Académique Documents
Professionnel Documents
Culture Documents
CAMPINAS
2016
ROBERTO ADRIAN MARKARIAN
CAMPINAS
2016
Folha de Aprovação
___________________________________________________________________
Profª. Drª. Fabiana Mantovani Gomes França – Orientadora
___________________________________________________________________
Profª. Drª. Flávia Lucisano Botelho do Amaral
___________________________________________________________________
Prof. Dr. Flávio Henrique Baggio Aguiar
___________________________________________________________________
Prof. Dr. Mario Alexandre Coelho Sinhoreti
___________________________________________________________________
Profª. Drª. Roberta Tarkany Basting
Tu vedi un blocco, pensa all'immagine:
l'immagine è dentro basta soltanto spogliarla
- Michelangelo Buonarotti-
Agradeço ao Prof. Dr. Marcelo Sperandio pelo auxílio com a versão em língua
inglesa deste trabalho.
Por fim, agradeço à minha esposa Deborah, que tanto me apoiou desde a
ideia de realizar este curso. Aceitou minhas ausências, me ajudou a superar as
dificuldades na elaboração da metodologia e quando precisei, seu pensamento
rápido e ponderado me ajudou a encontrar soluções para minhas dúvidas. Nestes
últimos anos além do papel de companheira, acabou se tornando também Profª.
MSc Deborah Galles, coautora deste trabalho, contribuindo para sua qualidade e
finalização no prazo adequado.
O objetivo deste trabalho foi mensurar a fenda na interface entre implantes dentários
e pilares customizados (interface pilar-implante IPI) fabricados com diferentes
métodos CAD/CAM, antes e após ciclagem mecânica, utilizando Microscopia
Eletrônica de Varredura (MEV), além de realizar uma avaliação morfológica dos
pilares. Um software CAD (Dental System, 3shape, Dinamarca) foi utilizado para
projetar um pilar personalizado para uma coroa aparafusada unitária, compatível
com um implante dentário com conexão protética tipo hexagono externo diâmetro de
plataforma 4.1mm. Um mesmo arquivo de estereolitografia resultante foi enviado
para fabricação usando quatro métodos CAD/CAM (n=10): fresagem e sinterização
de dióxido de zircônio (ZI), Co-Cr sinterizado a laser (SL), liga de Co-Cr denso
fresado (MF); fresagem e sinterização de liga de Co-Cr em pó aglutinado (MA).
Pilares de titânio pré-fabricados foram utilizados como controle (TI, AI 4151-Q, SIN -
São Paulo, Brazil). Cada pilar foi instalado sobre um implante dentário de 4.1x11 mm
(SA411, SIN, Brasil) inserido em bloco de alumínio. As medições foram realizadas
utilizando-se MEV (4000X) em quatro regiões da interface implante-pilar (IPI) com
distância relativa de 90° entre si. A média entre essas quatro medidas correspondeu
ao valor da amostra. Os espécimes foram envelhecidos mecanicamente (1.000.000
ciclos, 2 Hz, 100 N, 37° C) e a largura da IPI foi novamente medida utilizando a
mesma metodologia. Os dados foram analisados por Analise de variância com dois
fatores, seguido do teste de Tukey, com nível de significância de 5% (p<0,05). Antes
da ciclagem mecânica, os menores valores de fenda na IPI foram obtidos do grupo
TI (1,10 ± 0,31 µm), MA (1,1 ± 0,72 µm). Houve resultado significativamente maior
para o grupo MF (2,61 ± 1,13 µm). Medidas de fenda significativamente maiores
foram encontrados nos grupos SL (24,71 ± 15,29 µm) e ZI (11,75 ± 0,61 µm).
Após ciclagem mecânica, os menores valores de fenda foram obtidos do grupo TI
(2,29 ± 1,13 µm), MA (3,58 ± 1,80 µm) e MF (1,89 ± 0,98 µm). Observou-se um
resultado de adaptação significativamente maior para os grupos SL (18,40 ± 20,78
µm) e ZI (10,42 ± 0,80 µm). A ciclagem mecânica teve efeito estatisticamente
significante apenas nos espécimes MA, com o aumento da fenda na IPI. Os pilares
personalizados fabricados com MF e MA demonstraram menor fenda na IPI,
semelhantes aos obtidos com os pilares comerciais de titânio pré-fabricados após
ciclagem mecânica. A fenda de pilares personalizados elaborados pelos métodos SL
ou ZI, apresentaram resultados estatisticamente semelhantes entre si, e
permaneceram em níveis mais elevados, antes e depois da ciclagem mecânica. O
grupo SL apresentou deformações morfológicas e rugosidades que poderiam
interferir em seu desempenho clínico. O grupo MF apresentou resultados e
morfologia semelhantes ao resultados do grupo controle (TI). Conclui-se que o
método CAD/CAM, bem como os materiais utilizados podem influenciar na
morfologia dos pilares resultantes, bem como na precisão da adaptação na interface
implante-pilar.
The purpose of this work was to measure the microgap between dental implants and
custom abutments fabricated using different CAD/CAM methods, before and after
mechanical cycling, using Scanning Electron Microscopy (SEM). The CAD software
(Dental System, 3shape, Denmark) was used to design a custom abutment for a
single-unit screw-retained crown, compatible with a 4.1 external hexagon dental
implant. The same resulting stereolithography file was sent for manufacturing using
four CAD/CAM methods (N=40): milling and sintering of zirconium dioxide (ZO), Co-
Cr sintered by selective laser melting (SLM), fully sintered machined Co-Cr alloy
(MM); machined and sintered agglutinated Co-Cr alloy powder (AM). Prefabricated
titanium abutments were used as control (TI). Each abutment was installed onto a
dental implant measuring 4.1 x 11 mm (SA411, SIN, Brazil) inserted into an
aluminum block. Measurements were taken using SEM (4,000X) on four regions of
the implant-abutment interface (IAI), and relative distance of 90° from each other.
The specimens were mechanically aged (1,000,000 cycles, 2 Hz, 100N, 37°C) and
the IAI width was measured again using the same approach. Data were analyzed
using two-way ANOVA, followed by the Tukey test. After mechanical cycling, the best
adaptation results were obtained from group TI (2.29 ± 1.13 µm), AM (3.58 ± 1.80 µm)
and MM (1.89 ± 0.98 µm). A significantly worse adaptation outcome was observed
for the groups SLM (18.40 ± 20.78 µm) and ZO (10.42 ± 0.80 µm). Mechanical
cycling had a marked effect only on the AM specimens, which significantly increased
the microgap at the IAI. Custom abutments fabricated using MM and AM
demonstrated lower adaptation levels at the IAI, similar to those obtained with
commercial prefabricated titanium abutments after mechanical cycling. The
adaptation of custom abutments made by SLM or ZO methods remained in higher
levels, both before and after mechanical cycling.
aparência microscópica. .
SUMÁRIO
1. INTRODUÇÃO ................................................................................................ 09
2. OBJETIVOS .................................................................................................... 14
3. ARTIGO .......................................................................................................... 15
INTRODUÇÃO
até atingirem a forma planejada por computador. Por outro lado, as estratégias
opção pelo material a ser utilizado em cada caso depende do tipo de trabalho a
cirurgião dentista.
(Volpato et al., 2011); porém em alguns casos com maior exigência mecânica, há
10
titânio, ligas metálicas básicas ou nobres (Linkevicius & Vaitelis 2015; Svanborg
para prótese, por suas propriedades mecânicas, como a sua elevada resistência
benefício quando comparadas a ligas de metais nobres (Al Jabbari, 2014). Ligas
de Co-Cr podem ser utilizadas em estruturas para próteses sobre implante com
casos nos quais o desafio biomecânico seja mais acentuado, como em próteses
França et al. 2015), ou por estratégia aditiva (Koutsoukis et al., 2015; Tuna et al.,
2015).
ligas fundidas de Co-Cr (Becker et al., 1995; Dourandish et al., 2008; Stawarczyk
como sinterização a laser, utiliza como matéria prima metais como o Titânio e
Takaichi et al. 2015). A técnica SLM é capaz de produzir corpos mais complexos
(Oyagüe et al., 2012; Tuna et al., 2015), e pilares para implantes unitários
podem ter interface pilar-implante (IPI) por duas estratégias: i- Cimentando o pilar
o ajuste IAI seja ótimo, já que as peças são produzidas sob as especificações
precisão alcançada no nível da IPI e a resistência dos pilares obtidos por esses
Fernández et al., 2014; Joda et al., 2015). Enquanto que o primeiro método (i) é
suporte, que podem ser potencializadas caso exista uma desadaptação na interface
interface implante-pilar tem sido realizadas com sucesso por meio do microscópio
eletrônico de varredura (MEV) (Dias et al., 2012; Coelho et al., 2007; Solá-Ruíz et
et al. 2006; Cibirka et al. 2001; Assunção et al. 2009; Dittmer et al. 2012). Dado
forma segura.
14
2 OBJETIVOS
O objetivo deste trabalho foi avaliar por MEV a fenda entre implantes com
laser liga de Co-Cr em pó, fresagem de liga de Co-Cr denso, fresagem de liga de
3 ARTIGO
Roberto Adrian Markarian DDS, MS, PhD student, Deborah Pedroso Galles
ABSTRACT
Purpose: To measure the microgap between dental implants and custom abutments
fabricated using different CAD/CAM methods, before and after mechanical cycling,
Materials and Methods: The CAD software (Dental System, 3shape, Denmark) was
with a 4.1 external hexagon dental implant. The same resulting stereolithography file
was sent for manufacturing using four CAD/CAM methods (N=40): milling and
sintering of zirconium dioxide (ZO), Co-Cr sintered by selective laser melting (SLM),
fully sintered machined Co-Cr alloy (MM); machined and sintered agglutinated Co-Cr
alloy powder (AM). Prefabricated titanium abutments were used as control (TI). Each
abutment was installed onto a dental implant measuring 4.1 x 11 mm (SA411, SIN,
Brazil) inserted into an aluminum block. Measurements were taken using SEM
(4,000X) on four regions of the implant-abutment interface (IAI), and relative distance
of 90° from each other. The specimens were mechanically aged (1,000,000 cycles, 2
16
Hz, 100N, 37°C) and the IAI width was measured again using the same approach.
Data were analyzed using two-way ANOVA, followed by the Tukey test.
Results: After mechanical cycling, the best adaptation results were obtained from
group TI (2.29 ± 1.13 µm), AM (3.58 ± 1.80 µm) and MM (1.89 ± 0.98 µm). A
significantly worse adaptation outcome was observed for the groups SLM (18.40 ±
20.78 µm) and ZO (10.42 ± 0.80 µm). Mechanical cycling had a marked effect only
adaptation levels at the IAI, similar to those obtained with commercial prefabricated
made by SLM or ZO methods remained in higher levels, both before and after
mechanical cycling.
INTRODUCTION
blocks by drills or diamond burs until they have reached the desired shape.
enabled the use of such alloys with CAD/CAM technologies via subtractive
specially large CAD/CAM milling machines for dental laboratories are able to
produce structures from fully sintered blocks of Co-Cr on a large scale, due to the
Austria) and presented in the form of metal blocks composed of pressed powder
using less robust equipment.19 The processing steps involved in Co-Cr alloy
The additive strategy of selective laser melting (SLM), also known as laser
sintering, uses fine granules of metals such as titanium and Co-Cr as raw
successive vertical layers of material, the structure will reach its final volumetric
The SLM and sintered metal milling technologies have been studied for the
implants in the anterior region have shown good performance, which has been
external forces and estimate their clinical behavior, which can be achieved by
titanium based (Ti-based) abutment ; ii- directly designing the customized abutment
including the IAI. Meanwhile the former is more often indicated by implant
19
manufacturers, the latter could be used for some clinical solutions. 5,7
the IAI fit would be optimal, as parts are produced under original manufacturer’s
customized CAD/CAM abutment including the IAI could be manufactured under any
material and strategy used in desktop laboratory machines. Within these approaches,
concern may arise regarding the precision achieved at the IAI level and the
It is known that during function, masticatory forces generate stress not only
in the abutment, but also in the implant and supporting tissues, which can be
potentiated should the IAI be ill-fitting,35 thus resulting in bacterial leakage and
proven successful not only at measuring the IAI adaptation,37-39 but also at
The objective of this study was to evaluate by SEM the IAI adaptation of
effects of mechanical cycling on the abutments. The null hypothesis was that
Abutments
and anti-rotation platform. A 4.1 external hex implant analog (AN 4100, SIN - São
Paulo, Brazil) was scanned (D700, 3Shape A/S, Copenhagen, Denmark) and used
as the prototype connection. Using drawing software (Dental System, 3Shape A/S,
screwed-on restoration in the premolar region measuring 4.1 x 9.6 mm. The
abutment was meant to engage the implant hexagon and would require further
porcelain application (Fig 2). The same resulting stereolithography file format (STL)
was used to manufacture abutments using four different methods and, for each group,
ten identical abutments were made (Table 1). Titanium cylinders with anti-rotation
hexagon (AI 4151-Q, SIN - São Paulo, Brazil) were used as control abutments. Once
the abutments were obtained from the manufacturer, they were cleaned in a
milling machine (Roders, RXD5, Germany) was used. Zirconia milling was performed
in a pre-sintered stage with dimensions increased by 25%. This step was followed by
sintering of the structures in a specific furnace (In Fire, Sirona, Bensheim, Germany),
For the abutments in the SLM Group, the project was sent to a machining
center (CUBO, São Paulo, Brazil), which used an SLM machine (EOSINT M270 EOS
GmbH, Germany) and a Co-Cr alloy powder (EOS Cobalt Chrome SP2, EOS GmbH)
21
as raw material. The samples were then separated from their base using a diamond
disc.
For the preparation of the Co-Cr specimens (MM group), the STL file was sent
carved directly on a block of dense and completely sintered Co-Cr alloy, which
equipment (Ceramill Motion II, AmannGirrbach, Austria). The alloy was machined into
implants (n = 10) were used, measuring 4mm in diameter, 11mm in length and a
4.1-mm platform (SA411, SIN - Brazil). The implants were inserted individually
and 8-mm deep central orifice, using an implant insertion key (CCIT 20, SIN -
Then each abutment was fixed onto the implant using a square-headed
retaining screw (PTQ 2008 - SIN - Brazil) and manual torque of 32Ncm, using a
dedicated key (CQTM20, SIN - Brazil), manual a ratchet (TMEC, SIN, Brazil) and
Mechanical Cycling
a million mechanical fatigue cycles at a frequency of 2Hz under 100 N and 30-
Carlos, Brazil). The specimens were kept in saline solution at 37°C throughout
the experiment.
Adaptation measurements were taken at the IAI before and after mechanical
cycling at 4000x magnification using SEM. Each specimen received four markings
taken, corresponding to the distance between the bottom edge of each abutment and
the top edge of the implant. These measurements were always taken by the same
calibrated observer. The average of the four measurements (µm) for each specimen
observed qualitatively at the fitting face of the hexagon, before and after
Statistical analysis
between implants and abutments were evaluated by Shapiro-Wilk and Levene tests,
determine whether the mismatch between implants and abutments was influenced by
abutment type and mechanical cycling. For the breakdown of interactions, the Tukey
test was applied. Statistical calculations were performed on SPSS 20 (SPSS Inc.,
RESULTS
The Shapiro-Wilk and Levene tests indicated that the data did not conform to a
Breaking down the interactions using the Tukey test, it was observed that prior
to mechanical cycling lower adaptation values were found for control group (TI) and
AM, which were not significantly different between each other (Table 2). The highest
adaptation values were found in the zirconium oxide (ZO) and selective laser melting
(SLM) groups, which were not found to be significantly different from each other. The
adaptation values obtained from the machined fully sintered Co-Cr abutments (MM
group) showed intermediate values, which were significantly different from those
Upon completion of mechanical cycling, the Tukey test indicated that lower
adaptation values were obtained for the abutments from the MM, TI and AM groups,
with no significant differences between them. The zirconium oxide and laser sintered
Co-Cr abutments showed the highest misfit values and were not significantly different
between each other. Mechanical cycling had a significant effect on adaptation values
only for the agglutinated Co-Cr abutments, leading to a significant increase in misfit.
When inspecting adaptation at the IAI and hexagons by SEM, the lateral wall of
the titanium abutments (TI) appeared flattened and containing some grooves (Fig 5A).
These specimens had regular and well-defined edges that were not affected by
observed on the lateral wall of the abutments from the ZO group, with loss of
definition of the abutment edges (Fig 5C). After mechanical cycling, the abutment
edges did not change, however a smear layer could be seen on the hexagonal
The specimens from the SLM group showed moderate surface roughness as
well as irregularities on their walls and edges (Figs 5E and 5F). Inspection of the
interface revealed deformation of the abutment edges and the presence of lumps and
beads of material. There were not perceptive qualitative changes after mechanical
The lateral surface of the abutments from the MM group was flattened and
contained slight grooves and dimples inherent to the machining process, which was
very similar to the TI group (Fig 5G). The specimens presented regular and well-
defined edges, and were not apparently affected by mechanical cycling (Figs 5H and
9D). At the hex connection interface, some artifacts were observed such as chippings
and beads of material (Figs 9A and 9C). After mechanical cycling, these beads
disappeared leaving a smear layer deposited onto the interface surface (Fig 9D).
The abutments from the AM group showed a markedly rough lateral surface
with rounded edges (Fig 5I). Inspection of the interface revealed microtexture
patterns and loss of definition of the dihedral angles, despite the abutment appearing
the interface did not change after mechanical cycling (Figs 5J and 10D).
26
DISCUSSION
such high standard adaptation, including the use of additive or subtractive strategies,
the quality of the milling machine and the intrinsic properties of the preparation
material.3
CAD/CAM abutments in this study and the null hypothesis was therefore rejected.
The abutments from the TI group (control group) were machined by the same
manufacturer as the implants. This process followed strict specifications for maximum
precision of the implant-abutment set, obtaining average values in the order of 1 µm.
Similar mean values were obtained with the abutments from the agglutinated Co-Cr
powder (AM) group, which under inspection by SEM, were free from internal
irregularities and debris (Fig 10C). Intermediate adaptation values around 2.5 µm
were obtained for the fully sintered machined metal (MM) group, whose abutments
were prepared on a CNC machine and showed artifacts such as beads and debris at
the interface, which may have interfered with the quality of the initial adaptation (Fig
9C). The abutments from the groups made with selective laser melting (SLM) and
zirconium oxide (ZO) had the highest misfit values, greater than 10 µm. In the case of
(Fig 7C), but for the SLM abutments, a marked morphological change was observed,
such as roughness throughout the abutment surface at the interface region (Fig 8C)
commercial systems and classified the adaptation level as acceptable, good and
excellent, with values typically below 10 µm and average values below 4 µm. Results
below 10 µm were also found in the literature for fixed prostheses prepared by
mechanical loading, the abutments from the groups TI, MM, and AM showed similar
adaptation values to each other and all within the same levels as those reported in
the literature for external hex implant abutments.37,39 Apparently, metal abutments
accurate results. Moreover, although the abutments from groups ZO and SLM have
shown a decrease in average misfit, this was not significant and remained at high
levels, possibly because the causal factors of the initial misfit were maintained.
for external hexagon implants made in Co-Cr by CAD/CAM using SLM and milling
techniques, comparing them with cast abutments. They reported average adaptation
values in the order of 11.3 µm, 0.73 µm and 9.09 µm, respectively. They found that
the abutments prepared by SLM showed high superficial roughness and loss of
definition of the edges, with a positive correlation between internal roughness and
abutment adaptation.26
Such results corroborate those found in this study using SEM, which
demonstrated considerable roughness on the abutments from the SLM group, which
did not change after mechanical cycling, similarly to the agglutinated metal powder
abutments (AM). Clinically, this roughness could translate into greater bacterial
colonization on the unpolished inner surfaces of the abutment and enhance bacterial
28
count at the IAI (Figs 8C and 10C). The relevance of such bacterial accumulation
some extent.37 Jansen et al. studied microbial infiltration on thirteen different implant-
components can reduce bacterial leakage, though no system was able to prevent it
altogether.36
The lateral surface of the abutments from the MM group contained some
grooves and undulations, similarly to group TI (Fig 5G). The interface showed well-
defined geometry and angles (Fig 9A). The artifacts initially found at the IAI of the
the high wear resistance of the material.3,14 The friction between the components
during mechanical cycling was probably the cause of the smear layer and the small
chippings seen on the surface of the components. The same friction could have
some grooves on their walls from the CNC tools (Fig 5A), which apparently suffered
no qualitative changes after mechanical cycling (Fig 5B). The good results for the
titanium abutments observed in this study may encourage the use of Ti-based
which Ti-based abutments are used and onto which a monolithic crown7 or a
mesostructure4,5 can be designed. Therefore, at least in theory, the fit between the
implant and the prosthetic system would always be the best possible and less
brings out further questions in terms of their mechanical stability, the strength of the
cement layer between the interface and the mesostructure and bacterial colonization
29
of the cementation zone.5 According to the results from this study, theoretically, these
outcomes using techniques and materials from the groups MM and AM.
screws.
In this study, the simulation of 15 months of clinical use was performed on the
abutments,34 which did not affect the level of marginal fit in any of the groups, except
for group AM, thus causing the null hypothesis to be rejected. It is a novel material,
into which only few in vitro15,19,20 and no clinical studies have been performed. No
one knows for sure if the presence of organic agglutinating substances influence on
their ultimate strength, or the degree of porosity resulting from sintering. One of the
possible reasons for such findings could be the sintering process, which can cause
dimensional changes.21 In addition, the images from the interfaces suggest a certain
degree of horizontal misalignment of the components from this group (Fig 5J), which
could be detected at 4000x (Fig 5C). After mechanical cycling, no further damage to
the abutments was detected (Figs 7B and 5D), though smear deposits were
30
friction between components (Fig 7D), whose contents could not be determined. This
finding corroborate those by a study that reported debris, irregularities and increased
results due to their structural stability.9 Linkevicius & Vaitelis10 found that single-unit
titanium and zirconia abutments behave well clinically with esthetic advantages to
zirconia. The authors state that titanium should be the material of choice for stability
over time, but other materials may be used for restorations with direct application of
porcelain to zirconia and Co-Cr. The results from the present study corroborate these
previous findings since the best qualitative and quantitative results for adaptation
were obtained from titanium abutments. Conversely, reports show that single-unit
abutments made of zirconia for CAD/CAM systems performed just as well as their
In the present study, zirconia abutments had significantly poorer adaptation values
than the control group. Possible reasons for such difference may be related to certain
stages of the manufacturing process and intrinsic properties of the material.3 The
abutments from the titanium control group were processed on a large CNC milling
machine and no comparisons were made with other machines on the market.
heating furnace at high temperature to achieve the final sintering. Therefore, the
Milling and direct metal laser sintering using SLM strategies would be preferable
would not allow for finely detailed structures and, therefore, milling would be more
suitable to reproduce details.15 On the other hand, criticism to milling techniques may
include equipment vibrations and the use of drills that may become worn and unable
equal or slightly better than metal casting and certainly better compared to milling.25
However, in this study the laser sintered material showed high internal distortion and
sharp marginal surface roughness as well as irregularities of the walls and edges that
generated the most significant misfit seen in this study. Also the specimens in SLM
group presented not only higher fit measurements, but also a low consistency of the
would hinder its clinical application for prosthetic structures over implants.
cycling. One possible explanation for the distortions observed in the SLM group could
be the intensity of the laser. A study found that different laser intensities can generate
variable that may have influenced this group was the fact that the SLM process
generates residual stresses within the material due to rapid heating and cooling of
the alloy, which may even affect the accuracy of the final structure.23 For this reason,
an additional heating step should be necessary so that this residual stress could be
The vertical misfit of the components found in this study may not be the only
of the components and supporting tissues.35 The quality of the internal adaptation to
the abutment-implant system and the horizontal mismatch should also be further
the specimens.37,39
One may infer, therefore, that titanium remains the gold standard for single-unit
in select cases, especially when ceramics should be applied directly onto the coping
surface. The selective laser melting technique should be further investigated prior to
CONCLUSION
This work aimed to measure the microgap between external connection dental
implants and CAD/CAM abutments made in different materials: milled zirconia, Co-Cr
alloy sintered by selective laser melting, machined Co-Cr alloy fully sintered,
Based on the findings as well as the limitations of this study, the following
melting and zirconia remained in highest levels among the tested groups before and
abutments, which remained within the lowest levels among the tested groups, having
ACKOWLEDGEMENTS
The authors acknowledge the dental laboratory EsteticArt, São Paulo, Brazil, for
digital abutment design. Furthermore, they thank SIN, São Paulo, Brazil, for their
any entity that is commercially related to the products mentioned in this article.
REFERENCES
2012;28(1):3-12.
34
2014;2014.
Implants. 2015;30(6).
prosthodontics. 2014;6(2):138-45.
15. Li KC, Prior DJ, Waddell JN, Swain MV. Comparison of the microstructure and
16. de França DGB, Morais MHS, das Neves FD, Barbosa GA. Influence of
Selective Laser Melting Technique of Co-Cr Dental Alloys: A Review of Structure and
Prosthodontics. 2015;24(4):303-12.
18. Tuna SH, Pekmez NÖ, Kürkçüoğlu I. Corrosion resistance assessment of Co-
et al. A novel CAD/CAM base metal compared to conventional CoCrMo alloys: an in-
vitro study of the long-term metal-ceramic bond strength. Oral Health Dent Manag.
2014;13(2):446-52.
20. Lee D-H, Lee B-J, Kim S-H, Lee K-B. Shear bond strength of porcelain to a
new millable alloy and a conventional castable alloy. The Journal of prosthetic
dentistry. 2015;113(4):329-35.
melted CoCrW alloy for dental application. Materials Science and Engineering: C.
2015;49:517-25.
selective laser melting process for dental applications. Journal of the mechanical
2012;40(2):123-30.
37
85.
2015;26(12):1436-42.
31. Hecker DM, Eckert SE, Choi Y-G. Cyclic loading of implant-supported
dentistry. 2006;95(1):26-32.
32. Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the
2001;85(3):268-75.
38
33. Assunção WG, Dos Santos PH, Delben JA, Gomes ÉA, Barao VAR, Tabata
interface design on the yield force of the implant–abutment complex before and after
35. Markarian RA, Ueda C, Sendyk CL, Laganá DC, Souza RM. Stress
distribution after installation of fixed frameworks with marginal gaps over angled and
22.
36. Jansen VK, Conrads G, Richter E-J. Microbial leakage and marginal fit of the
1997;12(4).
37. Dias ECLdC, Bisognin EDC, Harari ND, Machado SJ, da Silva CP, de Almeida
five external-hex implant systems: an in vitro study. International Journal of Oral &
Fig 4 Test specimen following implant placement into aluminum blocks and abutment
adaptation.
44
Table 2 Mean and standard deviation of the adaptation values (µm) between implant
and the different abutments tested before and after mechanical cycling.
Uppercase and lowercase letters by the means represent the multicomparison of the
data by the Tukey test. Means followed by different uppercase letters indicate a
significant difference between types of abutments, considering the conditions before
and after mechanical cycling (comparisons within columns). Means followed by
different lowercase letters indicate a significant effect of mechanical cycling,
considering each type of abutment separately (comparison within rows).
Fig 5 Example of interfaces prior and after mechanical cycling for the different materials tested. (A) TI group before cycling. (B)
TI group after cycling. (C) ZO group before cycling. (D) ZO group after cycling. (E) SLM group before cycling. (F) SLM group
after cycling. (G) MM group before cycling. (H) MM group after cycling. (I) AM group before cycling. (J) AM group after cycling.
45
46
Fig 6 Qualitative analysis of an abutment from group TI before and after mechanical
cycling. (A) Hexagon before (50X). (B) Hexagon after mechanical cycling (50X). (C)
Hexagon before (250X). (D) Hexagon after mechanical cycling (250X).
47
Fig 7 Qualitative analysis of an abutment from group ZO before and after mechanical
cycling. (A) Hexagon before (50X). (B) Hexagon after mechanical cycling (50X). (C)
Hexagon before (250X). (D) Hexagon after mechanical cycling (250X).
48
Fig 8 Qualitative analysis of an abutment from group SLM before and after
mechanical cycling. (A) Hexagon before (50X). (B) Hexagon after mechanical cycling
(50X). (C) Hexagon before (250X). (D) Hexagon after mechanical cycling (250X).
49
4 CONSIDERAÇÕES FINAIS
testados, e similares aos obtidos com pilares de titânio pré-fabricados após ciclagem
mecânica.
REFERÊNCIAS
Assunção WG, Dos Santos PH, Delben JA, Gomes ÉA, Barao VAR, Tabata LF.
Effect of misfit on preload maintenance of retention screws of implant-supported
prostheses. Journal of materials engineering and performance. 2009;18(7):935-8.
Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the implant—
abutment interface after fatigue testing. The Journal of prosthetic dentistry.
2001;85(3):268-75.
de França DGB, Morais MHS, das Neves FD, Barbosa GA. Influence of CAD/CAM
on the fit accuracy of implant-supported zirconia and cobalt-chromium fixed dental
prostheses. The Journal of prosthetic dentistry. 2015;113(1):22-8.
Dias ECLdC, Bisognin EDC, Harari ND, Machado SJ, da Silva CP, de Almeida
Soares GD, et al. Evaluation of implant-abutment microgap and bacterial leakage in
five external-hex implant systems: an in vitro study. International Journal of Oral &
Maxillofacial Implants. 2012;27(2).
53
Hecker DM, Eckert SE, Choi Y-G. Cyclic loading of implant-supported prostheses:
comparison of gaps at the prosthetic-abutment interface when cycled abutments are
replaced with as-manufactured abutments. The Journal of prosthetic dentistry.
2006;95(1):26-32.
Jansen VK, Conrads G, Richter E-J. Microbial leakage and marginal fit of the implant-
abutment interface. International Journal of Oral & Maxillofacial Implants. 1997;12(4).
Lee D-H, Lee B-J, Kim S-H, Lee K-B. Shear bond strength of porcelain to a new
millable alloy and a conventional castable alloy. The Journal of prosthetic dentistry.
2015;113(4):329-35.
Li KC, Prior DJ, Waddell JN, Swain MV. Comparison of the microstructure and phase
stability of as-cast, CAD/CAM and powder metallurgy manufactured Co–Cr dental
alloys. Dental Materials. 2015;31(12):e306-e15.
Markarian RA, Ueda C, Sendyk CL, Laganá DC, Souza RM. Stress distribution after
installation of fixed frameworks with marginal gaps over angled and parallel implants:
a photoelastic analysis. Journal of Prosthodontics. 2007;16(2):117-22.
Tuna SH, Pekmez NÖ, Kürkçüoğlu I. Corrosion resistance assessment of Co-Cr alloy
frameworks fabricated by CAD/CAM milling, laser sintering, and casting methods.
The Journal of prosthetic dentistry. 2015;114(5):725-34.
Van Noort R. The future of dental devices is digital. Dental Materials. 2012;28(1):3-12.
Volpato CÂM, Bondioli F, ́Altoé Garbelotto LGD, Fredel MC. Application of zirconia in
dentistry: biological, mechanical and optical considerations: INTECH Open Access
Publisher; 2011.
55