Vous êtes sur la page 1sur 52

Ohaus Corporation

19A Chapin Road


P.O. Box 2033,
Pine Brook, NJ 07058-2033

MB45 MOISTURE ANALYZER

SERVICE MANUAL

Ohaus Corporation 19A Chapin Road, P.O. Box 2033, Pine Brook, NJ 07058-2033 (973) 377-9000
SERVICE MANUAL

MB45 MOISTURE ANALYZER

On/O
ff
Tes
t Me
nu
Dis
play MB
Setu 45
p

STA
RT/S
TO
P

Ente
r
Prin
t
TAR
E

Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ 07932-0900

The information contained in this manual is believed to be accurate at the time of


publication, but Ohaus Corporation assumes no liability arising from the use or misuse
of this material. Reproduction of this material is strictly prohibited.

Material in this manual is subject to change.

© Copyright 2006 Ohaus Corporation, all rights reserved.


® Registered trademark of Ohaus Corporation.
TABLE OF CONTENTS

CHAPTER 1 INTRODUCTION
Page
1.1 Introduction .......................................................................................... 1-1
1.2 Service Facilities .................................................................................... 1-1
1.3 Tools and Test Equipment Required ...................................................... 1-2
1.3.1 Special Tools .................................................................................. 1-2
1.3.2 Tools and Supplies ........................................................................ 1-2
1.4 Test Mass Required ................................................................................ 1-2
1.5 Specifications ......................................................................................... 1-2
1.5.1 Admissable Ambient Conditions ..................................................... 1-2

CHAPTER 2 OPERATION
2.1 Menu Structure .................................................................................. 2-1
2.2 Control Switch Functions................................................................... 2-2
2.3 Adjusting the Display Contrast and Brightness ................................. 2-3

CHAPTER 3 TROUBLESHOOTING
3.1 Troubleshooting ...................................................................................... 3-1
3.2 Troubleshooting Guide ........................................................................... 3-1
3.2.1 Diagnosis ........................................................................................ 3-1

CHAPTER 4 MAINTENANCE PROCEDURES


4.1 Preventive Maintenance ......................................................................... 4-1
4.1.1 Visual Inspection ............................................................................ 4-1
4.1.2 Preventive Maintenance Checklist.................................................. 4-1
4.2 Testing .......................................................................................... 4-2
4.2.1 Operational Test ............................................................................. 4-2
4.2.1.1 Display Test ...................................................................... 4-2
4.2.2 Weight Calibration .......................................................................... 4-3
4.2.3 Temperature Calibration ................................................................. 4-4
4.2.4 Performance Tests .......................................................................... 4-5
4.2.4.1 Precision Test ................................................................... 4-5
4.2.4.2 Off-Center Load Test ......................................................... 4-6
4.2.4.3 Centered Load Test ........................................................... 4-6
4.2.4.4 Visual Inspection ............................................................... 4-6
4.3 Repairing the Moisture Analyzer ............................................................. 4-6
4.3.1 Repairing the Drying Unit ............................................................... 4-7
4.3.1.1 Dismantling the Heating Module ....................................... 4-7
4.3.2 Replacing the Protective Glass and Reflector................................. 4-8
4.3.2.1 Replacing the Protective Glass and Reflector ................... 4-8
4.3.2.2 Installing Protective Glass and Reflector ........................... 4-8
4.3.3 Replacing the Halogen Lamp ......................................................... 4-8
4.3.3.1 Dismantle Halogen Lamp .................................................. 4-8
4.3.3.2 Installing the Halogen Lamp .............................................. 4-9
4.3.4 Replacing the Temperature Controller ............................................ 4-9
4.3.5 Replacing the Temperature Sensor .............................................. 4-10

i
TABLE OF CONTENTS (Cont.)
CHAPTER 4 MAINTENANCE PROCEDURES (Cont.)

4.3.6 Installing the Heating Module ....................................................... 4-10


4.3.7 Removing Housing Top ................................................................ 4-11
4.3.8 Removing the Membrane Keypad ................................................ 4-13
4.3.9 Installing the Membrane Keypad .................................................. 4-13
4.3.10 Replacing the Display ................................................................... 4-14
4.3.11 Replacing the Fan ........................................................................ 4-14
4.3.12 Replacing the CPU Board ............................................................ 4-15
4.3.13 Replacing the Power Supply Board .............................................. 4-16
4.3.14 Replacing the Weighing Cell ........................................................ 4-17
4.3.15 Replacing the Housing Base ........................................................ 4-18
4.3.16 Installing the Housing Top .............................................................. 4-19
4.4 Resetting the Thermal Overload Device ................................................ 4-20
4.5 Final Work Steps ................................................................................... 4-21

CHAPTER 5 DRAWINGS AND PARTS LISTS


5.1 Drawings and Parts Lists ........................................................................ 5-1
5.1.1 Top Housing ................................................................................... 5-2
5.1.2 Bottom Housing .............................................................................. 5-4
5.1.3 Dryer Unit ........................................................................................ 5-7

APPENDIX A SOFTWARE SERVICE TOOL INSTRUCTIONS


A.1 Software Service Tool Instructions ...................................................... A-1
A.2 Requirements ...................................................................................... A-1
A.3 Instructions and Notes ......................................................................... A-1
A.4 Error Messages ................................................................................... A-4

ii
TABLE OF CONTENTS (Cont.)

LIST OF TABLES

TABLE NO. TITLE PAGE NO.

1-1 Specifications ................................................................................. 1-2


2-1 Menu structure ................................................................................ 2-1
3-1 Diagnostic Guide ............................................................................ 3-2
4-1 Types of Performance Tests ........................................................... 4-5
5-1 Spare Parts List Top Housing ......................................................... 5-3
5-2 Spare Parts List Bottom Housing .................................................... 5-5
5-3 Spare Parts List Dryer Unit ............................................................. 5-7

LIST OF ILLUSTRATIONS

FIGURE NO. TITLE PAGE NO.

4-1 Access for Cleaning MB45 Moisture Analyzer ................................ 4-1


4-2 Display Test .................................................................................... 4-2
4-3 Dismantling the Heating Module ..................................................... 4-7
4-4 Removing the Heating Module ....................................................... 4-7
4-5 Replacing Protective Glass ............................................................ 4-8
4-6 Replacing Halogen Lamp ............................................................... 4-8
4-7 Installing the Halogen lamp ............................................................ 4-9
4-8 Replacing Temperature Controller.................................................. 4-9
4-9 Replacing Temperature Sensor .................................................... 4-10
4-10 Installing Heat module .................................................................. 4-10
4-11 Replacing Cover ........................................................................... 4-11
4-12 Remove Housing .......................................................................... 4-11
4-13 Remove Cable from Display ......................................................... 4-12
4-14 Remove Connectors ..................................................................... 4-12
4-15 Remove Membrane Keypad ......................................................... 4-13
4-16 Install Membrane Keypad ............................................................. 4-13
4-17 Replacing Display ........................................................................ 4-14
4-18 Replacing Fan .............................................................................. 4-14
4-19 Replacing CPU Board .................................................................. 4-15
4-20 Replacing Power Supply Board.................................................... 4-16
4-21 Replacing Weighing Cell .............................................................. 4-17
4-22 Replace Housing Base ................................................................. 4-18
4-23 Install Housing Top ....................................................................... 4-19
4-24 Plug in Cables .............................................................................. 4-19
4-25 Housing Top Reassembly .............................................................. 4-20
4-26 Resetting the Thermal Overload Device ........................................ 4-20

5-1 Top housing .................................................................................... 5-2


5-2 Bottom Housing .............................................................................. 5-4
5-3 Dryer Unit ........................................................................................ 5-6
iii
iv
CHAPTER 1 INTRODUCTION

1.1 INTRODUCTION
This service manual contains the information needed to perform routine maintenance and service on the Ohaus MB45
Moisture Analyzer.

The contents of this manual is contained in five chapters.

Chapter 1 Introduction - Contains information regarding service facilities, tools and test equipment, measuring
masses, and specifications.

Chapter 2 Operation - Contains information on the control functions of the Moisture Analyzer.

Chapter 3 Troubleshooting - Contains a diagnosis/diagnostics chart and error code table.

Chapter 4 Maintenance Procedures - Contains preventive maintenance procedures, performance tests and
adjustments, disassembly/replacement procedures, and calibration procedures.

Chapter 5 Drawings and Parts Lists - Contains exploded views of all assemblies identifying serviceable
components with parts lists.

Before servicing the Moisture Analyzer, you should be familiar with the Instruction Manual which is packed with every unit.

1.2 SERVICE FACILITIES


To service the Ohaus MB45 Moisture Analyzer, the service area should meet the following requirements:

DO NOT SERVICE the Moisture Analyzer:

• Next to open windows or doors causing drafts or rapid temperature changes.

• Near air conditioning or heat vents.

• Near vibrating, rotating or reciprocating equipment.

• Near magnetic fields or equipment that generates magnetic fields.

• On an unlevel work surface.

• Allow sufficient space around the instrument for ease of operation and keep away from radiating heat sources.

On/O
ff
Test
Menu
Displa
y MB
Setup 45

STAR
T/STO
P

Enter

Print
On/O
ff TARE
Test
Menu
Displa On/O
y MB ff
Setup 45 Test
Menu
Displa
STAR y MB
T/STO Setup 45
P

STAR
T/STO
P

Enter

Print
TARE Enter

Print
TARE

1-1
CHAPTER 1 INTRODUCTION

1.3 TOOLS AND TEST EQUIPMENT REQUIRED


This section contains information on tools and supplies required to properly service the MB45 Moisture Analyzer. These items
are listed as follows:

1.3.1 Special Tools


None required.

1.3.2 Tools and Supplies


1.Digital Voltmeter (DVM) - Input impedance of at least 10 megohms in the 1 Volt dc position.
2.Standard electronics tool kit
3.Soldering iron (50 watt) and solder (rosin core solder, not acid core).
4.Solder remover.
5.Ohaus Temperture Calibration Kit P/N 11113857
6.Soft, lint-free cleaning cloth
7.Anti-static bags for PCBs
8.Anti-static wrist strap and mat

1.4 TEST MASS REQUIRED


It is recommended that a ASTM Class 1 Tolerance mass be used. The mass required is 20 grams.

1.5 SPECIFICATIONS
Complete specificatons for the MB45 Moisture Analyzer are listed in Table 1-1. When an MB45 has been serviced, it must
meet the specifications listed in Table 1-1. Before servicing the Moisture Analyzer, determine what specifications are not
met.

TABLE 1-1. SPECIFICATIONS


Capacity 45g
Readability 0.001g, .01%
Heat Source Halogen bulb - 120 V or 240 V
Calibration External calibration mass-20g
Dimensions (DxWxH) (in/cm) 14 x 7.5 x 7/ 35.5 x 19 x 17.8
Operating Temperature Range 10°C to 40°C
Power Requirements 100V ac -127V ac or 230V ac - 240V ac, 50/60Hz
Weight (lb/kg) 9.8 / 4.5
Shipping Weight (lb/kg) 14.1 / 6.4

1.5.1 Admissible Ambient Conditions


Use only in closed rooms
Temperature range: 5 ºC to 40 ºC
Atmospheric humidity: 80% rh @ to 30 ºC
Voltage fluctuations: –15% +10%
Installation category: II
Pollution degree: 2
Power load: Max. 450 W during drying process
Current consumption: 4 A or 2 A, according to the heating element
Power supply voltage: 100 V – 120 V or 200 V – 240 V, 50/60 Hz
(the voltage is given by the heating element)
Power line fuse: 1 piece, 5 x 20 mm, T6,3 H 250 V

1-2
CHAPTER 2 OPERATION
2.1 MENU STRUCTURE
This chapter contains menu structure, control swich operation and display adjustments. For complete operation, refer to
the Instruction manual supplied with the unit. The menu structure is illustrated in the table below.

TABLE 2-1. MENU STRUCTURE


SETUP TEST MENU
Test Library
Setup
Edit Test
Lockout
01 Test Parameters
Weight Calibration
Load Test Test ID:
Temperature Calibration
Edit Test Profile:
Language
Delete Test Standard
English, Deutsch, Espaniol,
02 Fast
Francais, Italiano
Load Test Ramp
Beeper
Edit Test Step
Low, High, Off
Delete Test Dry Temp:
Time-Date
• Switch Off:
Format
• Timed- Switch Off
MM/DD/YR, DD/MM/YR
• Manual-Use Start/Stop
Set Date
• AFREE - Wt loss/time
Time format
50 A90 - <1MG in 90 seconds
12HR, 24HR
Load Test A60 - <1MG in 60 seconds
Set Time
Edit Test A30 - <1MG in 30 seconds
AM, PM
Delete Test Result:
RS232
% Moisture
Baud Rate
Custom Units
1200, 2400, 4800, 9600,
Grams
19,200
% Regain
Parity
% Solids
Odd, Even, None
Custom
Data Bits
Factor, Exponent, LSD
7, 8
Target Wt:
Stop Bits
Print Int:
1, 2
Off, 1, 3, 5, 10, 30 Sec
Handshake
1, 2, 5 Min.
None, XON-XOFF, RTS-CTS
Save Test:
Print
Print On/Off
GLP On/Off
Display
Contrast
Brightness
Factory Reset

2-1
CHAPTER 2 OPERATION
2.2 CONTROL SWITCH FUNCTIONS

1 6

On/Off 7
2 0%

Test Menu
8
Display

3
On/
Off
Test

9
Men
u
Setup Enter Print Dis
play MB
Setu 45
p

STA
RT/
STO
P

4 START/STOP TARE
10
Ente
r
12
11
Prin
t
TAR

5
E

11

No. Designation Function

1 Turns ON/OFF display and fan. ON returns with last screen. (All modes)

2 Test Menu button When pressed, selects Test Library screen.

3 Display button When pressed, will return to the current test or cycles through several display
screens (run mode).

4 Setup button When pressed, brings up analyzer setup menu.


5
5 Start/Stop button When pressed, either starts or stops drying process.

6 button When pressed, travels up through menu options and selects alphanumeric
characters.

7 button When pressed, travels to the left through displays.

8 button When pressed, travels to the right through displays.

9 button When pressed, travels down through menu options and selects alphanumeric
characters.

110 Enter/Print button When pressed, accepts selection/input (select mode) or sends print command
during a test (run mode).

11 Tare button When pressed, performs a tare function. Pressing during a test has no effect.

12 Leveling feet Used to level the Moisture Analyzer.

2-2
CHAPTER 2 OPERATION
2.3 ADJUSTING THE DISPLAY CONTRAST AND BRIGHTNESS
Enter the ANALYZER SETUP menu, this menu permits adjusting the contrast and brightness levels on the display screen.
To change the levels, proceed as follows:

ANALYZER SETUP Press the Setup button. Using the down arrow button, scroll to DISPLAY. Press
LANGUAGE the Enter button.
BEEPER
TIME-DATE
RS-232
PRINT
DISPLAY

Adjusting contrast
DISPLAY ADJUST
Using the up/down arrow buttons, adjust the screen contrast to the desired level.
CONTRAST: 60
The numeric value (0 to 100) is for reference only. Press the right arrow button
BRIGHT: 10
to advance to BRIGHT. Default = 40.
EXIT WITHOUT SAVING

Adjusting brightness
DISPLAY ADJUST
Using the up/down arrow buttons, adjust the screen brightness to the desired
CONTRAST: 60
level. The numeric value (0 to 10) is for reference only. Press the Enter button.
BRIGHT: 10
Default = 10.
EXIT WITHOUT SAVING

Exit without saving


DISPLAY ADJUST
If you do not wish to save settings, Use the right arrow button and scroll to EXIT
CONTRAST: 60
WITHOUT SAVING and press the Enter button. The display returns to
BRIGHT: 10
ANALYZER SETUP.
EXIT WITHOUT SAVING

2-3
CHAPTER 2 OPERATION

2-4
CHAPTER 3 TROUBLESHOOTING

3.1 TROUBLESHOOTING
This section of the manual specifies problem areas of the MB45 Moisture Analyzer which can occur. Information is contained
to isolate specific problems using the Troubleshooting Guide. Make certain that the work area is clean and use care when
handling components of the Analyzer.

3.2 TROUBLESHOOTING GUIDE


A Troubleshooting Guide is designed to help you locate the problem area quickly and easily. First locate the problem that
you are observing. The probable causes are listed with the most common cause first. Before attempting to repair the
Analyzer, read all chapters of this manual to familiarize yourself with the components and operation. Do not attempt repairs
unless you fully understand the operation of the Analyzer.

3.2.1 Diagnosis

1.Isolate and identify the symptom.

2.Review the probable causes in the order that they appear.

3.See the appropriate section of the manual to correct the malfunction.

4.Repair or replace the defective section of the Moisture Analyzer.

NOTE:
If more than one symptom is observed, it is necessary to
approach one area at a time, and also remember, that the
symptoms may be interrelated.

In the event that erratic or fluctuating weight readings are observed, it is necessary to isolate the problem to either the
mechanical area or the electronic area of the Moisture Analyzer.

If a problem arises that is not covered in this manual, contact:

Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ 07058-2033 USA
Tel: 973-377-9000
Fax: 973-944-7177

In the United States call toll free, 800-526-0659 between 8:00 a.m. and 6:00 p.m. EST.

3-1
CHAPTER 3 TROUBLESHOOTING
3.2 TROUBLESHOOTING GUIDE (Cont.)

TABLE 3-1. DIAGNOSTIC GUIDE


SYMPTOM PROBABLE CAUSE(S) REMEDY

Moisture analyzer cannot be switched Power cord not plugged in or no power Check
on. supply voltage.
The power supply is designed for
Moisture analyzer cannot be switched Defective main fuse. 100...240 VAC, but the drying unit only
on green LED on power supply board for 110 or 230VAC. Check whether the
does not light up. Defective power supply fuse. correct drying unit is installed. Check
the fuses on the back of the instru-
ment and the power supply board.

Moisture analyzer cannot be switched Cable connections inside instrument not Plug in or replace cable.
on. Green LED on power supply plugged in or defective.
board lights up.
Defective CPU board. Replace CPU board.

Moisture analyzer cannot be switched Defective power supply fuse. Replace power supply fuse on power
on green LED on power supply board supply board.
does not light up. Defective power supply board. Replace power supply board.

Some buttons do not respond. Defective keypad wiring. Replace membrane keypad.

Keypad does not function. Keypad not plugged in Check keypad.

Defective keypad wiring Replace membrane keypad.

Portions of the display do not light up. Defective display. Replace display.

Halogen lamp does not switch on Thermal overload device has tripped. Reset thermal overload device, see
after "Start". paragraph 4.4.

Defective temperature controller. Replace if resistance after resetting is


> 1 Ohm.
Defective halogen lamp. Replace if resistance > 20 Ohms
(check whether 100 or 230 VAC).

Temperature sensor not plugged in or Plug in or replace sensor.


defective.

Defective power supply board. Replace power supply board.

Defective CPU board. Replace CPU board.

Halogen lamp switches off before end Incorrect Halogen bulb Check voltage rating on bulb.
of test.

Temperature rises above 205 de- Defective temperature sensor. Replace temperature sensor.
grees C. Defective power supply board. Replace power supply board.
Defective CPU board. Replace CPU board.

Unstable weight indication. The air current which cools the weighing Clean the ventilation grill.
cell is blocked.

3-2
CHAPTER 4 MAINTENANCE PROCEDURES
4.1 PREVENTIVE MAINTENANCE
Ohaus MB45 Moisture Analyzers are precision instruments and should be carefully handled, stored in a clean dry area which
is dust free, and cleaned periodically. It is recommended that when a Analyzer has had chemicals or liquids spilled on it,
the Analyzer should be cleaned as soon as possible. Use warm water on a damp cloth to clean all exterior surfaces. Do
not leave a mass on the sample pan when the Analyzer is not in use.

When moving the unit from a storage area which is at a different temperature than the area where it is to be operated, allow
sufficient time for the mechanism to temperature stabilize. This time can vary quite a bit depending upon the temperature
differences. Allow one (1) hour for each 5 degrees Fahrenheit temperature change before using the Moisture Analyzer. Also,
after turning the Analyzer ON, allow five minutes after temperature stabilization for the internal electronics to stabilize.

4.1.1 Visual Inspection


When a Moisture Analyzer is first received for repairs, the unit should be checked visually for completeness.
— Check on completeness in accordance with standard equipment.
— Check on the condition of the following parts:
• Housing
• Keypad and display
• Draft shield with heat shield, pan handler and pan support
• Dryer unit with protective glass, Halogen radiator, reflector
• Fan grill (rear)
• Leveling feet and level
• Rating plate

4.1.2 Preventive Maintenance Checklist


On a regular basis, the MB45 Moisture Analyzer should be
inspected and checked as follows:

1.Remove power from the MB45.

2.Remove the Pan Support, Draftshield and


Heatshield and inspect the interior and exterior of
the Moisture Analyzer. Glass Shield

3.Clean the outside of the Moisture Analyzer using


a damp cloth with water or a mild household
Pan Support
detergent.
Draft Shield
4. Remove any debris which may have accumulated
under the pan or shields.
Heat Shield
5.Inspect the protec-
tive glass shield
and clean any spat-
tered material
which may have
accumulated on
the glass.

6.Inspect fan open-


ing for excessive Figure 4-1. Access for Cleaning MB45 Moisture
dust.
Analyzer.
4-1
CHAPTER 4 MAINTENANCE PROCEDURES
4.1.2 Preventive Maintenance Checklist (Cont.)

CAUTION

DO NOT USE CHEMICAL CLEANERS OR SOLVENTS OF ANY


TYPE. SOME CLEANERS ARE ABRASIVE AND MAY AFFECT THE
FINISH OF THE ANALYZER.

7.Check the Power Cord for broken or damaged insulation.

8.Make a visual inspection for faulty connectors, wiring, and loose hardware.

4.2 TESTING
Before servicing the MB45 Moisture Analyzer, an operational test and various performance tests should be made to ascertain
whether or not the unit meets specifications. Turn the Moisture Analyzer on and allow it warm up for at least one hour before
performing these tests. Make sure the test area is free from drafts and the surface that the Moisture Analyzer rests on is
level and vibration free. It is recommended that a ASTM Class 1 Tolerance mass be used for the performance tests and
adjustments.
4.2.1 Operational Test
1.Locate a suitable power source for the Moisture Analyzer. Check the voltage rating of the Moisture Analyzer.

2.Plug the Power Cord into a suitable power source.

NOTE: If possible, record original customer settings so that after repairs, the Moisture Analyzer can be returned in
a satisfactory operational mode.

4.2.1.1 Display Test

1.Turn the Moisture Analyzer on, a MB45 Moisture Analyzer display with the Ohaus logo should appear followed
by the software revision currently installed in the Moisture analyzer. After this display, a standard display
appears. This is a display test. Figure 4-2 is a full display test. The software revision will only appear after plugging
in the analyzer for the first time. Subsequent power on using the On/Off button will not display the software revision.

2.Press the TARE button. The display should indicate a zero weight.

Sr 1.06b
TEST ID: -DEFAULT-
TEMP/TIME: 100C / 10:00
TARG. WGT: 1 GRAMS
CLEAR
PAN
PRESS
0.000
TARE * GRAMS

Figure 4-2. Display Test.


4-2
CHAPTER 4 MAINTENANCE PROCEDURES
4.2.2 Weight Calibration
The Moisture Analyzer can be span calibrated with an external mass of 20 grams. Calibration of the Moisture Analyzer
balance is not absolutely necessary for a correct moisture determination as the measurement is relative. The balance
determines the weight of the sample before and after drying and the moisture is calculated on the basis of the ratio between
wet and dry weights.

Nevertheless, you should calibrate the built-in balance under the following conditions:
— If this is stipulated by your quality assurance system (GLP, GMP, ISO 9001).
— If you suspect the analyzer has been abused.
— Before and after all repairs.

To calibrate the Analyzer, proceed as follows:

ANALYZER SETUP Clear the pan handler, a sample pan should be in place, close the cover on the
LOCK OUT Moisture Analyzer.
WEIGHT CAL
TEMP CAL
LANGUAGE Press the Setup button. Using the arrow buttons, scroll to WEIGHT CAL. Press
BEEPER the Enter button.
TIME-DATE

The display screen highlights the action you should take.

Place the required weight on the sample pan, close the cover.
WEIGHT CAL
PLACE 20G MASS Follow the instructions on the display. The display indicates if the calibration was
successful.
TO ABORT
PRESS STRT/STP KEY

Press the Display button to return to display. To abort, press START/STOP


button.

4-3
CHAPTER 4 MAINTENANCE PROCEDURES
4.2.3 Temperature Calibration
You must have an Ohaus Temperature Calibration Kit, (Part Number 11113857) to perform this procedure. If the Moisture
Analyzer has been recently used, allow at least 30 minutes before performing calibration.

ANALYZER SETUP NOTE: The Temperature Calibration Kit is available as optional equipment.
LOCK OUT
WEIGHT CAL Press the Setup button. Using the arrow buttons, scroll to TEMP CAL.
TEMP CAL
LANGUAGE Press the Enter button. You are now prompted to remove the pan handler and
BEEPER pan support. Replace the pan handler and place a temperature calibration unit
TIME-DATE
on the pan handler.

NOTE: The unit will not calibrate with the pan support in place.
TEMP CAL
REMOVE PAN HANDLER
AND PAN SUPPORT

ABORT CALIBRATION

Press the Enter button to initiate the temperature calibration process. Follow the
screen prompts throughout the process.

The dryer unit is heated to a temperature of 100°C. You can observe the progress
on the display as the dryer temperature and count down period are displayed.
After 15 minutes, read the thermometer through the inspection window on the
cover and enter this temperature. Using the up/down arrow buttons, adjust the
display reading to agree with the thermometer. Use the left/right arrow buttons
TEMP CAL 100C
CURRENT TEMP: --- C to highlight ACCEPT NEW CAL, then press the Enter button. You have 10
TM TO CAL POINT: 15:00 MIN minutes to make this adjustment, otherwise the calibration is terminated.
ADJ CAL READING: 100 C
ACCEPTS NEW CAL
ABORT CALIBRATION
This is a two point adjustment (100°C and 160°C). The dryer unit now heats to
TEMP CAL 160C the second temperature (160°C). Adjustment of the temperature is defined by
CURRENT TEMP: 100 C two points. Proceed exactly as you did for the first temperature. After you have
TM TO CAL POINT: 15:00 MIN
set the display to agree with the thermometer, highlight ACCEPT NEW CAL,
ADJ TO CAL READING: 160 C then press the Enter button. The display returns to ANALYZER SETUP.
ACCEPTS NEW CAL
ABORT CALIBRATION

WARNING: Be careful when removing the temperature calibration unit from the
dryer unit as it can be very hot. Allow it to cool down by opening the cover before
removing.

Remove the calibration unit. Replace the pan support and the pan handler in their
proper positions.

4-4
CHAPTER 4 MAINTENANCE PROCEDURES
4.2.4 Performance Tests
Accurate performance of the MB45 Moisture Analyzer is determined by a series of three performance tests. Precision,
Off Center Load and Centered Load. The displayed readings are compared with the tolerances listed in Table 4-1. Tolerance
values are expressed in grams. Before starting the performance tests, level the Moisture Analyzer.

TABLE 4-1. TYPES OF PERFORMANCE TESTS

PERFORMANCE TEST TOLERANCE

Precision (Standard Deviation) 0.0015g

Off Center Load +0.004g

Centered Load +0.004g

The following performance tests are used to evaluate the Moisture Analyzer operation before and after repairs. Each Moisture
Analyzer tested must meet the requirements specified in each test as well as the specifications listed in Table 1-1. Before
proceeding with the following tests, all the procedures starting with paragraph 4.2 must have been accomplished on the
Moisture Analyzer first. Tolerance values are expressed in counts. A Moisture Analyzer which passes each of these three
tests meets the manufacturing specifications.

4.2.4.1 Precision Test


Precision is a word used in balance specifications meaning the Standard Deviation of a set of similar weight readings. To
determine whether an Analyzer meets the calculated Standard Deviation value in the Specification Table 1-1, perform the
following test:

Test
1.Tare the Analyzer. The reading on the display should be 0.000g.

2.Place a 20g mass on the center of the sample pan. Observe and record the reading when the stable indicator
come on.

3.Remove the mass. Reading should return to 0g +0.002g when the stable indicator is on.

4.Repeat this test for ten readings. If the standard deviation of the readings is less than 0.0015g, the Moisture
Analyzer passes the Precision Test.

CAUTION: Remove the mass from the pan after testing.

Adjustment
If the deviation for any set of readings (using the same mass placed on the center of the pan) is greater than 0.0015g, the
Moisture Analyzer does not meet the Precision Test specification. Inspect and correct the following areas:

1.Check for mechanical obstructions. Any foreign object touching any part of the moving pan linkage assembly
will cause a Moisture Analyzer to fail the Precision Test. Inspect and correct as necessary.

2.An error in the Off-Center Load Test can affect the results of the Precision Test. Inspect and correct if necessary.
See Off-Center Load Test.

3.Foreign material or debris located in the balance between the pan and the base will cause the Moisture Analyzer
to fail the test.

4.Environmental influences such as vibrations, drafts or a non-level surface can cause failures.

4-5
CHAPTER 4 MAINTENANCE PROCEDURES

4.2.4.2 Off-Center Load Test


The Off-Center Load Test is used to determine whether displayed weight values will be affected by moving the sample to
different areas of the sample pan.

(NOTE: Due to the use of relative weighing in moisture analysis, off center load performance is not critical.)

Test
1.Place a sample pan on the pan handler.

2.Place 1/2 of the Moisture Analyzer capacity (20 gramms) in the center of the sample pan.

3.Press the TARE button to return the reading to zero.

4.Move the mass halfway (between the center and the edge) to the front of the sample pan. Note any differences
in the displayed weight reading.

5.Repeat this test for the back, left, and right positions.

6.Maximum allowable change in displayed weight readings is +0.004g for each of the four positions. If this reading
is exceeded, it usually indicates a defective Weighing Cell.

4.2.4.3 Centered Load Test


This test is used to determine the linearity of the unit throughout its operating range.

NOTES:
1. Due to the use of relative weighing and limited use of the weighing range in moisture analysis, linearity
performance is not critical.

2. The Moisture Analyzer must pass the Off-Center Load Test and Precision Test before the Centered Load
Test may be performed.

Test
1.Place a 20g mass on the center of the sample pan and note the reading.

2.If the displayed weight is greater than 0.002g higher or lower than the value of the applied mass, and recalibrate
the Moisture Analyzer. The Moisture Analyzer should now be calibrated in grams.

3.After completing the above adjustments, check the linearity at 10g and 20g. The displayed weight readings should
be equal to the applied mass. Tolerance is +0.004g for each of the respective weight readings. If the displayed
weight values are in excess of tolerance count, the Weighing Cell must be replaced.

4.3 REPAIRING THE MOISTURE ANALYZER


Disconnect from power supply and allow heating module to cool !

Important: Electrostatic protection when servicing !

Electrostatic damage is difficult to detect, because the faults it causes are not clear cut. MOS switching transistors with
electrostatic damage have substantially higher, thermically unstable leakage currents. As a result, the Moisture Analyzer
balance drifts more and the display can fluctuate similar to when there is a draft.

To avoid electrostatic damage during production, conducting floors, controlled air humidity, and EMC mats are used. When
servicing is carried out it is therefore also advisable – as soon as the instrument is opened – to neutralize electrostatic
charges.

4-6
CHAPTER 4 MAINTENANCE PROCEDURES
4.3.1 Repairing the Drying Unit
4.3.1.1 Dismantling the Heating Module

1. Swivel drying unit upward.

2. Unscrew 4 screws (a).

3. Pull cover off toward back.

Figure 4-3. Dismantling the Heating Module.

4. Unscrew 2 screws (b).


b
5. Lift heating module out.

6. Disconnect heater cable at 2 connectors (c).

Figure 4-4. Removing the Heating Module.


4-7
CHAPTER 4 MAINTENANCE PROCEDURES
4.3.2 Replacing the Protective Glass and
Reflector e
Swivel drying unit up and unscrew cover (d).

4.3.2.1 Removing Protective Glass and


Reflector f
1. Pull reflector (e) up and out.

2. Pull protective glass (f) up and out.

4.3.2.2 Installing Protective Glass and


Reflector
1. Insert protective glass with opening pointing
down.

2. Insert reflector.

3. Screw cover on (d).

Figure 4-5. Replacing Protective Glass.


4.3.3 Replacing the Halogen Lamp

4.3.3.1 Dismantle Halogen Lamp


1. Dismantle the heating module.

2. Dismantle protective glass and reflector.

3. Use both forefingers to spread the clamping


springs to the left and right and at the same time
use both thumbs to lift the Halogen lamp care-
fully upward out of the clamping springs (the two
ends ot the Halogen lamp must not be bent
toward or away from each other, otherwise, the
glass tube will break).

Figure 4-6. Replacing Halogen Lamp.

4-8
CHAPTER 4 MAINTENANCE PROCEDURES
4.3.3.2 Installing the Halogen Lamp
a b
CAUTION: Finger oils can cause premature lamp failure. c
Only touch the black part of the Halogen lamp! If necessary,
clean the lamp with a weak organic solvent.

1. Hold the new Halogen lamp by the black ends


and insert it into the two metal supports (a) at the
back. If the metal supports have fallen out,
replace them so that the wider part of the fork (b)
is on top; the metal supports must press the
Halogen lamp gently against the protective glass
(c).

2. Lay the flat,connector side of the Halogen lamp


on the clamping springs at the front,use both
thumbs to push it about 2 mm toward the back
against the metal supports and then down to
engage in the clamping springs.
Figure 4-7. Installing the Halogen Lamp.
3. Reconnect the 2 connectors of the Halogen
lamp.

4. Install the protective glass and reflector.

5. Install the heating module.

4.3.4 Replacing the Thermal Overload


Device (Thermal cut-out switch)

NOTE: Before replacing the the thermal overload d


device, it may have tripped and has to be reset. Refer
to paragraph 4.4 to reset the thermal overload e
device.

1. Dismantle the heating module.

2. Use flat-nosed pliers to pull off the two cable


connectors. Figure 4-8. Replacing Thermal Overload Device.

3. Replace temperature controller (e). Do not


forget the spacing washer on the heat-conduct-
ing plate (d)! Do not tighten the washer too tight
(1.2 Nm).

4. Tighten nut with 1.2 Nm. If the temperature


controller has responded,reset with finger (do
not push spring under edge of base!).

4-9
CHAPTER 4 MAINTENANCE PROCEDURES

4.3.5 Replacing the Temperature Sensor

1. Dismantle heating module.

2. Pull off the connector (a).

3. Remove screw (b) and (c),lift temperature sen-


sor out of its holder and replace minding to c

pull the sensor cable through the clamp (d) d


also see Figure 4-10.
b a
4. Screw temperature sensor firmly back into place
and push connector (a) on again.

5. Install heating module.

Figure 4-9. Replacing Temperature Sensor.


4.3.6 Installing the Heating Module

1. Reconnect the 2 connecting cables (f) and (g) of


the heater.
e
2. Insert the heating module into the frame and
fasten with 2 screws (e).

The bare connector (f) carries the 220 V


power supply voltage and must therefore
f
always be installed as shown. It must not lie
outside the heating module like connector
(g).

Figure 4-10. Installing Heat Module.


4-10
CHAPTER 4 MAINTENANCE PROCEDURES
4.3.6 Installing the Heating Module (Cont.)

3. Replace cover and fasten with 4 screws (a).

4. Swivel drying unit closed.

Figure 4-11. Replacing Cover.

4.3.7 Removing Housing Top

1. Remove pan handler, pan support, and draft


shield.

2. Lift heat shield off.

3. Remove screw (a) and then lift the sample a


chamber, first at the front and then completely.

4. Remove the 4 screws (b) from the housing top.

Figure 4-12. Remove Housing.

4-11
CHAPTER 4 MAINTENANCE PROCEDURES
4.3.7 Removing Housing Top (Cont.)
5. Lift the front of the housing top then,using a
screwdriver, carefully unfasten the connector
socket of the keyboard cable (b)and pull off the
cable. Pull off the cable of the display (a).

Figure 4-13. Remove Cable from Display.

6. Lift the housing top until the level indicator (d)is


free, then swivel it into the vertical position
toward the back to rest against the open cover of
the drying unit.

7. Pull off the connectors for the fan (a), tempera-


ture sensor (b), and Halogen lamp (c).

8. Put the housing top aside out of the way.

a
c
b

Figure 4-14. Remove Connectors.

4-12
CHAPTER 4 MAINTENANCE PROCEDURES

4.3.8 Removing the Membrane Keypad

1. Remove the housing top.

2. Release the faulty membrane keypad from the


cable side (h).

3. Remove any traces of adhesive (if necessary,use


pure benzene).

WARNING
Benzene is highly flammable,
use only in small quantities.

Figure 4-15. Remove Membrane Keypad.

4.3.9 Installing the Membrane Keypad


Avoid making finger marks and getting dust on the display
and window of the membrane keypad!

1. Insert the cable of the new membrane keypad.

2. Pull the covering foil off the adhesive surface.

3. Insert the membrane keypad into the depression


in the housing top, starting only with the front,
right-hand corner and then with the right-hand
edge. Then, roll it toward the left into the depres-
sion and press it down.

4. Install the housing top. (See 4.3.16)

Figure 4-16. Install Membrane Keypad.

4-13
CHAPTER 4 MAINTENANCE PROCEDURES

4.3.10 Replacing the display


1. Remove housing top.

2. Unscrew display and remove.

3. Insert new display and fasten with screws.

4. Install housing top. (See 4.3.16)

Figure 4-17. Replacing Display.

4.3.11 Replacing the fan


1. Remove housing top.

2. Unscrew screening plate from housing top.

3. Slide cable (i) out through grommet (k), remove k


grommet (k) from screening plate together with m
fan connector.

4. Replace fan,taking care to install it in the direc-


tion indicated by the arrow printed on it (m).
p
5. Assemble the housing top: do not forget the
spring (p)! (The pin triggers the safety switchoff
if the housing top is opened during operation.) i

6. Install housing top. (See 4.3.16)

7. Test safety switchoff for correct functioning:

• Start drying so the the lamp is lit.

• Open the cover of the MB45: the lamp


should switch off immediately. If it does
not, check whether the spring (p) is pressing
the safety pin upward (q):

Figure 4-18. Replacing Fan.

4-14
CHAPTER 4 MAINTENANCE PROCEDURES
4.3.12 Replacing the CPU board
1. Remove housing top.

2. Remove screening plate (a).

3. Pull off the connector to the power supply


(b), carefully unfasten the socket connector of
the weighing cell cable (c), and pull the cable out.
a
4. Unscrew the CPU board and remove.

5. Transfer the cell EEPROM (d) from the old CPU


board, replacing the battery (e) with a new one if
b
necessary (3 V, Lithium,e.g. CR 2032).

6. Install new CPU board. e


d
7. Install screening plate (a). c

8. Re-insert cables (b)and (c) and fasten.

9. Install housing top. (See 4.3.16)

10. Perform weighing and temperature calibration.

11. After replacing the PC Board, refer to Appendix


A and follow the instructions to configure the new
PC Board.

Figure 4-19. Replacing CPU Board.

4-15
CHAPTER 4 MAINTENANCE PROCEDURES

4.3.13 Replacing the power supply board


1. Remove housing top.
f
2. Remove the screw (f) at the back of the screen-
ing plate. k

3. Pull off the 4 connectors for the CPU board i


(g), fan (i), temperature sensor (k) and lamp (m).

4. Remove screw to release ground protection


cable (h).
g
m
5. Remove the 4 screws from the power supply
board and the 2 screws (o) from the RS232
interface socket. Lift the power supply board at
the front and remove it. p

6. Install new power supply board:

• First tighten the 2 screws (o) of the h


RS232 interface socket then the 4 screws o
with toothed washers of the power supply
board.

• Screw the eye of the ground protection


cable and the toothed washer firmly under
the head of screw (h). r

7. Screw in the back screw (f) of the screening


plate.

8. Plug in the 4 connectors.

9. If necessary, replace the main fuse (r)T6.3 A/


250 V or power supply (p) 500 mA/250 V.

10. Install the housing top. (See 4.3.16)


Figure 4-20. Replacing Power Supply Board.

4-16
CHAPTER 4 MAINTENANCE PROCEDURES

4.3.14 Replacing the Weighing Cell

1. Remove housing top.

2. Remove screening plate (a)and seal (b).

3. Unfasten the socket connector (e) of the cell on


the CPU board and pull out the cable.

4. Unscrew weighing cell (c) and remove.

5. Install new weighing cell in reverse order. a

6. Replace cell EEPROM (d). b

7. Install screening plate.

8. Install housing top. (See 4.3.16)

9. Perform weight calibration. c


d

Figure 4-21. Replacing Weighing Cell.

4-17
CHAPTER 4 MAINTENANCE PROCEDURES

4.3.15 Replacing the Housing Base

1. Remove housing top.

2. Remove the following parts from the old housing


base and install them on the new one:
h
• Leveling screws

• CPU board (c).

• Power supply board (d).


g
• Weighing cell (e). e

• Level indicator (f)

• Seals (g)

• Screening plate (h)

3. Install housing top. (See 4.3.16)

Figure 4-22. Replace Housing Base.

4-18
CHAPTER 4 MAINTENANCE PROCEDURES

4.3.16 Installing the Housing Top

1. Lay the housing top toward the back on the


opened cover of the drying unit.

2. Insert connectors for fan (a), temperature sen-


sor (b), and Halogen lamp (c).

3. LIft the housing top until it is above the level


indicator (d), tilt it into the horizontal position,
then over the level indicator and seal and lower
it onto the housing base.

CAUTION: Verify that cables for fan (a), tem-


perature sensor (b) and Halogen lamp (c) are
clear from housing top when installing. Wires
can be damaged if not positioned properly and
can possibly cause damage to the Moisture
Analyzer.
d

a
c
b

Figure 4-23. Install Housing Top.

4. Plug in the cables of the display and


keyboard. Position the ferrite core (e) in the
opening in the screening plate. Fasten the
socket connector with a screwdriver (f). Close
the housing top.

Figure 4-24. Plug in Cables.


4-19
CHAPTER 4 MAINTENANCE PROCEDURES

4.3.16 Installing the Housing Top (Cont.)

5. Center the housing top on the housing base and


fasten the 4 screws (b).

6. Insert the sample chamber (a) starting at


rear, position at rear, then fasten with screw. a
7. Place heat shield into position.

8. Insert draft shield, pan support, and pan handler.

9. Close cover. b

Figure 4-25. Housing Top Re-assembly.

4.4 RESETTING THE THERMAL OVERLOAD


DEVICE
This procedure applies to units containing an
access hole that allows the thermal overload
protection device to be reset manually. When
the dryer has overheated and the thermal over-
load protection has responded, it may be reset
as follows:

1. Disconnect the unit from the power supply.

2. Lift the top cover straight up and remove the


access hole cover from the access hole using
a flat head screwdriver.

3. Reset the overload device with your finger. See


illustrations.

4. Replace the access hole cover and reconnect


unit to power supply.

Figure 4-26. Resetting Thermal Overload Device.


4-20
CHAPTER 4 MAINTENANCE PROCEDURES
4.5 FINAL WORK STEPS
– Set original customer configuration.

– Attach peripheral units.

– Perform weight and temperature calibration.

– Adjust dryer with Ohaus temperature calibration kit, if not available, drying with a coin as sample; 160 °C for 15
minutes: Display of moisture: 0.000 g ± 0.005g

4-21
CHAPTER 4 MAINTENANCE PROCEDURES

4-22
CHAPTER 5 DRAWINGS AND PARTS LISTS

5.1 DRAWINGS AND PARTS LISTS


This section of the manual contains exploded views of the top housing, bottom housing and dryer unit with associated parts
lists.
NOTE:
In all cases where a part is replaced, the Moisture Analyzer must be
thoroughly checked after the replacement is made. The Moisture
Analyzer MUST meet the parameters of all applicable specifications in
this manual.

If further technical information is needed, in the United States call Ohaus Aftermarket toll-free 1-800-526-0659 between 8.00
a.m. and 5:00 p.m. EST. An Ohaus factory service technician will be available to provide assistance. Outside the U.S.A.,
please contact:

Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ 07058-2033, USA
Tel: (973) 377-9000
Fax: (973) 944-7177

5-1
CHAPTER 5 DRAWINGS AND PARTS LISTS
5.1.1 Top Housing

4
2

9
7

11

8 10

35

Figure 5-1. Top Housing.

5-2
CHAPTER 5 DRAWINGS AND PARTS LISTS

TABLE 5-1. SPARE PARTS LIST TOP HOUSING

Item Description Part Number

1 Sample pan handler (accessory, 3 pieces) 214526

2 Sample pan holder 214642

3 Draft shield, clean out tray 214368

4 Heat shield 11113360

5 Sample chamber 11113359

6 Membrane keypad 11113356

7 Top housing with keypad (without display) 11113862

8 Display 11113772

9 Safety pin 11113367

10 Spring for safety pin 11600360

11 Fan 11113770

5-3
CHAPTER 5 DRAWINGS AND PARTS LISTS
5.1.2 Bottom Housing

12

13

14

24

15

22
19

17
23

18

21

16

20

Figure 5-2. Bottom Housing.

5-4
CHAPTER 5 DRAWINGS AND PARTS LISTS
TABLE 5-2. SPARE PARTS LIST BOTTOM HOUSING
Item Description Part Number
12 Seal for housing top 11113362

13 Weighing cell incl. EEPROM (Item 18) and seal 11113850

14 Seal for weighing cell 11600363

15 Power supply board 11113780

16 CPU board 11113853

17 Battery 3 V, Lithium, CR 2032 or equivalent See Note

18 Cell EEPROM 1) 11103757+SNR

19 Cable CPU – power supply 11113773

20 Bottom housing 11113350

21 Leveling screw 11101302

22 Level 11101335

23 Microfuse cap (fuse T6,3 A/250 V) 89290

24 Line cable (accessory) 6569-00, 120V US

24 Line cable (accessory) 76212-00, 220V EU

24 Line cable (accessory) 76448-00, 240V UK

24 Line cable (accessory) 76199-01, 240V AUS

1) The order has to contain the serial number SNR of the balance!

NOTE: battery is not available as a replacement part. Must source locally.

5-5
CHAPTER 5 DRAWINGS AND PARTS LISTS

5.1.3 Dryer Unit

25

36
26

27

28

32
29

30

33
35
34

Figure 5-3. Dryer Unit.

5-6
CHAPTER 5 DRAWINGS AND PARTS LISTS
TABLE 5-3. SPARE PARTS LIST DRYER UNIT

Item Description Part Number

25 Reflector 214382

26 Protective glass 214360

27 Halogen radiator110V 214417


Halogen radiator 230V 214418

28 Heating module110V 11113500


Heating module 230V 11113501

29 Temperature sensor 214349

30 Thermal overload device 214698

32 Dryer unit support 11113555

33 Dryer unit hinge left 11600358

34 Dryer unit hinge right 11600357

35 Dryer unit cable 11113771

36 Dryer unit cover 11113503

5-7
CHAPTER 5 DRAWINGS AND PARTS LISTS

5-8
APPENDIX A SOFTWARE SERVICE TOOL INSTRUCTIONS
A. SERVICE NOTES
SOFTWARE LEGEND
A.1 Software Service Tool Instructions MB45 GLOBAL 11670614x.MOT
This software tool is used when a new Main PCB is installed to:
1. Configure the Main PCB. MODEL
2. Enter balance serial number WHERE USED
3. Install the application software SOFTWARE
FILE REVISION

A.2 Requirements
1. File: MB45 Service Tool Install.zip

2. Application Software File:


a. MB45 Global 11670614x.mot
b. MB35 Europe 11670806x.mot
c. MB35 Latin America 802509712.mot
d. MB35 US 11670787x.mot
3. A PC running Windows 95 to XP

4. Interface cable Ohaus part number 80500525 (Serial extension cable)

A.3 Instructions and Notes


1. To open click on the zip file

2. Find the setup file (Setup.exe) and click on it.

3. Follow the instructions.

4. Copy the application software file to the desktop. This is not required but will make the file easier to find later.

5. It is advisable to hook up the balance to HyperTerminal and make sure it is communicating. You will need a standard
9 pin serial extension cable, Ohaus part number 80500525 to connect to your PC.

6. Find “MB45 Service Tool” in the program files and click it to start.

7. The screen below will open.

8. Match the serial port parameters to the balance. Balance defaults are shown above.

9. Click on the tab “Main Board Replacement”

A-1
APPENDIX A SOFTWARE SERVICE TOOL INSTRUCTIONS

10. Click on the button that matches the balance.

11. Enter the serial number of the balance.

12. Enter the TDNR number of the balance.

Note: The serial number and TDNR number are printed on a label on the back of the balance.

13. CAUTION – Confirm the information entered above. This programming can only be done once. If you have made
and error, click the button “Abort Data Entry” then reenter the data.

14. When the above information has been confirmed click the “Start” button.

The following screen will open to indicate the new main board has been configured.

15. Click on the “Download” Tab.

16. Browse to the required Application Software file (mot file).

A-2
APPENDIX A SOFTWARE SERVICE TOOL INSTRUCTIONS

17. Press the start button.

18. If there is a problem press the red button


a. The red button is intended to download the data for the very first time or to a blank EEPROM. It is also intended
to be used as a last resort if the start button fails.

b. Using the start button will save the existing user’s test library, using the red button will not.

c. After the red button is clicked you will see a screen that says “Please cycle power to scale…”. At this time unplug
the scale from the AC line for about 30 seconds and plug it back in. The download will then take about 3 minutes.
When it is complete the message “File download is SUCCESSFUL”(Checksum=xxxxxxx) will be displayed.
If the message “Application code download has FAILED” is shown click the OK button and click the red button
again. The second time it should work.

A-3
APPENDIX A SOFTWARE SERVICE TOOL INSTRUCTIONS

When the software has been downloaded the following screen will be shown.

A.4 Error Messages

Indicates the computer can not establish a connection to the balance.

Indicates the main board has already been configured. A main board can be configured only once and can not be
reconfigured.

Indicates the TDNR number does not match the balance configuration selected. Click “Abort Data Entry” then re-enter the
correct data.

A-4
MB45 MOISTURE ANALYZER SERVICE MANUAL P/N 80250897 A

Vous aimerez peut-être aussi