Académique Documents
Professionnel Documents
Culture Documents
User Manual
Schematic Diagram
Complete View
Radiator Head
Cable
Radiator Head
Connection
Adjustable
Radiator Head
Emission Field
Selection Knob
Arm for Radiator
Head
Housing with
Ventilation Opening
Carriage X-Base on
Rollers
1
Schematic Diagram
Back View
Connection for
Radiator Head
Cable
Housing with
Ventilation Panel
2
Schematic Diagram
Top View: Console
Power Display
Frequency Display
Digital Therapy Time
Display
Stand-by Display
Mains Display
Pulsed Display
Continuous Display
Operational Mode
Key (Pulsed /
Continuous)
Start/Stop Key
Therapy Time
Adjustment Keys
Intensity Control
Knob
Control Lamp
3
Micro 5
This User Manual is to be considered as part of the device itself and is therefore always
to be kept in close vicinity of the device. Careful attention to the User Manual is the
supposition for correct usage and handling of the device as well as for patient and
operator safety.
Table of Contents 4
1. Device Description 5
2. Assembly Instructions 5
3. Functional Test 6
5. Operation 8
6. Indications/Contraindications 12
8. Technical Information 14
9. Delivery / Accessories 15
10. Cleaning/Disinfecting 15
12. Maintenance 16
4
1. Device Description
What is Micro 5? Micro 5 is a microwave therapy device used to treat muscular-skeletal disorders as well
as inflamed conditions.
Device application is reserved for professional use only (i.e. medical doctors, medically
trained personnel).
How does it function? Microwave therapy results when electrical energy is transformed into warming energy in
the tissues. This penetrating warmth produced by microwave methods could not be
obtained with outer warmth applications (i.e. warm packs onto the skin), because it would
be channeled away by the blood circulation.
What are the advantages of Micro 5 is easy to apply and the Adjustable Radiator Head provides for three different
Micro 5? emission field settings for various treatment situations.
What other innovative Two selectable pulse frequencies provide a warming pulse, which is distinctly
Features are offered? perceptible.
Larger penetration depths can be achieved through this new pulse method, which allows
higher intensities to be used with constant medium power levels.
2. Assembly Instructions
Mounting the Arm for the Radiator Head Push the arm into the provided holder located on the back of the user console.
Mounting the Radiator Head Cable Insert the ball joint of the Radiator Head Cable into the holder on the arm and affix with
the locking screw. Plug the free cable end into the backside of the device using the
provided connection plug, which is located underneath the connection holder for the
Radiator Head Arm.
Mounting the Adjustable Radiator Head A Metal Feather Clip is provided on the connection plug in order to secure the
Adjustable Radiator Head.
The Adjustable Radiator Head can only be mounted or dismounted when this Clip is
removed. Afterwards, the ring (located underneath the Clip) can be pushed back to
allow the Radiator Head to be plugged in or out.
5
Mains Connection on the Device Micro 5 may only be connected to a safe current supply intended for medical rooms:
Application Group 1 in accordance with VDE 0107.
Functional Test After the device is correctly assembled, double check to insure that the Connection Cable
for the Radiator Head is plugged in perfectly and secured properly with the Metal
Feather Clip!
Proceed by performing a Functional Test for the device.
Attention! Pay close attention to insure exact cable connection and Feather Clip
position! The device may only be operated with correct Radiator Head Cable
connection! Incorrect connection could danger the environment, device, or persons in
the vicinity.
Attention! Remember that microwave exposure can penetrate walls and protection
shields, and could result in damage to other technical or medical devices. Choose the
position and location for the device carefully. Prior advice from the technical building
administrator is suggested. To avoid electrical interference, operate the device alone in
rooms with protection shields.
3. Functional Test
• Adjust the Emission Field Selection Knob on the Adjustable Radiator Head to the
“Circular Field” position, and turn on the device.
• After switching on the device, insure that all display elements (lights and LED digit
display) are functioning correctly and listen for the Acoustic Signal Tone that
indicates the end of the Automatic Self-Test.
• Press the Start/Stop Key and select the Pulsed Operational Mode (Pulsed
frequency 0.5 Hz) and a power selection of L 50 W (Low power).
• Move the Control Lamp (see Accessories on Page 15) through the emission field to
within approx. 10 cm in front of the Radiator Head. The light must light up in a
glimmer. Otherwise, assure that the device cannot be used further and contact
customer service.
6
4. Brief Start-up Procedure
Sequence of Operation Turn on device using the Mains Switch and observe the following: the Mains Display
lights up, an Acoustic Signal beeps indicating the end of the Automatic Self-Test, and the
Software Version Number is displayed. Micro 5 is ready for operation.
Position the Radiator Head towards or near the treatment area. Pay close attention to
insure that the correct Radiator Head Adjustment is selected and corresponds as closely
as possible to the treatment area. Follow the instructions for positioning the Radiator
Head precisely.
Select the appropriate Pulse Frequency and Operational Mode, “Pulsed or Continuous”.
Begin treatment by turning the Intensity Control Knob, observing the dosage guidelines.
7
5. Operation
To turn on device, switch the Mains Switch (located on the backside of the device) into
position I. An Automatic Self-Test follows. All display elements are tested.
An Acoustic Signal indicates a successful test run. The Software Version Number is
displayed. The device is now ready for operation.
Turn off device using the Mains Switch after treatment or during extended treatment
pauses.
The Stand-By Display blinks for 5 seconds when the Intensity Control Knob is turned
during the Automatic Self-Test.
Afterwards, the Stand-by Display is constantly lit and further dosages can be selected by
adjusting the Intensity Control Knob.
This display is lit when the Continuous Operational Mode is selected. After the
device has been switched on, the Continuous Operational Mode is activated.
This display is lit when the Pulsed Operational Mode has been selected.
The Power Display indicates the actual Microwave Power Output in Watt. The “Low
Power Operational Mode” is indicated by an “L” Display to the left.
Should an error or fault be detected, an “F” would be displayed in the Power Display
followed by an Error Number (See Chapter 13).
The Frequency Display indicates the selected pulse frequency in the Pulse
Operational Mode.
The Therapy Clock indicates the Pre-selected Therapy Time and the actual remaining
Therapy Time during the treatment. An Acoustic Signal indicates the end of the treatment
and the power setting returns to its original position.
When the device is turned on, the Therapy Time is pre-set for 10 minutes.
8
The Start/Stop Key interrupts the therapy during treatment. If it is pressed again, the
Therapy Time will re-set to its beginning value.
The Start/Stop Key must be pressed after the device is switched on to release the other
keys and Intensity Control Knob for operation.
The Operational Mode Key functions as a cyclic switch when pressed repeatedly,
allowing the user the choice between the following Operational Modes: -Continuous,
Pulsed 0.5 Hz Low Power, Pulsed 1 Hz Low Power, Pulsed 0.5 Hz, Pulsed 1 Hz. An „L“
in the Power Display indicates the Low-Power Operational Modes.
The keys and Intensity Control Knob are only released for operation once the
Start/Stop Key is pressed after the device has been turned on.
The device is in the Continuous Mode when it is switched on.
The Increase Therapy Time Key increases the Therapy Time in one-minute steps
each time when it is pressed; maximum is 30 minutes.
The Decrease Therapy Time Key decreases the Therapy Time in one-minute steps
each time when it is pressed until it reaches “0”. Treatment is not possible when the
therapy time reaches “0”.
The Intensity Control Knob allows the microwave power to be increased in steps of 2
(5 Watt) independently from the selected Operational Mode.
Turning the Intensity Control Knob clockwise increases the power; counter-clockwise
decreases the power.
A maximum of 250 Watts can be selected in Continuous Operational Mode; a maximum
of 125 W in Pulsed Operational Mode; a Maximum of 50 W in the Low-Power
Operational Mode.
After the device is ready for operation (Mains Light is on and the Standby Display is out)
and the Intensity Control Knob is turned for the first time, an Automatic Balancing Test is
performed during which the Standby Display blinks.
After 5 seconds, the Standby Display remains lit and the desired microwave power
dosage can be selected and the therapy treatment can begin.
9
Adjustable Radiator Head The Emission Field’s size can be adjusted to meet the requirements of the bodily area to
be treated.
The Emission Field should be enlarged or reduced to correspond as closely as possible
to the treatment area.
Radiator Head Positioning The Radiator Head should be positioned no further than 10 cm maximum from the bodily
area to be treated.
Attention! When positioning the Radiator Head and adjusting the Emission Field,
ensure that other bodily sections (eyes, genitals) are not unintentionally exposed. When
necessary, change the position of the patient or Radiator Head.
Attention! Pay close attention to ensure that the power output has been switched
off when positioning the Radiator Head in order to avoid accidental exposure to the
patient or therapist.
Attention! Special attention is required during treatment where the Radiator Head
is very close to the treatment area with respect to the restricted heat circulation.
Attention! Metal objects (such as jewelry, belt buckles, wallets, clothing with metal
threads, implants etc…) are prohibited from the radiation exposure field!
Observe the Warning Instructions in the User Manual when positioning the
Radiator Head.
10
Half Field – Right Adjustment
Dosage The power can be increased as long as the patient has a subjectively comfortable, warm
feeling. The dosage should not be increased during treatment, even if the patient’s
warmth perception decreases.
The therapist should continue to inquire into the patient’s subjective feeling during the
entire treatment. Should the treatment become uncomfortable or painful, the intensity
should be decreased or the treatment stopped.
The general dosage rule is as follows: The more acute the condition, the weaker the
dosage. Correspondingly, chronic conditions can be radiated with higher dosages. The
treatment duration is shorter with acute conditions and longer with chronic ones
(generally between 10 and 15 minutes). The Treatment Sessions can be daily with 6 –
12 treatments sessions per Treatment Series.
Protection Glasses (special accessory) are to be worn when treating the head area in
order to protect the eyes.
Patients with implanted cardiac pacemakers are contraindicated.
11
6. Indications / Contraindications
12
7. Warnings and Instructions
Pay close attention to insure § During the therapy session, the therapist is not in the emission field or exposure
that: area. Persons not to be treated should remain at least 1.5 meters away from the
Radiator Head.
§ Metal objects such as jewelry, clothing with metallic threads, belt buckles etc. must
be removed from the treatment area.
§ Metallic surfaces (couches, washbowl, device housings…) are not included in the
emission field.
§ The intensity is reduced to “0” when the Radiator Head is being positioned.
§ The Radiator Head is never directed toward the eyes or genitals. Even when
other bodily sections are to be treated, the Radiator Head must be positioned to
insure that the eyes and genitals are not included in the emission field or exposure
area.
§ The Radiator Head does not pass or be directed over the patient’s face; injury can
result from unobserved movements.
§ Bodily areas with implanted metallic materials (such as bone screws, snaps, artificial
joints, etc.) are contraindicated.
§ Patients with Impaired Circulation or Sensation are contraindicated
§ Patients with implanted pacemakers are not treated and not in the Diathermy
treatment rooms (such as where Micro 5 microwave therapy is being applied).
§ Hearing aids and other devices, which the patient has on his body, are removed
before treatment. Otherwise, they can malfunction or become permanently
damaged.
§ Patients being treated in the head region wear eye protection glasses.
§ The Radiator Head, Cable and Connection Cables are handled with the
necessary care. External defects (bumps and dents) can alter the exposure
characteristics.
§ The device’s housing is never opened, as internal high voltages can be life
threatening.
§ The device is never operated with a defective Radiator Head or Cable.
§ The device is only operated with a Closed Connection Plug that is secured
properly with a Metal Feather Clip.
§ The heat produced from the device can always circulate; the ventilation vents must
always be kept dirt and dust-deposit free to allow the air to circulate.
§ The device is not operated in humid environments (such as hydro-therapy rooms,
near Saunas, swimming pools, bathing rooms etc.).
§ That should fluids or foreign particles penetrate the housing, the plug is immediately
unplugged and the device is not operated until approved for operation by
authorized customer service.
§ Beams exposed from the Radiator Head do not directly hit electrical devices in the
near vicinity. The devices can malfunction or be damages irreparably.
§ Medical devices that are simultaneously used maintain a safety distance of at least 5
meters; should disturbances occur, stop therapy. We would be glad to supply
additional information concerning necessary shielding measures, when applicable.
13
8. Technical Information
Precision ± 20%
The power outputs refer to the average output; the peak output value is max. 1200 W.
Dimensions 33cm x 42cm x 72cm (excluding Radiator Head and Carriage Base)
33cm x 42cm x 83cm (excluding Radiator Head)
Weight 38.5kg
Symbol Explanation
Applicable Section Type B Non—ionizing Radiation
14
9. Delivery / Accessories
Switch off and unplug the Mains Cable before cleaning the device.
Cleaning The device can be cleaned with a standard plastic cleaner (diluted soapy water, dish
soap). Apply with a damp cloth. Fluids or moisture must not penetrate the device or plug
connections!
Disinfecting Standard disinfecting solution for medical equipment can be applied by wiping with a
damp cloth (not by spraying!). Follow the manufacturer’s instructions and dosages.
Never spray solution directly onto the device! Moisture must not be allowed
to penetrate the housing or plug connection!
The device must never be disinfected with steam, gas or gamma rays.
15
11. CE Conformity Declaration
This product bears the marking 0123 in accordance with the EG Guidelines for
Medical Products 93/42/EWG and fulfills the basic requirements stated in the Appendix of
these guidelines.
This device has been categorized as Class 11b in accordance with Appendix 1X of the
guidelines.
Micro 5 is manufactured and tested using the Quality Management System in accordance
with DIN ISO 9001/EN 29001 und EN 46001.
12. Maintenance
The Radiator Head and Cables are to be checked for damage before each application.
Check to insure that the connection for the Radiator Head Cable is securely fastened and
pay special attention to insure that the Metal Feather Clip is properly secured.
Check the ventilation vents regularly to insure that they are free from dirt and dust
deposits and are open to allow air circulation.
Insure that malfunctioning devices are not operated and are prevented from
further operation.
16
13. Malfunction / Error Indications
Although the device is switched on, the Check Mains Supply to insure that there is adequate current supply at the Mains
Mains Display Light does not come on: Connection Plug.
Check the Mains Plug Connections on the device and wall plugs.
Check the Mains Fuses located on the backside of the device.
The patient has no subjective warmth Check to insure that the device is installed correctly in accordance with Chapter 2
feeling of the User Manual.
Proceed with a Functional Test according to Chapter 3.
Should a malfunction exist, contact the customer service.
Error Indication Numbers The device performs an Automatic Self-Test during operation. Should an error or
malfunction be detected, the User is notified by an Error Indication Number, which is
displayed on the Power Display. Error Numbers begin with the letter “F”.
The therapy will be interrupted, and Acoustic Signal is given, and the user elements are
blocked.
Some errors can be corrected by simply switching the device off and re-starting.
Otherwise, the customer service is required.
F 42 The device has over-heated. Switch off device. Then, switch it back on and wait
approx. 2 minutes before turning the Intensity Control Knob. Operation should be
possible after this pause. Otherwise, contact the customer Service.
Other Error Indication Numbers Prevent the device from being operated and contact your customer service.
17
Zimmer
MedizinSysteme GmbH
Junkersstrasse 9
D-89231 Neu-Ulm
Tel. +49 731/9761-291
Fax +49 731/9761-299
export@zimmer.de
www.zimmer.de