Vous êtes sur la page 1sur 10

PhCh 136 Lec 2nd Exam - Only bonds that have dipole moment

1st Trans are able to absorb IR (hence the reason


April 3, 2019 why homonuclear molecules lack IR
absorption)
(Sorry guys hindi narecord ang konteng first part Types of vibrations involved in IR
ng IR Spectro kaya I will try my best to fill in the 1. Stretching Vibrations
first part based on information from references - Symmetric
such as Harvey or Skoog :( - Asymmetric
Infrared Spectroscopy 2. Bending Vibrations
- IR Region is from around 1 nm to - Twisting
around 1 mm - Wagging
- IR Spectro is typically associated only - Scissoring
with vibrational portion of IR (2.5 nm to - Rocking
25 nm) *Each of these types of vibrations will have a
- Important in the identification signature signal in the IR spectrum
(qualitative) of both organic and Another thing to look for in IR spectrum:
inorganic compounds because all Intensity of the band
molecular species absorb IR (Except for Intensity
homonuclear molecules such as O​2​, N​2​, - Tells the type of functional group
etc) present
- Important in quali also because every - ↑ dipole moment, ↑ intensity
molecule has a unique IR spectrum - ↑ electronegativity, ↑ intensity
(except for chiral molecules in the - ↑conjugation to polar double bond, ↑
crystalline state) (Skoog, et. al., 2013) intensity
- Wavelength is in terms of Energy Level (which is in terms of
wavenumbers, which is the reciprocal of wavenumbers)
the wavelength in cm - Molecules with similar functional groups
- Wavenumber is more preferred than will absorb photons of similar energies
wavelength because it is directly - The higher the atomic mass difference,
proportional to energy (higher the higher the energy of absorption
wavenumber corresponds to higher needed
energy) - Double bonds are stronger than single
IR Spectro for quantitative analysis? bonds and therefore absorb at higher
- It is less satisfactory than its UV/Vis energy than single bonds
counterpart since it offers less Is IR important in analytical chemistry?
sensitivity, less precision and higher - In certain monographs, IR spectra are
deviation from Beer’s Law needed in the identification of these
- IR radiation is less intense than UV/Vis, certain compounds
therefore more prone to stray radiations - Not all compounds need IR spectra
- However, IR MAY BE USED - There is no pattern that will tell you
quantitatively if precision requirements whether a compound will need an IR
are not that strict since it gives the spectra or not
advantage of being quite selective - Most of the time, monographs require
(Skoog, et. al., 2013) only physical characteristics or physical
Process of IR absorption constants for the identification of
- IR is lower energy than UV/Vis, compounds; sometimes, UV spectrum is
therefore it can only affect the stretching needed
and bending of bonds, as compared to - IR is usually used for identification of
UV/Vis wherein jumping of electrons to structure of compounds of unsure
a different energy state is concerned identity
Requirements for IR absorption IR in QC
- Only the frequencies of IR that match - IR is being looked at nowadays as a
the frequencies of the natural vibration potential QC test
of the molecule is absorbed
- This is because the sample preparation ● If the make up of the monochromator is not
in IR is as easy and fast as that of in as good, if the gratings are not as high (less
UV/Vis gratings = less resolution), they actually lose
- QC has to be as fast as possible some of the data. Ang nangyayari yung read
because it is a requirement in out from a dispersing type is a little cleaner.
manufacturing (wherein time is of the Nawawala kasi yung ibang data from the
essence) monochromator.
2 types of IR instruments
- One on the left side of the ppt is Source of radiation for IR: oxides of zirconium,
continuous or dispersive type of yttrium, or thorium
instrument ● Gives a wavelength from 2-15 micrometers
- One on the right is the FTIR or the ● Non-conducting at room temperature
Fourier Transform IR machine ● Alternative: glow bar → gives a less intense
Continuous IR type of radiation
- Earliest type
- Gives less information as compared to Monochromator
the FTIR ● Will only appear in the continuous or
- Not that common nowadays dispersive type of an IR instrument (older
- Sa FTIR kasi mas clean ang data and type of the instrument)
mas complete ● Most commonly used: grating type
- Continuous IR gives very general ● If the grooves are not as fine (mas
information malalawak yung distances), data loss will
How Continuous IR works come in because hindi talaga nahihiwa yung
- IR radiation is given off by a filament of wavelength coming from the sample.
zinc oxide or yttrium or thorium oxides (Remember: 2-15 micrometers lang yung
range ng IR)
All of the wavelengths associated with the IR ● It should be able to chop 2-15 micrometers
region will go to the sample. After that kapag into as many wavenumbers or wavelengths
nagkaroon na ng absorption (rotational to as possible.
vibrational energy levels), it will give out again ● Yung one micrometer, malaki na yun. It can
that light na icacapture ng monochromator one differentiate a C-O absorption and a C-Cl
wavelength at a time. absorption.
● Kapag mas marami [ang grooves/gratings],
FTIR mas okay.
● Still, the source will give out all the ● Mas maganda ang FTIR because it will not
wavelengths of that energy source have this disadvantage of using a
(zirconium oxide, yttrium oxide, or thorium monochromator. All of the data will be
oxide) then it will bring it to the sample captured. There is no filtering involved.
simultaneously. The sample will give the
data simultaneously. Sample cell
● Instead of getting the absorption data one ● Where the sample is placed
wavelength at a time, the moving mirror will ● If you have liquid sample, it should be as
capture all of the information simultaneously much as possible anhydrous in nature. It
also. can be tested as neat (undiluted form of the
● Mathematical equation of Fourier Transform sample; no added liquids to deliver the
-- inaayos yung data na simultaneously sample into the IR)
captured. ● It can also be diluted using carbon
● There is no data loss in the process tetrachloride or carbon disulfide in a
because everything is captured short-pathed cell.
simultaneously. There is no selection or ● It’s still possible to have problems with
filtering unlike that of continuous or saturation → dilute!
dispersed type. Pero essentially there is ● Carbon tetrachloride and carbon disulfide
filtering done by the monochromator. should not give any interfering response to
the rest of the spectra
or throw it to the moving mirror for the FTIR
Solids type.
● KBr disc ● From those mechanism, it will throw those
○ KBr will just be mixed thoroughly with signals into the detector.
the sample in a mortar and pestle ● There are several detectors right now, each
○ Sample is mixed in a particular ration with an advantage.
depending on the sensitivity of the ● The earliest ones are the transmission type
sample of the detector or the courier transform
○ E.g. 10 mg of the sample + 100 mg KBr which is still the most widely used. It uses a
→ 10% lang ng buong preparation signal from there which relays the radiation
○ standard source in which individual can be monitor in
within a one second of radiation can be
● NaCl monitor in a one second pulse of radiation.
○ Alternative when KBr cannot be used ● The important aspect here with courier
○ Preferred when the solid is in chloride transform is how fast it can generate the (?)
form so that halogen exchange will not ● So pag sinabing one second pulse(?), all the
occur → kailangan pareho yung salt data are captured in a matter of seconds.
form That’s why for FTRI instruments, the data
○ If hindi nakasalt form, there is no capture or the data generation will only
problem. Iniiwasan lang magkahalogen takes seconds.
exchange in the process. ● Pagkaprepare ng sample, paglagay dun sa
instrument, in a matter of second you can
Paano ba nagkakaroon ng halogen exchange? get the results. Again, no filtering involved.
● These materials are just mixed in a mortar ● But when a monochromator is used, the
and pestle. When these are made in a process can take from 15-30 minutes,
pellet, pressure is actually placed on the before a data is generated.
sample. ● ·Madalas naman hindi siya nagiging
● So, para kang gagawa ng waffle. Kapag problem, just because you are placing
gumagawa ka nun may presser ka. So samples which are relatively stable inside
ganun na ganun siya. Papaumbukin mo the instrument.
siya. And then, you really have to press on ● There is no heating involved in this. The light
that so that it becomes a hard pellet. will just go through the sample, a low energy
● Once you placed pressure, the tendency is light will just go through a sample. At one
you place enough energy for halogen point, data capturing lang ang magtatagal.
exchange to occur in the process. ● With the FTIR, you’ll just simply get all the
● Yung iba naman, there are some who would data coming from the sample.
prepare the solid materials, as a dispersion ● For the other one, yung isa pang type ng
in mineral oil. The process is the same. It is detector, Attenuated Total Reflectance IR. In
simply mixing the two components: mineral here, it occurs when a beam of radiation
oil and the sample and then placed it on the enters from a more dense (with a higher
sample cell. refractive index) into a less dense medium
● The other one, are the gasses where the (with a lower refractive index)
cells consist of a glass or metal body with ● So what happens is that, the sample, or the
two iron transparent ending. light coming from the radiation source, it will
● For pharmaceutical purposes, we don’t just heat the sample on one portion.
really encounter gasses. But for industry (potassium bromide).
who would still assay gasses in IR, then eto ● And then what happens is, kung isang
yung ginagawa, so possible din siya. portion lang, paano mo mapapadami yung
● From the sample cell, we go to the portion of signal that can reach the detector?
the detector. There are so many detectors ● So what happens is, it will try to heat the
right now that are available. sample on cellular points by placing the
● After IR absorption of the sample, the crystals with different refractive index, which
sample will either throw it to the means, it will bend the light again in this
monochromator for further dispersion of light type, on several points.
● So that, instead of just heating the sample active ingredient, which one is coming from
on one end, kapag binalik niya, natamaan all the other excipients?
na yung ibang fractions. ● The technique here is to always have a
● So ang magandang gawin ay is to get “baseline” data. Ibig sabihin may data na
signals from as many areas of the sample dapat kung ano yung mga ginagamit na
as possible by just one beam of light. excipients. Sila mismo may IR spectra na.
● So instead of irradiating the whole pellet, ● Pwedeng may library. Pwedeng may
kumbaga marerecycle lang muna yung characteristic siya na hindi macacapture lalo
portion na yun na nilight. Paulit-ulit lang siya pag nagiiba yung source nung excipient na
by the use of high refractive index crystals. yun
● So instead of irradiating the whole pellet, ● So, what can be done in a laboratory is you
kumbaga marerecycle lang muna yung actually build your own library. So ang
portion nay un na nilight. Paulit-ulit lang siya gagawin lang kahit literature kasi maraming
by the use of high refractive index crystals. pwedeng madownload for free is that to
● So yung sinasabing low dense medium, it check your own raw material.
will now become a sample, yung ● You then check it on your access library.
pangalawang media para magkaroon ng Kapag maganda yung matching edi good.
bending of light. But then again, you should still keep your
● And then, the other one, which is the Diffuse own library and then again, label from what
Reflectance IR. It is when light is reflected manufacturer it is. Because that might spell
off a matt surface, the light observed is of out the presence or absence of impurities.
the same intensity no matter what the angle So yun ung isang titingnan.
of observation. ● So from there, kapag meron ka nang library
● So what happens here is when the sample nung mga excipients, once you get your
prepared, as a powder, (potassium bromide sample, you can now subtract the baseline
pellet), you don’t expect the surface of the or the background coming from the
sample to be actually smooth. excipients of the creams and the tablets.
● Unlike kanina sa hitsura nung attenuated ● It can also be used as a method for
reflectance, it appears that the sample is just fingerprint test, films, coatings, and
a thin film on the sample cell but on packaging plastics.
actuality, hindi siya nangyayari na ganun.
● Parts also of the quality control, training
● During the preparation, as much as
possible, you try to level it properly but the
kasi dito sa college, we only do quality
reality is magkakaroon talaga siya ng control of dosage form. Part pa rin ng
rooms(?). So yan yung advantage na process is actually doing quality control of
inooffer ng diffuse reflectance. the packaging materials. Yung primary niya
● So, what happens is, instead of capturing all for example are the films and coatings in
these areas, ang mangyayari sa kanya is capsule. Or even halimbawa yung mga
isang region lang ang icacapture niya. Kaya tablets na nakablister pack. Lahat yun dapat
yung sinasabi diyan na no matter what the
dumadaan sa quality control.
angle of observation is macacapture niya
yung data.
● Imagine kapag gagawa ka ng plastic. You
● So yun ung mga common na ginagamit: can not prepare it in a solution. Di mo yun
Diffuse, Attenuated Total Reflectance or the magagawa using the usual techniques here
most common FTRI. in our college. Remember that these can be
● In terms of application, compound prepared in a pellet with potassium
identification is still the number application of bromide. There is no dissolution going on. It
IR. – is still part of ID tests of sampling is just mixing there. Magkakaroon ngayon
compound.
ng response yung films and the coatings,
● It is use as characterization of sample in a
solid and semi-solid for creams and tablets.
and even packaging plastics.
● If it is in dosage form, how will you able to
distinguish which signal is found from the
Another use of IR is detection of polymorph Determination of physico-chemical properties
formation using NIR
- Polymorph formation causes loss of - Water content: The presence of water in
pharmacologic activity the drug product causes noise or
- This may be detected only by IR (i.e. distortion in the fingerprint region
UV/Vis, etc cannot detect Advantages for NIR
polymorphism) - Good penetration properties which gives
Near Infrared Spectroscopy it its ability to analyze substances while
- Emerging technique in analysis in their original packaging (due to it
- Wavelength range of 700-2500 nm having higher energy as compared to
- It borders the visible region (which other wavelengths of IR)
technically extends up to 900 nm) - More intense radiation sources ( does
- Concept is still the same as that of not use the oxides of metals like
regular IR (vibrational or rotational thorium, yttrium and zirconium)
energy levels etc etc) - Rapid analysis as was mentioned
Applications of NIR previously
- Multicomponent analysis of dosage - Still on the validation phase
forms in a quantitative or Limitation of IR
semiquantitative manner - Expensive since demand is still not high
- Fingerprint check for the identity of a - Requires more method development
drug in QC of complex excipients (Just (One of the reasons why is because
like normal IR) analysts are still trying to figure out how
- Determination of physico-chemical to take into account the baseline
properties of drugs and excipients (parang blank) measurements for the
- Others in the ppt packaging materials. Different
Multicomponent analysis of dosage forms using manufacturers of packaging will
NIR definitely give off different signals, along
- Quantitative analysis may be done with different materials and different
without taking the drug out of its layers of packaging.)
packaging
- New technologies are being developed
such as NIR gun (portable NIR device Mass Spectroscopy
for easier and faster QC) - Another technique used for structural
- NIR can pass through the layers of elucidation
packaging of a drug Principle of Mass Spectroscopy
- This may also be used for the QC of raw - Charged fragments from the parent
materials that are still in their primary molecule (whose formation was caused
containers (in big barrels, etc) by bombardment of electrons) will move
- Main advantage lies in the ease and across the device through the use of
quickness kasi kapag kukuha pa ng electric or magnetic fields
sample sa barrel, may mga procedure - The difference in movement of these
pa (kelangan pa magrandom sampling, fragments is primarily (not always) due
etc) to the mass to charge ratio
- Currently being validated; not yet official ● On some types of mass spec
in compendia instruments, it’s not just the difference in
- Sometimes used in the preliminary the mass to charge ratios but also on
testing of raw materials. If the raw the kinetic energy that the molecule or
material fails the NIR gun, definitely may the fragments will carry along the
something wrong sa raw material. BUT if process.
the substance passes NIR, it must be ● But on most instruments, puro mass to
further supplemented with other tests in charge ratio lang ang tinitignan for the
order to conclude that the substance is spectrometer to capture.
of acceptable quality.
Parts of mass spec
● Inlet source - where the sample is - Bawal yung galing sa solid. Kasi if galing sa
introduced. The final introduction of the solid, hindi completely ma volatilized. May mga
sample can be done by attaching to a maiiwan na solid.
chromatographic instrument. - If it has to be introduced as an aqueous
Continuous Infusion (sample solution, Formic acid or Ammonium Formate
introduction) - it introduces the solution yung ginagamit.
continuously over a time. It uses a direct
insertion probe If the sample is introduced as liquid, the choice
of solvent is that:
-Yung pag drop niya, sobrang konti lang. 1.) Solubility
Why that small? 2.) Volatility
-As mentioned in the video, the sensitivity of the 3.) No solid should be deposited on the
mass spec goes as little as 10^-12 moles. sample introduction system
Sobrang liit niya. The instrument is highly valued
for trace analysis. For pharmaceutical purposes, Some of the sample will not go through the
trace analysis of impurities or degradation mass spec.
products. It’s even better on biological systems.
Examples: Therapeutic Drug monitoring, Pag may deposition of particles, cleaning is
ADMETox studies (Absorption, Distribution, performed by placing isopropyl alcohol and the
Metabolism, Excretion and Toxicology) surface is scrubbed using a microparticle
sandpaper. Parang Velvet using feel niya. All
For direct insertion probe- There is really no solid particles are scrubbed off so that in the
requirement except that the sample should be next introduction of the sample, hindi masama
delivered as liquid. yung solids na yun.
-Choose a liquid where it is soluble and the
readily volatilizes Another requirement should be the injection
should be in microliter quantities. For every
-Limonene- Dissolved in dichloromethane; eluate drop, it should be in microliter qualities.
Limonene is soluble in DCM; DCM has the
advantage of being easily volatilized.
● Excess volume of liquid from the
How hot is the area where all of this volatilization chromatographic system will be
occurs? vacuumed out so that only a
-The sample introduction point can have a representative amount will be put into
temperature as high as 250 degrees Celsius the mass spectrometer
● Atmospheric pressure chambers make it
Do we expect degradation at 250 degrees so that some ion sources in some cases
Celsius? do not need an initial vacuum
-Probably, it can have but to prevent environment; however, for the most part
degradation, the residence time of the sample most ion sources are in a vacuum
introduction system should be as short as ● It is essential to force the sample to
possible. enter the mass spec ASAP to prevent
How do you keep that short? degradation
- Everything should be in in vacuum. From here ● Mass specs use several vacuum pumps
pagka heat up, diretso sa mass spec. to keep the environment, well, in
Millisecond residence time. vacuum; takes time to establish a
vacuum eg 3 whole days
● In sample introduction system,
-Kailan, on the process, maibato kaagad yung fragmentation of molecules does not
solute papunta doon sa mass spec occur even with subjection to high heat;
-In cases where buffers are use, what is fragmentation occurs in ionization
recommended is that the component of the chamber—may be initial as some mass
buffer should be volatile also. specs are connected to one or even two
other mass specs in one system
○ Why should there be more
ionizations?
■ Higher resolution
■ To be able to
differentiate APIs from
impurities and other
degradation products or
just compounds in
general; one mass spec
will not be able to Different methods use different energy sources
differentiate well - low to bombard the sample/ionizing agent for
resolution mass spec desorption. The only way for the sample to be
fragmented is to place enough energy that the
Ionization methods electron or the functional group will be knocked
1. Gas phase off in the process.
- sample is vaporized then
ionized How will you know if the compound is
- Restricted to thermally stable undergoing degradation?
compounds unless you know It is indicated in references
the amount of degradation
products present. Hard ionization
- Small molecules (MW < 10 kDa) - Many molecules/fragments are created
because gas phase is often - Impart sufficient energy on the analyte
associated with hard ionization to leave the molecule in an excited state
sources in which relaxation results in the rupture
of bonds
2. Desorption - Not good for large molecules (as high as
- Sample (solid or liquid) is 100 kDa) because the spectra will be
converted directly to gaseous too complicated
ions - Molecular ion peak may be completely
- simultaneous ionization and gone or negligible
vaporization
- sample is usually placed in a Soft ionization
matrix since the samples are - Cause very little fragmentation and the
often hard to volatilize without spectra have few, if any, peaks besides
degrading them the molecular ion (M) peak
- Large molecules because
desorption is associated with If you have a reference standard or if you know
soft ionization the associated impurities will probably not have
the same molecular weight then field ionization
Gas Phase Instruments or sources may be used. Otherwise, use harder ionization
sources.

Electron Impact - earliest ionization source for


mass spec
- Still very much in use
- Hardest of the hard ionization sources
Desorption phase sources - Bombardment with a beam of energetic
electrons
- 70 electron-volts (eV) produced by a
radium or tungsten filament which are
accelerated towards a positive target or
the anode
- Sa video earlier, the electron passes - Methane’s natural progression is to
perpendicular to the movement of the stabilize itself
sample so that there is maximum impact - Magkakaroon ka ng charged
- Isipin niyo pag parallel lang sila methane and a proton
may impact naman pero mas - Yung charged methane can
less likely kasi they are all impact or can bombard another
moving in a single direction methane.
- Hindi siya yung magwawander - The initial collision of the
from one side to the other ‘tong electron with the methane gas
mga electrons na to will give enough energy to
- The stream coming from the unionize methane
chromatographic instrument (inlet) and - May excess energy to form additional
the stream from the ion source will ions
impact - If a sample hits a charged methane,
- At 70 eV, you immediately get to knock magkakaroon ng fragmentation
off one ​electron​ from the sample - Question: Bakit di na lang yung electron
- As long as it has not yet impact since electron din naman yung
reached the anode, it can be ginagamit pang bombard doon sa gas?
reused (since it still has high - Answer: If the analyst would
energy) to impact the fragments want lesser fragmentation,
resulting from the initial pipiliin mo yung chemical
fragmentation hanggang sa ionization. [T/N: pero pwede rin
magkaroon ng stable na naman ako ;) ] Ayaw mo lang
fragment yung sobrang taas na energy
bombarding the sample.
Chemical Ionization - hard ionization source - If the need is to have an energy
- Gaseous atoms of the sample are less than 70 eV, Chemical
ionized by collision with ions produced Ionization na yung gagamitin mo
by particle bombardment of an excess Negative Ion Chemical Ionization (NICI) - hindi
of a reagent gas chemical yung nambobombard sa sample
- Yung chemical na gagamitin for - Fake chemical ionization :(
bombardment dapat mataas yung - You still have the collision of the
energy kasi walang mangyayari pag electron with the reagent gas but the
normal reagent lang result is instead of the reagent gas
- Kailangan maglagay ng excess energy colliding with the sample, the electron
sa reagent gas so there should be an which collided with the reagent gas will
initial collision of high energy going collide with the sample.
towards the reagent gas - When the electron collides with the
- 2 types: Positive Ion Chemical Ionization reagent gas, some of the energy carried
(PICI) and Negative Ion Chemical by the electron will be transferred to
Ionization (NICI) this(?)
- Nag-iisa lang ang true chemical - Yung electron sa NICI, since it is
ionization (dun dun duuuuuun) decreased in energy, it will be the one
Positive Ion Chemical Ionization (PICI) that will bombard the sample
(Methane gas reacting with high-energy - These low-energy thermal
electrons) electrons are then captured by
- It is hard to imagine how to destabilize a molecules with high affinity for
methane molecule (looks very stable electrons
kasi simple molecule lang siya) but if - Yung difference niya sa electron impact
you place enough energy (e.g. electron is the amount of energy carried by the
impact ionization), you get to knock off electron
one electron from methane which - Naging chemical ionization ang NICI just
means nag-impart ka ng energy because you need a chemical to reduce
papunta doon sa methane the energy carried by the electron.
Otherwise, it is not actually the chemical ● Th 3000 volts is just for the initial
that bombards the sample ionization
- If the ion source makes use of a gas ● At this rate, the sample will already be
(e.g. methane), dadagdagan yung charged. Fragmentation has not yet
diagram ng gas tank para masabi na started. Bakit? Probably the sample is
yun yung ion source already charged, but the real
fragmentation occurs as it passes
Comparing the Hard Ionization Sources through the plates which are also
(Psoralen - furocoumarin for vitiligo mga chong electrostatically charged.
ko na nag-1-128) ● And then the charged molecules are
drawn into the mass spec by
Electron Impact - maraming fragments electrostatically charged plates, where
PICI and NICI - less fragments further fragmentation occurs
● Nitrogen gas placed on this portion for
Lahat naman ay gumagamit ng electrons drying purposes. Kasi kung titingnan,
while there is heat involved, to further
Electron Impact - directly utilized by the sample remove the solvent, dry nitrogen gas is
PICI - utilized by the reagent gas then reagent also continuously infused to the
gas bombards the sample electrospray ionization chamber.
NICI - electron has reduced energy by collision Continuous infusion means it will get the
with reagent gas then lower energy electron will nitrogen from the air, purify that, pass
bombard the sample the nitrogen going to the machine, and
then the rest is carbon dioxide and
Field Ionization sources oxygen.
- the answer for under the influence of a ● There is a hose at the back where
large electric field carbon dioxide and oxygen is released
- have 10 to 20 kilovolts of energy applied ➢ Why not just use a nitrogen
to 10 micrometer diameter wire where tank?
little fragmentation occur Because if you are doing a
continuous analysis,continuous
Thermospray drying is also involved. Even if
- the most commonly used ionization there is no equipment plugged,
source is the electrospray ionization nitrogen flow is continuous. 12
sources or ESI liters per min ang rate ng
- A charged aerosol is generated at infusion of nitrogen gas, which
atmospheric pressure and then the ensures removal of solvent. So
charged molecules are drawn into the ayun, mabilis lang ang buhay ng
mass spec by electrostatically charged nitrogen gas; on a long term
weights. basis, it is not economical.
- In the diagram, this ionization is usually ● Same mechanism applies with your TLC
in tandem with liquid chromatography applicator. There is a nitrogen tank on
technique. - the side, but not running at 12 L/min; it’s
- So anong nangyayari? a lot less. Its purpose is to dry the
● From the capillary outlet of liquid solvent off the sample.
chromatography, nakadirect siya ● So there is nitrogen gas for the sample,
papunta sa mass spec. In itself, the and then there is nitrogen gas again if
capillary is already subjected to a high the solvent is not completely vaporized
voltage of 3 kV (3000 volts). in the process.
● Charged up na yung sample palabas pa ● “Electro” portion- the electric field given
lang siya. Since 3000 volts siya, by the 3000 volts
kailangan niya ng sariling electrical line ● “Spray” portion- as soon as the sample
*And this is just the ionization source, gets out of the capillary, parang mist na
wala pa ng ibang energy from the other siyang lalabas papunta sa
parts of the mass spec* electrostatically charged plates.
* Nitrogen gas has 2 purposes: to Matrix -assisted Laser Desorption Ionization –
remove the solvent, and to create a fine also placed in a matrix that should absorb UV
mist of the sample radiation
- If you have instruments which involve - the energy from the laser is transferred to the
high heat and high volatilization, its is matrix, the excess energy ibibigay niya sa
really expected that some solids get molecule which is mixed with the matrix and the
deposited on any of these parts. transfer of energy will create fragments
- To clean this:
● If the nitrogen gas is not stable anymore For more information, you can read Mass
in terms of blowing the solvent off, Spec for Ionization Techniques for Pharmacy
remove sonication device Students (2​nd​edition) by David Watson. For
● A needle at the tip will cause the spray the IR and other parts of Mass Spectroscopy,
effect. It is easy to remove the needle read Pharmaceutical Analysis: A Textbook
but difficult to put back because there is for Pharmacy Students and Pharmaceutical
the danger of bending the needle if it Chemists also by David Watson (Same ang
hits a metal surface. nasa ppt ni ma’am sa mga nakalagay sa
- If you have an instrument which involves book na to).
high heat and volatilization, it is
expected that some solids get deposited - end -
in any of these parts.

Atmospheric Pressure Ionization – similar to ES,


but run at usual LC flow rates and used for less
polar molecules
- meron ka pa ring nitrogen gas to help vaporize
the solvent and to create a fine mist of the
sample and the solvent but yung creation of the
fine mist sa second source nangyayari
- the added nitrogen gas helps in treating the
less polar molecules
- at high liquid chromatography flow rates
kailangang maraming nitrogen to remove the
solvents kasi mataas ang volume load and
kailangan minimal amount lang ang makarating
sa mass spec para hindi masaturate ang mass
spec
- this occurs at atmospheric pressure
- the 3000 kilovolts is still there but it’s not a part
of the LC capillary and is a separate
mechanism; hindi simultaneous because of the
volume load

Thermospray – the volatilization is done by


heating the capillary

Fast Atom Bombardment – the sample should


be first run with a matrix
- for samples who do not easily volatilize
- one of the classic example of desorption phase
- the sample is mixed with glycerol to hold it
- the source of energy is a fast atom

Vous aimerez peut-être aussi