Vous êtes sur la page 1sur 4

SASSCO Intermediate Check Instruction Sheet

SAUDI ARABIAN SENSING SOLUTIONS CO., LTD.

Title: Intermediate Check Instruction Sheet Prepared by:


Equipment: 5520A Multi-Product Calibrator Omer Nadeem

Asset no: SS-002 Approved by:


Abdulwahab Almajed

Date of Intermediate Check -___________________

Instruction Checklist for Intermediate Check / Maintenance Check


(Mark the Figures for any Visible Defect)

Figure 01 : Front Panel Schematic (a)

Figure 02 : Rear Panel Schematic (a)

Instruction Sheet for Intermediate Check / Maintenance Checklist


Instruction Detail(a) Check Box(b) Remarks
Ports Connection Grip Check
(Front Panel).

Page 1 of 2
SASSCO Intermediate Check Instruction Sheet
SAUDI ARABIAN SENSING SOLUTIONS CO., LTD.

Rear Panel Check- Power Cable


Connection, Fuse Check, Port
Grip Check & Remote Cable
Connections Check

Earth Grounding Connection Grip


Check
Stabilization time of 30 Minutes
for the Equipment

Check zero value of the


equipment when fully stabilized

Functions check by selecting all


the menus

Buttons Functionality Test

Calibration Switch Disabled


Check / Rear Panel Switch
Intermediate Readings Check(c)
(a) Instruction details to be followed as it is, for any further calcification refer to manual in the Server
(b) Mark Check Box with Alphabets A: Accepted , AC: Accepted with Comments, R: Rejected, RE: Repaired
(c) Reading Data Sheet attached for filling.

Date:_________________ Check Done By: ___________________

Page 2 of 2
INTERMEDIATE CHECK DATA SHEET
Description Multi-Function Calibrator Certificate No.
Manufacturer Fluke Checked by
Model 5520A IC Date
Serial No. Next IC Date
INTERMEDIATE CHECK DATA SHEET
Limit
Function Tested Range Nominal UUT Reading
Min Max

Measure Mode Vdc Vdc Vdc Vdc


DC Voltage 0 to 20 Vdc 5.000 4.999 5.001
10.000 9.999 10.002
15.000 14.998 15.002
20.000 19.997 20.003

Vac Vac Vac Vac


AC Voltage 0 to 250 Vac 50.0 49.993 50.008
100.0 99.985 100.015
150.0 149.978 150.023
200.0 199.970 200.030
240.0 239.964 240.036

Ω Ω Ω Ω
Resistance 0 to 400 Ω 100.00 99.90 100.10
200.00 199.90 200.10
300.00 299.90 300.10
360.00 359.90 360.10

400 to 3200 Ω 500.0 499.0 501.0


1000.0 999.0 1001.0
2000.0 1999.0 2001.0
3000.0 2999.0 3001.0

mAdc mAdc mAdc mAdc


DC Current 0 to 24 mAdc 4.000 3.999 4.001
8.000 7.999 8.001
12.000 11.998 12.002
16.000 15.998 16.002
20.000 19.997 20.003
°C °C °C °C
Temperature -200 to 850 °C -100.0 -100.3 -99.7
PT100 0.0 -0.3 0.3
100.0 99.7 100.3
300.0 299.7 300.3
500.0 499.7 500.3
700.0 699.7 700.3
Type K -200 to 1372 °C -100.0 -100.4 -99.6
0.0 -0.4 0.4
250.0 249.6 250.4
500.0 499.6 500.4
1000.0 999.6 1000.4

Type J -210 to 1200 °C -100.0 -100.4 -99.6


0.0 -0.4 0.4
250.0 249.6 250.4
500.0 499.6 500.4
1000.0 999.6 1000.4

Type T -250 to 400 °C -100.0 -100.4 -99.6


0.0 -0.4 0.4
100.0 99.6 100.4
300.0 299.6 300.4
350.0 349.6 350.4

Page 1 of 2 QF-22, Rev.01


Limit
Function Tested Range Nominal UUT Reading
Min Max

Type E -270 to 1000 °C -100.0 -100.4 -99.6


0.0 -0.4 0.4
250.0 249.6 250.4
500.0 499.6 500.4
800.0 799.6 800.4

Source Mode Vdc Vdc Vdc Vdc


DC Voltage 0 to 20 Vdc 5.000 4.999 5.001
10.000 9.999 10.002
15.000 14.998 15.002
20.000 19.997 20.003

Ω Ω Ω Ω
Resistance 0 to 400 Ω 100.00 99.9 100.1
200.00 199.9 200.1
300.00 299.9 300.1
360.00 359.9 360.1

400 to 3200 Ω 500.0 499.0 501.0


1000.0 999.0 1001.0
2000.0 1999.0 2001.0
3000.0 2999.0 3001.0

mAdc mAdc mAdc mAdc


DC Current 0 to 24 mAdc 4.000 3.999 4.001
8.000 7.999 8.001
12.000 11.998 12.002
16.000 15.998 16.002
20.000 19.997 20.003
°C °C °C °C
Temperature -200 to 850 °C -100.0 -100.3 -99.7
PT100 0.0 -0.3 0.3
100.0 99.7 100.3
300.0 299.7 300.3
500.0 499.7 500.3
700.0 699.7 700.3

Type K -200 to 1372 °C -100.0 -100.4 -99.6


0.0 -0.4 0.4
250.0 249.6 250.4
500.0 499.6 500.4
1000.0 999.6 1000.4

Type J -210 to 1200 °C -100.0 -100.4 -99.6


0.0 -0.4 0.4
250.0 249.6 250.4
500.0 499.6 500.4
1000.0 999.6 1000.4

Type T -250 to 400 °C -100.0 -100.4 -99.6


0.0 -0.4 0.4
100.0 99.6 100.4
300.0 299.6 300.4
350.0 349.6 350.4

Type E -270 to 1000 °C -100.0 -100.4 -99.6


0.0 -0.4 0.4
250.0 249.6 250.4
500.0 499.6 500.4
800.0 799.6 800.4
Standards Used
Description ID No. Model No. Due Date Cert. No. Traceability
Multi-Product Calibrator 5520A NPL
Precision Multimeter 8846A NPL

Page 2 of 2 QF-22, Rev.01

Vous aimerez peut-être aussi