PhCh 136 Lec 2nd Exam - Only bonds that have dipole moment
1st Trans are able to absorb IR (hence the reason
April 3, 2019 why homonuclear molecules lack IR absorption) (Sorry guys hindi narecord ang konteng first part Types of vibrations involved in IR ng IR Spectro kaya I will try my best to fill in the 1. Stretching Vibrations first part based on information from references - Symmetric such as Harvey or Skoog :( - Asymmetric Infrared Spectroscopy 2. Bending Vibrations - IR Region is from around 1 nm to - Twisting around 1 mm - Wagging - IR Spectro is typically associated only - Scissoring with vibrational portion of IR (2.5 nm to - Rocking 25 nm) *Each of these types of vibrations will have a - Important in the identification signature signal in the IR spectrum (qualitative) of both organic and Another thing to look for in IR spectrum: inorganic compounds because all Intensity of the band molecular species absorb IR (Except for Intensity homonuclear molecules such as O2, N2, - Tells the type of functional group etc) present - Important in quali also because every - ↑ dipole moment, ↑ intensity molecule has a unique IR spectrum - ↑ electronegativity, ↑ intensity (except for chiral molecules in the - ↑conjugation to polar double bond, ↑ crystalline state) (Skoog, et. al., 2013) intensity - Wavelength is in terms of Energy Level (which is in terms of wavenumbers, which is the reciprocal of wavenumbers) the wavelength in cm - Molecules with similar functional groups - Wavenumber is more preferred than will absorb photons of similar energies wavelength because it is directly - The higher the atomic mass difference, proportional to energy (higher the higher the energy of absorption wavenumber corresponds to higher needed energy) - Double bonds are stronger than single IR Spectro for quantitative analysis? bonds and therefore absorb at higher - It is less satisfactory than its UV/Vis energy than single bonds counterpart since it offers less Is IR important in analytical chemistry? sensitivity, less precision and higher - In certain monographs, IR spectra are deviation from Beer’s Law needed in the identification of these - IR radiation is less intense than UV/Vis, certain compounds therefore more prone to stray radiations - Not all compounds need IR spectra - However, IR MAY BE USED - There is no pattern that will tell you quantitatively if precision requirements whether a compound will need an IR are not that strict since it gives the spectra or not advantage of being quite selective - Most of the time, monographs require (Skoog, et. al., 2013) only physical characteristics or physical Process of IR absorption constants for the identification of - IR is lower energy than UV/Vis, compounds; sometimes, UV spectrum is therefore it can only affect the stretching needed and bending of bonds, as compared to - IR is usually used for identification of UV/Vis wherein jumping of electrons to structure of compounds of unsure a different energy state is concerned identity Requirements for IR absorption IR in QC - Only the frequencies of IR that match - IR is being looked at nowadays as a the frequencies of the natural vibration potential QC test of the molecule is absorbed - This is because the sample preparation ● If the make up of the monochromator is not in IR is as easy and fast as that of in as good, if the gratings are not as high (less UV/Vis gratings = less resolution), they actually lose - QC has to be as fast as possible some of the data. Ang nangyayari yung read because it is a requirement in out from a dispersing type is a little cleaner. manufacturing (wherein time is of the Nawawala kasi yung ibang data from the essence) monochromator. 2 types of IR instruments - One on the left side of the ppt is Source of radiation for IR: oxides of zirconium, continuous or dispersive type of yttrium, or thorium instrument ● Gives a wavelength from 2-15 micrometers - One on the right is the FTIR or the ● Non-conducting at room temperature Fourier Transform IR machine ● Alternative: glow bar → gives a less intense Continuous IR type of radiation - Earliest type - Gives less information as compared to Monochromator the FTIR ● Will only appear in the continuous or - Not that common nowadays dispersive type of an IR instrument (older - Sa FTIR kasi mas clean ang data and type of the instrument) mas complete ● Most commonly used: grating type - Continuous IR gives very general ● If the grooves are not as fine (mas information malalawak yung distances), data loss will How Continuous IR works come in because hindi talaga nahihiwa yung - IR radiation is given off by a filament of wavelength coming from the sample. zinc oxide or yttrium or thorium oxides (Remember: 2-15 micrometers lang yung range ng IR) All of the wavelengths associated with the IR ● It should be able to chop 2-15 micrometers region will go to the sample. After that kapag into as many wavenumbers or wavelengths nagkaroon na ng absorption (rotational to as possible. vibrational energy levels), it will give out again ● Yung one micrometer, malaki na yun. It can that light na icacapture ng monochromator one differentiate a C-O absorption and a C-Cl wavelength at a time. absorption. ● Kapag mas marami [ang grooves/gratings], FTIR mas okay. ● Still, the source will give out all the ● Mas maganda ang FTIR because it will not wavelengths of that energy source have this disadvantage of using a (zirconium oxide, yttrium oxide, or thorium monochromator. All of the data will be oxide) then it will bring it to the sample captured. There is no filtering involved. simultaneously. The sample will give the data simultaneously. Sample cell ● Instead of getting the absorption data one ● Where the sample is placed wavelength at a time, the moving mirror will ● If you have liquid sample, it should be as capture all of the information simultaneously much as possible anhydrous in nature. It also. can be tested as neat (undiluted form of the ● Mathematical equation of Fourier Transform sample; no added liquids to deliver the -- inaayos yung data na simultaneously sample into the IR) captured. ● It can also be diluted using carbon ● There is no data loss in the process tetrachloride or carbon disulfide in a because everything is captured short-pathed cell. simultaneously. There is no selection or ● It’s still possible to have problems with filtering unlike that of continuous or saturation → dilute! dispersed type. Pero essentially there is ● Carbon tetrachloride and carbon disulfide filtering done by the monochromator. should not give any interfering response to the rest of the spectra or throw it to the moving mirror for the FTIR Solids type. ● KBr disc ● From those mechanism, it will throw those ○ KBr will just be mixed thoroughly with signals into the detector. the sample in a mortar and pestle ● There are several detectors right now, each ○ Sample is mixed in a particular ration with an advantage. depending on the sensitivity of the ● The earliest ones are the transmission type sample of the detector or the courier transform ○ E.g. 10 mg of the sample + 100 mg KBr which is still the most widely used. It uses a → 10% lang ng buong preparation signal from there which relays the radiation ○ standard source in which individual can be monitor in within a one second of radiation can be ● NaCl monitor in a one second pulse of radiation. ○ Alternative when KBr cannot be used ● The important aspect here with courier ○ Preferred when the solid is in chloride transform is how fast it can generate the (?) form so that halogen exchange will not ● So pag sinabing one second pulse(?), all the occur → kailangan pareho yung salt data are captured in a matter of seconds. form That’s why for FTRI instruments, the data ○ If hindi nakasalt form, there is no capture or the data generation will only problem. Iniiwasan lang magkahalogen takes seconds. exchange in the process. ● Pagkaprepare ng sample, paglagay dun sa instrument, in a matter of second you can Paano ba nagkakaroon ng halogen exchange? get the results. Again, no filtering involved. ● These materials are just mixed in a mortar ● But when a monochromator is used, the and pestle. When these are made in a process can take from 15-30 minutes, pellet, pressure is actually placed on the before a data is generated. sample. ● ·Madalas naman hindi siya nagiging ● So, para kang gagawa ng waffle. Kapag problem, just because you are placing gumagawa ka nun may presser ka. So samples which are relatively stable inside ganun na ganun siya. Papaumbukin mo the instrument. siya. And then, you really have to press on ● There is no heating involved in this. The light that so that it becomes a hard pellet. will just go through the sample, a low energy ● Once you placed pressure, the tendency is light will just go through a sample. At one you place enough energy for halogen point, data capturing lang ang magtatagal. exchange to occur in the process. ● With the FTIR, you’ll just simply get all the ● Yung iba naman, there are some who would data coming from the sample. prepare the solid materials, as a dispersion ● For the other one, yung isa pang type ng in mineral oil. The process is the same. It is detector, Attenuated Total Reflectance IR. In simply mixing the two components: mineral here, it occurs when a beam of radiation oil and the sample and then placed it on the enters from a more dense (with a higher sample cell. refractive index) into a less dense medium ● The other one, are the gasses where the (with a lower refractive index) cells consist of a glass or metal body with ● So what happens is that, the sample, or the two iron transparent ending. light coming from the radiation source, it will ● For pharmaceutical purposes, we don’t just heat the sample on one portion. really encounter gasses. But for industry (potassium bromide). who would still assay gasses in IR, then eto ● And then what happens is, kung isang yung ginagawa, so possible din siya. portion lang, paano mo mapapadami yung ● From the sample cell, we go to the portion of signal that can reach the detector? the detector. There are so many detectors ● So what happens is, it will try to heat the right now that are available. sample on cellular points by placing the ● After IR absorption of the sample, the crystals with different refractive index, which sample will either throw it to the means, it will bend the light again in this monochromator for further dispersion of light type, on several points. ● So that, instead of just heating the sample active ingredient, which one is coming from on one end, kapag binalik niya, natamaan all the other excipients? na yung ibang fractions. ● The technique here is to always have a ● So ang magandang gawin ay is to get “baseline” data. Ibig sabihin may data na signals from as many areas of the sample dapat kung ano yung mga ginagamit na as possible by just one beam of light. excipients. Sila mismo may IR spectra na. ● So instead of irradiating the whole pellet, ● Pwedeng may library. Pwedeng may kumbaga marerecycle lang muna yung characteristic siya na hindi macacapture lalo portion na yun na nilight. Paulit-ulit lang siya pag nagiiba yung source nung excipient na by the use of high refractive index crystals. yun ● So instead of irradiating the whole pellet, ● So, what can be done in a laboratory is you kumbaga marerecycle lang muna yung actually build your own library. So ang portion nay un na nilight. Paulit-ulit lang siya gagawin lang kahit literature kasi maraming by the use of high refractive index crystals. pwedeng madownload for free is that to ● So yung sinasabing low dense medium, it check your own raw material. will now become a sample, yung ● You then check it on your access library. pangalawang media para magkaroon ng Kapag maganda yung matching edi good. bending of light. But then again, you should still keep your ● And then, the other one, which is the Diffuse own library and then again, label from what Reflectance IR. It is when light is reflected manufacturer it is. Because that might spell off a matt surface, the light observed is of out the presence or absence of impurities. the same intensity no matter what the angle So yun ung isang titingnan. of observation. ● So from there, kapag meron ka nang library ● So what happens here is when the sample nung mga excipients, once you get your prepared, as a powder, (potassium bromide sample, you can now subtract the baseline pellet), you don’t expect the surface of the or the background coming from the sample to be actually smooth. excipients of the creams and the tablets. ● Unlike kanina sa hitsura nung attenuated ● It can also be used as a method for reflectance, it appears that the sample is just fingerprint test, films, coatings, and a thin film on the sample cell but on packaging plastics. actuality, hindi siya nangyayari na ganun. ● Parts also of the quality control, training ● During the preparation, as much as possible, you try to level it properly but the kasi dito sa college, we only do quality reality is magkakaroon talaga siya ng control of dosage form. Part pa rin ng rooms(?). So yan yung advantage na process is actually doing quality control of inooffer ng diffuse reflectance. the packaging materials. Yung primary niya ● So, what happens is, instead of capturing all for example are the films and coatings in these areas, ang mangyayari sa kanya is capsule. Or even halimbawa yung mga isang region lang ang icacapture niya. Kaya tablets na nakablister pack. Lahat yun dapat yung sinasabi diyan na no matter what the dumadaan sa quality control. angle of observation is macacapture niya yung data. ● Imagine kapag gagawa ka ng plastic. You ● So yun ung mga common na ginagamit: can not prepare it in a solution. Di mo yun Diffuse, Attenuated Total Reflectance or the magagawa using the usual techniques here most common FTRI. in our college. Remember that these can be ● In terms of application, compound prepared in a pellet with potassium identification is still the number application of bromide. There is no dissolution going on. It IR. – is still part of ID tests of sampling is just mixing there. Magkakaroon ngayon compound. ng response yung films and the coatings, ● It is use as characterization of sample in a solid and semi-solid for creams and tablets. and even packaging plastics. ● If it is in dosage form, how will you able to distinguish which signal is found from the Another use of IR is detection of polymorph Determination of physico-chemical properties formation using NIR - Polymorph formation causes loss of - Water content: The presence of water in pharmacologic activity the drug product causes noise or - This may be detected only by IR (i.e. distortion in the fingerprint region UV/Vis, etc cannot detect Advantages for NIR polymorphism) - Good penetration properties which gives Near Infrared Spectroscopy it its ability to analyze substances while - Emerging technique in analysis in their original packaging (due to it - Wavelength range of 700-2500 nm having higher energy as compared to - It borders the visible region (which other wavelengths of IR) technically extends up to 900 nm) - More intense radiation sources ( does - Concept is still the same as that of not use the oxides of metals like regular IR (vibrational or rotational thorium, yttrium and zirconium) energy levels etc etc) - Rapid analysis as was mentioned Applications of NIR previously - Multicomponent analysis of dosage - Still on the validation phase forms in a quantitative or Limitation of IR semiquantitative manner - Expensive since demand is still not high - Fingerprint check for the identity of a - Requires more method development drug in QC of complex excipients (Just (One of the reasons why is because like normal IR) analysts are still trying to figure out how - Determination of physico-chemical to take into account the baseline properties of drugs and excipients (parang blank) measurements for the - Others in the ppt packaging materials. Different Multicomponent analysis of dosage forms using manufacturers of packaging will NIR definitely give off different signals, along - Quantitative analysis may be done with different materials and different without taking the drug out of its layers of packaging.) packaging - New technologies are being developed such as NIR gun (portable NIR device Mass Spectroscopy for easier and faster QC) - Another technique used for structural - NIR can pass through the layers of elucidation packaging of a drug Principle of Mass Spectroscopy - This may also be used for the QC of raw - Charged fragments from the parent materials that are still in their primary molecule (whose formation was caused containers (in big barrels, etc) by bombardment of electrons) will move - Main advantage lies in the ease and across the device through the use of quickness kasi kapag kukuha pa ng electric or magnetic fields sample sa barrel, may mga procedure - The difference in movement of these pa (kelangan pa magrandom sampling, fragments is primarily (not always) due etc) to the mass to charge ratio - Currently being validated; not yet official ● On some types of mass spec in compendia instruments, it’s not just the difference in - Sometimes used in the preliminary the mass to charge ratios but also on testing of raw materials. If the raw the kinetic energy that the molecule or material fails the NIR gun, definitely may the fragments will carry along the something wrong sa raw material. BUT if process. the substance passes NIR, it must be ● But on most instruments, puro mass to further supplemented with other tests in charge ratio lang ang tinitignan for the order to conclude that the substance is spectrometer to capture. of acceptable quality. Parts of mass spec ● Inlet source - where the sample is - Bawal yung galing sa solid. Kasi if galing sa introduced. The final introduction of the solid, hindi completely ma volatilized. May mga sample can be done by attaching to a maiiwan na solid. chromatographic instrument. - If it has to be introduced as an aqueous Continuous Infusion (sample solution, Formic acid or Ammonium Formate introduction) - it introduces the solution yung ginagamit. continuously over a time. It uses a direct insertion probe If the sample is introduced as liquid, the choice of solvent is that: -Yung pag drop niya, sobrang konti lang. 1.) Solubility Why that small? 2.) Volatility -As mentioned in the video, the sensitivity of the 3.) No solid should be deposited on the mass spec goes as little as 10^-12 moles. sample introduction system Sobrang liit niya. The instrument is highly valued for trace analysis. For pharmaceutical purposes, Some of the sample will not go through the trace analysis of impurities or degradation mass spec. products. It’s even better on biological systems. Examples: Therapeutic Drug monitoring, Pag may deposition of particles, cleaning is ADMETox studies (Absorption, Distribution, performed by placing isopropyl alcohol and the Metabolism, Excretion and Toxicology) surface is scrubbed using a microparticle sandpaper. Parang Velvet using feel niya. All For direct insertion probe- There is really no solid particles are scrubbed off so that in the requirement except that the sample should be next introduction of the sample, hindi masama delivered as liquid. yung solids na yun. -Choose a liquid where it is soluble and the readily volatilizes Another requirement should be the injection should be in microliter quantities. For every -Limonene- Dissolved in dichloromethane; eluate drop, it should be in microliter qualities. Limonene is soluble in DCM; DCM has the advantage of being easily volatilized. ● Excess volume of liquid from the How hot is the area where all of this volatilization chromatographic system will be occurs? vacuumed out so that only a -The sample introduction point can have a representative amount will be put into temperature as high as 250 degrees Celsius the mass spectrometer ● Atmospheric pressure chambers make it Do we expect degradation at 250 degrees so that some ion sources in some cases Celsius? do not need an initial vacuum -Probably, it can have but to prevent environment; however, for the most part degradation, the residence time of the sample most ion sources are in a vacuum introduction system should be as short as ● It is essential to force the sample to possible. enter the mass spec ASAP to prevent How do you keep that short? degradation - Everything should be in in vacuum. From here ● Mass specs use several vacuum pumps pagka heat up, diretso sa mass spec. to keep the environment, well, in Millisecond residence time. vacuum; takes time to establish a vacuum eg 3 whole days ● In sample introduction system, -Kailan, on the process, maibato kaagad yung fragmentation of molecules does not solute papunta doon sa mass spec occur even with subjection to high heat; -In cases where buffers are use, what is fragmentation occurs in ionization recommended is that the component of the chamber—may be initial as some mass buffer should be volatile also. specs are connected to one or even two other mass specs in one system ○ Why should there be more ionizations? ■ Higher resolution ■ To be able to differentiate APIs from impurities and other degradation products or just compounds in general; one mass spec will not be able to Different methods use different energy sources differentiate well - low to bombard the sample/ionizing agent for resolution mass spec desorption. The only way for the sample to be fragmented is to place enough energy that the Ionization methods electron or the functional group will be knocked 1. Gas phase off in the process. - sample is vaporized then ionized How will you know if the compound is - Restricted to thermally stable undergoing degradation? compounds unless you know It is indicated in references the amount of degradation products present. Hard ionization - Small molecules (MW < 10 kDa) - Many molecules/fragments are created because gas phase is often - Impart sufficient energy on the analyte associated with hard ionization to leave the molecule in an excited state sources in which relaxation results in the rupture of bonds 2. Desorption - Not good for large molecules (as high as - Sample (solid or liquid) is 100 kDa) because the spectra will be converted directly to gaseous too complicated ions - Molecular ion peak may be completely - simultaneous ionization and gone or negligible vaporization - sample is usually placed in a Soft ionization matrix since the samples are - Cause very little fragmentation and the often hard to volatilize without spectra have few, if any, peaks besides degrading them the molecular ion (M) peak - Large molecules because desorption is associated with If you have a reference standard or if you know soft ionization the associated impurities will probably not have the same molecular weight then field ionization Gas Phase Instruments or sources may be used. Otherwise, use harder ionization sources.
Electron Impact - earliest ionization source for
mass spec - Still very much in use - Hardest of the hard ionization sources Desorption phase sources - Bombardment with a beam of energetic electrons - 70 electron-volts (eV) produced by a radium or tungsten filament which are accelerated towards a positive target or the anode - Sa video earlier, the electron passes - Methane’s natural progression is to perpendicular to the movement of the stabilize itself sample so that there is maximum impact - Magkakaroon ka ng charged - Isipin niyo pag parallel lang sila methane and a proton may impact naman pero mas - Yung charged methane can less likely kasi they are all impact or can bombard another moving in a single direction methane. - Hindi siya yung magwawander - The initial collision of the from one side to the other ‘tong electron with the methane gas mga electrons na to will give enough energy to - The stream coming from the unionize methane chromatographic instrument (inlet) and - May excess energy to form additional the stream from the ion source will ions impact - If a sample hits a charged methane, - At 70 eV, you immediately get to knock magkakaroon ng fragmentation off one electron from the sample - Question: Bakit di na lang yung electron - As long as it has not yet impact since electron din naman yung reached the anode, it can be ginagamit pang bombard doon sa gas? reused (since it still has high - Answer: If the analyst would energy) to impact the fragments want lesser fragmentation, resulting from the initial pipiliin mo yung chemical fragmentation hanggang sa ionization. [T/N: pero pwede rin magkaroon ng stable na naman ako ;) ] Ayaw mo lang fragment yung sobrang taas na energy bombarding the sample. Chemical Ionization - hard ionization source - If the need is to have an energy - Gaseous atoms of the sample are less than 70 eV, Chemical ionized by collision with ions produced Ionization na yung gagamitin mo by particle bombardment of an excess Negative Ion Chemical Ionization (NICI) - hindi of a reagent gas chemical yung nambobombard sa sample - Yung chemical na gagamitin for - Fake chemical ionization :( bombardment dapat mataas yung - You still have the collision of the energy kasi walang mangyayari pag electron with the reagent gas but the normal reagent lang result is instead of the reagent gas - Kailangan maglagay ng excess energy colliding with the sample, the electron sa reagent gas so there should be an which collided with the reagent gas will initial collision of high energy going collide with the sample. towards the reagent gas - When the electron collides with the - 2 types: Positive Ion Chemical Ionization reagent gas, some of the energy carried (PICI) and Negative Ion Chemical by the electron will be transferred to Ionization (NICI) this(?) - Nag-iisa lang ang true chemical - Yung electron sa NICI, since it is ionization (dun dun duuuuuun) decreased in energy, it will be the one Positive Ion Chemical Ionization (PICI) that will bombard the sample (Methane gas reacting with high-energy - These low-energy thermal electrons) electrons are then captured by - It is hard to imagine how to destabilize a molecules with high affinity for methane molecule (looks very stable electrons kasi simple molecule lang siya) but if - Yung difference niya sa electron impact you place enough energy (e.g. electron is the amount of energy carried by the impact ionization), you get to knock off electron one electron from methane which - Naging chemical ionization ang NICI just means nag-impart ka ng energy because you need a chemical to reduce papunta doon sa methane the energy carried by the electron. Otherwise, it is not actually the chemical ● Th 3000 volts is just for the initial that bombards the sample ionization - If the ion source makes use of a gas ● At this rate, the sample will already be (e.g. methane), dadagdagan yung charged. Fragmentation has not yet diagram ng gas tank para masabi na started. Bakit? Probably the sample is yun yung ion source already charged, but the real fragmentation occurs as it passes Comparing the Hard Ionization Sources through the plates which are also (Psoralen - furocoumarin for vitiligo mga chong electrostatically charged. ko na nag-1-128) ● And then the charged molecules are drawn into the mass spec by Electron Impact - maraming fragments electrostatically charged plates, where PICI and NICI - less fragments further fragmentation occurs ● Nitrogen gas placed on this portion for Lahat naman ay gumagamit ng electrons drying purposes. Kasi kung titingnan, while there is heat involved, to further Electron Impact - directly utilized by the sample remove the solvent, dry nitrogen gas is PICI - utilized by the reagent gas then reagent also continuously infused to the gas bombards the sample electrospray ionization chamber. NICI - electron has reduced energy by collision Continuous infusion means it will get the with reagent gas then lower energy electron will nitrogen from the air, purify that, pass bombard the sample the nitrogen going to the machine, and then the rest is carbon dioxide and Field Ionization sources oxygen. - the answer for under the influence of a ● There is a hose at the back where large electric field carbon dioxide and oxygen is released - have 10 to 20 kilovolts of energy applied ➢ Why not just use a nitrogen to 10 micrometer diameter wire where tank? little fragmentation occur Because if you are doing a continuous analysis,continuous Thermospray drying is also involved. Even if - the most commonly used ionization there is no equipment plugged, source is the electrospray ionization nitrogen flow is continuous. 12 sources or ESI liters per min ang rate ng - A charged aerosol is generated at infusion of nitrogen gas, which atmospheric pressure and then the ensures removal of solvent. So charged molecules are drawn into the ayun, mabilis lang ang buhay ng mass spec by electrostatically charged nitrogen gas; on a long term weights. basis, it is not economical. - In the diagram, this ionization is usually ● Same mechanism applies with your TLC in tandem with liquid chromatography applicator. There is a nitrogen tank on technique. - the side, but not running at 12 L/min; it’s - So anong nangyayari? a lot less. Its purpose is to dry the ● From the capillary outlet of liquid solvent off the sample. chromatography, nakadirect siya ● So there is nitrogen gas for the sample, papunta sa mass spec. In itself, the and then there is nitrogen gas again if capillary is already subjected to a high the solvent is not completely vaporized voltage of 3 kV (3000 volts). in the process. ● Charged up na yung sample palabas pa ● “Electro” portion- the electric field given lang siya. Since 3000 volts siya, by the 3000 volts kailangan niya ng sariling electrical line ● “Spray” portion- as soon as the sample *And this is just the ionization source, gets out of the capillary, parang mist na wala pa ng ibang energy from the other siyang lalabas papunta sa parts of the mass spec* electrostatically charged plates. * Nitrogen gas has 2 purposes: to Matrix -assisted Laser Desorption Ionization – remove the solvent, and to create a fine also placed in a matrix that should absorb UV mist of the sample radiation - If you have instruments which involve - the energy from the laser is transferred to the high heat and high volatilization, its is matrix, the excess energy ibibigay niya sa really expected that some solids get molecule which is mixed with the matrix and the deposited on any of these parts. transfer of energy will create fragments - To clean this: ● If the nitrogen gas is not stable anymore For more information, you can read Mass in terms of blowing the solvent off, Spec for Ionization Techniques for Pharmacy remove sonication device Students (2ndedition) by David Watson. For ● A needle at the tip will cause the spray the IR and other parts of Mass Spectroscopy, effect. It is easy to remove the needle read Pharmaceutical Analysis: A Textbook but difficult to put back because there is for Pharmacy Students and Pharmaceutical the danger of bending the needle if it Chemists also by David Watson (Same ang hits a metal surface. nasa ppt ni ma’am sa mga nakalagay sa - If you have an instrument which involves book na to). high heat and volatilization, it is expected that some solids get deposited - end - in any of these parts.
Atmospheric Pressure Ionization – similar to ES,
but run at usual LC flow rates and used for less polar molecules - meron ka pa ring nitrogen gas to help vaporize the solvent and to create a fine mist of the sample and the solvent but yung creation of the fine mist sa second source nangyayari - the added nitrogen gas helps in treating the less polar molecules - at high liquid chromatography flow rates kailangang maraming nitrogen to remove the solvents kasi mataas ang volume load and kailangan minimal amount lang ang makarating sa mass spec para hindi masaturate ang mass spec - this occurs at atmospheric pressure - the 3000 kilovolts is still there but it’s not a part of the LC capillary and is a separate mechanism; hindi simultaneous because of the volume load
Thermospray – the volatilization is done by
heating the capillary
Fast Atom Bombardment – the sample should
be first run with a matrix - for samples who do not easily volatilize - one of the classic example of desorption phase - the sample is mixed with glycerol to hold it - the source of energy is a fast atom