Vous êtes sur la page 1sur 6

Diffusion Coefficients of Ferri- and Ferrocyanide

Ions in Aqueous Media,


Using Twin-Electrode Thin-Layer Electrochemistry
S. J. Konopka and Bruce McDuffie
Department of Chemistry, State University of New York at Binghamton, Binghamton, N. Y. 13901

The diffusion coefficients, D0 and Dr, of the species experimental problems (18-21). The twin-electrode thin-
ferri- and ferrocyanide, respectively, in aqueous KCI layer method proposed by Anderson and Reilley (19) for
media were determined by twin-electrode thin-layer
electrochemistry using a micrometer-type thin-layer measuring the D values of a redox couple is an absolute
cell. The values obtained for D0 and DR at 25° C in method based on measurement of the limiting steady-state
LOOM KCI were 0.726 (±0.011) X IQ-5 and 0.667(±0.014) X current, iss, followed by coulometric analysis of the thin
10-5 cmVsec, respectively, covering a concentration layer. The iss between closely spaced twin working elec-
range from 0.61 to 6.36mM in total electroactive species. trodes is given by the equation:
Identical values within experimental uncertainty were
found in 0.10M KCI medium. Criteria for the reliability
of this absolute method for measuring the D values of ~nFAC~ 2DqDr
See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.

a soluble, stable redox pair are presented, and com- l _Dq + Dr_
parisons of D values with those obtained by other
workers are made. where n is the number of electrons transferred in the redox
reaction, F is the value of the Faraday, A is the projected
Various methods have been devised for determining the electrode area, C is the total concentration of electroactive
Downloaded via TUFTS UNIV on July 12, 2018 at 13:45:10 (UTC).

diffusion coefficients (D values) of species in solution, such as species in the thin layer, / is the thickness of solution between
the use of the calibrated diaphragm cell (1-3), radiochemical the two working electrodes, and D0 and DI{ are the diffusion
tracers (4, 5), electrical conductance (6), and optical methods coefficients of the oxidized and reduced species, respectively
(7). For electroactive species in an excess of supporting (19, 22).
electrolyte, electrochemical methods often have been used. D values are calculated according to the following equations
Such methods, for the most part relative methods requiring (19):
calibration with a substance of known D value, have in-
cluded Cottrell-type (8-13), chronopotentiometric (12,14,15), iss(c)l2
Do (2)
and rotating disk (16) measurements at platinum electrodes, IqT
and polarographic measurements at the dropping mercury
electrode (11, 17). Results by these electrochemical methods iss{a)l2
Dr (3)
may differ from one another by 10-20% (13, 77). Macero
and Rulfs (13) have attributed these differences largely to
uncertainties in the effective electrode area and to the necessity where /33(C) and iSS(a) are the limiting steady-state currents at
for using a suitable reference ion. the cathode and anode, respectively, Qc is the charge
Thin-layer steady-state methods are characterized by mathe- required to reduce all of the oxidized species between the thin-
matical simplicity and relative freedom from time-dependent layer electrodes once the condition of /33 has been reached, and
Qa is the charge required to oxidize all of the reduced species
in the thin layer at the iss condition. The values of iSS(c) and
(1) J. H. Northrup and M. L. Anson, J. Physiol. Chem., 10, 523
(1929). iSS(a) will be equal if there is negligible electrolysis current along

(2) R. H. Stokes, J. Amer. Chem. Soc., 72, 763 (1950). with the steady-state current. From Equations 2 and 3, it is
(3) C. L. Rulfs, ibid., 76, 2071 (1954). evident that the D values can be determined independently
(4) J. H. Wang and F. M. Polestra, ibid., p 1584. of n, A, and C.
(5) T. A., Miller, B. Prater, J. K. Lee, and R. N. Adams, ibid., 87, From Equation 1 and Faraday’s law for the coulometry of
121 (1965).
(6) H. S. Harned and R. L. Nuttall, ibid., 69, 736 (1947). thin layers of solution (23, 24), straight lines should be ob-
(7) G. Kegeler and L. J. Cosling, ibid., p 2916. tained if the experimental values of l/z'3S(c), 1 /iss(a), Qc and Qa
(8) . A. Laitinen and I. M. Kolthoff, ibid., 61, 3344 (1939). are each plotted against the /-setting of the micrometer, giving
(9) . A. Laitinen and I. M. Kolthoff, /. Phys. Chem., 45, 1061 the respective slopes, Si/i33<C), Si/usia), $q„ and SQa. Using
(1941).
(10) I. M. Kolthoff and J. J. Lingane, “Polarography,” 2nd ed.,
Interscience, New York, N. Y., 1952, pp 409-411.
(11) M. von Stackelberg, M. Pilgram, and V. Toome, Z. Elektro- (18) R. C. Bowers and A. M. Wilson, J. Amer. Chem. Soc., 80, 2968
chem., 57, 342 (1953). (1958).
(12) P. J. Lingane, Anal. Chem., 36, 1723 (1964). (19) L. B. Anderson and C. N. Reilley, J. Electroanal. Chem., 10,
(13) D. J. Macero and C. L. Rulfs, J. Amer. Chem. Soc., 81, 2942 295(1965).
(1959). (20) H. Dahms, ibid., 11, 62 (1966).
(14) C. N. Reilley, G. W. Everett, and R. H. Johns, Anal. Chem., (21) A. T. Hubbard and F. C. Anson, in “Electroanalytical Chem-
27,483 (1955). istry,” A. J. Bard, Ed., Vol. 4, Marcel Dekker, New York, N.Y.,
(15) D. M. Oglesby, S. V. Omang, and C. N. Reilley, ibid., 37,1312 1970, pp 129-210.
(1965). (22) B. McDuffie, L. B. Anderson, and C. N. Reilley, Anal.
(16) A. J. Arvia, J. C. Bazan, and J. S. W. Carrozza, Electrochim. Chem., 38, 883 (1966).
Acta, 13, 81 (1968). (23) C. R. Christensen and F. C. Anson, ibid., 35, 205 (1963).
(17) L. Meites, “Polarographic Techniques,” 2nd ed., Interscience, (24) L. B, Anderson, B. McDuffie, and C. N. Reilley, J. Electro-
New York, N.Y., 1965, pp 145-150. anal. Chem., 12, All (1966).

ANALYTICAL CHEMISTRY, VOL. 42, NO. 14, DECEMBER 1970 · 1741


these slopes instead of absolute values, Equations 2 and 3 between the electrodes) by measuring the resistance, R, of a
become: 10~4M NaClG4 solution at various micrometer settings as
follows: a 5 kHz alternating current signal of 5 V rms
Do = ---

(4) amplitude was passed to ground through a bridge circuit con-


2($ui„J(SJ taining in one arm, a 100-K resistor followed by a variable
1
resistor and variable capacitor in series, and in the other arm,
Dr -
a 100-K resistor, and the twin electrode cell. The bridge
(5)
2(Sm„J(SQa) was balanced to within ±0.5 mV using a Tektronix Model
Thus the D values are determined independently of a knowl- 502-A oscilloscope with differential input channels as a null
detector. With this, arrangement, R values from 800 to
edge of the absolute layer thickness, although accurate
2400 ohms were measured with a precision of ±0.5%.
measurement of increments in / are required for accurate
determination of the various slopes. Assuming a direct proportionality between R and /, the least
squares best straight line corresponding to the R data at the
In the present study, the thin-layer method is critically various settings was determined by a computer program.
evaluated using the ferri-ferrocyanide couple in aqueous The relative standard deviation of the slope constituted the
KC1 as a model redox system. Several modifications in check for linearity of the micrometer scale, and the zero-
technique and procedure are introduced to develop the in- resistance intercept gave the setting corresponding to /-
herent potential of the method. zero, with its standard deviation. (See Figure 2 and dis-
cussion below.)
EXPERIMENTAL Determination of Electrode Configuration. Cottrell-type
A micrometer-type thin-layer electrochemical cell with twin chronoamperometric experiments were performed to estab-
working electrodes, each of measured projected area 0.278 lish the optimum electrode configuration, in view of solution
cm2, was used (15, 19). Before each set of measurements density considerations (8). With a deaerated ferricyanide-
with a fresh portion of solution, the platinum surfaces were KC1 solution in the cell, the top electrode was set at the limit-
polished to a mirror-like consistency using gamma alumina ing anodic potential and adjusted so the solution layer thick-
on a Gamal polishing cloth (Fisher Scientific Co., No. 12-282). ness was 0.2 inch. After the background current became
The Teflon (Du Pont) collar and cup were flush with the elec- constant, the top electrode was stepped to the limiting re-
trode surfaces to minimize edge effects while still allowing the duction potential and its i-t curve recorded, the other twin
attainment of /-values as small as 30 microns. electrode being at ambient potential. The layer was flushed
A microscope with a filar micrometer eyepiece (Gaertner out, by raising and lowering the upper electrode several times,
No. mllO-A), was positioned horizontally about 6 inches and the same procedure was followed at the bottom elec-
from the micrometer scale of the cell and used to ensure trode. Then a ferrocyanide solution was put in the cell and
reproducibility of micrometer settings within ±0.05 micron analogous i-t curves for oxidation were recorded at the two
(µ). In each experiment, layer thicknesses were changed by electrodes, with the same solution thickness. The results
increments equivalent to the smallest division of the microm- (see Figure 1 ,a and b) are interpreted below.
eter scale, i.e., 25.4 ± 0.1 µ, the accuracy of these increments Procedure for D Value Determinations. i„w and Qa were
being limited more by the tolerance in placement of scale measured first, as follows: the micrometer was carefully
lines on the micrometer barrel than by the reproducibility adjusted to obtain the thickest desired layer, i.e., an /-value
of setting the micrometer at a given line. in the range 80-100 µ. The upper electrode was set at the
Philbrick operational amplifiers were used (15, 19) to limiting anodic potential to pre-oxidize the solution layer,
achieve potential control of the two working electrodes, and the other thin-layer electrode being at ambient potential.
to measure and integrate current. All potentials were mea- After 15 seconds, the bottom electrode (for the ferri-ferro-
sured vs. an SCE reference electrode which was in contact cyanide system) was adjusted to the limiting cathodic poten-
with the cell solution via a fiber-tip salt bridge containing tial, which initiated the limiting steady-state current, measured
the supporting electrolyte. The auxiliary electrode was a at the anode as iSS{a). The current was allowed to flow for 1
platinum wire ring. A dual pen recorder, Sargent model minute, enabling its constancy to be observed. The cathode
DSRG, was used to record current and coulometric data. was then disconnected and simultaneously integration of the
Limiting potentials were obtained by the method of thin- electrolysis current at the anode was begun. The Qa curve
layer steady-state voltammetry (TLSSV) as follows: The was recorded for 90 seconds, after which the cell was returned
electrode for measurement of steady-state current was fixed to ambient potential. The solution between the electrodes
at an extreme anodic or cathodic potential, and the other was flushed out by raising and lowering the upper electrode,
working electrode was scanned slowly (100 mV/min) from a the micrometer was set to a layer precisely 0.001 inch thinner
potential of zero steady-state current to a potential giving the than the first, and /„(«) and Qa were measured again. The
limiting steady-state current, iss, i.e., to a limiting cathodic process was repeated at successively thinner / settings. This
or anodic potential, respectively. A sensitive x-y recorder first set of measurements at the various / values was dis-
was used to record steady-state current vs. E curves. carded as a preconditioning set, then at least four sets after the
Reagent grade KC1 and distilled demineralized water were first were run, at each of at least three l values. Immediately
used to prepare a large quantity of LOOM stock supporting after the i„w and Qa data were obtained, iSS(c) and Qc were
electrolyte solution, which was diluted to O.lOOMas necessary. measured in an exactly analogous fashion by monitoring at
Weighed amounts of reagent grade KsFe(CN)e or K4Fe(CN)e· the cathode and disconnecting the anode for coulometry.
3H20 were added to dry flasks and diluted to volume with An identical sequence of /-values was used.
supporting electrolyte to obtain the desired concentration of The procedure described above constitutes one experiment
electroactive species. Solutions containing electroactive or run. To allow for edge effects, the value of Q for each
species were used the same day as prepared. All solutions Q-t curve was obtained by linear extrapolation of the coulo-
were deaerated with presaturated nitrogen prior to use and metric curve to zero time using the tangent to the curve in the
kept under a blanket of nitrogen during experiments (15). 60-75 second region. All data were corrected for background
Experiments were conducted at 25.0 ± 0.1 °C in a constant effects. The slopes of the 1 /iss-l and Q-l lines were deter-
temperature room, the temperature of the air and solutions mined by a least squares computer program, which also gave
being monitored with a calibrated thermistor probe. the / intercepts and the relative standard deviations of the
Resistance Method for Calibration of Micrometer. The slopes and intercepts. Then D0 and DR with their cumula-
micrometer cell was calibrated for linearity and for /-zero tive standard deviations were calculated from Equations 4
(i.e., the setting corresponding to zero solution thickness and 5.

1742 e ANALYTICAL CHEMISTRY, VOL. 42, NO. 14, DECEMBER 1970


Figure 1. Effect of electrode configuration on currents in chronoamperometric experiments
a. Reduction of 3.6 mM ferricyanide, at potential of —0.05 b. Oxidation of 4 mM ferrocyanide, at potential of +0.50
V vs. SCE V vs. SCE
Curves A and B show response of top and bottom electrodes, respectively; layer thickness approximately 0.2 inch. Curve C is a
theoretical curve for diffusion to an unshielded circular, planar electrode (26), of area 0.278 cm2, at stated concentrations

RESULTS AND DISCUSSION stirring (8, 12). In the cases of electrode configurations giv-
Using aImM ferricyanide solution in 1M KC1, the TLSSV ing rise to inverse density gradients (curve A, Figure la;
curve wasNernstian for a 1-electron transfer reaction, scan- curve B, Figure 16) the larger currents are observed, with
a reproducible fluctuation in the ferrocyanide case such as the
ning in either the anodic or cathodic direction, with a half-
current potential, Eim)S), of +0.23 V, in agreement with the irregularities noted previously (8). The optimum electrode
formal standard potential of this system (25). From the configuration is defined as that giving minimum current in the
TLSSV curves, +0.50 and —0.05 V vs. the SCE were limiting chronoamperometric experiments (in the present case, top
anodic and cathodic potentials, respectively. The same electrode anode, bottom electrode cathode). Furthermore,
values were found in a 0.10M KC1 medium. it is assumed that convective stirring will be even less when
Laitinen and Kolthoff (8) pointed out that in electrolysis at the electrodes are very close together in the thin-layer experi-
horizontal shielded planar electrodes, if the direction of diffu- ments. To test this assumption, z'8S for the optimum electrode
sion was such that the less dense material (i.e., ferricyanide) configuration was compared with the z'88 for the inverse con-
formed above the material of greater density (i.e., ferrocy- figuration: values only about 2% higher were found in the
anide), agreement with theory for linear diffusion was ob- latter case; the sum of the coulometric slopes was unchanged,
tained. If the direction of diffusion was reversed, producing in accord with the concept of mass balance, but the slope ratio
an inverse density gradient, large and irregular currents were was changed significantly, giving rise to different D values.
obtained. Analogously, the experiment on determination of The closeness of the z88 values confirms the recent work of
optimum electrode configuration was run using the unshielded Schmidt-Weinmar that convection is minimized in solution
electrodes of the thin-layer cell. For the reduction of ferri- layers less than 200 µ thick (27).
cyanide (Figure la), the smaller current (curve B) is obtained Table I presents a comparison of /-zero values from the
at the bottom electrode, while for oxidation of ferrocyanide linear extrapolation of 1 /z'88<C) and 1 /iSs(a) data, illustrated in
(Figure 16), after about one minute, the smaller current Figure 2. The agreement of /-zeroes with each other and with
(curve A) is observed at the top electrode, in general agree- the /-zero from resistance measurements (Figure 2) indicates
ment with Laitinen and Kolthoff’s work. However, these the reproducibility of handling the cell, since it is taken apart
smaller currents are higher than the theoretical values (curve after each experiment in order to repolish the electrodes. It
C) for diffusion to an unshielded planar circular electrode at has been observed that if nonlimiting potentials are used (e.g.,
short times (26), presumably because of some convective the usual limiting potentials, but with electron transfer slowed

(25) Ref. 10, p 480. (27) H. G. Schmidt-Weinmar, Ber. Bunsenges. Phys. Chem., 71, 97
(26) Z. G. Soos and P. S. Lingane, J. Phys. Chem., 68, 3821 (1964). (1967).

ANALYTICAL CHEMISTRY, VOL. 42, NO. 14, DECEMBER 1970 · 1743


0.0100 2500
Table I. Reproducibility of /-Zero Measurements
Data No. of
plot Medium expts" /-zero6
av ± std dev
l//»s(o) 1MKC1 10 0.136817 ± 0.000018
0.1MKC1 3 0.136798 ± 0.000003
l//ss(c) 1MKC1 10 0.136822 ± 0.000015
0.1MKC1 3 0.136822 ± 0.000032

For concentrations of ferri- and ferrocyanide, see
Table II.
6
Micrometer setting in inches corresponding to solution layer of
zero thickness.

Table II. Dependence of Steady-State Current


on Concentration
C, mM" Sl/Mo) Kh
1.00M KC1 Medium
0.30 17.89 5.37
0.61= 8.69 5.38
1.10 4.89 5.36
1.10 4.98 5.48
1.36 3.93 5.32
1.55 3.40 5.26
5.02= 1.06 5.33
5.09 1.05 5.32
5.23d 1.04 5.45
6.36 0.85 5.43
Figure 2. Comparison of methods for calibration of av 5.37
twin-electrode thin-layer cell std dev 0.07 (1.2%)
1/iOs) vs. I data; 5.23 mM ferrocyanide in l.OOM KC1. R 0.10M KC1 Medium
vs. I data: 10-4M Na004 solution 0.91 6.00 5.46
5.01d 1.06 5.32
5.09 1.07 5.45
by contamination of the electrode surfaces) the currents will av 5.41
be less than theoretical, and the /-intercept from the 1 //„ std dev 0.08
vs. I plots will be at a much smaller micrometer setting than “
100 % ferricyanide except as otherwise noted.
the /-zero obtained from the resistance measurements. Slight 6
K [Sin„(»)] [C], See text.
=

contamination will produce a shift in the /-intercept several =


Equimolar ferri-ferrocyanide mixture.
orders of magnitude greater than the standard deviation for
d
Ferrocyanide solution.
the measurement. This effect can be tested easily by measur-
ing the steady-state current at various /-settings without clean-
ing the electrodes, then cleaning the electrodes and repeating (±0.05) X 104 and 2.70 (±0.04) X 104 coulomb cm2 mole-1
the measurements on another portion of the same solution. for 1.00 and 0.10M KC1, respectively, were within 1 % of the
The currents in the first case will be much lower than those in value calculated from the measured electrode area.
the second case, resulting in widely varying /-zeroes. The individual diffusion coefficients found for the ferri-
According to Equation 1, /„ is proportional to C. Al- and ferrocyanide species, with respective relative standard
ternatively using Sinss, one obtains: deviations, are presented in Table III. Each D value, calcu-
1 lated from Equations 4 and 5, is based on the slopes of two
DSi/UtC] (6) least squares data plots. The standard deviation, sd, of
nFA Du
each D value is the square root of the sum of the squares of
where the right hand term, hereafter designated K, is a cali- the relative standard deviations of the two slopes used. In
bration constant for a given redox system and twin-electrode 1M KC1 the value of sd0 ranged from 0.5 to 2.0% and sdr
cell, Du being defined as a weighted diffusion coefficient from 0.9 to 3.2%, the averages being 1.4 and 1.6%, respec-
equal to 2D0D¡¡/(D0 + Dr). Table II lists the values of K tively. The agreement of these individual sd values with the
obtained for a range of concentrations in the two KC1 media. standard deviations of the groups of D values, 1.5% and 2.1%,
The results indicate that the current is a true steady-state respectively, indicates that there is no significant variance
current in agreement with Equations 1 and 6. Since K is between experiments. Similar comparisons of standard
the same within experimental error in 1.00 as well as 0.1 OM deviations can be made using the 0.1M KC1 data. (The last
KC1, the value of Dw is the same for the ferri-ferrocyanide two results in 1MKC1 were obtained approximately 6 months
system in both media. after the other results, indicating the stability and reproduci-
From mass balance considerations, the coulometric data bility of the cell and the method over a period of time.)
should follow the relationship From Table III it is evident that the diffusion coefficients are
nFA independent of the total concentration of electroactive species,
[SQa + SgJ/C =
.
(7) within the range of concentrations used, in accord with Pick’s
Thus, to establish the reliability of the coulometric measure- first law. Furthermore, the D values are independent of the
ments, the quantity nFA was calculated from Equation 7 initial concentration ratio of oxidized to reduced species.
for each experiment. The average values so obtained, 2.71 With this redox system, both D0 and Dr could be obtained

1744 · ANALYTICAL CHEMISTRY, VOL. 42, NO. 14, DECEMBER 1970


Table III. Diffusion Coefficients of Ferri- and Table IV. Comparison with Selected Literature Values for
Ferrocyanide at 25 °C Ferri- and Ferrocyanide at 25 °C
Concn'1 Do X 10s Du X 10= Do X 10= Dr X 10=
mM :m2 sec-1
i ¡¡DO, % cm2 sec-1 SDR, % Method Medium cm2 sec-1 cm2 sec-1 Ref
1.00M KC1 medium Cottrell, absolute: 0.1MKC1 0.842 0.717 8
0.61= 0.731 2.0
0.1MKC1 0.836 0.687 13
1.9 0.671 1.9 0.1MKC1 0.763 0.650 11
1.10 0.731
0.718 1.4 0.697 2.1
0.1MKC1 0.670 12
1.10
1.36 0.719 1.7 0.673 3.2 Chronopotentio-
1.55 0.737 1.7 0.656 2.6 metric 0.1MKC1 0.697 12
0.718 0.7 0.667 0.9 Pt, relative to Pb 0.1MKC1 0.713 0.638 10
5.02=
0.705 1.6 0.666 1.0 Rot. Pt disk 1.0MKC1 0.677 0.588 16
5.09
5.23= 0.721 0.8 0.668 1.2 Thin layer 1.0MKC1 0.726 0.667 This work
6.36 0.741 1.9 0.641 1.1 Thin layer 0.1MKC1 0.72 0.67 This work
4.47= 0.730 0.5 0.668 1.0
5.00 0.738 0.9 0.664 1.2 difference between KC1 and KNCL as supporting electrolytes.
av 0.726 av 1.4 av 0.667 av 1.6 A comparison of results with those obtained by other meth-
std devd 0.011 (1.5%) std dev 0.014(2.1%) ods is given in Table IV. The first set of results (8) may be
0.10M KC1 medium high because of electromigration effects in the Cottrell ex-
0.91 0.739 1.7 0.681 2.1 periment, since a 5mM electroactive species concentration
5.01 =
0.717 0.7 0.656 1.6 was used with only a 0.1M supporting electrolyte concentra-
5.09 0.703 1.2 0.660 1.3 tion. [In our method, although these same concentrations
av 0.720 av 1.2 av 0.666 av 1.7 were used in two experiments, no appreciable electromigration
std devd 0.018 (2.5%) std dev 0.013(2.0%) effect was observed.] The high value of Do in the next set

100% ferricyanide except as otherwise noted. (13) may be traced to the neglect of solution density effects;
b
Equimolar ferri-ferrocyanide mixture.
=
Ferrocyanide solution. ferricyanide was reduced at an electrode where the direction
d
Standard deviation of the group of measurements, not standard of diffusion was upward, forming the heavier ferrocyanide
deviation of the mean. and thus increased current due to convection. The remaining
results show reasonable agreement except for the D values
obtained using a rotating platinum disk electrode (16),
at a total concentration of electroactive species of lmM or where nonlimiting potentials may have been used.
greater. At lower concentrations, the iss values were still Effect of Nonparallelism of Electrodes. Although the work-
reliable, but Q determinations became erratic. Electrode ing electrode faces are considered to be parallel to each other,
interferences may be responsible for erratic coulometric
nonparallelism could introduce errors in the steady-state
results at low concentrations. current. .This problem was treated by Anderson and Reilley
The D values obtained in 1M KC1 give Stoke’s law radii for square facing electrodes (19). For circular electrodes,
(IT) of 3.38 A and 3.68 Á for the diffusing ferri- and ferro- graphical integration yields the following results (same nota-
cyanide species, respectively. Both ions tend to ion pair with tion as in reference 19): 1 //8, will be 6.4% less than for per-
K+ (28, 29), and ferrocyanide appears to form an ion triplet
fectly parallel surfaces if / 2q, 2.7% less if /
=
3q, and 1.5%
=

to some extent (29). A small part of the observed difference less if l =


4q. [Corresponding decreases for the square
in Stoke’s law radii could be attributed to ion triplet formation electrode case (19) are 9%, 3.8%, and 2.1%, respectively.
in the ferrocyanide case, but a larger part of the difference must From the micrometer setting at which the cell “shorted" in
have some other explanation, such as a greater extent of hydra- the present work, it appears that the thinnest layer used for
tion of the ferrocyanide species over that of the ferricyanide. data collection was at / > 2q, thicker settings corresponding
The calibration constant, K, can be used with n, F, and the
roughly to / > 3>q and / > 4q. Thus nonparallelism could
measured value of A to calculate the quantity, Dw. Since Dw have caused the slopes of the 1 ¡iss vs. /lines to be high, and the
is defined as 2D0DrI(D0 + Dr), it can be calculated from the D values low, by a maximum of 3 %, only slightly larger than
experimentally determined values of D0 and DR which have the standard deviation of the experimental results. For
no dependence on C, A, or n. From the average values for increased accuracy, the degree of nonparallelism or tilt could
the 1.00M KC1 medium (Table II), Dw = 0.694 X 10"5 be reduced by a more careful polishing and assembling tech-
cm2 sec-1, whereas from the average D values (Table III),
nique.
Dw = 0.695 X 10-6 cm2 sec-1. In 0.10M KC1 the corre-
Coulometry with Elimination of Charging Current. For
sponding values are 0.689 and 0.692 X 10-5 cm2 sec-1, re- thin-layer potential-step coulometry in a cell with only one
spectively. This agreement demonstrates that the twin- working electrode, Faraday’s law takes the form (24)
electrode thin-layer method is an absolute method: i.e„
D values can be obtained for a soluble, stable redox pair with- Q =
tlFAlC + Qads + Qd.l.
out a knowledge of n, A, and C and without the use of a where Qaae and Qa.i. represent the contributions caused by
reference ion. reaction of adsorbed species and by change in double layer
In one exploratory experiment, from steady-state current
capacitance, respectively. A plot of Q vs. I gives an intercept
data, a Dv; value of 0.709 X 10-5 cm2 sec-1 was obtained in at /-zero equal to Qais + Qd.i., neither of which is usually
O.IMKNOj, [This datum was obtained five years earlier in known. [The slope, Sq, equal to nFAC, provides the basis
another laboratory using a 2mM ferri-ferrocyanide mixture for determining C.]
in the same type of thin layer cell] This result indicates little
By contrast, in the coulometric method used here for D
value determinations, there should be no significant contribu-
(28) J. C. James and C. B. Monk, Trans. Faraday Soc., 46, 1041
(1950). tion from charging current. The electrode used for coulom-
(29) S. R. Cohen and R. A. Plane, J. Phys. Chem., 61, 1096 (1957). etry after initiation of the steady-state current remains at

ANALYTICAL CHEMISTRY, VOL. 42, NO. 14, DECEMBER 1970 · 1745


constant potential and has immediately adjacent to it in the electroactive species at the anode and at the cathode appear
solution nearly the same concentration of electroactive species to be 2.0 X 10-9 and 2.6 X 10-9, respectively. These
(as 100% O or 100% R) before and after coulometry. Thus results would indicate several monolayers of electroactive
one would expect no appreciable change in the double layer material. Obviously further studies of this aspect of the
capacitance of this working electrode. From a Q-l plot, system are needed.
the number of coulombs at /-zero (obtained from /S5 or R data)
can be expected to give Qaas directly without any Qua. term:
the intercept from the Qa-l data should give the amount of Received for review June 1, 1970. Accepted September 17,
species R previously adsorbed at the cathode, and the Qc-l 1970. Presented in part at the Pittsburgh Conference on
intercept should give the amount of species O previously Analytical Chemistry and Applied Spectroscopy, Cleveland,
adsorbed at the anode, assuming that the time required for Ohio, March 5, 1970. The authors acknowledge financial
desorption of these species is short compared to the duration support of the National Science Foundation under Grant
of the individual Q-t curves. At the higher concentrations NSF-G9925 and the Research Foundation, State University of
used in the present studies, the amounts of adsorption of New York, under Grants 40-223A, FRF-40-035, and 40-223B.

Method and Apparatus for Determining


Helium Content of Gas Mixtures
David E. Emerson and Richard L. Kaplan1
Division of Helium, Bureau of Mines, V. S. Department of the Interior, Amarillo, Texas 79106

A method and apparatus are described for determining assure proper payment. Approximately 2000 samples per
the helium content of gaseous mixtures. The work was for the Helium Conservation Program must be analyzed.
done so that the Bureau of Mines could more accurately year
and economically analyze helium-containing natural Because the previous method (5) was time consuming and
gases, crude helium purchased from private indus- required frequent calibrations, the present apparatus was
tries, and helium in gases used in research. Activated developed.
coconut charcoal Is utilized at liquid nitrogen tempera- This paper describes the new apparatus and procedure for
ture to adsorb components other than helium in the
mixture. A transducer is then used to determine the determining the helium content of gaseous mixtures. This
helium pressure. Twenty or more analyses with a method simplifies the calculations by eliminating the necessity
standard deviation of ±0.04% can be made in an 8-hour of correcting volume, temperature, and pressure. The pres-
day. sure of helium in the unknown sample is compared with the
pressure of high-purity helium at the same conditions. This
Activated coconut charcoal at liquid nitrogen temperature
results in a primary standard method of determining the
(77 °K) is used to adsorb all gases except helium and neon, helium content of samples that contain from 0.1 to 100% he-
and the neon concentration is usually negligible. This ad- lium.
sorptive property of activated charcoal was discovered by
Dewar (1) in 1875. Cady and McFarland (2) utilized acti- APPARATUS
vated coconut charcoal at liquid air temperatures to determine
the helium content of natural gas. The Bureau of Mines The gas flow and the essential elements of the analyzer are
shown schematically in Figure 1. Vacuum valve A is a pneu-
adopted this method with minor modifications as reported
matic on-off valve. Sample inject valve B is an 8-port pneu-
by Anderson (3). Frost in 1946 (4) also utilized activated matic valve and is shown in the sample flush and sample in-
charcoal to analyze concentrations of helium below 10%.
ject positions in Figure 1. Valves C-F are 3-port pneumatic
Frost, Kirkland, and Emerson (J) described an apparatus valves, and the air-control valves G-K are 3-port solenoid valves.
and procedure for determining the helium content of gases Metering valves L-N control the gas flow to flowmeters
containing 10% or more helium. However, these methods 4 and 7. The pressure is measured by a 0-1 pound-pér-
require corrections for volumes, pressures, and temperatures, square-inch-differential transducer, 1, with its electrical span
and the apparatus must be calibrated with weighed primary set to give 0 to 1.5 volts for the pressure of helium in the sam-
standard mixtures (6) to obtain accurate results. ple. The voltage is then converted by a voltage-to-frequency
The Bureau of Mines purchases crude helium from private converter and counter. The resulting reading for a 100%
helium sample is approximately 150,000 counts on the digital
industry and must accurately determine the helium content to
readout, 3, for about 0.2 psid. Flowmeters 4 and 7 are dif-
1
Present address, Health and Safety, Bureau of Mines, U. S. ferential pressure gauges that have been calibrated by using a
Department of the Interior, Denver, Colo. 80225 predetermined length of Vie-inch capillary tubing to obtain
the pressure range required (7). Charcoal trap, 8, contains
(1) J. Dewar, Nature (London), 12, 217-218 (1875). 3 grams of 50-60 mesh activated coconut charcoal. The in-
(2) Hamilton P. Cady and David F. McFarland, J. Amer. Chem.
Soc., 29, 1523-1536(1907).
let line to the charcoal is as short as possible (6 inches) to
(3) C. C. Anderson, U. S. Bur. Mines, Inform. Cir., 6796 (1934). minimize the volume of Na pusher gas that must be extracted
(4) E. M. Frost, Jr., U. S. Bur. Mines Rep. Invest., 3899 (1946). by cryogenic pumping and diffusion. Volume tank, 10 (500
(5) E. M. Frost, C. G. Kirkland, and D. E. Emerson, ibid., 6545
(1964). (7) G. W. Munns, Jr., and V. J. Frilette, J. Gas Chromatogr., 3,
(6) J. E. Miller, A. J. Carroll, and D. E. Emerson, ibid., 6674(1965). 145-146 (1965).

1746 · ANALYTICAL CHEMISTRY, VOL. 42, NO. 14, DECEMBER 1970

Vous aimerez peut-être aussi