Académique Documents
Professionnel Documents
Culture Documents
1 Approved
Consultant Signature
2 Accepted with Comments, Resubmit,
Name
Work may proceed
3 Rejected, Resubmit, Work May Not Date
Proceed
4 Information only, Reviewed Not
Required, Work May Proceed
ONGC Signature
Name (ONGC)
Date
Note: Approval/ Review by PMC/OWNER/Engineering Consultant shall not absolve the CONTRACTOR of his CONTRACTUAL obligations.
CONTRACTOR is fully responsible for complying with the requirement of the CONTRACT & other relevant documents & codes / standards.
Acceptance of any document shall not be interpreted as permitting any deviation & shall in no way relieve / absolve the CONTRACTOR of
his CONTRACTUAL obligations under the terms of the CONTRACT, for correct design & execution of works.
HOLD LIST
No Section Description
REMARKS – This Document shall be read in conjunction with APPLUS Rtd - Doc Ref : UT 19 510
Revision: 2 – Title: MANUAL ULTRASONIC TESTING PROCEDURE
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
Contractor : McDermott
Scope : Manual Ultrasonic Testing of welds including thickness, lamination and repair areas.
Validity : This procedure is valid only for Manual Ultrasonic Testing of material with a surface
temperature up to 60°C
Development – Revisions
Revision No. Prepared By Date Approved By Date Amendment Details
Contents
1 Purpose 5
2 Scope 5
3 References 5
Applus+ RTD Documents 5
Codes and Standards 5
Pipeline Contractor / Company Specific Documents 5
Abbreviations and Definitions 5
4 Personnel Qualifications 6
5 Safety 6
6 Equipment 7
Ultrasonic Flaw Detector 7
Periodical Verifications 7
Probes 7
International Calibration Block 8
Project Specific Reference Block 8
Couplant 8
7 Calibration 9
Calibration Control 9
Time Base Calibration 9
Reference Sensitivity 10
Transfer Correction 10
Scanning Sensitivity 10
Signal to Noise Ratio 10
8 Testing 11
Surface Preparation 11
Probe Movement, Direction and Scanning Speed 11
Probe Angles 11
Parent Material Testing 11
Weld Testing 11
9 Acceptance Criteria 12
10 Reporting 12
Project Specific Reference Block 13
Reference Block PIN: 2019|8.625|12.7|137762-MUT-301 13
Reference Block PIN: 2019|15|20.6|137762-MUT-302 14
Probe Movement and Direction 15
Example of Report 16
Acceptance Criteria 17
Page 4 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
1 Purpose
The purpose of this procedure is to detail the methodology used for the Ultrasonic Testing (UT), based on the
requirements of Pipeline contractor / Company specific documents No. 2020-01391D-01-SY-CRT-0001 and the
Offshore Standard DNVGL-ST-F101: 2017 during the KG-DWN 98/2 Project.
2 Scope
This procedure details the methods for UT of welds including thickness measurement, lamination check and
repair areas of the Carbon steel pipes, diameter 6” up to 24” and wall thickness between 10 mm up to 25 mm.
3 References
Applus+ RTD Documents
• PIPL-12.03.0004 Rev.0 Control Procedure for Ultrasonic Flaw Detectors.
• NL-10 Written practice for the training, qualification, certification and authorization
of NDT personnel.
Page 5 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
4 Personnel Qualifications
Personnel shall be qualified in accordance to Applus+ RTD regional “Written Practice” for the training,
qualification, certification and authorization of NDT Personnel.
Only personnel qualified to UT Level 2 as minimum, ISO 9712 shall interpret the test results.
5 Safety
All work shall be performed in accordance with Applus+ RTD safety policies including all national, local safety
rules and regulations including any Client Specific Health and Safety programs. All personnel shall be aware
that they have a responsibility to work safely and prevent others working in an unsafe manner.
All portable electronic equipment shall be electrically tested in accordance with Applus+ RTD/Client
regulations.
The equipment is non-intrinsically safe and the data acquisition unit can be damaged by water. Both these
points shall be taken into account when planning or performing data collection.
Batteries shall not be replaced in any area of live plant or where gas or other explosive materials may be
present.
All technicians shall wear proper PPE in the manner that they were designed to be used. Any defective PPE
shall be replaced immediately.
Information regarding the Applus+ RTD occupational Health and Safety program can be obtained from the
Project Health and Safety Management Plan.
Page 6 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
6 Equipment
Ultrasonic Flaw Detector
An ultrasonic flaw detector with a DAC curve facility, with valid verification report according to the control
procedure PIPL-12.03.0004, and shall have the following minimum capabilities:
• Be applicable for the pulse echo technique and for the double probe technique.
• Cover as a minimum the frequency range from 1 to 6 MHz.
• Have a calibrated gain regulator with maximum 2 dB per step over a range of at least 60 dB.
• Have a flat screen accessible from the front for direct plotting of reference curves or be equipped with
digital DAC- display presentation of user-defined curves.
• Allow echoes with amplitudes of 5% of FSH to be clearly detectable under test conditions.
Periodical Verifications
The Amplitude Control, Screen Height and Time Base Linearity shall be checked for correct verification at the
beginning of each period of extended use or every 3 months, whichever is less according to the control
procedure PIPL-12.03.0004.
Probes
All probes shall be checked before and after use and at approximately four-hourly intervals for correct angles,
beam alignment and surface. Probes not in compliance to the requirements described in section 7.1 of this
procedure shall be corrected or withdrawn from service.
Page 7 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
Couplant
Water or Ultrasonic Gel shall be used as couplant and the same coupling medium is to be used for both
calibration and testing.
Manufacturers data sheets/COSHH sheets shall be available upon request, ensuring that the couplant utilized
will have no detrimental effect to the pipe material or the MUT Operator and that all traces of couplant utilized
will be removed post inspection works.
Page 8 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
7 Calibration
Calibration Control
Interval
Before and after use and at approximately four-hourly intervals, the range scale, probe angle and primary gain
must be checked and confirmed. Checks shall also be carried out whenever a system parameter is changed or
changes in the equivalent settings are suspected.
Correction
If deviation is found to be larger than 2% of range scale, or 4 dB of primary gain setting or 2° of nominal
probe angle, the equipment shall be re-calibrated and the testing carried out with the equipment over the
previous period shall be repeated. Re-calibration shall be performed whenever the equipment has been out
function for any reason including on-off and whenever there is any doubt concerning proper function of the
equipment.
Page 9 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
Reference Sensitivity
The angled beam probes sensitivity shall be based on a minimum of a 3-point reference level DAC, ensuring
the whole weld volume is inspected.
The initial sensitivity setting shall be established over half skip distance by obtaining a minimum response of
80% FSH from the SDH. Without changing this sensitivity setting, obtain maximum amplitudes from the SDHs
between half and full skip distance. Mark the peaks of each maximized echo on the screen.
Connect the screen marks to provide the DAC.
If only one side of the weld is accessible for testing, the reference sensitivity shall be increase with 6dB.
Transfer Correction
Due to the use of a project specific reference block a transfer correction is not required.
Scanning Sensitivity
Scanning sensitivity shall be reference sensitivity (section 7.3) + 6 dB (additional scanning sensitivity).
For the evaluation of an in imperfection the +6 dB (additional scanning sensitivity) shall be removed.
Page 10 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
8 Testing
Surface Preparation
The temperature difference during calibration and testing shall be within +/- 14°C. Temperature of the object
under testing shall not exceed 60°C.
The base material, both sides of the weld, shall be cleaned by either power wire brushing and/or filing to
remove all weld spatters, dirt, rust, loose scale and other irregularities that may interfere with the testing. The
ultrasonic operator shall determine whether the surfaces have been adequately cleaned.
Probe Angles
The probe angles to be chosen shall be according the DNVGL-ST-F101, table D-2 as shown below.
Weld Testing
Using the probes detailed in section 6.3 and following sections 6 to 8 investigate any relevant indications
exceeding 20% DAC reference curve (evaluation level).
All indications exceeding 20% of the DAC reference curve are to be evaluated and all indications exceeding
50% DAC are to be reported. Investigation is to be performed to the extent that the operator can determine
the shape and location of the indication.
Ultrasonic inspection must be considered from all surfaces i.e. ID and OD where possible. Weld configurations
may exist where access for ultrasonics is restricted i.e. flange welds, the UT report must clearly identify these
ultrasonic restriction(s).
Indication length evaluation:
• Establish the maximum amplitude on the indication,
• Move the probe parallel with the weld until the amplitude drops to the evaluation level (DAC -6dB) and
mark position,
• Establish the maximum amplitude on the indication,
• Move the probe parallel with the weld until the amplitude drops on the other side of the indication to the
evaluation level (DAC -6dB) height and mark position,
• The indication length is the distance between the marked points.
Weld configurations may exist where access for ultrasonic is restricted i.e. flange welds, the UT Report must
clearly identify these ultrasonic restriction(s). Ultrasonic inspection must be considered from all surfaces i.e.
ID and OD.
Page 11 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
9 Acceptance Criteria
The acceptance criteria shall be according DNVGL-ST-F101, table D-6 as shown in Attachment D.
10 Reporting
For each inspection all information shown on the report format in Attachment - C shall be recorded.
Page 12 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
Page 13 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
Page 14 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
Page 15 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
Example of Report
Page 16 of 17
Doc Ref : UT 19 510
Revision : 2
MCD Ref. :
Date : 02-08-2019
Title : Manual Ultrasonic Testing Procedure
Project : KG-DWN 98/2
Acceptance Criteria
Page 17 of 17