ImageJ Operating Instructions

ddRCompact HD-3000

ImageJ Operating Instructions


HD-3000

ddRCompact

Revision Date Initials Reason for changes

00a 2008-04-10 UHR First draft

00b 2008-05-19 UHR Editorial amendments (RIN)

Release date: 5/19/2008 UHR Pages 1 / 4 Document ID: SHT_37_160_010_00b_D

Copyright 2005 Swissray Medical AG All rights reserved


ImageJ Operating Instructions
ddRCompact HD-3000

1 Version
ImageJ is a public domain Java image-processing program inspired by NIH Image of
the National Institute of Mental Health (NIMH). Version 1.37v of ImageJ was used for
this documentation.

2 Defect Map
2.1 General information
One defect map image each in which all defective pixels are marked is required for
Detector MODE #1 (Default) and MODE #2 (Binning).

If no Defect Map images are available, or automatic image generation does not
function, a Defect Map image of this sort can be created with ImageJ using the
following description.

All defective pixels have the value 65535. All pixels that are OK have the value 0.

2.2 Creating the image


• In the <File> menu, click <New> and then <Image...>.
¾ The <New Image...> window is displayed.

• In the <New Image...> window, enter the following values for the MODE desired:

MODE #1 (Default) MODE #2 (Binning)

Name: DefectMap4096x4126 Name: DefectMap2048x2063

• Click <OK>.
¾ The image is displayed in a new window.
All pixels have the value 0.

Release date: 5/19/2008 UHR Pages 2 / 4 Document ID: SHT_37_160_010_00b_D

Copyright 2005 Swissray Medical AG All rights reserved


ImageJ Operating Instructions
ddRCompact HD-3000

2.3 Drawing defects


Assume the image in MODE # 1 has a column error that begins at point x1=100 and
y1=100, and ends at point x2=100 and y2=4117. To draw this line, the corresponding
pixels first have to be marked.

• In the <Edit> menu, click <Selection> and then <Specify...>.


¾ The <Specify> window is displayed.

• In the <Specify> window, enter the following values to mark the column defect
previously described:

<Width> is the width of the line and <Height> is the length


of the line. The length of the line can be calculated with y2 –
y1 + 1.
<X Coordinate> and <Y Coordinate> are position values for
the starting point (x1/x2).

• Click <OK>.
¾ The line is marked in yellow.

Release date: 5/19/2008 UHR Pages 3 / 4 Document ID: SHT_37_160_010_00b_D

Copyright 2005 Swissray Medical AG All rights reserved


ImageJ Operating Instructions
ddRCompact HD-3000
2.4 Filling the marked area
• In the <Process> menu, click <Math> and then <Add...>.
¾ A new window called <Add> opens.

• Enter the following values in the <Add> window:

• Click <OK>.
¾ The line is marked with the value 65535.

• For all other defects, repeat the steps described above using the respective
coordinates, until all defects are marked.

2.5 Saving the image


The image can be saved once all defects are marked as described above.

• In the <File> menu, click <Save As> and then <Raw Data...>.

• Save the image in the folder desired with <Save>.