SPECIFICATIONS
FOR
DIAGNOSTIC ULTRASOUND SYSTEM
MODEL SSA-580A
(2A730-516EN*A)
Applicable System
Production Schedule : '06, Apr.
YEM55600A-1
No. 2A730-516EN*A
IMPORTANT!
2. The contents of this manual are subject to change without prior notice
and without our legal obligation.
C-1 YEM55600A-1 *
No. 2A730-516EN*A
REVISION RECORD
R-1 YEM55600A-1 *
No. 2A730-516EN*A
CONTENTS
Page
-a- YEM55600A-1
No. 2A730-516EN*A
CONTENTS - continued
Page
-b- YEM55600A-1 *
No. 2A730-516EN*A
1. OUTLINE
(1) This next-generation full-digital diagnostic ultrasound system has a user-friendly interface and
provides high-sensitivity and high-performance, enabling the acquisition of high-quality images.
(3) The dedicated data management board makes it possible to easily add advanced application
software packages, DICOM software, and many other functions.
1-1 YEM55600A-1 *
No. 2A730-516EN*A
2. FEATURES
This system employs full-digital transmission and reception circuits, which were previously
available only in top-of-the-line systems. Higher definition ultrasound beams can thus be
generated, leading to higher image quality.
Multiple frequencies can be selected using a single transducer . The optimal frequency can be
selected according to the region to be examined, the purpose of examination, and the physical
size of the patient, enabling higher throughput.
The switches are arranged in a circular arc around the trackball (which is the home position) so
that they can be operated easily with one hand. The use of sensitive hard keys allows easy
operation even during examinations.
A 15-inch LCD monitor is employed. The color monitor is provided with a mechanism
permitting tilting according to the operator's eye level, reducing operator fatigue.
Peripheral devices such as the filing unit and video printer can be mounted within the compact
body of the system, permitting the system to be moved swiftly within the hospital.
A wide range of system configurations, including systems mainly intended for abdominal and
obstetrical/gynecological examinations or systems supporting cardiovascular examinations,
are available. The systems can easily be upgraded using optional kits.
2-1 YEM55600A-1 *
No. 2A730-516EN*A
3. BASIC CONFIGURATION
Basic units
Basic configuration Basic main unit
System unit
T&R unit
Power supply unit
LCD monitor
Operation panel
Accessories
One set of operation manuals
Transducer cable hanger
Printer cable
Standard transducer
3-1 YEM55600A-1
No. 2A730-516EN*A
3-2 YEM55600A-1
No. 2A730-516EN*A
3-3 YEM55600A-1
No. 2A730-516EN*A
NOTE: For *1, 2D1 to 2D3 in the frequency columns support the settings for Multi freq of the
preset 2D menu.
3-4 YEM55600A-1
No. 2A730-516EN*A
: Can be used
: Not recommended but can be used
: Can be used only when the B-PSP and CDI-PSP/QSP kit is installed
3-5 YEM55600A-1
No. 2A730-516EN*A
Advanced
Components Premium version Basic
version
Main unit
LCD monitor
Peripheral unit rack
Sector Transducer & CW Support kit - -
(UISP-580A)
Sector Transducer Support kit - -
(UIPC-580A)
B-PSP and CDI-PSP/QSP kit -
(UIQS-580A)
Reference Signal kit - -
(UJUR-580A)
Data Management board -
(UIDM-580A)
3-6 YEM55600A-1 *
No. 2A730-516EN*A
4. PERFORMANCE
(2) Preset types specific to the system (common to all applications) : 1 type
(4) Service preset types specific to the system (common to all applications) : 1 type
(1) 2D : (R)
(2) 2D+2D : (R)+(F), (F)+(R)
(3) M+2D* : (R)+(R), (R)+(F)
(4) M : (R)
(5) PWD+2D* : (R)+(R), (R)+(F)
(6) PWD : (R)
(7) M+PWD+2D : (R)+(R)+(R), (R)+(R)+(F)
(8) M+PWD : (R)+(R)
(9) SCWD+2D* : (R)+(F), (F)+(R)
(10) SCWD : (R)
(11) CWD (pencil) : (R)
(12) CDI : (R)
(13) CDI+CDI : (R)+(F), (F)+(R)
(14) MCDI+CDI* : (R)+(R), (R)+(F), (F)+(R)
(15) MCDI : (R)
(16) PWD+CDI* : (R)+(R), (R)+(F), (F)+(R)
(17) SCWD+CDI* : (R)+(F), (F)+(R)
(18) MCDI+PWD+CDI : (R)+(R)+(F), (F)+(F)+(R)
(19) MCDI+PWD : (R)+(R)
(20) Cardiac-gated 2D : (SR)
(21) Cardiac-gated 2D+2D : (SR)+(SR), (SR)+(R), (R)+(SR), (SR)+(F), (F)+(SR)
(22) Cardiac-gated CDI : (SR)
(23) Cardiac-gated CDI+CDI : (SR)+(R), (R)+(SR), (SR)+(SR), (SR)+(F), (F)+(SR)
(24) 2D+CDI : (R)+(R)
(25) Other cardiac-gated scan modes : Same as scan modes (3), (5), (7), (14), (16), and (17)
above except that (SR) is available in place of 2D or
CDI
For scan modes marked *, display formats in which M, PWD, or SCWD is displayed in the lower
part and 2D or CDI is displayed in the upper part are also available.
4-1 YEM55600A-1
No. 2A730-516EN*A
1 2D
2D
(R)
2 2D+2D
2D 2D 2D 2D
(R) (F) (F) (R)
3 M+2D
M 2D M 2D
(R) (R) (R) (F)
2D 2D
(R) (F)
M M
(R) (R)
4 M
M
(R)
5 PWD+2D
PWD 2D PWD 2D
(R) (R) (R) (F)
2D 2D
(R) (F)
PWD PWD
(R) (R)
4-2 YEM55600A-1
No. 2A730-516EN*A
6 PWD
PWD
(R)
7 M+PWD+2D
PWD PWD
(R) (R)
M 2D M 2D
(R) (R) (R) (F)
8 M+PWD
PWD
(R)
M
(R)
9 SCWD+2D
CWD 2D CWD 2D
(R) (F) (F) (R)
2D 2D
(R) (F)
CWD CWD
(F) (R)
10 SCWD
CWD
(R)
11 CWD (pencil)
CWD
(R)
12 CDI
CDI
(R)
4-3 YEM55600A-1
No. 2A730-516EN*A
13 CDI+CDI
14 MCDI+CDI
15 MCDI
MCDI
(R)
16 PWD+CDI
17 SCWD+CDI
CDI CDI
(F) (R)
CWD CWD
(R) (F)
4-4 YEM55600A-1
No. 2A730-516EN*A
18 MCDI+PWD+CDI
PWD PWD
(R) (F)
MCDI CDI MCDI CDI
(R) (F) (F) (R)
19 MCDI+PWD
PWD
(R)
MCDI
(R)
20 Cardiac-gated 2D
2D
(SR)
21 Cardiac-gated 2D + 2D
2D 2D 2D 2D
(SR) (R) (R) (SR)
2D 2D 2D 2D
(SR) (F) (F) (SR)
2D 2D
(SR) (SR)
22 Cardiac-gated CDI
CDI
(SR)
4-5 YEM55600A-1
No. 2A730-516EN*A
CDI CDI
(SR) (SR)
24 2D+CDI
2D CDI
(R) (R)
For display formats in which M, PWD, or SCWD is displayed in the lower part and 2D or CDI is
displayed in the upper part, the images can be displayed with a magnification ratio of 1:1 or 1:2.
4-6 YEM55600A-1
No. 2A730-516EN*A
(3) Density
Types of density : L, H
It is possible to set the density to HIGH ON or OFF for each
transducer.
(6) PAN/EXPAND
4-7 YEM55600A-1
No. 2A730-516EN*A
Focus
Multifrequency
2D : 3 types
THI : 2 types (maximum)
PWD/SCWD : 2 types (1 type for SCWD)
CDI : 2 types
Reception method
(11) Adjusting acoustic output : Possible in the range from 10% to 100%, in 5% steps
4-8 YEM55600A-1
No. 2A730-516EN*A
(13) M mode
Number of channels : 1 CH
Display method : Moving bar
Scroll speed : 4 types 1, 2, 4, 8 s/screen
GAIN correction : -20 dB to +20 dB (Not exceeding the GAIN range in (9))
Sample rate : 1 sample/ms
TIME MARK : T.MARK can be turned ON/OFF in D/M.
POSITION MARK : P.MARK can be turned ON/OFF in 2D.
Display processing : Max hold
M map : 8 types (including OFF) (Controlled independently of 2D/D.)
M postprocessing : 8 types (Controlled independently of 2D/D.)
(2) Pulse repetition frequency : PWD: 1.3 kHz to 12.5 kHz (equal steps)
CWD: 4.0 kHz to 47.6 kHz (equal steps)
PWD : 7 types
CWD : 7 types
Method : FFT
No. of data items : 32 to 128
Contrast : 4 types (independent setting for PWD and CWD)
Pixel ratio : 7 types
FFT averaging : 8 types (independent setting for PWD and CWD)
D postprocessing : 8 types (Controlled independently of 2D/M.)
(11) Audio : Stereo output (blood flow toward and away from the
transducer)
4-9 YEM55600A-1
No. 2A730-516EN*A
(17) 2D-mode refresh : Allows updating of 2D mode images that were frozen during
scanning in D mode only.
(20) ANGLE MARK : This mark is used to measure the angle of the direction of the
velocity with respect to the ultrasound beam. (0° to 80°)
(22) Intelligent Doppler : Automatically adjusts the velocity range (Auto Scale Mode).
4-10 YEM55600A-1
No. 2A730-516EN*A
(5) Color image processing : Different combinations of the following parameters can
be selected for each anatomical region.
Heart : 4 types
PV : 3 types
Superficial : 1 type
Transcranial : 1 type
Abdomen : 2 types
Obstetrics/gynecology : 3 types
User : 6 types
4-11 YEM55600A-1 *
No. 2A730-516EN*A
5. DISPLAY-RELATED FEATURES
Swivel of the first arm : 120° to the right to 180° to the left
ID : 64 characters 1 line
Comment : 60 characters
YYYY/MM/DD
MM/DD/YYYY
DD/MM/YYYY
hh:mm:ss AM (PM)
hh:mm:ss (24-hour clock)
Heart rate
Character area display : Characters can be displayed at the desired position on the
image.
Up to 31 characters can be entered at one time.
5-1 YEM55600A-1
No. 2A730-516EN*A
Automatic data
Can be turned ON/OFF: Frame rate, GAIN, DR, cardiac phase display, color filter, Doppler
filter, color PRF, cine frame, other measurement values
Displayed only when parameters are changed: 2D-IP, C-IP, DEPTH, scale number
Auto-annotation
Heart : 35 types
Abdomen : 29 types
Obstetrics/gynecology : 30 types
PV : 40 types
(5) Pictograms
5-2 YEM55600A-1 *
No. 2A730-516EN*A
6. REFERENCE SIGNALS
(1) Type
Phonocardiogram
(2) Display
M mode : 3 CH
6-1 YEM55600A-1 *
No. 2A730-516EN*A
2D mode : 512 images (number of rasters per frame 256 or less, high
density)
CDI mode : 512 images (number of rasters per frame 256 or less, high
density)
MCDI mode : 15 s/1-s sweep, 30 s/2-s sweep, 60 s/4-s sweep, 120 s/8-s
sweep, stored independently of the 2D-mode setting.
PWD/CWD mode : 15 s/1-s sweep, 30 s/2-s sweep, 60 s/4-s sweep, 120 s/8-s
sweep, stored independently of the 2D-mode setting.
Playback mode
2D 2D-DUAL M (D) + 2D (CDI) M (D)
Record mode
2D
2D-DUAL
M (D) + 2D (CDI)
M (D)
ECG sync.
7-1 YEM55600A-1 *
No. 2A730-516EN*A
Distance : 8 CH
Angle : 8 CH
Volume (ellipse) : 1 CH
Cross : 8 CH
Parallel : 8 CH
Joint : 8 CH
Mean IMT : 8 CH
Trace length : 8 CH
Distance : 8 CH
Time : 8 CH
Slope : 8 CH
Velocity : 8 CH
Acceleration : 8 CH
Time : 8 CH
Velocity trace : 8 CH
Velocity histogram : 2 CH
Velocity : 8 CH
8-1 YEM55600A-1 *
No. 2A730-516EN*A
(4) OB measurements
Up to 16 items can be measured, including fetal growth, gestational age, estimated delivery
date, etc.
Trend graph function
9-1 YEM55600A-1
No. 2A730-516EN*A
9-2 YEM55600A-1 *
No. 2A730-516EN*A
Body mark
Annotation
Date/time
Visual customize (editing of colors and character size for measurements, annotations, and
body marks)
10-1 YEM55600A-1 *
No. 2A730-516EN*A
(5) CD
11-1 YEM55600A-1 *
No. 2A730-516EN*A
12. SYSTEMATIZATION
12-1 YEM55600A-1 *
No. 2A730-516EN*A
Color : 1
B/W : 1
Input : 2
Output : 2
Output : 1
(5) VGA
Output : 1
Output : 2
Input : 1 (R+L)
Output : 1 (R+L)
VCR control
DVD recorder control
(10) USB : 4 (2 for the main CPU, 2 for the image processing unit)
13-1 YEM55600A-1 *
No. 2A730-516EN*A
(1) Log on
14-1 YEM55600A-1 *
No. 2A730-516EN*A
(1) Line voltage : <1> 100 VAC to 127 VAC (rating indication)
<2> 220 VAC to 240 VAC
CLASS I EQUIPMENT
CONTINUOUS OPERATION
15-1 YEM55600A-1 *
No. 2A730-516EN*A
(2) Mass : Approximately 100 kg (peripheral units and the TEE stand not
included)
16-1 YEM55600A-1 *
No. 2A730-516EN*A
(1) General : IEC60601-1 (1988), AMD No. 1 (1991), AMD No. 2 (1995)
(3) Acoustic power : Information for Manufacturers Seeking Marketing Clearance of Diagnostic
Ultrasound Systems and Transducers (1997)
17-1 YEM55600A-1 *
No. 2A730-516EN*A
493 780
1372
Unit: mm
18-1 YEM55600A-1
No. 2A730-516EN*A
GAIN VOLUME
POSI. BALANCE
TRANS
PHYSIO APPLI EXT
DUCER
18-2 YEM55600A-1 E