When operating Olympus Innov-X FPXRF analyzers, great care should be taken to protect the
instrument detector. The detector is the most fragile and also very expensive piece of hardware
inside the instrument and to avoid expensive repairs, users should follow the guidelines in this
document.
1. Place the instrument lying on its side on your desk. Carefully place the screwdriver on the
screw and support / cover the window with your other hand (Figure 2). Open both screws on
the window hinge.
Innov-X Africa, PostNet Suite 126, Private Bag X1, Edenglen 1613, South Africa
Tel: +27 (0) 87 701 3165 | Fax: +27 (0) 11 450 0321 | Cell No: +27 (0) 82 409 1646
Email: info@innovxafrica.co.za Web: www.innovxafrica.com
Page 1 of 2
DETECTOR CARE AND WINDOW REPLACEMENT
2. While replacing the window, be careful not to drop anything inside the probe head. Place the
new window in place as below, close the hinge and tighten the screws. Again, support / cover
the window with your other hand.