Vous êtes sur la page 1sur 9

Anal. Chem.

1994,66, 3916-3924

Particle Beam Aqueous Sample Introduction for Hollow


Cathode Atomic Emission Spectroscopy
Jianzhang You, James C. Fanning, and R. Kenneth Marcus'
Department of Chemistry, Howard L. Hunter Chemical Laboratory, Clemson University,
Clemson, South Carolina 29634- 1905

The design and preliminary operating characteristics of a the range of applications of the devices to virtually any solid
particle beam/hollow cathode atomic emission spectrometry sample type, making them among the most diverse of the
(PB/HC-AES) device is described. By use of a high-efficiency solids analysis techniques.
thermal-concentric nebulization system coupled to a particle Paralleling the developments in other areas of atomic
beam LC/MS interface, analyte particles originating in aqueous spectrometry, where techniques best suited for aerosol/vapor
solution are introduced into a heated hollow cathode glow introduction are being investigated with solids introduction
discharge for subsequent vaporization and excitation. The schemes, many laboratories are now looking for efficient means
Thermabeam (Extrel Corp.) interface permits liquid introduc- of applying GD sources to the analysis of solution-based
tion in either flow injection or chromatographic mode. Helium samples. A long-term goal in this laboratory is the develop-
and argon are evaluated as discharge gases over a range of ment of a single spectrochemical device for the analysis of
pressures and discharge currents, with He chosen as the most solid and solution samples. Solution samples have been
appropriate for subsequent analytical studies. Nebulization analyzed indirectly with glow discharge devices through the
temperature, liquid (solvent) flow rate, and vaporization sputter atomization of solution residue~.~-IOIn this method,
temperature are evaluated in order to determine their effects aliquots of the sample solution (1-100 pL) are applied to the
on analyte emission characteristics. Optimized sample in- cathode (either a planar or hollow cathode) surface and dried
troduction and He plasma operation conditions are employed under an infrared lamp. The resultant residue is atomized by
to study the analytical characteristics of the PB/HC-AES cathodic sputtering with the atomic population sampled by
system. Background equivalent concentrations (BECs) for atomic a b ~ o r p t i o n ,emission,'
~.~ fluorescence,8 or mass spec-
Na and Cs (as nitrates) in aqueous solutions are 0.53 and 0.74 t r o m e t r i e ~ .The
~ ? ~GD
~ methods for solution residue analysis
ppm, respectively. Subsequent studies of cesium at ppm levels have been overshadowed by flame atomic absorption and
in diluted (1OOOX) simulated nuclear waste solutions (5.0 M plasma emissiontechniques because of their wide applicability,
NaNO3 and 0.1 M KOH) indicated that the addition of excess sensitivity, and high sample throughput.
chloride as a counterion, rather than nitrate, greatly enhances
In order to improve the sample throughput of GD devices
-
detectability. BEC values for Cs in aqueous solution were
lowered by a factor of 10, while for the diluted simulant the
values were lowered by >lOOX, with detection limits in the
for solution sample analysis, it is desirable to have some means
of continuous, or at least segmented, flow for sample
introduction. Furthermore, if a glow discharge source were
latter found to be -8 ppb.
attached to the end of a liquid chromatography (LC) column,
species could be separated and introduced into the excitation
Glow discharge (GD) devices have become valuable tools region for subsequent analysis by optical or mass spectrometries
for direct solids elemental analysis of a wide variety of (MS). Because of the low pressures and powers employed in
materials, from bulk metals to thin-film systems.' Glow GD devices (resulting in T,,, <lo00 Kl1 ), the discharges are
discharges provide efficient atomization, excitation, and quite inefficient at aqueous sample desolvation and vaporiza-
ionization of sample material and have been employed tion in comparison to atmospheric pressure flames and plasmas.
extensively in atomic absorption, emission, and mass spec- Even if solvent evaporation could be achieved within the
trometric studiese2 In addition to the analysis of metals and discharge volume, the presence of water-related gaseous species
alloys, sample preparation methodologies have been developed (assuming aqueous sample introduction) in the plasma can
for the analysis of nonconducting powders by compaction in cause band emission in AES and lower the metastable atom
a conducting host m a t r i ~ .More
~ recently, radio frequency populations in the negative glow region, quenching analyte
glow discharge devices have been developed for the direct
analysis of electrically insulating materials without matrix ( 5 ) Bruhn, C. G.; Harrison, W. W. Anal. Chem. 1978,50, 16-21.
m~dification.~ The use of radio frequency power has widened (6) Chakrabarti, C. L.; Headrick, K. L.; Bertels, P. C.; Back, M. H. J. Anal. At.
Specrrom. 1988, 3, 713-723.
(7) Chen, F.; Williams, J. C. Anal. Chem. 1990, 62, 489-495.
(1) Harrison, W. W.; Barshick, C. M.; Kiingler, J. A.;Ratiiff, P. A,; Mei, Y. Anal. (8) Smith, B. W.; Omenetto, N.; Winefordner, J. D. Spectrochim. Acta, Part 8
Chem. 1990, 62, 943-949A. 1984, 398, 1389-1393.
(2) Marcus, R. K. Glow Discharge Spectroscopies; Plenum Press: New York, (9) Jakubowski, N.; Stuewer, D.; Tolg, G. Spectrochim. Acta, Parr 8 1991,468,
1993. 155-163.
(3) Dogan, L.; Laqua, K.; Massmann, H. Specfrochim. Acta, Part 8 1972, 278, (10) Barshick, C. M.; Duckworth, D. C.; Smith, D. H.J. Am. SOC.MassSpectrom.
65-88. 1993, 4, 47-53.
(4) Marcus, R. K. J . Anal. At. Spectrom. 1993, 8, 935-943. (11) West, C . D.; Human, H. G. C. Spectrochim. Acra. PartB 1976,318,81-92.

3910 Analytical Chemisrty, Vol. 66, No. 22, November 15, 1994 0003-2700/94/0366-39 16$04.50/0
0 1994 American Chemical Society
ionization.12 Therefore, any liquid sample introduction
approach must include a high-efficiency desolvation step and
t hv
N ~ ~ Cathode
Hollow
~ ~ ,
subsequentsolvent vapor removal prior to analyte introduction
to the discharge source. Complications due to solvent Macor
background and its effects on spectral quality are also major
concerns when one is attempting to introduce liquids into
conventional (organic) mass spectrometric ion sources.13 For
these reasons, it is logical to pursue methods of liquid sample
introduction that have been developed for liquid chromatog-
raphy/mass spectrometry (LC/MS) for the coupling of LC
(or other liquid delivery systems) with the glow discharge.
The implementationof LC/MS approaches to liquid sample

'a'
introduction into GD devices were first described by Brackett Particle
and Vickers in 1983.14 That source provided continuous \ Themoblock
Assembly
sample introduction into a glow discharge with a chain
conveyor, wherein the sample was deposited on the moving Beam t
chain (straight wire and link types), passed through a reduced
pressure oven to affect solvent evaporation, and then passed Discharge
across the GD cathode. In this way, the metal chain became Gas
part of the cathode and the analyte residue was sputter-ablated. Figure 1. Diagrammatic representation of the plasma source region
of the particle beam/hollow cathode glow discharge atomic emission
Difficulties in reproducibility in the initial studies were device.
overcome by use of a bead chain, as this allowed for even
segmentation of the sample on each of the beads? The latter
studies suggested that liquid delivery rates of up to 200 pL ablation and by losses in the diffuse plasma structure. Analyte
min-I could be tolerated before excessive solvent loading of emission intensities were found to optimize under those
the plasma led to instabilities. A sputter atomization step conditions wherein the aluminum target was heated through
was confirmed by the fact that little correlation was found sputtering, rather than those conditions which produce
between detection limits (ranging from 0.8 to 3.2 pg mL-I for optimum photon emission from the plasma glow.
Pb, Cd, B, and Cu) and compound volatility. Alternatively, We present here a preliminary evaluation illustrating the
compound decomposition in the desolvation furnace, resulting practicality of interfacing a particle beam device with a heated
in memory effects, was found to be an important consideration. hollow cathode (HC) glow discharge and the subsequent
In consideringalternativeintroduction schemes, the results measurement of the atomic population by emission spec-
of Brackett and VickersI4J5further point to the need to employ trometry (PB/HC-AB). By use of the heated hollow cathode
an interface that has a very high efficiency for solvent vapor volume, the analyte particles are vaporized into the gas phase
removal. Particle beam devices are finding increased usage as free atoms. The influence of discharge gas identity (Ar or
in mass spectrometric analyses of LC separations because of He), vaporization temperature, and discharge conditions are
their ability to operate at reasonable liquid flow rates while evaluated for sodium and cesium analytes. The practical
subsequently removing solvent vapors to levels that permit application of the particle beam approach is illustrated by the
electron impact ionization.16J7 In this laboratory we have analysis of cesium at ppm levels in diluted simulated nuclear
employed a particle beam LC/MS interface (Thermabeam, waste solutions containing initially high levels of salt ( 5 M
Extrel Corp., Pittsburgh, PA) for liquid sample introduction NaN03 and 0.1 M KOH). Signal suppression induced by
into a GD device. The Thermabeam utilizes a thermal- the presence of the salt is overcome by the addition of small
concentric nebulizer to convert the eluent from the LC column amounts of HCl. Data presented here suggest that the PB/
into a finely dispersed aerosol. The aerosol is desolvated and GD-AES approach may indeed be a viable technique for the
passed through a momentum separator, producing a particle elemental analysis of volume-limited or chromatographically
beam that is directed into the source volume. A previous introduced samples.
study illustrated the feasibility of the particle beam introduc-
tion approach for aqueous sample analysis by glow discharge EXPERIMENTAL SECTION
atomic emissionspectrometry.'* In that work, analyte particles
Glow Discharge Source. A diagrammatic representation
were effectively collected on a planar cathode and the atomic
of the hollow cathode source region of the particle beam/
population was produced through cathodic sputtering. The
hollow cathode glow discharge interface is shown in Figure
analytical sensitivity of this approach was limited to the single
1. The hollow cathode is contained within a stainless steel
ppm level due to the inefficient, nonquantitative sample
block (termed the thermoblock) which is mounted onto a 6
(12)Hess, K. R.;Harrison, W. W. Anal. Chem. 1988,60,691-696. in. diameter flange, within a stainless steel five-way cross.
( I 3) Covey, T.R.;Lee, E. D.; Bruins, A. P.; Henion, J. D. Anal. Chem. 1986,58,
1451 A-1461A. Use of the cross allows for placement of the optical sampling
(14)Brackett, J. M.; Vickers, T. J. Specfmchfm.Acta, Part B 1983,38B,979-985. window, vacuum pumping port, discharge gas inlet, electrical
(1 5) Brackett,J. M.; Vickers,T. J. Spectrochim. Acfa. Parr B 1984,39B,837-841. feedthroughs, and particle beam interface on the appropriate
(16)Willoughby, R. C.; Browner, R. F. Anal. Chem. 1984.56.2626-2631.
(17)Singh, R. P.; Brindle, I. D.; Jones, T. R. B.; Miller, J. M.; Chiba, M. J. Am. flanges. The plasma is operated by using helium or argon as
Sa.Mass Spectrom. 1993,4,898-905. the discharge gas with the bulk chamber pressure monitored
(18)Strange, C. M.; Marcus, R. K. Spcfrochim. Ada, Parr B 1991,46B,517-
526. by means of a thermocouple gauge (DV-4D, Teledyne
To
Monochromator

Momentum

5-way Cross
c-----
Source Housing

Stage 1 Stage 2 U
Pump Pump
t
Discharge
Gas
Flgure 2. Complete Thermabeam nebullzer and partlcle beam Interface coupled to the heated hollow cathode glow discharge atomic emission
source.

Hastings-Raydist, Hampton, VA). The discharge is powered directly into a stainless steel desolvation chamber (3.4 cm i.d.,
by a Kepco (Flushing, NY) Model BHK 2000 supply operating 10.1 cm long) that is encircled by electrical heating tapes.
in a constant-current mode. The desolvated aerosol passes through a two-stage momentum
The implementationof the heated hollow cathode discharge separator that "skims" light-weight gas molecules from the
volume is integral to the enhanced performance over the flow and also serves as a differential pumping device to obtain
previous PB/GD approach. The titanium hollow cathode (3 a beam consisting only of particles. The interface is flange
mm diameter, 17 mm length) is housed within the 28 X 28 mounted opposite from the GD source, with the exit aperture
X 60 mm3 stainless steel thermoblock, heated by a pair of fitted into the side of the thermoblock, -14 mm from the
commercially available cartridge heaters (Model SC 25 15, heated surface. No attempt has been made to modify the
Scientific Instrument Services, Ringoes, NJ) which are commercial particle beam nebulizer or momentum separator.
powered by a laboratory transformer. In this way, the Solution Preparation and Delivery. For the purposes of
temperature of the hollow cathode can be regulated up to 220 evaluatingthe feasibility of the particle beam/hollow cathode
OC, as measured by a W-Re thermocouple. Desolvated coupling,aqueous stock solutions of NaN03 and CsNO3 were
analyte particles are introduced to the thermoblock through prepared with deionized, distilled water from reagent grade
the momentum separator (described in the next section) inorganic salts. Simulated nuclear waste solutions were also
through a 1.5 mm diameter orifice, wherein they collide with prepared from reagent grade salts, having a nominal com-
the heated wall and are vaporized. Volatilized analyte is then position of 5.0 M NaN03 and 0.1 M KOH. The solvent
swept into the hollow cathode discharge by a perpendicular delivery system was a Waters (Division of Millipore, Milford,
flow of discharge gas and subjected to excitation collisions MA) Model 5 10 high-performance liquid chromatography
within the discharge. The separation of the vaporization and pump. The solutions were passed through a 0.76 mm 0.d.
excitation processes allows for optimization of each without stainless steel tubing with no attempt made at performing
compromise to theother, which was not thecase in the previous chromatographic separations. Solution flow rates are adjust-
studies.'* able at intervals of 0.1 mL min-' up to a maximum of 1.5 mL
Particle Beam Interface. Figure 2 is an expanded drawing m i d . Throughout these studies, samples were introduced
of the Thermabeam/glow discharge interface geometry. The into the flowing aqueous solvent stream in a flow injection
Thermabeam is a particle beam type of LC-MS interface mode, with 200 pL injection volumes.
that employs a thermal-concentric nebulizer to generate a Optical System. Atomic emission produced from within
finely dispersed aerosol, a heated expansion (spray) chamber the hollow cathode discharge was focused by a plano-convex,
for desolvation of the solvent, and a momentum separator to fused silica lens onto the entrance slit of a scanning mono-
inject dry particles of the analyte into the source volume. The chromator. The spectrometer employed is a 0.24 m Czerny-
liquid sample is passed into the nebulizer through a fused Turner mount having a 2400 gr/mm holographic grating
silica capillary (0.15 mm i.d.) within a stainless steel capillary (Digikrom 240, CVI Laser Corp., Albuquerque, NM).
(0.50 mm i.d.). Helium gas is passed through the steel tubing Spectral scanning, wavelength selection, and slit widths are
and flowsaround the capillary tip and thus forms a concentric easily controlled via the monochromator control interface
nebulizer. A potential difference between the ends of the provided by the manufacturer. A photomultipliertube (RCA
stainless steel tubing resistively heats the tubing, effectively 1P28) and analog photometer (Model 110, Pacific Instru-
heating the He nebulizer gas, adding a thermal component ments, Concord, CA) are employed to detect and record
to the nebulization action at the tip. The aerosol is introduced spectral intensities. Emission intensity data are recorded by

3918 A n a ~ c a l l m l s t t y Vol.
, 66, No. 22, November 75, 1994
an X/Y recorder (Model 200, Houston Instruments, Austin,
TX), with transient peak heights used in all tabulations and
calculations.

RESULTS AND DISCUSSION .-c,


Role of Discharge Parameters. As in the case with all GD v)
C
Q,
devices, the parameters of discharge current and pressure,
along with the discharge gas identity, play key roles in the -
c,
C

observed emission characteristics of the PB/HC-AES system.


Different from other GD sources, though, is the fact that the
atomic population in the plasma need only to be excited, rather
than also being produced by sputtering. In this evaluation,
both helium and argon were investigated as discharge gases,
as they represent the most commonly applied gases for hollow
cathode emission and mass spectrometric applications, re-
spectively. The primary difference lies in the ability of He
to excite atoms to high-lying levels due to its higher ionization
potential and the resultant higher discharge voltages, while
Ar is a much better sputtering gas while still efficiently
exciting/ionizing most elements of the periodic table. Of
particular interest in the comparison of the two gases was the
response of analyte emission intensities over a range of
discharge conditions and the relative complexityof the emission
spectra.
Figure3 illustrates theresponseofNa (I) 589.0nmemission
intensity over a range of discharge conditions for the (a) He
and (b) Ar discharge gases. Throughout this study, the
NaNO3 concentration was maintained at a value of 0.1%,
with a liquid flow rate of 0.5 mL min-l. As seen in both plots,
thedischarge current has a direct effect on emission intensities,
reflecting the typical response of electron number densities
found in hollow cathode discharges. l 9 A proportional response
of analyte emission to increases in discharge current is not
generally observed in glow discharges where sputtering effects Figure 3. Effect of discharge pressure and current on Na (I) 589.0
are involved,20q21even though electron densities do increase.22 nm emission intensity for (a, top) heliumand (b, bottom)argondischarge
gases: 0.1 % NaN03 solution; 0.5 mL min-I solution flow rate.
The respective responses to changes in discharge gas pressure
are very different for He and Ar, with the former showing a
more well-defined range of optimum operation. A comparable case in the previous studies of the PB/GD combination,'* the
set of experiments employing Cs as the analyte yielded similar overall magnitude of band emission is not prohibitively high,
response characteristics. The differences between the gases roughly of the same magnitude as the Na (I) 589.0 nm emission
are likely explained by the physical nature of the plasmas, from the 10 ppm N a N 0 3 test solution. The low amounts of
wherein the energy within the negative glow (and indeed the band structure are a testament to the efficiency of the
shape) of the He plasma is more concentrated within the center desolvation and particle beam separation characteristics of
of the cathode, while the Ar plasma is visually more diffuse. the Thermabeam interface. Neither the atomic or molecular
Studies by Kirichenko et al.23in hollow cathode discharges spectral contributions of the discharge gases nor the residual
have found as well that differences exist in plasma shapes and water are considered detrimental. Given the possible wide
character between He and Ar plasmas. range of excitation energies that might be presented in
The spectral comparison between He and Ar lies in the applications of this particle beam/hollow cathode device, a
overall spectral complexity due to the gases themselves. With decision has been made to employ solely He as the discharge
respect to atomic transition overlaps, He produces a far simpler gas.
spectrum than Ar, particularly in the spectral region between Role of Nebulization Parameters. As with any technique
400 and 500 nm. On the other hand, the higher excitation wherein solution nebulization is employed, one must be
energy afforded by He does tend to produce slightly more concerned with the role that solution/solvent introduction has
band emission from OH' and air-related species. As was the on plasma performance. In the Thermabeam interface, the
key parameters affecting nebulization efficiency are the
(19) Howorka, F.; Pahl, M. Z . Naturforsch. Teil A 1972, 27, 1425. temperature at the nebulizer tip and the liquid flow rate. In
( 2 0 ) Boumans, P. W. J. M.Anal. Chem. 1972, 44, 1219-1228.
(21) Leis, F.; Broekaert, J. A. C.; Laqua, K. Spectrochim. Acta, Part E 1987,428,
practice, He nebulization gas flow rate and the desolvation
1169-1 176.
(22) Fang, D.; Marcus, R. K. Spectrochim. Acta, Parr E 1990,458, 1053-1074.
(23) Kirichenko, V. I.; Tkachenko, V. M.; Tyutyunnik, V. B. Sov. Phys. Tech Phys.
1976, 21, 1080-1086.
-
chamber temperature are also important. Previous experience
has suggested that He flow rates over 50 mL m i d do not
substantially affect nebulization characteristics, while opti-

Analytical Chemistry. Vol. 66, No. 22, November 15, 1994 3919
mum desolvation efficiency (>95%) can be achieved with
chamber wall temperatures of >220 "C.18 The role of
nebulization (tip) temperature was assessed by use of a
thermocouple attached at the exit end of the stainless steel
capillary. This temperature is not the temperature directly
applied to the solvent but a measure of the heating of the He
nebulizer gas. Helium, having a very high thermal conduc-
tivity, transfers the heat to the fused silica capillary indirectly
heating the solvent. Additional thermal energy is imparted
as the hot He gas passes the silica capillary exit, producing
the combination thermal-concentric nebulization character.
Figure 4a depicts the observed, integrated Na (I) 589.0 nm 0.0I ' I ' I , I , I , I
160 180 200 220 240 260
emission intensity as a function of the nebulizer tip tempera-
ture. As would be expected, the efficiency of nebulization (as Tip Temperature ('C)
observed through emission intensity) increases with tempera-
ture, up to a maximum of 220 "C. Above temperatures of
220 OC,the nebulizer begins to sporadically eject solvent and
vapor from the silica as vaporization appears to occur from
within the capillary rather than at the end. Additionally,
clogging of the capillary becomes an issue for high-solids
(>O. 1%) solutions. For the remainder of the studies described
here, a nebulizer tip temperature of 220 "C is maintained.
Similar to other nebulizer-based solution introduction
schemes, the liquid flow rate is a critical parameter as it will
ultimately limit both the nebulization efficiency and the
response of the source (plasma/flame) characteristics. More
so than any other source, the effect of introduced solvent vapors
into the hollow cathode region can be detrimental, even at Flow Rate (ml/min)
very low (C10 ppm) levels.12 As seen in Figure 4b, the response
of integrated Na (I) intensity is inversely related to increases
in solvent delivery rate. The reasons for this response can be *.O 1
evaluated in terms of the nebulization, desolvation, and 1.8-
particle/vapor introduction stages. In the nebulization step,
higher flow rates would be expected to lower the efficiency 1.6-
as more thermal energy is required for aerosol generation.
This was evident by the fact that the current passing through 1.4-
the steel heating capillary had to be increased with increasing
flow rate in order to maintain a constant tip temperature of 1.2-
220 OC. Even though the temperature was held constant, the
1 .o-
reduced residence time of the solution in the capillary and in
the aerosol generation region at the tip would be expected to
lower the overall efficiency. Likewise, heavy solvent load
0.0
100
I
150 200 250 300
within the desolvation chamber resulted in a need to increase
Block Temperature ("c)
the heating current there to maintain a constant temperature.
Figure 4. Roles of nebulization and vaporization parameters on Na
Although the walls of the chamber were kept hot, surely the (I) 589.0 nm emission Intensity (0.1 % NaN03solution): (a, top) nebulizer
desolvation step also suffers at high flow rates. Finally, in the tip temperature: (b, middle) solvent flow rate; (c, bottom) vaporization
event of reduced desolvation efficiency, more water vapor temperature.
would be expected to pass to the discharge, quenching the
plasma energetics. This effect is the least likely culprit as no and nebulizer gas flow rates, nebulizer temperature, and solvent
(expected) change in discharge voltage was observed under identity (volatility).
these conditions. This observation is consistent with the One additional experimental variable involved with the
previous studies and confirms the excellent vapor-filtering particle beam interface is the vaporization temperature that
characteristics of the particle beam interface. While more the introduced particles experience. The previous studies
detailed studies of the nebulization and desolvation qualities showed quite clearly that the the inherent sputtering in the
of the Thermabeam device are in order, it has been generally glow dischargedevice occurred at rates insufficient to maintain
observed that moderate flow rates of -0.5 mL min-' are less the temporal quality of flow injected or chromatographically
susceptible to clogging and will be employed throughout the introduced analytes; Le., the residence time was too long.
remainder of the work presented here. Future studies will Thermal vaporization has long been employed in organic LC/
include a thorough evaluation of the relationship between liquid MS applicationsof particle beam devices.I6 The present design

3920 AnalyticalChemistry, Vol. 66,No. 22, November 75, 1994


I I I
0.0 1 .o 2.0
Time (min)
Flgure 5. Temporal profile of Na (I) 589.0 nm emission intensity for
a 200 pL injection of a 10 ppm Na solution: solution flow rate, 0.5 mL Na Concentration (ppm)
min-I.
8
y - 0.86 + 1,17x RAZ - 0.961

follows this lead as the analyte particles are introduced


perpendicularly to the hollow cathode axis, flash vaporized, 6-
and swept by the discharge gas flow into the plasma. In such
an approach, the temperature within the thermoblock is key
4-
to efficient particle vaporization and analyte introduction.
Figure 4c illustrates the role of block temperature on the
integrated Na (I) intensity under the nebulization conditions
2-
described above. As can be seen, the vaporization process is
highly dependent on temperature. This is the expected
response characteristic as the particles would require some I
O ! ' , ' I ' I ' I ' ! '
minimum energy for volatilization, with the efficiency 0 1 2 3 4 5 6
increasing to the point of completevaporization. Interestingly,
the response shows a rollover at the highest temperatures, Cs Concentration (ppm)
rather than the expected plateau. The reasons for this response Figure 6. Analytical response curves obtained for aqueous (a, top)
are not clear at this point and require further evaluation. It NaN03(589.0 nm) and (b, bottom) CsN03 (852.1 nm) solutions. Error
bars representthe range of values obtainedfor three consecutive 200
seems obvious that the absolute threshold temperature pL inJections.
(analogous to an appearance temperature in GF-AAS) will
be a function of the sample particle identity and could affect
and also having established its feasibility for proposed
substantial matrix effects. Future studies will look at this
chromatographic applications, an evaluation of the basic
variable for a wide range of metal/anion combinations. Given
calibration characteristics was in order. Initial efforts focused
the nature of the response curve, it seems apparent that the
on the generation of working curves for simple solutions of
highest practical temperature will allow for the most uniform
aqueous Na and Cs, each in the form of nitrates. Shown in
sample introduction.
Figure 6 are the obtained calibration curves (transient peak
An important characteristic of any proposed chromato-
height) over concentration ranges of between 0.001 and 10
graphic detector is a low degree of dispersion as indicated by
ppm for sodium (a) and 0.05 and 5 ppm for cesium (b).
the temporal response of the detected signal. Evidence of the
Indicated on the respective plots are the least-squares linear
flow dispersion induced by the particle beam introduction is
regression data as well as the sample-to-sample variability
provided in Figure 5. Shown is a strip chart trace of the
obtained for triplicate injections. While these preliminary
signal generated by the injection of 200 pLof NaN03 solution,
calibration plots cover limited concentration ranges, reasonable
which was 10 ppm in Na. The 0.5 mL min-' flow rate
linearity is observed, particularly in the case of sodium.
corresponds to an ideal signal transient time of 24 s. As can
Background equivalent concentrations calculated from the
be seen, the temporal profile exhibits a very sharp leading
working curves were also encouraging. In comparison to the
edge, with the signal approaching the baseline level in -20
initial work with the planar glow discharge arrangement,24
s. It is suggested from this response that the volatilization
the sodium sensitivity is quite good with a BEC = 0.5 ppm,
process is relatively free from memory effects and that the
with the BEC from the cesium regression equal to 0.7 ppm.
particle and vapor transport processes do not suffer from
From these preliminary calibration data, detection limits (DL
excessive dead volume. In addition, the plasma stability of
= hblank/m, nblank = 10) are calculated to be 0.05 and 0.1
the current design is very good as well. This transient signal
ppm, respectively for Na and Cs. Given the rather early stage
illustrates the desired response qualities necessary for future
of instrument development, and a rather simplistic optical
implementation of this device in chromatographic sampling
and detection system, we consider these to be quite reasonable
applications.
figures of merit. It should be emphasized that while cesium
Analytical Characteristics. Having optimized the plasma
and nebulization characteristics of the PB/HC-AES device, (24) Strange, C. M. Masters Thesis,Clemson University, Clemson, SC, 1992.

Analytical Chemistry, Vol. 66, No. 22, November IS, 1994 3921
is in general a strong emitter at the 852.1 nm wavelength, its 4
analytical characteristics in atomic emission spectrometry are y = 0.22+ 2.56e-2x R A 2 = 0.974

poor due to the very low responsivity of photomultiplier tubes


and low transmission characteristics of optical spectrometers
that are designed for optimum performance at wavelengths
below 500 nm (as is the instrumentation employed here).
Further analytical evaluation into the possible applicability
of the PB/HC-AES approach was pursued by way of a
collaborative project with the Westinghouse Savannah River
Co. (Aiken, SC) at the Savannah River Site (SRS), where
a major nuclear waste disposal program is underway. The
primary waste mixture has an approximate bulk makeup of
5 M NaN03,O. 1 M KOH (pH 14) with most heavy metal ions 0 20 40 60 80 100 120
precipitated as hydroxides or oxides. The primary radioactive Cs Concentration (ppm)
species present in the solution is I3'Cs at a concentration of
-20 ppm. The objective of the collaborative project was for
the inorganic chemist (J.C.F.) to develop methods to efficiently
remove Cs+ from solution. In order to assess the respective
efficiencies of the developed procedures in simulated waste
I y = 1.14 + 1.88e-Zx RA.2 = 0.994

solutions, in-house analytical methods were needed.


The bulk solution is obviously quite caustic and would be
classified in most instances as having a high degreeof dissolved
solids. The combination of high alkalinity, high solids, and
having a high concentration of easily ionizable elements (EIEs)
poses particular challenges for traditional atomic spectrometric
methods such as flame AAS or ICP emission or mass
spectrometries. 'The high solids content would generally
dictate a minimum dilution factor of lOOX for standard 0 20 40 60 80 100 120
pneumatic nebulization systems, though the high level of EIEs Cs Concentration (ppb)
would dictate this step as well, even for nebulizers with high
solids capabilities. The constricted nature of the fused silica 2.61 I

(150 wm i.d.) introduction path for the Thermabeam nebulizer,


2.4-
along with the aforementioned possibility to desolvate along
the capillary path, required that a dilution factor of 500- 2.2 -
lOOOX be employed to efficiently nebulize the test solution in
this case without the deleterious effects of clogging. 2.0 -
The initial analytical characterization was carried out by
1 .a-
generating a working curve for Cs doped into the stock
simulated waste solution (lOOOX dilution). Figure 7a shows 1.6-
the obtained response curve taken under the same nebulization
and plasma conditions as those employed in the neat solution
in Figure 6b. Immediately obvious in these data is the large
suppression of both the analytical signal intensities as well as 1*2b ' 20 ' 40 ' 60 ' 80 ' 100' * !O
a lowering of the sensitivity (slope) relative to the neat solution Cs Concentration (ppb)
curve, with the background equivalent concentration increasing Flgure 7. Analytical response curves obtained for Cs (I) 852.1 nm
to -9 ppm. At this point it is not immediately obvious where emission in solutions composed of (a, top) 5 M NaN03/0.1 M KOH
in the complete nebulization/desolvation/vaporization/ matrix (lOOOX dilution in D I H20),(b, middle) aqueous CsN03 with HCI
addition (5:1), and (c, bottom) 5 M NaN03/0.1 M KOH matrix (lOOOX
excitation sequence that the matrix effect comes into play. dilution in DI H20) with HCI addition (51). Error bars represent the
Physical characteristics that might be expected to change range of values obtained for three consecutive 200 pL injections.
relative to the neat solution include the solution viscosity,
analyte particle size, particle volatility, and, subsequently,
plasma excitation conditions. In a rather simple test to to the final particle characteristics, either size or volatility, as
examine the role of pH in the matrix effect, small aliquots of the chloride ion from HC1 addition is in great excess in
HC1 were added to the test solutions and the resultant Cs (I) comparison to the nitrate ion within the sample matrix.
emission intensities monitored. It was quickly found that a Panels b and c of Figure 7 further illustrate the profound
5: 1 mixture of the test solution and concentrated HC1produced effect of HCl addition (5:1, sample/acid) on the analytical
a great enhancement in the analyte emission intensity, whereas response of Cs (I) emission for the aqueous CsNO3 and Cs-
the use of H N 0 3 produced no effect at all (even though the doped separation solutions, respectively. Note first the
separation solutions were neutralized by the respective acid concentrations depicted in the calibration curves, now in the
additions). These results point to an effect that is more related ppb range. For the aqueous solution, the slope of the response

3922 Analytical Chemistry, Vol. 66, No. 22, November 15, 1994
curve exhibits far better detectability and a > 1OX lower BEC of 10-3 1 ppb, in aqueous solution. Depending on the amount
value (60 ppb) relative to theoriginal data (Figure 6b). Similar of dilution required to nebulize/atomize the waste solution
effects were seen as well for HCl addition to the aqueous effectively, the detection limits for each of these methods are
NaN03 test solutions. Figure 7c illustrates a much more more or less comparable. In any case, the approach described
dramatic change in response curve for the separation solution here seems to hold a great deal of promise, having sensitivities

-
case. In this instance, the BEC value decreases by 2 orders
of magnitude, from -9 ppm to 100 ppb. The far better
calibration characteristicsfor the latter case, in which the pH
comparable to the aforementioned methods, while having the
additional advantage of ready chromatographic sampling
capabilities of relatively small sample volumes (200 pL).
change and the extent of anion (N03-) compensation are far Detailed evaluations of nebulization and transport parameters
greater than for aqueous solution, do not clearly differentiate are hoped to improve the detection limits reported here as
the actual mechanism for the removal of the matrix effect. well as reduce the required sample volumes.
The definite increase in both Cs (I) and Na (I) emission
intensities upon addition of excess C1- clearly illustrates the CONCLUSIONS
fact that matrix effects exist in either the nebulization, Preliminary data have been presented illustrating the basic
desolvation, particle transport and vaporization, or plasma operating characteristics of the combination of a particle beam
excitation processes. Comparison of the physical character- LC/MS interface with a heated hollow cathode discharge
istics of the respective chloride and nitrate salts reveals source for the analysis of aqueous samples (PB/HC-AES).
somewhat surprising qualities. An initial assumption that The combination of efficient desolvation and solvent removal
the chloride salts have higher volatilities upon impaction in properties of the particle beam interface allows for the
the heated block was not correct, as the boiling points of the presentation of analyte particles to the hollow cathode source
metal chlorides are substantially higher than the nitrates in without the deleterious effects of residual water vapor loading
both cases. Therefore one must turn to the possibility of on the plasma. In addition, the use of an HPLC pump and
analyte losses in the desolvation process in the case of the flow injection sample introduction illustrate excellent temporal
nitrates. That is, the nitrates are sufficiently volatile that qualities with regard to future applications as chromatographic
desolvated salts vaporize enroute to the source volume and detectors. The use of the heated hollow cathode source to
are lost in the momentum separator, as small particles (vapors) affect particle vaporization and subsequent analyte excitation
are not efficiently transported. The respective solubilities of was found to be a far superior approach than the planar cathode
the salts would also point to transport-related phenomena. In geometry investigated previously.’8
both cases, the chlorides have higher solubilities, and therefore Plasma operation parameters (current and pressure) were
form larger particles than the nitrates. These transport-related evaluated for their effects on Na (I) atomic emission intensities
matrix effects will be assessed in future experiments involving for both helium and argon discharge gases. While the emission
the collection of solute particles and subsequent microscopic intensities derived with the respective gases were relatively
observation of particle size and densities. Additionally, a far equivalent, the overall spectral purity and ability to excite
ranging counterion study is in order to assess the role of anion high-lying transitions were found to be more advantageous
identity on plasma excitation efficiency, as this may also for the He gas. The nebulization parameters of solution flow
contribute to the observed matrix effect@). rate and nebulizer tip temperature were investigated, with
Regardless of the precise mechanism(s) involved, it is clear low flow rates being found to be more efficient and tip
that the addition of HCl provides a different counterion temperaturesof -220 OC most optimal. Lower temperatures
environment for all of the solutions tested in this study. The were found to be less efficient at producing dry aerosols, while
sensitivities exhibited here (DLcs = 8 ppb in aqueous CsN03 higher temperatures wereapt tocreate clogging in thenebulizer
and 10 ppb for the diluted simulated waste) are indicative of capillary due to premature desolvation. Evaluation of the
role of particle vaporization temperature indicated that block
the effectiveness of the chloride addition in normalizing the
temperatures of >-200 OC were required to provide sub-
Cs response from vastly different matrices. These detection
stantial vaporization. Under the optimized conditions, injec-
limits are difficult to compare with well-established optical
tions of Na analyte (10 ppm) exhibited temporal characteristics
methods of Cs analysis, as “typical” values for ICP-AES and
that indicated no appreciable dead volume contributions to
furnace AAS are not included in general texts. Detection
band broadening.
limits of 8 and 20 ppb can be found for flame atomic emission
and a b s o r p t i ~ n .Characteristic
~~ masses for furnace AAS The basic analytical characteristics of the PB/HC-AES
are reported to be 0.3 ppb (20 pL samples).25 Figures of approach were evaluated for Na and Cs analytes. Calibration
ICP-AES are not widely published due to the aforementioned curves derived from aqueous standards prepared from the
difficulties with PMT detection at 852.1 nm and the fact that respective nitrate salts exhibit good linearity and calibration
Cs is nearly 100%ionized in the plasma, with the ion emitting characteristics, with BECs found to be at the tenths of parts-
in vacuum-UV region of the spectrum. Thompson and per-million level. Practical application of the technique was
Barnes26have included a detection limit for Cs in a tabulation demonstrated by the analysis of Cs in diluted simulated nuclear
listing it in a category of elements with values in the range waste solutions, which contained high levels of sodium and
potassium salts. While the heavy salt matrix severelydepressed
(25) Ingle, J. D., Jr.; Crouch, S. R. Spectrochemical Analysis; Prentice Hall:
theemission characteristics of Cs (even after lOOOX dilution),
Englewood Cliffs, NJ, 1988; Chapters 8 and 10. it was found that a nominal addition of HCl resulted in
(26) Thompson, M.; Barnes, R. M. In Inductively Coupled Plasmas in Analytical enhanced detectability for both aqueous and separation
Atomic Spectrometry, 2nd ed.; Montaser, A,, Golightly, D. W., Eds.; VCH
Publishers: New York, 1992; Chapter 5, p 265. solutions. This suggests that chloride ion acts as a preferential

Analytical Chemistry, Vol. 66, No. 22, November 15, 1994 3923
counterion over nitrate and seemingly results in an analyte technique described here will not likely supplant current
particle with more favorable transport characteristics. The methods of bulk solution analyses, such as flame atomic
resultant BECs for Cs in aqueous and salt solutions were absorption or ICP atomic emission, it is believed that the
reduced by 1-2 orders of magnitude relative to the native unique qualities described here will find selected applications.
solutions without HC1 addition. Detection limits for Cs in
the diluted separation solution were determined to be 10
-
parts per billion ( 10 ppm in the original sample).
- ACKNOWLEDGMENT
Financial support from the National Science Foundation
The results presented here give very positive evidence that under Grant CHE-9117152 and the Department of Energy,
of particle beam sample introduction into a heated hollow Office of Technology Development, through the Westinghouse
cathode device warrants further exploration. The abilities to Savannah River Co. in cooperation with the South Carolina
work in a flow injection mode with small sample volumes or Universities Research and Education Foundation is greatly
to directly sample chromatographic eluents are very attractive appreciated. Donation of the Thermabeam nebulizer and
features. Future studies will focus on evaluating the role of particle beam interface by Extrel, Division of Millipore, is
nebulization parameters (liquid and gas flow rates, capillary also appreciated.
size, and solvent makeup) and counterion identity on analyte
response. Ultimately, implementation of ion chromatography
to affect ionic species separations is envisioned. Additional Received for review May 3, 1994. Accepted August 18, 1994."
studies are currently underway to investigate the PB/HC
approach for elemental mass spectrometry. While the a Abstract published in Advance ACS Absfracfs.October 1, 1994.

3924 Analytical Chemistry, Vol. 66,No. 22, November 15, 1994

Vous aimerez peut-être aussi