0 FIELD MODIFICATION INSTRUCTIONS

9.1 INDEX

Page FMI Number Date Title

9‐3 1 ‐ 1733 12/06/2006 Solera Opus Software Release v. 1.3.0

Solera Titan / Opus Service Manual FIELD MODIFICATION INSTRUCTIONS


P/N D0792 REV. A 9-1
THIS PAGE IS INTENTIONALLY LEFT BLANK 

FIELD MODIFICATION INSTRUCTIONS Solera Titan / Opus Service Manual


9-2 P/N D0792 REV. A
FMI #1‐1733 SOLERA OPUS SOFTWARE RELEASE VERSION 1.3.0, 06 DEC 2006

Purpose

The purpose of this Field Modification Instruction (FMI) is to provide notification of the release of 
version 1.3.0 software for the Solera Opus system. The addition of software version 1.3.0 to existing 
Solera Opus systems should be accomplished on an “as needed” basis. 

Notable changes within version 1.3.0 software include the addition of LimeLight handpiece support. 
The LimeLight handpiece (already supported by the Xeo platform) emits light in the range of 
520‐1100 nm and is intended to treat Benign Pigmented Lesions. Additional changes include several 
Snowflake / Sunshine mode display interface tweaks and a modified error reporting strategy that 
should reduce the occurrence of E339. 

Type

Routine (next service call):___
Mandatory (safety ‐ immediate):___
As Needed:_X_

Units Affected

All Solera Opus systems

Materials Needed

U16 flash program pn. 7000255 rE ‐ Main DSP v1.3.0 (cksum 6892) 
U46 flash program pn. 7000256 rE ‐ Safety DSP v1.3.0 (cksum C2CE) 
U17 flash program pn. 7000254 rD ‐ Flash Memory v1.3.0 (cksum EF14) 
J‐Link flash tools and USB cable 

Instructions

1. After updating U16 and U46 of the control board and U17 of the display board, test the operation 
of all available delivery devices. 

The methods used to load (flash) the updated software onto the control and display boards are 
outlined in several procedures located in the Work Instruction folder of the Service Tool CD. 
There are no significant changes in the flashing procedures, other than the updated flash code. 

Handpiece test parameters and expected results are detailed in a series of informational bulletins 
located in the Delivery Device folder of the Service Tool CD. Carefully position the sapphire win‐
dow of the handpiece parallel to, and approximately 1‐2 mm from, the energy detector. Fire sev‐
eral shots while holding the handpiece steady to determine the correct output energy. 
Solera Titan / Opus Service Manual FIELD MODIFICATION INSTRUCTIONS
P/N D0792 REV. A 9-3
2. Demonstrate the updated functionality of the system to the customer. 

The Service Engineer must document on the Service Report form (984005847) upon successful 
completion of the Field Modification Instructions. Include the system serial number, FMI number, 
name of the person performing the modification, and date of completion. The Service Report form is 
to be faxed to Cutera at 1‐415‐330‐2444, Attn: FMI Completion.

FIELD MODIFICATION INSTRUCTIONS Solera Titan / Opus Service Manual


9-4 P/N D0792 REV. A