REX SERIES

TROUBLE SHOOTING

This instruction is to provide you with guide line of trouble shooting


flow on REX series.

LISTEM Corporation
1- REX T/S
DISPLAY TEST (1)

POWER SUPPLY 1) Put Demo IC into U1 on Main Console B/D.


Connection 2) Connect the power of Power supply to J5 on Main Console B/D.

* How to adjust RAD LCD DISPLAY


1) Adjust the brightness of characters in Left LCD(Small one) by
RAD. LCD adjusting VR1 on LCD PCB. (Contrast Adjustment-Precise
DISPLAY Operation adjustment required.)
2) Adjust the brightness of Left LCD(Small one) by adjusting VR4 on
LCD PCB. (Back Light Adjustment-Medium level adjustment
required.)
3) Adjust the brightness of characters in Right LCD(Big one) by
adjusting VR2 on LCD PCB. (Contrast Adjustment-Precise
adjustment required.)
4) Adjust the brightness of Right LCD(Big one) by adjusting VR3 on
LCD PCB. (Back Light Adjustment-Medium level adjustment
required.)

* How to adjust FLUORO LCD DISPLAY


1) Adjust the brightness of characters in Left LCD(Small one) by
FLUORO. LCD adjusting VR1 on LCD PCB. (Contrast Adjustment-Precise
DISPLAY adjustment required.)
2) Adjust the brightness of Right LCD(Big one) by adjusting VR2
on LCD PCB. (Contrast Adjustment-Precise adjustment
required.)

A) Check the condition of solder of LCD power line and mis-insertion


of circuit elements.
B) Replace LCD, U1,U2 and VR1,2,3,4

(A) RAD LCD DISPLAY (B) FLUORO LCD DISPLAY

2- REX T/S
MEMBRANE TEST (2)

Switch Operation 1) Check the condition of Membrane switch by pressing each


switch on Membrane panel.
2) Check the condition of LED Lamp operation when pressing
each switch.
NO

A) Replace Membrane sheet after checking of switch condition.


B) Replace Main Console B/D.
C) Replce CN1,CN2 and FLAT CABLE

Repetitive Operation 1) Check again LCD PCB B/D.


2) Check again Main Console B/D.

<Radiography Console> <R/F Console>

3- REX T/S
RELAY2 BOARD TEST (3)
1) Normal Operation of K10 onRELAY2 B/D.
Turn on the Power
A) Check if input voltage is 12VAC on J4-5,6
NO (K9,K10 Operation voltage)
B) Check if input voltage is 220VAC on J4-1,2.
C) Check if output voltage is 220VAC on J4-3,4
(Control Operation voltage)
D) Check mis-inserted elements.(BD1,TLP5604)
E) Replace K10 Relay.(12VAC)

1) Point of contact of K10 comes off when K9 on RELAY2 B/D


Turn off the Power is operating.
NO A) Check the condition of point of contact of Membrane switchs
B) Replace K9 Relay(12VAC)

During Display 1) After operation of K8, K2 on RELAY2 B/D, K1, K2 on Main B/D is
Transmitting operating with K4 operation.

NO A) Check if input voltage is +24VDC on J3-1,2.


(Relay operation voltage)
B) Check if MPOWER of J1-5 become GND.
C) Replace Main console B/D.
D) Replace CN1,CN2 & FLAT CABLE

1) When selecting BUCKY1 : operation of K4 2) When selecting BUCKY2 : operation of K5


3) When exposing X-ray : operation of K3 4) When power on : operation of K7
5) When switching from Small to Large : operation of K8
* When it is not working as the above : Output a signal from J7 on the H.T Control.
Usually it is output the approved voltage 24VDC. When the above 1)~5) is selected on
Display Control, it is dropped to 0VDC(GND).
* Check whether U20(ULN2803), U14(82C55), U9(TIBPAL) and program IC U3(27C512) is operating
normally or not and, if necessary, replace them by new one.

4- REX T/S
RELAY1 BOARD TEST (4)

During DISPLAY 1) Operating in order K3(Pre charge)---K1(Main)---K2(Charge) and


TRANSMITTTING after then, K3 Relay comes off.
2) R-Series : K1(ON) – MG1(ON) – K3(ON) – K2(ON) –MG2(ON) –
K3(OFF)
3) RF-Series : K1(ON), K3(ON) – MG1(ON) – K2(ON) – MG2(ON)
* When K3 doesn’t work : Check whether J1-5 has GND.
Check whether J8-5 on H.T Control has +24VDC and then, GND.
* When K1 doesn’t work : Check whether J1-3 has GND.
Check whether J8-3 on H.T Control has +24VDC and then, GND.
* When K2 doesn’t work : Check whether J1-4 has GND.
Check whether J8-4 has +24VDC and then, GND.

NO A) Check wiring condition of J1 on RELAY1 B/D.


B) Check Diode polarity & mis-insertion.
C) Check input voltage of J1(Check if J1-3,4,5 become GND.)
D) Check if input voltage is +24VDC on J1-1,2.
E) Check if polarity of J1-1,2 is reversed.

5- REX T/S
POWER BOARD TEST (5)

1) Put GND of Tester device on CON1-1 on Power B/D.


During DISPLAY 2) Put another probe on the point you want to check.
TRANSMITTTING 3) Adjust VR7 so that the voltage between CON1-3 can be +5VDC.
4) Adjust VR2 so that the voltage between CON1-7 can be +12VDC.
5) Adjust VR4 so that the voltage between CON1-5 can be -12VDC.

POWER BOARD A) Check input voltage of CON1.


Voltage Adjustment (20VAC,0VAC,20VAC // 15VAC,0VAC)
B) Check the condition of F1,F2 Fuse in Power B/D.
C) Check if there is any misinserted elements.
NO D) Check variable point of VR1 & VR4 is located in center.
E) In case +5VDC,+12VDC,-12VDC voltage become low
after turning the power.
: Check the condition of element on Filament B/D, HT control
B/D and whether there is any short-circuits.

6- REX T/S
FILAMENT BOARD TEST (6)

Voltage measurement 1) Insert the GND of tester into J9-4.


2) Put the rest terminals of tester to J4-3(filament board)
3) When programming at display console, about +18VDCis output.
When finishing the program, about +16VDC is maintained
(According to the length of HT cable, the voltage is changed a little
by a basis on 200mA)
Wave measurement 4) Locate the GND terminal of scope on GND point in PCB and the rest
terminal on heat sink of Q1 and Q2(IRF630)
5) Adjust variable resistance R1 to 10KHZ.
6) adjust the variable resistance R4 and then save not to be shape of
spherical wave and peak

NO A) Check if NF1 input voltage is 50 VAC


B) Check if D1 output voltage and J2-1,3 is +60VDC
C) Check if F1,F2 fuse is disconnected or not.
D) Turn the variable resistance R38 to the right(2/3 location)
E) Replace U3,4,5,(TL082)
F) In case the voltage between GND-J4 No.1 outputs +60~+70VDC
* Caution : DO NOT “EXPOSURE”
1> Check if Q7 is damaged, and then replace it.
2> Check if ARRAY TR PCB is defective, and then replace it.
3> Check if the PCB lag of Connector J4-1,2,3 has bad contact or
disconnection
4> Check if the HV cable contact connected to “HV transformer-
Tube-Change box” is bad (HV Cable Bushing should be
combined with HV transformer bushing tightly)

7- REX T/S
H.T CONTROL TEST (7)

1) LED1,2,3,4 is Off for a moment and On


Power On (Indicates the state of CPU RESET)
2) When the program starts process, LED 1 is on a light.

NO
A) Check if the parts are inserted wrong.
B) Check if Q4(2N2907A),Q5(2N222A) is defective and replace it
C) Check the output signal wave in the No 15 pin of U1(80C552) is

IPM , FEED BACK okay and replace it.

Voltage and frequency D) Check if the power board output voltage(+5VDC,+12VDC,-

measurement 12VDC) is okay and the fuse is disconnected.


1) Set the frequency between GND-TP9 to 30KHZ by adjusting VR 1
(Tube 1)
2) Set the frequency between GND-TP9 to 10KHZ by adjusting VR 2
(Tube 2- Fluoro)
3) Set the voltage between GND-TP2 to 3.0VDC by adjusting VR 3
(Tube 1)

8- REX T/S
A) Check U38(TL082) and replace it
NO

1) Set the voltage between GND-TP17 to 1VDC(after READY


ROTOR signal Voltage status-LED3 on) by adjusting VR 13
adjustment(ROTSEN) 2) Before READY(Rotor sensing) finish, make sure the voltage
should not be more than 5V.(approx.3V). After Rotor sensing,
the voltage will be approx.1VDC.

A) Check the voltage of TP17(after READY), and adjust VR13.


B) Check the CT1 operation of the main power board and

NO the connector contact badness


C) Check the ROTOR input voltage (Ready 220VAC input /

Filament Signal Voltage after ready – 60VAC input)

adjustment (FILSEN) Warning! : In case of High speed system such as


REX550,650,850, fix ROTOR input voltage to 220VAC)
D) Check if F7 fuse is disconnected
E) Check if the Rotor line W,B,G is disconnected and wrong
wiring

1) All process will be taken by CPU automatically, no need to adjust.

A) Check the U5(TL082) of filament board and replace it.

NO B) U22-1,7 should be LOW

9- REX T/S
1) Adjust READY TIME in “Rotor Delay” setting mode of control
READY console.(See Parameter setting mode section)
Time Adjustment
2) Adjust READY TIME of Spot Radiography in “Spot Delay” setting
mode of control console. (See Parameter setting mode section)

READY FINISH
1) When READY button lights on control console, ROTOR sensing is
finished.(Ready status-LED 3 on).

NO A) Place a tester to TP17 to check ROTOR output voltage


B) And then, increase output voltage to 1VDC by adjusting VR13 during
operating READY button. (Counter-clockwise : Increase)

10- REX T/S


IPM DRIVE PCB, IPM GATE WAVE TEST (8)
1) Inset Drive PCB into IPM DRIVE PCB test tool
IPM DRIVE TEST 2) Connect J1,J2 connectors.
3) Put the tester on the end of GND and +15VDC and set + 15VDC by
adjusting VR1 and VR2 (In case of REX125 and 325, no need to adjust
NO it. Different PCB applied for both models.)
4) Check if LEC is switching in turn by pressing switches.
A) When +15VDC is not be output, check D1,D2(DF06) U1,U2(LM317)
and replace it.
B) When LED is not be switching in turn, check PC1,2,3,4 and replace it.
C) Replace PC(HCPL4503) in the side that the switching is inferior

1) When Espousing, there must be feed back wave at GND and +KV,-
KV,+mA,-mA
2) If the wave is not measured, check it as follows.
A) The drive signal must be output at U38-9,10 as follows.
Repeat Operation

B) The drive signal must be output at U38-9,10 as follows.

11- REX T/S


* When the A) wave is not be output at EXPOSURE
1) Check if KOUT,EXP,KSY signal is input
2) Check U38(TL594) and replace it.
NO
* When the B) wave is not be output at EXPOSURE
1) Check U37(AD201G) and replace it.
2) Check U41(ULN2004) and replace it.
3) Check IPM DRIVE PCB and replace it.
4) Check IPM and replace it.
5) Check HV transformer and replace it.

<HT control B/D>

12- REX T/S


TERMINAL FUNCTION AND VOLTAGE MEASUREMENT(9)

Function and Voltage output – R-Series

1) TB2-1,2 : When selecting Bucky1 on Display Console and pressing Ready, 110VAC
voltage is output between the terminal 1 and 2.
* When it doesn’t work : follow the check instructions of Relay 2 Board Test.
2) TB2-3,4 : When selecting Bucky1 on Display Console and pressing Ready, 110VAC
voltage is output between the terminal 3 and 4.
* When it doesn’t work : follow the check instructions of Relay 2 Board Test.
3) TB2-5,6 : 24VAC voltage is output to operate the collimator.
* When it doesn’t work : check whether F1 fuse on Main Power Board is disconnected or not
and, if necessary, replace it by new one.
4) TB2-7,8 : 30VDC LOCK voltage is output to operate Stand, Table and Bucky Stand.
* When it doesn’t work : check whether F1 fuse on Main Power Board is disconnected or not
and, if necessary, replace it by new one.
Check out the operation of MG1. (LOCK voltage is approved with the point of contact of
MG1)
5) TB2-7,8,9,10 : 110VAC voltage is output to operate Stand and Table.
* When it doesn’t work : Check out the operation of MG2. (LOCK voltage is approved with
the point of contact of MG2)
6) TB2-11,12 : It is supplied Power to operate the board of the optional photo-fluorography
and so on. (The voltage to operate photo-fluorographic relay on Interface board)
7) TB2-15 : It is Exposure Signal and for photo-fluorography.
8) TB2-16,17,18 : It is output the voltage to operate the rotor of tube.
* When it doesn’t work : follow the check instructions of Rotor signal voltage adjustment on
H.T control Board Test(7).
9) TB2-19,20 : It is Temp Sensor and detects the overload of tube – This is optional and
when connecting, it works normally. ( Usually the contact point of thermometer sensor in
the tube is connected to the rotor wire and when overloading of tube, it is treated the rotor
error.)

Function and Voltage output – RF-Series

1) TB2-1,2,3 : It is output the voltage to operate the rotor of tube2.(Tube under the table –
for Fluoroscopy)
* When it doesn’t work : follow the check instructions of Rotor signal voltage adjustment on
H.T control Board Test(7).
2) TB2-4,5,6 : When selecting Tube 2, it is supplied Power to operate I.I, Power Supply,
Monitor, Camera and so on to be inflicted the voltage.

13- REX T/S


3) TB2-7,8 : It is Thermometer Sensor and detects the overload of tube – This is optional
and when connecting, it works normally. ( Usually the contact point of thermometer sensor
in the tube is connected to the rotor wire and when overloading of tube, it is treated the
rotor error.)
4) TB2-9,10 : When selecting Bucky1 on Display Console and pressing Ready, 110VAC
voltage is output between the terminal 9 and 10.
* When it doesn’t work : follow the check instructions of Relay 2 Board Test.
5) TB2-11,12 : When selecting Bucky2 on Display Console and pressing Ready, 110VAC
voltage is output between the terminal 11 and 12.
* When it doesn’t work : follow the check instructions of Relay 2 Board Test.
6) TB2-13,14 : 24VAC voltage is output to operate the collimator.
* When it doesn’t work : check whether F1 fuse on Main Power Board is disconnected or not
and, if necessary, replace it by new one.
7) TB2-15,16 : 30VDC LOCK voltage is output to operate Stand, Table and Bucky Stand.
* When it doesn’t work : check whether F1 fuse on Main Power Board is disconnected or not
and, if necessary, replace it by new one.
Check out the operation of MG1. (LOCK voltage is approved with the point of contact of
MG1)
8) TB2-17,18,19,20 : 110VAC voltage is output to operate Stand and Table.
* When it doesn’t work : Check out the operation of MG2. (110VAC voltage is approved with
the point of contact of MG2)
9) TB2-21,22 : It is supplied Power to operate the board of the optional photo-fluorography
and so on. (The voltage to operate photo-fluorographic relay on Interface board)
10) TB2-23 : It is Exposure Signal and Signal Wire for photo-fluorography and the other
options.
8) TB2-24,25,26 : It is output the voltage to operate the rotor of tube 1.(Tube on the
tubestand – for Radiography)
* When it doesn’t work : follow the check instructions of Rotor signal voltage adjustment on
H.T control Board Test(7).
9) TB2-27,28 : It is Thermometer Sensor and detects the overload of tube – This is optional
and when connecting, it works normally. ( Usually the contact point of thermometer sensor
in the tube is connected to the rotor wire and when overloading of tube, it is treated the
rotor error.)

14- REX T/S


REX-R SERIES CALIBRATION METHOD (2)

Membrane Switch Function

(그림변경)

1) POWER OFF Switch 2) POWER ON Switch


3) Small Focus LED Indicator 4) Large Focus LED Indicator
5) Over Load LED Indicator 6) Error LED Indicator
7) Kvp Increase Switch 8) Kvp Decrease Switch
9) mA Increase Switch 10) mA Decrease Switch
11) mAs / sec Increase Switch 12) mAs/sec Decrease Switch
13) Density Increase Switch (no use) 14) Density Decrease Switch (no use)
15) Tube1 Selection Switch 16) NO Bucky Selection Switch
17) Bucky1 Selection Switch 18) Bucky2 Selection Switch
19) Tube2 Selection Switch 20) mAs / sec Change Switch
21) Pre-exposure (Ready) Switch 22) Exposure (x-ray) Switch
23) Pre-exposure (Ready) LED Indicator 24) Exposure (x-ray) LED Indicator

15- REX T/S


B) Connect DIVIDER to H.T.T and Tube.
C) Connect DISPLAY and Communication cable of DIVIDER
D) Put the power to D-3.

* KV Offset : ROM DATA (no change) - - - - + , - RAM DATA (changeable)


* mA Offset : ROM DATA (no chagne) - - - - + , - RAM DATA (changeable)
* Exp.mA : mA feedback wave can be flat. ( change only plus(+) side )

16- REX T/S


kV & mA CALIBRATION

CALIBRATION SETTING
1) Power ON
2) Follow the above Section 6.3.1 procedure
Calibration mode will start.
3) Adjust kVp value to 40kVp by using kV UP/DOWN button, mA value to 200mA by using mA UP/DOWN
button and time to 0.05sec.
4) Start calibration on the basis of 200mA
5) Prepare the calibration data sheet.

CALIBRATION BY DYNALYZER lll (D-lll)

1) To calibrate by using D-lll, connect Display and Divider of D-llI as below


- Connect with Divider between HT transformer and tube.
- Connect with the correspondence cable between Display and Divider.
- Supply the power to D-lll.

* kV Offset : ROM DATA(Non-changeable) ----- +, - RAM DATA(Changeable)


* mA Offset : ROM DATA(Non-changeable) ----- +, - RAM DATA(Changeable)
* EXP mA : mA feedback wave can be flat.( possible to change to plus(+) side only)
(Factory Adjustment Only)

17- REX T/S


Fixed Changeable

(1) Kvp DATA Increase Switch (2) Kvp DATA Decrease Switch
(3) mA DATA Increase Switch (4) mA DATA Decrease Switch
(5) EXP mA Increase Switch (6) EXP mA Decrease Switch
(7) DATA Value Save Switch

Calibration with
Oscilloscope

18- REX T/S


REX-R/F SERIES Calibration Method (4)

Membrane Switch Function

Left LCD

Right LCD

1)POWER OFF Switch 2) POWER ON Switch


3) Small Focus LED Indicator 4) Large Focus LED Indicator
5) Over Load LED Indicator 6) Error LED Indicator
7) Kvp Increase Switch 8) Kvp Decrease Switch
9) mA Increase Switch 10) mA Decrease Switch
11) mAs / sec Increase Switch 12) mAs/sec Decrease Switch
13) Density Increase Switch (no use) 14) Density Decrease Switch (no use)
15) Tube1 Selection Switch 16) NO Bucky Selection Switch
17) Bucky1 Selection Switch 18) Bucky2 Selection Switch
19) Tube2 Selection Switch 20) mAs / sec Change Switch
21) Pre-exposure (Ready) Switch 22) Exposure (x-ray) Switch
23) Pre-exposure (Ready) LED Indicator 24) Exposure (x-ray) LED Indicator
25) Fluoro Kvp Increase Switch 26) Fluoro Kvp Decrease Switch
27) Fluoro mA Increase Switch 28) Fluoro mA Decrease Switch
29) Fluoro Timer Reset Switch 30) Image Enlargement Normal Switch(no use)
31) Image Enlargement MAG1 Switch(no use) 32) Image Enlargement MAG2 Switch(no use)
33) Automatic Brightness Control( ABC Switch(no use))

19- REX T/S


Calibration Setting Method

A) Connect DISPLAY and DIVIDER of D-3 to calibrate with D-3


like below.

B) Connect DIVIDER to H.T.T and Tube.


C) Connect DISPLAY and Communication cable of DIVIDER
D) Put the power to D-3.
* KV Offset : ROM DATA (no change) - - - - + , - RAM DATA (changeable)
* mA Offset : ROM DATA (no change) - - - + , - RAM DATA (changeable)
* mA -COMP : mA feedback wave can be flat. ( change only plus(+) side )

20- REX T/S


(A) TUBE1 RAD CALIBRATION MODE (B) TUBE2 RAD CALIBRATION MODE

Calibration with
an Oscilloscope

Feedback terminal

A) Put Oscilloscope’s GND on PCB’s GND.


B) Place Oscilloscope’s terminal on +KV , -KV, and press 22 switch in the middle of pressing
21 switch of the right LCD to check KVP(10KVP per 1V)
C) Place Oscilloscope’s terminal on +mA , -mA, and press 22 switch in the middle of
pressing 21 switch of the right LCD to check mA(10mA per 1V)

21- REX T/S


* TUBE2 FL CAL MODE CALIBRATION METHOD ( FLUORO MODE)

J1-9:FLU mode calibration


On / Off selection terminal

(C) TUBE2 FL(FLUORO) CALIBRATION MODE

22- REX T/S


Calibration with
an Oscilloscope

Feedback terminal

A) Put Oscilloscope’s GND on PCB’s GND.


B) Place Oscilloscope’s terminal on +KV , -KV, and press 22 switch in the middle
of pressing 21 switch of the right LCD to check KVP(10KVP per 1V)
C) Place Oscilloscope’s terminal on +mA , -mA, and press 22 switch in the middle of
pressing 21 switch of the right LCD to check mA ( 2mA per 0.1V)

23- REX T/S


Error Message & Troubleshooting

* REX-R(F) SERIES
CODE ERROR Checking point
HVC connecting point. Filament board defect, and refer TEST
Filament value is LOW Error in
01 A setting voltage of U22-3. Output voltage of U22-1 in H.T
preheating
Output voltage of U11-2, and replace if faulty
After Ready, Filament value is Low Filament board. U22,30,25 in H.T. Replace if faulty
02
Error
HVC connecting point. Filament board defect, and refer TEST
Filament value is HIGH Error in
03 A setting voltage of U22-6. Output voltage of U22-7 in H.T
preheating
Output voltage of U11-2, and replace if faulty
After Ready, Filament value is High Filament board. U22,30,25 in H.T. Replace if faulty
04
Error
Rotor driven voltage is 220VAC,60VAC input and output voltage
CT1 and fuse 7 cutting in MAIN POWER BOARD
05 ROTOR ERROR
A setting voltage of U27-3,12. Output voltage of U27-1,14 in H.T
U27-2,13 ROTSEN Voltage input in H.T
Checking main fuse1,2 cutting. MG1,MG2 operation checking
06 CHARGE ERROR Program down. P2 connector dislocation in CHARGE CPB
Part broken(D1,D2,U1,Resistance) in CHARGE PCB
No relevance (TEMP SENSOR built in Tube)
07 OVER HEAT(TEMP SENSOR) ERROR
Disconnection of TEMP wire on TB1 terminal
Connecting point between J1 and J2 of communication line
08 COMMUNICATION ERROR
Defect of TIBPAL IC and program IC
09 KV LOW ERROR
10 KV HIGH ERROR
Checking IPM DRIVE PCB voltage. Replace it if faulty
Checking the feedback wave of +KV,-KV,+mA,-mA
00 IPM FAULT ERROR
Checking U37,38,42 IC in H.T
Defect of high voltage PCB inside H.T.T
Checking light of LED1,2,3,4 on H.T control.
In case the program does not go on (check CPU RESET is possible.)
without ERROR code. Checking defect of Q4,Q5 unit in H.T control
Break of CPU(U1) unit in H.T control
Fuse1, fuse2 cutting in POWER BOARD
Unit broken or short in Filament board

24- REX T/S


CALIBRATION WHEN REPLACING THE FILAMENT BOARD WITH MAS METER(6)

* When the filament board is defective and new one is replaced, follow the below steps.

1. Each board has some differences of voltage. Accordingly, refer to the below standard
datasheet when replacing filament board by new one.

* In case of High Voltage Cable HVC-6M (Test Point : between GND and the connector J4-3)

mA step Initial value Ready (Preparation)


25mA 13.84VDC 29.5VDC
50mA 13.84VDC 32.0VDC
75mA 13.84VDC 33.5VDC
100mA 13.84VDC 34.5VDC
150mA 13.74VDC 39.0VDC
200mA 13.74VDC 40.7VDC
250mA 13.74VDC 41.9VDC
300mA 13.74VDC 43.0VDC
400mA 13.74VDC 44.3VDC
500mA 13.74VDC 45.5VDC

* The above values may have some differences in accordance with the length of High Voltage
Cable or the condition of calibration.

2. With reference of the above values, make measure the voltage of the old filament board
in advance and write down the values. Replace the old board by new one.

3. Test Point on PCB of the replaced board : Make place the tester between GND and J4-3
and set up the display condition to 200mA/70kV. When the ready, adjust its value to the same as
the voltage of PCB on the old filament board with the pot resistor ‘R60’.

25- REX T/S


* Follow the above steps to obtain the more precise mA/mAs values.
(mA, mAs Calibraion with mAs Meter)

1. Make preparation mA/mAs Meter.


2. Disconnect the wires and connect the terminal of mAs Meter to the below connector (A).

3. Select the display condition to 200mA/70kV/20mAs.


4. Make reset mAs meter and press the exposure switch. Check whether the established
mAs value is displayed. Otherwise, adjust its value to the same as the established
20mAs on the display with the pot resistor ‘R60’ of the filament board.
5. Make establishment each mAs(each mA X 0.1sec) as the above steps. Adjust mA Data
on the display console and make calibration as REX-R SERIES CALIBRATION(2) and
REX-R/F SERIES CALIBRATION(4)