Académique Documents
Professionnel Documents
Culture Documents
Analytical methodology
Nuclear Instruments and Methods in Physics Research B77 (1993) 95-109
North-Holland
NIIIMB
Beam Interactions
with Materials&Atoms
Methods of standardization in quantitative micro-PIXE analysis are reviewed and various issues that bear on analytical accuracy
are explored; pertinent recent work on Si(Li) X-ray detector response is included and some geochemical examples are drawn upon.
Extension of the GUPIX software to deal with multilayer targets, including secondary fluorescence within and between layers, is
reported, analytical examples include alloy foils and multilayer solar cell structures.
0168-583X/93/$06.00 0 1993 - Elsevier Science Publishers B.V. All rights reserved III. ANALYTICAL METHODOLOGY
96 J.L. Campbellet al. / PIXE microanalysisof thick specimens
creased; a wide range of specimen types may be han- the microcoulomb (l.&), we define lr(Z, Cz,,) as the
dled with a very small suite of well-chosen standards. computed or theoretical X-ray intensity per PC per
The greatest flexibility is achieved by reverting to a steradian and per unit concentration. The Cz,, are the
“first principles” approach that abandons standards concentrations of the major elements which comprise
entirely. This too is possible in p,-PIXE, given the the specimen matrix and thus determine ME. Then the
simple physics and the excellent data base. But it measured intensity for Q ~J,Cof charge is
demands absolute measurement of proton beam flux,
which is less straightforward than superficial examina-
tion might suggest. At an early stage in thick target There are several important points to note. First, the
PIXE’s development, Clayton 131 demonstrated lo- IE, viz. the terms 0, &. and tz, factor out; they will
15% systematic errors in such analyses based on first therefore cancel in the comparison of specimen (SP)
principles. The discrepancies were a function of speci- and standard (ST) for X-rays of a particular element
men type, and Clayton corrected them by analyzing z:
reference materials of the appropriate type. By so
doing, he effectively demonstrated the efficacy of the IS’(Z) c;p I,“‘(Z, Cz,,)
___=-
third of Newbury’s approaches, as outlined in the pre- IS’(Z) c;= I,“(& Cz,,) . (4)
vious paragraph.
In addition, measurement of Q needs to be correct
2.2. The matrix correction approach only in relative terms between specimen and standard.
The ME are represented by the term I&Z, Cz,,),
The PIXE version of eq. (1) has appeared fre- which involves both the compound stopping power and
quently in the literature; it results from an integration the compound attenuation coefficient, which are con-
of the X-ray intensities recorded from each point on centration-weighted sums’ of these quantities over the
the linear path, within the specimen, of the decelerat- major constituents. If the C,,, subset is known, e.g.
ing proton: from prior EPMA analysis, then the I,(Z, C,,) are
immediately calculable, and eq. (4) offers a one-step
solution for the trace element concentration subset
CZ,t. The further advantage of this ratio approach for
each Z-value of interest is the additional cancellation
of fluorescence yields and relative X-ray line intensi-
ties, and the partial cancellation of ionization cross-
X
sections within the integrals. The degree of partial
cancellation of stopping power and attenuation coeffi-
(2) cients will increase with the similarity of the specimen
In this equation, for the K shell case: Z(Z) is the and the standard. We have discussed the accuracy of
intensity in a particular K X-ray line, usually but not the data base in some detail elsewhere [4] and with one
necessarily the dominant K, or Kal; Np is the number exception will not repeat that here. We simply note
of protons of energy E, incident at angle a! to the that the data base is sufficiently well known that a
normal upon the specimen, and uz(E) the ionization generic, as opposed to a specific, similarity between
cross-section for the K shell; wz is the corresponding specimen and standard will suffice. The exception per-
fluorescence yield and b, the fraction of K X-rays in tains to elements of very low atomic number, which are
the selected line; N, is Avogadro’s number; A, is the of increasing interest to PIXE analysts, and will be
atomic mass of element Z; S(E) is the proton stopping dealt with in section 3.4.
power; 6ro is the X-ray take-off angle; flci is the The situation is more complex, but entirely tractable,
absolute efficiency of detection, which involves both when the major element concentrations Cz, are un-
Si(Li) detector solid angle and intrinsic efficiency of known, a situation that we encounter with ‘increasing
the silicon crystal; tZ is the attenuation of the X-rays in frequency in our geochemical work. The approach here
any absorber (filter) placed between specimen and is to begin by estimating a set of values for the Cz,,,
detector; p/p is the attenuation coefficient within the compute the corresponding I,(Z, C,,,), fit the meas-
specimen matrix for the X-ray line of interest. ured X-ray spectrum by least squares to obtain I(Z),
The L and M-shell equivalents are essentially the and via eq. (4) deduce a new set of C,,,. This process
same, albeit with the additional complication of sub- is iterated until a self-consistent set of Cz,, is ob-
shells; they thus involve the Coster-Kronig probabili- tained. Computing overheads have long ceased to be a
ties that transfer the proton-induced vacancies among deterrent to this inclusion of one iterative process
subshells of a major shell prior to X-ray emission. (spectrum fit) within the iterations of another (con-
The practical unit of integrated beam charge being centration determination).
J.L. Campb~~ et at. / PIXE ~icroa~a~~isofthick specimens 91
Eq. (4) presents a choice as regards the standard. The characteristic X-rays of major elements may
We indicated above that, given an accurate data base induce significant secondary fluorescence both of trace
from which to calculate ME, only a general as opposed elements and other major elements. We have incorpo-
to a close resemblance was necessary between speci- rated a full treatment of secondary fluorescence in
men and standard. In this spirit we use synthetic sul- GUPIX92. Its accuracy has been examined elsewhere
fide minerals, containing known quantities of one or [S] and we have further comments later in this regard.
two trace elements, to standardize measurements upon
a wide variety of oxide, silicate and other minerals.
However if the concentration of one major element in 3. Major element anaiysis: practical aspects
the specimen is known (e.g. from EPMA) then the
specimen itself may serve as standard; an example is Micro-PIXE has to date been concerned much more
the analysis by Rogers et al. [5] of the NIST standard frequently with trace elements than with major ele-
610 (trace metals in glass), effected relative to the ments, the latter frequently being predetermined by
major constituent, calcium. This confers the further EPMA in a complementary use of the two methods.
advantage of cancelling any residual errors in proton However, geochemists (for example) who have to pur-
stopping powers, which are now identical in specimen chase time outside their laboratory for both techniques
and standard. may well be accommodated in full by micro-PIXE, to
The iterative solution discussed above for major their obvious economic benefit.
elements has been incorporated in the GUPIX soft- The advantages and disadvantages of the standard-
ware package (GUPIX92 is the current version) for ization approach embodied in eq. (4) i.e. of an ap-
fitting spectra and providing concentrations 161.A use- proach based on computation of ME and cancellation
ful concept, due originally to Lagarde et al. [7], is that of IE are now explored. The examples used are mainly
the matrix may include an “invisible element”, i.e. one from our own materials science and geochemistry expe-
whose X-rays are too low in energy to appear in the rience.
spectrum. The condition is then imposed that all con-
centrations sum to 100% and the invisible element 3.1. Attenuation and secondary fluorescence uncertain-
concentration is iterated with the others in the compu- ties
tation of concentrations. GUPIX92 allows there to be
more than one invisible element, with the necessary Pure single-element standards are convenient and
proviso that their ~ncentrations remain in a specified attractive in that they preferentially transmit their own
ratio; this would permit for example the presence of characteristic X-ray lines, thus minimizing the impact
SiO, as a defined species. EPMA practice suggests a of X-ray attenuation coefficient errors on the analysis.
useful variant for the specific case of the wide range of In our work on silver and platinum-group elements as
silicate minerals where the elements are all present in traces in sulfide minerals, it is sometimes necessary to
oxide form. This is to deduce the oxygen contribution examine major element variations. An example is iron
for each oxide by applying the stoichiometric ratio to and nickel in pentlandite, where there is interest in a
each cation concentration; one then can compare the smuggested [91 linear correlation between the trace Pd
total of all element concentrations to 100% as a fur- content and the Ni/Fe ratio. The major and trace
ther check of the analysis. analyses can be effected in a single measurement with
Table 1
Results of PIXE analyses of Fe/Ni alloys. Cases A use theoretical p/p for Fe; B uses measured I( /p
Standard Element Reference Concentration [%I 10
PIXE (A) PIXE (B)
NIST SRM 1159 (thick target) Si 0.5 0.24 0.24 0.01
Cr 0.069 0.07 0.07 0.003
Mn 0.3 0.29 0.29 0.01
Fe 51.0 51.7 51.2 0.1
Ni 48.2 47.3 48.4 0.1
lnvar FOBS a (10 pm thickness) Ti - 0.1 0.1 0.005
Cr 0.12 0.12 0.004
Mn 0.34 0.34 0.007
Fe 64.0 63.2 63.4 0.1
Ni 36.0 34.2 35.3 0.1
a Mn + Si + C < 1%. hence Fe and Ni slightly below nominal values.
3 MeV protons, using a 0.35 mm Al absorber to prefer- gies above and within 1 keV of an absorption edge. A
entially suppress the dominant Ni and Fe peaks. Due detailed experimental study of this region for iron [II]
to the absorber, no Si X-rays are detected and so Si is suggests a p/p value for NiK, that is 6% above
treated as an “invisible element”, the total Csi + C, theory. Fig. 1 presents a comparison of the measured
+ C, being 100%. The standards are pure iron and and theoretical p/p values in the vicinity of the iron K
nickel targets. absorption edge. When the g/p values from ref. [ll]
This appears straightforward, but we felt that the are substituted, together with the corresponding photo-
procedure warranted testing. An appropriate way to do electric cross-sections, in the GUPIX data base [6],
this is by analysis of the NIST nickel-iron alloy stand- agreement between our data and the recommended
ard reference material (SRM 1159>, since it has 48% of concentrations improves significantly. We do not have
nickel and 51% of iron. We find consistently (see table similarly extensive experimental p/p values for attenu-
1) that we underestimate the nickel concentration and ation in nickel, and so the correction of the data base
overestimate that of iron. This is unlikely to arise from is necessarily incomplete.
inhomogeneity in the standard, since the microbeam is The point that we wish to demonstrate is simply
rastered over an area 0.4 x 0.4 mm. We have observed PIXE’s susceptibility to error in attenuation coeffi-
the effect before [8], in conventional broad-beam cients at X-ray energies near absorption edges. This
thick-target PIXE with a different detector. And (see kind of error is likely to be encountered for other
section 5 below) we observe it also in thin Invar alloy element pairs spaced by one or two Z-values. GUPIX
foils. has therefore been modified to provide a user-friendly
This discrepancy is not attributable to either ioniza- means by which a user may substitute preferred experi-
tion cross-sections or stopping powers since we are mental p/p values for those in the data base. In
dealing with elements which are separated by only two addition, these observations argue against the use of
units of atomic number; any discrepancies would be semiempirical formulae for attenuation coefficients,
the same for both, and would tend to cancel. Suspicion since these cannot accommodate the oscillatory be-
therefore falls on the attenuation coefficients p/p, and haviour (see fig. 1) near an absorption edge.
specifically on the largest of these, that for the nickel
K, line in iron, since this quantity strongly influences 3.2. Light major element amlysis at low proton energy
both the matrix attenuation for NiK, and the sec-
ondary fluorescence of iron. While the p/p theoretical If the matrix is comprised mainly of low-Z elements,
data base used here [lo] is felt to be accurate generally then the presence of any absorber between specimen
to 2-3%, its authors warn of anomalies at X-ray ener- and detector is precluded and the intense X-rays of the
Energy (keV)
Fig. I. Difference A between measured and XCOM data base values of X-ray attenuation coefficient in iron at energies near the K
absorption edge.
J.L. Campbell et al. / PIXE microanalysis of thick specimens 99
trum in fig. 2) and bustamite where the mineral albite ENERGY (KeV)
was employed as a standard for the elements sodium, Fig. 2. PIXE spectrum of sanidine recorded using 0.75 MeV
aluminum and silicon; single-element standards were protons.
used for higher-Z elements. The oxygen concentration
was found by assigning the normal valence to each
cation species and deducing the oxygen corresponding to the content of sodium, aluminum, etc. Not only is
there good agreement between recommended and
Imeasured concentrations for the various elements, in-
cluding oxygen, but the total element content in each
Table 2 case is close to lOO%, which provides a check on
u-PIXE analyses of the minerals bustamite and sanidine using consistency.
an albite a standard; R = recommended, M = measured The advent of ultrathin window (UTW) detectors
Element Concentration [%I
permits recording of the oxygen lines and averts the
need for an invisible element in the computation. The
Bustamite Sanidine
two prices paid for this complete coverage of the X-ray
R M R M spectrum would be (i) a smaller detector (since only
0 38.39 38.5 kO.4 46.28 45.4kO.3 small-diameter windows are feasible) and (ii) entry of
Na 0.04 0.11+0.1 2.23 2.2 * 0.15 scattered protons. The first is of no matter for major
Mg 0.13 0.14kO.03 - 0 elements since count rates remain adequate, but the
AI 0.02 0 9.93 9.710.12 decreased efficiency would be a detriment in trace
Si 22.46 22.3 +O.l 30.23 29.7 i 0.2 analysis of the same specimen. Musket [12] has demon-
K 0.05 0 10.05 10.3 + 0.2 strated removal of the scattered protons in conven-
Ca 13.56 13.8 +0.3 0.14 0
tional PIXE by interposing an arrangement of compact
Mn 18.83 19.6 +0.9 - 0
Fe 6.32 6.1 +0.9 - 0
permanent magnets (see fig. 3) between specimen and
Zn 0.20 0 _ 0
detector. It remains to be determined if this solution is
Ba 0 0.98 1.0+0.2 compatible with a microbeam. If it proved to be com-
patible, one could conduct both major and trace ele-
Total 100 100.6 99.85 98.3 ment analysis with 3 MeV protons, which would be
a AIbite concentrations are: 48.76% 0; 8.60% Na; 10.34% Al; much more convenient than the mode described here,
32.03% Si; 0.18% K 0.09% Ca. which involves switching between two energies.
Fig. 3. Compact magnetic deflection unit for removal of The GUPIX data base employs the most sophisti-
scattered beam particles from the axis of a UTW Si(Li) cated ECPSSR K ionization cross-sections available to
detector [El. date [17], together with Bambynek’s 1984 compilation
J.L. Campbell et al. / PIXE microanalysis of thick specimens 101
_”
4 6 6 10 12
ENERGY (keV)
Fig. 6. PIXE spectrum of a Cd,Hg,_,Te film deposited on an indium-tin-oxide coated glass substrate.
400 jc j” jca
BS
303
,” 250
:
I
c
3 200
0
” 150
‘..
:_ ., ... ,.., ..‘, ‘. .‘. ..
50
I ,_. ..’
0
1500.3 L750.0 2000.0 2250.0 2500.0 2750.0 3ooo.a
energy/ keV
energy/ keV r-
Fig. 7. Backscatter and PIXE spectra of a fly-ash particle [23].
J.L. Campbell et al. / PIXE microanalysis qf thick specimens 103
Table 3 Table 4
Variation in major element concentrations in pelletized X-ray intensity variation among ten replicate PIXE analyses
CCIJ-1 standard of synthetic pyrrhotites (Fess) containing trace elements at
0.1% concentration
Element Concentration [%I Standard deviation [%I
Internal External Beam spot diameter Standard deviation a [%I
[wnl Fe Cu Se Pd
Fe 30.80 0.36 2.4
cu 24.71 0.21 2.4 Standard A: Fe,S (Cu, Pd)
Zn 3.22 0.18 10.1 5x5 cTh 1.82 2.25 2.29
a, 0.29 1.84 2.10
50x50 uh 0.43 1.68 3.67
0.28 1.82 2.08
deviation from simple counting statistics. But the Standard B: Fe,S (Se, P:)
standard deviations for ten replicate measurements 5X5 "h 1.38 10.67 1.79
(see table 3) were considerably larger, and clearly sug- “e 0.27 1.27 2.48
gested a much greater heterogeneity for zinc than for 50X.50 uh 0.69 6.68 2.63
iron and copper. fie 0.27 1.22 2.44
Recognizing the need to homogenize geological a (TV= observed standard deviation, a, = expected standard
standard materials on the scale of the volume sampled deviation ( = square root of counts).
by the beam, Ryan et al. [26] fused US Geological
Survey rock standards to form glasses. In two cases
(BCR-1 and AGV-l), EPMA analysis demonstrated tion of heterogeneity. The GSP-1 material was
homogeneity in major and minor elements, but some markedly heterogeneous due to the distribution of
trace elements measured by p-PIXE at eight spots minute grains of zircon and quartz.
showed spreads in concentration more than twice the For these reasons we have preferred to conduct our
statistically expected range. The summed spectrum gave geochemical work with synthetic sulfide minerals pre-
concentrations that agreed to 1% on average with pared by CANMET [27]. But even these simple sys-
recommended values, lending credence to the observa- tems can be heterogeneous. Table 4 shows the varia-
(b)
105
ti
g lo4
iI
V
p: lo5
E
100 r IO0 t.
14 16 16 20 14 16 16 20
14 16 16 20 14 16 16 20
ENERGY ENERGY
Fig. 9. Details of fits to a zirconium PIXE spectrum: (a) no scattering step is included in the line shape, and yttrium is included in
the element list; (b) a scatter step is included.
J.L. Campbellet al. / PIXE microanalysisof thickspecimens 105
References [19] H. Paul and J. Sacher, At. Data and Nucl. Data Tables
42 (1989) 105.
[I] D.E. Newbury, J. Trace and Microprobe Tech. 32 (1986) [20] M.O. Krause, J. Phys. Chem. Ref. Data 8 (1979) 307.
103. [21] Y.C. Yu, M.R. McNeir, D.L. Weathers, J.L. Duggan,
[2] M. Hyvonen-Dabek, J. Raisanen and J.T. Dabek, J. Ra- F.D. McDaniel and G. Lapicki, Phys. Rev. A9 (1991)
dioanal. Chem. 63 (1981) 163. 5702.
[3] E. Clayton, Nucl. Instr. and Meth. 191 (19811567. [22] M.F.C. Willemsen and A.E.T. Kuiper, Nucl. Instr. and
[4] J.L. Campbell, W.J. Teesdale and J.-X. Wang, Nucl. Meth. B61 (1991) 213.
Instr. and Meth. B50 (1990) 189. [23] M. Jaksic, G. Grime, J. Henderson and F. Watt, Nucl.
[5] P.S.Z. Rogers, C.J. Duffy and T.M. Benjamin, Nucl. Instr. and Meth. B54 (199i) 491.
Instr. and Meth. B22 (1987) 133. [24] G.H. Faye, W.S. Bowman and R. Sutarno, CANMET
[6] J.A. Maxwell, J.L. Campbell and W.J. Teesdale, Nucl. Rep. 79-16, Copper concentrate CCU-1, a certified refer-
Instr. and Meth. B43 (1988) 218. ence material (Dept. of Energy, Mines and Resources,
[7] G. Lagarde, J. Cailleret and C. Heitz, Nucl. Instr. and Ottawa, Canada).
Meth. 205 (1983) 545. [25] G.I. Johansson, X-Ray Spectrom. 11 (1982) 194.
[8] J.L. Campbell, J.-X. Wang, J.A. Maxwell and W.J. Tees- [26] C.G. Ryan, D.R. Cousens, S.H. Sie, W.L. Griffin, G.F.
dale, Nucl. Instr. and Meth. B43 (1989) 539. Suter and E. Clayton, Nucl. Instr. and Meth. B47 (1990)
[9] G.K. Czamanske, V.E. Kunilov, L.J. Cabri, M.L. Zientek, 55.
L.C. Calk and A.P.Likhachev, Can. Mineralogist 30 (1992) [27] L.J. Cabri, J.L. Campbell, J.H.G. Laflamme, R.G. Leigh,
249. J.A. Maxwell and J.D. Scott, Can. Mineralogist 23 (1985)
[lo] M.J. Berger and J.H. Hubbell, National Bureau of Stan- 133.
dards Rep. 87-3597 (NIST, Gaithersburg, 1987). [28] C.P. Swann and S.J. Fleming, Nucl. Instr. and Meth. B49
[ll] N.K. Del Grande and A.J. Oliver, Bull. Am. Phys. Sot. (1990) 65.
16 (1971) 545. [29] F. Hei and P. Van Espen, Anal. Chem. 63 (1991) 2237.
[12] R.G. Musket, Nucl. Instr. and Meth. B15 (1986) 735. [30] J. Rickards and E.P. Zironi, Nucl. Instr. and Meth. B29
[13] R.G. Musket, National Bureau of Standards Spec. Pub. (1987) 527.
604 (NIST, Gaithersburg, 1981). 1311W. Reuter, A. Lurio, F. Cardone and J. Ziegler, J. Appl.
[14] J.L. Campbell and J.-X. Wang, X-Ray Spectrom. 21 Phys. 46 (1975) 3194.
(1992) 223. [32] G. Demortier, S. Mathot and B. Van Gystaeyen, Nucl.
[U] M. Krumrey, E. Tegeler and G. Ulm, Rev. Sci. Instr. 60 Instr. and Meth. B49 (1990) 46.
(1989) 2287. [33] C.G. Ryan, D.R. Cousens, C.A. Heinrich, W.L. Griffin,
[16] J.L. Campbell and J.-X. Wang, X-Ray Spectrom. 20 S.H. Sie and T.P. Mernagh, Nucl. Instr. and Meth. B54
(1991) 191. (1991) 292.
[17] M.H. Chen and B. Crasemann, At. Data Nucl. Data [34] J.L. Campbell, J.A. Maxwell and W.J. Teesdale, to be
Tables 33 (1985) 217; published.
M.H. Chen, private communication.
[18] J.H. Hubbell, National Institute of Standards and Tech-
nology Rep. 89-4144 (NIST, Gaithersburg, 1989).