Académique Documents
Professionnel Documents
Culture Documents
OPHTHALMOMETER
OM-4
Thank you for purchasing the TOPCON Ophthalmometer OM-4. To get the from the best use
from the instrument, please carefully read these instructions and place it in a convenient location
for future reference.
PRECAUTIONS
1. Always use or keep this instrument where it will be unaffected by heat, humid-
ity or dust, and avoid exposing the instrument to direct sunlight.
2. Check that all cables are correctly and firmly connected.
3. Never touch the lens or prism surfaces with your finger or with any hard ob-
ject.
4. Always turn the power source off before disconnecting the cables or when re-
placing the bulb.
5. When disconnecting the cables, do not use excessive force and never attempt
to unplug the connection by directly pulling on the cable.
6. If any trouble occurs with your instrument or its accessories, first refer to the
troubleshooting guide in this manual and carry out the checks listed. If nothing
is found by your check, then ask your dealer or TOPCON to service it.
7. Always turn the power source off and replace the cover on the instrument
when it is not in use.
CONTENTS
1. NOMENCLATURE .................................................................................................................... 1
2. ASSEMBLY ............................................................................................................................... 3
2.1 Components...................................................................................................................... 3
2.2 Assembly Procedure ........................................................................................................ 4
3. OPERATION PROCEDURES................................................................................................... 6
3.1 Preparations...................................................................................................................... 6
3.2 Practice measurements with the test ball ......................................................................... 6
3.3 Measurement of the radius of curvature of the cornea..................................................... 8
3.4 Measurement of contact lenses...................................................................................... 10
4. MAINTENANCE ...................................................................................................................... 11
4.1 Exchange of the lamp ..................................................................................................... 11
4.2 Exchange of the fuse ...................................................................................................... 11
4.3 Cleaning the instrument .................................................................................................. 11
4.4 General care ................................................................................................................... 11
6. FEATURES.............................................................................................................................. 12
7. SPECIFICATIONS................................................................................................................... 13
1
1. NOMENCLATURE
Fig. 1
2
Fig. 2
3
2. ASSEMBLY
2.1 Components
The following items are contained in a packing box:
Instruction manual 1 each
Conversion table 1 each
Dust cover 1 each
Table top 1 each
Accessory box 1 set
Rail covers 1 pair
Power cord 1 each
Chin rest and head rest 1 each
Screwdriver 1 each
Measuring head and base 1 each
Fig. 3 Fig. 4
Fig. 5 Fig. 6
4
• The contents of the accessory box are: only screwed into their screwholes temporarily and
Spanner (not supplied with the table for ophthalmic can be unscrewed very easily. Use them to fix the
stand) 1 each table on top of the elevating mechanism of the
Silicone cloth 1 each adjustable instrument table, as illustrated in Fig. 8.
Lamps (spares) 9 each
Contact lens holder 1 each b) Installing the ophthalmic stand. Simply insert
Test ball (R 7.50) 1 each the pole protruding from the bottom surface of
Fuse 2 each the table into the socket on the lower instru-
for IEC 0.5A 2 each ment arm and tighten the fixing screw. as
6A 1 each shown in Fig. 9.
Chin rest pads 1 pack
Column 1 each
Cleaning brush 1 each
Pad fixing pin 2 each
Fig. 9
Fig. 10
Fig. 8
5
Voltage selector
for IEC
Fig. 11
Fig. 14
Fig. 12
6) Check the light
5) Connecting the Electrical Cords Switch on the power switch pilot lamp, mire
Switch off the power switch (5). lamp and ill scale are lighted.
Then, connect the power cord (3) and con- and check whether illumination lamp for scale
nector cord (1) to the room receptacle and are lighted.
connector respectively as in Fig. 13 and 14.
6
3. OPERATION PROCEDURES
3.1 Preparations 6) Set 0°, or 180°, of the protractor scale (15) to
1) Use a room which is dimly lighted since a room the black-colored horizontal index (17) as in Fig.
in total darkness will be too dark to make the re- 18.
quired adjustments.
If a special room cannot be prepared, use a room
without direct sunlight and cover the windows with
dark thick curtains.
Furthermore, Iocate the instrument so that the pa-
tient will not be looking towards the window.
2) Connect the power cord (3) to the room recepta-
cle, if it has not been connected yet.
3) Switching on the power switch (5) and check
whether the pilot.
4) Set the illumination control switch (6) to N-
Position.
5) Adjust the instrument to the observer’s eye.
Fig.18
First, draw out the eyepiece (13) as far as possible,
by revolving it counter-clockwise to its limit.
3.2 Practice Measurements with the Test
Next, place a white paper before the mire plate (23)
Ball
at the position of the patient’s eye, as in Fig. 15 and
1) Attachment of the column
a blurred black circle image will be seen, as in Fig.
Set up the column (27) in the opening available on
1 6. Revolve the eyepiece slowly in the clockwise
the chin rest. Then, place the test ball (26) on top
direction and stop when the circle image is seen
of the column, as in Fig. 19.
distinctly as double circle, as in Fig. 17.
Fig. 19
Fig. 16 Fig. 17
7
ball on the same line of sight. as in Fig. 20. The mire image in the center of the field of view will
This operation will position the test ball and instru- be seen doubled. as in Fig. 22, when the instru-
ment (measuring head) at the correct height. ment is not positioned at the correct distance to the
test ball.
Fig. 20 Fig. 22
Operate the main body with hold the joystick softly The instrument will be set to the correct distance
like. as shown in Fig. 21. And following operating is when the double images. in the center of the field
operated in holding the joystick: of view. are coincided into a single image by ad-
• Elevation control of instrument justing the joystick. The coincided image should be
(Rotating the joystick) located approximately in the center of the field of
• Roughly horizontal positioning of the instru-
view, at this time. but, if it is not, use joystick and
center it properly so that it is concentric with the
ment. (Slide the main body with holding the joy-
double images in the center of the field of view. as
stick)
in Fig. 23.
• Accurately horizontal positioning of the instru-
ment (Lean the joystick)
Fig. 23
3) Measurement
Three mire images, as in Fig. 23, are to be seen in
Fig. 21 the eyepiece at this time, and the horizontal knob
(11) and vertical knob (12) should be revolved to
Next, use the joystick and position the measuring coincide the I mark of the left mire image with the
head in front of the test ball. while looking into the. mark of the center mire image, and the
eyepiece, so that the mire images show up in the mark of the top mire image with the mark of
field of view. the center image, which will be the result of the
8
measurement (see Fig. 24), to be read as follows: head rest (22), as in Fig. 25.
Fig. 24
Fig. 28
Fig. 27
3) Measurements
a) When there is corneal astigmatism.
Turn the axis rotating handle (9). as in Fig. 28,
when corneal astigmatism exists, the mark
of the top mire image and the mark of the
center mire image will be seen displaced, as in
Fig. 29.
Fig. 30
Fig. 32 Fig. 33
4. MAINTENANCE
Fuse holder
4.1 Exchange of the Lamp
The mire place (23) and lamp will be quite hot,
should the lamp go out while the instrument is in
use. Therefore, let them cool down a little bit and
then replace lamp in the following manner:
(1) After switching off the Power switch (5) of the
Power supply unit (4), disconnect the Power
cord (3).
(2) Take a firm hold of the leveling pin (16) and
revolve it about two or three times in the coun-
ter-clockwise direction. to loosen.
for IEC
(3) Pull out the mire plate next which will expose
the lamp. Revolve it in the counter-clockwise di-
rection, too, and detach. Fuse holder
(4) Insert the replacement bulb and revolve it
clockwise until it stops and is securely attached.
(5) Replace the mire plate carefully. The mire plate
is notched, therefore, line up the notched sec-
tion with an alignment pin inside the instrument.
Push it in fully and then screw in the leveling pin
which will fix the mire plate securely.
(6) Connect the power cord, Switch on the Power
switch, and check whether illumination lamps
are lighted.
Fig. 36
Fig. 35
6. FEATURES
6.1 Internal Reading Scales
The millimetre scales for the radius of curvature
(bottom figures) and the diopter scales for the cor-
neal refractive power (top fi9ures) of both horizon-
tal(H) and vertical(V) axes, are seen in the field of
view, together with the mire images, during the
measurement.
7. SPECIFICATIONS
Measuring Head
Type ........................................................................... Sutcliffe type
Magnification .............................................................. 20x
Eyepiece adjustment range ........................................ 0 to ±5 dptr
Measuring range
Base
Power Unit
Primary ....................................................................... AC 100V, 120V, 220V, 240V;
................................................................................... adjustable with voltage selector
Frequency................................................................... 50/60 Hz
Secondary .................................................................. AC 6V and 7.5V; adjustable with switch
Power Consumption ................................................... 30VA
OM-4
TOPCON CORPORATION
75-1 Hasunuma-cho,Itabashi-ku,Tokyo,174-8580 Japan.
Phone:3-3558-2520 Fax:3-3960-4214