Académique Documents
Professionnel Documents
Culture Documents
gehealthcare.com
Technical
Publication
Direction 5275601-100
Revision 2
GE Healthcare
LING LONG Medical Diagnostic Radiography
XR 6000 Service Manual
0459
Page 2
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
Legal Notes
TRADEMARKS
All other products and their name brands are the trademarks of their respective holders.
COPYRIGHTS
All Material, Copyright © 2007 by General Electric Company, Inc. All rights reserved. The material
presented and contained herein may not be reproduced in any form or manner, without the written
permission of General Electric Company, Inc.
Important Precautions
LANGUAGE
警告 • 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
• このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでくださ
い。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者さんが、
感電や機械的又はその他の危険により負傷する可能性があります。
(KO) • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우, 번역 서비스를 제공하는 것
은 고객의 책임입니다.
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage
In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation
company will not pay a claim for damage if an inspection is not requested within this 14 day period.
Call GEHC Global Parts 1-800-548-3366 and select option 8, immediately after damage is found.
At this time be ready to supply name of carrier, delivery date, consignee name, freight or express
bill number, item damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section S of the Policy And
Procedures Bulletins.
14 July 1993
IMPORTANT...X-RAY PROTECTION
X-ray equipment, if not properly used, may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Healthcare
Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and of any other local authorities, and take adequate steps to protect against
injury.
The equipment is sold with the understanding that the General Electric Company, Healthcare
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.
Preface
Publication Conventions
Standardized conventions for representing information is a uniform way of communicating
information to a reader in a consistent manner. Conventions are used so that the reader can easily
recognize the actions or decisions that must be made. There are a number of character and
paragraph styles used in this publication to accomplish this task. Please become familiar with them
before proceeding forward.
It’s important that you read and understand hazard statements, and not just ignore them.
Section 1.0
Safety & Hazard Information
Proper product safety labeling allows a person to safely use or service a product. The format and
style for safety communications reflected in this publication represents the harmonization of IEC/
ISO 3864 and ANSI Z535 standards.
Within this publication, different paragraph and character styles are used to indicated potential
hazards. Paragraph prefixes, such as hazard, caution, danger and warning, are used to identify
important safety information. Text (Hazard) styles are applied to the paragraph contents that are
applicable to each specific safety statement.
Section 2.0
Publication Conventions
2.1 General Paragraph and Character Styles
Prefixes are used to highlight important non-safety related information. Paragraph prefixes (such
as Purpose, Example, Comment or Note) are used to identify important but non-safety related
information. Text styles are also applied to text within each paragraph modified by the specific
prefix.
The current section and its title The current chapter and its title
are always shown in the footer of are always shown in the footer of
the left (even) page. the right (odd) page.
An exclamation point in a triangle is used Paragraphs preceeded by a symbol
to indicate important information to the user. (e.g. bullets) contain information that
Paragraphs preceeded by Alphanumeric has no specific order.
characters (e.g. numbers) contain infor-
mation that must be followed in a specific order.
Headers and footers in this publication are designed to allow you to quickly identify your location.
The document part number and revision number appears in every header on every page. Odd
numbered page footers indicate the current chapter, its title and current page number. Even page
footers show the current section and its title, as well as the current page number.
2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys)
Different character styles are used to indicate actions requiring the reader to press either a hard or
soft button, switch or key. Physical hardware, such as buttons and switches, are called hard keys
because they are hard wired or mechanical in nature. A keyboard or on/off switch would be a hard
key. Software or computer generated buttons are called soft keys because they are software
generated. Software driven menu buttons are an example of such keys. Soft and hard keys are
represented differently in this publication.
Example: Hard A power switch ON/OFF or a keyboard key like ENTER is indicated by applying a character style
Keys that uses both over and under-lined bold text that is bold. This is a hard key.
Example: Soft Whereas the computer MENU button that you would click with your mouse or touch with your hand
Keys uses over and under-lined regular text. This is a soft key.
Table of Contents
Preface
Publication Conventions ...................................................................................... 15
Section 1.0
Safety & Hazard Information ........................................................................... 15
1.1 Hazard Messages............................................................................................................ 15
1.2 Text Format of Signal Words ........................................................................................... 15
1.3 Symbols and Pictorials Used ........................................................................................... 16
Section 2.0
Publication Conventions ................................................................................. 17
2.1 General Paragraph and Character Styles........................................................................ 17
2.2 Page Layout..................................................................................................................... 17
2.3 Computer Screen Output/Input Text Character Styles .................................................... 18
2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys) ........................................... 18
Section 1.0
Introduction
The goal of this document is to guide the service representative through an entire X-ray room
installation and to provide Diagnostic and Calibration information for troubleshooting and repair.
The following On-site Requirements must be accounted for prior to beginning installation of the
system.
Section 2.0
Objective and Scope of this Manual
This direction will lead you in a correct sequence through the manuals needed to install the
equipment. Following the suggested sequence will result in an orderly installation. However,
extenuating circumstances may dictate a deviation from the recommended sequence. Feel free to
deviate, but keep track of your progress by checking off the tasks in the flowchart as you complete
them.
Check equipment according to the packing list for your system against the main catalog numbers
in the FDO (Field Delivery Order). Also, verify that all the manuals are present in the binders
according to the Table of Contents.
2.3 Interconnection
Interconnection of products by furnished cables or contractor supplied wiring is indicated by
numbered MIS runs in the installation sections. Run interconnect charts are available, which
describe the connection and function of each wire.
For LING LONG XR 6000 System, refer to the following:
• LING LONG XR 6000 System Schematics (Chapter 10)
2.4 Presentation
The procedures in this steering manual are presented in jobcard forms.
Section 3.0
On-Site Requirements
3.1 EMC Requirements
If electromagnetic compliance is mandatory for the local country where this system is installed, this
equipment is allowed to be installed only in X-ray protected rooms, which provide an attenuation of
at least 12 dB for radio disturbances from 30 MHz to 1 GHz.
3.3 Documentation
Directions are necessary for correct installation of the LING LONG XR 6000 System. Procedures
from each document will be used and reflect the latest information available at this time.
Not all procedures in this manual apply to all sites. Before you begin installation, cross out the
procedures that do not apply to your site.
Section 4.0
Description of System
Standard
3
1 4
Cabinet
Operator Console WLS
Standard
Section 1.0
System Installation Guide
11 Attach Warning Labels and Rating Plates People Required: 1 Minutes to Complete: 15
Attach all the warning labels and detent marks.
If your country requires the product rating plate to be in the language of that country, follow the
instructions on document #45474336. Paste the appropriate language label over the English rating plate.
Lead is used in this system to provide required radiation shielding, such as in tube, collimator,
please pay attention and do not touch it directly.
Section 2.0
Installation Checklist
Pre-Installation Checklist
Delivery Date: Sales Person:
Customer: FDO No.: Room #
Equipment:
2.) Floor is strong enough for intended equipment and mounting methods
approved?
3.) Delivery route accommodates all intended equipment?
8.) Support structures installed for floor and wall mounted equipment?
Interconnections Completed
1.) Signal cable, power and grounding plans produced?
4.) Flexible, stranded wire provided for System input power connection?
5.) System “feeder” power cables pulled and sufficient length available at
disconnect box for connections?
General Completed
1.) Walls and floor clear of all obstructions?
Comments:
Inspection Date(s):
3.) Cable shield clamps installed and tightened (how many?, where?).
4.) All mounting bolts installed and tightened (floor and wall).
10.) List of any installation drawing deviations (all deviations should be signed by customer
and reviewed by the installation specialist).
Warning labels
Rating plates
23.) Has system documentation (paper and CDROM) been given to the customer?
24.) Has local physics inspection been completed. Some states/sites may not allow use of
system for patients until this is done.
25.) Has backup of pristine system been performed? (Patient Data Base as well as total
system backup)
Pre-Installation
Installation checklist
Section 3.0
Before You Begin
Check all the crates appearances, preventive collision labels and preventive tilting labels.
If the center of the preventive collision label is GREEN, it means the system is OK without any
collision during shippment.
If the center of the preventive collision label is RED, it means the system is not OK with collision
during shippment. Please open the crates to check system carefully.
Green Red
Figure 2-1 Preventive Collision Label
If the center of the preventive tilting label is GREEN, it means the system is OK without any tilting
during shippment.
If the center of the preventive tilting label is RED, it means the system is not OK with tilting during
shippment. Please open the crates to check system carefully.
Green Red
Figure 2-2 Preventive Tilting Label
Section 4.0
Integrated Table
4.1 Radiography Table
• Do not discard any packing materials such as envelopes, boxes, bags until all parts are
accounted for against the packing list.
• Check the mechanical conditions and external appearances of all parts for possible damages
or missing items.
• The distribution center or factory must be notified immediately of any damage or shortage of
parts.
Note: Unpacking OPEN ME FIRST box (Pink Box) in Crate #1 at first.
Take out the Documents: Pre-Installation Manual (5275600-100), Service Manual (5275601-100),
Operator Manual (5275602-100), and Quick Start (5275603-100). Read Pre-Installation Manual
and Service Manual carefully before system installation and calibration.
2.) Remove the side shipping container, and table top is bound on it. See Figure 2-5.
Side Shipping Container
with Table Top
3.) Loosen the remaining fibulaes, and remove the top and other sides of shipping containers
together. See Figure 2-6 and Figure 2-7.
Ramps
9.) Demount the two fixed screws, and remove the table front cover.
10.) Demount the four bolts which fixed the table to the crate base.
11.) Adjust caster bolts to lower the casters down the table base level. See Figure 2-16.
12.) Pull the table down the ramp and to the field.
13.) Unpack the table top, and take it into the field.
4.1.2 Installation
1.) Refer to the room layout in Pre-Installation Manual (5275600-100), confirm the location of the
table with the operator, make marks and move the table away.
2.) Drill four holes at the marks, and then clean the holes of all dust and chips.
3.) Move the table, let the mounting holes on the table align the holes in the floor, and adjust caster
bolts to upper the casters above the table base level.
9.) Take out all the cables inside the tube stand rail, and route the cables.
Note: Suggest to route the cables from the table food end.
Table Table
Foot Head
End End
10.) Demount the cabinet tray with two screws each side and pull it out.
Pull Here
Figure 2-28 Pull Out Cabinet Tray
11.) Demount generator cover with four screws and PDU cover with two screws, and put them
aside.
4.2.1 Unpacking
• Do not discard any packing materials such as envelopes, boxes, bags until all parts are
accounted for against the packing list.
• Check the mechanical conditions and external appearances of all parts for possible damages
or missing items.
• The distribution center or factory must be notified immediately of any damage or shortage of
parts.
1.) Loosen all the 6 fibulaes, and remove the top and sides of shipping containers together. See
Figure 2-30.
4.2.2 Installation
1.) Demount the rubber bumper and leveling bolt on one side of the tube stand rail.
Note: Suggest to install the tube stand from the table head end (if the cables are routed from the table
food end).
Table
Head
End
6.) Demount tube stand back cover, connect the tube stand cables (TM1-XS105 cable 5193103,
No1 cable 5197244, and Grouding cables 5193106 / 5193107), and remount the back cover.
5139106/5139107
5139103 - XP105
5139103 - XS106
8.) Press the tube arm to the appropriate position that are easy to install the tube and collimator,
and insert the carriage fixed bolt again to fix the carriage.
2.) Loosen the four screws of the upper clamp of tube fork, and remove the upper clamp of tube
fork.
To prevent the clamps from any impact. In case impact occurred, it should be verified that the
clamp has enough strength before installation.
Upper Clamps
Lower Clamps
Notice the tube installation direction. If the direction is wrong, the tube focal spot will
be inaccurate, and the tube may be falling down if the clamps are loose.
6.) Adjust the tube to align the RED indication label with 0 degree, and lock the tube.
Locks
Spacer
Collimator
1.) Demount the two stop plates and take out the collimator flange.
Stop
Plates
Collimator
Flange
Suggest to demount the two stop plates totally, but NOT only loosen them. Or the stop
plates will crash the glass window during collimator installation.
Glass Window
Drop into
the hole
Collimator Flange
Spacer
Control Handle
4.) Mount collimator onto the collimator flange using the two stop plates.
Note: In order to securely mount the collimator, please add Loctite 242 to the screws on the stop
plates.
6.) Route the cable through cable supporing bracket to the tube side.
Note: DO NOT fix the cables on supporting bracket or plate before the cable connections on tube
side.
7.) Prior to the connection of the high-voltage cables, remove the old silicone compound on the
plug surfaces with clean dry cloth and check to verify there are as cracks, flaws, or traces of
creeping discharge on the surfaces.
Note: When the traces of creeping discharge are found, be sure to replace the defective parts.
Grinding off the traces only will not be effective.
8.) Coat the tip of the cable plugs with the silicone compound about five to seven grams, that is,
to the extent the tip is hidden under the heaped up compound. At this time, the side of the cable
plug should not be greased.
9.) Remove the cap covered the socket of the tube housing and wipe the inside softly with a clean
dry cloth. DO NOT use any chlorinated solvents to clean any parts.
10.) Connect the anode and cathode HV cables (2269002-5) to the X-ray tube.
Check the polarity of HV cables carefully, and DO NOT connect the two HV cables
reversedly.
Connect the HV cables in the correct polarity with sufficient care not to produce
clearance between the cable plug flange and the cable socket at the insertion of the
plugs. Presence of such clearance or incomplete silicone compound coating will cause
creeping discharge and normal operation of tube will be disturbed.
15.) Connect the control box cables to the connecting plate, fix the plate onto control box, ty-rap
the cables and fix with a cupule, and mount the control box cover.
5193107 5197244-E
b. Install the
control box cover.
16.) Install the cable supporting plate, route and ty-rap the cables, and fix them onto the plate.
Be careful of the cable moving range, and make sure the cable will not be damaged during
tube movement.
Fix
Supporting Plate
17.) Mount the cable supporting bracket onto the back of the tube stand, put the cables onto the
supporting bracket, but DO NOT fix the cables.
DO NOT FIX
4.5 Counterweight
Adjust the counterweight by adding or removing couterweight blocks until the up/down operation
force is balanced.
Limit
Position Fixer
Cassette Tray
Side Cover
2.) Install the ion chamber and grid, connect table ion chamber (5167409) and 4 m cable
(5166493), and install the cassette tray side cover back.
Block is in the
lower place
Cable
Routing
5167409
Cassette Tray
Side Cover
Connect to JEDI
Cable
Routing
5166493
AEC1
Limit
Position Fixer
Block is in the
upper place
2.) Connect and route Door Light (pin 4, 6) and Gate Command (pin 1, 2) Cables (hospital
provide). For details, please refer to Room Power Supply in Chapter 4 of Pre-Installation
Manual (5275600-100).
Figure 2-72 Connect and Route Door Light and Gate Command Cables
3.) If wall stand ion chamber is ordered, connect and route the ion chamber 24 m cable (5166654)
and grounding cable (5193110) to JEDI.
Service Laptop
System Console
AEC1 - Table
Cable Routing
5166654/5193110
Push Here
2.) Fix the cabinet tray to the table base with two screws.
2.) Check the table top leveling. If it is not level, loosen the mounting anchors, and adjust the
leveling bolts.
Section 5.0
System Console
Route the console cables to the control room.
JEDI Generator
Communication Cable Routing
Cable (5182354) 5182354/5183903
Console Power
Cable (5183903)
System
Part No. Part Name The Other End
Console
5184620 Handswitch J1 Handswitch
5194292/5241079 RS232 Service Cable / ISP Download Cable J2 Service Laptop
5183903 Console Power Cable J3 PDU - TM3
5182354 JEDI Generator Communication Cable J4 PDU - J1
System
Part No. Part Name The Other End
Console
5184620 Handswitch J1 Handswitch
5194292/5241079 RS232 Service Cable / ISP Download Cable J2 Service Laptop
5183903 Console Power Cable J3 PDU - TM3
5182354 JEDI Generator Communication Cable J4 PDU - J1
3.) Demount console stand bottom cover with two screws. Confirm the location of the console with
the operator, make marks and move the console away.
4.) Drill two holes at the marks, and then clean the holes of all dust and chips.
5.) Move the console, and let the mounting holes on the console foot panel align the holes in the
floor.
6.) Insert the supplied two M6 anchors and washers, and hammer the anchors into the holes.
7.) Tighten the anchors.
8.) Mount the bottom cover back.
Section 6.0
Collimator Alignment
1.) Power up the system. Check LEDs status. Refer to Chapter 4 - Functional Checks - 3.1
System Power Up.
2.) Move the tube to the lowest position above the table top, open collimator light, confirm the
center of the light field projected by the collimator with an article, such as a screw.
4.) Make sure the bias of the center is less than 3 mm, and make sure the collimator beam can
align the center point of the cassette tray handle.
Bias in less
than 3 mm
Collimator
beam
Handle Center
Figure 2-86 Collimator Beam Align Cassette Tray Handle Center Point
5.) If the requirements above cannot be met, adjust the three screws on the tube fork to align the
collimator.
Section 7.0
Standard Wall Stand
7.1 Unpacking
• Do not discard any packing materials such as envelopes, boxes, bags until all parts are
accounted for against the packing list.
• Check the mechanical conditions and external appearances of all parts for possible damages
or missing items.
• The distribution center or factory must be notified immediately of any damage or shortage of
parts.
1.) Loosen all the 6 fibulaes, and remove the top and sides of shipping containers together. See
Figure 2-88.
Rear
Front
7.) Adjust caster bolts to lower the casters down the wall stand base level. See Figure 2-93. Pull
wall stand into the field.
Front
Rear
7.2 Installation
1.) Refer to the room layout in Pre-Installation Manual (5275600-100), and confirm the location of
the wall stand with the operator. Move wall stand and its template to the location, adjust caster
4.) Rotate the tube facing to the wall stand front panel.
5.) Open the collimator light, move the tube from SID 1.5 m to 1.8 m, and make sure that the bias
between the cross hair projected by the collimator and the cross hair in the wall stand front
panel is less than 3 mm in two orthogonal directions.
Note: When the collimator blades are closed to one line, it should be overlapped with the wall stand
cross hair.
Measure the rectangle’s length and width at 1.5 SID and 1.8 SID to make sure the bias is less
than 3 mm.
6.) If the requirements above cannot be met, adjust wall stand location or the four leveling bolts.
7.) Drill two holes through the two holes on wall stand base, and clean holes of all dust and chips.
8.) Insert the supplied two M10 anchors and washers, and hammer the anchors into the holes.
9.) Tighten the anchors to fix the wall stand to the floor.
Cable Routing
From JEDI
Figure 2-101 Routing Ion Chamber Cables
Cable Routing
Note: If 60 line or 78 line grid is ordered, use the wide edge of the fixers to the bottom and narrow
edge to the top.
Section 8.0
Final Cover Installation
After system configuration and calibration, ty-rap all the cables, and install the protective covers.
2.) Insert SID ruler through the cover, and install the ruler.
3.) Install tube stand side covers, and tighten the tube arm protective cover.
Note: Notice the side covers directions.
The No1 Cable (5197244) should be routed inside the right side cover.
Section 9.0
Labels and Rating Plates
Attach English/multilingual labels to overlay the English/Chinese labels attached in factory.
9.3 Cabinet
Section 10.0
Installation Checklist
Purpose: Installation checklist to confirm that all critical installation steps have been completed.
3.) Cable shield clamps installed and tightened (how many, where).
4.) All mounting bolts installed and tightened (floor and wall).
Section 11.0
Regulatory Tests and Final Steps
11.1 AEC Tests
11.1.1.1 Overview
At 50kV or more, exposures shall not exceed 600mAs per exposure.
11.1.2.1 Overview
At 50 kV or more, minimum exposures must be less than 16 msec.or less than the time to deliver 5
mAs, whichever time is greater.
11.1.2.2 Procedure
1.) Align the tube with the image receptor. Lateral Detent ON. Select 50kV, AEC (area 1 selected).
2.) Set the SID to 100 cm.
3.) Remove all phantoms.
4.) Acquire an AEC exposure.
5.) Record the exposure time:
Ion Chamber Area 1 _____________ msec
Ion Chamber Area 2 _____________ msec
Ion Chamber Area 3 _____________ msec
6.) Verify that the exposure time is less than 16 msec, or less than 5 mAs.
7.) Repeat the above procedure for each detector in the system.
Record the exposure time:
Ion Chamber Area 1 _____________ msec
Ion Chamber Area 2 _____________ msec
Ion Chamber Area 3 _____________ msec
8.) Verify that the exposure time is less than 16 msec, or less than 5 mAs.
11.1.3.1 Overview
• For any specific combination of selected technique factors, the estimated coefficient of
variation (CV) of radiation exposures shall be no greater than 0.05. HHS rejection limit is 0.045.
• Test Equipment: Radiation meter with integrate mode, 3.8 cm thickness of Aluminum.
11.1.3.2 Procedure
1.) Align the tube with the image receptor. Lateral Detent ON.
2.) SID = 100 cm.
3.) For Digital table place 3.8 cm aluminum on the table, center in the collimator light.
4.) Center radiation probe 30 cm above the table top.
5.) Collimate to a field size of 13 cm square.
6.) Select 80 kV, AEC mode and mA such that the exposure will be at least 0.1 second long.
7.) Measure the exposure in mR for each of the 10 exposures. After each exposure, recycle the
technique (switch to other values and then return them to the original values). Also vary the
time between exposures to record any filament heating variations.
Test Results:
Exposure # kV mA Time mR
1 _______ _______ _______ _______
2 _______ _______ _______ _______
3 _______ _______ _______ _______
4 _______ _______ _______ _______
5 _______ _______ _______ _______
6 _______ _______ _______ _______
7 _______ _______ _______ _______
8 _______ _______ _______ _______
9 _______ _______ _______ _______
10 _______ _______ _______ _______
Exposure # kV mA Time mR
1 _______ _______ _______ _______
2 _______ _______ _______ _______
3 _______ _______ _______ _______
4 _______ _______ _______ _______
5 _______ _______ _______ _______
6 _______ _______ _______ _______
7 _______ _______ _______ _______
8 _______ _______ _______ _______
Section 1.0
Configuration & Calibration List
Section 2.0
Configuration & Calibration Contents
2.1 X-Ray Tube Seasoning
When warming up the tube, always take the precautionary measures to protect personnel from X-
ray radiation.
When a new x-ray tube is put into operation for the first time or when it has not been in operation
for more than one month, the seasoning procedures Seasoning Schedule at installation (See
Figure 3-1) shall be carried out to assure subsequent trouble-free operation. This procedure
establishes the proper relation to its new environment with the equipment in used. The action
establishes a favorable distribution of the electrical charges and Electrostatic stresses in the
insulation system of the tube and the associated equipment.
2.1.1 Requirements
1.) Use the larger focal spot of the tube, and the seasoning will be done with the circuit specified
in technical data.
2.) The operating tube voltage and current should be monitored all exposures during this
seasoning operation.
3.) The anode rotation should be utilized at normal speed (3000 or 3600 min-1).
4.) Do not exceed either the generator or the tube ratings.
2.1.2 Procedure
Effecting exposure at near peak potential and current without prior tube warm up will
damage the tube.
To maximize tube life, perform the following tube seasoning procedure.
1.) Provide continuous input of small tube current (Approx.1 mA) and increase the tube voltage
stepwise according to the time of the Seasoning Schedule at installation (See Figure 3-1).
For any X-ray equipment without the fluoroscopy circuit, perform seasoning of short-time input
from 50 kV.
2.) Use the larger focus, rotate the anode, set the tube current to 50% of the maximum permissible
tube current at maximum tube voltage, apply short-time input once a minute with exposure
time of 0.1s, and increase the tube voltage stepwise.
When unstable, states of the tube current are observed in the course of increasing tube
voltage, reduce the tube voltage to the extent that no variation occurs, and then increase it
more slowly than previously.
3.) Stain like impact traces might appear on the X-ray tube target surface by the slight discharge
in the course of the seasoning.
The phenomena are one process to recover the withstanding voltage performance at that time,
Even if unstable states appear in the course of the seasoning, the stable operation is obtained
at the maximum tube voltage of succeeding seasoning, the tube unit can be used without any
interference to its electrical performance.
(The maximum tube testing voltage must be strictly limited to the maximum rated tube voltage
of the tube unit.)
4.) When the X-ray tube unit is used after a dormant period of a week or more, use the tube unit
after performing the Seasoning Schedule at Starting (See Figure 3-1).
5.) Every day morning, when you start to use X-ray tube, you are requested to warm-up the tube
to your necessary high voltage condition for stabilized equipment operation.
a.) At least 80kV (Step 1) is requested for all operation.
b.) Need voltage up to your necessary daily voltage (Step 2 or higher step).
c.) If discharge happened, re-warm up again under the previous condition.
Steps Voltage Current Focus Expose Time Cooling Time after Number of
(condition) [kV] [mA] [sec.] Exposure [sec.] cycle
Note 1: If you use over 120 kV, up voltage by 5 kV to your necessary voltage.
2.2.1 Supplies
None.
2.2.2 Tools
• JEDI service software (the software is a DOS application)
• Console R232 cable (5194292)
• Computer with floppy or CD drive
2.2.4 Prerequisites
Basic knowledge of PC based computer laptop, DOS and usage of Microsoft WINDOWS.
2.2.5 Personnel
1 person.
2.2.6 Time
15 minutes.
2.2.7 Procedure
The software will load the help screen information contained in Asiane .msg file.
• Wait or press space bar to jump to screen.
The Application screen will appear:
"English text Version 1.2" is displayed 5 sec.,
then display shows the last used parameters.
Note: If the serial line to the generator is not operating (e.g.: cable not correctly connected, I/F or kV
Control board down, Generator not ready or powered OFF), further screens will not be
displayed and no function keys will operate. Correct this problem before continuing.
: Status area. Displayed by Service Software, Enlighten or blinking key controlled by JEDI
Software.
Display of function keys usable with menu displayed. Commands are remembered in help area:
F3: Increase mA or mAs (3 points or 2 points) F4: Decrease mA or mAs (3 points or 2 points)
F9: Darkness correction + (w/t AEC Cell) F0 (F10): Darkness correction - (w/t AEC Cell)
MENU 1
[F1] [F3] [F5] [F7]
HEAT ANOD HIGH AEC
AUTO, AUTO,
3,4,5 3k, 9k
MENU 2
[F3] [F5] [F7]
LIST NEXP DATE
MENU 3
[F1] [F5]
TECH MODE
MENU 4
[F5] [F7]
AEC CAL
MENU 5
[F1] [F7]
RAMS Vx Px.x DSPL
[F1] [F3]
CKS MCL
2.3.1 Supplies
None.
2.3.4 Prerequisites
The service terminal must be configured and software installed (refer to 2.2 - Service Software User
Guide).
2.3.5 Personnel
1 person.
2.3.6 Time
15 minutes.
2.3.7 Procedure
2.3.7.2 Verify
According to the tube used in the installation, verify if 50 kW database loaded for tube E7843X
• The maximum mA displayed on the console must be 630 mA for 80 kV.
The word "Workstation" is incorrectly used and displayed in the JEDI software screens.
"Workstation" actually refers to the word "Receptor".
2.4.1 Supplies
None.
2.4.2 Tools
• JEDI service software (the software is a DOS application)
• Console R232 cable (5194292)
• Computer with floppy or CD drive
2.4.5 Personnel
1 person.
2.4.6 Time
15 minutes.
2.4.7 Procedure
2.4.7.1 Introduction
Receptor programming uses function keys indicated on screen for the following :
• receptor select,
• tube select,
• define nominal number of exposures for 2-point mode operation,
• AEC : types of chamber and number of pickup fields in each,
• Screen-film combination select, (if only one Select A).
<-
R : Return Q : Exit
Nominal number
Receptor Tube Tomo tech & time of exposures In Type of AEC
2-point mode
Table 3-1 Receptor Parameters (Example for RAD with Wall Bucky and two ionisation
chambers)
2.5.1 Supplies
None.
2.5.2 Tools
• Sensitometer
• Densitometer
• Set of five plexiglass plates (5 cm x 25 cm x 25 cm)
For 5 cm, it is 2 inches
OR
• A plastic bucket with a minimum height of 25 cm (for water)
• JEDI service software (the software is a DOS application)
• Console R232 cable (5194292)
• Computer with floppy or CD drive
2.5.4 Personnel
1 person.
2.5.5 Time
1 hour.
2.5.6 Prerequisites
• Working knowledge of service terminal and installed software,
• Service terminal must be configured and software installed,
• Workstation configuration done,
• AEC option must be installed.
2.5.7 PROCEDURE
2.5.7.3 Principle
The automation of this calibration is to use the optical density entered after one exposure or four
exposures to predict and propose the parameter for the next exposures. This leads to a reduction
in the number of exposures to run to find mAs, kV and reference AEC for a given optical density.
INIT
KVEF acquisition screen
No
END
INITIALIZATION
KVEF is the main compensation coefficient for film darkening : its value varies significantly with kV
The MENU4 F7 key of the laptop service enters the calibration menu.
A
Software : first kV step 80kV or next kV step (selected by Key F5-F6
(steps are 80kV, 100kV, max kV, 40kV and 60kV)
Operator : Start a serie of exposures for this kV step
Keys a & B are flickering
NO
Is V flickering ?
YES
Operator : Valid the kV step with V key when flickered
NO YES
Screen key light on?
B
Figure 3-6 KVEF Parameter Acquisition Flow Diagram (Continued)
End of kvef calibration. If first screen-film See descriptions of time and grid
calibration , select, if needed, time calibrations below
1.) Press T key and input Destination Density, then validates it with W key. Pressing T key
deselect Destination Density. Run the first exposures proposed by the software, develop the
films and measure the optical density. For each exposure, the software records the AEC signal
from the ionization chamber. The operator then selects the exposure number with F1/F2 keys,
enters the associated read optical density with key pad, and validates it with W key.
Note: There is no Beep during the exposure! The laptop Icon will light during X-ray.
2.) When the optical densities have been entered, the software calculate the parameters for the
next exposure regarding the result of the previous ones (AEC signal from ionization chamber
and optical density).
3.) The operator should run the next exposure proposed by the software, develop the film and
measure the optical density. The operator then selects the exposure number with F1/F2 keys
and enters the associated read optical density with the keypad, and validates it with the W key.
The software records the AEC signal from the ionization chamber. At each optical density
entered, the software evaluates if those values permit it to find the mAs0, which provides the
target density for this screen-film combination and the kV. When the mAs0 is found, the
software calculates the KVEF associated, then the V key starts to flicker to ask the operator to
validate the calibration by pressing the V keystroke.
4.) Select another kV and repeat the previous operation for all kV steps
Note: Don't forget to change the phantom thickness when changing kV settings!
When all kV values are calibrated, the software extend the result and calculates the equivalent mAs
(for an optical density equal to the target value) to the kV,mA plan.
At the end of the first screen-film KVEF calibration , the operator selects the KT time calibration with
the V key after the V key has been illuminated.
CHECKSUM VALIDATION
The checksum must be validated at the end of the calibration. See 2.6 - Checksum Validation.
SET KVEF
• Use the function key F1 to select sub-routine KVEF.
SET KT
• Use function key F5 to select sub-routine KG.
• Use the following keys to select the screen-film combination : A for combination 1, B for
combination 2, C for combination3. The screen-film combination selected is displayed in
reverse video.
• Select time step number with F1/F2 keys. (N = 0 T = 10ms, N = 1 T = 100ms, N = 2 T = 500,
N = 3 T = 1s, N = 4 T = 2s, N = 5 T = 3.2s).
• If necessary, select time with F3/F4 keys.
• Set KTIME to the known value with F9/F10 keys.
• Validate each new setting with the W key.
• Press the R key when the settings have been completed.
SET KG
• Use the function key F3 to select sub-routine KG.
• Use the function keys F5/F6 to scroll grid (grid number corresponds to the ionization chamber
number). Set KG to the known value using F9/F10 keys
• Validate each new setting with the W key
• Press the R key when settings have been completed.
CHECKSUM VALIDATION
The checksum must be validated after settings have been completed; refer to 2.6 - Checksum
Validation.
2.6.1 Supplies
None.
2.6.2 Tools
• JEDI service software (the software is a DOS application)
• Console R232 cable (5194292)
• Computer with floppy or CD drive
2.6.5 Personnel
1 person.
2.6.6 Time
5 minutes.
2.6.7 PROCEDURE
Resulting display:
MCLR is used to erase all setup parameters. Function key [F3] corresponding to this function
should not be used.
2.7.1 Supplies
None.
2.7.2 Tools
• JEDI service software (the software is a DOS application)
• Console R232 cable (5194292)
• Computer with floppy or CD drive
2.7.4 Prerequisites
1.) The service terminal must be configured and the software installed (refer to 2.2 - Service
Software User Guide).
2.) All calibrations must have been completed.
2.7.5 Personnel
1 person.
2.7.6 Time
15 minutes.
2.7.7 Applicability
Setup parameters must be logged after the setup procedure has been completed, or whenever
these parameters are changed. This enables service engineers to key in parameters rapidly, i.e.
when changing the KV Control Board in the event of a battery outage.
2.7.8 Procedure
MESSAGE ERROR
Error communication Error during transfer. Try a new transfer from the beginning.
Note: Ensure ASIAN.exe software running under DOS and not
under WINDOWS.
Section 1.0
Introduction
Functional Checks describes the procedures necessary to determine whether the LING LONG
XR 6000 system (Refer to Figure 4-1) is operating correctly and performing within specifications.
All the LING LONG XR 6000 system functions should conform to safety regulations, local codes
or other applicable standards.
• Operational Checks
• Performance Checks
Operational Checks may be performed by the customer or trained personnel. These checks only
indicate that the unit is functioning correctly at an operational level. They do not indicate that the
unit is
performing within specifications.
Performance Checks must be performed by trained personnel only. These checks verify that
the unit is performing to specification. It is recommended to complete these checks during and
after installation, during periodic maintenance, and when the equipment is replaced or modified.
Before performing Functional Checks all System Configuration and Calibrations must be
completed. Refer to Chapter 3 - - Configuraton & Calibration.
10 3
2
9
4
8
7 6
Figure 4-1 LING LONG XR 6000 System
Section 2.0
Tools and Test Equipment
In addition to the standard service tool kit the following tools and test equipment will be necessary
to perform the functional checks described in this chapter:
• DVM
• Metric Tape Measure
• KEITHLEY 35080A kV peak meter (or equivalent)
• RADCAL MDH2025 dosimeter (or equivalent) with 20x5-60 probe
Section 3.0
Operational Checks
3.1 System Power Up
Power up the LING LONG XR 6000 system, check that all the corresponding LEDs are displaying
correctly.
Green
Red
1 2
6 3 4 6
Section 4.0
Performance Checks
4.1 Console
1.) Check the buttons functions.
a.) Power up.
b.) Press every button except power up and off button, a beep sound will be heard.
2.) Check the APR Save function.
a.) Change any parameter of the factory default APR database.
b.) Press the APR Save button for more than 3 seconds, and the system console will
a.) Press the buttons together for more than 10 seconds. System console will
retrieve the factory default APR database. After the retrieve is done, system console will
give a beep sound to inform the operators.
b.) Power off.
c.) Power up.
d.) Check the technique parameters.
4.2.1 Requirements
The following image is gotten in a plane that is perpendicular to x-ray field. Verify below
requirements are met.
|a1| + |a2| <= 1.7% SID and |b1| + |b2| <= 1.7% SID.
b1
b2
a2
X-ray Field
|b1| + |b2|
Note: If the x-ray field is left shifted, move the flange right.
Note: If the x-ray field is right shifted, move the flange left.
13.) Re-install the collimator to the moved flange, re-test the misalignment value between x-ray
field and light field.
14.) If the requirement |a1| + |a2| <= 1.7% SID and |b1| + |b2| <= 1.7% SID has been met, make
marks on the flange interface between the tube and collimator as below.
Note: Please oriente the collimator flange by referring to the thin and thick lines.
15.) If the misalignment is not within the system requirements, please repeat step 13) to 14) until
the accuracy is within the specification and then make marks when the accuracy finally meet
the requirements.
4.3.1 Requirements
Mis-alignment of centers of x-ray field and image receptor shall not exceed 1.7% SID. |d1| <= 0.017
SID.
d1
After replace the tube or JEDI or some parts in JEDI, please do the tests below from 4.5 to 4.8.
4.5.1 Requirements
Range: 40kVp ~ 150kVp;
Minimum adjusting step: 1kV;
Page 158 Section 4.0 Performance Checks
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
Accuracy: less than or equial to +/-(3%+2kV)
4.5.2 Procedure
1.) System power up, select radiography mA at 200mA to 630mA, set 100ms, Large focal spot,
select exposure kV as following table.
2.) Take exposure with large spot and record the data of console display. The variance should be
less than or equal to +/-(3%+2kV).
40
80
150
4.6.1 Requirements
Range: 10mA~630mA separate adjustment, 10.0, 12.5, 16.0, 20.0, 25.0, 32.0, 40.0, 50.0, 63.0,
80.0, 100.0, 125.0, 160.0, 200.0, 250.0, 320.0, 400.0, 500.0, 630.0;
Large Spot: 10~630mA;
Small Spot: 10~250mA;
Accuracy: less than or equal to +/-15%.
4.6.2 Procedure
1.) System power up, select three parameter mode (mA,kV,time) on system console. Manually
adjust exposure mA as following table, exposure at 150kV from 10 to 400mA, 80kV for 630mA,
set 100ms exposure time.
2.) Take exposure by handle switch with Large Spot, record the data of console display.
Note: The mA used shall be minimum as allowed and mSec > 20.
10 mA
200 mA
400 mA N/A
630 mA N/A
4.7.1 Requirements
Range: 1.0ms~6300ms;
Separate adjustment: 1.0, 1.25, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10.0, 12.5, 16.0, 20.0, 25.0,
32.0, 40.0, 50.0, 63.0, 80.0, 100.0, 125.0, 160.0, 200.0, 250.0, 320.0, 400.0, 500.0, 630.0, 800,
1000, 1250, 1600, 2000, 2500, 3200, 4000, 5000, 6300;
4.7.2 Procedure
System power up, select radiography exposure parameter 80kV, 100mA. Take exposure and
record the data of console display. The variance should be less than or equal to +/-(10%+1ms).
4.8.1 Requirements
Range: 0.5mAs~630mAs;
Separate adjustment: 0.5, 0.63, 0.8, 1.0, 1.25, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10.0, 12.5, 16.0,
20.0, 25.0, 32.0, 40.0, 50.0, 63.0, 80.0, 100.0, 125.0, 160.0, 200.0, 250.0, 320.0, 400.0, 500.0, 630;
Accuracy: less than or equal to +/-(10%+0.2mAs).
4.8.2 Procedure
1.) System power up, select two-parameter exposure mode, and select the exposure mAs as
following table at 80kV.
2.) Take exposure and record the data of console display. The variance should be less than or
equal to +/-(10%+0.2mAs).
10
40
63
Warm up or season the tube before exposures of high kV or mA can be taken appropriately.
Section 1.0
Theory
1.1 Glossary
Glossary of terms used in this document:
Term Definition
CPU Control Processor Unit. Microprocessor and peripherals which run the
software/firmware
EPLD Erasable Programmable Logic Device.
EMC Electro Magnetic Compatibility. The EMC function prevents the generator
from polluting the power source.
FPGA Field Programmable Gate Array. It is programmed by the CPU core after
the reset and handles all the exposure control logic including the system
interface real-time lines.
HV RIPPLE High voltage variations due to inverter current pulses. Typically a few
percent.
STATE Software or hardware function which handles the state of a system and
MACHINE authorize to go to the next state upon reception of specific events.
Llr HV power inverter serial resonant current; current in the serial inductor.
2.3 Applications
1 Point Mode X X X
0 Point Mode X X
AEC X X X
Tomography X X
AET X X
Continuous/Pulsed X X X
Fluoroscopy
Rad Exposures X
Cinema 30 fr/s X
Cinema 90 fr/s X
ABC X X
Variable mA Scans X
Low mA Fluoroscopy X
Legend:
• AEC - Automatic Exposure Control
• AET - Automatic Tomographic Exposure
• ABC - Automatic Brightness Control in fluoroscopy
2.4 Architecture
Refer to Illustration 19: Jedi Generator / Functional Architecture
The Jedi family is composed of 3 elements:
2.4.1 A Kernel
• High voltage chain composed of kV control, HV power inverter and HV tank
• Anode rotation function
• Tube filaments heater function
• Control bus for communication between the functions
• DC bus for power distribution to each function
• Input voltage to DC conversion: AC/DC function
• Low voltage power supply
• Application software, running on the kV control board
These functions are the Jedi core. They are present in all versions of the generator.
A function can be unique for all products, or can have several different releases based on product
specification.
Example: The anode rotation function is available in 2 releases:
• low speed rotation for applications where the tube has a max rotation of 3000 rpm
• high speed/low speed rotation for applications where at least one of the tubes can use 8000-
10000 rpm
The control bus is unique.
Inverter
Dual Snub LVPS Board
(for TIGER) AC/DC
Tube Board
HV
Gate
Board
+-
CT
Inter KV Control
face Board EMC Filter
(Optional)
Heater Board
System Interface
Rotation Box for the rad
Board product
Tube Grid/Bias
Selection Control
Rotation Rotation
Capacitor Capacitor
(optional)
Auxiliaries Optional
Box Boxes
Section 1.0
Exception Handling
JEDI software performs auto-test at power up and continuously monitors the correct operation of its
functions during application. Any malfunction is stored in the JEDI error log. Errors found can only
be reported if the generator is powered on and functional.
1.1 Diagnostics
There are different levels of diagnostics.
Section 2.0
Error Codes, Diagnostics & Troubleshooting
2.1 Introduction
This diagnostics section is to introduce the JEDI generator with the following information:
• Power-On diagnostics LED indication.
• List all the potential error codes that can be issued by JEDI Generator.
• Provide error code explanation, potential cause and recommended action.
• List of diagnostics aids and explanation of diagnostics.
kV control board
S0..................S7
_ ~ ~ o~o~o~o~
S0.........................S7
o
The 8 LED's (S0...S7) are lit successively in both directions (Scrolling as indicated by the
arrows): the power up diagnostics are completed, kV control board is up and running.
S0.....................S7
o o o o
One out of two is lit: Data base checksum problem. An error code is logged. Refer to error code
description.
S0.....................S7
o oo
S7, S6, S5 are lit along with either S0 or S1 or S2 (depending of the type of FPGA downOne
out of two is lit: Data base checksum problem. An error code is logged. Refer to error code
description.
• When an application error occurs (Not PRD)
K K e.g: No communication
The simplified error code is displayed on the Leds. They blink; when the error is cleared (by a
return to the standby mode for example), the 8 LEDs are lit successively.
Heater board LED status: (See central listing)
After the power on diagnostics, heater board LEDs DS1 and DS2 are lit successively. Any
different status correspond to an abnormal situation. An error code is logged. Refer to error
code description.
Rotation board LED status: (See central listing)
After the power on diagnostics, rotation board LED DS5 is blinking. Any different status
correspond to an abnormal situation. An error code is logged. Refer to error code description.
2.5 Diagnostics
This chapter describes diagnostics tools based on error codes and specific diagnostics.
Before any manual intervention, ensure the main power is off. Apply lock out-tag out procedure for
your own safety when manipulating inside the equipment is required.
*
LEDs Neon's
*
DC bus, 400 ... 800 V
DS 1
DS 3
DIRECTION 5275601-100, REVISION 2
+ M -
*
F1
INVERTER. & LLC. Circuit DS 1 - 15V DC bus, 160V
+ 15V
(120 to 200V)
Gate Command Board
DS 300
*
LVPS
*
I/F AC/DC NE 1 DS 2
Console PUPT-
or
DS7-6
DS 1, 2
- 15 V
-15V, +15V + - DS 3, 2, 1
AEC + 15 V Rotor +5, -15, +15
kV ctrl Board
- 15 V "VCC": +5V DS 7
*
N1 (230 V)
* + 15 V
EMC Flt.
3ph, 380 - 480 VAC
Figure 6-1 JEDI Visual Power Supply Distribution for LING LONG XR 6000
PDU
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
2.6.1 Errors
Section 3.0
Console Troubleshooting & Error Codes
Adjust KV
No Connector of J4 on Yes
KV can be changed?
console is loosing?
Yes
No
8QSOXJ+6FRQQHFWRU
Press the button.
No
Is LED (DS4) lighting or Yes Jedi
Replace Power up 15s
flashing on TAV board? troubleshooting
console after power off
No
Replace TAV
board
Yes
Yes
Operation each button.
Check and record which
button is invalid.
Yes
Replace console
main board
No
Power off console, then
power on it after 15 sec
No
No
Replace console
main board
Section 4.0
PDU Troubleshooting
4.1 PDU Troubleshooting
Y
N
N
Replace Console Y Check DSP1 output is
or Console Cable 24 V or not
Replace F1
Replace DSP1
No EM lock power
Check LED1 is light or N Close BK or
or
not hostipital BK
No collimator power
Y
N
Replace EM lock N
Y Check DSP2 output is
or
24 V or not
Replace Collimator
Check CT2 is close or Y
Replace F2
N not
Replace DSP2
N
Check K1 is close or Y
Replace CT2
not
N
Y
Y
Y
Replace K1
Section 2.0
Preventive Maintenance
The following pages list the preventive maintenance detailed activities.
2.1.1 Supplies
Cleaning liquids and compounds.
Touch-up paint.
2.1.2 Tools
None.
2.1.5 Personnel
1 person.
2.1.6 Time
15 minutes.
2.1.7 Procedure
1.) Remove power from the system.
2.) Carefully clean all surfaces of the system. Note any paint damage as you clean the system.
3.) Apply touch-up paint to any paint-damage areas.
4.) Reapply power to the system.
2.2.1 Supplies
None.
2.2.2 Tools
None.
2.2.4 Prerequisites
None.
2.2.5 Personnel
1 person.
2.2.6 Time
15 minutes.
2.2.7 Procedure
1.) Remove power from the system.
2.) Visually check around the moving parts of the system for any debris, wear or damage. This
includes tabletop, tube stand, XRT mount, collimator moving parts and the wall stand if
present.
3.) If any debris, wear or damage is observed, replace or repair the affected areas.
4.) Reapply power to the system.
2.3.1 Supplies
None.
2.3.2 Tools
None.
2.3.4 Prerequisites
None.
2.3.5 Personnel
1 person.
2.3.6 Time
45 minutes.
2.3.7 Procedure
Follow the functional checks specified in Chapter 4 - Functional Checks of this Service Manual.
2.4.1 Supplies
None.
2.4.2 Tools
Measuring tape.
2.4.4 Prerequisites
None.
2.4.5 Personnel
1 person.
2.4.6 Time
45 minutes.
2.5.1 Supplies
Grease.
Light Machine Oil.
2.5.2 Tools
None.
2.5.4 Prerequisites
None.
2.5.6 Time
45 minutes.
2.5.7 Procedure
1.) Surface finish.
Remove the dust from the table top and table topside rails.
2.) Bearings (longitudinal and transversal).
Remove the dust.
3.) Bearing tracks and rails (longitudinal and transversal).
Remove the dust from the bearing tracks and rails. Add the light machine oil on the tracks and
rails.
4.) Rubber bumper end stops (longitudinal and transversal).
Verify the bumper stops have no cracks and one fixed tightly.
2.6.1 Supplies
Grease.
Light Machine Oil.
2.6.2 Tools
None.
2.6.4 Prerequisites
None.
2.6.5 Personnel
1 person.
2.6.6 Time
45 minutes.
2.6.7 Procedure
1.) Electrical connector and cable
Confirm properly connected.
2.) Cassette handle assembly, push button and hand holder
Check the push button function. Check the movement with the handle functions.
3.) Bearings and tracks (front and rear)
2.7.1 Supplies
Grease.
Light Machine Oil.
2.7.2 Tools
None.
2.7.4 Prerequisites
None.
2.7.5 Personnel
1 person.
2.7.6 Time
45 minutes.
2.7.7 Procedure
1.) Counterweight chain (Tube stand)
Check the chain is fixed tightly.
2.) Carriage roller bearing inspection
Corresponding movement operates smoothly.
3.) Vertical lock
Confirm proper function.
2.8.1 Supplies
Film.
Error-code recording form.
X-ray counter recording form.
2.8.4 Prerequisites
None.
2.8.5 Personnel
1 person.
2.8.6 Time
60 minutes.
2.8.7 Procedure
mA accuracy
Run the exposures in table below and check if the mA value on post display (Z key) is between
the min and max value.
If not , run the same exposure 10 times with exposure time set to 12.5ms and then check again.
2.9.1 Supplies
None.
2.9.2 Tools
None.
2.9.4 Prerequisites
None.
2.9.5 Personnel
1 person.
2.9.6 Time
30 minutes.
2.9.7 Procedure
Follow the process specified from Step 7 to Step 11 in Chapter 2 - Physical Installation - 4.3 X-Ray
Tube of this Service Manual.
2.10.1 Supplies
None.
2.10.2 Tools
5.5 nut driver or socket wrench with ratchet.
2.10.4 Prerequisites
Upload the database that is stored in the RAM before starting the replacement procedure
Refer to Chapter 3 - Configuraton & Calibration - 2.7 Backup and Restore Database Parameters.
2.10.5 Personnel
1 person.
2.10.7 Procedure
This NV RAM contains a LITHIUM CARBON MonoFlouride (LiCFx) battery, used to provide the
battery backup for the static RAM and the time keeper device.Electrical Characteristics: 3 V - 49
mAh type BR 1225.Resend the former component to your local GEMS representative for recycling.
Section 1.0
Renewal Parts List
Note: Refer to Section 2.0 for information regarding JEDI Generator Renewal Parts.
Item Part No. Description Quantity/
unit
1 5183178 Pioneer console assembly 1
2 5199452 Pioneer console E-board 1
3 5184620 Handswitch 1
4 5182285 TAV board 1
5 5195772 Relay DC24V 1
6 5195778 Power supply 150W 1
7 5195779 Power supply 320W 1
8 5181727 Long electromagnetism brake assembly C (tube stand 1
movement)
9 5182277 Line wheel 2
10 5195615 Circle electrical bracket A (up/down) 1
11 5191715 Tube stand top cover (50kW) 1
12 5181910 Pioneer tube support assembly (without clamps) 1
13 5181374 Tube fork clamps 1
14 5184986 Rotundity electricity brake assembly B (tube rotate) 1
15 5180983 Pioneer tube cover assembly (mechanical parts) 1
16 5215489 Table top 1
17 5184226 Long electricity brake assembly B (table top lognitudinal) 2
18 5184407 Table front cover 1
19 5184488 Long electricity brake assembly A (table top transverse) 3
20 5184750 Lock handle (table cassette tray) 1
21 5184895 Foot Pedal Tweak spring 2
22 5196542 Pioneer WS cassette tray spring kit (spring 5193827, twist 2
spring 5193856)
23 5184245 Wall Stand bottom cover 1
24 5184272 Wall Stand top cover 1
25 5195279 Switch VS10N051C2 1
26 5193099 MC3 MC4 cable 1
27 5193102 MC8 cable 1
28 5195277 Key switch 5
29 5195278 Foot pedal switch 1
30 5196543 Bearing kit 1 GB/T276 6001-2Z(021-0141) + 5181681 6
31 5196544 Bearing kit 2 GB/T276 6001-2Z (021-0141) + 5182435 4
32 021-0103 Bearing GB/T276 6200-2Z 12
33 021-0129 Bearing GB/T276 6203-2Z 2
34 021-0102 Bearing GB/T276 6000-2Z 19
35 021-0122 Bearing GB/T276 609-2Z 3
36 5184698 Bearing outer circle A 4
37 5184364 Nylon outer circle B 12
38 5183903 Pioneer console power cable 1
39 5182354 Pioneer generator console cable 1
40 5212348 Table 78L 10:1 100 cm grid 1
41 2259298-75 Table 60L 10:1 100 cm grid 1
42 5192454 Tube (50kW): Toshiba E7843X 1
43 5213401 Wall stand with 78L 10:1 180 cm grid 1
Item 39
Item 38
Item 3
Page 222
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
Item 4
Item 7
Item 53
Item 49 Item 48
Item 47
Item 6
Item 50
(F1~F5)
Item 54
Item 52
Item 11
Item 51
Item 31
Item 33 Item 33
Item 31
Item 30
Page 224
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
Item 9
Item 10
Item 34
Item 46
Item 11
Item 42
Item 45
Item 16
Item 18
Item 15
Item 28
Item 13
Item 14
Item 12
Two
Four Totally
Four Totally
Item 32
Page 226
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
Item 26
Item 27
Item 17
Item 34, 37
Item 55
Item 19
Item 19
Item 35
Item 19
Item 25
Page 228
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
Item 56
Item 57
Item 28
Item 20
Item 24
Item 34
Item 23
Page 230
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
3.) Demount the back cover with four screws (two in the back cover, and two in side covers).
Another Screw
5.) Remount back cover, rotate collimator back, and tighten the fix knob.
Page 232
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
1.3 Rotundity Electricity Brake Assembly (5184986) Replacement
Note: About disassembly/reassembly of the tube assembly, please refer to Chapter 2 - Physical
Installation - Section 4.0 Integrated Table - 4.3 X-Ray Tube ~ 4.4 Control Box and Collimator.
1.) Power off.
2.) Lower the tube assembly to the location which is convenient for operation.
3.) Disconnect all the cables of X-ray tube, collimator and control box.
4.) Rotate the tube to 180 degree, and let the X-ray window upturned.
5.) Uninstall the collimator.
6.) Uninstall the control box.
7.) Rotate the tube back.
8.) Uninstall the two upper clamps of the tube fork.
9.) Remove the tube, and put it away carefully.
10.) Loosen the three fix screws on the tube arm, and demount the tube fork.
11.) Demount the fix screw.
13.) Demount the cable clamp, disconnect the cable, and take out the rotundity electricity brake
assembly.
Page 234
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
14.) Replace the brake, seal the cable, and fix it with cable clamp.
15.) Refer to the above steps to re-install the parts.
Note: Please refer to Chapter 2 - Physical Installation - Section 6.0 Collimator Alignment to do alignment.
3.) Demount the handswitch socket off from the handswitch mounting plate.
4.) If console stand is ordered, please demount console off from the console stand. If not, please
skip this step.
Page 236
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
5.) Install the handswitch socket which is demounted before to the mounting plate on the new
console.
6.) Connect the related cables.
7.) If console stand is ordered, mount the new console onto the console stand. If not, please skip
this step.
8.) Power up.
9.) Switch to configuration mode and set the site’s system configuration.
- 32kW or 50kW
- WS AEC or not
- Table AEC or not
- Have WS or not
- Dose display or not
10.) Retrieve factory default APR database.
11.) Exit configuration mode, and reset system.
Page 238
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
Pull Here
Figure 8-5 Pull Out Cabinet Tray
Push Here
Section 2.0
JEDI Generator Renewal Parts
2.1 Introduction
This Section provides information for the identification of Renewal Parts for the JEDI 50 R 1T
generator.
FRU name
Inverter PW Assembly
2187001 Part number for
• Inverter Capa Assembly ordering
2203900
JEDI 50 R 1T
2212259-2
Module Auxiliaries 50 R 1T
2215808
System cable
2214511
Module Power 50 R 1T
2229597
CAN Cable
2221983
Equipped AC/DC 50 R 1T
2221722
Service Tools
2228546
Interface Module 50 R 1T
2205629-2
Power Unit
Auxiliary Unit
TOP VIEW
Interface
Unit
AC/DC Unit
Jedi Standard
HV Tank
REAR VIEW
Figure 8-7 JEDI 50 R 1T Generator Location of Main Assemblies
Miscellaneous Items:
First aid spares: Services tools
1 short circuit cable Service software, 2227656
1 fuse 15 A 700 VDC RS232 Cable, 2207478
1 fuse 10 A 250 V
10 M4-8 mm screws
10 tie-wraps
AC/DC Unit
Interface Unit
Auxiliary Unit
Figure 8-8 JEDI 50 R 1T AC/DC Unit, Interface Unit & Auxiliary Unit
A
Non-Volatile parameters
RAM
2159419
FRU: 2
Job Card PM JEDI 15-2
Programmed kV Control
Board
2214606-2
FRU: 1
Job Card D/R JEDI 17-26
DC_BUS N to INVERTER
DC_BUS P to INVERTER
DC_BUS to Auxiliary
8 Unit (Rotation Board)
DC_BUS to
Power Unit
J1 J3
Not Used 9pts (Gate Command
J4 AEC Board DC_BUS+ Board)
DC_BUS- J1J2J3
PH3PH2PH1 DC_BUS M
Not Used 9
pts
J7 16
pts J5
Not Used PH1
9
pts
J5
Room Door PH2
Interface 9
pts J6 J2&TP2J1&TP1
W22 9pts 9pts PH3
9
pts
J8
EMC Board AC/DC Board
Service
Laptop
25
pts
J9
Console
I/F Board
1
64
J10 J3 pts
16
pts
2
AC/DC Unit
9
AEC1 pts
3 9
AEC2 pts THREE PHASES
Interface Unit + Ground from PDU
From
KV Control
board
Keyed
Connections
DC_BUS N
DC_BUS P 70 o C
DC_BUS to ACDC Unit J1
to Tube
DC_BUS to ACDC Unit J3
to Tube Rotor
NP
anode cathode 1 1 1
11 9 6 5 4 3 1
CAN 7
5 6 26pts
J1 CF1 CF2
Keyed
Connections
Section 3.0
JEDI Generator Disassembly/Reassembly
3.1 Introduction
This section provides information to dismount and remount FRU (Field Replaceable Unit).
3.2 Tools
Serviceability of Jedi Generator requires the following tools:
F I
C
E
a
D G
B H
Digital Multimeter (A)
Phillips Screwdriver (B)
Allen Key (C)
Screwdriver With Interchangeable Socket Head (D)
Long Extension for Ratchet Wrench (E)
Short Extension for Ratchet Wrench (F)
Cutting Pliers (G)
High Voltage Wrench (H)
Ratchet Wrench (I)
Socket Wrench (J)
Hexagonal Head Socket Set (K)
3.3.1 Supplies
None
3.3.2 Tools
• 5.5 mm nut driver or socket wrench with ratchet.
• 3 mm Allen wrench.
• 13 mm socket with ratchet.
• C-clamp for HV connections.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.3.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.3.5 Personnel
1 person.
3.3.6 Time
1 hour 30 minutes.
3.3.7 Disassembly
1.) To release the Power Unit, follow the instructions of Disassembly / Reassembly procedure of
3.4 Inverter Mid-Power Assembly.
Observe the Caution relating to dangerous voltages, and the Warning relating to the weight of
the generator!
Do not remove the two EMC panels from the top of the Power Unit.
3.) Disconnect the short ribbon cable 2 connecting J2 on the Gate Command board (Inverter
Assembly) to J1 on the KV Measure Board (HV Tank).
4.) Disconnect the wider short ribbon cable 3 connecting J2 on the KV Control Board to J2 on
the KV Measure board (HV Tank).
5.) Using a 13 mm socket wrench, remove the three nuts 4 attaching the primary coil flat cables
Take great care not to let screws fall into the HV connector receptacles. Cover them with
plastic caps.
6
7.) Remove the 5.5 mm screws: three at the front side 6 (two at right, one at left) and three at the
right side 7 .
8.) At the rear side, remove the eight 3 mm Allen screws 8 which secure the heat sink of the
Inverter Mid Power Assembly to the HV Tank.
9.) Take off the assembly constitued by five parts: EMC covers, Coils Inverter Assy, Capa Inverter
Assy, Mid Power Assy and KV Control Board. The HV Tank is now free 9 .
3.3.8 Reassembly
Reverse the disassembly procedure to reassemble and replace the unit.
1.) Mount the flat cable J1 on the kV Measure Board of the new HV Tank (refer to (refer to
Disassembly section 2 ).
2.) Reassemble the HV Tank with the other elements of the Power Unit (refer to Disassembly
section 9 ).
3.) Mount the eight 3 mm Allen screws to secure the heat sink of the Inverter Mid Power Assembly
6.) Using the 13 mm socket wrench, connect the primary coil flat cables to the three capacitors 1
The two flat cables with white bands on them go both together on the middle capacitor
with the parallel resonant cable (refer to section Reassembly 2 ).
7.) Reconnect the cable to J3 on the KV Measure Board (refer to section Disassembly 1 ).
Tighten the nuts to 8 N.m (80 kg.cm).
8.) Reconnect the wider short ribbon cable from J2 on the kV Control board to J2 on the kV
Measure Board (refer to section Disassembly 3 ).
9.) Mount the two 5.5 hex head screws securing the cover to the HV Tank on the HV outlet side
(refer to section Disassembly 5 ) and the three screws on the other side (refer to section
Disassembly 7 ).
Tighten the screws between 1.5 to 2 N.m (15 to 20 kg.cm).
10.) Mount the Power Unit on the support plate.
3.3.9 Verify
Connect the service laptop (refer to Chapter 3 - Configuraton & Calibration - 2.2 Service Software
User Guide) and proceed to "HV no load diagnostics" to ensure correct operation.
Skip steps "Gate Command and Inverter in Short Circuit" diagnostics.
3.4.1 Supplies
None
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.THESE LEDS INDICATE THE PRESENCE OF DANGEROUS
VOLTAGE ON THE POWER SUPPLY CAPACITORS.
Look through the front of the power unit; observe leds ds 1 at the top of the dual snub board
and ds 300 at the right hand end of the gate command board. if either is lit, wait until both
are off before proceeding further.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.4.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.4.5 Personnel
1 person.
3.4.6 Time
2 hours.
3
1
1
3
4
4 2 2
TOP VIEW
FRONT VIEW
1 5 1 AC/DC UNIT
2 POWER UNIT
4
3 INTERFACE UNIT
4 AUXILIARY UNIT
2
3 5 SUPPORT PLATE
TABLE
TABLE BUCKY
BUCKY
WALL
WALL BUCKY
BUCKY
DOOR
DOOR
LAPTOP
LAP TOP
CONSOLE
CONSOLE
AEC2
AEC2 AEC1
AEC1
1.) Disconnect the connectors J2 and J3 and ground cable on the top of the generator 1 .
2.) Remove four 5.5 mm hex head securing screws holding the Auxiliary Unit on the support plate
2 .
3.) Disconnect the flat ribbon cable 3 , without tearing, between the Auxiliary Unit and the
Interface Unit.
4.) Remove the Auxiliary Unit.
2
1
3
1.) The Interface Unit appears after removing the Auxiliary Unit. Release the flat ribbon
cable 1 ,making it sliding along the cover panel 2 of the Interface Unit.
2.) Unscrew twelve 5.5 mm screws that retain the cover panel and remove this one 2 .
3.) On the I/F Rad Board, disconnect the flat wide ribbon cable J10 3 linked to the KV Control
Board.
THE POWER UNIT WEIGHTS 30 KILOGRAMS (66 POUNDS). DO NOT TRY TO LIFT IT
WITHOUT ASSISTANCE.
4.) Unscrew twelve 5.5 mm screws securing the Interface Unit box to the Power box.
5.) Taking care of the flat wide ribbon cable 3 ,remove the Interface Unit box. The components
of the Power Unit appear: coils assembly, capacitors assembly, KV Control Board.
1 2
5 4
1.) On the top of the generator, the two connectors 1 are fixed mecanically on the support
plate. Unscrew connectors screws in order to release the 4 power cables 2 coming from
the Power Unit and the AC/DC Unit.
2.) Remove the screw securing the ground cable on the AC/DC Unit. Disconnect the cable 3
3.) Remove the four screws 5 on the support plate in order to release the Power Unit. This one
is then free.
1.) Remove the 5.5 mm hex head screws which secure the two EMC cover panels 1 at each
side of the Power Unit; remove the panels.
2.) With a 10 mm socket wrench with ratchet 2 ,remove the three hex head screws securing the
cables Q100C 3 , Q100E and Q200C on the Inverter Mid Power Assembly.
7
A
6
3.) Disconnect flat ribbon cables from J1 and J3 on the Gate Command Board 4 and
disconnnect J4 at the right hand side of the board.
4.) With 10 mm wrench, disconnect the parallel Coil connection from the Current Transformer
pass through 5 and with 13 mm wrench the other connection from the Capacitor
Assembly 6 .
5.) Remove the eight 3 mm Allen screws 7 which secure the Inverter Mid Power Assembly to
the HV Tank.
3.4.8 Reassembly
1.) Present the new Inverter assembly. Put back the eight 3 mm Allen screws to secure the
Inverter Mid Power assembly to the HV Tank (refer to section 3.4 Inverter Mid-Power
Assembly 7 ).
Tighten them firmly to torque 2 N.m (20 kg.cm).
2.) Remount all cables that were removed:
a.) Remount the three Coil cables on the Inverter IGBT's with three 5.5 mm screws.
Tighten them firmly to torque 3 N.m (30 kg.cm).
d.) With the 10 mm screw and flat and lock washer, reconnect the shorter cable of the parallel
coil on the Inverter assembly onto the right hand side current transformer (Ti1/100) pass
through. A .
Tighten to 3 N.m (30 kg.cm) torque.
e.) With the 13 mm screw, with flat and lock washer, reconnect the longer cable of the coil
on the central capacitor (C3UF) of the Capacitor assembly, together with the two primary
Parallel Coil
Q100C Q200C
C C
Q100E
Inverter Assembly
E E
Coil Assembly
+ -
+ DC Ti1/40 Ti1/100 - DC
Capa Assembly
C3UF
HV Tank
4.) Remount the two EMC panels on the top of Power Unit with the 5.5 hex head screws.
Tighten all the cover screws to torque between 1.5 to 2 N.m (15 to 20 kg.cm)
5.) Assemble the Power Unit with the AC/DC Unit on the support plate with 5.5 mm hex head
screws. Put back in place the disconnected cables: ground, DC bus (refer to section 3.4
Inverter Mid-Power Assembly).
6.) Remount the Interface Unit on the support plate then the Auxiliary Unit. Follow the inverse
procedure indicated in the sections and .
7.) Remount the generator inside the table.
3.4.9 Verify
Run HV LOAD DIAGNOSTIC - "Inverter in Short Circuit Diagnostics" and "Inverter Gate Command
Diagnostic" are not pre-requested.
Verify correct operation on the whole application range.
3.5.1 Supplies
None
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.5.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.5.5 Personnel
1 person.
3.5.6 Time
15 minutes.
3.5.7 Disassembly
The I/F Board Rad is situed in the Interface Unit. To reach this Unit, the Auxiliary Unit has to be
released from the generator. Follow the instructions of 3.4 Inverter Mid-Power Assembly.
1 4
1.) Disconnect the flat ribbon cable 1 and the two cables 2 , if existing, coming from the AEC
Board.
2.) Disconnect the flat wide ribbon cable J10 coming from the KV Control Board 3 .
4.) Remove the ten 5.5 mm screws 5 securing the Interface Board Rad; remove the board.
3.5.8 Reassembly
1.) Reverse the disassembly procedure to reassemble the unit. Tighten screws to torque between
1.5 and 2 N.m (15 to 20 kg.cm).
2.) Power the system on and verify that the system functions properly.
3.6.1 Supplies
None
3.6.2 Tools
• 5.5 mm nut driver or socket wrench with ratchet.
• 10 mm Allen wrench.
• 13 mm socket with ratchet.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.6.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
The KV Control board contains a saved memory Database with essentially the following
parameters:
• Tube configuration parameters, AEC calibration parameters, and Room configuration
(workstation, AEC, screen film pairs).
• Filament aging variable
• Error Log
• Recorded parameter data
If it is possible, backup database first. Else, the diagram hereafter helps you to determine the action
to take for your situation.
Save Non
Configura*
tion
Calibration Volatile RAM
KV Control
Out of Order
Database
DB Is
Upload
Database Yes Upload No original DB No
Possible? saved?
(*)
Yes
Replace Replace Replace
KV Control KV Control KV Control
SW Download SW Yes SW
No Yes OK?
OK? Software OK?
Yes No No
Download Download
Software Software
Calibrate if
applicable
Reset
Database:
* Date
* Time
* Clear Error
log
Operational
Checks
Save Non
Volatile RAM
Database
3.6.5 Personnel
1 person.
3.6.7 Disassembly
1.) Disconnect the cables from J1, J2 and J3 on the kV Control Board 1 .
2.) Remove the seven 5.5 mm hex head screws which secure the kV Control Board; remove the
board 2 .
3.6.8 Reassembly
Reverse the disassembly procedure to reinstall the KV Control board and the Interface Unit.
Tighten 5.5 mm screws to torque between 1.5 and 2 N.m (15 to 20 kg.cm).
3.6.9 Verify
1.) Connect the Service Laptop (refer to Chapter 3 - Configuraton & Calibration - 2.2 Service
Software User Guide).
2.) Reapply power and verify that the 8 LEDs on KV Control board are scrolling.
3.) Verify software version (service floppy, menu 5)
- If it corresponds to the software version on the floppy delivered with the equipment (or
upgraded by FMI), proceed to Database Restore (refer to Chapter 3 - Configuraton &
Calibration - 2.7 Backup and Restore Database Parameters).
- If not, download the software and proceed to Chapter 9 - Service Tool Job Cards - Section
1.0 Software Download / Upgrade.
4.) Restore Database (refer to Chapter 3 - Configuraton & Calibration - 2.7 Backup and Restore
Database Parameters).
Chapter 8 - Renewal Parts Page 265
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
- Up-to-date database if available
- Most recent or originally saved Data Base if available
- Default one from the floppy delivered with the equipment in the Service Tool set.
Alternately, you may try to swap the NVRAM chip (U32) from the former to the new KV Control
board to keep all the parameters. Refer to Chapter 7 - Planned Maintenance - 2.10 Replace
Generator NonVolatile Parameters RAM“Replace Non Volatile Parameters RAM” JobCard. If this
chip is not involved in the failure or the cause for replacement, it will allow to keep up-to-date data.
5.) Calibrate:
- Perform several exposures on both focus in several workstation, with exposure time
longer than 12.5 ms.
If default database was restored, it is necessary to resume room set up and AEC calibration (if
present) (refer to Chapter 3 - Configuraton & Calibration - 2.5 Automatic AEC Calibration). Then
perform the checksum (refer to Chapter 3 - Configuraton & Calibration - 2.6 Checksum Validation).
6.) Save Database on the service floppy delivered with the equipment (refer to Chapter 3 -
Configuraton & Calibration - 2.7 Backup and Restore Database Parameters).
3.7.1 Supplies
None.
3.7.2 Tools
5.5 mm nut driver or socket wrench with ratchet.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.7.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHTS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.7.5 Personnel
1 person.
3.7.6 Time
45 minutes.
4.) Remove the four 5.5 mm screws securing the AC/DC Unit to the support plate 4 . The AC/
1.) Remove ten 5.5 mm hex head screws which retain the cover. Remove that one.
2.) Remove the three nuts 1 securing the phase cables and three screws 2 retaining the EMC
board to the rectifier block.
3.) Unscrew nine 5.5 mm hex head screws 3 and remove the EMC Board 4 .
3.7.8 Reassembly
Reverse the disassembly procedure to reinstall the EMC board.
Tighten 5.5 mm screws to torque between 1.5 and 2 N.m (15 to 20 kg.cm).
Ensure that U1 screw is tighten to the torque between 1.5 and 2 N.m (15 to 20 kg.cm). Refer
to Disassembly 5 .
Remount the generator inside the table.
3.7.9 Verify
Reapply power and verify the presence of 3 phases.
Verify that the system functions correctly.
3.8.1 Supplies
None.
3.8.2 Tools
• 5.5 mm socket, with ratchet if possible (to remove cover and board)
• 10 mm socket with ratchet (to remove DC and rectifier cables)
Page 268 Section 2.0 JEDI Generator Renewal Parts
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
• Large Philips screw driver (to remove cables from the rectifier block).
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
1.) Insure that the neon (orange) indicator light on the board has gone out.
2.) Be careful touching components on this board; some become very hot.
3.) To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.8.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHTS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.8.5 Disassembly
Note: The AC/DC 3 Phases Mid Power Assembly is made up of two main elements: AC/DC Board 3
Phases and Rectifier Block 3 Phases.
4.) Remove the four 5.5 mm screws securing the AC/DC Unit to the support plate 4 . The AC/
1.) Remove fourteen 5.5 mm hex head screws which retain the cover. Remove that one.
2.) Remove the two DC cables fastened with 10 mm nuts on the board 1 .
3.) Remove two screws retaining the AC/DC board to the rectifier block 2 .
4.) Remove seven 5.5 mm hex head screws 3 and lift out the AC/DC Board 4 .
1.) The Rectifier Block 1 appears after removing the EMC Board (refer to 3.7 EMC Board 3
Phases) and the AC/DC Board.
2.) With the Allen key 3 mm, unscrew the 2 screws 2 holding the rectifier to the chassis and
remove the block.
3.8.6 Reassembly
1.) Reverse the disassembly procedure to reinstall the AC/DC 3 PH Mid Power Assembly and the
Verify that the + and - DC Bus cables are not reversed to prevent damage to the Inverter
The Rectifier Block has to be placed with the two poles in regard of the AC/DC Board
Be careful to preserve the special grease that is spread between the Rectifier Block and
the chassis. This grease, called: Conductive Surface Compound, provides both thermal
and electrical conductivity that is important in reducing failure of the rectifier.
3.) Remount the generator inside the table.
3.8.7 Verify
Reapply power and verify that the system functions correctly.
3.9.1 Supplies
None.
3.9.2 Tools
• 5.5 mm nut driver or socket wrench with ratchet.
• 3 mm Allen wrench.
• 13 mm socket with ratchet.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.9.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHTS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.9.5 Personnel
1 person.
3.9.7 Disassembly
THE POWER UNIT IS NOW FREE BUT IT WEIGHTS 30 KILOGRAMS (66 POUNDS). DO NOT
TRY TO LIFT IT WITHOUT ASSISTANCE.
1.) On the Power Unit, remove the 5.5 mm hex head screws 1 which secure the L-shaped EMC
2.) With a 10 mm socket wrench with ratchet 3 , remove the three hex head screws 4 securing
7 7
6 6
3.) Remove the 10 mm screw 5 on the left current transformer Ti1/40 pass through, top of the
Capacitor assembly. Retain the washers and release the cable.
4.) Disconnect the two cables linking coils to capacitors at right and left 6 Use a socket wrench
to remove screws.
5.) On each side 7 , remove one 5.5 mm screw holding the coils support.
3.9.8 Reassembly
1.) Present the new Inverter Coil assembly and mount one 5.5 mm hex head screw on each side
of the frame to hold the Coil assembly (refer to 3.9 Inverter Coil Mid Power Assembly
removed 7 .
Tighten to torque between 1.5 to 2 N.m (15 to 20 kg.cm)
2.) Remount all cables that were removed.
a.) Remount the three Coil cables on the Inverter Mid Power Assembly (IGBTS) with three
c.) Using a socket wrench, connect the two cables linking coils to capacitors at right and left
(refer to section Disassembly 6 .
Tighten to torque between 1.5 to 2 N.m (15 to 20 kg.cm).
3.) Refer to Disassembly / Reassembly procedure for 3.4 Inverter Mid-Power Assembly.
3.10.1 Supplies
None.
3.10.2 Tools
• 5.5 mm nut driver or socket wrench with ratchet.
• 10 mm allen wrench.
• 13 mm socket with ratchet.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.10.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHTS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.10.5 Personnel
1 person.
3.10.6 Time
3 hours.
3.10.7 Disassembly
THE POWER UNIT IS NOW FREE BUT IT WEIGHTS 30 KILOGRAMS (66 POUNDS). DO NOT
TRY TO LIFT IT WITHOUT ASSISTANCE.
1
2
1.) On the Power Unit, remove the 5.5 mm hex head screws 1 which secure the L-shaped EMC
2.) Remove the cable from connector J1 3 on the Gate Command Board.
6 5 6
8 8
4.) Remove the 10 mm 5 and 13 mm screws 6 on the top side , and the 13 mm screws 7 on
the lower side of the Capacitor assembly. Release the cables.
5.) Remove the four left and right 5.5 mm screws 8 holding the Capacitor assembly.
3.10.8 Reassembly
1.) Present the new capacitor assembly.
2.) Mount two 5.5 mm hex head screws to hold the Capacitor assembly (refer to section
Disassembly 8 ) ).
3.) Reconnect the cable J1 on the Gate Command Board (refer to section Disassembly 3 ).
4.) Attach the shorter cable (parallel coil) on the upper right current transformer (Ti1/100)
pass through with a 10 mm screw 1 and its washers (refer to section Disassembly 5 .
Tighten to torque 3 N.m (30 kg.cm).
5.) Remount the DC+ cable and the corresponding serial coil cable together on the upper left
side of the capacitor assembly with a 13 mm screw and its washers (refer to section
Disassembly 6 . Tighten to torque 8 N.m (80 kg.cm).
6.) Remount the DC- cable and the corresponding serial coil cable together on the upper
right side of the capacitor assembly with a 13 mm screw and its washers (refer to section
Disassembly 6 . Tighten to torque 8 N.m (80 kg.cm).
7.) Remount the remaining serial coil cable to the upper left current transformer (Ti1/40) pass
through with a 10 mm screw 1 and its washers (refer to section Disassembly 5 . Tighten
to torque 3 N.m (30 kg.cm).
8.) Run the ”INVERTER IN SHORT CIRCUIT” diagnostic:
Q100E
E Inverter Assembly
E
Coil Assembly
+ -
+ DC Ti1/40 Ti1/100 - DC
Capa Assembly
c.) Reassemble the Power Unit with AC/DC Unit, Interface Unit and Auxiliary Unit (refer to
3.4 Inverter Mid-Power Assembly).
d.) Run “inverter short circuit diagnostics”.
9.) After ”INVERTER SHORT CIRCUIT” diagnostics, disassemble the Auxiliary Unit and the
Interface Unit (see section Disassembly). Remove the short circuit cable.
10.) On the central capacitor (C3UF), attach together the two flat wires of the primary coils of the
HV transformer that have a white ring 2 and the longer wire of Parallel Coil with a 13 mm
screw and its washers. Tighten to torque 8 N.m (80 kg.cm).
3 3
11.) Mount the two remaining flat cables 3 from the HV Tank with 13 mm screws and their
12.) Reconnect the Filament cable J3 on the HV Tank (refer to section Disassembly 4 .
13.) Mount the two last remaining 5.5 mm hex head screws on the left and right side of the frame
(refer to section Disassembly).
Tighten to torque 1.5 to 2 N.m (15 to 20 kg.cm).
14.) Reassemble the Power Unit. Mount it with the AC/DC Unit, the Interface Unit and the Auxiliary
Unit on the support plate (Refer to 3.4 Inverter Mid-Power Assembly).
3.10.9 Verify
It is recommended to perform the “No load HV“ diagnostics test. Verify correct operation on the
whole application range.
3.11.1 Supplies
None.
3.11.2 Tools
5.5 mm hex nut driver or socket wrench with ratchet.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
1.) Insure that the Neon N1 (orange) indicator light on the LVPS Board has gone out.
2.) Be careful touching components on the LVPS Board. Some become very hot.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.11.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHTS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.11.5 Personnel
1 person.
3.11.6 Time
15 minutes.
1.) Disconnect the connectors J2 and J3 and the ground cable on the top of the generator 1 .
2.) Remove four 5.5 mm hex head securing screws 2 olding the Auxiliary Unit 4 on the support
plate 5 .
3.) Disconnect the flat ribbon cable 3 without tearing, between the Auxiliary Unit 4 and the
Interface Unit 6 .
4.) Remove the Auxiliary Unit.
1.) The Auxiliary Unit being free, unscrew ten 5.5 hex head screws and remove its cover.
2.) Disconnect the 3 cables 1 , 2 , 3 from the LV Power Supply 3 Phases board 4 .
3.) Unscrew the six 5.5 mm hex head screws and remove the LV Power Supply 3 Phase Board
4 .
3.11.8 Reassembly
Reverse the disassembly procedures to reinstall the LV Power Supply 3 Phases Board.
Verify that the following cables are not reversed before power ON (refer to section 3.11.7.2
Remove the LV Power Supply Board:
3.11.9 Verify
Verify that the system functions correctly.
3.12.1 Supplies
None.
3.12.2 Tools
5.5 mm hex nut driver or socket wrench with ratchet.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
1.) Insure that the Neon DS3 (green) indicator light on the Filament Board has gone out.
2.) Be careful touching components on the Filament Board . Some become very hot.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.12.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHTS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.12.5 Personnel
1 person.
3.12.6 Time
15 minutes.
1.) Disconnect the connectors J2 and J3 and the ground cable on the top of the generator 1 .
2.) Remove four 5.5 mm hex head securing screws 2 holding the Auxiliary Unit 4 on the
support plate 5 .
3.) Disconnect the flat ribbon cable 3 , without tearing, between the Auxiliary Unit 4 and the
Interface Unit 6 .
4.) Remove the Auxiliary Unit.
1.) The Auxiliary Unit being free, unscrew ten 5.5 mm hex head screws and remove its cover.
2.) Disconnect the ribbon cable 1 between the Filament Board 4 and the LVPS Board 5
3.) Unscrew the six 5.5 mm hex head screws and remove the Filament Board 1 INV 4 .
3.12.8 Reassembly
Reverse the disassembly procedures to reinstall the Filament Board 1 INV.
Verify that cables CF4 (160 V) and CF3 (0 V) 2 are not reversed before power ON (refer to
section 3.12.7.2 Remove the Filament Board).
Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
Remount the cover. Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
Remount the auxiliary unit on the generator support plate.
Remount the generator inside the table.
3.12.9 Verify
Verify that the system functions correctly.
Run Heater Function Diagnostics.
3.13.1 Supplies
None.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
1.) Insure that the Neon DS7 (orange) indicator light on the Rotor Board has gone out.
2.) Be careful touching components on the Rotor Board. Some become very hot.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.13.3 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHTS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.13.4 Personnel
1 person.
3.13.5 Time
15 minutes.
3.13.6 Disassembly
1.) Disconnect the connectors J2 and J3 and the ground cable on the top of the generator 1 .
2.) Remove four 5.5 mm hex head securing screws 2 holding the Auxiliary Unit 4 on the
support plate 5 .
3.) Disconnect the flat ribbon cable 3 , without tearing, between the Auxiliary Unit 4 and the
Interface Unit 6 .
4.) Remove the Auxiliary Unit.
1.) The Auxiliary Unit being free, unscrew ten 5.5 mm hex head screws and remove its cover.
2.) Disconnect the ribbon cable 1 between the Rotor Board High Speed 4 and the LVPS
3.) Unscrew two 5.5 mm hex head screws that are in holes 3 below the surface of the board
and that secure an IGBT.
4.) Unscrew the eight 5.5 mm hex head screws and remove the Rotor Board High Speed 4 .
Be careful to preserve the special grease that is spread between the IGBT and a heat sink
under the Rotation Board. This grease, called "Conductive Surface Compound", provides
both thermal and electrical conductivity that are important in reducing failure of the IGBT.
3.13.7 Reassembly
Reverse the disassembly procedure to reinstall the Rotor Board High Speed.
Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
Ensure that cables are correctly connected. Ensure that marks on the cables match the
marks on the board.
Remount the cover. Tighten the 5.5 mm hex head screws to 1.5 to 2 N.m (15 to 20 kg.cm).
Remount the auxiliary unit on the generator support plate.
Remount the generator inside the table.
3.14.1 Supplies
None.
3.14.2 Tools
5.5 mm nut driver or socket wrench with ratchet.
BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY LOCK
OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING INSIDE THE
EQUIPMENT IS REQUIRED.
To protect the circuit boards, do not forget to wear the antistatic bracelet.
3.14.4 Prerequisites
Remove the generator from the table and put it on its bottom side on the floor or on a work bench.
THE WHOLE GENERATOR WEIGHTS 63 KILOGRAMS (139 POUNDS). TWO PEOPLE ARE
REQUIRED TO CARRY IT.
3.14.5 Personnel
1 person.
3.14.6 Time
10 minutes.
1.) Disconnect the connectors J2 and J3 and the ground cable on the top of the generator 1 .
2.) Remove four 5.5 mm hex head securing screws 2 holding the Auxiliary Unit 4 on the
support plate 5 .
3.) Disconnect the flat ribbon cable 3 , without tearing, between the Auxiliary Unit 4 and the
Interface Unit 6 .
4.) Remove the Auxiliary Unit.
1.) After removing the Auxiliary Unit, the Interface Unit appears. Release the flat ribbon cable
2.) Unscrew twelve 5.5 mm screws that retain the cover panel and remove this one 2 .
4 6
8
3.) Disconnect the flat ribbon cable from J5 3 between the I/F Board RAD 7 and the AEC
Board 8 .
4.) Disconnect the connectors J2 4 and J1 5 and Test Points TP1 and TP2 6 on the AEC
Board.
5.) Remove the six 5.5 mm screws securing the AEC Board 8 .
3.14.8 Reassembly
Reverse the disassembly procedure to reassemble the unit. Tighten screws to torque between 1.5
and 2 N.m (15 to 20 kg.cm).
3.14.9 Verify
Reapply power. Proceed to ”AEC diagnostics” to ensure correct operation.
Section 1.0
Software Download / Upgrade
1.1 Supplies
None.
1.2 Tools
• JEDI service software (the software is a DOS application)
• Console R232 cable (5194292)
• Computer with floppy or CD drive
1.4 Prerequisites
Basic knowledge of PC based computer laptop, DOS and usage of Microsoft WINDOWS.
1.5 Personnel
1 person.
1.6 Time
15 minutes.
1.7 Procedure
3.) Press button together more than 5 seconds to enter the HUB model;
4.) Restart the computer in DOS mode (for software download or database backup, it is required
to run ASIAN.exe under DOS to avoid communication problem).
With the mouse, click on START, SHUTDOWN, RESTART COMPUTER UNDER DOS
MODE.
5.) Run the program:
Under DOS:
* Select Disk drive from where you want to run the program:
* C: [enter]
* key in: “ASIAN” [enter]
6.) From the laptop software main screen, type “F7: restore memory”.
7.) Type “appl.mx”, then enter (or C:\JEDI\appl.mx).
Wait for the count to start. It may take 20 seconds before starting. The software code transfer
will take about 20 minutes. Download is complete when all the LEDs on kV Control board are
ON and laptop indicates download is complete.
Normally, the kV Control board will reset at the end of transfer and all the LEDs will scroll.
8.) Verify the software version is P3.2, behind “RAM” text (or P3.X in case of new software);
9.) Select menu 2 with the arrows, select LIST, and clear generator error log. (^H, then W);
10.) To exit program, press Q, or R to return then ESC, or ESC twice.
Section 2.0
Tube Replacement Software Procedure
2.1 Supplies
None.
2.2 Tools
• JEDI service software (the software is a DOS application)
• Console R232 cable (5194292)
• Computer with floppy or CD drive
2.4 Prerequisites
Basic knowledge of PC based computer laptop, DOS and usage of Microsoft WINDOWS.
Knowledge of Service Software (refer to Chapter 3 - - Configuraton & Calibration - 2.2 - Service
Software User Guide).
2.5 Personnel
1 person.
2.7 Procedure
3.) Press button together more than 5 seconds to enter the HUB model;
4.) Restart the computer in DOS mode (for software download or database backup, it is required
to run ASIAN.exe under DOS to avoid communication problem).
With the mouse, click on START, SHUTDOWN, RESTART COMPUTER UNDER DOS
MODE.
5.) Run the program:
Under DOS:
* Select Disk drive from where you want to run the program:
* C: [enter]
* key in: “ASIAN” [enter]
Note: Before resetting all the memory, it is advised to record them on table 2.7.4 - Generator Counter
Reading Record, found at the end of this Job Card.
4.) Enter ^H, press W to confirm.
All the tube related Data will be erased; only Generator related Data will be kept.
The filament ageing parameter will be reset to default value.
Serial number:
Software version:
Options: (AEC)
X-Ray system
description:
7.) On the application screen, press button together for more than 5 seconds to
Page 299
1 2 3 4 5 6 7 8
A A
Grounding
Grounding
B B
Tube Stand
System Cabinet
Power Supply Tube
P1 P1 24 VDC Console
PDU
P2 TM2
C Hospital Power P2
24 VDC C
P3 P3 Terminal
Supply GND
Grounding
P1 P2 P3
GND
Stud
3 Phase
Chapter 10 - Schematics
Stud
TM3 Ele-Magnetic
Brakes
Rotor/Tube fan/Pressure SW
P1 P2 P3 J2/J3 Tube
HV +
HV + HV +
HV -
D HV - HV -
D
Xray-On Light J6
24 VDC
Chapter 10 - Schematics
Generator Collimator
Door Interlock J6
Jedi
Table
Switches
E 24 VDC
E
AEC2
AEC1
EM Brakes
J8
J9
TM1
5166493
RS232
Ion-Chamber
RTL
RS485/
RS232
5166654
Ion-Chamber
F Wallstand F
Serial / RTL
Grounding Line
Others System Cosole
J1
H TITLE
Pioneer System Architecture
REV H
FIRST MADE FOR: Pioneer
2
GEHL MEDICAL SYSTEMS 51888818 BLK SHT/CONT ON
1/2
1 2 3 4 5 6 7 8
Section 1.0
GE HEALTHCARE
MIS Map
LING LONG XR 6000 SERVICE MANUAL
1 2 3 4 5 6 7 8
A A
Service Laptop
Handswtich 5194292/
J1
Console J2 5241079 Serial port
5194507
B B
J4 J3
5184620
5182354
5183903
5196643
5182354
5196454
UVW
laptop
console
GND
E Stud GND 5183561 E
Stud
TM1
Stud
GND
TM1
DOOR Interlock AEC1
5166493
AEC1
Room Light J6
F F
5193106
5193103
Jedi R1T 50
2212259-2/4
J2
XP105
Stud
GND
5195936 Rotation
J3 Anode
HV +
2269002-5
HV +
G 2269002-5 Tube Stand G
HV - HV -
5183392/
5184093
DIRECTION 5275601-100, REVISION 2
5166654
AEC2
H TITLE
Pioneer 500 System Architecture
REV H
FIRST MADE FOR: Pioneer 500
2
GEHL MEDICAL SYSTEMS 51888818 BLK SHT/CONT ON
2/2
1 2 3 4 5 6 7 8
GE HEALTHCARE
Page 300
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
1.1 Jedi 50 R 1T Cabling
Section 2.0
Central Listings
2.1 kV Measure Board
This board forms part of the oil seal of the High Voltage Tank. It can only be removed at the
factory. The Field Replaceable Unit (FRU) is the complete HV Tank.
HIGH VOLTAGE: DO NOT GO INTO GENERATOR UNTIL INDICATOR DS3 GOES OUT.
DC
160v LF Common
SF
HIGH VOLTAGE: DO NOT GO INTO GENERATOR UNTIL INDICATOR DS6 AND DS7 (NEON-
ORANGE) GO OUT.
Auxiliary
DC bus Common
Main
DS6
DC bus
2.3.2 J2 Wiring
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
115V
sense
Fan, 115V AC
Line input
Fan*, 115V AC X-ray tube
Line output
Thermal safety
HIGH VOLTAGE AND HOT SURFACE: DO NOT TOUCH BOARD UNTIL INDICATOR NE1
(NEON - ORANGE) GOES OUT.
+ 400
DC bus
- 400
160V
DC
0
HIGH VOLTAGE: DO NOT TOUCH BOARD UNTIL DS300 ON THIS BOARD AND DS1 ON
DUAL SNUB BOARD ARE OUT.
DC bus
DS300 - Neon (Orange) -
Presence of voltage on DC
bus for Flyback Converter to
create power supplies for
both Gate Commands
ILR
DS102
* Insulated Gate
Bipolar Transistor
HIGH VOLTAGE: DO NOT GO INTO GENERATOR UNTIL INDICATOR DS1 (NEON - ORANGE)
GOES OUT.
2.7.2 Indicators
REFERENCE DESIGNATION
U32 NON VOLATILE PARAMETERS RAM (BATTERY
INCLUDED)
HIGH VOLTAGE: DO NOT GO INTO GENERATOR UNTIL DS1 (NEON - ORANGE) GOES OUT.
N1 - Orange (neon) -
Indicates presence of 100 ON & VAC
VAC for the
Pin 9
Table Bucky
Pin 1
Pin 9
Wall Bucky
Pin 1
Pin 9
Pin 1
Service Laptop
TUBE
PUPT - Yellow -
Data being sent to the console
Table Bucky
Pin 1
Pin 9
Wall Bucky
Pin 1
To AEC board
Pin 9
Room Door & Light I/F
Pin 1
DS4: CAN TRANSMIT
(yellow)
Service Laptop
DS5: EXT.CAN
TRANSMIT (yellow)
DS6: Receive Data
from console (yellow)
Console
OPTIONAL
Presence of strap -
TUBE 1
Testpoint Measurement
SIGNAL RANGE
TESTPOINT MEASUREMENT
SIGNAL RANGE
TP1 230 VDC 230 VDC _5%
TP2 230 VDC 230 VDC _5%
TP3 230 VDC 230 VDC _5%
TP4 230 VDC 230 VDC _5%
TP5 AEC assignment reference 0 to 10 V
TP6 10 VDC (reference) 10 VDC _1%
TP7 AEC return voltage
TP8 AEC1 return voltage
TP9 AEC2 return voltage
TP10 AEC3 return voltage
TP11 AEC4 return voltage
Section 3.0
Schematics
3.1 PDU Connection Diagram
6
MC8 EM lock7XEH VWDQG
- Cassette tray VZLWFK༨ ᅧ
( $&9
ൎပ࢈ڑนШΞ҈Ԥ (
Ѥӊ٤Еᄗವ Αै6FDOH ϫॸ 0DWHULDO ζੋ҉स )LQLVK ϠѩΩۜ
7ROHUDQFH
&ROOLPDWRU
ಁޙఐ ໎ճఐ ಌଛఐ ы 7LWOH
'HVLJQ'DWH &KHFN'DWH $SSURYDO'DWH
DIRECTION 5275601-100, REVISION 2
3LRQHHU7DEOH(OHFWULFDO
ଫҕԨ୷ၐस
$&9 7KHRU\'UDZLQJ
ำߑफι ۜ 3DUW 1R ͑·ۜ ྚۜ
) ඹဈԨ୷ܟྤ९ಁပຫ٤ഓ 'RF7\SH 5HY 3DJH )
*(+/ 0(',&$/ 6<67(06 6&+ 6&+
2
GE HEALTHCARE
Page 318
3.2
LING LONG XR 6000 SERVICE MANUAL
Page 319
&74 7DEOH ͑·ۜ ٗ؟ઝఘ ٗ؟რ ఐ
5HY 'HVFULSWLRQ 0DGH E\ 'DWH
3(
'LPHQVLRQ
;3
;6
इ
&74
܃
3(
۱
͂
܃ ܃
ڕ౨ຏၮՎԨҨත 0&
:
;3
;6
۱ ۱
;3
;6
۱ /'
ᠭܻ 12 6 12
& &
:
% ڕ༨ᅧၮՎԨҨත 0& %
;6
;3
۱
܃ ܃ ࠄ̟ࣅᄥۧᆦߑ
ोᅘഃ଼ၮՎԨҨත 0&
:
۱ ۱
ࣅᄥۧ
;3
;6
ᠭܻ
܃
۱
& &
͂
इ
ܻ
:
;6
3(
:
3(
;3 :
ᠭܻ
܃
۱
۱
܃
इ
ܻ
ोᅘ70ᆦߑ
Chapter 10 - Schematics
ोᅘ :
܃ ܃
:
' : :
'
0& 0& ҕੋ۸ၮՎԨҨත
܃
۱
3(
܃
۱
1& ۱ ۱
: ܃ ۱
&20 6
܃
܃
۱
܃ :
&20 1&
۱ 3( 0& 0& ҕੋᆚၮՎԨҨත
ᠭܻ
3(
6
܃
۱
۱
इ
ܻ
ଫ̟ۧ
3(
3( 70 ۱ ۱ ۱
( :
ଫۧԨҨත
ଫඞ ҕ (
ᠭܻ
܃
۱
:
ܻ
͂
0&
इ
ҕ70ᆦߑ
Ѥӊ٤Еᄗವ Αै6FDOH ϫॸ 0DWHULDO ζੋ҉स )LQLVK ϠѩΩۜ
3(
3(
: : : 7ROHUDQFH
܃
۱
ಁޙఐ ໎ճఐ ಌଛఐ ы 7LWOH
܃
۱
DIRECTION 5275601-100, REVISION 2
ܻ
ᠭܻ
इ
͂
۱
܃
ޏ೬ྑௐġ ำߑफι ۜ 3DUW 1R ͑·ۜ ྚۜ
) 3(
ඹဈԨ୷ܟྤ९ಁပຫ٤ഓ 'RF7\SH 5HY 3DJH )
c̟ᄜྑௐၽອԅॕࠄ୶౨Ⴔඬճԅॕࠄ୶γஆd 5197241ADW
3'870ᆦߑ 3'870 *(+/ 0(',&$/ 6<67(06 $': 3
cၽॕࠄْڑನēᅟ࿉ْڑԅࠄԤူຬऍ౨ԅຬۜճڑຂd
GE HEALTHCARE
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
1 2 3 4 5 6 7 8
I1 I8
I5
LPC2294 MEMORY
A POWER
ARM_JTAG<5..0> DATA<31..0>
DIRECTION 5275601-100, REVISION 2
ARM_IO<30..0>
CPLD_JTAG<3..0> ADDR<23..0>
CPLD_CLK
ARM_RESET ARM_IO<30..0>
ARM_CON<9..0> FEED_DOG
B
I3
EPM570
I4
ADDR<23..0>
CPLD_CLK ARM_IO<30..0>
ISP_ENTER ISP_ENTER
ARM_RESET
ARM_IO<30..0> CPLD_IO<33..0>
LED<6..1> ARM_RESET
FEED_DOG
HCLK
Chapter 10 - Schematics
I7
LCD I6
JTAG
DATA<31..0>
ARM_RESET ARM_JTAG<5..0>
ADDR<23..0>
CPLD_JTAG<3..0>
ARM_CON<9..0> ARM_IO<30..0>
HCLK ARM_RESET
A61001C Console Control Board Circuit Diagram
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
Page 321
LING LONG XR 6000 SERVICE MANUAL
Page 322
GE HEALTHCARE
1 2 3 4 5 6 7 8
5C6> ARM_RESET
A
DIRECTION 5275601-100, REVISION 2
ARM_IO<30..0> 4D3>
IO
U13
74LVC-LCX14
R1 3V3 1V8
CPLD_CLK 12 B A 13
OUT
33.2
U1
0.125W
VCC=3V3;GND=GND LPC2292FBD144
1%
V18(1..0)
3V3 143 V18A
V3(9..0)
Y1 U13 14 V3A
74LVC-LCX14
FL3 ARM_RESET 135 RESET_N TD1 22 CAN1_TX 26
BLM31A601S C4
1 2 4 VCC VOUT 3 9 A B 8 CPU_CLK 142 XTAL1
141 XTAL2
600 OHM 1 2 100P
R2 ENABLE GND 50V 0 PO0_TXD0 42 91 CS0 0
C1 C2 C3 P0_0_TXD0_PWM1 P1_0_CS0
50V 5% 1 PO1_RXD0 49 90 OE 9
1U 10N 1K VCC=3V3;GND=GND P0_1_RXD0_PWM3_EINT0 P1_1_OE
16V 50V 0.1U 0.125W 10MHZ 2 PO2_SCL 50 P0_2_SCL_CAP0_0 ARM_CON<9..0> 4C7>
OUT
B 10% 10% 10% 1% 3 PO3_SDA 58 P0_3_SDA_MAT0_0_EINT1 P1_16_TRACEPKT0 34
4 PO4_SCK0 59 P0_4_SCK0_CAP0_1 P1_17_TRACEPKT1 24
5 PO5_MISO0 61 P0_5_MISO0_MAT0_1 P1_18_TRACEPKT2 15
6 PO6_MOSI0 68 P0_6_MOSI0_CAP0_2 P1_19_TRACEPKT3 7
7 PO7_SSEL 69 P0_7_SSEL0_PWM2_EINT2 P1_20_TRACESYNC 102
GND 8 PO8_TXD1 75 P0_8_TXD1_PWM4 P1_21_PIPESTAT0 95
9 PO9_RXD1 76 P0_9_RXD1_PWM6_EINT3 P1_22_PIPESTAT1 86
10 PO10_RTS1 78 P0_10_RTS1_CAP1_0 P1_23_PIPESTAT2 82
11 PO11_CTS1 83 P0_11_CTS1_CAP1_1 P1_24_TRACECLK 70
12 PO12_DSR1 84 P0_12_DSR1_MAT1_0 P1_25_EXTIN0 60
GND 13 PO13_DTR1 85 P0_13_DTR1_MAT1_1 P1_26_RTCK 52 ARM_RTCK 0
14 PO14_DCD1 92 P0_14_DCD1_EINT1 P1_27_TDO 144 ARM_TDO 1
15 PO15_EINT2 99 P0_15_RI1_EINT2 P1_28_TDI 140 ARM_TDI 2
16 PO16_EINT0 100 P0_16_EINT0_MAT0_2_CAP0_2 P1_29_TCK 126 ARM_TCK 3
17 PO17_SCK1 101 P0_17_CAP1_2_SCK1_MAT1_2 P1_30_TMS 113 ARM_TMS 4
18 PO18_MISO1 121 P0_18_CAP1_3_MISO1_MAT1_3 P1_31_TRST 43 ARM_TRST 5
19 PO19_MOSI1 122 P0_19_MAT1_2_MOSI1_CAP1_2
20 PO20_SSEL1 123 P0_20_MAT1_3_SSEL1_EINT3
21 PO21_PWM5 4
0
0.125W 0%
A
R3 AGND GND
17 PO17_SCK1 CPLD_TCK 1
0
0.125W 0%
R4
18 PO18_MISO1 CPLD_TDO 0
0
0.125W 0%
R5
19 PO19_MOSI1 CPLD_TDI 3
0
0.125W 0%
D
R6
20 PO20_SSEL1 CPLD_TMS 2 CPLD_JTAG<3..0>
IO
0
0.125W 0%
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
1 2 3 4 5 6 7 8
A
DIRECTION 5275601-100, REVISION 2
DATA<31..0> ADDR<23..0>
IO OUT
U1
LPC2292FBD144
Chapter 10 - Schematics
10K R14 SYM 2 OF 2
C6
0.1W
10K 0.1U
0.1% 3V3 ARM_CON<9..0>
0.1W 50V OUT 3B8>
0.1% 10%
C R19 3V3 3V3
10K
R16
0.1W
A
0.1% 10K
AGND 0.1W
BOOT1 DATA<27> 27 0.1%
1V8 R10
29 26 DATA<26> BOOT0
10K
C9 C11 C13 C15 C16 C17 C18 C19
0.1W R15
C7 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
0.1%
10K 0.1U 50V 50V 50V 50V 50V 50V 50V 50V
0.1W 50V 10% 10% 10% 10% 10% 10% 10% 10%
0.1% 10%
A
BOOT 1:0 = 11 INTERNAL BOOTUP
AGND
12V R11 C10 C12 C14
30 0.1U 0.1U 0.1U
1V8 50V 50V 50V
10K
0.1W 10% 10% 10%
R12 C8
0.1% 2K GND
0.1U
0.125W 50V
D 1% 10%
A
AGND
ARM_IO<30..0> 3A8<>
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
Page 323
LING LONG XR 6000 SERVICE MANUAL
Page 324
GE HEALTHCARE
1 2 3 4 5 6 7 8
A
ARM RESET & WATCH DOG
DIRECTION 5275601-100, REVISION 2
VCC
VCC
R9
10K
0.125W
1%
3V3
J3
B
FEED_DOG 1 2 C23
IN R21
0.1U
10K U24 50V
R20
1%
JUMPER 0.125W MAX6301 10%
R22
1%
2.49K
0.125W
20K
S1 1%
0.125W
KT11B1JM
8 VCC
1 1 3 3 1 RESET_IN
6 WDI
2 2 4 4 5 WDS RESET_N 7 ARM_RESET 3A4<
OUT
3 SRT
4 SWT GND 2
EXTENDED MODE
R18
C20 C21 C22
5V
1K 2.2U 47N 4.7N
C TWD=2.67*CSW*500=6.3S
GND GND
GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
1 2 3 4 5 6 7 8
R24 R28
7 3
NLED<6..1>
10K 10K
3
1
2
4
5
6
0.125W 0.125W
CPLD_IO<33..0> 1% 1%
IO
R25 R29
3
1
9
0
2
4
5
6
10
11
12
13
14
15
16
17
18
19
6 2
20
21
22
23
10K 10K
A 0.125W 0.125W
DIRECTION 5275601-100, REVISION 2
1% 1%
R26 3V3 R30 3V3
5 1
101
102
103
104
105
106
107
108
73
74
75
76
77
78
79
80
81
84
85
86
87
88
93
94
95
96
97
98
3V3 3V3
10K 10K
0.125W 0.125W
1% 1%
GND R27 R31
IO_73
IO_74
IO_75
IO_76
IO_77
IO_78
IO_79
IO_80
IO_81
IO_84
IO_85
IO_86
IO_87
IO_88
IO_93
IO_94
IO_95
IO_96
IO_97
IO_98
IO_101
IO_102
IO_103
IO_104
IO_105
IO_106
IO_107
IO_108
4 0
R23
10K 10K 10K
VCCINT<3..0>
0.125W 0.125W 0.125W
VCCIO1<3..0>
1% 1% 1%
VCCIO2<3..0> S2
CHS_04TB
C24
C26
C28
C30
50V 50V 50V 50V 8 IO_11 IO_133 27
OUT
10% 10% 10% 10% 9 DATA<9> 12 IO_12 IO_134 134 28
10 DATA<10> 13 IO_13 IO_137 137 29 IN
11 DATA<11> 14 IO_14 IO_138 138 30
12 DATA<12> 15 IO_15 IO_139 139 31 GND
Chapter 10 - Schematics
13 DATA<13> 16 IO_16 IO_140 140 32 U16
14 DATA<14> 21 IO_21 IO_141 141 33 74ABT244 VCC
15 DATA<15> 22 IO_22 IO_142 142 PO22
0.1U 0.1U 0.1U 0.1U 0 23 IO_23 IO_143 143 FEED_DOG
24 144 ISP_ENTER 1
C25
C27
C29
C31
50V 50V 50V 50V 1 IO_24 IO_144 OE_N C32
10% 10% 10% 10% 2 27 IO_27 0.1U
C 3 28 IO_28 GNDI0<7..0> 6 2 A1 Y1 18 6 50V
4 29 IO_29 GNDINT<3..0> 5 4 A2 Y2 16 5 10%
5 30 IO_30 4 6 A3 Y3 14 4
6 31 IO_31 3 8 A4 Y4 12 3
7 32 IO_32
GND
GND SWITCH<7..0> VCC=VCC;GND=GND
U16
IO_37
IO_38
IO_39
IO_40
IO_41
IO_42
IO_43
IO_44
IO_45
IO_48
IO_49
IO_50
IO_51
IO_52
IO_53
IO_55
IO_57
IO_58
IO_59
IO_62
IO_63
IO_66
IO_67
IO_68
IO_69
IO_70
IO_71
IO_72
74ABT244
VCC
GND
37
38
39
70
71
72
40
41
42
43
44
45
48
49
50
51
52
53
55
57
58
59
62
63
66
67
68
69
15 A3 Y3 5 0.125W
PO21_PWM5 17 A4 Y4 3 1 2 1%
RED
1
VCC=VCC;GND=GND
7
0
2
4
5
6
8
9
ARM_CON<9..0>
IN
LED<6..1>
OUT
10
11
12
13
4
5
6
23
24
ARM_IO<30..0>
IO
D GND
22
21
3
1
7
0
2
4
5
6
8
ADDR<23..0>
IN
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
Page 325
LING LONG XR 6000 SERVICE MANUAL
Page 326
GE HEALTHCARE
1 2 3 4 5 6 7 8
PDU24V
POWER CONNECTOR
R33
3 J6
15K
0.125W
10C6> XRAY_ON_LIGHT 4 4 1 1
1% 5 2 PSW
5 2 7B5>
6 6 3 3 DOORIN TERLOCK 7C1<
C D1 12V
A R36
DIRECTION 5275601-100, REVISION 2
2 B LSM340J K1
Q1 40V
AZ742-2C-24D 24V_IN PIN
4.99K MMBT3904 GND
E RIGHT_ANGLE
CR1
0.125W
U14 WIRE_TO_BOARD
R47
1%
SMBJ5.0 1% 1 FL7
20K
COIL1 MC1489A SHIELD
6.7V 2 BLM31P500S
0.125W
COIL2 R46
9B3< OVER_CURRENT 3 1 1 2
4 3
COM<1> CLOSED<1> O IN
15K 50 OHM
OPEN<1> 5
7 0.125W
2
RESP_CTRL
PDUGND PDUGND COM<0> CLOSED<0> 6
VCC=VCC;GND=GND 1%
OPEN<0> 8 C38
R37 0.1U
1 50V
10%
15K PDU24V
J4
0.125W
F1 D3
1% D2 MURS320T3
LSM340J K2 K3 GND 3 3 1 1
C 40V 4 2
R38 G6B_2114P_US24VDC G6B_2114P_US24VDC 4 2 2A
0 B 200V
Q2
4.99K MMBT3904 8 8
E COILP COILP PIN
0.125W 1 1
COILN COILN RIGHT_ANGLE
CR2 1%
WIRE_TO_BOARD
SMBJ5.0 3 OPEN1 OPEN2 4 3 OPEN1 OPEN2 4 PSW 7A8<
6.7V 6 5 6 5
B CLOSED1 CLOSED2 CLOSED1 CLOSED2
PDUGND
PDUGND
PDUGND
ON/OFF_KEY<3..0> 11D3<> J7
FL5
BLM31P500S
1 2 1 1
VCC 7D1< PREP 2 2
3V3 50 OHM 3 3
7D1< EXP 4 4
R39
1%
1
1
FL4
20K
MC1489A
BLM31P500S 11 9 X_RAY_LIGHT
0.125W
R44 A1 Y1 10D4<
7A8> DOORIN TERLOCK 1 2 13 11 13 A2 Y2 7 CPLD_IO<0> 0 CPLD_INTERLOCK MH6
IN O 15 A3 Y3 5 CPLD_IO<1> 1 CPLD_PREP MHP4_50TB10_00
50 OHM 15K
1
12
MHP4_50TB10_00
1% VCC=VCC;GND=GND
1
VCC=3V3;GND=GND
U14
R40
1%
CPLD_IO<3> 3
20K
MC1489A
0.125W
R42 MH3 MH7 C39 R48
7C7> PREP 10 8 MHP4_50TB10_00 MHP4_50TB10_00 22N 1MEG
1
1
IN O 250V 0.125W
12V 15K 20% 1%
C33
0.125W RESP_CTRL
9
0.1U GND
1% VCC=VCC;GND=GND
50V MH4 MH8
10% MHP4_50TB10_00 MHP4_50TB10_00
1
1
U14
R41
1%
GND
20K
MC1489A
GND
0.125W
D R43
7C7> EXP 4 6
IN O
15K
C34 11C8<> 10B1<> 9B3<> CPLD_IO<33..0> SHIELD
0.125W RESP_CTRL IO
5
0.1U
1% VCC=VCC;GND=GND
50V
10%
GE Healthcare
Company Restricted
GND
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
1 2 3 4 5 6 7 8
A
DIRECTION 5275601-100, REVISION 2
TP2
TIN
VCC R53 R54
10K 1K ARM_IO<30..0> 9B1<>
IO
U2 0.125W 0.125W
DCR010505U 1% 1% CAN_VCC
TIN
HCPL0720-060 50V GND
VCC 10%
TP1
5 VREF
1 VDD1 VDD2 8 8 S 7 CAN_H 10C6<>
CANH
CAN_GND
26 CAN1_TX 2 VI VO 6 1 TXD
GND 3V3
R49 C42 C46 4 5 4 RXD CANL 6 CAN_L 10C8<>
U12 0.1U GND1 GND2
74VHC123A 3V3 50V
20K 1U 10% VCC=CAN_VCC;GND=CAN_GND
0.125W 16V
1% 10% 6
Chapter 10 - Schematics
CEXT R55 CAN_GND
R57
1%
DS1
121
7 REXT_CEXT Q 5 332
HSMY-C650 0.125W
0.125W
9 A_N 1%
10 B Q_N 12 1 2 GND U4
9C8< ARM_RESET 11 RD HCPL0720-060
IN
YELLOW VCC
C
R52 VCC=3V3;GND=GND 8 VDD2 VDD1 1
R50 0 R56
3V3 10K 0.125W 25 CAN1_RX 6 VO VI 2
0.125W 0%
1% CAN1_TX 26 C47 5 4 C48 332
CAN1_RX 25 0.125W
0.1U GND2 GND1 0.1U
C41 50V 50V 1%
0.1U R51 10% 10%
50V 10K
10% 0.125W
1%
CAN_GND
GND
GND
GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
Page 327
LING LONG XR 6000 SERVICE MANUAL
Page 328
GE HEALTHCARE
1 2 3 4 5 6 7 8
VCC
3V3
C57
0.1U
50V C60
10% 0.1U
50V
VCC 10%
A 3V3
DIRECTION 5275601-100, REVISION 2
3V3
VCC GND
C53 RS485_TX+ 10C8< R64 C58 GND
0.1U U12 3V3
50V R58 U18 74VHC123A
10% 4.99K 20K 1U
MAX488E R60
0.125W 110 0.125W 16V R72
1% 0.1W 1% 10% 14 332
CEXT
1 0.1% 15 13 DS5 0.125W
VCC REXT_CEXT Q
Y 5 HSMY-C650 1%
U8
GND 74LVC_LCX244 DATA_OUT 3 DI 1 A_N
Z 6 RS485_TX- 10C6< 2 B Q_N 4 1 2
10C6> RS485_RX+ 8 A 3 RD
OE_N 1 RO 2 DATA_IN YELLOW
R59 7 B
DATA_OUT 18 Y1 A1 2 PO8_TXD1 8 110 GND 4 R65 VCC=3V3;GND=GND
9 PO9_RXD1 16 Y2 A2 4 DATA_IN 0.1W 10K R69
14 Y3 A3 6 0.1% 0.125W 0
12 Y4 A4 8 OVER_CURRENT 7A5> 1% 0.125W
10C8> RS485_RX- 0%
VCC=3V3;GND=GND PO8_TXD1
PO9_RXD1
GND
B R67
8B6<> ARM_IO<30..0> GND 10K
0.125W
1%
24
CPLD_IO<33..0> 7D4<> 10B1<> 11C8<>
GND
ISP_ENTER
OUT
3V3
3V3 3V3
DTR
0.125W
C56 R63 D6
C C52 C55 1U PMBD914
R68
1%
20K
16 Q3
20K
VCC
MMBT3904 0
0.125W
2 VP E 0.125W
6 VN D4
ISP_TXD 0 C +0.05
PMBD914
1 C1P 85V B C59
3 Q4
C1M 0.1U
4 MMBT3904 50V
C2P
R61
E
1%
5 10%
20K
C2M D5
0.125W
PMBD914
C51 C54 0 PO0_TXD0 11 T1IN T1OUT 14 85V
0.1U 0.1U 10 T2IN T2OUT 7 GND
50V 50V 1 ISP_RXD 13 R1IN R1OUT 12 PO1_RXD0 1
10% 10% 8 R2IN R2OUT 9 RTS GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
1 2 3 4 5 6 7 8
VCC VCC
VCC
GND
B 19 OE_N
4 CPLD_IO<4> 11 A1 Y1 9 EXPOSURE_ENABLE P1
1%
1%
332
332
5 CPLD_IO<5> 13 A2 Y2 7 RESET I150
R83
R84
0.125W
0.125W
10A6> XRAY_ON/OFF 15 A3 Y3 5
6 CPLD_IO<6> 17 A4 Y4 3 XRAY_CMD BUZZER
1
VCC=3V3;GND=GND BUZ 2
GND 12V
7 CPLD_IO<7> FL6
VCC BLM31P500S
8 CPLD_IO<8> EXPOSURE 1 2
J11
50 OHM
9 CPLD_IO<9> PREPARE C64
GND
Chapter 10 - Schematics
0.1U
50V 3V3 1 1 2 2
10% 3 3 4 4 BUZ_CMD 10C1<
GND 10A8> XRAY_CMD+ 5 5 6 6 XRAY_ON/OFF+ 10A3<
10A6> RESET+ 7 7 8 8 EXPOSURE_ENABLE+ 10A3>
U13 U13 R78 R79 R82 9B3< RS485_RX+ 9 9 10 10 RS485_TX+ 9A5>
74LVC-LCX14 74LVC-LCX14 GND 10K 10K 10K 11 11 12 12
C 0.125W 0.125W 0.125W 8B8<> CAN_H 13 13 14 14 EXPOSURE_SW 10C6>
U7 1% 1% 1% 2111738 10C6> PREPARE_SW 15 15 16 16
1 2 11 10 74LVC_LCX244 18PIN 17 17 18 18 XRAY_CMD- 10A8>
A B A B
Q6 10B3< XRAY_ON/OFF- 19 19 20 20 RESET- 10A6>
ULN2803 10A3> EXPOSURE_ENABLE- 21 21 22 22 RS485_RX- 9B3<
12V 1 OE_N 9A5> RS485_TX- 23 23 24 24 CAN_L 8C8<>
VCC=3V3;GND=GND VCC=3V3;GND=GND 25 25 26 26
2 A1 Y1 18 1 I1 O1 18 EXPOSURE_SW 10C8<
4 A2 Y2 16 2 I2 O2 17 PREPARE_SW 10C6<
U20 6 A3 Y3 14 CPLD_IO<10> 10 3 I3 O3 16 XRAY_ON_LIGHT 7A7< PIN
R70
1%
8 12 4 15 BUZ
20K
MC1489A A4 Y4 I4 O4 RIGHT_ANGLE
5 14
0.125W
R75 I5 O5 RIBBON_CABLE_STANDARD
10C8< BUZ_CMD 1 3 6 I6 O6 13
IN O VCC=3V3;GND=GND 7 I7 O7 12 24V GENERATOR CONNECTOR
15K
8 I8 O8 11
0.125W
2
RESP_CTRL 10
1% VCC=VCC;GND=GND
COMMON CAN_GND
11 CPLD_IO<11> GND 9
D
GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
Page 329
LING LONG XR 6000 SERVICE MANUAL
Page 330
GE HEALTHCARE
1 2 3 4 5 6 7 8
A
R91 R102 R104 100V 50V
D8 10%
C67 R97 MMSD4148
10K 10K 10K
0.1U U23
50V
0.125W 0.125W 0.125W
74ABT244
10%
1% 10K 1% 1% GND
100V
0.125W
R92 R105 D9
1% 1
U26 MMSD4148 OE_N
74LVC_LCX244 R98
GND
10K 10K 18 2 CPLD_IO<23> 23
Y1 A1
0.125W 0.125W 16 4 CPLD_IO<22> 22
100V Y2 A2
1 1% 10K 1% 14 6 CPLD_IO<21> 21
OE_N D10 Y3 A3
0.125W 12 8 CPLD_IO<20> 20
R94 R106 MMSD4148 Y4 A4
12 CPLD_IO<12> 18 2 1%
Y1 A1
13 CPLD_IO<13> 16 Y2 A2 4
14 CPLD_IO<14> 14 6 10K 10K
Y3 A3 KEYBOARD CONNECTOR 100V VCC=VCC;GND=GND
15 CPLD_IO<15> 12 8 0.125W 0.125W
Y4 A4 J12 D11 U23
1% GND GND 1%
MMSD4148 74ABT244
1 1 2 2
VCC=3V3;GND=GND 3 3 4 4
U26 5 5 6 6 100V OE_N 19
74LVC_LCX244 7 7 8 8 D12
9 9 10 10 MMSD4148 9 Y1 A1 11 CPLD_IO<19> 19
11 11 12 12 7 Y2 A2 13 CPLD_IO<18> 18
OE_N 19 XRAY ON R99 13 13 14 14 5 Y3 A3 15
B 15 15 16 16 100V 3 Y4 A4 17
VCC DS3 16 CPLD_IO<16> 9 11 VCC 17 18
Y1 A1 17 18
HSMG-C650 17 CPLD_IO<17> 7 13 150 19 20
Y2 A2 19 20
5 15 0.125W 21 22
R34 Y3 A3 21 22 VCC=VCC;GND=GND
2 1 3 17 1% 23 24
Y4 A4 READY R93 23 24
25 25 26 26
332 27 28
GREEN 27 28
0.125W PDU24V 150 29 30
VCC=3V3;GND=GND 29 30
1% 0.125W 31 32
31 32
DS4 1% GND
POWER ON R100
HSMY-C650
R35 PIN
2 1 1.21K
RIGHT_ANGLE
GND 0.125W
DUAL_ROW_HEADER
332 1%
YELLOW TUBE WARM R95
0.125W
1%
332
0.125W
332
C 0.125W
3
1
6
5
4
2
LED<6..1> 1%
IN
CPLD_IO<33..0> 9B3<>
7D4<>
10B1<>
3
ON
2
1
OFF
7B2<> ON/OFF_KEY<3..0> 0
RS232<3..0> 9D7<>
J8
3
1
0
2
1 1 2 2 ISP_RXD
ISP_TXD 3 3 4 4
5 5 6 6 RTS
DTR 7 7 8 8
D 9 9 10 10
PIN
RIGHT_ANGLE
RIBBON_CABLE_LOW_PROFILE
GE Healthcare
Company Restricted
GND
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
1 2 3 4 5 6 7 8
TP7
POWER SUPPLY TIN
5V TO 3.3V
R116
1K
VCC 0.125W D15 TP9
DIRECTION 5275601-100, REVISION 2
2
R109 U9 VR2 1K
24V_IN 0.125W 3.16K LM2676 DS6 LT1529-3.3 0.125W
1% 0.1W
FL1 HSMG-C650 1%
ZJYS51R5-2P 0.1% GREEN
1 4 2 INPUT SWIT_OUT 1 C75 L1 5 VIN VOUT 1
1
6 FEEDBACK BOOST 3 4 SHDN_N SENSE 2
CR3 7 ON_OFF 33U
SM6T27A C69 C71 R110 GND 4 10N 20% R113 GND1 3 C83
27V 2 3 100U 0.1U 1K TAB_GND 8 50V D13 5A CR4 499 C78 C80 GND2 6 100U
50V 10% LSM340J SMBJ5.0
35V
10%
0.125W 0.125W
C85
22U
50V 10U 0.1U 16V
200 OHM 20% 10% 1% 40V 6.7V 1% 16V 50V 10%
+80%-20% 10% 3.3V
PDUGND TP5
B
TP4 24V TO 12V 1V8
12V
TIN R108 5V TO 1.8V R117
TIN
TP8
5.11K 1K
R111
0.125W 0.125W
2
9.09K U10
24V_IN 1% 0.1W VCC 1%
LM2676 DS7 D16
FL2 0.1% HSMG-C650 BAS116
ZJYS51R5-2P GREEN
1 4 2 INPUT SWIT_OUT 1 C76 L2
R118
1
6 FEEDBACK BOOST 3 75V
7 ON_OFF 33U 1.4K
C70 C72 R112 GND 4 10N 20% R114 VR1 0.1W
50V
Chapter 10 - Schematics
2 3 100U 0.1U 1K TAB_GND 8 D14 5A CR5 1K LT1963EQ 1%
50V 10% LSM340J P6SMB18A
35V
10%
0.125W 0.125W
C86
22U
50V
200 OHM 20% 10% 1% 40V 18V 1%
R115 2 VIN VOUT 4 10U
1 5
C84
0.1W
1%
0.1U
C77
50V 10U
TP10
C79
10% 16V
+80%-20%
12V TO LCD TIN
U25 J1
FL8
12V ZJYS51R5-2P GND
1 4 1 1 3 3 1 1 G1 G1
2 2 G2 G2 24V_IN
4 4 FL9 24V
ZJYS51R5-2P
2 3 2 2 5 5 1 4
PIN
RIGHT_ANGLE C81 C82
200 OHM CXA-L10L
GND C73 C74 WIRE_TO_BOARD 100U
OUTPUT HV 0.1U
4.7U 0.1U 2 3 50V 50V
50V 50V TWO PARTS AS CLOSE AS POSSIBLE 10% 20%
10% 10% SHIELD 200 OHM
D
PDUGND
GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
power
1 2 3 4 5 6 7 8
Page 331
LING LONG XR 6000 SERVICE MANUAL
Page 332
GE HEALTHCARE
1 2 3 4 5 6 7 8
ARM_JTAG<5..0>
IO
3V3
DIRECTION 5275601-100, REVISION 2
C88
50V
1% 1% 1% 1% 1% J5 10%
1 1 2 2
5 ARM_TRST 3 3 4 4
2 ARM_TDI 5 5 6 6
4 ARM_TMS 7 7 8 8
3 ARM_TCK 9 9 10 10
0 ARM_RTCK 11 11 12 12
1 ARM_TDO 13 13 14 14
ARM_RESET 15 15 16 16
IN
17 17 18 18
R123 19 19 20 20
10K
0.125W
1% PIN
STRAIGHT
DUAL_ROW_HEADER_BREAK
B GND GND
3V3
C
3V3
J2
2 2 1 1 CPLD_TCK 1
4 4 3 3 CPLD_TDO 0
6 6 5 5 CPLD_TMS 2
8 8 7 7
0.1U 10 10 9 9 CPLD_TDI 3
C87
50V
10%
PIN
STRAIGHT
DUAL_ROW_HEADER_BREAK
GND
CPLD_JTAG<3..0>
IO
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
1 2 3 4 5 6 7 8
3V3 U15
S1D13700F02
ARM_CON<9..0>
IN
COREVDD<2..0>
HIOVDD<2..0>
NIOVDD<1..0>
A
ADDR<23..0> 3 CS3 43 CS_N
DIRECTION 5275601-100, REVISION 2
IN
9 OE 41 RD_N ARM_IO<30..0>
IO
4 WE 42 WR_N
61 54 PO16_EINT0 DATA<31..0>
16
AS_N WAIT_N IO
0 ADDR<0> 16 A0 D0 52 DATA<0> 0
1 ADDR<1> 15 A1 D1 51 DATA<1> 1
2 ADDR<2> 14 A2 D2 50 DATA<2> 2
3 ADDR<3> 13 A3 D3 49 DATA<3> 3
4 ADDR<4> 11 A4 D4 47 DATA<4> 4
5 ADDR<5> 10 A5 D5 46 DATA<5> 5
6 ADDR<6> 9 A6 D6 45 DATA<6> 6
7 ADDR<7> 8 A7 D7 44 DATA<7> 7
8 ADDR<8> 6 A8 CYCLE TIME
9 ADDR<9> 5 A9
10 ADDR<10> 4 A10 FPDAT0 21 FPD<0> 14D4<
11 ADDR<11> 3 A11 FPDAT1 20 FPD<1> 14D4< FPSHIFT:CLOCK INPUT = 16 : 1
U13 12 ADDR<12> 2 A12 FPDAT2 19 FPD<2> 14D4<
74LVC-LCX14 13 ADDR<13> 64 A13 FPDAT3 18 FPD<3> 14D4<
14 ADDR<14> 63 A14
R134 15 ADDR<15> 62 A15
HCLK 6 B A 5 FPSHIFT 23 CP 14C4<
OUT
ARM_RESET 36 RESET_N XECL 24
33.2 IN
YSCL 29
0.125W 56 26 LOAD
VCC=3V3;GND=GND CNF0 FPLINE 14C4<
1% 57 30 FRM
B CNF1 FPFRAME 14C4<
Y2 U13 58 CNF2 MOD_OUT 27
3V3 74LVC-LCX14 59 CNF3 YDIS 31 DISP 14C4<
FL10 60 CNF4
BLM31A601S R135
1 2 4 VCC VOUT 3 3 A B 4 39 CLKI
600 OHM 1 2 33.2 35 XCG1
ENABLE GND 0.125W 34
VCC=3V3;GND=GND XCD1
C89 C90 C91 1%
R132
1U 10N 0.1U 1K 40MHZ 37 SCANEN
16V 50V 50V 0.125W 38 TSTEN VSS<4..0>
10% 10% 10% 1%
GND
Chapter 10 - Schematics
GND
GND
3V3
C 3V3
J14
1%
R125
10K
8 8
9 GND
0.125W
9
10 10
R133 11 11
CW FPD<3> 12
14B7> 12
14B7> FPD<2> 13 13
10K 14B7> FPD<1> 14 14
500MW 14B7> FPD<0> 15 15
10% 16 16
TP_LCDGND 17 17
18 18
1%
R136
15K
TIN 19 19 G1 G1
0.125W
20 20 G2 G2
D SOCKET
RIGHT_ANGLE
FLEX_CABLE_ZIF_BOTTOM
GND
GND SHIELD
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
Page 333
LING LONG XR 6000 SERVICE MANUAL
Page 334
GE HEALTHCARE
1 2 3 4 5 6 7 8
A
DIRECTION 5275601-100, REVISION 2
3V3
3V3
C97
0.1U
B 50V
U11 10%
24C256
R137 R138
10K 10K
0.125W 7 WP 0.125W GND
1% 1%
2 PO2_SCL 6 SCL
1 A0 SDA 5
2 A1
3 A2
VCC=3V3;GND=GND
GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
memory
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
3.4
GE HEALTHCARE
1 2 3 4 5 6 7 8
POWER SUPPLY
A TP1 12V TO 5V A
VCC
TIN
DIRECTION 5275601-100, REVISION 2
12V_IN
12V
R1
3.16K VARIOUS
0.125W U2
1% LM2676 R4
F1 FL1 511
FAST_ACTION ZJYS51R5-2P 0.125W
1 4 2 INPUT SWIT_OUT 1 C7 L1 1%
6 FEEDBACK BOOST 3
1A 7 ON_OFF 33U
63V 4 10N
GND 20%
2 3 TAB_GND 8 50V 5A
10%
2
C3 C4 R2
200 OHM 100U 0.1U 1K C9 DS1
35V 50V 0.125W D1 10U D2 HLMP-1790
20% 10% 1% LSM340J 35V MMSZ5235B GREEN
TP2 40V 10%
6.8V
TIN
1 POWER GOOD
B B
P
12V_GND
GND
Jedi Generator TAV Interface Board
ISOLATED 5V
TP3
Chapter 10 - Schematics
VCC RS232_VCC
C R3 TIN
C
U1 10K
DCR010505U 0.125W
1%
1 VS1 VOUT 14
2 VS2
ERROR_N 17
28 SYNCIN VREC 12
26 0VIN1 ENABLE 16
C1 C2 27 0VIN2 0VOUT 13
10U 0.47U C6 C8
35V 16V 47U 0.47U
10% 10% 5V C5 35V 16V
1U 10% 10%
25V TP4
10%
TIN
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
Page 335
LING LONG XR 6000 SERVICE MANUAL
Page 336
GE HEALTHCARE
1 2 3 4 5 6 7 8
VCC
RS232<1..2>
CONSOLE_RX+ 1
RS232_VCC U5 U7
C30 HCPL0720-060 MAX488E
0.1U C32 R17
U3 50V 0.1U 110
C11 C13 MAX3221E 10% 1 VDD1 VDD2 8 50V 1 VCC 0.1W
0.1U 0.1U 10% Y 5 0.1%
50V 50V RXD 2 VI VO 6 RS485_TX 3 DI
10% 10% 15 VCC Z 6 CONSOLE_RX- 3
3 VP 4 5 2 CONSOLE_TX+ 8 A
7 VN GND1 GND2 RO 2 RS485_RX
7 B
2 C1P R16 GND 4
4 C1N 110
5 C2P GND 0.1W
6 C2N 0.1%
C10 C12 RS232_GND
0.1U 0.1U TXD 11 T1IN T1OUT 13 RS232_TX 1 4 CONSOLE_TX-
50V 50V RS232_RX 8 R1IN R1OUT 9 RXD
10% 10% RS232_VCC VCC
1 EN_N INVALID_N 10
B 12 FORCEON FORCEOFF_N 16 B
GND
GND 14 C29
0.1U U6
50V HCPL0720-060 C31
10% 0.1U
2 RS232_RX 50V RS485<1..4>
8 VDD2 VDD1 1 10%
TXD 6 VO VI 2 RS485_RX
R6 R8 5 4
RS232_VCC 20K 0 GND2 GND1
0.125W 0.125W
1% +0.05
GND
U4 R11 U4
20K 1U 511 20K 1U
74HC123 74HC123 R18
0.125W 16V 0.125W 0.125W 16V 511
R5 R13
1K 1% 10% 1% 1K 1% 10% 0.125W
0.125W 14 CEXT 0.125W 6 CEXT 1%
1% 15 13 DS2 1% 7 5 DS3
REXT_CEXT Q REXT_CEXT Q
HLMP-1719 HLMP-1719
1 A_N 9 A_N
2 B Q_N 4 1 2 10 B Q_N 12 1 2
3 RD 11 RD
YELLOW YELLOW
VCC=VCC;GND=GND VCC=VCC;GND=GND
TX LED RX LED
D D
REV 2 01July04
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
1 2 3 4 5 6 7 8
RS485<1..4>
12V_IN
J1
1 1
1 CONSOLE_RX+ 2 2 14 14EXPOSURE_SW 2
2 CONSOLE_TX+ 3 3 15 15PREPARE_SW 3
4 4 16 16ON_SW 4
5 5 17 17CONSOLE_RX- 3
1 EXT_CAN_H 6 6 18 18CONSOLE_TX- 4
7 7 19 19EXT_CAN_L 5
1 JEDI_TX+ 8 8 20 20JEDI_TX- 3
2 JEDI_RX+ 9 9 21 21JEDI_RX- 4
B 4 JEDI_ON_SW 10 10 22 22JEDI_EXP_SW 6 B
7 JEDI_PREP_SW 11 11 23 23
5 EXT_CAN_L 12 12 24 24 EXT_CAN_H 1
2 RS232_RX 13 13 25 25 RS232_TX
G1 G1
G2
1
G2
RS232<1..2>
SOCKET
RIGHT_ANGLE
STD_DENSITY
Chapter 10 - Schematics
RS232_GND
C C
MH1 MH3
VCC VCC VCC VCC VCC VCC VCC VCC MHP4_50TB10_00 MHP4_50TB10_00
1
1
U8
74ACT00
1 A
Y 3
2 B C16 C17 C18 C19 C20 C21 C23 C24
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U C22
50V 50V 50V 50V 50V 50V 50V 50V 22N
VCC=VCC;GND=GND 10% 10% 10% 10% 10% 10% 10% 10% 250V
20%
R19 MH2 MH4
20K MHP4_50TB10_00 MHP4_50TB10_00
1
1
0.125W
1% GND SHIELD
GND
D D
REV 2 01July04
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
Page 337
LING LONG XR 6000 SERVICE MANUAL
Page 338
GE HEALTHCARE
1 2 3 4 5 6 7 8
CONSOLE_JEDI<1..7>
A 12V 12V A
K1 K2
DIRECTION 5275601-100, REVISION 2
G6RN_1_12VDC G6RN_1_12VDC
1 COIL1 1 COIL1
5 COIL2 5 COIL2
CLOSED 4 CLOSED 4
2 COM OPN 3 JEDI_EXP_SW 6 2 COM OPN 3 JEDI_PREP_SW 7
R21 D3 R25 D4
20K BAS116 20K BAS116
0.125W 75V 0.125W 75V
1% 1%
2 EXPOSURE_SW 3 PREPARE_SW
RS232_GND RS232_GND
U10
MC1489A
R24
U10 U8 4 6
B MC1489A 74ACT00 IN O B
15K
R22
0.125W RESP_CTRL
5
1 3 12 A C28
11 1% VCC=VCC;GND=GND
IN O Y 0.1U
15K 13 50V
B
0.125W RESP_CTRL
2
10%
1% C26 VCC=VCC;GND=GND
VCC=VCC;GND=GND
0.1U
50V
10%
GND
GND VCC
VCC
R20
1%
0.125W
392K
VCC=VCC;GND=GND 1%
0.125W
9 A_N
10 12 VCC=VCC;GND=GND
B Q_N
11 RD
C25
VCC=VCC;GND=GND
1U U9
16V 74HC123
10% GND
14 CEXT
15 REXT_CEXT Q 13
1 A_N
2 B Q_N 4
3 RD
VCC=VCC;GND=GND
PREPARE,EXPORSURE AND INDICATOR
D D
GND
REV 2 01July04
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
1 2 3 4 5 6 7 8
LING LONG XR 6000 SERVICE MANUAL
GE HEALTHCARE
DIRECTION 5275601-100, REVISION 2 LING LONG XR 6000 SERVICE MANUAL
3.5 I/F Board RAD Electrical
1 2 3 4 5 6 7 8
LVPS board
220P 330
CR19
DIRECTION 5275601-100, REVISION 2
2KV 1W
SSTA206S
10% 5%
A 600V A
CR9 ISLOCALPART=TRUE
+400_LINE
SMCJ43A CR20 CR3
50.3V SSTA206S DSEI12-12A
ZENERS_CHAIN 600V
0.25AMP V160DC
0.25AMP 1200V 0.75AMP
CR4 MID_ZD
2
SMCJ43A 11A
50.3V CR21 V160AC ISLOCALPART=TRUE V160DC 160V
R10
1W
5%
SMCJ43A
47K
ZENERS_CHAIN 0.75AMP L3 0.75AMP
1
0.25AMP 0.25AMP 0.25AMP 50.3V TAB
STARTUP_CHAIN CR8 ZENERS_CHAIN 1U STRAIGHT
SMCJ43A 2.5 TURN PIN
R33
2
50.3V CR16 CR22 CF4
ZENERS_CHAIN SMCJ43A C30 C6 C8 SMCJ43A 47K
R13
1W
5%
50.3V
47K
0.25AMP 100V LOW_VOLTAGE 1W
1
0.25AMP 0.75AMP 5% A1 1
2
2
50.3V SMCJ43A 20% 20% 20% CR18
0.25AMP 50.3V SMCJ43A
R20
GREEN
1W
5%
100V
47K
0.25AMP 0.25AMP ZENERS_CHAIN
1
1
HLMP-1790
A1 1
DCBUS_P STARTUP_CHAIN 0.25AMP CR15 0.75AMP
ZENERS_CHAIN SMCJ43A LOW_VOLTAGE
2
50.3V TAB
B B
2PIN L4 STRAIGHT
R21
R23
5%
5%
DCBUS_N ISLOCALPART=TRUE 1U PIN
100K
100K
0.5W
0.5W
1
0.25AMP R27 DCBUS_N 0.25AMP 2.5 TURN CF3
CR13
BAW56 0.75AMP
100 1 K1 LOW_VOLTAGE
0.1W 2 1 A 3 ISLOCALPART=TRUE
C25 1% 2 K2 T1 POWER_P15V
C23
CR2 10A P15V
0.1U 0.25AMP C19 ISLOCALPART=TRUE
50V NE1 DSS10-01A ISLOCALPART=TRUE
STARTUP_CHAIN 0.47U 75V ISLOCALPART=TRUE LOW_VOLTAGE L1 LOW_VOLTAGE
10% N76530 2 2 1 1
0.1U 50V
C7
25V
20%
NEON 50V 10% 1AMP 1U 1AMP
100U
C24 10% 22 19
19 100V R3
CF2 2.5 TURN
100P 1 OHM
PIN R31 C16
1MEG 50V 3 3 18 18 0.5MW
STRAIGHT R30 0.25AMP DCBUS_N ISLOCALPART=TRUE 0.47U
5% 5%
C4
C1
10%
50V
ISLOCALPART=TRUE
18V
0.1W
25V
20%
25V
20%
C12
TAB 0.25AMP
2.5A 20K LOW_VOLTAGE 1717 50V
1500U
1500U
0.47U
0.1W
CR12
SMCJ16A
Chapter 10 - Schematics
2X3.3MH 1% 1% U5 HEATSINK 16 1AMP 1.5AMP
1 A1 R26 UC3844A CR17
44 16
ISLOCALPART=TRUE 70V C31 C13
K2
5 5 15
15
BAV99 0.5AMP C 220P 0.47U
150K 7 8 2KV 50V
L2 VIN VREF R25 DCBUS_N
C 0.1W 4 6
AK
3 G 10% C
F1
FILTER RT_CT VOUT 14 14 10%
1% Q1
C3
C2
10%
50V
18V
1 2 2 1
25V
20%
25V
20%
C18
3 22 IRG4PH10U
CR7
SMCJ16A
1 A
5% 0.25AMP 0.25AMP LOW_VOLTAGE LOW_VOLTAGE
3 4 DCBUS_N SW_SNUB 9
9 12 12 ISLOCALPART=TRUE
ISLOCALPART=TRUE DCBUS_N 0.5AMP
R14
1W
100V 2 1
5%
330
F2
ISLOCALPART=TRUE
0.25AMP LOW_VOLTAGE
R38
R12
CF1 R24
20K 1 6 1 OHM
CR11 0.25AMP
1
6
R37
1W
1% 0.25AMP
1W
5%
330
2 3
0.1
R5
-400V_LINE
R4
1W
5%
330
REV 2 01July04
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.
1 2 3 4 5 6 7 8
Page 365
LING LONG XR 6000 SERVICE MANUAL
Page 366
GE HEALTHCARE
1 2 3 4 5 6 7 8
P15V
C11
0.1U
DIRECTION 5275601-100, REVISION 2
50V
A 10%
LOW_VOLTAGE A
LOW_VOLTAGE
C14 VCC
0.1U 0.25AMP
50V 0.25AMP
10%
LOW_VOLTAGE C9
0.1U
VCC VCC
0.25AMP 50V
M15V 10%
R6 R9 LOW_VOLTAGE
10K 2K
CR6 0.1W 0.1W 0.25AMP
BAW56 LOW_VOLTAGE 0.25AMP 1% 1%
0.5AMP LOW_VOLTAGE 0.25AMP R11 LOW_VOLTAGE R8 LOW_VOLTAGE
K1 1
LOW_VOLTAGE 0.25AMP 0.25AMP
FORWARD_MEASURE
3 A POWER_P15V
K2 2 10K 100K
R18 LOW_VOLTAGE 0.25AMP
0.1W 0.1W
2
2K LOW_VOLTAGE
0.1W 1% VCC 1%
75V DS1
0.25AMP 1% 0.25AMP HLMP 1709 R2 J1
C10 LOW_VOLTAGE ISLOCALPART=TRUE 1K
0.47U R16 0.1W
RED
1
B 50V 2K R15 P15V VCC
LOW_VOLTAGE U2 1% 1 1 2 2 B
8
10% 0.1W 20K MAX485 3 3 4 4
1% 0.1W U3 5 6
VP
LM311 0.25AMP 5 6 LOW_VOLTAGE
LOW_VOLTAGE 1% 2 7 7 8 8
0.25AMP LOW_VOLTAGE 7 _MAINS_DROP 4 DI RO 1 9 9 10 10
R17 3 VOUT 3 DE 11 11 12 12
1K C17 0.25AMP LOW_VOLTAGE 6 A 13 13 14 14
0.1W 0.1U 0.25AMP 7 B 15 15 16 16
1 GND
1%
4 VN
6
5 BAL
50V 2 RE_N 17 17 18 18
10% C15 19 19 20 20
BAL_STR
R19 0.1U 21 21 22 22
1K 50V VCC=VCC;GND=GND_CHASSIS 23 23 24 24
0.1W 10% 25 25 26 26
1% LOW_VOLTAGE
PIN
STRAIGHT
M15V RIBBON_CABLE_STANDARD
C C
P15V
1AMP
LOW_VOLTAGE LOW_VOLTAGE
U4
0.25AMP MC78M05CDT VCC
R1
1 VIN VOUT 3
10 C22 4 C21
GND
1W
0.47U 0.1U
5% 50V 50V
10% 5V 0.75AMP 10%
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.
1 2 3 4 5 6 7 8
Section 1.0
Scope
This section provides detailed information about each MIS run in the form of individual MIS cable
charts.
Each of the following charts defines a single interconnection in the system. The charts are arranged
in the order, according to MIS numbers. In some cases, a particular MIS number is repeated, except
for a letter added at the end. Such a set of charts use the same cable, and carry essentially the
same signals, however the applications for each chart is different - the point of connection is
different (at one or both ends) for each chart in the set.
Section 2.0
How to Use
Each MIS cable shown on the MIS map has its own MIS chart. Each chart identifies all signals
present on the cable, and the specific pins associated with each signal. See Figure 11-1 for an
example.
Section 3.0
MIS Charts
3.1 System Console
Handswitch J1
COM
1 1
PREP
2 2
COM
3 3
EXP
4 4
J2 Serial Port
Receive Data
2 to console
Transmit Data
3 from console
Signal Ground
5 GND
J2 Serial Port
Receive Data
2 to console
Transmit Data
3 from console
Data Terminal Ready
4 from console
Signal Ground
5 GND
Request to Send
7 from console
TM3 J3
1 DC24V
1 Console Power Input
4 E
4
5 Power Issue A
5
Power Issue B
6 6
PSW
7 7 RTL (0-DC24V; Active: 0V)
J4 J1
1 12V_GND
1
9 RS485_RX+ 2
10 RS485_TX+ 3
11 12V_GND
4
12 12V_IN
5
CAN_H
13 6
Exposure_SW
14 14
Prepare_SW
15 15
ON_SW
17 16
RS485_RX-
22 17
RS485_TX-
23 18
CAN_L
24 19
JEDI TAV
J9 J1
11 RS232-GND
7
10 RS485_TX+ 8
9 RS485_RX+ 9
17 ON_SW
10
15 Prepare_SW
11
CAN_L
24 12
RS485_TX-
23 20
RS485_RX-
22 21
Exposure_SW
14 22
CAN_H
13 24
JEDI TAV
J8 J1
3 TX
25
2 RX 13
5 RS232-GND 23
PDU Generator
380 VAC
TM1 Un P1
380 VAC
TM1 Vn P2
380 VAC
TM1 Wn P3
PDU Generator
E
Grouding Bar G
3.3 Table
TM1 XP105
1 1
2 2
3 3
4 4
XP106 1 5
XP106 2 6
PDU Table
E
Grouding Bar PE1
E
PE1 PE2
E
PE2 PE3
E
PE5 PE6
24 V
5 3
0V
6 4
foot-1
7 8
24 V
8 2
0V
9 1
24V
1 1
0V
2 2
XP105 3
Collimator Connector 4
Section 1.0
Compliance Statement
This equipment complies with IEC60601-1-2 Edition 2 (2001) EMC standard for medical devices.
This equipment generates, uses, and can radiate radio frequency energy. The equipment may
cause radio frequency interference to other medical and non-medical devices and radio
communications.
To provide reasonable protection against such interference, this product complies the radiated
emission as per CISPR11 Group1 Class A standard limits.
However, there is no guarantee that interference will not occur in a particular installation.
If this equipment is found to cause interference (which may be determined by turning the equipment
on and off), the user (or qualified service personnel) should attempt to correct the problem by one
or more of the following measure(s):
• Reorient or relocate the affected device(s)
• Increase the separation between the equipment and the affected device (see recommended
separation distances)
• Power the equipment from a source different from that of the affected device
• Consult the point of purchase or service representative for further suggestions
Use of accessories, transducers, cables and other parts other than those specified by the
manufacturer of this equipment may result in increased emissions or decreased immunity of the
equipment.
The manufacturer is not responsible for any interference caused by using other than recommended
interconnect cables or by unauthorized changes or modifications to this equipment. Unauthorized
changes or modifications could void the users' authority to operate the equipment.
All interconnect cables to peripheral devices must be shielded and properly grounded, except when
technologically prohibited. Use of cables not properly shielded and grounded may result in the
equipment causing radio frequency interference.
The LING LONG XR 6000 is predominantly intended for use in non-domestic environments, and
not directly connected to the Public Mains Network that supplies buildings used for domestic
purposes.
The compatible accessories must be used within the recommended operating conditions outlined
in the operation manuals. In addition to calibration and warm-up, other devices must be reset before
and after use to ensure accurate dose measurements. Sustained exposure to electromagnetic
fields (exceeding the test conditions) may cause false measurements. Failure to follow the
recommended use may cause false measurements.
The magnetic field environment from a MRI device located nearby is a risk of interference.
All of the above are required to achieve the Electromagnetic Compatibility for a typical installation
of the LING LONG XR 6000. Further detailed data & requirements are described in the Use
Recommendations and Installation Recommendations sections.
Section 2.0
Compatibility Tables
This equipment complies with IEC60601-1-2 Edition 2 (2001) EMC standard for medical devices.
The LING LONG XR 6000 is suitable to be used in an electromagnetic environment, as per the
limits & recommendations described in the tables hereafter:
• Emission Compliance level & limits (Table 1).
• Immunity Compliance level & recommendations to maintain equipment clinical utility (see
Table 2 and Table 3).
Note: This system complies with above mentioned EMC standard when used with supplied cables.
If different cable lengths are required, contact a qualified service representative for advice.
Guidance and Manufacturer’s declaration - electromagnetic immunity. The LING LONG XR 6000 is suitable for use in
the specified electromagnetic environment. The customer or the user of the LING LONG XR 6000 should assure that
it is used in an electromagnetic environment as described below.
Immunity Test IEC 60601-1-2 Compliance Electromagnetic Environment Guidance
Test Level Level
Conducted RF, 3 Vrms, 150 kHz to V1 = 3Vrms Portable and mobile RF communications equipment
IEC 61000-4-6 80 MHz should be used no closer to any part of the LING LONG
Radiated RF 3 V/m, 80 MHz to E1 = 3V/m XR 6000, including cables, than the recommended
IEC 61000-4-3 2.5 GHz separation distance calculated from the equation
appropriate for the frequency of the transmitter.
Recommended Separation Distance
d = 1.2
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the LING LONG XR 6000 is used exceeds
the applicable RF compliance level above, the LING LONG XR 6000 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the LING
LONG XR 6000.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects, and people.
Table 3 LING LONG XR 6000 Electromagnetic Immunity (Continued)
Recommended separation distances between portable and mobile RF communications equipment and the LING
LONG XR 6000
The LING LONG XR 6000 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the LING LONG XR 6000 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the LING LONG XR 6000 as recommended below, according to the maximum output power of the communications
equipment.
Rated Maximum Output Power (P) Separation distance according to frequency of transmitter
of Transmitter Watts (W)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
100 12 12 23
For transmitters rated at a maximum output power not listed above, the separation distance can be estimated using
the equation in the corresponding column, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer. V1 and V2 are the COMPLIANCE LEVELS for the IEC 61000-
4-6 test and E1 is the COMPLIANCE LEVEL for the IEC 61000-4-3 test. V1 and V2 are in V and E1 is in V/m.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Table 4 Recommended Separation Distances
Section 3.0
Use Recommendations
This product complies with IEC 60601-1-2 Edition 2 (2001) EMC standard for medical devices and
with radio frequency emission requirements per CISPR11 Group1 Class A standard limits. The
LING LONG XR 6000 is predominantly intended for use in hospitals.
Do not use devices which intentionally transmit RF Signals (Cellular Phones, Transceivers, or
Radio Controlled Products) in the vicinity of this equipment as it may cause performance outside
the published specifications. Keep the power to these types of devices turned off when near this
equipment.
Adhering to the distance separation recommended in Table 4, between 150 kHz & 2.5 GHz, will
reduce disturbances recorded at the image level but may not eliminate all disturbances. However,
when installed and operated as specified herein, the system will maintain its essential performance
by continuing to safely acquire controlled radiological X-ray exposures in a mobile radiography
environment.
Section 4.0
Installation Recommendations
This system complies with above mentioned EMC standard when used with supplied cables.
In order to minimize interference risks, the following requirements shall apply.
Installation Date:
cancel
F.E. Name: First PM Date:
check
Nota : After filling, this PM report shall be inserted in the system Logbook on site, use a new PM paper report each year
Time
Assembly Frequency 1 Year 2 Year 3 Year 4 Year 5 Year 6 Year 7 Year 8 Year
(H )
Replace NonVolatile RAM on the kV Control
To be replace 8 years 0.5
Board (2159419)
System General
Maintenance Item Frequency Time 1 Year 2 Year 3 Year 4 Year 5 Year 6 Year 7 Year 8 Year
Gene GENERATOR
Maintenance Item Frequency Time 1 Year 2 Year 3 Year 4 Year 5 Year 6 Year 7 Year 8 Year
dates Comments
Intervention Date:…………………………………
Intervention Date:…………………………………
FE Sign Off:…………………………………