Académique Documents
Professionnel Documents
Culture Documents
CASE STUDY
SPE Method Development for Pesticides
OBJECTIVE
The objective of this work was to develop a suitable SPE method for pesticide standards, to include
Aldicarb Pestanal Standard, Aldicarb-Sulfone Pestanal Standard and Oxamyl Pestanal Standard.
CONCLUSIONS
The resulting LCMS chromatograms showed good retention of the compounds and revealed that
methanol with 100 mM ammonium acetate is an effective elution solvent. 94% or greater recovery was
observed for all three pesticides using methanol with 100 mM ammonium acetate. Acetonitrile with 10
mM ammonium acetate also exhibited good results, with 90% or greater recovery in the first elute step.
www.jordilabs.com Page 1
Final Report
Company Name
Date:
Released by:
Dr. Mark Jordi
President
Jordi Labs LLC
Client Name
Company Name
Address
Please find enclosed the test results for your samples described as:
Objective
The objective of this work is to develop a suitable SPE method for the pesticide standards
provided.
Screening runs of the samples by GCMS showed significant degradation of the target
compounds preventing accurate quantitation. We therefore used LCMS to analyze the
starting, load, wash and elute solutions prepared by SPE.
The resulting LCMS chromatograms showed good retention of the compounds and revealed
that methanol with 100 mM ammonium acetate is an effective elution solvent. 94% or greater
recovery was observed for all three pesticides using methanol with 100 mM ammonium
acetate. Acetonitrile with 10 mM ammonium acetate also exhibited good results, with 90% or
greater recovery in the first elute step.
For analytes that are easily ionized, LCMS can provide improved detection because the
sample is not vaporized at high temperature prior to chromatographic separation. We
therefore continued the analysis using LCMS.
LCMS
The SPE starting, load, wash, and elute solutions were analyzed by LCMS in order to
determine their SPE retention behavior. The chromatogram of the SPE starting solution can
be found in Figure 1. The three standards were found to be adequately separated using a
gradient from 5% organic to 100% organic.
Aldicarb-
Sulfone
o-aminoazotoluene
Aldicarb
Oxamyl
Figure 1: LCMS chromatogram of the SPE starting solution identifying the three
standard peaks
The relative peak areas for each standard are provided in Table 1. Recovery values are
provided as well, and are based on the peak area relative to the starting solution.
Table 1: Standard Peak Areas and % Recovery Values – Methanol SPE
Nearly 100% of the standards were retained on the loading step, and only a small percentage
was observed to elute with additional washing. A majority of the sample (≥ 90%) was
recovered during the first elute step.
For this analysis, approximately 40 ppm o-aminoazotoluene was added to each solution as an
internal standard. The peak areas in the above table are scaled based on the o-
aminoazotoluene response.
Analysis Conditions
This section of a Jordi report provides information on the methods used including
instrument type, temperatures, solvents, sample preparation, etc. The specific conditions
have been removed for this case study.
Closing Comments
Deformulation of an unknown material is intended to provide a best estimate of the chemical
nature of the sample. All chemical structures are supported by the evidence presented but are
subject to revision upon receipt of additional evidence. Additional factors such as material
processing conditions may also affect final material properties.
Jordi Labs’ reports are issued solely for the use of the clients to whom they are addressed. No
quotations from reports or use of the Jordi name is permitted except as authorized in writing.
The liability of Jordi Labs with respect to the services rendered shall be limited to the amount
of consideration paid for such services and do not include any consequential damages.
Jordi Labs specializes in polymer testing and has 30 years experience doing complete
polymer deformulations. We are one of the few labs in the country specialized in this type of
testing. We will work closely with you to help explain your test results and solve your
problem. We appreciate your business and are looking forward to speaking with you
concerning these results.
Sincerely,
Mark Jordi
Mark Jordi, Ph. D.
President
Jordi Labs LLC