Académique Documents
Professionnel Documents
Culture Documents
1.0 PURPOSE
2.0 APPLICATION
4.0 GENERAL
4.2 A proper calibration standard is one made of the same or similar material
with the same or similar heat treatment as the material to be inspected.
TABLE 4.3
Range of RSC UT Calibrataion standards
Item No : PTSCOS, ROTARY-4/06-6/2000 Page 1 of 5
Section : Rotary Shoulder 4/6 Approval : Date:1st June 2000
5.0 PROCEDURE
5.1 Prior to calibration, visually examine the plastic wedge to assure adequate
couplant between the wedge and transducer. If air bubbles are observed,
disassemble the transducer from the plastic wedge and apply sufficient
couplant to the mounting surface of the wedge. Reassemble the transducer
to the wedge, again observing that complete coupling occurs.
5.2 Activate the couplant valve and verify that couplant flows freely from the
outlets in the plastic wedge.
5.3 Calibration of the ultrasonic testing system shall be performed each time the
flaw detector is turned on, prior to performing the initial inspection, and each
time a change in calibration standard is required. Verification of calibration
against the standard shall be performed after every tenth connection
inspected and at the end of each job.
5.4 Using the proper calibration standard, obtain “peak” signal from the .010”
notch at maximum distance bu adjusting the angle of the transducer.
5.5 Set system sensitivity to obtain a 25% amplitude signal from the .010” notch
at the maximum distance.
5.6 Using the three .010” notches at varying distances, set the time varied gain
(TVG).
5.7 Return to the .010” notch at maximum distance and adjust the gain as
required to obtain a 25% amplitude signal.
5.8 Set the gate start as close as possible to the interface signal. This should be
approximately 1 ½” .
5.9 Adjust gate width to sufficiently cover the entire length of the threads to be
inspected.
5.10 Adjust gate level to approximately 5% less than the signal amplitude of the .
010” notch at Maximum distance ( refer to Paragragh 5.7).
5.11 Adjust the reject level as required to eliminate excess grass along the
baseline. Do not exceed 20% reject level.
5.12 All indications in excess of the gate threshold level shall be marked for further
investigation and disposition.
5.13 All indications marked shall be investigated with magnetic particle or
penetrant inspection, as applicable. Disposition shall be performed in
accordance with FI-012. Any ultrasonic indication that cannot be verified by
other means shall be dispositioned as follows :
A. If the signal amplitude is less than 50% of full screen height (FSH), the
disposition shall be “use as is”
B. If the signal amplitudeis equal to or greater than 50% of FSH, the
disposition shall be “reject”.
5.14 After disposition, all parts should be marked as applicable for their particular
classification. A report shall be generated detailing the inspection results.
Remaining couplant should be removed from the test surfaces.
TABLE 5.4
MATERIAL in / us cm / us in / us cm / us
ALUMINUM . 2490 . 6320 . 1230 . 3130
AL 1100-0 ( 2SO ) . 2500 . 6350 . 1220 . 3130
AL 2014 ( 14S) . 2490 . 6320 . 1210 . 3070
AL 2024 T4 ( 24ST ) . 2510 . 6370 . 1240 . 3160
AL 2117 T4 ( 17ST ) . 2560 . 6500 . 1230 . 3120
Item No : PTSCOS, ROTARY-4/06-6/2000 Page 3 of 5
Section : Rotary Shoulder 4/6 Approval : Date:1st June 2000
MATERIAL in / us cm / us in / us cm /
us
STEEL, 4150 RC64 . 2290 . 5820 . 1090 . 2770
STEEL, 4340 . 2300 . 5850 . 1280 . 3240
ACRYLIC . 1050 . 2670 . 0441 . 1120
BAKELITE . 1020 . 2590
CELLULOSE ACETATE . 0965 . 2450
DELRIN . 1000 . 2540
LUCITE . 1050 . 2670 . 0496 . 1260
NYNON . 1030 . 2620
PHENOLIC . 0560 . 1420
PLEXIGLASS ( UVA) . 1090 . 2760 . 0563 . 1430
PVC . 0905 . 2299
TEFLON . 0531 . 1350