Académique Documents
Professionnel Documents
Culture Documents
1 Introduction .............................................................................. 5
1.1 Important user information .................................................................................6
1.2 Regulatory information .........................................................................................7
1.3 WAVE system configurations .............................................................................9
1.4 WAVE Bioreactor 2/10 system ........................................................................ 11
1.5 WAVE Bioreactor 20/50 system ..................................................................... 12
1.6 Cellbag ........................................................................................................................ 13
1.7 WAVEPOD .................................................................................................................. 14
1.8 External instrument modules .......................................................................... 16
1.9 Control software .................................................................................................... 23
3 Installation.............................................................................. 33
3.1 Site requirements .................................................................................................. 33
3.2 Unpacking ................................................................................................................. 33
3.3 Installation of WAVE Bioreactor 2/10 .......................................................... 34
3.4 Installation of WAVE Bioreactor 20/50 ........................................................ 37
3.5 Installation of WAVEPOD .................................................................................... 41
3.6 Setting up external instrument modules .................................................... 45
3.7 Transport ................................................................................................................... 47
3.8 Spare parts and accessories ............................................................................ 47
4 Control systems..................................................................... 49
4.1 WAVE Bioreactor 2/10 control system ........................................................ 49
4.2 WAVE Bioreactor 20/50 control system ..................................................... 56
4.3 WAVEPOD control system ................................................................................. 66
4.4 External instrument module control systems ......................................... 83
5 Operation ................................................................................ 95
5.1 Operation overview .............................................................................................. 95
5.2 Preparations before starting a culture ........................................................ 96
5.3 Performing a run ..................................................................................................102
5.4 Procedures after a run ......................................................................................109
6 Maintenance......................................................................... 111
6.1 General .....................................................................................................................111
6.2 Safety switch inspection procedure ...........................................................112
6.3 Replacement of fuses ........................................................................................113
6.4 Calibration ..............................................................................................................114
1 Introduction
Purpose of the Operating
Instructions
The Operating Instructions provide you with the instructions needed to handle WAVE
Bioreactor 2/10 and 20/50 systems in a safe way.
Prerequisites
In order to operate WAVE Bioreactor 2/10 and 20/50 systems safely and according to
the intended purpose the following prerequisites must be met:
• You should be acquainted with the use of general laboratory equipment and with
handling of biological materials.
• You must read the Safety Instructions in Chapter 2 of these Operating Instructions.
• The system should be installed according to the instructions in Chapter 3 of these
Operating Instructions.
In this chapter
This chapter contains important user information and a general description of
WAVE Bioreactor 2/10, WAVE Bioreactor 20/50, WAVEPOD™, and external instrument
modules, and their intended use.
All users must read the Safety Instructions in Chapter 2 of these Operating Instructions
before installing, using or maintaining the system.
Do not operate WAVE Bioreactor 2/10 and 20/50 systems in any other way than
described in the user documentation. If you do, you may be exposed to hazards that
can lead to personal injury and you may cause damage to the equipment.
Intended use
The WAVE Bioreactor is a cell culture device. Culture medium and cells are loaded in to
a single use, presterilized bag known as a Cellbag™. The Cellbag is then placed on an
electric rocking base, the rocking unit.
WAVE Bioreactor 2/10 and 20/50 systems shall not be used in any clinical procedures,
or for diagnostic purposes.
Safety notices
These Operating Instructions contain WARNINGS, CAUTIONS and NOTICES concerning
the use of the product, with meanings as defined below.
WARNING
WARNING indicates a hazardous situation which, if not avoided, could
result in death or serious injury. It is important not to proceed until all
stated conditions are met and clearly understood.
CAUTION
CAUTION indicates a hazardous situation which, if not avoided, could
result in minor or moderate injury. It is important not to proceed until all
stated conditions are met and clearly understood.
NOTICE
NOTICE indicates instructions that must be followed to avoid damage to
the product or other equipment.
Tip: A tip contains useful information that can improve or optimize your procedures.
Typographical conventions
Software texts and commands are identified by bold italic text. A colon is used to
separate menu levels (e.g., File:Open refers to the Open option in the File menu).
Manufacturing information
Requirement Content
CE Conformity
Directive Title
International standards
CE marking
The CE marking and the corresponding Declaration of Conformity is valid for the
instrument when it is:
• used as a stand-alone unit, or
• used in the same state as it was delivered from GE Healthcare, except for
alterations described in the user documentation or explicitly authorized by
GE Healthcare.
Regulatory compliance of
connected equipment
Any equipment connected to WAVE Bioreactor 2/10 system and WAVE Bioreactor 20/
50 system should meet the safety requirements of EN 61010-1/IEC61010-1 or relevant
harmonized standards. Within the European Union, connected equipment must be
CE-marked.
CO2 concentration
Perfusion culture
O2 concentration
(measurement)
Temperature
Dissolved O2
pH control
External
Rocking
Bioreactor instrument
base unit module
BASE2/10EH + +
BASE2/10EH DOOPT20 + + +
BASE2/10EH CO2MIX20 + + +
BASE2/10EH O2MIX20 + + +
BASE2/10EH PERFCONT2E + + +
BASE20/50EHT + +
BASE20/50EHT-CO2 + + +
BASE20/50EHT-O2 + + +
BASE20/50EHT-L + + +
BASE20/50EHT DOOPT20 + + +
BASE20/50EHT CO2MIX20 + + +
BASE20/50EHT O2MIX20 + + +
• Cellbag holder
Illustration of
WAVE Bioreactor 2/10
The illustration below shows the main parts of the WAVE Bioreactor 2/10.
4
3
2
6
1
Part Description
3 LCD display
4 Rocking unit
5 Cellbag holder
• Cellbag holder
• WAVEPOD (optional)
Illustration of
WAVE Bioreactor 20/50
The illustration below shows the main parts of the WAVE Bioreactor 20/50.
Part Description
3 Rocking unit
1.6 Cellbag
Introduction
The Cellbag is a disposable bioreactor chamber. It is presterilized by gamma radiation
and is intended for single use. The Cellbag is partially filled with culture media,
inoculated, and placed on the bioreactor base unit.
Illustration
The illustration below shows the location of the main parts of the Cellbag.
" #
&
Part Description
1 Cellbag rod
2 Inoculation/harvest lines
5 Oxywell2™
1.7 WAVEPOD
Introduction
The WAVEPOD instrument module integrates the instrumentation associated with
WAVE Bioreactor 20/50. This includes pH, dissolved oxygen, and CO2/O2 gas mixing
controls. Up to four of instrument modules can be installed into WAVEPOD at the same
time to provide desired functionality. The WAVE Bioreactor can be operated from the
WAVEPOD Touch screen.
The WAVEPOD can be placed on either side of a WAVE Bioreactor and is available in
both left and right facing versions.
Module Description
Airflow controller Controls the flow of air through the Cellbag. The Airflow
controller is required if the CO2 or O2 controllers are
used
Illustration
The illustration below shows the location of the main parts of WAVEPOD.
3
5
2
Part Description
1 Bay 1 - pH module
2 Bay 2 - DO module
5 Touch screen
Name Function
3
4
5
3 Probe connector
1 2 3
4
8 5
7 6
5 CO2 ON switch
1 2
10 9 8 7 6 5
1 2
3
4
5
6
9 8
1 2
9 8 7 6 5
1
3
2 RUN switch
4
3
5
2
7 6
4 Process controller
2 Safety instructions
This chapter describes safety compliance, safety labels, general safety precautions,
emergency procedures, power failure and recycling of WAVE instruments.
• Personal protections
• System operation
• Maintenance
General precautions
WARNING
Do not operate the WAVE systems in any other way than described in
the WAVE user documentation.
WARNING
Operation and user maintenance of the WAVE system should be
performed by properly trained personnel only.
WARNING
Do not use any accessories not supplied or recommended by GE
Healthcare.
WARNING
Only WAVE approved Cellbags shall be used on the equipment.
WARNING
The WAVE Bioreactors are not designed to handle flammable fluids. The
WAVE Bioreactor are not approved for work in a potentially explosive
atmosphere.
Personal protection
WARNING
To avoid hazardous situations when working with the WAVE systems,
take the following measures for personal protection.
WARNING
Always use appropriate personal protective equipment during
operation and maintenance of WAVE systems.
WARNING
Hazardous substances. When using hazardous chemical and
biological agents, take all suitable protective measures, such as
wearing protective glasses and gloves resistant to the substances used.
Follow local and/or national regulations for safe operation and
maintenance of the system.
WARNING
Spread of biological agents. The operator has to take all necessary
actions to avoid spreading hazardous biological agents in the vicinity of
the instrument. The facility should comply with the national code of
practice for biosafety.
WARNING
Emergency stop. Position the WAVE system so that the power switch is
easily accessible for power shut off. The power switch is located at the
rear side of the instrument.
WARNING
Protective ground. The WAVE instruments must always be connected
to a grounded power outlet.
WARNING
Heavy object. Filled Cellbags have a considerable weight and heavy
lifts must be done with care. Use 1 person per multiple of 15 kg weight,
for example 3 persons for 30 to 45 kg. All lifting and moving must be
performed in accordance with local regulations.
CAUTION
Ensure that all tubing, hoses and cables are placed so that the risk or
tripping accidents is minimized.
CAUTION
Make sure that there is enough free space around the instrument for
the rocking motion.
CAUTION
The safety switches of the base unit must be functionally tested after
the installation of the equipment or after the instrument has been
transported and every 6 months from then on. See Chapter 6 for proper
method of testing.
System operation
WARNING
Biohazard. Make sure that the Cellbag is sealed before and during the
fermenting process.
WARNING
Electrical shock hazard after spillage. If there is a risk that large
volumes of spilled liquid may penetrate the casing of the WAVE
instrument, immediately switch off the instrument, disconnect the
power cord, and contact an authorized service engineer.
CAUTION
Pinch hazard. The rocking unit is a safety hazard. Remain clear of all
moving parts during operation. Do not work on the bag or rocking unit
unless the rocking motion is stopped.
CAUTION
Overheating is possible if the unit is operated without a liquid-
containing bag. The Cellbag holder temperature may exceed 60°C. Do
not touch.
CAUTION
Leakage risk of biological substances. Before every use, check all
hoses for signs of cracking or tears. None of the air hoses should
contain liquids of any kind.
CAUTION
Remove any spillage on the floor immediately to minimize the risk for
slipping accidents.
CAUTION
Pinch hazard when using PERFCONT. Do not run the feed and harvest
pump with the door open
Maintenance
WARNING
Electrical shock hazard. All repairs should be done by service
personnel authorized by GE Healthcare. Do not open any covers or
replace parts unless specifically stated in the user documentation.
WARNING
Disconnect power. Always disconnect power from the instrument
before performing any maintenance task.
WARNING
Always clean the equipment a well ventilated area. Never douse or
immerse any part of the unit with any liquid. When cleaning is required,
use only water and alcohol.
WARNING
Only spare parts that are approved or supplied by GE Healthcare may
be used for maintaining or servicing the system.
2.2 Labels
This section describes safety labels and labels concerning hazardous substances that
are attached to the WAVE instruments.
Label Description
Label Description
Emergency procedures
In an emergency situation, do as follows to stop the run:
Step Action
1 Switch off the power to the instrument by pressing the power switch to
the O position.
Power failure
In the event of power failure, the run is immediately interrupted. If the AUTOSTART
option is set to ON, the operation is automatically resumed on power up. For further
information, please refer to the user documentation.
Safety switches
Safety switches are located on top of the bioreactor base units. If any of the safety
switches is hit, the rocking motion stops and rocking unit moves to a level position. If a
safety switch has been hit, power must be switched off and then on again to reset the
safety switch.
Recycling of hazardous
substances
WAVE systems contain hazardous substances. Detailed information is available from
your local GE Healthcare representative.
Disposal of electrical
components
Waste of electrical and electronic equipment must not be disposed as unsorted
municipal waste and must be collected separately. Please contact an authorized
representative of the manufacturer for information concerning the decommissioning of
your equipment.
3 Installation
This chapter provides information regarding installation of WAVE Bioreactor 2/10,
WAVE Bioreactor 20/50, WAVEPOD, and external instrument modules.
For further information on installation, refer to the user documentation listed in
Section 8.2.
3.2 Unpacking
Unpack the equipment and place it on a stable surface. If applicable, adjust the leveling
feet.
Check the equipment for any apparent damage before starting installation. Document
any damage carefully and contact your GE Healthcare representative.
1 2 3 4 5 6 7
9 8
5 ALARM connector
Installing PERFCONT2E
perfusion module (optional)
An optional perfusion module is available for the WAVE Bioreactor 2/10. This modules
provides weight-based control of feed and harvest.
Follow the instruction below to install the perfusion module.
Step Action
1 Place the perfusion module on a stable surface. Adjust the levelling feet.
2 Place the WAVE Bioreactor 2/10 on top of the perfusion module. Make
sure all four feet are properly placed on the stainless-steel platform.
3 Connect the supplied PERFcable (DB9-F) to the WAVE Bioreactor 2/10 rear
panel port marked SCALE and to the rear of the perfusion module. Secure
the cable at both ends using the thumbscrews.
4 Connect tubing from the feed and harvest containers to the Cellbag.
Make sure the flow direction corresponds to the arrows marked on the
peristaltic pumps and that the tubing is properly pinched in the pump
rollers.
5 Open any clamps on the feed and harvest lines. Swing the clear covers to
the right on the pump override buttons, and press the buttons to test and
prime the pumps.
Step Action
2 Place the Cellbag holder on top of the stainless-steel rocker with the
heater cable towards the rear of the rocker.
3 Plug the blue heater cable into the rear panel connector labeled HEATER.
Twist the lock nut to secure.
4 Plug the metal end of the yellow temperature sensor cable into the
connector on the rear panel labeled TEMP. Twist the metal collar to lock
the plug. Plug the other end of the cable to the temperature sensor
attached to the Cellbag holder.
5 Check that the voltage selector is set to correct voltage. If not, pry out the
jumper using a screwdriver, adjust the arrow so that it is pointing to the
desired line voltage, and reinsert the jumper.
Step Action
6 Connect the supplied electrical cord to the power inlet and to a power
outlet.
7 Press the power switch on the rear panel to the I position to turn on the
instrument.
Result: The LCD screen lights up and the instrument starts. The instrument
tilts to its home position and when initialization is completed the main
screen is displayed.
Step Action
1 Connect the supplied air tube between the AIR OUT port located on the
left side of the bioreactor and the inlet filter of the Cellbag.
NOTICE
Pressure at the AIR IN port must not exceed 1 psig. Exceeding this limit
may cause rupture of the Cellbag. Provide a suitable safety pressure
relief device.
10
1 2 3 4 5 6 7 8 9
Installation of KIT20EHT(D) or
KIT50EHT(D)
KIT20EHT(D) or KIT50EHT(D) are used with the WAVE Bioreactor 20/50 rocking unit. Each
kit includes the hardware required to properly mount a Cellbag holder and a Heater pad
to control the operating temperature.
Follow the instruction below to install KIT20EHT(D) or KIT50EHT(D).
Step Action
1 Place the Top plate onto the bioreactor base unit and align the four
mounting holes. Make sure that the opening for cables is on the left when
facing the front of the bioreactor.
2 Fasten the Top plate by screwing the four mounting screws through the
mounting holes using an Allen key.
Step Action
1 Place the Heater pad on the Top plate. Make sure that the Heater pad is
oriented with the white side up and with the power connector fitting into
the slot on the left side of the Top plate.
2 Plug the blue heater cable into the rear panel connector labeled HEATER.
3 Place the stainless steel holder on top of the Heater pad and snap it in.
6 Check that the temperature display on the main screen reads the
ambient temperature.
Connecting power
Follow the instruction below to connect power to the bioreactor.
Step Action
1 Connect the supplied power cord to the power inlet and to a grounded
power outlet.
2 Press the power switch on the rear panel to the I position to turn on the
instrument.
Result: The Touch screen lights up and the instrument starts. The
instrument tilts to its home position and when initialization is completed
the main screen is displayed.
Step Action
1 Connect the supplied air tube between the Air out port located on the
side panel of the bioreactor and the inlet filter of the Cellbag.
Step Action
5 Change the date and time as needed. Press ↵ to accept each entry.
6 Press SET to accept the new date and time or BACK to abort.
Result: The main screen is displayed.
Setting up LCELL20/50EHT
optional weight sensor
The optional weight sensor is mounted under the Cellbag holder and provides the
weight of the Cellbag holder and the Cellbag.
To connect the LCELL20/50EHT to the WAVE Bioreactor 20/50, connect the yellow 4-pin
cable from the LCELL20/50EHT to the connector labeled LOAD CELL located on the rear
panel of the bioreactor.
Equipment required
The items listed below are recommended to facilitate the installation of WAVEPOD.
Some items are facility sources and other items can be purchased from GE Healthcare.
Facilities
• Oxygen (O2) supply source - pressure regulated to 0.7 to 1 bar (10 to 15 psig)
• Carbon dioxide (CO2) supply source - pressure regulated to 0.7 to 1 bar (10 to 15
psig)
• pH probe
• DOOPT probe
Connecting power
Follow the instruction below to connect power to WAVEPOD.
Step Action
Step Action
Step Action
Step Action
Step Action
2 Remove the DOOPT probe from its packaging. The probe can be left in its
protective tube until a calibration of the probe is needed.
3 Connect the DOOPT probe to the PROBE connector located on the front
panel of the monitor. Make sure that the ridge on the plug is lined up with
the notch on the connector. Push the plug in and twist to lock it in
position.
Step Action
3 Set the external CO2 source between 0.7 and 1 bar (10 and 15 psig).
Connect the external CO2 source to the CO2 IN port on the front panel.
4 If installing CO2MIX20-R, adjust the CO2 pressure to 0.4 bar using the
regulator knob and the and the gauge located on the front panel of the
instrument. The regulator knob must be pulled to out to unlock.
5 If ambient air is to be mixed with the CO2 stream, leave the AIR IN port
unconnected. Otherwise, connect the desired air mixture to the AIR IN
port. External gas pressure must be regulated to between 0.1 and 0.2 bar
(1 and 3 psig).
6 Connect the MIX OUT port to the inlet filter on the Cellbag using the
tubing provided. The CO2 /air mixture is pumped out from this port.
Step Action
2 Press the power switch to the on position. Allow five minutes for the
instrument to warm up and the reading to stabilize.
3 Set the external O2 source between 0.7 and 1 bar (10 and 15 psig).
Connect the external O2 source to the O2 IN port on the front panel.
5 If ambient air is to be mixed with the O2 stream, leave the AIR IN port
unconnected. Otherwise, connect the desired air mixture to the AIR IN
port. External gas pressure must be regulated to between 0.1 and 0.2 bar
(1 and 3 psig).
6 Connect the MIX OUT port to the inlet filter on the Cellbag using the
tubing provided. The O2 /air mixture is pumped out from this port.
Step Action
2 Connect the other end of the cable to the FEED PUMP or HARVEST PUMP
connector on the rear panel of the LOADCONT20 or
WAVE Bioreactor 20/50.
Standalone use
Step Action
1 Connect the power cord to the power connector located on the rear
panel of PUMP20 and to a grounded power outlet.
2 Connect the local shorting plug to the DB9 connector on the rear panel of
PUMP20. This will permit local operation.
Step Action
3 Connect the acid/base PUMP20 units to pH20 using the DB9 connectors
labeled ACID and BASE located on the rear panel of pH20 and the
supplied cables.
3.7 Transport
Follow the instruction below to move WAVE Bioreactor 2/10 or WAVE Bioreactor 20/50.
Step Action
3 Lift the instrument by the lifting handles. The instrument should be moved
by two persons.
4 Control systems
This chapter describes basic operation of embedded control systems in the WAVE
Bioreactor base unit and external instrument modules. For some functions there are
different control options. For detailed instructions see relevant user documentation
listed in Section 8.2 Literature‚ on page 142.
Item Function
SP Setpoint values
In this example 21/6º (rpm/rock angle), 0.20 (lpm) and 32.0 (degC)
ALARM! ALARM!
• A flashing ALARM! indicates a new alarm not yet
acknowledged.
• A steady ALARM! indicates an alarm that has been
acknowledged but not fixed.
• No ALARM! visible indicates that there is no current ALARM!
Rocking control
Step Action
1 Press ROCK button to start rocking. When rocking is on, the ROCK LED is
lit.
2 Press the ROCK up/down arrow buttons to change the rocking speed
(range 2 to 40 rpm). The selected rocking speed is shown on the setpoint
(SP) row.
Note: If rocking is obstructed by a hand or other object, the safety switches stop the
rocking instantaneously. Reset power to resume operation.
Aeration control
.
Step Action
1 Press AIR button to start aeration. When aeration is on, the AIR LED is lit.
2 Press the AIR up/down arrow buttons to change the air rate (range 0.01
to 0.50 lpm). The selected air rate is shown on the setpoint (SP) row.
3 Press AIR button to stop aeration if aeration is on.
Note: The air pump automatically shuts down if the outlet pressure exceeds 3 inch H2O
for more than 10 seconds. This is to prevent over pressurizing the bag in the event
of clogging. The air flow resumes when the pressure returns to normal.
Heating control
.
Step Action
1 Press TEMP button to start heating. When heating is on, the TEMP LED is
lit.
Step Action
Note: The temperature system automatically shuts down if the unit is not rocking. This
is to prevent hotspots due to stagnant liquid motion. The temperature control
resumes once the unit is rocking again.
Step Action
1 Press PERF button to start the perfusion system. When perfusion is on,
the PERF LED is lit.
shot The weight loss (or gain) in grams during each feed or
harvest operation.
SP Setpoint values
In this example 21/6º (rpm/rock angle), 0.20 (lpm) and 32.0
(degC)
4.1.4 Alarms
.
Step Action
1 A flashing ALARM! in the LCD display (see Fig 4-1) indicates a new alarm
not yet acknowledged.
2 Press ALM button to view alarm summary at any time. An example with a
temperature alarm (TMPDEV, temperature deviation) is shown.
4 To fix the Alarm, see Alarm messages‚ on page 120 for a list of alarm
messages and make appropriate corrective actions.
5 Press ALM button again to return to the previous view in the LCD display.
• Press SETUP Enter (↵) button to save a parameter value and open next menu.
• Press SETUP MENU ESC button while in a menu to return to the display main
screen.
Table 4-1. Operation parameters available using SETUP MENU ESC button.
SET ANGLE
- Change the rocking angle
(range 2° to 9°)
SET PERFUSION
- Enable/disable perfusion option
SET TARE
- Tare the perfusion load cell
CLEAR MEMORY
- Clear stored cumulative feed/harvest amounts
SET WEIGHT SP
- Set the weight control setpoint
CALIB TEMPERATURE
- Calibrate the temperature sensor
STOP POSITION
- Set the sample stop position (default 9º)
UNIT ADDRESS
- Set the Unit Address
(range 101 to 110)
AUTOSTART
-If the AUTOSTART option =1 controls are
automatically restored to the ON state on power
up after shutdown or power lost.
CAL LEVEL
- Set the calibration level. This allows adjusting the
level.
LEVEL@deg/10= 1 means 1 divide by 10 which
will give an 0.1 offset.
• The black boxes are function buttons. Pressing one of these will bring up the
corresponding detail screen.
• If alarm conditions exist, then the ALARM button will be displayed at the bottom of
the screen. If the ALARM is new and has not been acknowledged, then the ALARM
button will be flashing.
1 2 3 4
10
11
5 6 7 8 9
Part Description Part Description
Figure 4-2. Main screen example showing a Dual option EHTD system. Available functions varies
depending on system configuration
Each of the function has a screen which opens by pressing respective button. For details
of control system functions, refer to the WAVE Bioreactor 20/50 Operator Manual
Function Description
TREND Press the TREND button to get a graph for the parameter
over a 10 minute period.
ALARM See Section 4.2.8.
BACK Press the BACK button to return from current screen to the
main screen.
Step Action
3 Set the rocking speed set point by pressing the SP box. Enter desired set
point (2 to 40 rpm) in the data entry keypad that opens and press the ↵
key.
4 Set the rocking angle by pressing the angle box. Enter desired set point (2
to 12 degrees) in the data entry keypad that opens and press the ↵ key.
5 Press ROCK to start rocking and check that the unit rocks properly.
Note: When using a Dual system with a large single bag (20L, 22L or 50L) use Single
setup operation..
Step Action
2 The HEATER screen opens (the figure shows a Dual setup screen with two
sets of commands, LEFT and RIGHT, see Section 4.2.9 to change setup).
3 Set the temperature set point by pressing the SP box. Enter desired set
point (0°C to 45°C) in the data entry keypad that opens and press the ↵
key.
Step Action
Step Action
3 Set the net weight set point by pressing the SP box. Enter desired set
point (0.2 to 25 kg) in the data entry keypad that opens and press the ↵
key.
6 To set net weight to zero, press TARE. The TARE NOW? screen opens.
• Press TARE NOW to set the net weight to zero. This cannot be
undone.
• Press CANCEL to return to the WEIGTH screen.
Step Action
8 Press the RESET button to zero out both the feed and harvest cumulative
amounts.
• Set the total feed by day (L/d) pressing the FEED box. Enter desired
set point in the data entry keypad that opens and press the ↵ key.
• Set the feed volume per shot by pressing the Shot ml box. Enter
desired set point in the data entry keypad that opens and press the ↵
key.
• The period in minutes between each shot is calculated by values set
for total feed by day and feed volume per shot.
Step Action
11 Set the net weight set point by pressing the SP box. Enter desired set
point (0.2 to 25 kg) in the data entry keypad that opens and press the ↵
key.
Step Action
2 The AIRPUMP screen opens (the figure shows a Dual setup screen)
3 Set the aeration set point by pressing the desired SP box. Enter desired
set point (0 to 0.5 Lpm) in the data entry keypad that opens and press the
↵ key.
4 Press ON (left or right in Dual setup) to start the air pump.
5 Press STOP (appears in the same position as ON) to stop the air pump.
Air flow is required for the CO2 controller to work properly. For convenience, airflow
controls are provided on the right side of the CO2 screen. For information of controlling
the airflow, see Section 4.2.7
.
Step Action
3 Set the CO2 concentration set point by pressing the desired SP box. Enter
desired set point (0% to 15%) in the data entry keypad that opens and
press the ↵ key.
5 Press STOP (appears in the same position as ON) to stop the air pump.
Note: CO2 gas must be connected to the right side of the unit. The pressure must be
regulated to between 10 and 15 psig and capable of delivering at least 0.2 Lpm.
4.2.7 O2 concentration /
aeration control
Note: O2 concentration control requires BASE20/50EHTD-O2 instrument configuration).
Air flow is required for the O2 controller to work properly. For convenience, airflow
controls are provided on the right side of the O2 screen. For information of controlling
the airflow, see Section 4.2.7.
Step Action
3 Set the O2 concentration set point by pressing the desired SP box. Enter
desired set point (21% to 50%) in the data entry keypad that opens and
press the ↵ key.
5 Press STOP (appears in the same position as ON) to stop the air pump.
Note: O2 gas must be connected to the right side of the unit. The pressure must be
regulated to between 10 and 15 psig and capable of delivering at least 0.2 Lpm.
4.2.8 Alarms
An new alarm is indicated by:
• A flashing ALARM button appears on the bottom of the displayed screen.
• The screen color changes to red. This indicates that the alarm has not been
acknowledged.
• An audible beeper sounds. The beeper can be disabled, see WAVE Bioreactor 20/
50 Operator Manual.
To handle an alarm
All alarms have a specified code and are time-stamped to facilitate troubleshooting
(Alarm messages‚ on page 126).
Step Action
Step Action
2 The ALARMS screen opens. This action also acknowledges the alarm and
returns the screen color to green.
3 On the ALARMS screen each alarm is recorded. Each alarm message has:
• The date and time the alarm occurred or when it was cleared.
• Alarm status (Alm = not cleared, OK = cleared).
• The alarm code number, and short description, see Alarm messages‚
on page 126 (underscore between).
5 To access a help screen for alarms, press desired alarm to highlight it.
and press HELP.
6 To get a list of currently active alarms, press the Active Alarms button.
8 Once an alarm has been acknowledged, the ALARM button at the bottom
of the main screen stops flashing. It remains visible as long as any alarm
condition exists.
9 To view the alarm history and active alarm list, press the ALARM button
in the main screen (if visible) or accessed press MENU and then ALARM in
the MENU screen.
10 The alarm history is stored in the Touch screen. The alarm history is
maintained even if power is lost. The last 1000 alarms are retained. If the
Touch screen is interchanged between base units, the alarm history is
lost.
Step Action
11 Autoacknowlegde:
Normally the user must acknowledge any new alarm by pressing the
ALARM button to view the alarm history screen. If the AUTOACK feature
is set to ON on the SETUP:OPTIONS screen then all alarms are
automatically acknowledged. No warnings are generated but the alarm
is still logged.
12 Alarm contacts
If any alarm condition exists, this triggers the UNIT ALARM. This sets the
external alarm contact. The UNIT ALARM goes OFF only if no alarm
conditions are present.
Step Action
3 Press the ALARM to open the ALARMS screen described in Section 4.2.8.
Step Action
Click on the Enter Password box and enter password (default password
is 2050) followed by the ↵ key.
2
1
9
3 13
4 10 14
5
11 15
6
12 16
7 17
8 19
18
Part Description Part Description
Figure 4-3. WAVEPOD main screen. Green/red bars indicate that actual values are within/outside
setpoint value limits.
Function Description
TREND Press the TREND button to get a graph for the parameter
over a 10 minute period. WAVEPOD does not maintain long-
term trends.
A typical TREND screen is shown below.
BACK Press the BACK button to return to the main screen from
current screen.
4.3.2 pH control
..
Step Action
Step Action
5 Press ACID BASE/CO2 BASE control option selection to select mode for
pH control.
• ACID BASE option:
pH control by Acid/Base pumps.
• CO2 BASE option:
pH control by CO2 concentration and a Base pump.
Step Action
Step Action
3 DO Probe bar graph indicates the condition of the DO probe. When the
bar graph is green, the probe is functional. When the green bar reaches
the bottom, the graph turns red indicating the probe should be replaced.
Step Action
Step Action
2 The O2MIX CONTROL screen opens. Air flow is required for the O2
controller to work properly. Airflow controls are provided on the right side
of the O2MIX CONTROL screen (REMOTE BASE)
Step Action
Step Action
2 The CO2MIX CONTROL screen opens. Air flow is required for the CO2
controller to work properly. Airflow controls are provided on the right side
of the CO2MIX CONTROL screen (REMOTE BASE)
Step Action
5 Optional:
The CO2 setpoint can be set remotely by another controller (RSP, Remote
Set Point). When the option is active, a flashing message, RSP from pH, is
displayed.
6 Press the Aeration (MIX OUT) setpoint CO2 concentration SP display box.
• A data entry keypad opens.
• Enter a new setpoint (range 0 to 1.00 Lpm).
• Press the ENT key to accept the new setpoint, CLR to clear the
display and ESC to abort data entry.
4.3.8 Alarms
All alarms are time-stamped and have a short description to facilitate troubleshooting,
see (Alarm messages‚ on page 131) for a list of alarms.
The alarm beeper sounds every time a new alarm occurs. The beeper can be disabled
(ALM BEEPER OFF) from the SETUP OPTIONS screen.
Step Action
2 Press the ALARMS button to access the alarm history screen. This action
also acknowledges the alarm and makes the ALARMS button turn yellow
and stop flashing. The ALARMS button remains visible until all alarms
have cleared.
Step Action
Each alarm entry has the date and time, alarm number, description and
status (red=current unacknowledged alarms, blue=reviewed alarms,
green=cleared alarms).
The alarm history is retained on power failure or shutdown. The last 1000
alarms are retained.
• Use the up and down scroll arrows to page through the alarm log.
• The alarm history can be permanently cleared from the SETUP menu
(MENU:SETUP: CLEAR ALARM LOG).
• To review an alarm, press the alarm message. This brings up a pop-
up window with more information about the alarm. The alarm
message turns blue indicating that alarm has been viewed.
4 The ALARM LOG can be reviewed at any time by pressing the ALARMS
button from the main screen (Fig 4-3) or by pressing the REVIEW ALARMS
button in the SYSTEM INFO screen.
Step Action
Step Action
• Press the XXXX button and enter the password and press ENT.
• Press the PROCEED button which is displayed.
4 The SETUP screen opens. Options available in the SETUP screen depends
on the specific system configuration. For detailed description of setting
parameters, see the WAVEPOD Operator Manual.
5 Click OPTIONS button to set Bioreactor base unit type, autostart after
power failure (on/off), autoacknowledge of Alarms (on/off) and Alarm
beeper (on/off).
7 Click CLEAR ALARM LOG button to clear the stored alarm log. The alarm
log cannot be retrieved.
Step Action
9 Click pH button.
Step Action
Step Action
4.3.10 Help
Step Action
Figure 4-4. Large Process controller (left) and small Process controller (right).
NOTICE
Sensitive DO probe. Do not touch the tip of the probe with your finger or
any object. Do not attempt to clean or dry the tip with a cloth. Simply
shake dry
Step Action
1 Connect the probe to the connector on the front panel. Make sure that
the ridge on the plug is lined up with the notch on the connector. Push the
plug in and twist to lock it in position, do not force the plug.
2 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
Step Action
6 The Process control module show the dissolved oxygen value (red).
Note: Please allow the unit to scan the probe, be patient.
7 Calibrate the DO probe before use, see Section 6.4 for instructions
9 For optimum accuracy readjust the probe after insertion into the bag
(possible if the Cellbag has not been inoculated). With aeration and
rocking this should correspond to the 100% saturation point.
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
Step Action
3 Set the air flow setpoint by the up/down arrow buttons on the airflow
controller. The setpoint is shown in green. The actual airflow is shown in
the red. Control range is 0 to 0.5 liters/minute.
4 Air is drawn from the AIR IN inlet connector located on the front of the
instrument. Any special gas mixture can be connected here in place of
room air. External gas pressure must be regulated to between 0.1 and 0.2
bar (1 to 3 psig).
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
3 Set the air flow setpoint by the up/down arrow buttons on the airflow
controller. The setpoint is shown in green. The actual airflow is shown in
the red. Control range is 0 to 0.5 liters/minute.
4 Press the CO2 ON switch to switch on the CO2 controller. The switch lights
up indicating that CO2 control is active.
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
4 Air is drawn from the AIR IN inlet connector located on the front of the
instrument. Any special gas mixture can be connected here in place of
room air. External gas pressure must be regulated to between 0.1 and 0.2
bar (1 to 3 psig).
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
3 Press the CO2 button to switch on the CO2 controller. The button lights up
indicating that CO2 control is active.
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
Step Action
3 Set the air flow setpoint by the up/down arrow buttons on the airflow
controller. The setpoint is shown in green. The actual airflow is shown in
the red. Control range is 0 to 0.5 liters/minute.
4 Air is drawn from the AIR IN inlet connector located on the front of the
instrument. Any special gas mixture can be connected here in place of
room air. External gas pressure must be regulated to between 0.1 and 0.2
bar (1 to 3 psig).
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
3 Set the air flow setpoint by the up/down arrow buttons on the airflow
controller. The setpoint is shown in green. The actual airflow is shown in
the red. Control range is 0 to 0.5 liters/minute.
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module.
Step Action
1 Feed the tubing so that the inlet side is located at the lower clip and the
outlet in the upper clip. Rotate the pumphead and ensure the tubing is
securely held with no slack. The roller tension should be adjusted. Adjust
the clips so that the tubing is not pulled into the pump when the head
rotates.
Note: Pump always runs in a clockwise direction.
2 Press the PUMP ON switch to turn power on. The PUMP ON button is lit.
• If connected to a LOADCONT20 or BASE20/50EHT, the pump does not
run unless this device sends a REMOTE ON signal through the DB9
connecting cable. The pump runs at the speed set on the panel digital
indicator.
• If the shorting plug is installed, then the pump runs at the speed set
on the panel digital indicator.
3 Press the RUN button to run the pump at full speed as long as the button
is held down. PUMP ON switch indicator light must be on (green). This
function is useful for priming tubing.
5 The pump speed can be set 10% to 100% by pushing the +/- buttons on
the digital speed control.
General recommendations
For stable pH control it is necessary to cycle the addition pumps so that the acid or
base has sufficient time to mix properly. The pH20 controller has a cycle time of a
preset duration (default= 30 seconds). Part of this cycle is the on time when the acid
or base pump is energized. For the remainder of the cycle the pumps are off and
the controller waits until the cycle time is complete.
• If, at the end of the cycle, the pH is still not at the setpoint, the controller initiates
another add/wait cycle.
• If the pH is at the setpoint, the controller does nothing. A small deadband around
the setpoint (default = 0.1 pH) is used to prevent excessive pump chatter.
The cycle time can be adjusted if the system requires more mixing time. The
amount of acid or base added per cycle is proportional to the difference between
the setpoint and the actual pH. The Gain constant adjusts the sensitivity. This Gain
can be decreased if insufficient acid or base is being added per cycle. It can also be
increased if very concentrated acid or base is being used and significant overshoot
is observed.
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module. Make sure that the BASE ENBL button is off.
3 Make sure that the pump direction is correct and turn all local switches
on. Route the base tubing through the pump and connect to the Cellbag
container. Open all valves.
4 Set the desired pH setpoint on the controller and press the BASE ENBL
button to activate base control.
5 The base pump turns on whenever the pH drops 0.1 unit below the pH
setpoint.
Step Action
1 Turn the power switch on (I). The power switch is located on the back of
the instrument module. Make sure that the ACID ENBL button is off.
3 Make sure that the pump direction is correct and turn all local switches
on. Route the base tubing through the pump and connect to the Cellbag
container. Open all valves.
Step Action
4 Set the desired pH setpoint on the controller and press the ACID ENBL
button to activate base control.
5 The acid pump turns on whenever the pH exceeds 0.1 unit below the pH
setpoint.
Step Action
2 Press the up/down arrow buttons to change the setpoint. The controller
responds once the button is released.
Step Action
1 Press the Process controller Advance button Tto access the menu.
Press theT button repeatedly until the desired parameter (pB1 or pB2)
is displayed on the green LED. Increase the pB values to reduce the
amount of acid/base added per shot.
• Decrease the pB values to increase the amount of acid/base per
shot.
2 Press the up/down arrow buttons to change the displayed gain. Press the
α button to return to normal operation.
can be set independently. Ct1 is the base cycle time and Ct2 is the acid cycle time.
The cycle time is in seconds (default=30.0 seconds).
Step Action
1 Press the Process controller Advance button Tto access the menu.
Press the T button repeatedly until the desired parameter (Ct1 or Ct2)
is displayed on the green LED.
2 Press the up/down arrow buttons to change the displayed cycle time.
Press the ∞ button to return to normal operation.
Step Action
Pins 1 and 9 on the DB9 connector are the emitter and collector of an open
collector output respectably. Whenever the controller needs a pump to run, the
output is turned on. The output load current must not exceed 100mA.
User supplied pumps can be also connected to the pH20 acid/base external
instrument module by using the optional Power relay interface.
5 Operation
This chapter describes basic operation of WAVE Bioreactor 2/10 and WAVE Bioreactor
20/50 systems. For detailed instructions see relevant user documentation listed in
Section 8.2 Literature‚ on page 142.
Workflow
1 Preparations before start
7 Start culturing
Step Action
1 Make sure that any desired external instrument units are correctly
installed.
2 Turn the Bioreactor base unit power switch on (I). It is located at the rear
of the instrument.
3 Optional:
Turn desired installed external instrument units power switches on (I).
Note: If the safety switches have been hit, power must be turned off (O)
and on (I) to reset the safety switch.
5 After initialization, the main screen is displayed on the LCD screen (WAVE
Bioreactor 2/10) or Touch screen (WAVE Bioreactor 20/50).
Step Action
Step Action
Step Action
3 Assemble Cellbag(s)
For a description of Cellbags with inlets and valves, see Section 1.6.
• Select Cellbag size depending on experiment demands and
instrument setting. For Dual operation two Cellbags are used. See
Table 5-1.
• Remove the selected disposable Cellbag(s) from the protective plastic
over wrap bags.
For WAVE Bioreactor 20/50:
• Open the guides on the
Cellbag holder by lifting the
cam levers located on either
side of the Cellbag holder.
• Place the Cellbag into the
Cellbag holder and push each
of the two plastic rods into the
corresponding open guide.
Then press the cam levers
down to lock the Cellbag into
the Cellbag holder.
• Place the Cellbag on the
Cellbag holder with the
sampling ports towards the
front.
For WAVE Bioreactor 2/10:
• Lay Cellbag on the Cellbag holder and push the rod on each end into
the gray plastic clips located on either side.
The rods should snap into the clips and secure the Cellbag firmly in
the Cellbag holder.
Note: WAVE Bioreactor 20/50 is equipped with barcode reader. A warning message is
generated if no bag is detected or if the Cellbag barcode is not valid.
Using any Cellbag that does not have a valid barcode is not permitted on WAVE
equipment and will void any warranty or service claims.
Step Action
1 Connect the aeration air line from the Air out port located on the left side
of the instrument (dual mode systems has an additional Air out port
located on the right side of the instrument) to the inlet filter on the
Cellbag (Section 1.6).
3 Make sure both inlet and outlet filter tubing clamps are open.
Step Action
3 Plug the filter heater cable into the heater connector(s) located on the
rear panel.
Operating conditions
recommendations
Each cell line and medium requires some optimization of operating conditions. The
rocking rate should be set at the minimum rate that provides mixing and oxygen
transfer without excessive foaming, see Table 5-2 for typical rocking and aerating rates.
Table 5-2. Typical operating conditions (WAVE Bioreactor 20/50 system).
100 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Operation 5
Preparations before starting a culture 5.2
• Any rocking rate higher than 6 rocks per minute is sufficient for particle
suspension and bulk mixing.
• Measuring dissolved oxygen is the best way to determine the required rocking
rate. This may be done by off-line or on-line measurement. See Section 5.3.4.
• For very foamy medium, the rock angle should be reduced to 4 to 5 degrees.
• For cells with high oxygen demand, such as insect cells, the rock angle can be
increased to 10 degrees.
• If extensive amount of foam is generated reduce the rocking speed. Note however
that is critical that the rocking speed be sufficient to generate a visible surface
wave.
• It is usual practice to add 0.01 mg/liter of pluronic F-68 to the culture medium to
minimize protein degradation associated with the effects of foaming. Most
commercial cell culture media already contains pluronic.
Excessive foaming also occurs if the Cellbag is not rigidly inflated. Check that you have
sufficient airflow and that the pressure relief valve is functioning
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 101
5 Operation
5.3 Performing a run
Step Action
1 Calibrate pH and DO probes before each run, see Section 6.4 Calibration‚
on page 114.
3 Verify that the Cellbag is firmly inflated and secured to the Cellbag holder.
4 Verify that the Cellbag holder is firmly latched on the Bioreactor base unit
Top plate. The Cellbag should be taut, but not creased.
5 Verify that air is released through the outlet pressure relief vent by
pressing down gently on the chamber and observing a release of air
through the outlet pressure relief valve.
For instructions to monitor and set operating conditions for the current system
configuration, see relevant parts of Chapter 4 Control systems‚ on page 49.
Note: Continue airflow to the headspace to keep the Cellbag rigidly inflated. This is
important otherwise excessive foam is generated.
Step Action
1 Ensure that the Cellbag is inflated prior to filling with medium in order to
reduce foaming.
102 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Operation 5
Performing a run 5.3
Step Action
6 Adjust the rocking speed if required by changing the set point value.
• For rocking speed recommendations, see Operating conditions
recommendations‚ on page 100.
• Increase the speed if no wave is visible.
Reduce the speed if there is excessive foaming.
Step Action
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 103
5 Operation
5.3 Performing a run
Note: Issues with the systems or set culturing conditions are acknowledged by an
alarm. An new alarm is indicated by a flashing ALARM button in the main screen
of the Touch screen. The Touch screen color also changes to red and an audible
beeper sounds. See Chapter 4 Control systems‚ on page 49 for handling Alarms.
Step Action
1 Make sure that desired operating conditions are set, see Operating
conditions recommendations‚ on page 100.
• It is possible to adjust rocking rate, rocking angle, aeration, and
temperature,
• For some system configurations it is also possible to monitor and/or
control dissolved oxygen, pH, CO2 concentration, O2 concentration
and conditions for performing perfusion culturing.
Step Action
104 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Operation 5
Performing a run 5.3
Step Action
5 Release the tubing clamp and withdraw a sample into the syringe. You
may need to push down on the Cellbag to force the liquid up the sample
tube.
6 Remove the syringe and wipe the top of the sampling connector again
with 70% ethanol and replace the dust cap.
7 Pinch the sampling connector tubing a few times to ensure that any
liquid in the tubing drains back into the Cellbag.
Note: The sampling device has a fairly small orifice. If you are working with large
microcarriers or large cell aggregates do not use the sampling connector.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 105
5 Operation
5.3 Performing a run
The WAVE system is designed to provide an excess of oxygen for most culture systems.
To ensure desired levels of dissolved oxygen it is important to measure the levels of
dissolved oxygen and adjust the dissolved oxygen concentration if required.
Step Action
Note: Dissolved CO2 concentrations are influenced in the inverse manner as dissolved
oxygen.
pH control
pH control can be achieved by:
• use of pH control module (WAVEPOD or standalone unit), which enables on-line
control of pH in the Cellbag.
106 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Operation 5
Performing a run 5.3
Scaling up culturing
WARNING
Heavy object. Filled Cellbags have a considerable weight and heavy
lifts must be done with care. Use 1 person per multiple of 15 kg weight,
for example 3 persons for 30 to 45 kg.
All lifting and moving must be performed in accordance with local
regulations.
Cellbags have a large range in operating volume. This allows an inoculum scaleup and
eliminates tedious sequential transfers. Start at low volume and add fresh medium to
the Cellbag as the cells grow. Up to a 1:10 expansion is possible in a single chamber.
A typical inoculum sequence is:
Step Action
In this sequence, only one transfer was made in going from 100 mL to 10 liters of
culture. This saves time and reduces the risk of contamination. This scheme can also be
used to scale-up inoculum for larger conventional bioreactors.
For instructions to monitor and set operating conditions for the current system
configuration, see relevant parts of Chapter 4 Control systems‚ on page 49.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 107
5 Operation
5.3 Performing a run
Note: To avoid possible oxygen depletion, the culture medium exchange operation
should be performed in less than one hour.
Step Action
1 Stop the rocking and aeration and clamp off the inlet and outlet filters.
2 Remove the Cellbag holder from the rocking unit and place it in the
vertical position resting against a support. Allow the cells or microcarriers
10 to 15 minutes to settle.
3 Connect tubing to the harvest line on the Cellbag. The other end of this
tubing should be connected to a sterile collection vessel.
7 Open the inlet and outlet filter clamps and restart aeration and rocking.
Perfusion culture
During perfusion culture cell-free harvest can be withdrawn and fresh medium can
be added continuously.
Requirements:
For instructions to set perfusion culture operating conditions for the current
system configuration, see relevant parts of Chapter 4 Control systems‚ on page 49,
for control system instructions.
108 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Operation 5
Procedures after a run 5.4
For instructions to monitor and set operating conditions for the current system
configuration, see relevant parts of Chapter 4 Control systems‚ on page 49.
Note: The system has minimal turnaround time. Simply remove the completed batch
and place a new Cellbag holder and Cellbag on the instrument.
Step Action
1 Stop the rocking, aeration and other control and monitoring systems.
4 Disconnect any other tubes and sensors still connected to the Cellbag.
6 Remove the Cellbag holder containing the Cellbag from the base unit for
processing at any suitable location.
Step Action
1 Turn the power switches of all used units off (O). Power switches are
generally located at the rear of respective unit.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 109
5 Operation
5.4 Procedures after a run
110 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Maintenance 6
General 6.1
6 Maintenance
This chapter provides instructions for routine maintenance.
6.1 General
Regular maintenance of the WAVE systems is essential for reliable results.
WARNING
Electrical shock hazard. All repairs should be done by service
personnel authorized by GE Healthcare. Do not open any covers or
replace parts unless specifically stated in the user documentation.
WARNING
Disconnect power. Always disconnect power from the instrument
before performing any maintenance task.
NOTICE
Cleaning. Keep the instruments dry and clean. The instruments must be
turned off and unplugged before cleaning begins. Clean the exterior of
the instruments with a damp cloth with water and, if required, alcohol.
Do not use abrasive cleaners. Water should not be applied directly to the
instruments. Make sure that the instruments are completely dry before
plugging them in.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 111
6 Maintenance
6.2 Safety switch inspection procedure
Instruction
Step Action
2 Set the instrument rocking speed to 10 rpm. Set the angle to 12 degrees
and turn rocker function on.
3 While the instrument is rocking, place one end of the 300 mm long piece
of tubing between the safety switch and the top rocker plate at front of
the instrument.
Result: The instrument should lightly squeeze the tubing, return to level
and then stop rocking. Within 30 seconds the background lighting of the
Touch screen should turn red and flash. The following message should be
displayed: “SAFETY BUMPER SWITCH HIT” “CYCLE POWER TO RESET”.
Note: If the instrument reacts differently than described, stop the
inspection and contact Technical support. Do not use the
equipment as the safety switches may not be working properly.
112 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Maintenance 6
Replacement of fuses 6.3
WARNING
Disconnect power. Always disconnect power from the instrument
before replacing fuses.
WARNING
If a fuse requires repeated replacement, do not continue to use the
instrument. Contact an authorized service engineer.
Refer to Section 8.1 Specifications‚ on page 139 for information about the fuse types
and rating.
WARNING
For continued protection from fire hazard, replace only with same type
and rating of fuse.
Instruction
Follow the instruction below to replace fuses in WAVE Bioreactor 2/10, WAVE Bioreactor
20/50 and WAVEPOD.
Step Action
2 Pry off the fuse cover from the power entry module.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 113
6 Maintenance
6.4 Calibration
6.4 Calibration
Before starting culturing, the pH probe and the DOOPT probe should be calibrated.
For further information on calibration of instrument modules, refer to WAVE Bioreactor
System 20/50EHT Operator Manual, WAVEPOD Operator Manual and WAVE Bioreactor
Instrumentation Manual.
pH calibration
Follow the instruction below to perform a single point calibration of the pH probe using
either WAVEPOD or pH20.
Step Actions
DO calibration
Calibration of the DOOPT probe requires a two-point method. Perform zero oxygen
calibration and 100% air saturation calibration according to the instructions below.
Preparations
Follow the instruction below to prepare for DO calibration.
Step Actions
114 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Maintenance 6
Calibration 6.4
Step Actions
2 Press the CALIB button. In the pop-up window, press the Set 0% button.
Result: A pop-up window displays SETTING 0%. When the pop-up
window closes the 0% oxygen point is set.
If the screen displays CAL FAIL, re-calibration is needed.
Follow the instruction below to perform a 100% air saturation calibration using
WAVEPOD.
Step Actions
1 Remove the probe from the zero oxygen solution. Rinse the probe in
water and shake dry. Allow the probe to equilibrate in air. Do not touch
the probe tip or attempt to dry it with a cloth.
2 When the reading is stable (within ±2%) in 1 to 2 minutes, press the CALIB
button. In the pop-up window, press the Set 0% button.
Result: A pop-up window displays SETTING 100%. When the pop-up
window closes the 100% oxygen point is set.
If the screen displays CAL FAIL, re-calibration is needed.
Step Actions
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 115
6 Maintenance
6.4 Calibration
Step Actions
5 When the phase angle reading is stable (within ±0.25ª), press the PUSH
FOR MENU button.
Result: The display shows TRY0 0 (no of scans), 000.00% (actual reading).
If the display shows NEED 0% CAL, re-calibration is needed.
Follow the instruction below to perform a 100% air saturation calibration using
DOOPT20.
Step Actions
1 Remove the probe from the zero oxygen solution. Rinse the probe in
water and shake dry. Allow the probe to equilibrate in air. Do not touch
the probe tip or attempt to dry it with a cloth.
2 When the reading is stable (within ±2%) in 1 to 2 minutes, press the PUSH
FOR MENU button.
Result: The display shows --MENU-- 1.CALIB.
6 When the phase angle reading is stable (within ±0.25ª), press the PUSH
FOR MENU button.
Result: The display shows TRY0 0 (no of scans), 100.00% (actual reading).
If the display shows NEED 100% CAL, re-calibration is needed.
116 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
WAVE Bioreactor 2/10 7.1
7 Troubleshooting
The Cellbag appears to Too high airflow Check that the airflow to the
be overinflated. bioreactor does not exceed
0.5 Lpm.
The Cellbag should be
inflated so that it Faulty relief valve Check that air is passing out of
requires force to the pressure relief valve: attach a
squeeze it. However, it short length of tubing to the
should not be so exhaust vent and immerse the
pressurized that tubing in water to a 1 cm depth.
creases form near the Bubbles should appear
attachment points. indicating flow. If no flow is
observed, remove the pressure
relief valve. The outlet filter may
be plugged and removing the
pressure relief valve may allow
continued operation.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 117
7 Troubleshooting
7.1 WAVE Bioreactor 2/10
The Cellbag appears to Too low airflow Check that there is sufficient air
be underinflated. flow to the inlet filter.
An underinflated Inlet air supply Check that you have connected
Cellbag will generate wrongly connected the inlet air supply line to the
excessive foam and inlet filter (does not have the
poor mixing. pressure relief valve).
118 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
WAVE Bioreactor 2/10 7.1
The rocking unit does Faulty safety Do not operate the instrument in
not initialize properly switches this condition. Contact Technical
support.
The error message The heater is Check that the heater cord is
HTROUT is displayed disconnected from firmly connected.
the WAVE Bioreactor
base unit
Heating is not working Faulty rocking unit Check that the rocking unit is
and no error message rocking. The heater is
is displayed automatically turned off when
the rocking unit is not rocking.
Starting the rocking will
automatically switch the heater
on if temperature control is
enabled.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 119
7 Troubleshooting
7.1 WAVE Bioreactor 2/10
Alarm messages
Message Description
NOFEED If the shot size is not fed after 20 minutes the alarm is triggered
and the perfusion is turned off.
NOHARV If the shot size is not harvested after 20 minutes the alarm is
triggered and the perfusion is turned off.
120 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
WAVE Bioreactor 20/50 7.2
The Cellbag appears to Too high airflow Check that the airflow to the
be overinflated. bioreactor does not exceed 0.5
Lpm.
The Cellbag should be
inflated so that it Faulty relief valve Check that air is passing out of
requires force to the pressure relief valve: attach
squeeze it. However, it a short length of tubing to the
should not be so exhaust vent and immerse the
pressurized that tubing in water to a 1 cm depth.
creases form near the Bubbles should appear
attachment points. indicating flow. If no flow is
observed, remove the pressure
relief valve. The outlet filter may
be plugged and removing the
pressure relief valve may allow
continued operation.
The Cellbag appears to Too low airflow Check that there is sufficient air
be underinflated. flow to the inlet filter.
An underinflated Inlet air supply Check that you have connected
Cellbag will generate wrongly connected the inlet air supply line to the
excessive foam and inlet filter (does not have the
poor mixing. pressure relief valve).
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 121
7 Troubleshooting
7.2 WAVE Bioreactor 20/50
Temperature control
122 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
WAVE Bioreactor 20/50 7.2
Slow heating The heater cable is Check that the heater cord is
not connected firmly connected.
Remember, the heater
is intended to control The heater is not • Turn the heater on.
temperature using turned on
gentle heating. It takes • Check that the HTR ON
over 1 hour to heat 10 indicator is flashing.
liters from room • Check that the heater power
temperature to 37°C. bars are increasing or at
maximum.
• When using 20 L or larger
Cellbags, make sure that
both heaters are running
(LEFT+RGHT setting) on
HEATER control screen.
The temperature The Cellbag does not Make sure that the temperature
control is not cover the sensor is properly positioned
functioning or the temperature sensor under the Cellbag.
displayed temperature
The temperature Check that the heater is correctly
appears to be
sensor is not positioned and that the Cellbag
incorrect.
properly installed holder is latched securely.
The HTR FAIL alarm is The temperature WARNING! Do not touch the
activated exceeds 60ªC Cellbag holder.
Check that the sensor is
correctly installed and that a
Cellbag is present. In the event of
an overheat condition, the
system will reset automatically
once the heater cools down
sufficiently.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 123
7 Troubleshooting
7.2 WAVE Bioreactor 20/50
Rocking unit
Rocking unit stops The rocking unit is Clear the obstruction and turn
rocking and the mechanically the power off and on to resume
SAFETY SWITCH restricted from the rocking.
TRIPPED alarm is moving
activated
Aeration controller
No airflow, and a high The airpump has Disconnect the Air in and Air out
pressure alarm is shut down due to tubing and locate the blockage.
activated high pressure
The airpump is running Blocked air tubing Disconnect the inlet and outlet
but the airflow shows connections on the side panels. If
zero the airflow is restored then check
if the Air in or Air out tubing is
blocked.
The airflow fluctuates, Restriction in the air Disconnect the inlet and outlet
and a high pressure tubing connections on the side panels. If
alarm is activated. the airflow is restored then check
for restrictions in the Air in or Air
Airflow should stay
out tubing.
constant within ± 0.02
Lpm. Clogged Cellbag Transfer the contents to another
exhaust vent filter Cellbag.
124 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
WAVE Bioreactor 20/50 7.2
Weight controller
CO2 controller
Erratic CO2 control CO2 gas not properly Check that the CO2 gas to the
connected unit is connected and at the
correct pressure.
O2 controller
Erratic O2 control O2 gas not properly Check that the O2 gas to the unit
connected is connected and at the correct
pressure.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 125
7 Troubleshooting
7.2 WAVE Bioreactor 20/50
A message is displayed Faulty bar code Make sure that the Cellbag is
stating that no Cellbag reader valid and correctly installed.
is detected, that the If the problem persists, contact
Cellbag is incorrectly GE Healthcare Technical
installed in the holder, support.
or that the Cellbag bar
code is not valid.
Alarm messages
The table below lists the alarm messages that can appear on the
WAVE Bioreactor 20/50 Touch screen.
126 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
WAVE Bioreactor 20/50 7.2
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 127
7 Troubleshooting
7.3 WAVEPOD
7.3 WAVEPOD
General problems
The table below gives a brief guide of how to solve general problems that might occur
when using WAVEPOD.
0 ADC overflow
4 Low amplitude
128 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
WAVEPOD 7.3
The table below gives a brief guide of how to solve problems that might occur when
using the DO (Dissolved Oxygen) controller.
Low amplitude alarm Probe not properly Check that the probe is
connected or faulty connected correctly and make
probe sure that the plug is secure. Cycle
power to the instrument to reset.
On the WAVEPOD Touch screen
overview screen, press DO OPT
to display the DETAIL screen for
the display for the DO controller.
Check the signal amplitude. Less
than 2000 indicate a bad probe.
If the error persists, replace the
probe.
CO2 MIX
The table below gives a brief guide of how to solve problems that might occur when
using the CO2 controller.
The display shows a A minor deviation If the reading is not within the
CO2 concentration in from the expected tolerance, the CO2 offset can be
air which deviates from value is not unusual. adjusted, see WAVEPOD
the expected value The tolerance of the Operator Manual.
(0.0%). CO2 sensor is
±0.38% in the range
0% to 7.5% CO2.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 129
7 Troubleshooting
7.3 WAVEPOD
CO2 reading keeps The CO2 gas supply Check that the green CO2 inlet
drifting down pressure is too low pressure light is lit. The CO2
supply pressure must between
10 and 15 psig.
Faulty CO2 control The CO2 supply is not Verify that CO2 is connected to
properly connected the correct inlet port. Check that
the air pump is on and that the
flow rate is between 0.1 and 0.5
liters/minute. Check that the CO2
switch is on.
The Cellbag does not No airflow Verify that there is airflow to the
stay inflated Cellbag.
130 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
WAVEPOD 7.3
O2 MIX
The table below gives a brief guide of how to solve problems that might occur when
using the O2 controller.
The display shows a O2 A minor deviation If the reading is not within the
concentration in air from the expected tolerance, the O2 offset can be
which deviates from value is not unusual. adjusted, see WAVEPOD
the expected value The tolerance of the Operator Manual.
(21.0%). O2 sensor is ±0.6% in
the range 21% to
40% O2.
O2 reading keeps The O2 gas supply Check that the green O2 inlet
drifting down pressure is too low pressure light is lit. The O2 supply
pressure must between 10 and
15 psig.
Alarm messages
The table below lists the alarm messages that can appear on the WAVEPOD Touch
screen.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 131
7 Troubleshooting
7.3 WAVEPOD
132 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
External instrument modules 7.4
Not getting 0% Probes not properly Check that the DOOPT probe is
connected disconnected and that the
temperature probe is connected
(red LED is not lit).
Not getting 100% Faulty DOOPT probe Check that the temperature
probe is connected (red LED is
not lit). Check that zero is
obtained when the DOOPT probe
is disconnected. If it is still not
possible to get zero with the
SPAN knob, then replace the
DOOPT probe.
Low amplitude alarm The probe is not Check that the probe is
properly connected connected correctly and make
sure that the plug is secure. Cycle
power to the instrument to reset.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 133
7 Troubleshooting
7.4 External instrument modules
NEED 0% CAL or NEED The probe requires Calibrate the probe. If the
100% CAL is displayed calibration problem persists, replace the
probe.
The display shows a A minor deviation If the reading is not within the
CO2 concentration in from the expected tolerance, the CO2 offset can be
air which deviates from value is not unusual. adjusted, see WAVE Bioreactor
the expected value The tolerance of the Instrumentation Manual.
(0.0%). CO2 sensor is
±0.38% in the range
0% to 7.5% CO2.
CO2 reading keeps The CO2 gas supply Check that the green CO2 inlet
drifting down pressure is too low pressure light is lit. The CO2
supply pressure must between
10 and 15 psig.
134 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
External instrument modules 7.4
Faulty CO2 control The CO2 supply is not Verify that CO2 is connected to
properly connected the correct inlet port. Check that
the air pump is on and that the
flow rate is between 0.1 and 0.5
liters/minute. Check that the CO2
switch is on.
The Cellbag does not No air flow Verify that there is airflow to the
stay inflated Cellbag.
The display shows a O2 A minor deviation If the reading is not within the
concentration in air from the expected tolerance, the O2 offset can be
which deviates from value is not unusual. adjusted, see WAVE Bioreactor
the expected value The tolerance of the Instrumentation Manual.
(21.0%). O2 sensor is ±0.6% in
the range 21% to
40% O2.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 135
7 Troubleshooting
7.4 External instrument modules
O2 reading keeps The O2 gas supply Check that the green O2 inlet
drifting down pressure is too low pressure light is lit. The O2 supply
pressure must between 10 and
15 psig.
Faulty O2 control The CO2 supply is not Verify that O2 is connected to the
properly connected correct inlet port. Check that the
O2 supply pressure is set to 0.7
bar minimum. Check that the air
pump is on and that the flow rate
is between 0.1 and 0.5 liters/
minute. Check that the O2 switch
is on.
PUMP20
The PUMP ON button The pump is not • Make sure that the PUMP20
on the front panel of powered is plugged in.
PUMP20 is not lit
• If using a WAVE Bioreactor
20/50 for power, check that
the bioreactor is turned on.
• For standalone use, make
sure that PUMP20 is properly
connected to an external
power supply.
PUMP20 does not run A shorting plug is not For standalone use, make sure
connected that a shorting plug is connected
to the DB9 connector on the rear
panel of PUMP20
136 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Troubleshooting 7
External instrument modules 7.4
The pump speed does The speed settings • Check the speed settings on
not change are not correct the front panel of the
PUMP20.
• Press the RUN button to run
the pump all full speed
regardless of the speed
setting.
pH20
The message Err is The pH probe is not Make sure the pH probe is
displayed on the connected properly connected.
screen
The pH probe is Replace the pH electrode if it is
faulty damaged.
The acid pump does The pump is not Make sure that the switches on
not run turned on the pump are on and that the
pump is powered.
The acid pump is not Make sure that the ACID ENBL
enabled to run switch on pH20 is pressed and lit
automatically up.
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 137
7 Troubleshooting
7.4 External instrument modules
The base pump does The pump is not Make sure that the switches on
not run turned on the pump are on and that the
pump is powered.
pH overshoots The gain value is too Adjust the gain to a lower value.
high
Constant pH is not Obstructed tubing Check that the acid pump tubing
maintained is open.
138 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Reference information 8
Specifications 8.1
8 Reference information
This chapter contains technical data, regulatory and other information.
8.1 Specifications
WAVE Bioreactor 2/10
Parameter Value
Ingression protection IP 20
Power 145/210 VA
Maximum current 4A
Weight 4.2 kg
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 139
8 Reference information
8.1 Specifications
Parameter Value
Ingression protection IP 20
Power 630 VA
Weight 16 kg
140 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Reference information 8
Specifications 8.1
WAVEPOD
Parameter Value
Power 200 VA
Weight 8.6 kg
DOOPT20
Parameter Value
Weight 1.5 kg
Parameter Value
Weight 2.7 kg
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 141
8 Reference information
8.2 Literature
Parameter Value
Weight 2.7 kg
PUMP20
Parameter Value
Weight 1.0 kg
pH20
Parameter Value
Weight 1.5 kg
8.2 Literature
For further information regarding WAVE Bioreactor 2/10, WAVE Bioreactor 20/50,
WAVEPOD, and the external instrument modules, refer to the following:
• WAVE Bioreactor System BASE 2/10 EH Operator Manual
142 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
Reference information 8
Ordering information 8.3
WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA 143
8 Reference information
8.3 Ordering information
144 WAVE Bioreactor 2/10 and 20/50 systems Operating Instructions 28-9641-23 AA
For local office contact information, visit GE, imagination at work and GE monogram are trademarks of General
Electric Company.
www.gelifesciences.com/contact , Wave Bioreactor, WAVEPOD, and Cellbag are trademarks of GE
Healthcare companies.
GE Healthcare Bio-Sciences AB All third party trademarks are the property of their respective owners.
Björkgatan 30 © 2009 General Electric Company—All rights reserved.
751 84 Uppsala First published Nov. 2009
Sweden All goods and services are sold subject to the terms and conditions of sale
of the company within GE Healthcare which supplies them. A copy of these
terms and conditions is available on request. Contact your local GE
Healthcare representative for the most current information.
www.gelifesciences.com/wave
GE Healthcare UK Ltd
Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK
imagination at work
28-9641-23 AA 11/2009