Document No.:
Chapter 1 Introduction 4
Page 2 of 29
VALIDATION PROTOCOL APPROVAL:
*Note: If the validation is executed by a service provider, then the engineer who does the
validation is the performer, and customer is the first verifier, reviewer and approver.
Page 3 of 29
Chapter 1 Introduction
Scope
This protocol is to validate theBIOBASE BIODUSTRY (SHANDONG)Co., LTd Climate
Incubator at initial installation, modification and relocation.
Model _________________________________
Located _________________________________
Used to _________________________________
Revalidation
System relocation
Responsibilities
Person overseeing the design will perform the qualification and record the information.
The responsible engineer will verify the records and write the report.
Quality Assurance will review and approve the Protocol and Report.
Procedure
Page 4 of 29
The installation and operational qualifications (IQ & OQ) will be conducted by physically
installing and operating the system, comparing it to relevant documentation for
appropriate functioning and documenting results on the attached data sheets for future
change control. Performance qualification (PQ) will be performed by validating resistivity
of the product water from the system. TOC and microbial burden tests are optional.
1.Scope of application
Widely used in microbial cell culture, seed germination, seedling test, plant cultivation, and
insects, small animal husbandry etc.. Accurately simulate different environmental and climatic
conditions.
Test Equipment
To insure test equipment is documented and copies of calibration certificates are present.
Ensure all test equipment is documented in the table below and copies of calibration
certificates are appended to this action.
illuminance meter
Watch
The calibration of all test equipment used shall be traceable to national standards (or an
international equivalent).
The calibration status of the equipment shall be valid (within calibration period, prior to
next calibration).
Acceptance criteria:
Page 5 of 29
All of the above are properly calibrated and accepted.
Accepted:
Comments & corrective actions required: (detail section and test number)
Reviewed:
Page 6 of 29
Chapter 2 Installation Qualification
OBJECTIVE
To ensure that the system installed conforms to the purchase specifications and the
manufacturer’s User Guide, and to document the information that the equipment meets
specifications.
SCOPE
To be performed at time of installation, modification, and relocation.
RESPONSIBILITY
Person overseeing the installation will perform the qualification and record the
information.
The responsible engineer will verify the records and write the report.
Quality Assurance will review and approve the IQ Protocol and Report.
Page 7 of 29
DESCRIPTION OF THE SYSTEM BEING INSTALLED:
ITEM CONTENT
Page 8 of 29
SYSTEM SPECIFICATIONS
Model
BJPX-A250
Parameter
2. Relative humidity:≤80%RH
Page 9 of 29
IQ CHECKLIST
Packaging
System and Accessories
Documentation
Installation Environment
System Installation
Page 10 of 29
INSTALLATION QUALIFICATION
A. Packaging
No Forced Entry/Damage
Model
Biobase
Manufacturer Biodustry(Shandong) Co.,
Ltd.
C. Documents
Page 11 of 29
Description Acceptance Criteria Result Located
(Conform/Deviation) in
User Manual One Set
Page 12 of 29
D. Installation Environment
Result
Description Acceptance Criteria
(Pass/Fail/Deviation)
Power 220VAC 10%,50Hz or 110
VAC, 60 HZ
Ambient temp. ±5~35℃
Relative Humidity ≤80%RH
Conclusion:
E. System Installation
Description Acceptance Criteria Result
(Pass/Fail/Deviation)
The surrounding without strong
1 vibration and corrosive gases
Should be installed in a
4 well-ventilated place
Page 13 of 29
Inspected by: Date:
Page 14 of 29
DEVIATION REPORT
Deviation:
Suggested Remedy:
Justification on Acceptance:
□ Accepted □ Rejected
Page 15 of 29
ISSUES/COMMTS
Page 16 of 29
IQ CONCLUSIONS:
System Installation Qualification is completed and ready for the Operation Qualification.
Signature_____________________________ Date__________________________
Approved by Title__________________________
(Print name)
Approved by Title__________________________
(Print name)
Signature ______________________________Date__________________________
Approved by Title__________________________
(Print name)
Signature ______________________________Date__________________________
Page 17 of 29
Chapter 3 Operation Qualification
OBJECTIVE
To determine that the system/equipment operates according to specifications, and to
record all relevant information and data to demonstrate it functions as expected.
SCOPE
To be performed after installation, modification or relocation, after the Installation
Qualification has been completed.
RESPONSIBILITY
Person responsible for operating the system/equipment will perform the qualification
and record the information.
The supervisor will supervise the study, verify the completion of the records, write the
deviation report and the Operational Qualification Report.
Quality Assurance will review and approve the OQ Protocol and Report.
Page 18 of 29
APPROVAL FOR OPERATION QUALIFICATION (OQ)
OQ Protocol is Completed
Signature Date:
DOCUMENT CHECK
TRAINING RECORDS
Course on SOP# Staff Name Date
Page 19 of 29
OQ Checklist
System Operations
Temperature setting
Lighting & humidity
Defrost
Page 20 of 29
OPERATION QUALIFICATION
A. Handing Precautions
Results
Conclusion
Panel:
Page 21 of 29
Item Confirmation Process
1. :Main interface and query interface, click to switch
query interface and the main interface; setting
interface long press to enter the program;
Setting interface, click the down switch value,
long press to exit the setting interface;
Conclusion
Page 22 of 29
C. Introduction to use
Example:
equipment needed to start work after 120 minutes, for 10 cycles, each cycle set in
paragraph 2, first paragraph of the need to work for 720 minutes, the temperature control at
30 °C, the illumination of 6, and humidity control at 60% RH; firstSec 1200 minutes, the
temperature is controlled at 10 °C, light intensity for four, humidity control at 50% RH.
b) Click on the ">" key to move the cursor to switch to the value of
the total cycle, using the "∧" and "∨" key to modify the set
value of the total cycle of 10; then click on the "SET" button, the
cursor is in the total number of segments the value, modify the
settings of the total number of segments 2; then click on the
"SET" button, the cursor on the value of the appointment, to
modify the appointment time is 120 minutes;
c) then click on the "SET" button, enter the first paragraph set
interface, the cursor on the value of the temperature, the
temperature setting is modified to 30 °C; click on the "SET"
key, the cursor on the value of the humidity, themodify the
humidity setting to 60% RH, and then click on the "SET" button,
the cursor is in the value of the illumination to modify
illumination setting is 6; then click on the "SET" button, the
cursor on the value of the time, to modifyThe time setting is 720
minutes;
d) and then click on the "SET" button to enter the set state in
paragraph 2, as described above, followed by the setting of
temperature, humidity, light intensity, and time was revised to
10 °C, 50% RH, 4; last long press the” SET"button for 4
seconds to exit the setting interface, return to the main interface,
set;
Page 23 of 29
e) long press "R / S" button for 2 seconds, the controller starts
timing of the appointment time running status display:
Appointment timing ... 120 minutes, the program began to run
after the first paragraph;
g) after the end of the first paragraph of the timing to the second
stage, after the end of the second paragraph of the timing, enter
the first paragraph of the second cycle, and so on; after the end of
the second paragraph of the timing of when the 10th cycle, the
end of the program is running, close all out put, buzzer (press any
key to mute).
Conclusion
D. Parameter description:
Temperature parameters: password (the main interface, long press”>” 3 seconds to enter the
password interface)
Page 24 of 29
Parameter Parameter Function Range (factory
name parameters)
When measuring temperature> set
Over-temperature temperature + over-temperature alarm,
alarm instrument warning lights are bright, cut 0.0~50.0 (5.0) ℃
heating protection relay, open refrigeration
buzzer sounds;
Zero adjustment Zero adjustment = mercury thermometer -99.~99.9(0) ℃
readings - Display temperature;
When the deviation of the actual
temperature, adjust the parameter values to
Full-scale tune correct;
Whole -999~999(0)
Full-scale adjustment = 1000 * (mercury
thermometer readings - Displays
temperature) / temperature display
When measuring temperature> = set
Off heating
temperature + off heat, turn off the heating -50.0~50.0(2.0) ℃
output
When the control mode is 0, if the set
temperature <environmental temperature
normally open temperature, the compressor
Normally open
is automatically switched to the normally -50.0~50.0
temperature
open type; when the control mode 1, if the (10.0) ℃
set temperature <normally open
temperature, the compressor is
automatically switchedis often open.
When the control mode is 2, the opening
Open point of the refrigerant for: measuring
refrigeration temperature> = the set temperature + open -20.0~20.0(0.0) ℃
refrigeration.
Page 25 of 29
Conclusion:
E Defrost parameters: password (the main interface at the same time long press
"SET" and "∨” 3 seconds to enter the password interface)
Range
Parameter Parameter Function
(factory
name
parameters)
Interval of 0.0 ~ 9.0 ℃ defrost interval time in hours 0~200(12)
one h
Interval of 9.1 ~ 15.0 ℃ defrost interval time in hours 0~200(24)
two h
Interval of Greater than 15. 0 ℃ when the defrost interval 0~200(72)
three time in hours h
Defrost a Output of 0.0 ~ 9.0 ℃ defrost time, in seconds 0~300(60)
S
Defrost 9.1 ~ 15.0 ℃ defrost output time in seconds 0~300(60)
two S
Defrost Output of greater than 15.0 ℃ defrost time, in 0~300(60)
three seconds S
0: No manual defrost function;
Manual
defrost 1:Main interface, long press "∨" to 8 0~1(0)
seconds, start defrosting, defrost time based on
the parameter: defrost, defrost, defrost three;
Page 26 of 29
DEVIATION REPORT
Deviation:
Suggested Remedy:
Justification on Acceptance:
□ Accepted □ Rejected
Page 27 of 29
ISSUES/COMMENTS
Page 28 of 29
OQ CONCLUSIONS
The Climate Incubator (serial number: ) has passed
Operation Qualification per approved OQ protocol. All items inspected meet user’s and
manufacturer‘s specifications.
Signature_____________________________ Date__________________________
Approved by Title__________________________
(Print name)
Approved by Title__________________________
(Print name)
Signature ______________________________Date__________________________
Approved by Title__________________________
(Print name)
Signature ______________________________Date__________________________