Flushing water pressure gauge installation
ARA 12.9.2003
Approved by Date
JKK 12.9.2003

Contents
1. General ..........................................................................................................................................................2
1.1 Installation ..........................................................................................................................................2

READER COMMENTS
Flushing water pressure gauge installation
Code 10000002629e (prev. code 710 3174) Revision C Page 2 (3)

1. General
This document shows installation of the flushing water pressure gauge onto
slurry multiplexer type MXA 4645 used in association with Courier 6SL analyzer
system.

Flushing Water Pressure Gauge unit provides water pressure contact


information for the Courier 6 analyzer system. Automatic Window Changer
utilizes this contact for successful window change process. Flushing Water
Pressure Gauge comprises of Water Pressure Sensor and required water hose
fittings.

1.1 Installation
Only one flushing water pressure gauge is required for one analyzer system. In
case the system has more than one slurry multiplexers the flushing water
pressure gauge is installed onto water supply hose of multiplexer one.

The switch contact is normally closed but in case the water pressure is lost or
the signal cable is cut the contact opens. The contact is wired onto the input #7
(WPS) of the Main ICM module at the PCS 4000.

Besides the physical installation the user parameters need to be defined for
effectively use of the flushing water pressure gauge contact. The definition is
done at lines 25 and 26 of the MC an MCFL sheet of the User parameters Excel
file. At line 25 “Flush water pressure sensor in use” parameter needs to defined
as “1” and at line 26 “ Flush water pressure signal filtering time” can be defined
as 3 seconds.

At Physical IO sheet the right polarity of Input #7 (FWS LI) has already been
defined by default.

For physical installation of the flushing water pressure sensor see two figures
on this document.
Flushing water pressure gauge installation
Code 10000002629e (prev. code 710 3174) Revision C Page 3 (3)

Figure 1

Figure 2
TELEFAX MESSAGE DATE: Page 1

READER COMMENTS
The design team of Outotec is committed to ensure the quality and usefulness of this manual
and the COURIER analyzer system. Your comments, recommendations and suggestions will
help make our continuous effort more effective.

Please take few moments to write your comments either directly on the form provided or on a
photocopy of it.

Your assistance is greatly appreciated.


Thank you for choosing the COURIER analyzer system.

Outotec, Analyzers
Fax to: FAX: INT + 358 205 292998

or mail to: P.O. Box 84, FI-02201 Espoo, Finland

or e-mail to: analyzers@outotec.com

Attn: Manager, Analyzer Sales Support

From:

Analyzer Serial Number:

Manual:

Flushing water pressure gauge installation


Code 10000002629e (prev. code 710 3174) Revision C

( Please identify applicable software versions and individual


documents )

Comments /
recommendations:

Number of additional pages ( )

Vous aimerez peut-être aussi