Vous êtes sur la page 1sur 16

BRITISH STANDARD BS EN

60601-2-40:
1998
IEC
60601-2-40:
1998

Medical electrical BS 5724-2.40:


Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

1998

equipment —
Part 2-40: Particular requirements for
safety — Specification for
electromyographs and evoked response
equipment

The European Standard EN 60601-2-40:1998 has the status of a


British Standard

ICS 11.040.50
BS EN 60601-2-40:1998

National foreword

This British Standard is the English language version of EN 60601-2-40:1998.


It is identical with IEC 60601-2-40:1998.
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

The UK participation in its preparation was entrusted to Technical Committee


CH/99, Neurological electrodiagnostic, which has the responsibility to:
— aid enquirers to understand the text;
— present to the responsible international/European committee any
enquiries on the interpretation, or proposals for change, and keep the UK
interests informed;
— monitor related international and European developments and
promulgate them in the UK.
A list of organizations represented on this committee can be obtained on
request to its secretary.
From 1 January 1997, all IEC publications have the number 60000 added to
the old number. For instance, IEC 27-1 has been renumbered as IEC 60027-1.
For a period of time during the change over from one numbering system to the
other, publications may contain identifiers from both systems.
Cross-references
Attention is drawn to the fact that CEN and CENELEC Standards normally
include an annex which lists normative references to international
publications with their corresponding European publications. The British
Standards which implement these international or European publications may
be found in the BSI Standards Catalogue under the section entitled
“International Standards Correspondence Index”, or by using the “Find”
facility of the BSI Standards Electronic Catalogue.
A British Standard does not purport to include all the necessary provisions of
a contract. Users of British Standards are responsible for their correct
application.

Compliance with a British Standard does not of itself confer immunity


from legal obligations.

Summary of pages
This document comprises a front cover, an inside front cover, pages i and ii,
the EN title page, pages 2 to 11 and a back cover.
This standard has been updated (see copyright date) and may have had
amendments incorporated. This will be indicated in the amendment table on
the inside front cover.

This British Standard, having


been prepared under the Amendments issued since publication
direction of the Health and
Environment Sector Board, Amd. No. Date Comments
was published under the
authority of the Standards
Board and comes
into effect on
15 June 1998

© BSI 05-1999

ISBN 0 580 29950 3


Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

© BSI 05-1999
Foreword
National foreword
Contents

Text of EN 60601-2-40
Inside front cover
Page

3
2
BS EN 60601-2-40:1998

i
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

ii
blank
EUROPEAN STANDARD EN 60601-2-40
NORME EUROPÉENNE
April 1998
EUROPÄISCHE NORM
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

ICS 11.040.50

Descriptors: Medical electrical equipment, electromyographs, evoked response equipment, safety requirements, protection against
electric shock, protection against mechanical hazard, radiation protection, fire protection, environmental conditions

English version

Medical electrical equipment Part 2-40: Particular


requirements for the safety of electromyographs and evoked
response equipment
(IEC 60601-2-40:1998)

Appareils électromédicaux Medizinische elektrische Geräte


Partie 2-40: Régles particulières de Teil 2-40: Besondere Festlegungen für die
sécurité pour les électromyographes Sicherheit von Elektromyographen und
et les appareils à potentiel évoqué Geräten für evozierte Potentiale
(CEI 60601-2-40:1998) (IEC 60601-2-40:1998)

This European Standard was approved by CENELEC on 1998-04-01.


CENELEC members are bound to comply with the CEN/CENELEC Internal
Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the Central Secretariat or to any
CENELEC member.
This European Standard exists in three official versions (English, French,
German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the
Central Secretariat has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria,
Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece,
Iceland, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Spain,
Sweden, Switzerland and United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
Central Secretariat: rue de Stassart 35, B-1050 Brussels

© 1998 CENELEC — All rights of exploitation in any form and by any means reserved worldwide for
CENELEC members.
Ref. No. EN 60601-2-40:1998 E
EN 60601-2-40:1998

Foreword Contents
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

The text of document 62D/255/FDIS, future Page


edition 1 of IEC 60601-2-40, prepared by SC 62D, Foreword 2
Electromedical equipment, of IEC TC 62, Electrical
Introduction 3
equipment in medical practice, was submitted to the
IEC-CENELEC parallel vote and was approved by Section 1. General
CENELEC as EN 60601-2-40 on 1998-04-01. 1 Scope and object 3
The following dates were fixed: 2 Terminology and definitions 4
— latest date by which the 5 Classification 4
EN has to be implemented 6 Identification, marking and documents 4
at national level by
7 Power input 5
publication of an identical
national standard or by Section 2. Environmental conditions
endorsement (dop) 1999-01-01 Section 3. Protection against electric
shock hazards
— latest date by which the
national standards 14 Requirements related to classification 5
conflicting with the EN 20 Dielectric strength 5
have to be withdrawn (dow) 2001-01-01 Section 4. Protection against mechanical hazards
Annexes designated “normative” are part of the Section 5. Protection against hazards from
body of the standard. unwanted or excessive radiation
Annexes designated “informative” are given for 36 Electromagnetic compatibility 6
information only. Section 6. Protection against hazards of
In this standard, Annex ZA is normative and ignition of flammable anaesthetic mixtures
Annex AA and Annex ZB are informative. Section 7. Protection against excessive
Annex ZA and Annex ZB have been added by temperatures and other safety hazards
CENELEC. 42 Excessive temperatures 6
Endorsement notice 46 Human errors 6
Section 8. Accuracy of operating data and
The text of the International Standard
protection against hazardous output
IEC 60601-2-40:1998 was approved by CENELEC
as a European Standard without any modification. 50 Accuracy of operating data 7
51 Protection against hazardous output 7
Section 9. Abnormal operation and fault
conditions; environmental tests
Section 10. Constructional requirements
Annex L References — Publications
mentioned in this Standard 9
Annex AA (informative) General
guidance and rationale 9
Annex ZA (normative) Normative references
to international publications with their
corresponding European publications 11
Annex ZB (informative) Other international
publications mentioned in this
standard with the references of
the relevant European publications 11
Figure 101 — Layout of testing
arrangements (see 36.201.1.7
and 36.202.2.2 d)] 8

2 © BSI 05-1999
EN 60601-2-40:1998

Introduction 1.3 Particular Standards


Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

This Particular Standard amends and supplements Addition:


IEC 60601-1 (second edition, 1988): Medical This Particular Standard refers to IEC 60601-1
electrical equipment — Part 1: General requirements (1988): Medical electrical equipment —
for safety, as amended by its amendment 1 (1991) Part 1: General requirements for safety as
and amendment 2 (1995), hereinafter referred to as amended by its amendment 1 (1991) and
the General Standard (see 1.3). amendment 2 (1995).
The requirements are followed by specifications for For brevity, IEC 60601-1 is referred to in this
the relevant tests. Particular Standard either as the “General
Following the decision taken by subcommittee 62D Standard” or as the “General Requirement(s)”.
at the meeting in Washington in 1979, a “General The numbering of sections, clauses and subclauses
guidance and rationale” section giving some of this Particular Standard corresponds to that of
explanatory notes, where appropriate, about the the General Standard. The changes to the text of the
more important requirements is included in General Standard are specified by the use of the
Annex AA. following words:
Clauses or subclauses for which there are “Replacement” means that the clause or subclause
explanatory notes in Annex AA are marked with an of the General Standard is replaced completely by
asterisk (*). the text of this Particular Standard.
It is considered that a knowledge of the reasons for “Addition” means that the text of the Particular
these requirements will not only facilitate the Standard is additional to the requirements of the
proper application of the Standard but will, in due General Standard.
course, expedite any revision necessitated by
“Amendment” means that the clause or subclause of
changes in clinical practice or as a result of
the General Standard is amended as indicated by
developments in technology. However, this annex
the text of this Particular Standard.
does not form part of the requirements of this
Standard. Subclauses or figures which are additional to those
of the General Standard are numbered starting
Section 1. General from 101, additional annexes are lettered AA,
BB, etc., and additional items aa), bb), etc.
The clauses and subclauses of this section of the The term “this Standard” is used to make references
General Standard apply except as follows: to the General Standard and this Particular
Standard taken together.
1 Scope and object Where there is no corresponding section, clause or
This clause of the General Standard applies except subclause in this Particular Standard, the section,
as follows: clause or subclause of the General Standard,
although possibly not relevant, applies without
1.1 Scope modification; where it is intended that any part of
Addition: the General Standard, although possibly relevant,
is not to be applied, a statement to that effect is
This Particular Standard applies to given in this Particular Standard.
ELECTROMYOGRAPHS as defined in 2.1.101 and
The requirements of this Particular Standard take
EVOKED RESPONSE EQUIPMENT as defined in 2.1.102,
priority over those of the General Standard.
hereinafter referred to as EQUIPMENT.
1.5 Collateral Standards
1.2 Object
Addition:
Replacement:
Collateral Standards IEC 60601-1-1 as amended by
The object of this Particular Standard is to establish
particular requirements for the safety of its amendment 1 and IEC 60601-1-2 have been
ELECTROMYOGRAPHS and EVOKED RESPONSE
taken into consideration in this Particular Standard
(see 6.8.2 and clause 36).
EQUIPMENT as defined in 2.1.101 and 2.1.102,
respectively.

© BSI 05-1999 3
EN 60601-2-40:1998

2 Terminology and definitions 5 Classification


Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

This clause of the General Standard applies except This clause of the General Standard applies except
as follows: as follows:
Additional definitions: 5.2 *According to the degree of protection against
2.1.101 electric shock:
ELECTROMYOGRAPH Amendment:
MEDICAL ELECTRICAL EQUIPMENT for the detection Delete TYPE B APPLIED PART.
and analysis of biopotentials accompanying nerve 5.6 *According to the mode of operation:
and muscle action, either spontaneously, Amendment:
intentionally or evoked by electrical or other
stimulation Delete all except CONTINUOUS OPERATION.

2.1.102 6 Identification, marking and


EVOKED RESPONSE EQUIPMENT
documents
MEDICAL ELECTRICAL EQUIPMENT for the detection
This clause of the General Standard applies except
and analysis of biopotentials resulting from an
as follows:
evoking stimulus. The stimulus may be electrical,
tactile, auditory, visual, olfactory, etc. 6.1 Marking on the outside of EQUIPMENT or
EQUIPMENT parts
2.1.103
ELECTRICAL STIMULATOR j) Power input
parts of EQUIPMENT for the application of electric Amendment:
currents via electrodes in direct contact with the Replace the fourth paragraph by the following:
PATIENT; for the evoking of biopotentials or other The RATED power input of mains operated
action EQUIPMENT shall be the maximum power input
2.1.104 averaged over any period of 5 s under the conditions
PULSE DURATION set out in item aa) of 7.1.
duration of the electrical stimulus pulse WAVEFORM p) *Output
at 50 % of the peak amplitude Addition:
2.1.105 EQUIPMENT capable of delivering electrical stimulus
WAVEFORM outputs into a load resistance of 1 000 7 in excess
variations in magnitude of an electrical stimulus of 10 mA r.m.s. or 10 V r.m.s. averaged over any
output (in either voltage or current) as a function of period of 5 s shall be marked near the electrode
time appearing in the APPLIED PART of the connections with the symbol No. 14 (see Annex D of
ELECTRICAL STIMULATOR
the General Standard).
2.1.106 6.7 Indicator lights and push-buttons
AUDITORY STIMULATOR Addition:
parts of EQUIPMENT for the application of sound See also 51.102.
pressure from a transducer (headphone, bone 6.8 ACCOMPANYING DOCUMENTS
conductor or free-field) to the ear(s) of the PATIENT,
for the evoking of biopotentials or other action 6.8.2 Instructions for use

2.1.107 Additional item:


VISUAL STIMULATOR The instructions for use shall additionally contain
the following:
parts of EQUIPMENT or SYSTEM for the application of
electromagnetic radiation in the visible spectrum aa) *Information on the output WAVEFORM(S),
from a transducer to the eyes of the patient, for the including any d.c. component, PULSE DURATIONS,
evoking of biopotentials or other action pulse repetition frequencies, maximum
amplitude of output voltage and/or current, and
2.1.108
the effect of load impedance on the demanded
BIOPOTENTIAL INPUT PART
parameters.
APPLIED PART(S) of EQUIPMENT or SYSTEM for the bb) *Advice on the size of electrodes to be used
collection of biopotentials and the method of application for each particular
type of examination for which the ELECTRICAL
STIMULATOR is intended.

4 © BSI 05-1999
EN 60601-2-40:1998

cc) Advice on any necessary precautions to be 7 Power input


taken when the output contains a d.c. component.
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

This clause of the General Standard applies except


dd) *Advice that a PATIENT with an implanted as follows:
electronic device (for example a cardiac
7.1
pacemaker) should not be subjected to electrical
stimulation unless specialist medical opinion has Additional item:
first been obtained. aa) The power input is measured with a load
resistance of the lowest value specified in the
ee) Advice to avoid trans-thoracic stimulation, for
technical description (see 6.8.3), and with any
example maintenance of anode and cathode
output controls set to give maximum power input.
stimulating sites in close proximity.
ff) A warning on the following potential SAFETY Section 2. Environmental conditions
HAZARDS:
• connection of a PATIENT to a h.f. surgical The clauses and subclauses of this section of the
equipment and to an ELECTROMYOGRAPH or General Standard apply.
EVOKED RESPONSE EQUIPMENT simultaneously
may result in burns at the site of the Section 3. Protection against electric
ELECTRICAL STIMULATOR or BIOPOTENTIAL shock hazards
INPUT PART electrodes and possible damage to
the ELECTRICAL STIMULATOR or biological The clauses and subclauses of the General Standard
amplifiers; apply except as follows:
• operation in close proximity (for 14 Requirements related to
example 1 m) to a shortwave or microwave
classification
therapy equipment may produce instability in
the ELECTRICAL STIMULATOR output. This clause of the General Standard applies except
gg) *For EQUIPMENT capable of delivering output as follows:
values in excess of 10 mA r.m.s. or 10 V r.m.s. 14.6 TYPES B, BF and CF APPLIED PARTS*
into the specified load impedance (see 6.8.3), Replacement:
averaged over 1 s, or having an energy greater
APPLIED PARTS of ELECTRICAL STIMULATORS, VISUAL
than 10 mJ per pulse into the specified load
STIMULATORS and AUDITORY STIMULATORS shall be
impedance:
TYPE BF APPLIED PARTS or TYPE CF APPLIED PARTS.
• information on the maximum output values
allowed for the electrodes recommended by 20 Dielectric strength
the manufacturer for use with the ELECTRICAL
STIMULATOR.
This clause of the General Standard applies except
as follows:
• advice that current densities for any
electrodes exceeding 2 mA r.m.s./cm2 may 20.2 Requirements for EQUIPMENT with an
APPLIED PART
require the special attention of the OPERATOR.
hh) When a mains powered video monitor, not Amendment:
complying with IEC 60601-1, is to be used as part B-b: Between BIOPOTENTIAL INPUT PARTS and
of the VISUAL STIMULATOR, the instructions other energized APPLIED PARTS, for
for use shall explain how to comply with example stimulators, the insulation shall
IEC 60601-1-1. be DOUBLE INSULATION OR REINFORCED
ii) *Advice to avoid accidental contact between INSULATION.
connected but unapplied APPLIED PARTS and Insulation is not required between
other conductive parts including those connected different BIOPOTENTIAL INPUT PARTS if
to protective earth. intended for use on a single PATIENT.
6.8.3 Technical description Between stimulator APPLIED PARTS of
Additional item: differing modalities, for example
aa) The technical description shall specify the ELECTRICAL STIMULATOR AND AUDITORY
parameters mentioned in 6.8.2 aa). The range of STIMULATOR, the insulation shall be
load impedance for which these parameters are DOUBLE INSULATION or REINFORCED
valid shall be specified. INSULATION.

© BSI 05-1999 5
EN 60601-2-40:1998

Section 4. Protection against 36.202.2 Radiated radiofrequency


mechanical hazards electromagnetic fields
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

36.202.2.2 Test conditions*


The clauses and subclauses of this section of the
Amendment:
General Standard apply.
Replace item d) by the following:
Section 5. Protection against hazards d) For the radiated radiofrequency electromagnetic
from unwanted or excessive radiation field test, all relevant electrodes shall be connected
and applied to the contents of a 1 000 ml capacity
The clauses and subclauses of this section of the normal saline filled phantom, positioned
General Standard apply except as follows: within 400 mm of the EQUIPMENT
(see Figure 101).
36 Electromagnetic compatibility
Collateral Standard for EMC IEC 60601-1-2 applies Section 6. Protection against hazards
except as follows: of ignition of flammable anaesthetic
36.201 EMISSIONS
mixtures
36.201.1.7 The clauses and subclauses of this section of the
Replacement: General Standard apply.
For the radiated radiofrequency emissions test, all
relevant electrodes shall be connected and applied to
Section 7. Protection against excessive
the contents of a 1 000 ml capacity, normal saline temperatures and other safety
filled phantom, positioned within 400 mm of the hazards
EQUIPMENT (see Figure 101).
The clauses and subclauses of this section of the
36.202 IMMUNITY General Standard apply except as follows:
Amendment:
Replace the fourth paragraph by the following: 42 Excessive temperatures
Compliance with the requirements given in 36.202.1 This clause of the General Standard applies except
to 36.202.6 shall be checked by verifying that, under as follows:
specified conditions, the EQUIPMENT and/or SYSTEMS Addition:
do not deliver stimuli exceeding more than ± 10 % of
Compliance with the requirements for maximum
those set by the OPERATOR for intensity, amplitude, temperatures specified in the General Standard
PULSE DURATION or repetition rate.
shall be checked under the conditions specified
Disturbance of the display during the test in 7.1 aa).
of 36.202.2.1 does not constitute a non-compliance
with the requirements of this Standard. 46 Human errors*
After completion of the immunity tests, the Replacement:
EQUIPMENT and/or SYSTEMS shall continue to
comply with the PATIENT, PATIENT AUXILIARY and The ELECTRICAL STIMULATOR shall not become
EARTH LEAKAGE CURRENT requirements.
unsafe if the output is switched on inadvertently
with open- or short-circuit electrodes, even if such
36.202.1 ELECTROSTATIC DISCHARGE an operation is considered to be misuse.
Addition: Compliance is checked by operating the ELECTRICAL
Electrostatic discharge tests shall include all STIMULATOR at maximum output settings for 5 min
connectors and terminals intended to form part of with the electrodes open-circuited and for 5 min with
the PATIENT circuit during NORMAL USE as well as all the electrodes short-circuited. After these tests all
other accessible surfaces, control knobs etc. safety requirements of this Standard shall be
Connectors and terminals other than PATIENT satisfied.
electrode connectors shall not be tested.

6 © BSI 05-1999
EN 60601-2-40:1998

Section 8. Accuracy of operating data 51.102 ELECTRICAL STIMULATOR output


and protection against hazardous indicator*
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

output EQUIPMENT which can deliver into a load resistance


of 1 000 7 an output in excess of 10 mA r.m.s.
The clauses and subclauses of this section of the or 10 V r.m.s. or pulses having an energy
General Standard apply except as follows: exceeding 10 mJ per pulse, shall provide visual
indication that the ELECTRICAL STIMULATOR is
50 Accuracy of operating data delivering stimuli, or is primed to deliver stimuli.
This clause of the General Standard applies except The colour of the visual indicator shall be yellow.
as follows: Compliance is checked by inspection and by a
functional test.
50.1 Marking of controls and instruments*
51.103 Limitation of ELECTRICAL STIMULATOR
Replacement:
output parameters*
An output control for the ELECTRICAL STIMULATOR
shall be incorporated which will control the The pulse energy with a load resistance of 1 000 7
ELECTRICAL STIMULATOR output from minimum to
shall not exceed 50 mJ per pulse.
maximum of the range continuously, or in discrete Compliance is checked by measurement.
increments of not more than 1 mA amplitude or 5 V 51.104 Limitation of VISUAL STIMULATOR output
amplitude per increment. At its minimum setting, parameters
the output shall not exceed 2 % of that available at
the maximum setting of the control. When the transducer of the VISUAL STIMULATOR
consists of light-emitting diodes (LEDs), the
Compliance is checked by inspection and radiation density shall not exceed the limits laid
measurement, using the load impedance which is the down in IEC 60825-1.
least favourable within the range specified in the
ACCOMPANYING DOCUMENTS. 51.105 Limitation of AUDITORY STIMULATOR
output parameters*
50.2 Accuracy of controls and instruments*
If a continuous output is available, for example for
Replacement: masking, the overall sound level should not
The values of PULSE DURATIONS, pulse repetition exceed 125 dB HTL.
frequencies and amplitudes, as described in the Compliance is checked by measurement.
ACCOMPANYING DOCUMENTS or indicated on the
EQUIPMENT (see 6.8.2), shall not deviate by more Section 9. Abnormal operation and
than 30 %, when measured with an error not
fault conditions; environmental tests
exceeding ± 10 % into a load resistance within the
range specified in the ACCOMPANYING DOCUMENTS The clauses and subclauses of this section of the
(see 6.8.3). General Standard apply.
Compliance is checked by measurement.
Section 10. Constructional
51 Protection against hazardous requirements
output
The clauses and subclauses of this section of the
This clause of the General Standard applies except
General Standard apply.
as follows:
Additional subclauses:
51.101 Supply voltage fluctuations*
Supply voltage fluctuations of ± 10 % of the
NOMINAL voltage shall not affect the ELECTRICAL
STIMULATOR output amplitude, PULSE DURATION or
pulse repetition frequency by more than ± 10 %.
Compliance is checked by measurement.

© BSI 05-1999 7
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

8
EN 60601-2-40:1998

Figure 101 — Layout of testing arrangements [see 36.201.1.7 and 36.202.2.2 d)]

© BSI 05-1999
EN 60601-2-40:1998

The annexes of the General Standard apply except as follows:


Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

Annex L References — Publications mentioned in this Standard


This annex of the General Standard applies except as follows:
IEC Standards
Addition:
IEC 60601-1:1988, Medical electrical equipment — Part 1: General requirements for safety.
Amendment 1:1991
Amendment 2:1995
IEC 60601-1-1:1992, Medical electrical equipment — Part 1: General requirements for safety — 1. Collateral
standard: Safety requirements for medical electrical systems.
Amendment 1:1995
IEC 60601-1-2:1993, Medical electrical equipment — Part 1: General requirements for safety — 2. Collateral
standard: Electromagnetic compatibility — Requirements and tests.
IEC 60645-3:1994, Audiometers — Part 3: Auditory test signals of short duration for audiometric and
neuro-otological purposes.
IEC 60825-1:1993, Safety of laser products — Part 1: Equipment classification, requirements and user’s
guide.

Annex AA (informative)
General guidance and rationale
This annex provides a concise rationale for the important requirements of this Standard and is intended
for those who are familiar with the subject of the Standard but who have not participated in its
development. An understanding of the reasons for the main requirements is considered to be essential for
the proper application of the Standard. Furthermore, as clinical practice and technology change, it is
believed that a rationale for the present requirements will facilitate any revision of the Standard
necessitated by these developments.
Guidance and rationale for particular subclauses
5.2 The APPLIED PART of the ELECTRICAL STIMULATOR should be isolated to avoid unwanted
current paths through the PATIENT due to the capacitance or a possible conductive
connection to earth.
5.6 The EQUIPMENT is usually operated with one PATIENT for periods up to 1 h and may
subsequently be used immediately with the next PATIENT. Therefore it should be
suitable for CONTINUOUS OPERATION.
6.1 p) The USER is particularly alerted to consult the instructions for use, because of the
higher output allowed.
6.8.2 aa) Because of the danger of tissue necrosis, any d.c. components of the WAVEFORMS
should be declared.
bb) Electrodes of inadequate size or unsuitable application could provoke skin reactions
or burns.
dd) Interference to the implanted devices by the stimulating current could create a
SAFETY HAZARD.

gg) The USER should be warned that stimulation with excessive current densities may
be a SAFETY HAZARD to the PATIENT.
ii) APPLIED PARTS connected to a PATIENT CIRCUIT, but not connected to the PATIENT,
should be prevented from contacting other conductive parts in order to preserve the
PATIENT CIRCUIT electrical isolation.

14.6 See rationale on 5.2.

© BSI 05-1999 9
EN 60601-2-40:1998

36.202.2.2 The use of the phantom is intended to standardize the testing for EMC, particularly for
susceptibility.
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

Contamination of the display resulting from the radiated radiofrequency


electromagnetic field tests is not considered to be hazardous and, therefore, not a
non-compliance.
Instructions for use should specify maximum electromagnetic fields within which all
performance specifications will be met, if these are necessarily lower than the levels
specified for the tests of 36.202.2.1.
46 Switching on the ELECTRICAL STIMULATOR inadvertently is considered to be a normal
occurrence since the ELECTRICAL STIMULATOR is likely to be short-circuited or
open-circuited accidentally during use due to movements of the electrodes and/or the
PATIENT.

50.1 A small increase in output amplitude may produce a disproportionate stimulus to the
PATIENT. A control which enables the USER to adjust the output amplitude smoothly or
in small increments is considered to be an important safety feature. Limitation of the
output available at the minimum setting of the output control enables the USER to
commence stimulation from a low output level.
50.2 An accuracy of ± 30 % is considered to provide adequate safety, since the values selected
are mainly determined by the electrophysiological responses and the subjective reaction
of the PATIENT.
51.101 Supply voltage fluctuations not exceeding the limit of the General Standard should not
influence the output parameters excessively.
51.102 The indication should advise the USER that the ELECTRICAL STIMULATOR is delivering
stimuli, or that the ELECTRICAL STIMULATOR is armed to deliver stimuli as a result of
further action by the USER, for example manual triggering.
51.103 Experience shows that the limits specified allow all known diagnostic applications to be
carried out without exceeding the allowable value.
51.105 The value of 125 dB HTL is taken from IEC 60645-3.

10 © BSI 05-1999
EN 60601-2-40:1998

Annex ZA (normative)
Normative references to international publications with their corresponding
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

European publications
This European Standard incorporates by dated or undated reference, provisions from other publications.
These normative references are cited at the appropriate places in the text and the publications are listed
hereafter. For dated references, subsequent amendments to or revisions of any of these publications apply
to this European Standard only when incorporated in it by amendment or revision. For undated references
the latest edition of the publication referred to applies (including amendments).
NOTE When an international publication has been modified by common modifications, indicated by (mod) the relevant EN/HD
applies.
Publication Year Title EN/HD Year

Addition to Annex ZA of EN 60601-1:1990/A2:1995:


IEC 60601-1 1988 Medical electrical equipment EN 60601-1 1990
Part 1: General requirements for safety + corr. July 1994
A1 1991 A1 1993
+ corr. July 1994
A2 1995 A2a 1995
A13 1996
IEC 60601-1-1 1992 1. Collateral standard: Safety requirements EN 60601-1-1 1993
for medical electrical systems
A1 1995 A1 1996
IEC 60601-1-2 1993 2. Collateral standard: Electromagnetic EN 60601-1-2 1993
compatibility — Requirements and tests
IEC 60825-1 1993 Safety of laser products EN 60825-1 1994
Part 1: Equipment classification, + corr. February 1995
requirements and user’s guide + A11 1996
+ corr. July 1997
a A2 includes corrigendum June 1995 to IEC 60601-1:1988/A2.

Annex ZB (informative)
Other international publications mentioned in this standard with the
references of the relevant European publications

Publication Year Title EN/HD Year

Addition to Annex ZB of EN 60601-1:1990/A2:1995:


IEC 60645-3 1994 Audiometers EN 60645-3 1995
Part 3: Auditory test signals of short
duration for audiometric and neuro-otological
purposes

© BSI 05-1999 11
BS EN
60601-2-40:
1998 BSI — British Standards Institution
IEC
BSI is the independent national body responsible for preparing
60601-2-40: British Standards. It presents the UK view on standards in Europe and at the
1998 international level. It is incorporated by Royal Charter.
BS 5724-2.40:
Revisions
Licensed Copy: Mr. Universiti Teknologi Malaysia User, Universiti Teknologi Malaysia, 18/08/2011 06:58, Uncontrolled Copy, (c) BSI

1998
British Standards are updated by amendment or revision. Users of
British Standards should make sure that they possess the latest amendments or
editions.

It is the constant aim of BSI to improve the quality of our products and services.
We would be grateful if anyone finding an inaccuracy or ambiguity while using
this British Standard would inform the Secretary of the technical committee
responsible, the identity of which can be found on the inside front cover.
Tel: 020 8996 9000. Fax: 020 8996 7400.

BSI offers members an individual updating service called PLUS which ensures
that subscribers automatically receive the latest editions of standards.

Buying standards

Orders for all BSI, international and foreign standards publications should be
addressed to Customer Services. Tel: 020 8996 9001. Fax: 020 8996 7001.

In response to orders for international standards, it is BSI policy to supply the


BSI implementation of those that have been published as British Standards,
unless otherwise requested.

Information on standards

BSI provides a wide range of information on national, European and


international standards through its Library and its Technical Help to Exporters
Service. Various BSI electronic information services are also available which give
details on all its products and services. Contact the Information Centre.
Tel: 020 8996 7111. Fax: 020 8996 7048.

Subscribing members of BSI are kept up to date with standards developments


and receive substantial discounts on the purchase price of standards. For details
of these and other benefits contact Membership Administration.
Tel: 020 8996 7002. Fax: 020 8996 7001.

Copyright

Copyright subsists in all BSI publications. BSI also holds the copyright, in the
UK, of the publications of the international standardization bodies. Except as
permitted under the Copyright, Designs and Patents Act 1988 no extract may be
reproduced, stored in a retrieval system or transmitted in any form or by any
means – electronic, photocopying, recording or otherwise – without prior written
permission from BSI.

This does not preclude the free use, in the course of implementing the standard,
of necessary details such as symbols, and size, type or grade designations. If these
details are to be used for any other purpose than implementation then the prior
written permission of BSI must be obtained.

If permission is granted, the terms may include royalty payments or a licensing


agreement. Details and advice can be obtained from the Copyright Manager.
BSI Tel: 020 8996 7070.
389 Chiswick High Road
London
W4 4AL

Vous aimerez peut-être aussi