ddR System Stand
Minimum room height 2.60 m / 102.4”
Optimum room height 2.80 m / 110.2”
C-Arm rotation -35° … +135° motorized
Detector swivel range -45° … +45° motorized, Shift function
Central beam – floor - distance 46 cm - 190 cm / 18.1” – 74.8” motorized
Floor – detector – distance min. 63 cm / 24.8” motorized
Source image distance (SID) 103 cm - 180 cm / 40.6” – 70.9”
Tube rotation ddRCompact™ +12° … -12° motorized
ddRCompact™ Plus: +12° … -180° motorized
HD-3000™ Detector
Technology High Definition Silicon Solid State
Cesium iodide scintillator (CsI:TI)
A/D conversion 14 bit / 16’384 grey levels
Active detector area 43 cm x 43 cm / 17” x 17”
Spatial resolution up to 3.3 lp/mm
Active Pixel matrix 4096 x 4096 pixels
Pixel size 103 microns
Sensitivity up to 400 speed film/screen combination
X-ray HV Generator
Power 80 kW at 100 kHz
Automatic exposure control 3-field ionization chamber
X-ray Tube
Optitop 150/40/80 HC-100L Double focus, high speed rotating anode tube
Focal spots 0.6 / 1.0 mm
Heat capacity 763 000 HU
Collimator
Multileaf collimator AL 02 Operation motorized (remote controlled) or
manual
Rotation +/- 45° manual
Integrated internal filters 0.1 mm Cu / 0.2 mm Cu / 0.3 mm Cu
Total filtration (X-ray tube + collimator) 2.5 mm Al equivalent
DAP meter (dose area product) optional
Workstation
SwissVision® XG for ddRCompact™ with DICOM Worklist
Environmental Conditions
According to IEC / EN 60601-1, UL 60601-1
Operating
Ambient temperature +10°C to +40°C
Recommended room temperature +22°C to +25°C
Relative air humidity 30% to 75%, non condensing
Air pressure 700 hPa to 1060 hPa
Storage / Transport
Ambient temperature 0°C to +50°C
Relative air humidity 10% to 80%, non condensing
Air pressure 700 hPa to 1060 hPa
Main Supply
Power consumption ddR system 1 x 208 / 230 VAC, 10 Amps, 50 Hz - 60 Hz
Power consumption X-ray generator 3 x 400 VAC, 5 Amps (160 Amps peak) Europe or
3 x 480 VAC, 5 Amps (140 Amps peak) USA
Weight
Overall 720 kg / 1587 lb
4.1 Dimensions
Our units and equipment’s must only be installed, serviced and put into operation by qualified