AQM60

Environmental
Station
Section Two:
Service Guide
V5 Rev. 1

Aeroqual Limited
109 Valley Road, Mt Eden, Auckland 1024, New Zealand
Phone: +64 9 623 3013
Fax: +64 9 623 3012
Web: www.aeroqual.com
Contents
1.0 Service ...................................................................................................................................................................................... 1
1.1 Service Schedule ...................................................................................................................................................................... 1
1.2 Service Procedures .................................................................................................................................................................. 2
1.2.1 System Service ............................................................................................................................................................... 2
1.2.2 Sampling Lines ............................................................................................................................................................. 35
1.2.3 Controller Module ......................................................................................................................................................... 37
1.2.4 Gas Treatment Module ................................................................................................................................................. 49
1.2.5 Gas Sensor Module ...................................................................................................................................................... 63
1.2.6 Particle Monitor Module ................................................................................................................................................ 69

Aeroqual AQM 60 User Guide 2


1.0 Service
Predictive diagnostic functions including data acquisition and event logging built into the Aeroqual AQM60 allow
the user to determine when repairs are necessary without performing unnecessary, preventive maintenance
procedures.

There is, however, a minimal number of simple procedures that, when performed regularly, will ensure that the
monitor continues to operate accurately and reliably over its lifetime.

1.1 Service Schedule


An indicative service schedule is shown below. Depending on the local conditions and your data integrity
requirements, you may wish to tailor the schedule to your specific application:

SECTION TASK FREQUENCY


System Service Inspect the mains power cable for damage 1x / YEAR
System Service Inspect the enclosure for damage or abnormalities 1x / YEAR
System Service Measure the flow rate at the sample inlet 1-4x / YEAR
System Service Perform manual Gas Sensor zero check / calibration 1-4x / YEAR

System Service Perform manual Gas Sensor span check / calibration 1-4x / YEAR

System Service Perform a factory Particle Monitor span calibration 1 / 2 YEARS

System Service Inspect the event log 1x / MONTH


Sampling Lines Replace the inlet filter 1-4x / MONTH
Sampling Lines Inspect all connections and tubing 1-4x / YEAR
Controller Verify the controller display is functional 1x / YEAR
Controller Verify the serial communications are functional 1-4x / YEAR
Controller Verify all of the sensor modules are logging correctly 1-4x / YEAR
Controller Verify the SD card is logging correctly 1x / MONTH
Controller Verify the controller parameters are as expected 1-4x / YEAR
Gas Treatment Module Measure the flow rate of the gas treatment module pump 1x / YEAR
Gas Treatment Module Leak test the gas treatment module pump 1x / YEAR
Gas Treatment Module Verify that the solenoid is switching 1x / YEAR
Gas Treatment Module Replace scrubber media 1-4x / YEAR
Gas Sensor Module Measure the flow rate 1x / YEAR
Gas Sensor Module Leak test (pumped type) 1x / YEAR
Leak test (non-pumped type)
Gas Sensor Module Verify the gas sensor module parameters are as expected 1-4x / YEAR
Particle Monitor Module Measure the flow rate 1x / YEAR
Particle Monitor Module Clean the cyclone 1-4x / YEAR
Particle Monitor Module Perform a fibre span check 4x / YEAR
Particle Monitor Module Clean the particle monitor filters 1/ 2 YEARS

Aeroqual AQM 60 User Guide 1


1.2 Service Procedures
This section details the service procedures outlined in the service schedule.

NOTE: The service procedures in this manual are restricted to qualified service representatives.

1.2.1 System Service

Title: Site Report


Item: System Service
Code: SYS001
Preparation: N/A
Specifications: N/A
Procedure:

(a) Record the Monitor information.


HINT: Note the Monitor ID.
HINT: Note the GSM modem number (if applicable).
HINT: Note the IP address (if applicable).
HINT: Note the SD card number.
(b) Record the Site information.
HINT: Note the address.
HINT: Note the GPS position.
HINT: Note the type of site (urban, rural, industrial, roadside, fenceline etc…).
HINT: Note the local pollutants (urban, traffic, petrochemical plant, geothermal, combustion sources etc…).
HINT: Note the expected pollutant levels (background, low, medium, high).
(c) Record the Mounting information.
HINT: Note the monitor placement (inside or outside).
HINT: Note the height the monitor is mounted above the ground.
HINT: Note the mounting structure (pole, wall, roof, other etc…).
HINT: Note what type of solar protection is present (none, shade panels, shaded site, north/south facing).
HINT: Note the security measures (wire cage surround, key accessible site only, hidden, none etc…).
(d) Record the type of service you are performing.
(e) Record where the service has taken place (on-site or at the factory).
(f) Take a photograph of the site.
(g) Record the time and date.
(h) Record the ambient temperature.
(i) Record the ambient humidity.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 2


SYSTEM SERVICE – INSPECT THE MAINS POWER CABLE FOR DAMAGE
Title: Inspect the mains power cable for damage.
Item: System Service
Code: SYS002
Preparation: N/A
Specifications: N/A
Procedure:

(a) Inspect the mains power cable leading to the junction box for damage.
HINT: Check the cable has not been pulled from the strain relief.
HINT: Check the insulation for signs of mechanical damage.
HINT: Check the insulation for signs of electrical damage.
HINT: Check the insulation for signs of heat damage.
WARNING: If damage is observed an assessment regarding the safety of the system must be made and
steps to prevent any personal injury and any further damage to the instrument should be taken.
(b) Record observations in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 3


SYSTEM SERVICE – INSPECT THE ENCLOSURE FOR DAMAGE OR ABNORMALITIES
Title: Inspect enclosure for damage or abnormalities.
Item: System Service
Code: SYS003
Preparation: N/A
Specifications: N/A
Procedure:

(a) Inspect enclosure for external damage or abnormalities.


HINT: Check to see if the enclosure has been pierced.
HINT: Check to see if externally mounted sensors have been damaged.
HINT: Check the door is sealed correctly.
HINT: Check all of the O-rings and glands are sealed correctly.
HINT: Check the enclosure is securely fastened to the mounting bracket.
(b) Inspect enclosure internally for damage.
(c) Inspect the insulation for damage.
HINT: Check the insulation is adhered securely to the enclosure.
HINT: Check the insulation is not torn or compressed.
HINT: Check the insulation is not wet.
(d) Inspect the mounting of all components.
HINT: Check the modules are securely located on the DIN rail.
HINT: Check the transit fasteners are in place.
(e) Inspect inside the enclosure for dust.
(f) Inspect inside the enclosure for moisture.
(g) Record observations in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 4


SYSTEM SERVICE – MEASURE THE FLOW RATE OF THE SAMPLE INLET
Title: Measure the flow rate of the sample inlet.
Item: System Service
Code: SYS004
Preparation: Rotameter.
Specifications: N/A
Procedure:

(a) Connect the rotameter to the Sample Inlet of the monitor.


HINT: Leave the air inlet side of the rotameter open to atmosphere.
HINT: Connect the air outlet side of the rotameter to the Sample Inlet.

(b) Turn on the power to the AQM.


(c) Open the Aeroqual AQM software.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 5


(d) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

(e) Ensure that the zero scrubber is turned OFF.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 6


(f) If the zero scrubber is turned ON, turn off the zero air scrubber.

(g) Measure the flow rate of the air through the sample inlet.
HINT: This is the flow rate of the sample inlet, not the flow rate of the sample inlet through the scrubber.
HINT: If you have an Ozone module be aware that the Ozone module pump cycles on and off about every
thirty seconds. Wait for the Ozone module pump to turn ON before taking a measurement.
(h) Verify the value is within specification.
(i) Record the measurement in the instrument Log Book.
(j) Turn ON the zero scrubber.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 7


(k) Ensure the zero air scrubber is turned ON.

(l) Measure the flow rate of the air through the sample inlet.
HINT: This is the flow rate of the sample inlet through the scrubber.
HINT: If you have an Ozone module be aware that the Ozone module pump cycles on and off about every
thirty seconds. Wait for the Ozone module pump to turn ON before taking a measurement.
(m) Verify the value is within specification.
(n) Record the measurements in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 8


SYSTEM SERVICE – PERFORM A ZERO CHECK
Title: Perform a zero check.
Item: System Service
Code: SYS005
Preparation: Zero air source.
Specifications: N/A
Procedure:

(a) Ensure the AQM is standing upright with the door closed.
(b) Turn on the power to the AQM.
(c) Open the Aeroqual AQM software.
(d) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

(e) Ensure that the zero scrubber is turned OFF.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 9


(f) If the zero scrubber is turned ON, turn OFF the zero air scrubber.

(g) Open the real-time logging table using the Data  Table  Real Time… menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 10


(h) Start data logging using the File  Start Data Logging menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 11


(i) Connect the Zero Air source to the Sample Inlet.
HINT: Use a T-piece to connect to the Gas Sensor module inlet to allow excess flow/pressure to bleed out
the exhaust.
HINT: The zero air can be humidified using the Nafion membrane humidifier as long as there are low levels
of VOCs present in the room.

(j) Turn on the zero air source.


HINT: The zero air flow rate should be at least 1.5 LPM.
(k) Sample the zero air until stable readings are obtained.
HINT: This could take approximately 30 minutes.
(l) Note the zero values in the instrument Log Book.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 12


(m) Turn ON the zero scrubber.

(n) Ensure the zero air scrubber is turned ON.

(o) Sample the zero air until stable readings are obtained.
HINT: This could take approximately 30 minutes.
(p) Note the zero values through the zero scrubber in the instrument Log Book.
(q) Verify the sample readings and the zero scrubber readings are the same.
HINT: If the readings are different then you may need to replace your scrubber media.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 13


(r) Turn OFF the zero scrubber.

(s) Verify the zero scrubber is OFF.

(t) If the sensor baseline readings are outside the specified acceptable range then a zero calibration should
be performed.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 14


SYSTEM SERVICE – PERFORM A ZERO CALIBRATION
Title: Perform a zero calibration.
Item: System Service
Code: SYS006
Preparation: Zero air source.
Specifications: N/A
Procedure:

(a) Ensure the AQM is standing upright with the door closed.
(b) Turn on the power to the AQM.
(c) Open the Aeroqual AQM software.
(d) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

(e) Ensure that the zero scrubber is turned OFF.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 15


(f) If the zero scrubber is turned ON, turn OFF the zero air scrubber.

(g) Open the operations dialog using the Setup  Operations menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 16


(h) Turn the Auto Zero function to Auto Zero Function OFF and click Save.

(i) Open the real-time logging table using the Data  Table  Real Time… menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 17


(j) Start data logging using the File  Start Data Logging menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 18


(k) Connect the Zero Air source to the Sample Inlet.
HINT: Use a T-piece to connect to the Gas Sensor module inlet to allow excess flow/pressure to bleed out
the exhaust.
HINT: The zero air can be humidified using the Nafion membrane humidifier as long as there are low levels
of VOCs present in the room.

(l) Turn on the zero air source.


HINT: The zero air flow rate should be at least 1.5 LPM.
(m) Sample the zero air until stable readings are obtained.
HINT: This could take approximately 30 minutes.
(n) Note the zero values in the instrument Log Book.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 19


(o) Turn ON the zero scrubber.

(p) Ensure the zero air scrubber is turned ON.

(q) Sample the zero air until stable readings are obtained.
HINT: This could take approximately 30 minutes.
(r) Note the zero values through the zero scrubber in the instrument Log Book.
(s) Verify the sample readings and the zero scrubber readings are the same.
HINT: If the readings are different then you may need to replace your scrubber media.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 20


(t) Turn OFF the zero scrubber.

(u) Verify the zero scrubber is OFF.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 21


(v) Begin the zero calibration using the Tools  Zero Cal menu.
HINT: You should head the zero air scrubber turn on.

(w) Wait for the zero calibration to end.


HINT: This takes approximately 20 minutes.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 22


SYSTEM SERVICE – PERFORM A SPAN CHECK
Title: Perform a Span Check.
Item: System Service
Code: SYS007
Preparation: Calibrated Span Gas Source
Specifications: N/A
Procedure:

(a) Ensure the AQM is standing upright with the door closed.
(b) Turn on the power to the AQM.
(c) Open the Aeroqual AQM software.
(d) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

(e) Ensure that the zero scrubber is turned OFF.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 23


(f) If the zero scrubber is turned ON, turn OFF the zero air scrubber.

(g) Open the operations dialog using the Setup  Operations menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 24


(h) Turn the Auto Zero function to Auto Zero Function OFF and click Save.

(i) Open the real-time logging table using the Data  Table  Real Time… menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 25


(j) Start data logging using the File  Start Data Logging menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 26


(k) Connect the Calibrated Span Gas source to the Sample Inlet.
HINT: Use a T-piece to connect to the Gas Sensor module inlet to allow excess flow/pressure to bleed out
the exhaust.
HINT: The calibrated span gas can be humidified using the Nafion membrane humidifier as long as there are
low levels of VOCs present in the room.

(l) Turn on the calibrated span gas source.


HINT: The calibrated span gas flow rate should be at least 1.5 LPM.
(m) Sample the calibrated span gas until stable readings are obtained.
HINT: This could take approximately 30 minutes.
(n) Note the span values in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 27


SYSTEM SERVICE – PERFORM A SPAN CALIBRATION
Title: Perform a Span Calibration.
Item: System Service
Code: SYS008
Preparation:
Specifications: N/A
Procedure:

(a) Ensure the AQM is standing upright with the door closed.
(b) Turn on the power to the AQM.
(c) Open the Aeroqual AQM software.
(d) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

(e) Ensure that the zero scrubber is turned OFF.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 28


(f) If the zero scrubber is turned ON, turn off the zero air scrubber.

(g) Open the real-time logging table using the Data  Table  Real Time… menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 29


(h) Start data logging using the File  Start Data Logging menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 30


(i) Connect the Calibrated Span Gas source to the Sample Inlet.
HINT: Use a T-piece to connect to the Gas Sensor module inlet to allow excess flow/pressure to bleed out
the exhaust.
HINT: The calibrated span gas can be humidified using the Nafion membrane humidifier as long as there are
low levels of VOCs present in the room.

(j) Turn on the calibrated span gas source.


HINT: The calibrated span gas flow rate should be at least 1.5 LPM.
(k) Sample the calibrated span gas until stable readings are obtained.
HINT: This could take approximately 30 minutes.
(l) Note the span values in the instrument Log Book.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 31


(m) Select Tools  Modify Factors…

(n) Enter the system password.


HINT: The factory set user password is password.

(o) Adjust the gain factor for the sensor you are calibrating until its reading matches span calibration gas
mixture and click OK.

(p) Repeat steps (g) to (m) for each span gas.


(q) Record the Gain Factors in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 32


SYSTEM SERVICE – INSPECT THE EVENT LOG
Title: Inspect the Event Log.
Item: System Service
Code: SYS009
Preparation: N/A
Specifications: N/A
Procedure:

(a) Open Aeroqual AQM software.


(b) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

HINT: You can do this remotely using a GSM modem by selecting File  Dial GSM Modem…

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 33


(c) Download the event log from the SD data card.

(d) Check the event log for events which may require intervention.
HINT: In many cases the event log can be used to determine whether a site visit is necessary.

Event Log Examples Meaning


Pump failure
O3 :P.F. 2008/11/12 23:45 O3 module pump failure on 12
Nov 2008 at 23:45
Sensor failure
NO2 :S.F. 2008/11/12 23:45 NO2 module sensor failure on 12
Nov 2008 at 23:45
AQM Power turned on at 10:12
Power on: 2008/11/12 10:12 am on 12 November 2008
Zero Calibration started at 11:24
Zero Cal at 2008/11/12 11:24 on 12 November 2008

AQM configuration saved at 1:12


Config at 2008/11/12 13:12 pm on 12 November 2008
0x01,05,30,40,50,B0,F8,00,00,00,00,00,00,00,00,00,00 Sensor codes listed
O3 module no data report at
O3 : NR 2008/11/12 12:42 12:42 pm on 12 November 2008
NO2 scrubber failed at 3:36 pm
NO2 Scrubber FAILED: 2008/11/12 15:36 on 12 November 2008

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 34


1.2.2 Sampling Lines
Title: Replace the inlet filter.
Item: Sampling Lines
Code: SMP001
Preparation: 5um particulate inlet filter.
Specifications: N/A
Procedure:

(a) Remove the inlet filter.


(b) Insert the new inlet filter.
(c) Ensure both connections are securely fastened.
(d) Record the filter change in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 35


SAMPLING LINES – INSPECT ALL CONNECTIONS AND TUBING
Title: Inspect all connections and tubing
Item: Sampling Lines
Code: SMP002
Preparation: 5um particulate inlet filter.
Specifications: N/A
Procedure:

(a) Inspect all of the inlet and outlet connections inside the AQM.
HINT: All connections should either be connected or be blocked by a Luer cap.
(b) Inspect the sample and exhaust tubing inside the AQM.
HINT: Check for condensation inside the tubing.
HINT: Check for dust inside the tubing.
(c) Record any observations in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 36


1.2.3 Controller Module

CONTROLLER MODULE – CHECK THE CONTROLLER DISPLAY IS FUNCTIONAL


Title: Check the controller display is functional
Item: Controller Module
Code: CON001
Preparation: N/A
Specifications: N/A
Procedure:

(a) Open the door of the AQM60.


(b) Observe the display on the controller.
HINT: This should be scrolling with sensor readings.
(c) Note your findings in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 37


CONTROLLER MODULE – CHECK THAT THE RS232 COMMUNICATIONS ARE FUNCTIONAL
Title: Check that the RS232 communications are functional
Item: Controller Module
Code: CON002
Preparation: N/A
Specifications: N/A
Procedure:

(a) Open Aeroqual AQM software.


(b) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

HINT: You can do this remotely using a GSM modem by selecting File  Dial GSM Modem…

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 38


(c) When connected to the module the Search Monitor Successful dialog box appears. This means the
RS232 communications are working as expected.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 39


CONTROLLER MODULE – CHECK THAT ALL OF THE SENSOR MODULES ARE LOGGING
CORRECTLY
Title: Check that all of the sensor modules are logging
Item: Controller Module
Code: CON003
Preparation: N/A
Specifications: N/A
Procedure:

(a) Open Aeroqual AQM software.


(b) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

HINT: You can do this remotely using a GSM modem by selecting File  Dial GSM Modem…

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 40


(c) Check the modules listed match the modules fitted (refer to CON005 for details).
(d) Open the real-time logging table using the Data  Table  Real Time… menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 41


(e) Start data logging using the File  Start Data Logging menu.

(f) Wait for approximately two minutes until you receive a reading.
HINT: All sensors should report a reading.
(g) Record your findings in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 42


CONTROLLER MODULE – CHECK THAT THE SD DATA CARD IS LOGGING CORRECTLY
Title: Check that the SD data card is logging correctly.
Item: Controller Module
Code: CON004
Preparation: N/A
Specifications: N/A
Procedure:

(a) Open Aeroqual AQM software.


(b) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

HINT: You can do this remotely using a GSM modem by selecting File  Dial GSM Modem…

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 43


(c) Start data logging using the File  Start Data Logging menu.

(d) Wait for five minutes so that one or two measurements will have been logged to the SD card.
(e) Open the Download File dialog using the File  Download Files….menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 44


(f) Choose the file with today’s date is in the list and click SAVE.

(g) Open the file and verify that the data is being logged correctly.
(h) Record your findings in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 45


CONTROLLER MODULE – CHECK THE PARAMETERS ARE AS EXPECTED
Title: Check the parameters are as expected.
Item: Controller Module
Code: CON005
Preparation: N/A.
Specifications: N/A
Procedure:

(a) Open Aeroqual AQM software.


(b) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

HINT: You can do this remotely using a GSM modem by selecting File  Dial GSM Modem…

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 46


(c) Open the Configuration View Dialog using the Tools  View Configuration… menu.

(d) Check that the sensor list is correct.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 47


(e) Open the View Factors Dialog using the Tools  View Factors… menu.

(f) Check that the factors are correct.

(g) Record your findings in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 48


1.2.4 Gas Treatment Module

GAS TREATMENT MODULE – MEASURE THE FLOW RATE OF THE GAS TREATMENT
MODULE PUMP
Title: Measure the flow rate of the gas treatment module pump.
Item: Gas Treatment Module
Code: GTM001
Preparation: Rotameter.
Specifications: N/A
Procedure:

(a) Disconnect the vacuum inlet connection of the Gas Treatment module.
(b) Disconnect the pressure outlet connection of the Gas Treatment module.
(c) Connect the rotameter to the Gas Treatment module.
HINT: Leave the air inlet side of the rotameter open to atmosphere.
HINT: Connect the air outlet side of the rotameter to the Gas Treatment module vacuum side.
(d) Turn on the power to the AQM.
(e) Measure the flow rate of the air through the gas treatment module pump.
(f) Verify the value is within specification.
(g) Record the measurement in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 49


GAS TREATMENT MODULE – LEAK TEST THE GAS TREATMENT MODULE PUMP
Title: Leak test the gas treatment module pump.
Item: Gas Treatment Module
Code: GTM002
Preparation: Rotameter.
Luer cap.
Specifications: N/A
Procedure:

(a) Disconnect the vacuum inlet connection of the Gas Treatment module.
(b) Disconnect the pressure outlet connection of the Gas Treatment module.
(c) Connect the rotameter to the Gas Treatment module.
HINT: Connect the air inlet side of the rotameter to the Gas Treatment module pressure side.
HINT: Leave the air outlet side of the rotameter open to atmosphere.
(d) Turn on the power to the AQM.
(e) Block the inlet port of the Gas Treatment module using the Luer cap.
(f) Verify that the flow rate shown on the rotameter drops to zero.
(g) Record your observations in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 50


GAS TREATMENT MODULE – CHECK THAT THE SOLENOID IS SWITCHING
Title: Check that the solenoid is switching.
Item: Gas Treatment Module
Code: GTM003
Preparation: 2x rotameter, pump.
Specifications: N/A
Procedure:

(a) Disconnect the vacuum inlet connection of the Gas Treatment module.
(b) Disconnect the pressure outlet connection of the Gas Treatment module.
(c) Disconnect the sample inlet connection of the Gas Treatment module.
(d) Disconnect the zero inlet connection of the Gas Treatment module.
(e) Disconnect all of the sample manifold outlet connections of the Gas Treatment module.
(f) Connect the two rotameters to the Gas Treatment module.
HINT: Connect the air outlet side of rotameter one to the Gas Treatment module sample inlet.
HINT: Connect the air outlet side of rotameter two to the Gas Treatment module zero air inlet.
HINT: Leave the air inlet side of both rotameters open to atmosphere.
(g) Block all but one of the sample manifold outlet connections of the Gas Treatment module using Luer
caps.
(h) Connect the pump to the one unblocked sample manifold outlet connection.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 51


(i) Open the Aeroqual AQM software.
(j) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

(k) Ensure that the zero scrubber is turned OFF.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 52


(l) If the zero scrubber is turned ON, turn off the zero air scrubber.

(m) Measure the flow rate of the air through the sample inlet.
(n) Measure the flow rate of the air through the zero air inlet.
HINT: This should be zero.
(o) Verify the value is within specification.
(p) Record the measurement in the instrument Log Book.
(q) Turn ON the zero scrubber.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 53


(r) Ensure the zero air scrubber is turned ON.

(s) Measure the flow rate of the air through the sample inlet.
HINT: This should be zero.
(t) Measure the flow rate of the air through the zero air inlet.
(u) Verify the value is within specification.
(v) Record the measurements in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 54


GAS TREATMENT MODULE – REPLACE SCRUBBER MEDIA
Title: Replace Scrubber Media
Item: Gas Treatment Module
Code: TM004
GT
Preparation:
Specifications: A
N/A
Procedure:

e Gas Treatment module.


(a) Remove the cover from the
HINT: Remove the two screws shown
s in the photograph.

(b) Undo luer fitting marked in red and separate the cover from the Gas Treatment Module
M base.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 55


(c) Pull the large scrubber filter out of the housing slightly to reveal the mounting scre
ews.

(d) Remove the two screws annd then the zero air scrubber assembly.
HINT: You will need to remove
e the soft tubing which connects the zero air scrubbe
er assembly to the Gas
Treatment Module cover.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 56


(e) Replace the used zero air scrubber
s assembly with the new zero air scrubber asse
embly.
(f) Connect the zero air scrubb ber assembly to the gas treatment module cover.
(g) Reconnect the luer fitting.

(h) Refit the cover.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 57


END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 58


GAS TREATMENT MODULE – LEAK TEST THE GTM SAMPLE AND SCRUBBER INLETS.
Title: Leak test the GTM sample and scrubber inlets.
Item: Gas Treatment Module
Code: GTM005
Preparation: 1x rotameter, pump.
Specifications: N/A
Procedure:

(a) Disconnect the vacuum inlet connection of the Gas Treatment module.
(b) Disconnect the pressure outlet connection of the Gas Treatment module.
(c) Disconnect the sample inlet connection of the Gas Treatment module.
(d) Disconnect the zero inlet connection of the Gas Treatment module.
(e) Disconnect all of the sample manifold outlet connections of the Gas Treatment module.
(f) Block both the sample and the zero inlet connection of the Gas Treatment module using Luer caps.
(g) Block all but one of the sample manifold outlet connections of the Gas Treatment module using Luer
caps.
(h) Connect the pump to the one unblocked sample manifold outlet connection.
(i) Connect the rotameter to the outlet of the pump.
HINT: Connect the air inlet side of the rotameter to the pump outlet.
HINT: Leave the air outlet side of the rotameter open to atmosphere.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 59


(j) Open the Aeroqual AQM software.
(k) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

(l) Ensure that the zero scrubber is turned OFF.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 60


(m) If the zero scrubber is turned ON, turn off the zero air scrubber.

(n) Measure the flow rate of the air on the rotameter.


HINT: This should be zero – if not there is a leak in the sample inlet portion of the Gas Treatment Module
pneumatics.
(o) Record the measurement in the instrument Log Book.
(p) Turn ON the zero scrubber.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 61


(q) Ensure the zero air scrubber is turned ON.

(r) Measure the flow rate of the air through the zero air inlet.
HINT: This should be zero – if not there is a leak in the zero scrubber inlet portion of the Gas Treatment
Module pneumatics.
(s) Record the measurement in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 62


1.2.5 Gas Sensor Module

GAS SENSOR MODULE – MEASURE THE FLOW RATE OF THE GAS SENSOR MODULE
Title: Measure the flow rate of the gas sensor module.
Item: Gas Sensor Module
Code: GSM001
Preparation: Rotameter.
Luer cap.
Specifications: N/A
Procedure:

(a) Disconnect the sample inlet connection from the Gas Sensor module.
(b) Connect the rotameter to the Gas Sensor module.
HINT: Leave the air inlet side of the rotameter open to atmosphere.
HINT: Connect the air outlet side of the rotameter to the Gas Sensor module inlet.
(c) Turn on the power to the AQM.
(d) Measure the flow rate of the air through the sensor module.
HINT: If you are measuring the flow rate through an Ozone module be aware that the Ozone module pump
cycles on and off about every thirty seconds. Wait for the Ozone module pump to turn ON before taking a
measurement.
(e) Verify the value is within specification.
(f) Record the measurement in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 63


GAS SENSOR MODULE – LEAK TEST THE GAS SENSOR MODULE (PUMPED TYPE)
Title: Leak test the gas sensor module (pumped type).
Item: Gas Sensor Module
Code: GSM002
Preparation: Rotameter.
Luer cap.
Specifications: N/A
Procedure:

(a) Disconnect the sample inlet connection from the Gas Sensor module.
(b) Connect the rotameter to the Gas Sensor module outlet.
HINT: Connect the air inlet side of the rotameter to the Gas Sensor module outlet.
HINT: Leave the air outlet side of the rotameter open to atmosphere.
(c) Turn on the power to the AQM.
(d) Block the inlet port of the Gas Sensor module using the Luer cap.
(e) Verify that the flow rate shown on the rotameter drops to zero.
HINT: This can take a number of minutes depending on the type of module.
HINT: If you are measuring the flow rate through an Ozone module be aware that the Ozone module pump
cycles on and off about every thirty seconds. Wait for the Ozone module pump to turn ON before taking a
measurement.
(f) Record your observations in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 64


GAS SENSOR MODULE – LEAK TEST THE GAS SENSOR MODULE (NON-PUMPED TYPE)
Title: Leak test the gas sensor module (non-pumped type).
Item: Gas Sensor Module
Code: GSM003
Preparation: Rotameter.
Pump.
Luer cap.
Specifications: N/A
Procedure:

(a) Disconnect the sample inlet connection from the Gas Sensor module.
(b) Connect the pump to the Gas Sensor module.
HINT: Connect the vacuum side of the pump to the Gas Sensor module outlet.
(c) Connect the rotameter to the pump.
HINT: Connect the pressure side of the pump to the air inlet side of the rotameter.
HINT: Leave the air outlet side of the rotameter open to atmosphere.
(d) Block the inlet port of the Gas Sensor module using the Luer cap.
(e) Verify that the flow rate shown on the rotameter drops to zero.
HINT: This can take a number of minutes depending on the type of module.
(f) Record your observations in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 65


GAS SENSOR MODULE – VERIFY THE GAS SENSOR MODULE PARAMETERS ARE AS
EXPECTED
Title: Verify the gas sensor module parameters are as expected.
Item: Gas Sensor Module
Code: GSM004
Preparation: Computer.
NOTE: this procedure can be performed remotely.
Specifications: N/A
Procedure:

(a) Open Aeroqual AQM software.


(b) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

HINT: You can do this remotely using a GSM modem by selecting File  Dial GSM Modem…

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 66


(c) Select Diagnostics  Sensor Module Settings…

(d) Enter the diagnostic password.


HINT: The factory set diagnostic password is george.

(e) Click on View All Settings. The settings will be downloaded from the controller.

(f) Verify the sensor settings against the settings recorded in the instrument log book.
HINT: Parameters may have changed during a user zero calibration. Verify that these are within the
acceptable range.
(g) Record your observations in the instrument Log Book

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 67


GAS SENSOR MODULE – REPLACE GAS SENSOR MODULE
Title: Replace gas sensor module.
Item: Gas Sensor Module
Code: GSM006
Preparation: Replacement gas sensor module.
Specifications: N/A
Procedure:

(a) Turn off the AQM.


(b) Disconnect the bus cable from the Gas Sensor module.
(c) Disconnect the sample inlet connection from the Gas Sensor module.
(d) Disconnect the air outlet from the Gas Sensor module.
(e) Remove the transit clip from the module.
(f) Locate the orange DIN rail clip and move the clip in the downward direction using a small screw driver.
(g) With the clip loosened, gently lift out the Gas Sensor module.
(h) Note the date and the serial number of the replacement Gas Sensor module in the instrument Log Book.
(i) Insert the new Gas Sensor module and locate it on the DIN rail.
(j) Fit the transit clip to the module.
(k) Connect the air outlet to the Gas Sensor module.
HINT: Ensure the tube fitting is correctly connected to prevent leakage.
(l) Connect the sample inlet connection to the Gas Sensor module.
HINT: Ensure the tube fitting is correctly connected to prevent leakage.
(m) Connect the bus cable to the Gas Sensor module.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 68


1.2.6 Particle Monitor Module

PARTICLE MONITOR MODULE – MEASURE THE FLOW RATE OF THE PARTICLE MONITOR
MODULE
Title: Measure the Flow Rate of the Particle Monitor Module.
Item: Particle Monitor Module
Code: PMM001
Preparation:
Specifications: Flow rate: 2.0 +/- 0.1 L/min
Procedure:

(a) Remove the TSP head and the sharp cut cyclone from the particle monitor.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 69


(b) Connect the rotameter to the particle monitor.
HINT: Leave the air inlet side of the rotameter open to atmosphere.
HINT: Connect the air outlet side of the rotameter to the External Inlet.

(c) Measure the flow rate of the air through the sensor module.
(d) Verify the value is within specification.
(e) Record the measurement in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 70


PARTICLE MONITOR MODULE – CLEAN THE PARTICLE MONITOR CYCLONE
Title: Clean the Particle Monitor Cyclone.
Item: Particle Monitor Module
Code: PMM002
Preparation:
Specifications: N/A
Procedure:

(a) Remove the TSP head and the sharp cut cyclone from the particle monitor.

(b) Open the sharp cut cyclone.


(c) With a cloth clean the cyclone.
(d) Close the sharp cut cyclone.
(e) Replace the sharp cut cyclone and the TSP head ensuring that there are no leaks.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 71


PARTICLE MONITOR MODULE – PERFORM A FIBRE SPAN CHECK
Title: Perform a Fibre Span Check.
Item: Particle Monitor Module
Code: PMM003
Preparation:
Specifications: Span measurement must be within 20% of the Fibre Span Reference
Reading (supplied on the calibration certificate).
Procedure:

(a) Open Aeroqual AQM software.


(b) Connect to the AQM by selecting File  Search Monitor…
HINT: This is achieved using your computers serial port.

(c) Open the real-time logging table using the Data  Table  Real Time… menu.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 72


(d) Start data logging using the File  Start Data Logging menu.

(e) Wait for approximately two minutes until you receive a reading.
HINT: All sensors should report a reading.
(f) Set the Particle Monitor Span Switch to ON.
(g) Wait for approximately four minutes until you receive two readings.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 73


(h) Verify that the Particle Monitor span reading is within the acceptable range.
HINT: The value of the span check reading is quoted in the calibration certificate supplied with the unit (Fibre
Span Reference Reading).
HINT: The span check value is affected by heat and movement of the fibre, thus it can only be used as
detection of major component failures such as the laser, photo detector or lens.
(i) Record your measurements in the instrument Log Book.

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 74


PARTICLE MONITOR MODULE – CHANGE THE PARTICLE MONITOR
Title: Change the Particle Monitor filters.
Item: Particle Monitor Module
Code: PMM004
Preparation:
Specifications: N/A
Procedure:

(a) Remove the filters from the Particle Monitor engine.


HINT: these are on the left hand side of the engine.

CONTINUED OVER PAGE

Aeroqual AQM 60 User Guide 75


(b) Change the filters and replace in the Particle Monitor engine.
HINT: The filter on the left in the photograph below goes into the filter location shown on the left in the
photograph (previous).

END OF SERVICE PROCEDURE

Aeroqual AQM 60 User Guide 76