Data from CARESCAPE R860 and external lab results are used to provide the following values:
• PAO2 – Alveolar partial pressure of oxygen
• AaDO2 –Alveolar arterial oxygen difference
• Pa/FiO2 – Oxygenation index
• PaO2/PAO2 – Alveolar arterial oxygen pressure difference
• CO – Cardiac output
• OI – Oxygen Index
• SpO2/FiO2 – Saturation FiO2 Ratio
• Vd/Vt – Dead space ventilation
• Vd – Dead space volume
• VA – Alveolar ventilation
Tube Compensation
• Type of compensation: Electronic tube compensation
• Compensation for: Endotracheal and tracheostomy tubes
• Tube diameter: 5 to 10 mm
• Level of compensation: 25 to 100%
Non-Invasive Ventilation (NIV) (optional)
• Mask ventilation: Yes
• Integrated unique leak recognition algorithm
Automatic Patient Detection (APD)
• Patient re-connection: Automatic detection in standby
• Detection by: Back pressure to Bias-flow
Lung Protection Strategy:
Lung Protection strategy: an integrated approach
The lung is a delicate structure that is highly sensitive. Mechanical ventilation can be detrimental
if too much volume or pressure is delivered. And patients affected by lung disease are highly
susceptible to ventilation-associated lung injury.1
A Lung Protection strategy can help ensure that different zones receive the treatment they need
The CARESCAPE R860 was designed with a simple and universal navigation scheme. The Lung
Protection tools calculate FRC, providing the clinician with data for determining the optimal
PEEP setting for the patient.