Académique Documents
Professionnel Documents
Culture Documents
IMR4000
The compact flue gas analyzer
for industrial applications
© M&C® TechGroup Gentics® GmbH // +49 71 32/96 06-0 Technical specifications and illustrations are without 01.15
1/3
gentics@mc-techgroup.com // www.mc-gentics.com obligation, subject to modifications. 1.00.00
// IMR4000 // INDUSTRIAL; LABORATORY; R&D
Up to 12 gases can be measured simulta- The oxygen content in the flue gas is neces-
neously. Four additional components can be sary for calculation of variables such as ex-
displayed. The physical dimensions of the cess air. The simultaneous measurement of
analyzer permit the use even in hard-to-ac- flue gas volume and velocity is available as
cess areas. an option.
The most common applications are: Measuring results can be printed on the
built-in printer as ‘standard protocol’ or
Emission analysis ‘mean value’. Data can also be transferred
Furnace optimization via the RS232 serial interface, an USB 2.0 or
a 0…10 V (0/4…20 mA) analogue signal port.*
Environmental monitoring
Service revision; e.g. in power stations
TECHNICAL DATA
SMALLEST LARGEST
COMPONENT METHOD MAESURING MEASURING RESOLUTION ACCURACY
RANGE RANGE
± 0,3%
O2 (Oxygen) electrochem. sensor 0 ... 20,95 Vol.-% 0 ... 20,95 Vol.-% 0,01 Vol.-%
absolute
CO (Carbon monoxide) 0 ... 75 mg/m³ 0 ... 5 Vol.-%
NO (Nitric oxide) 0 ... 200 mg/m³ 0 ... 5.000 mg/m³
< 100 mg/m³: 0,1 mg
max. ± 3%
NO2 (Nitric dioxide)* electrochem. sensor 0 ... 100 mg/m³ 0 ... 500 mg/m³ > 100 mg/m³: 1,0 mg
of range
SO2 (Sulphur dioxide)* 0 ... 75 mg/m³ 0 ... 5.000 mg/m³
H2S (Hydrogen sulfide)* 0 ... 60 mg/m³ 0 ... 300 mg/m³
H2 (Hydrogen)* TCD 0 ... 20 Vol.-% 0 ... 100 Vol.-% 0,1 Vol.-% ± 3%
± 0,3%
O2 (Oxygen)* Paramagnetic sensor 0 ... 20,95 Vol.-% 0 ... 100 Vol.-% 0,01 Vol.-%
absolute
CxHy (Hydro carbons)* 0 ... 0,2 Vol.-% 0 ... 100 Vol.-%
CO2 (Carbon dioxide)* Infrared sensor 0 ... 20 Vol.-% 0 ... 100 Vol.-% 0,1 Vol.-% ± 2%
CO (Carbon monoxide)* 0 ... 20 Vol.-% 0 ... 100 Vol.-%
CxHy (Hydro carbons)*
Pelistor sensor 0 ... 4,4 Vol.-% CH4 10 ppm < 5%
(< LEL)
°C (flue gas temperature) K-type thermocouple 0 ... 500 °C 0 ... 1.605 °C 1K ± 1K
© M&C® TechGroup Gentics® GmbH // +49 71 32/96 06-0 Technical specifications and illustrations are without 01.15
2/3
gentics@mc-techgroup.com // www.mc-gentics.com obligation, subject to modifications. 1.00.00
// IMR4000 // INDUSTRIAL; LABORATORY; R&D
SMALLEST LARGEST
VARIABLE METHOD MAESURING MEASURING RESOLUTION ACCURACY
RANGE RANGE
-20°...120 °C -20°...120 °C
°C (Ambient air temperature) K-type thermocouple 1K ±1K
-4° ... 248 °F -4° ... 248 °F
± 20 hPa ± 300 hPa
hPa Pressure*/Draft Semiconductor sensor 0,01 ± 3%
(mm H2O) (mm H2O)
± 0,3%
CO2 (Carbon dioxide)** calculated 0...CO2 max. *** 0,1 Vol-%
absolute
λ (Lambda)/excess air calculated 1,0...99,9 0,01 ± 2%
v (flue gas velocity)* Semiconductor sensor 0...5 m/s 0...25 m/s 0,001 m/s ± 2%
Filter paper method Volume-regulated suction pump
Soot (soot number)*
according to DIN 51402 1,63 l/min ± 0,07 l/min
The analyzer complies with EN 50379-2, TÜV approved
* Option
** No CO2 calculation, if CO2 sensor is installed DIN EN 50379-2
*** Depending on fuel
© M&C® TechGroup Gentics® GmbH // +49 71 32/96 06-0 Technical specifications and illustrations are without 01.15
3/3
gentics@mc-techgroup.com // www.mc-gentics.com obligation, subject to modifications. 1.00.00