Specification Sheet
S/5 Avance®
Innovating with you, shaping exceptional care
Features
• Complete patient monitoring capabilities: respiratory gas,
hemodynamic and adequacy of anesthesia
• Our state of the art electronic gas mixer with pneumatic
back-up control
• Full color displays
Exceptional design
• Generous storage and work surface space
• Bi-level work surface illumination
• Integrated cable and tube management
Physical specifications
Dimensions
Height: 134.5 cm/52.9 in
Width: 72 cm/28.3 in
Depth: 73 cm/28.7 in
Weight: Approximately 125 kg/275 lb
Top shelf
Weight limit: 34 kg/75 lb
Width: 66 cm/26 in
Depth: 40 cm/15.75 in
Work surface
Height: 81.7 cm/32.2 in
Size: 2160 cm2/334 in2
DIN rail
Side of
machine: 34.5 cm/13.6 in
Drawers (internal dimensions)
Height: 17.5 cm/6.9 in
Width: 33 cm/13 in
Depth: 26.5 cm/10.4 in
Absorber bag arm (optional)
Arm length: 30.5 cm/12 in
Bag arm height
(adjustable): 87 cm/34.3 in
104 cm/40.9 in
Casters
Diameter: 12.5 cm/5 in
Brakes: Individual locking front casters
Ventilator operating specifications
Pressure
(Plimit) range: 12 to 100 cm H2O
XXxxx Xxxx
8:48
(increments of 1 cm H2O)
1:58
Paw 40 cmH20
15
l/min
15
l/min
20 P
a 28
Ppeak
40
Pmean
8
Pressure
Vent On: Volume Control
w PEEP
Flow 120
0
C
O EtCO2 mmHg
5.0
FiCO2 mmHg
Off
Rate: 4 to 100 breaths per
2
/ 26 Off 2
N2O
1.0
O2
1.0
CO2 5
0
O
2
45
EtO2 %
Off
Off 49
FiO2 %
Off
21
minute for Volume Control
Gases
Minute volume
range: 0 to 99.9 L/min
Pressure
(Pinspired)
range: 5 to 60 cm H2O
(increments of 1 cm H2O)
5 to 1500 mL volume delivery
Ventilator operating specifications,
continued
Alarm settings
Ventilator performance
Tidal volume
Pressure range (VTE): Low: OFF, 0 to 1500 mL
at inlet: 240 kPa to 700 kPa/ High: 20 to 1600 mL, OFF
35 psig to 100 psig
Minute volume
Peak gas (VE): Low: OFF, 0 to 10 L/min
flow: 120 L/min + fresh gas flow High: 0 to 30 L/min, OFF
Flow valve Inspired oxygen
range: 1 to 120 L/min (FiO2): Low: 18 to 100%
High: 19 to 100%, OFF
Flow
compensation Apnea alarm: Mechanical ventilation ON:
range: 200 mL/min to 15 L/min < 5 mL breath measured
in 30 seconds
Ventilator accuracy Mechanical ventilation OFF:
< 5 mL breath measured
Delivery/monitoring accuracy
in 30 seconds
Volume
Low airway
delivery: > 210 mL=better than 7%
pressure: 4 cm H2O above PEEP
< 210 mL=better than 15 mL
< 60 mL=better than 10 mL High pressure: 12 to 100 cm H2O
(increments of 1 cm H2O)
Pressure
delivery: ±10% or ±3 cm H2O Sustained airway
pressure: Mechanical ventilation ON:
PEEP delivery: ±1.5 cm H2O
Plimit < 30 cm H2O, the
Volume sustained limit is 6 cm H2O
monitoring: > 210 mL=better than 9% Plimit 30 to 60 cm H2O,
< 210 mL=better than 18 mL the sustained limit is 20%
< 60 mL=better than 10 mL of Plimit
Pressure Plimit > 60 cm H2O, the
monitoring: ±5% or ±2 cm H2O sustained limit is 12 cm H2O
PEEP and mechanical
ventilation ON:
Sustained limit increases by
PEEP minus 2 cm H2O
Mechanical ventilation OFF:
Plimit ≤ 60 cm H2O,
the sustained limit is 50%
of Plimit
Plimit > 60 cm H2O, the
sustained limit is 30 cm H2O
Subatmospheric
pressure: Paw < -10 cm H2O
Alarm silence
countdown
timer: 120 to 0 seconds
Ventilator components Anesthetic agent delivery
Flow transducer Delivery
Type: Variable orifice flow sensor Vaporizers: Tec 5, Tec 6 Plus, Tec 7
Dimensions: 22 mm OD and 15 mm ID Number of
positions: 2
Location: Inspiratory outlet and
expiratory inlet Mounting: Tool-free installation
Selectatec® manifold inter-
(Optional autoclavable sensor available)
locks and isolates vaporizers
Oxygen sensor
Type: Optional galvanic fuel cell
or paramagnetic with MGAI
option up to
g es.
rnin all tim cence ents.
ep t au use in anae in co
• KeDo not for able used itor.
nl
of
Waright atclave. e presetic agunction
th sth
ns. ance
.
Ventilator screen
t serv
Nex
18
16
14
12
10
8 7
6 5 4 3 2 1
Battery backup c 6P
lus
Te
AGENT
LOW
UP
WARM
Tec 7
S/5 Compact Airway modules
General Respiration rate (RR)
M-CAiO, M-CAiOV, M-CAiOVX module Measurement
software version 3.2 or higher range: 4 to 60 breaths per minute
Size (WxDxH): 75 x 215 x 112 mm/ Detection
2.9 x 8.4 x 4.4 in criteria: 1% variation in CO2
Weight: 1.6 kg/3.7 lb Adjustable low and high alarm limits for
respiration rate; alarm for apnea
Sampling rate: 200 mL/min ±20 mL
Patient Oxygen (O2)
Automatic compensation for atmospheric
pressure variation (500 to 800 mmHg) FiO2: Inspired O2 concentration
temperature and CO2/N2O and CO2/O2 col-
EtO2: End-tidal O2 concentration
lision broadening effect. Parameter display
update interval typically breath-by-breath. FiO2-EtO2: Inspired-expired difference
Functional alarms for blocked sample line, O2 waveform
D-fend check and D-fend replacement.
Measurement
Non-disturbing gases: range: 0 to 100 %
Ethanol, acetone, methane, nitrogen, nitric Accuracy: ±2 vol %*
oxide, carbon monoxide, water vapor:
Datex-Ohmeda differential paramagnetic
Maximum effect sensor
on readings: CO2 < 0.2 vol %;
O2 < 2 vol % Adjustable low and high alarm limits for
FiO2 and EtO2; alarm for FiO2 < 18%
Carbon dioxide (CO2)
Nitrous Oxide (N2O)
EtCO2: End-tidal CO2 concentration
Measurement
FiCO2: Inspired CO2 concentration range: 0 to 100%
CO2 waveform Accuracy: ±3 vol %*
Measurement FiN2O > 82% alarm
range: 0 to 15% (0 to 15 kPa,
* Typical value
0 to 113 mmHg) ** Measurement not valid with O2 and N2O mixtures
O2 controls O2 concentration
range: 25% to 100%
Method: N2O shut off with loss of
O2 pressure O2 concentration
accuracy: ±5% V/V for flows < 1 L/min
Supply failure
±2.5% setting for flows
alarm: Range: 193 kPa to 221 kPa/
> 1 L/min
28 psig to 32 psig
Sounds at maximum volume Electronic mixer
every 10 seconds response time: 500mS
(10% to 90% flow step)
O2 flush: Range: > 35 L/min
Compensation: Temperature and
Alternate O2 (safety flow)
atmospheric pressure
Range: 500 mL/min minimum to compensated to standard
10 L/min conditions of 20°C
and 101.3 kPa
Indicator: Flow tube
Hypoxic guard: Electronic
Indicator
accuracy: ±5% full scale Materials
All materials in contact with patient breath-
ing gases are free of natural rubber latex.
Environmental specifications Breathing circuit specifications
System operation Operational modes
Temperature: 10° to 40°C/50° to 104°F Breathing circuit is circle mode; SCGO
option converts to open circuit mode
Humidity: 15 to 95% relative humidity
(non-condensing) Carbon dioxide absorbent canister
per IEC 68-2-3
Absorbent
Altitude: -440 to 3565 m/ capacity: 800 g
500 to 800 mmHg
Integrated expiratory limb water reservoír
System storage
Ports and connectors
Temperature: -25° to 60°C/-13° to 140°F
Exhalation: 22 mm OD ISO
Humidity: 10 to 95% relative humidity 15 mm ID taper
(non-condensing) per
Inhalation: 22 mm OD ISO
IEC 68-2-3
15 mm ID taper
Altitude: -440 to 5860 m/
Bag port: 22 mm OD
375 to 800 mmHg
Bag-to-Ventilator switch
Oxygen cell
storage: -15° to 50°C/5° to 122°F Type: Bi-stable
10 to 95% relative humidity Control: Controls ventilator and
500 to 800 mmHg direction of breathing gas
Electromagnetic compatibility within the circuit
GE Healthcare
Madison, Wisconsin U.S.A.
www.gehealthcare.com
General Electric Company reserves the right to make changes in specifications and features
shown herein, or discontinue the product described at any time without notice or obligation.
Contact your GE Representative for the most current information.
©2004 General Electric Company
Datex-Ohmeda, Inc., a General Electric company going to market as GE Healthcare.