Académique Documents
Professionnel Documents
Culture Documents
The diffusion coefficients, D0 and Dr, of the species experimental problems (18-21). The twin-electrode thin-
ferri- and ferrocyanide, respectively, in aqueous KCI layer method proposed by Anderson and Reilley (19) for
media were determined by twin-electrode thin-layer
electrochemistry using a micrometer-type thin-layer measuring the D values of a redox couple is an absolute
cell. The values obtained for D0 and DR at 25° C in method based on measurement of the limiting steady-state
LOOM KCI were 0.726 (±0.011) X IQ-5 and 0.667(±0.014) X current, iss, followed by coulometric analysis of the thin
10-5 cmVsec, respectively, covering a concentration layer. The iss between closely spaced twin working elec-
range from 0.61 to 6.36mM in total electroactive species. trodes is given by the equation:
Identical values within experimental uncertainty were
found in 0.10M KCI medium. Criteria for the reliability
of this absolute method for measuring the D values of ~nFAC~ 2DqDr
See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.
a soluble, stable redox pair are presented, and com- l _Dq + Dr_
parisons of D values with those obtained by other
workers are made. where n is the number of electrons transferred in the redox
reaction, F is the value of the Faraday, A is the projected
Various methods have been devised for determining the electrode area, C is the total concentration of electroactive
Downloaded via TUFTS UNIV on July 12, 2018 at 13:45:10 (UTC).
diffusion coefficients (D values) of species in solution, such as species in the thin layer, / is the thickness of solution between
the use of the calibrated diaphragm cell (1-3), radiochemical the two working electrodes, and D0 and DI{ are the diffusion
tracers (4, 5), electrical conductance (6), and optical methods coefficients of the oxidized and reduced species, respectively
(7). For electroactive species in an excess of supporting (19, 22).
electrolyte, electrochemical methods often have been used. D values are calculated according to the following equations
Such methods, for the most part relative methods requiring (19):
calibration with a substance of known D value, have in-
cluded Cottrell-type (8-13), chronopotentiometric (12,14,15), iss(c)l2
Do (2)
and rotating disk (16) measurements at platinum electrodes, IqT
and polarographic measurements at the dropping mercury
electrode (11, 17). Results by these electrochemical methods iss{a)l2
Dr (3)
may differ from one another by 10-20% (13, 77). Macero
and Rulfs (13) have attributed these differences largely to
uncertainties in the effective electrode area and to the necessity where /33(C) and iSS(a) are the limiting steady-state currents at
for using a suitable reference ion. the cathode and anode, respectively, Qc is the charge
Thin-layer steady-state methods are characterized by mathe- required to reduce all of the oxidized species between the thin-
matical simplicity and relative freedom from time-dependent layer electrodes once the condition of /33 has been reached, and
Qa is the charge required to oxidize all of the reduced species
in the thin layer at the iss condition. The values of iSS(c) and
(1) J. H. Northrup and M. L. Anson, J. Physiol. Chem., 10, 523
(1929). iSS(a) will be equal if there is negligible electrolysis current along
(2) R. H. Stokes, J. Amer. Chem. Soc., 72, 763 (1950). with the steady-state current. From Equations 2 and 3, it is
(3) C. L. Rulfs, ibid., 76, 2071 (1954). evident that the D values can be determined independently
(4) J. H. Wang and F. M. Polestra, ibid., p 1584. of n, A, and C.
(5) T. A., Miller, B. Prater, J. K. Lee, and R. N. Adams, ibid., 87, From Equation 1 and Faraday’s law for the coulometry of
121 (1965).
(6) H. S. Harned and R. L. Nuttall, ibid., 69, 736 (1947). thin layers of solution (23, 24), straight lines should be ob-
(7) G. Kegeler and L. J. Cosling, ibid., p 2916. tained if the experimental values of l/z'3S(c), 1 /iss(a), Qc and Qa
(8) . A. Laitinen and I. M. Kolthoff, ibid., 61, 3344 (1939). are each plotted against the /-setting of the micrometer, giving
(9) . A. Laitinen and I. M. Kolthoff, /. Phys. Chem., 45, 1061 the respective slopes, Si/i33<C), Si/usia), $q„ and SQa. Using
(1941).
(10) I. M. Kolthoff and J. J. Lingane, “Polarography,” 2nd ed.,
Interscience, New York, N. Y., 1952, pp 409-411.
(11) M. von Stackelberg, M. Pilgram, and V. Toome, Z. Elektro- (18) R. C. Bowers and A. M. Wilson, J. Amer. Chem. Soc., 80, 2968
chem., 57, 342 (1953). (1958).
(12) P. J. Lingane, Anal. Chem., 36, 1723 (1964). (19) L. B. Anderson and C. N. Reilley, J. Electroanal. Chem., 10,
(13) D. J. Macero and C. L. Rulfs, J. Amer. Chem. Soc., 81, 2942 295(1965).
(1959). (20) H. Dahms, ibid., 11, 62 (1966).
(14) C. N. Reilley, G. W. Everett, and R. H. Johns, Anal. Chem., (21) A. T. Hubbard and F. C. Anson, in “Electroanalytical Chem-
27,483 (1955). istry,” A. J. Bard, Ed., Vol. 4, Marcel Dekker, New York, N.Y.,
(15) D. M. Oglesby, S. V. Omang, and C. N. Reilley, ibid., 37,1312 1970, pp 129-210.
(1965). (22) B. McDuffie, L. B. Anderson, and C. N. Reilley, Anal.
(16) A. J. Arvia, J. C. Bazan, and J. S. W. Carrozza, Electrochim. Chem., 38, 883 (1966).
Acta, 13, 81 (1968). (23) C. R. Christensen and F. C. Anson, ibid., 35, 205 (1963).
(17) L. Meites, “Polarographic Techniques,” 2nd ed., Interscience, (24) L. B, Anderson, B. McDuffie, and C. N. Reilley, J. Electro-
New York, N.Y., 1965, pp 145-150. anal. Chem., 12, All (1966).
RESULTS AND DISCUSSION stirring (8, 12). In the cases of electrode configurations giv-
Using aImM ferricyanide solution in 1M KC1, the TLSSV ing rise to inverse density gradients (curve A, Figure la;
curve wasNernstian for a 1-electron transfer reaction, scan- curve B, Figure 16) the larger currents are observed, with
a reproducible fluctuation in the ferrocyanide case such as the
ning in either the anodic or cathodic direction, with a half-
current potential, Eim)S), of +0.23 V, in agreement with the irregularities noted previously (8). The optimum electrode
formal standard potential of this system (25). From the configuration is defined as that giving minimum current in the
TLSSV curves, +0.50 and —0.05 V vs. the SCE were limiting chronoamperometric experiments (in the present case, top
anodic and cathodic potentials, respectively. The same electrode anode, bottom electrode cathode). Furthermore,
values were found in a 0.10M KC1 medium. it is assumed that convective stirring will be even less when
Laitinen and Kolthoff (8) pointed out that in electrolysis at the electrodes are very close together in the thin-layer experi-
horizontal shielded planar electrodes, if the direction of diffu- ments. To test this assumption, z'8S for the optimum electrode
sion was such that the less dense material (i.e., ferricyanide) configuration was compared with the z'88 for the inverse con-
formed above the material of greater density (i.e., ferrocy- figuration: values only about 2% higher were found in the
anide), agreement with theory for linear diffusion was ob- latter case; the sum of the coulometric slopes was unchanged,
tained. If the direction of diffusion was reversed, producing in accord with the concept of mass balance, but the slope ratio
an inverse density gradient, large and irregular currents were was changed significantly, giving rise to different D values.
obtained. Analogously, the experiment on determination of The closeness of the z88 values confirms the recent work of
optimum electrode configuration was run using the unshielded Schmidt-Weinmar that convection is minimized in solution
electrodes of the thin-layer cell. For the reduction of ferri- layers less than 200 µ thick (27).
cyanide (Figure la), the smaller current (curve B) is obtained Table I presents a comparison of /-zero values from the
at the bottom electrode, while for oxidation of ferrocyanide linear extrapolation of 1 /z'88<C) and 1 /iSs(a) data, illustrated in
(Figure 16), after about one minute, the smaller current Figure 2. The agreement of /-zeroes with each other and with
(curve A) is observed at the top electrode, in general agree- the /-zero from resistance measurements (Figure 2) indicates
ment with Laitinen and Kolthoff’s work. However, these the reproducibility of handling the cell, since it is taken apart
smaller currents are higher than the theoretical values (curve after each experiment in order to repolish the electrodes. It
C) for diffusion to an unshielded planar circular electrode at has been observed that if nonlimiting potentials are used (e.g.,
short times (26), presumably because of some convective the usual limiting potentials, but with electron transfer slowed
(25) Ref. 10, p 480. (27) H. G. Schmidt-Weinmar, Ber. Bunsenges. Phys. Chem., 71, 97
(26) Z. G. Soos and P. S. Lingane, J. Phys. Chem., 68, 3821 (1964). (1967).
A method and apparatus are described for determining assure proper payment. Approximately 2000 samples per
the helium content of gaseous mixtures. The work was for the Helium Conservation Program must be analyzed.
done so that the Bureau of Mines could more accurately year
and economically analyze helium-containing natural Because the previous method (5) was time consuming and
gases, crude helium purchased from private indus- required frequent calibrations, the present apparatus was
tries, and helium in gases used in research. Activated developed.
coconut charcoal Is utilized at liquid nitrogen tempera- This paper describes the new apparatus and procedure for
ture to adsorb components other than helium in the
mixture. A transducer is then used to determine the determining the helium content of gaseous mixtures. This
helium pressure. Twenty or more analyses with a method simplifies the calculations by eliminating the necessity
standard deviation of ±0.04% can be made in an 8-hour of correcting volume, temperature, and pressure. The pres-
day. sure of helium in the unknown sample is compared with the
pressure of high-purity helium at the same conditions. This
Activated coconut charcoal at liquid nitrogen temperature
results in a primary standard method of determining the
(77 °K) is used to adsorb all gases except helium and neon, helium content of samples that contain from 0.1 to 100% he-
and the neon concentration is usually negligible. This ad- lium.
sorptive property of activated charcoal was discovered by
Dewar (1) in 1875. Cady and McFarland (2) utilized acti- APPARATUS
vated coconut charcoal at liquid air temperatures to determine
the helium content of natural gas. The Bureau of Mines The gas flow and the essential elements of the analyzer are
shown schematically in Figure 1. Vacuum valve A is a pneu-
adopted this method with minor modifications as reported
matic on-off valve. Sample inject valve B is an 8-port pneu-
by Anderson (3). Frost in 1946 (4) also utilized activated matic valve and is shown in the sample flush and sample in-
charcoal to analyze concentrations of helium below 10%.
ject positions in Figure 1. Valves C-F are 3-port pneumatic
Frost, Kirkland, and Emerson (J) described an apparatus valves, and the air-control valves G-K are 3-port solenoid valves.
and procedure for determining the helium content of gases Metering valves L-N control the gas flow to flowmeters
containing 10% or more helium. However, these methods 4 and 7. The pressure is measured by a 0-1 pound-pér-
require corrections for volumes, pressures, and temperatures, square-inch-differential transducer, 1, with its electrical span
and the apparatus must be calibrated with weighed primary set to give 0 to 1.5 volts for the pressure of helium in the sam-
standard mixtures (6) to obtain accurate results. ple. The voltage is then converted by a voltage-to-frequency
The Bureau of Mines purchases crude helium from private converter and counter. The resulting reading for a 100%
helium sample is approximately 150,000 counts on the digital
industry and must accurately determine the helium content to
readout, 3, for about 0.2 psid. Flowmeters 4 and 7 are dif-
1
Present address, Health and Safety, Bureau of Mines, U. S. ferential pressure gauges that have been calibrated by using a
Department of the Interior, Denver, Colo. 80225 predetermined length of Vie-inch capillary tubing to obtain
the pressure range required (7). Charcoal trap, 8, contains
(1) J. Dewar, Nature (London), 12, 217-218 (1875). 3 grams of 50-60 mesh activated coconut charcoal. The in-
(2) Hamilton P. Cady and David F. McFarland, J. Amer. Chem.
Soc., 29, 1523-1536(1907).
let line to the charcoal is as short as possible (6 inches) to
(3) C. C. Anderson, U. S. Bur. Mines, Inform. Cir., 6796 (1934). minimize the volume of Na pusher gas that must be extracted
(4) E. M. Frost, Jr., U. S. Bur. Mines Rep. Invest., 3899 (1946). by cryogenic pumping and diffusion. Volume tank, 10 (500
(5) E. M. Frost, C. G. Kirkland, and D. E. Emerson, ibid., 6545
(1964). (7) G. W. Munns, Jr., and V. J. Frilette, J. Gas Chromatogr., 3,
(6) J. E. Miller, A. J. Carroll, and D. E. Emerson, ibid., 6674(1965). 145-146 (1965).