Académique Documents
Professionnel Documents
Culture Documents
Department of Pharmaceutical Chemistry, Smt Kishoritai Bhoyar College of Pharmacy, Kamptee, Nagpur,
Maharashtra, India
Pharmaceutical grade testosterone standard was Method IV (second order derivative UV-
obtained as generous gift sample by Alkem spectrophotometry): The zero order absorption
laboratory Pvt. Ltd. Mumbai, Maharashtra, India. spectrum of Testosterone was transformed to
HPLC grade Acetonitrile purchase from MERCK second order derivative and the amplitudes were
Life Science Pvt. Ltd. and distilled water was recorded at 258.6 nm and 239.6 nm. The
used. The pharmaceutical dosage form used in selections of wavelength for the proposed
this study is 1%W/W testosterone gel formulation methods are shown in Figure 2 (a-c).
manufactured by sun pharmaceutical Ind. Ltd.
Abs
To determine the content of Testosterone from
0.4 marketed gel formulation; average content per
sachet was calculated. An amount of gel
0.2 equivalent to 100.0 mg of testosterone was
weighed and transferred to a 10.0 mL volumetric
0 flask. To it sufficient amount of acetonitrile was
200 250 300 350 400
Wavelength [nm] added, sonicated for 30 min and the solution was
diluted up to mark with the same solvent. The
Figure 2 (a): Zero order spectrum of content of the flask was filtered through
testosterone showing AUC between selected Whatmann filter paper (No. 41). From the filtrate
wavelength. solution pipette out 1.0 ml and add 10.0 ml
volumetric flask and make up the volume up to
0.07
the mark and get the final concentration of
0.05
10μg/ml. The absorbance was measured at
selected wavelengths as described above and
concentrations in the sample were determined
using regression equation (method 1); Comparing
Abs
0
the AUC of standard and sample at selected
wavelength (method 2) and comparing the
derivative absorbance of standard and sample at
selected wavelength (method 3 and 4)
respectively.
-0.05
200 250 300 350
Wavelength [nm]
Validation of methods
Figure 2 (b): First order derivative spectrum. The proposed method was validated as per ICH
0.002
guidelines.
0.001
Accuracy
0
The accuracy of proposed method was
Abs
ascertained on the basis of recovery studies
-0.001
performed by standard addition method. The pre-
analyzed gel equivalent to 100 mg was weighed;
-0.002 a known amount of standard drug was added at
200 250 300 350
Wavelength [nm] different levels covering the range 50-150% of
target. The resultant solutions were then analyzed
Figure 2 (c): Second order spectrums. by the proposed methods. At each level, triplicate
samples were analyzed to check repeatability and
Preparation of calibration curve from the data, the amount of standard drug
Appropriate dilutions of standard stock solution recovered was estimated.
were made using acetonitrile to get final Precision
concentration in the range of 5-25 μg/ml.
Absorbance and area under curve were measured The precision of the analytical method expresses Commented [SM1]:
of each prepared solution at above selected the closeness of agreement between a series of
wavelengths. The calibration curve was plotted measurements obtained from multiple sampling
between concentrations vs. absorbance/AUC, of the same homogeneous sample under the
prescribed conditions. The precision of the
methods can be studied as; intra-day variation, The % amount estimated from gel formulation
inter-day variation studies. Intra-day study was indicates that there was no interference from
carried out by analyzing the 10 μg/ml of sample excipients present in it (Table 1).
for five times within the day while in inter-day
study same solution was analyzed on five Method validation
different days.
Developed methods were validated for linearity,
Ruggedness accuracy, precision, ruggedness and sensitivity as
per the ICH guidelines.
Ruggedness of proposed methods was performed
to examine effect of non-procedure related Accuracy
factors such as instruments, wavelength, and
analysts. For this study Testosterone (10μg/mL) Results of recovery studies are shown in Table 2.
was analyzed by proposed methods using two The % RSD value was found to be less than 2
different analyst, two different wavelength, and indicating that the methods are accurate.
two different UV-spectrophotometers (Jasco V-
630 and Shimadzu-1700) restraining similar Linearity
operational and environmental conditions.
From the linear regression data, it was found that
Linearity the linearity curve showed good linear
relationship over the concentration range of 80-
The linearity of an analytical procedure is the 120% of label claim. The linearity curves are
ability to obtain test results that are directly shown in Figure 3 (a-f).
proportional to the concentration (amount) of an
analyte in the sample within a given range. Five Ruggedness
solutions of sample of different percent of label
claim (80-120%) were prepared, analyzed by The results of ruggedness study were found in the
proposed methods and the obtained data were acceptable range with% RSD values less than 2
utilized to plot linearity curve as Percent of label by all proposed methods as shown in Table 3a and
claim vs absorbance. 3b. The results showed no statistical differences
between different operators and instruments
Results and Discussion suggesting that the developed methods were
rugged.
Testosterone was found to be highly soluble in
acetonitrile and stable in acetonitrile, hence Precision
working standard solutions were prepared of
desired concentrations were prepared throughout The precision of the method was expressed in
experimentation. All developed methods were terms of% RSD. The obtained results showed
found to obey Beer’s-lambert’s law in the reproducibility of the assay. The% RSD values
concentration range of 5-25 μg/mL with were found within limit indicates that the
correlation coefficient value less than 1. Further methods were found to be precise. Results are
the proposed methods were applied to the phar- shown in Table 4(a) and 4(b).
maceutical formulation for assay of testosterone
in gel formulation. The recovery study was Statistical analysis of results
carried out at three different levels 50-150%. All
developed methods were validated as per ICH Paired t-test
guidelines.
Paired t-test is the method of validating a new
Analysis of marketed formulation procedure is to compare the results using sample
of varying compositions with the values obtained
The percentage amount of testosterone estimated by an accepted method are shown in Table 5
from gel formulation using method I to IV was (a).11,12
found to be M-I= 101.8, M-II=99.6, M-III=101.2
& 99.0 and M-IV=100.2 & 100.8 respectively.
% label claim
Sr. Different Wt. of gel MI M II M III M IV
No. Analyst (mg) 237.4 AUC 250.4 226.4 258.5 239.6
1. Analyst-I 101.74 100.02 101.74 99.33 100.50 100.77
2. Analyst-II ~100.0 102.02 100.67 102.02 101.58 99.96 102.00
Mean 101.5 100.34 101.5 100 99.5 101.38
±SD 0.707 0.45 0.707 1.414 0.707 0.869
% RSD 0.697 0.458 0.697 1.414 0.711 0.857
1.2 1
1 y = 0.0074x + 0.1098 0.8 y = 0.0095x - 0.2355
Absorbance
Absorbance
R² = 0.993 R² = 0.9971
0.8
0.6
0.6
0.4
0.4
0.2 0.2
0 0
0 50 100 150 0 50 100 150
A B
Concentration (μg/ml) Concentration (μg/ml)
0
0.05 -0.005 0 50 100 150
0.04
Absorbance
0.001 0
0 50 100 150
0.0008 -0.0002
-0.0004 y = -6E-06x - 0.0002
Absorbance
Absorbance
0.0006 R² = 0.9914
-0.0006
0.0004 y = 7E-06x - 1E-05
R² = 0.999 -0.0008
0.0002
-0.001
0
0 50 100 150 -0.0012
E Concentration (μg/ml) F Concentration (μg/ml)
Figure 3 (A-F): Plot of linearity and range for method I-IV. (A) M-I: Plot of linearity and range,
(B) M-II plot of AUC, (C) M-III first order derivative at 250.4 nm, (D) M-III first order
derivative at 226.2 nm, (E) M-IV (258.6), (F) M-IV (239.6).
the comprise of each methods. The significance 3. Luetjens CM, Weinbauer GF. Chapter 2:
is finding on to the two methods is (M1 vs. M3b Testosterone: Biosynthesis, transport,
& M3a vs. M3b) but the remaining methods there metabolism and (non-genomic) actions. In:
is not outcome to the significance value (Figure Nieschlag E, Behre HM, Nieschlag S, eds.
4). The results are tabulated in the Table 5 (b). Testosterone: Action, Deficiency,
Substitution (4th Ed.). Cambridge:
150
Cambridge University Press; 2012: 15–32.
4. Dabbs M, Dabbs JM. Heroes, rogues, and
lovers: testosterone and behavior. New York:
100 * McGraw-Hill, 2000.
% Claim
11. Mendham, RC. Denney, JD, Barnes, M. 12. Verma RM. Analytical chemistry, Theory
Thomas, B. Sivasankar Vogel textbook of and practice third edition. 30-31.
qualitative chemical analysis 6th edition.
2013: 115-119.