Vous êtes sur la page 1sur 6

PAGEl I6

Lu orutory
GALIBRATION CERTIFICATE NO 53879 1

Company:

Equipment: Digital Megohmmeter Megabras @,


model MD - i0KVx
Serialnumber: 18J0306

Calibration date: 1011012018

Calibration Procedure
The used calibration procedure is based on the recommendations of the standard ISO I IEC 17025, on the technical
instruction PQ MD 08 and on operating manuals of instruments involved. The calibration was performed by method of
comparison with the used standards.

Used standards
Description Manufacturer / Model Serial number Certificate Issuing Calibration Due

6%-digit digital multimeter Fluke / 8845A 1 869003 309598 ESC' 0111312019


Standard resistance box Megabras/CPR-20G-E 049G 1 581 59-1 01 ¡PT2 05t04t2019
Digital kilovoltmeter Megabras/KV25e oG 1009 F 1 58176-1 01 IPT2 0510512019

r ESC Everett Service Center.


-
'z IPT
- lnstituto de Pesquisas Tecnológicas de Sáo Paulo - Traceable by RBC - Brazilian Calibration Network.

Environmental conditions during the procedure


Room temperature: . 24!1oC
Relative humidity : 40¡4%

Calibration results are contained in the accompanying tables, which compare the values indicated by the unit under test with
the reference values of the standards.

A This certificate moy not ,be reproduced other fhsn in fuil.

THNTICH COTPORATION
7330 NW éó'!h §T I Miomi, rL 33róó I USA
Phone: *l 305 9380389 N Fo* +l 78640171ó5 llechsupporl@tenlech.us lwww.tentech.com
PAGEZI6

Lm ürstory
CALIBRATION CERTIFICATE NO 53879 1

1- Voltages test calibration


MPE (Maximum Permissible Error): t (3o/o of the nominal value)
Unit Nominal value Measured value Error MPE
500 -5 t15
600 -8 t18
700 -4 t21
800 -2 t24
V 900 1 t27
1 000 1 t30
1 500 1509 I t45 Steps 500V
2000 2010 10 ,+60
2500 2510 10 t75
3.00 3.01 0.01 r0.09
3.50 3.50 0.00 t0.1 1
4.00 4.01 0.01 ¡0.12
4.50 4.51 0.01 ¡0.14
5.00 5.00 0.00 t0.1 5
5.50 5.50 0.00 t0.17
6.00 6.00 0.00 t0.1 8
KV 6.50 6.51 0.01 t0.20
7.00 7.00 0.00 ¡0.21
7.50 7.51 0.01 ¡0.23
8.00 8.00 0.00 ¡0.24
8.50 8.50 0.00 t0.26
9.00 9.00 0.00 t0.27
9.50 9.50 0.00 t0.29
10.00 10.03 0.03 10.30

2- Voltmeter calibration AC/DC


Range (15 - 600) V
MPE (Maximum Permissible Error): t (5% of reading + 3 digits)
Unit Standard value Current Measured value Error MPE
20.1 22 19 !4
110.4
AC
110 -0.4 t9
200.9 200 -0.9 t13
573.3 569 -4.3 t32
V
18.3 19 0.7 !4
97.2 96 -1 .2 t8
DC
243.1 250 6.9 t15
494.5 506 11 .5 ¡28

A This certificote rnoy not ,be reproduced ot,her thsn in full.

TENTf;CH CÜR?ÜT},Ñ'ÜN
7330 Nv/ $&th ST ¡ Wtiami, FL 33X 6ó | USA
Phone: -L I 3ü5 938 ü38V I Fr:x: + 1 7 86 4ü1 7 X 65 | techs uppürtffi)tentech. us wv,rw.tentech.(úm
I
PAGE3/6

Ln #rmtmry
CALIBRATION GERTIFICATE NO 53879 1

3- Ammeter calibration
MPE (Maximum Permissible Error): + (10% of reading + 3 digits).
Unit Standard value Measured value Error MPE
83.0 83.5 0.5 ¡11
18 2 0.0 t03
uA
5.08 5 0.0 ¡4
7.82 I 0.0 t04
mA 1.944 1.94 -0.004 ¡0.23

4- Calibration of capacitance ranges


MPE (Maximum Permissible Error): t (1OYo of reading + 3 digits).
Unit Standard value Measured value E rror MPE
97.6 100 2.42 t13
238.5 231 -7.51 t27
nF
498.4 490 -8.4 t53
995.7 1000 4.3 11 03
1.970 2.00 0.030 ¡0.23
pF
8.1 50 8.30 0.1 50 t0.82

5- Calibration of resistance ranges with voltages up to 4 kV


MPE (Maximum Permissible Error): t (5% of measured value)
Standard MV MV MV MV MV
Unit Error Error Error Error Error MPE
Value 500 v 1kv 2kv 3kv 4kv
2.00 2.02 0.02 2.01 0.01 2.00 0.00 2.00 0.00 2.00 0.00 t0.1 0
10.00 10.04 0.04 9.99 -0.01 10.02 0.02 10.02 0.02 10.01 0.01 t0.50
20 00 20.0 0.00 1g.g -0.1 0 19.9 -0.1 0 20.0 0.00 20.0 0.00 t1 .0
MO
100.00 101 .4 1.40 100.8 0.80 100.2 0.20 1 00.1 0.10 I 00.1 0.10 t5.0
500.00 505 5.00 502 2.00 501 1 .00 501 1 .00 501 1.00 t25
1 000 00 1012 12.00 1 007 7.00 1 003 3.00 1 003 3.00 1002 2.00 r50
5.000 5.06 0.060 5.02 0.020 5.01 0.010 5.01 0.010 5.01 0.010 !0.25
10 000 10.10 0.1 00 10.06 0.060 10.05 0.050 10.01 0.010 10.01 0.010 t0.50
GO 50.000 49.8 -0.200 50.0 0.000 50.0 0.000 50.0 0.000 50.0 0.000 x2.5
100.000 100 0.000 100 0.000 100 0.000 100 0.000 100 0.000 t5
200 200 0.00 200 0.00 200 0.00 200 0.00 200 0.00 t10

A This certificate moy not ,be reproduced other fhsn in full.

TINTTCH CÜRIDÜRÑ'ÜN
733ü NV/ b(tth ST 'i !'ltion'ti, FL 3316b I USA
Phrsne: -L I 3ú5 938 ü389 t, F*x: -r 1 V 8ú 40 1 7 X 65
| techsuppo't*Jitentech. us I www.tentech,norn
PAGE4I6

Lm ffirmtsry
CALIBRATION CERTIFICATE NO 53879 1

6- Calibration of resistance ranges with voltages up to g kV


MPE (Maximum Permissible Error): t (5% of measured value)
Standard MV MV MV MV MV
Unit Error Error Errotr Error Error MPE
Value 5kv 6kv 7kv 8kv 9kv
2.00 2.00 0.00 2.00 0.00 2.00 0.00 2.00 0.00 2.00 0.00 t0.1 0
10.00 10 0.00 10 0.00 10 0.00 10 0.00 10 0.00 t0.50
20 00 20.0 0.00 20.0 0.00 20.0 0.00 20.0 0.00 20.0 0.00 r1.0
MO
100 00 100.0 0.00 100.0 0.00 100.0 0.00 100.0 0.00 100.0 0.00 t5.0
500.00 501 1 .00 500 0.00 500 0.00 498 -2.00 498 -2.00 t25
1000.00 1002 2.00 1002 2.00 1002 2.00 1 001 1 .00 1 001 1 .00 t50
5.000 5.01 0.010 5.01 0.010 5.00 0.000 5.00 0.000 5.00 0.000 ¡0.25
10 000 10.00 0.000 10.00 0.000 10.00 0.000 9.99 -0.010 9.99 -0.010 t0.50
GO 50.000 50.0 0.000 50.0 0.000 50.0 0.000 50.1 0.1 00 50.1 0.1 00 t2.5
100.000 100 0.000 100 0.000 100 0.000 100 0.000 100 0.000 t5
200 200 0.00 200 0.00 200 0.00 200 0.00 200 0.00 t10

7- calibration of resistance ranges with voltage 10 kV


MPE (Maximum Permissible Error): t (5% of measured value)
Standard Meas. Value
Unit Error MPE
Value 10 kv
2.00 2.00 0.00 t0.1 0
10 00 10 0.00 t0.50
20.00 20.0 0.00 t1.0
MO
100 00 r00 0.00 t5.0
500.00 498 -2.00 ¡25
1 000 00 1 001 1.00 t50
4 980 4.99 0.010 ¡0.25
10 000 9.99 -0.010 t0.50
GO 50 000 50.1 0 100 ¡2.5
100.000 r00 0 000 t5
200 200 0.00 t10

A This certificate mtry not be reproduced other fhsn in full.

T[NTECH CORPORATION
7330 NW ééth ST I Miomi, rL33róó I U§A
Phone: *13059380389 lFox: +'l 78640171ó5 ltechsuppori@lentech.us lwww.tentech,com
PAGEs/6

Ln ürmtory
GALIBRATION CERTIFICATE NO 53879 1

Observations
'This calibration certificated is valid only for the tested unit, not being extensive to any lots, even if it is similar.
. The error shown in the tables is obtained by subtracting the measured value minus the standard value.
'The MPE (Maximum Permissible Error) is calculated according to the manufacturer specifications.
. Legend for abbreviations used (MV = Measured Value).

Uncertainty
The conventional true values (standards) are affected with the following uncertainties:

Items (1X2) Voltage: + 0.3o/o


Item (3) Current: t 0.3%
ltem (4) Capacitance: t 3.0%
Items (5XOX7) Resistance: r 1.0%

The uncertainty expanded informed is based on a standardized uncertainty combined multiplied by an inclusion factor K=2,
for a level of trust of approximately 95%.

RESULT
The instrument tested meets the accuracy specifications declared bv the manufacturer in the user ouide.

A This certificate moy not ;be reproduced other than in full.

TEN?TCá-1 CÜftPÜp.^Ñ'ÜN
733ü NV,/ b&th ST i ?,Aismi, FL 331 óú I USA
Phone: *'1 3ü5 ?38 ü389 | Fr:x: + 1 786 4fr1 V 165 | techsuppüfirr.fJ)tentech.us I w''rrw.te ntech.rorn
PAGE6/6

Ln 0rotory
CALIBRATION CERTIFICATE NO 53879 1

Calibration date. October 10, 2018

Calibration performed by: Supervisioned by:

Luis Chan Alejandro Rivero

A This certificote rnoy not be reproduced o.lher than in futl"

TENTECH COftPORATION
rt33róó | U§A
7330 NW é6'rh ST I Miomi,
Phone: *1 3059380389 lFqx: *l 78é401 71ó5 ltechsupport@renlech.us lwww.tentech.com