writing and reading Return to factory to us. Yes
impossible)
Analog output correction
data abnormal. (Same data Check the data before correction again. No
before correction existed.)
Zero/span adjustment
abnormal. Adjust zero and span again. No
(Data at zero > data at span)
Program to 0.001 when the emissivity
Remote emissivity programming value is overshooting
programming exceeding the from the minimum value and 1.999 No
programmable range. when it isovershooting from the
maximum value.

Automatic emissivity Program the emissivity value to1.000.


calculation exceeding the However, this programming is not No
programmable range. stored.

*1 : Displayed only IR-CAE, IR-CAG, IR-CAP, IR-CAR, IR-CAT and IR-CAU.

On the items with “Yes” in the output column, the contact signal for the abnormal
Caution condition is outputted from the contact output terminals (OFF at abnormal
condition).
This output is available when “Self-diagnostic abnormal” is selected in
[6.5.5 Contacts output item selection] in the engineering mode.

5.2 Overflow/underflow
● “oFL” is displayed when the measured temperature is higher than (the maximum value of
measuring range + 20)ºC.
● “uFL” is displayed when the measured temperature is lower than (the minimum value of
measuring range – 20)ºC.
[ Overflow ] [ Underflow ]

- 26 -
INST.No.INE-453-P3 Software Ver.1.0□

6. List of starting up modes


6.1 Starting up modes
Keys Screen Remarks Paragraph
Temperature measurement
Measurement 1
(Power on) and parameter programming
mode 3
in the operator mode
Press SEL and ENT keys simultaneously in Engineering Parameter programming in 2.2
measurement mode screen. mode the engineering mode 4
Press SEL, and ENT keys simultaneously Locking/release of function
Key lock 4.13
for 2 seconds in the measurement mode. keys

6.2 Screens
Screen Outline
∙Measurement starts by power on.
∙By pressing SEL key, the screen moves to the parameter programming screen (for
Measurement emissivity value, signal modulation, alarm and other parameters) in the operator mode.
mode (Ref. [3 Operator mode])
∙The screen automatically returns to the measurement mode screen when the programming
up to [3.5Alarm programming] is completed or if any key is not pressed for 1minute.
∙By pressing SEL and ENT keys simultaneously, the screen moves to the parameter
programming screen (for analog output scaling, analog dummy output, hold function,
analog output correction, optional analog input, optional communications and other
Engineering
parameters) in the engineering mode. (Ref. [5 Engineering mode])
mode
∙The screen automatically returns to the measurement mode screen when the programming
up to [4.10 Analog output correction programming] is completed (except the thermometer
with optional specifications) or if any key is not pressed for 1minute.
∙The key lock is used not to change the parameters being programmed.
∙By pressing SEL, and ENT keys simultaneously for 2 seconds in the measurement
Key lock mode, the function keys are locked.
∙For releasing, pressing SEL, and ENT keys simultaneously for 2 seconds again.
(Ref. [6.6 Key lock])
∙The laser (option) is targeted or stopped.
Laser ∙By pressing SEL and keys simultaneously for 2 seconds in the measurement mode, the
targeting laser is targeted.
(option) ∙For releasing, pressing SEL and keys simultaneously for 2 seconds again.
(Ref. [4.13 Laser targeting])

7. Parameter programming/selection
7.1 Parameter programming/selection (operating mode)
Parameter item Sub display Parameter Default Paragraph
Emissivity ε 1.999 to 0.050 1.000 3.1
(Sub marker)
Automatic Auto 0 to 6280ºC Minimum value of 3.2
emissivity calculation measuring range
Signal modulation mode Modu dELy、PEAk dELy 3.3
Modulation degree *1 tAu 0.0 to 99.9 (second) or 0.0 second 3.4.1
dEc 0.00 to 99.9 (second) 0ºC/second 3.4.2
0, 2, 5, 10ºC (ºF)/second
Alarm type A.Mod no: Alarm disable Hi 3.5.1
Hi: High alarm
Lo: low alarm
Alarm set - point ALM 0 to 6280ºC Maximum value of 3.5.2
measuring range
Minimum value of
measuring range
*1: The parameters for the modulation degree differ by the signal modulation mode selection.
(Ref. [6.4.3 Signal modulation mode selection.])

-27-
INST.No.INE-453-P3 Software Ver.1.0□

7. Parameter programming/selection
7.2 Parameter programming/selection (engineering mode)
Parameter item Sub display Parameter Default Chapter
Analog output scaling OutL or OutL: Minimum value (0 to 6280ºC) Minimum and 4.1
OutH OutH: Maximum value (0 to 6280ºC) maximum value of
measuring range
Analog dummy output OutC 0 to 100% 0% 4.2
Hold functions type HoLd no: Hold function disable no 4.3.1
PEAk: Peak hold
SAMP: Sample hold
Peak hold reset type H.rSt no: Reset disable no 4.3.2
in: Internal - time reset
Et:External -remote contact reset
Peak hold reset time r.tiM 0.0 to 99.9 seconds 0.0 second 4.3.3
Measurement unit unit C, F C 4.4
Modulation time tAu.d Decimal point selection 2 4.5
constant unit
Contact output dout no: No contact output ALM 4.6
ALM: High/low temperature alarm
Err: Self-diagnostic abnormal
Automatic emissivity E.Aut no: Calculation disable no 4.7
calculation in: Internal - Key
Et: External – Analog input
Measuring wevelength dEtc 1coL,tP,inGA,Si 1coL 4.8
selection
Zero/span CAL Minimum and 4.9
adjustments ZEro 0 to 6280ºC maximum value of
SPAn measuring range
Analog output CP.n Number of analog output 0 4.10.1
correction no correction data: 0, 2 to 8 1 4.10.2
in Analog output correction data No.: 1 to 8 0ºC 4.10.3
Out Data before analog output 0ºC 4.10.4
End correction: 0 to 6280ºC cont 4.10.5
Data after analog output
correction: 0 to 6280ºC
Analog output correction programming
Cont: Continuous entering
Canc:Exit from programming
Go: Correction execution
Analog input Ain no: Analog input disable no 4.11.1
programming ErMt: Remote emissivity programming
EAut: Automatic emissivity calculaiton
Emissivity scaling AinL Remote emissibity Ain L : 0.0050 4.11.2.
for remote emissivity AinH programming:ε = 0.001 to 1.999 Ain M: 1.999
Temperature scaling AinL Automatic emissivity calculation Minimum and 4.11.3
for automatic emissivity AinH :Temp = 0 to 6280ºC maximum value of
calculation measuring range
Communications Adr Address: 1 to 32 1 4.12.1
SPd Communications speed: 9.6 4.12.2
4.8 … 4800bps
9.6 … 9600bps
19.2 … 19200bps
Prot Protocol mode rtu: MODBUS rtu 4.12.3
Priv: IR-CAexclusive
Key lock Lock Lock: Locking the function keys None 4.13

-28-
32-8, KUMANO-CHO, ITABAS
HI-KU, TOKYO 173-8632

Telephone: +81-3-3956-2171
Facsimile: +81-3-3956-0915
Web site http://www.chino.co.jp/

INE-453-P3 Sep-'04 IR-CA Series Single color radiation thermometer [Software volume]