: 15297001
LaboPol-25
Instruction Manual
Spare Parts and Diagrams
LaboPol-21/ -25
Instruction Manual
Always state Serial No and Voltage/frequency if you have technical questions or when ordering spare parts.
You will find the Serial No. and Voltage on the type plate of the machine itself. We may also need the Date
and Article No of the manual. This information is found on the front cover.
The following restrictions should be observed, as violation of the restrictions may cause cancellation of Struers
legal obligations:
Instruction Manuals: Struers Instruction Manual may only be used in connection with Struers equipment
covered by the Instruction Manual.
Service Manuals: Struers Service Manual may only be used by a trained technician authorised by Struers.
The Service Manual may only be used in connection with Struers equipment covered by the Service Manual.
Struers assumes no responsibility for errors in the manual text/illustrations. The information in this manual is
subject to changes without notice. The manual may mention accessories or parts not included in the present
version of the equipment.
Original instructions. The contents of this manual is the property of Struers. Reproduction of any part of this
manual without the written permission of Struers is not allowed.
Struers
Pederstrupvej 84
DK-2750 Ballerup
Denmark
Telephone +45 44 600 800
Fax +45 44 600 801
LaboPol-21/ -25
Instruction Manual
LaboPol
Safety Precaution Sheet
To be read carefully
before use
1. The operator(s) should be fully aware of the use of the machine
according to the Instruction Manual.
3. Be sure that the actual voltage corresponds to the voltage stated on the
back of the machine. The machine must be earthed.
7. The machine must be disconnected from the mains prior to any service.
Wait until residual potential on the capacitors is discharged.
8. Do not cycle mains power more than once every three minutes.
Damage to the drive will result.
The equipment should only be used for its intended purpose and as detailed in the Instruction Manual.
The equipment is designed for use with consumables supplied by Struers. If subjected to misuse, improper
installation, alteration, neglect, accident or improper repair, Struers will accept no responsibility for damage(s)
to the user or the equipment.
Dismantling of any part of the equipment, during service or repair, should always be performed by a qualified
technician (electromechanical, electronic, mechanical, pneumatic, etc.).
LaboPol-21/ -25
Instruction Manual
Disposal
Equipment marked with a WEEE symbol contain electrical and
electronic components and must not be disposed of as general
waste.
Please contact your local authorities for information on the correct
method of disposal in accordance with national legislation.
LaboPol-21/ -25
Instruction Manual
User’s Guide
Table of Contents Page
2. Operation
Getting Acquainted with LaboPol ....................................................... 5
Mounting a Disc ................................................................................. 5
Starting the Machine .......................................................................... 5
Stopping the Machine ........................................................................ 5
1
LaboPol-21/ -25
Instruction Manual
1. Getting Started
Checking the Contents of In the packing box you should find the following parts:
Packing 1 LaboPol-21/ -25, complete with inlet hose
1 Power cable
2 Splash rings
1 Outlet hose
1 Hose clamp ø 25-40 mm
1 Hose nipple
1 Union nut, 1/2”
1 Gasket
1 Hose clamp
1 Set of Instruction Manuals
Unpacking LaboPol Remove the three screws, securing LaboPol onto the bottom of
the crate, using a 13 mm pipe spanner.
Lift LaboPol off the bottom plate and place it on a stable table.
Remove the two transportation support nuts at the bottom of
LaboPol using a 10 mm fork spanner (only on specific models).
Water pressure
Specimen mover On
Max. 10 bar Probenhalter
Wasserdruck
Pression d'eau
Min 1 bar Porte-èchantillon Off
Fuse 4 AT
Water outlet
Water inlet
On/off switch for specimen mover
Power supply socket for specimen mover
Communication socket for specimen mover
Fuse
Main switch
Power inlet
2
LaboPol-21/ -25
Instruction Manual
Supplying Power Always remember to switch the power off when installing electrical
equipment.
IMPORTANT
Check that the mains voltage corresponds to the voltage stated
on the type plate on the back of the machine.
Single-phase Supply The 2-pin (European Schuko) plug is for use on single-phase
connections.
2-phase Supply The 3-pin (North American NEMA) plug is for use on 2-phase power
connections.
Connection to the Machine Both cables are on the other end equipped with an IEC 320 cable
connector that has to be connected to the LaboPol.
WARNING!
The output voltage from this cable is 200 – 240V and not 110V.
DO NOT use this cable to connect equipment that use an 110V power
supply. Failure to adhere to this may result in material damage.
DANGER!
The machine must be earthed
IMPORTANT
If LaboPol has not been connected to a specimen mover, the switch
“specimen mover” must be put into off position.
Otherwise LaboPol cannot be started.
3
LaboPol-21/ -25
Instruction Manual
Supplying Water Connect the outlet hose to the water outlet on the back of the
Water Outlet machine and secure the hose using the hose clamp.
Lead the other end of the outlet hose to the drain. Be absolutely
sure to place the outlet hose with a steady slope to prevent the
water from being discharged too slowly, which may cause
overflow or blocking of the water outlet. Avoid sharp bends on
the outlet hose.
Water Inlet Connect the inlet hose to the water tap and secure with a hose
clamp. The fittings (gasket, union nut and reduction ring)
supplied with the machine may also be used for the connection.
4
LaboPol-21/ -25
Instruction Manual
2. Operation
Getting Acquainted with Take a moment to familiarise yourself with the location and names of
LaboPol the LaboPol components.
-25
LaboPol
Start switch
Stop switch
Turntable
Water tap
Speed selector (LaboPol-25 only)
Mounting a Disc Place the preparation disc on the turntable and turn it until the
three pins of the preparation disc engage with the holes in the
turntable.
5
LaboPol-21/ -25
Instruction Manual
Reference Guide
Table of Contents Page
1. Advanced Operations
Selection of Disc ...............................................................................7
Mounting of Specimen Mover (Optional) ...........................................7
4. Trouble-Shooting .......................................... 10
5. Maintenance
Daily Service ...................................................................................11
Weekly Service ...............................................................................11
6
LaboPol-21/ -25
Instruction Manual
1. Advanced Operations
Selection of Disc LaboPol is designed for use with wet grinding and polishing discs.
The diameter of wet grinding discs with ring should be 230 or
250 mm. When using polishing cloths, MD-Allegro or SiC-paper/ SiC-
foil, discs with 200 mm or 250 mm dia. can be used.
Please Note
To mount the specimen mover,
see the LaboForce Instruction Manual.
7
LaboPol-21/ -25
Instruction Manual
Metalog Guide™
A complete guide to materialographic specimen preparation.
www.struers.com/KNOWLEDGE/Metalog Guide.
8
LaboPol-21/ -25
Instruction Manual
Remember...
Struers offers a comprehensive range of consumables for grinding and
polishing. Please ask for separate leaflets.
9
LaboPol-21/ -25
Instruction Manual
4. Trouble-Shooting
Error Cause Action
Noise when the The belt is not sufficiently tight Call a Struers service technician.
machine starts or the The belt must be replaced.
machine will not pull
The machine does not - The main switch is off - Turn the main switch on
operate when the start - The fuse at the rear of LaboPol has blown - Replace the fuse
switch is pressed. - The switch for the specimen mover at the - Set the switch to “off”
back of LaboPol is set to “on”, but no
specimen mover is connected.
Water is not draining - Drain hose squeezed - Straighten the hose
away - Drain hose clogged - Clean the hose
- Drain hose does not slope away - Adjust the hose to an even
slope
Water dripping The water does not drain away See above
underneath the
machine
Cooling water stops - Water tap on mains closed - Turn the water on
- Built-in water tap closed - Turn the water on
- Built in water tap blocked - Clean water tap
- Filter at the water inlet blocked - Clean filter
The preparation disc Dirt underneath the disc or on the turntable Clean the contact face between
vibrates the disc and the turntable.
Continuous, irregular Coupling on either the specimen holder/mover Please contact a Struers Service
wear on a plate or the specimen mover head of the Technician to replace the
grinding/polishing polishing machine is worn. coupling.
surface.
10
LaboPol-21/ -25
Instruction Manual
5. Maintenance
Daily Service Clean all accessible surfaces with a moist cloth.
Weekly Service Remove the splash ring and the preparation disc, and remove all
dirt from the tray and the drain tube.
11
LaboPol-21/ -25
Instruction Manual
6. Technical Data
Subject Specifications
LaboPol-21 LaboPol-25
Disc Diameter 200-250mm / 8-10" 200-250mm / 8-10"
Speed 250 rpm 50-500 rpm
Rotational direction Counter-clockwise
Motor 250 W 250 W
Torque at Continuous 9.5Nm 9.5Nm @ 250 rpm
disc Max. >12.0Nm >12.0Nm@ 250 rpm
Safety standards Please refer to the Declaration of Conformity
12
529 LaboPol-21
529F_2016
The above has been declared according to the global method, module A
Dansk Overensstemmelseserklæring
Fabrikant, Struers A/S
ansvarlig for Teknisk Pederstrupvej 84
Dossier DK-2750 Ballerup, Danmark
Telefon 44 600 800
erklærer herved, at Produktnavn: LaboPol-21
Type nr.: 529
Maskintype: Slibe og polér maskine
S
586D_2015
The above has been declared according to the global method, module A
Dansk Overensstemmelseserklæring S
Fabrikant, Struers A/S
ansvarlig for Teknisk Pederstrupvej 84
Dossier DK-2750 Ballerup, Danmark
Telefon 44 600 800
erklærer herved, at Produktnavn: LaboPol/-25
Type nr.: 586
Maskintype: Slibe og polér maskine
Spare Parts
and Diagrams
Table of contents Drawing
Struers assumes no responsibility for errors in the manual text/illustrations. The information in this manual is
subject to changes without notice. The manual may mention accessories or parts not included in the present
version of the equipment. The contents of this manual is the property of Struers. Reproduction of any part of
this manual without the written permission of Struers is not allowed.
Struers
Pederstrupvej 84
DK-2750 Ballerup
Denmark
Telephone +45 44 600 800
1
LaboPol-21/ -25
Spare Parts and Diagrams
2
LaboPol-21/ -25
Spare Parts and Diagrams
3
1 2 3 4 5 6 7 8
P os P A R T NU M B E R Q T Y.
N o.
5 15180605 Bundramme, 1
A A
bearbejdet_ej faerdig
10 15180200 Axle for sample disk 1
20 2BK00045_sporkugleleje-6004- 1
2RSR
Rustbeskyttes m ed 2GF60020 Boelgefjeder
Tellus S32 olie på 30 28x34.5x0.5 1
øverste flade og sider 2BK00040_sporkugleleje-6003-
40 2RSR 1
50 15180520 Remskive ø230, 1
bearbejdet
B Hård drivpasning 10 60 2ZF10520 key 5x5x20 1 B
70 2ZA20010 Skive 10 DIN 9021 A 1
Drivpasning A2
80 2TR70620 MK skrue M6x20 A2 1
20
Drivpasning
Let drivpasning
30
C C
40
D D
5 50 60 80 70
E
Loctite 222
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15180050 Lejekonstruktion, monteret C
B
1 2 3 4 5 6 7 8
A A
120 170 13 0 60 80 70 90 Rotate when inserting
11 0
B
100 B
20
40
Do not grease
C C
20
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 14920065 Water tap, assembly H
1 2 3 4 5 6 7 8
QS1 X1
2x5AMP
W1 F1
4 BK1
L1
A A
MUST BE SHORT CIRCUIT
PROTECTED WITH EXTERNAL 2
FUSES ACCORDING TO LOCAL BK2
L2
REGULATIONS
F = 16Amp MAXIMUM F2 PE YE/GN
PE
3 5 3 1
3x1mm²
POWER ON
REAR
PLATE S2 K1
START
6 4 2
XPE1 4
X3
W3 SELECT
S3 XPE2
1 REAR
FROM FT2 LaboForce-1/-3 PLATE
3 BK5 2
W4 5x1mm²
LaboForce-3 FT2
2 X5
THERMAL PROTECTION M2 PE
BK5 BK2 BK1
W4
Comment: 3x0,75mm²
REAR *
PLATE C1
t>155C M U5
XPE2
2~
THERMAL
PROTECTION
C M1 C
T2 180W
W4 BK4 YE/GN W4 M1
A2
K1
XPE3
A1
BK2
BL = BLUE
XPE5 XPE6 XPE4 XPE3
110-120V/60Hz. 30µF LaboPol-21
GY = GREY BOTTOM FRONT BOTTOM REAR 220-240V/50Hz. 10µF
Thursday, September 13, 2001 Size CAGE CODE DWG NO Rev
WH = WHITE PLATE PLATE FRAME PLATE 220-240V/60Hz. 8µF
A2 15293100
Monday, December 11, 2000 SCALE Sheet of
DEM/DEM 1 1
1 2 3 4 5 6 7 8
X1 A1 M1
W1 R1
220k, 1W 2x 4,0 AT
FREQUENCY INVERTER MOTOR
4
L1 L1 L1 Plug-In Module Note:
200-240 V / 50-60 Hz EPM (Machine setup) U U
2
N / L2 L2 N / L2
V V
F1 + F2 PE 3-Phase motor
PE PE
PE
W W 240V / 60Hz
Power Entry Module
with main power switch
MUST BE CONNECTED PE PE
TO A PROTECTIVE
7 (Common)
20 (+12V dc)
EARTH CONNECTION
External Fuse
Max. 16 Amp. XPE1 XPE2
+temp.
K14
K12
E28
E1
E2
E3
MOUNTING MOUNTING
(120C)
9
PLATE PLATE
Note:
On LaboPol-21/25
U is connected to V and
V is connected to U
2B 2
M1 motor data:
LaboPol-1: 4-poles, 250W (1/3 H.P.)
LaboPol-2: 4-poles, 250W (1/3 H.P.)
LaboPol-21: 4-poles, 250W (1/3 H.P.)
LaboPol-25: 4-poles, 250W (1/3 H.P.)
LaboPol-4: 8-poles, 80W (1/9 H.P.)
LaboPol-5: 4-poles, 250W (1/3 H.P.)
LaboPol-6: 4-poles, 250W (1/3 H.P.)
LaboPol-35: 4-poles, 370W (1/2 H.P.)
K1
+
1K S4 (on back)
X2
EXTERNAL EQUIPMENT
ACTIVE / NON ACTIVE
-
POWER OUTLET TO X3
EXTERNAL EQUIPMENT
1
1 3
S4
2
1A
XPE3
MOUNTING
STOP SIGNAL FROM
PLATE
XPE4
EXTERNAL
MOUNTING EQUIPMENT
1 A 2 PLATE
2 E.G.: THERMAL PROTECTION IN SAMPLE MOVER
1 3 3 4 1 2
STRUERS A/S
Rev.B: FTH 31-10-2007 VALHOEJS ALLÈ 176
R1 resistor added in Power Switch X1 DK-2610 ROEDOVRE
DENMARK
Rev.C: FTH 28-08-2009 PHONE: +45 3670 3500
LaboPol-35 added
XPE5 XPE6 XPE7 XPE8 XPE9 XPE10 XPE11 XPE12
LABOPOL-SERIES, CIRCUIT DIAGRAM
MOUNTING BOTTOM BOTTOM ALU. BOTTOM FRONT BOTTOM DISK
PLATE PLATE PLATE CHASSIS PLATE PLATE PLATE FOUNDATION Size CAGE Code DWG NO Rev
(MOTOR PLATE) A2 <Cage Code> 15863100 C
Scale Sheet
Friday, August 28, 2009 FTH / FTH of
1 1
X1 A1 M1
R1
220k, 1W 2x 4,0 AT
X10 FREQUENCY INVERTER W2 MOTOR
L BK1 1 BK3 1.5 mm2
L1 Note: M1 motor data:
200-240 V
BK4
Plug-In Module U BK1 U LaboPol-1: 4-poles, 250W (1/3 H.P.)
N
50-60 Hz BK2 LaboPol-2: 4-poles, 250W (1/3 H.P.)
EPM (Machine setup)
IN 2 BK5 V BK2 V LaboPol-21: 4-poles, 250W (1/3 H.P.)
F1 + F2 PE YE/GN
N / L2 3-Phase motor LaboPol-25: 4-poles, 250W (1/3 H.P.)
240V / 60Hz LaboPol-4: 8-poles, 80W (1/9 H.P.)
W BK3 W LaboPol-5: 4-poles, 250W (1/3 H.P.)
Power Entry Module
with main power switch
YE/GN
PE LaboPol-6: 4-poles, 250W (1/3 H.P.)
PE YL/GN PE LaboPol-35: 4-poles, 370W (1/2 H.P.)
7 (Common)
20 (+12V dc)
Note:
On LaboPol-21/25
XPE1 BK1 is connected to V and +temp.
K14
K12
E28
E1
E3
E2
X2 MOUNTING
BK2 is connected to U
(120C)
9
PLATE
XPE2
MOUNTING
PLATE
L BK7
200-240 V
50-60 Hz N BK6
OUT
PE
W4
YE/GN 8x 0.25 R1
Strap Strap 1K
BN WH GN RD BU YE GY 0.25W
K1 1K
-
+
1 2 3 4 5 6 7 8 9 10 11 12 X11
XPE3
MOUNTING
PLATE
0.75, BU12
0.75, BU1
W3
2x 0.75
RD2
YE/GN YE/GN YE/GN
RD1
0.75, BU8
0.75, BU6
W5 W6
XPE5 XPE6 XPE7 XPE8 XPE9 XPE10
3x 0.25 4x 0.5 0.75, BU11
MOUNTING BOTTOM BOTTOM ALU. BOTTOM FRONT 0.75, BU10
PLATE PLATE PLATE CHASSIS PLATE PLATE
(MOTOR PLATE) BN
WH 0.75, BU8
LABOPOL-21/25 ONLY GN 0.75, BU7
YE/GN
X3
BU BN BK BE 1
YE/GN 2
XPE11 XPE12 1 2 3
BOTTOM DISK
1 A 2 3 4 1 2 1 1A 2 2B
PLATE FOUNDATION
STOP SIGNAL FROM
EXTERNAL EQUIPMENT
XPE4 E.G.: THERMAL PROTECTION IN SAMPLE MOVER
MOUNTING
RP1 (on front) S3 (on front) S1 (on front) S2 (on front) S4 (on back) PLATE
SET SPEED LOW / HIGH START STOP Ext. Equipm.
SPEED YES / NO
ALL WIRINGS AWG-16, EXCEPT OTHERWISE MARKED.