Sound Level Meter Model UT351 and UT352 (hereafter the surface of the Meter when servicing. No abrasive
Conforms to Standards of European Union
UT351/352
referred to as “the Meter”) is a stable, safe and reliable and solvent should be used to prevent the surface of the
sound level meter. The Meter is suitable to use in noise Meter from corrosion, damage and accident.
control, quality control, health care and all different kind of Constantly check the battery as it may leak when it has The Meter Structure (see figure 1)
SOUND LEVEL METER environmental noise testing. For example: factory, road, been using for some time, replace the battery as soon
as leaking appears. A leaking battery will damage the
family, musical instrument and all kind of places which
OPERATING MANUAL need noise testing. Meter.
Unpacking Inspection
Open the package case and take out the Meter. Check
the following items carefully to see any missing or
damaged part:
1 2 3
B.Display Symbols (see figure 2) No. Meaning Side Panel (see figure 3) Measurement Operation and Functional
1 Data Store is full Buttons
2 Date and Time display Below table indicated for information about the functional
3 Data Store button operations
14 13 12 11 10 9 4 Decibel Button Operation Performed
5 C-Weighting HOLD Turn the Meter on and off. Press once
15 8 to turn the Meter on. Press and hold for
6 A-Weighting
around 1 second to turn the Meter off.
7 Sound value display
16 7 8 Range display
HOLD feature: During sound measure
ment, press once to freeze the current
9 Over range reading in the display. Press the button
17 6 10 Slow response again to resume normal operation.
11 Fast response External DC6V A/C During sound measurement, press
5
18 12 Low battery display DC Output Terminal
A/C button to select “A” or “C” frequency
13 Data Hold is on weighting. With “A” weighting selected,
4 the frequency response of the Meter is
19 14 Auto power off enabled
AC Output Terminal
similar to the response of the human ear.
3 15 Under range CAL potentiometer “A” weighting is commonly used for envi
16 Analogue bar graph display
20 ronmental or hearing conservation
17 Symbol of Sound Pressure Level programs ‘C” Weighting is a much flatter
2 response and is suitable for the sound
18 Auto ranging enabled
21 22 23 1 19 Date display figure 3
level analysis of machines, engines, etc.
Most noise measurements are performed
20 Time display using “A” Weighting and SLOW
21 Maximum value display 1. DC Output Terminal: DC analogue signal output.
Response.
Output impedance is around 100Ω (10mV/dB)
figure 2 22 Minimum value display For the Model UT352, press A/C button
2. AC Output Terminal: AC analogue signal output.
23 Data Store enabled also could recall and clear data:
Output impedance is around 600Ω (0.707V/ each
range scale) RECALL:
Press and hold this button, the LCD
3. CAL potentiometer: Calibration
displays the last data stored and the
4. External DC6V:Using power adaptor DC6V, output
index number.
plug ( 3.5) to plug in the terminal. It can use 4pcs
Press button to recall additional
of 1.5V batteries or power adaptor to power up the stored reading.
Meter Press button to exit the RECALL
mode
CLR:
Press and hold A/C button when turning
on the Meter until the LCD displays CLR
and RECORD. All the data stored will
be cleared.
4 5 6 7
8 9 10 11
A. Decibel dB Maintenance
This section provides basic maintenance information Warning
Function Range Resolution Accuracy Remarks including battery replacement instruction. To avoid false readings replace the battery as
30~80dB soon as the battery indicator appears.
Frequency Response Warning To replace battery:
A-Weighting and 50~100dB
Do not attempt to repair or service your Meter unless 1.Press the Meter power off
C-Weighting 0.1dB ±1.5dB
60~110dB 31.5~8kHz you are qualified to do so and have the relevant 2.Remove the screw from the battery compartment, and
calibration, performance test, and service information. then take out the battery door from the battery compart
80~130dB
To avoid electrical shock or damage to the Meter, do ment.
FAST Sampling Time: 125 microsecond not get water inside the case. 3.Remove the battery from the battery compartment.
Sampling Rate
4.Replace the battery with 4pcs new 1.5V AA batteries
SLOW Sampling Time: 1 second
In order not to affect the Meter accuracy or damge to 5.Rejoin the battery door and the battery compartment,
1 dB per scale, the Meter, do not open the Meter housing and install the screw
Analogue Bar Graph 30~130dB 1dB Sampling time: 200 times per second