7000 Series Respiratory Humidifier


(PMH7000)

Specification Sheet
Modes

Available modes IPPV Preset airway temp 40°C, chamber temp 37°C
NPPV For NIV use, airway temp 34°C, chamber temp 31°C
Manual User adjustable temperature settings:

Patient side: 30°C to 40°C (increments 1°C)


Chamber side: 26°C to 43°C (increments 1°C)
Set Temperature Range

Patient Side 30°C to 40°C (increments 1°C)


Chamber Side 26°C to 43°C (increments 1°C)
Temperature setting of patient side vs. chamber side temperature is in relationship within the range of -4°C to +3°C.
Temperature Display

Three digit 14mm 7 segment LED (Light Emitting Diode) display.


Patient Side Temperature Display Range 10°C to 45°C increments 0.1°C
Chamber Side Temperature Display Range 10°C to 55°C increments 0.1°C
Alarms

Patient Temperature High. Exceeds set temperature by 2°C or >43°C

Low; 4°C below set temperature


Chamber Temperature High. Measured temperature exceeds set temperature by 4°C
for more than 20 mins, or the measured temperature exceeds set
temperature by 10°C

Low. Measured temperature is below set temperature by 4°C


for more than 20 mins, or the measured temperature is below set
temperature by 10°C
Temperature Probes Faulty, Disconnected
Heated Wire Faulty, Disconnected
Heater Plate Measured chamber temperature does not rise by more than 2°C
in 20 minutes from start up
Physical Dimensions

Size (Without Humidification Chamber Installed) Height - 145mm / Width - 140mm / Depth – 180mm
Weight 1.7 kg. (approx.)
Intersurgical
7000 Series Respiratory Humidifier
(PMH7000)
Power Requirements
REF Operating Voltage Line Frequency
7000000 230V ~ +/- 10% 50Hz
7000001 100V ~ +/- 10% 50 / 60Hz
7000002 230V ~ +/- 10% 50Hz

Power Consumption 250 VA Maximum


Current Rating 230V - 1.1A 100V - 2. 5A
Heater Plate Power 150 Watts
Heater Wire Power Inspiratory 44 Watts (Max. DC24V, 1.8A), regulated.
Expiratory 44 Watts (Max. DC24V, 1.8A), regulated.
Heater Plate Temperature 100°C Thermostat regulated (normal operation).
130°C Thermostat limited (AC power disabled).
Heater Wire Resistance 11 to 50 Ω (ohms)
MDD Classification IIb
Electric Shock Classification: Class I, type B Applied Part
Protection against Water Classification: IPX1
Mode of Operation Classification: Continuous Operation
Environmental

Operating Temperature 10°C to 40°C (50°F to 104°F)


Recommended Ambient Temperature 20°C to 26°C (68°F to 79°F)
Storage Temperature -20°C to 60°C (-4°F to 140°F)
Operating Relative Humidity 30 to 75% Non-condensing
Storage Relative Humidity 10 to 95% Non-condensing
Atmospheric Pressure 700 to 1060 hPa
Standards Applied

JIS Medical Electrical Equipment, JIS Electrical Devices, ISO8185, UL60601-1 (CAN/CSA-C22.2 No. 601.1-M90),
IEC60601-1, IEC60601-2.
REF Product Description
7000000 7000 Series Respiratory Humidifier (UK)
7000001 7000 Series Respiratory Humidifier (JPN)
7000002 7000 Series Respiratory Humidifier (EU1): [ fr, es, it, de, nl ]
5601000 Electrical adaptor lead (Dual heated wire system)
5600000 Electrical adaptor lead (Single heated wire system)
7700000 Dual temperature probe assembly
Manufactured by:
Pacific Medico Co. Ltd, 2-6-4 Hongo, Bunkyo-ku, Tokyo 1133-0033 Japan
Issue 3: TS 02.13
Distributed by:
United Kingdom (Head Office): Crane House, Molly Millars Lane, Wokingham, Berkshire, RG41 2RZ
T: +44 (0)118 9656 300 F: +44 (0)118 965 6356 info@intersurgical.com | www.intersurgical.com
France Deutschland España Pocc South Africa Japan
info@intersurgical.fr info@intersurgical.de info@intersurgical-es.com info@intersurgical.ru info@intersurgical.co.za info@intersurgical.co.za
Portugal Nederland Lietuva Česká Republika Philippines
info@intersurgical.pt info@intersurgical.nl info@intersurgical.lt info@intersurgical.cz info@intersurgical.ph

Vous aimerez peut-être aussi