Vous êtes sur la page 1sur 2

Voltage measurements

made easy
Application Note

Transient spikes can create some


serious problems when they
occur in sensitive electronic
equipment. Not only can voltage
spikes cause electronic compo-
nents to fail, but also circuitry
can be falsely triggered and pro-
duce unexpected results. And if
that’s not bad enough, the tran-
sient’s unpredictable nature
presents a considerable chal-
lenge to the technician trying to
detect the elusive signal and
locate its source.
In the past, transient problems
required a trained eye and a
complicated oscilloscope for
detection — not always a con-
venient or practical solution.
But today, technicians can
employ a high-performance digi-
tal multimeter (DMM) with the
latest recording features to track
down even the most difficult
transient problems.

Not your daddy’s DMM


Both the Fluke 87 and Fluke 189
offer a MIN MAX function, a
recording feature that allows the In order to track down tran- Only a matter of time
DMM to record the highest and sients, the DMM has to be able
lowest values. When the function to “capture” a very short change Transients can be pretty short in
is enabled, the DMM stores the in measured value and still pro- duration. How short a signal the
first reading it takes into two duce a reading that represents DMM can detect and measure
registers. One register for high the magnitude or peak value of depends on the DMM’s response
values (MAX) and one for low that change. For example, if the time, which is the length of time
(MIN). From this point on, each DMM is measuring a steady dc an input must stay at a new
measurement of the DMM is value, and a transient voltage value to be recorded. Obviously,
compared to the values in these spike is placed on the dc value, a shorter response time captures
two registers. Whenever a meas- the DMM needs to display the shorter events.
ured value is higher than the maximum voltage deviation Both the Fluke 87 and 189
value in the MAX register or caused by the transient. Also DMMs have selectable MIN MAX
lower than the value in the MIN keep in mind, the transient could response times. For tracking
register, the DMM beeps and the either increase or decrease the down transient problems, a DMM
new value is stored. steady dc voltage. needs to have very fast response

From the Fluke Digital Library @ www.fluke.com/library


times. When using a Fluke 87, the At this point, the MIN MAX
fastest MIN MAX response time is recording mode has been acti-
captured when the “PEAK MIN vated and the DMM is comparing
MAX” mode is enabled. The each new measurement against
same feature is labeled “FAST the MIN MAX registers and stor-
MN MX” on the Fluke 189. In ing any new values it detects.
both these meters, these selec- When the relay with the open
tions set the DMM’s response diode de-energizes, a transient
time to 250 microseconds. may be detected by the DMM.
To see the recorded values,
perform the following.
For the Fluke 87:
Press the to see the
maximum value measured. Press
it again to see the minimum
value measured.
For the Fluke 189:
With the maximum value already
displayed in the primary display,
each press of steps
through the minimum (MIN),
F i gure 1. average (AVG) and back to the
maximum (MAX) reading.
Again, the present measured
Step by step To capture and measure this value is always displayed in the
transient, set up the DMM as secondary display while cycling
To demonstrate the use of the follows:
MIN MAX function in detecting through the MIN MAX values.
1. Select the dc voltage mode. It may also be helpful to know
and measuring transients, con-
sider the example shown in Fig- 2. Plug the red test lead into the that the amount of time that has
ure one. The circuit is part of a V jack and the black test elapsed since the MIN MAX
control circuit that, among other lead in the COM jack. mode was activated to the last
things, controls some dc relays. 3. Connect the black test lead to recorded MAX or MIN value is
The diode across the coils of the the dc common of our control also displayed.
relays is used to short out any circuit and the red test lead The MIN MAX function is a
electrical energy created by the on the 12-volt power line very handy addition to a DMM.
collapsing magnetic field in the going to the relays. You will find it very helpful for
relay coil when the voltage is identifying intermittent problems,
removed. Assume for this exam- 4. Press the until the such as transients. The Peak or
ple that one of the diodes has desired range is selected. This Fast MIN MAX functions greatly
opened. Without the diode to should be done to get the enhance DMMs, making your
short out the energy released by best accuracy before entering work easier and more productive.
the collapsing field around the the MIN MAX mode.
coil, the energy must be dissi- 5. Activate the MIN MAX mode.
pated by the remaining circuitry For the Fluke 87:
of our control circuit. With the
Press the .
coil now acting like a source, the
voltage it creates will be added Fluke. Keeping your world
Press the to enter the
to the supply voltage, possibly up and running.
PEAK MIN MAX mode until
causing a spike on the dc supply
1 mS is displayed.
line. As this transient propagates
through the circuitry, it may For the Fluke 189:
Fluke Corporation
cause erratic and intermittent Activate FAST MN MX by PO Box 9090, Everett, WA USA 98206
behavior in our control circuit. pressing Fluke Europe B.V.
PO Box 1186, 5602 BD
then press . Eindhoven, The Netherlands
For more information call:
The maximum stored value In the U.S.A. (800) 443-5853 or
is presented in the primary Fax (425) 446-5116
display while the secondary In Europe/M-East/Africa (31 40) 2 675 200 or
display indicates the present Fax (31 40) 2 675 222
In Canada (800) 36-FLUKE or
measured value. Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2004 Fluke Corporation. All rights reserved.
Printed in U.S.A. 2/2004 2138412 A-ENG-N Rev A
2 Fluke Corporation Voltage measurements made easy

Vous aimerez peut-être aussi