Version history
Version Release date
3228/0115 28.08.18 Initial Version in English BREL
a
V1.0 02.10.2018 Initial Version in English KARL STORZ
1 Upgrade instruction
1.1 Required equipment for upgrading the C-MAC PM
8403XD C-MAC PM (device to be updated)
8403XDA Rechargeable lithium battery for C-MAC PM
8403XDP Connecting cable
Latest software release 747VXXX according valid release notes of KARL STORZ
Compatible C-MAC camera device (e.g. 8403HX Videolaryngoscope)
PC with USB port
Minimum system requirements
o Operating system: Windows 7 or higher
o USB port: USB 2.0 HighSpeed (480 Mbps)
Attention
The safety hints and instructions of the KARL STORZ In-
struction manual must be known and followed any time.
Note that the latest by KARL STORZ released software ver-
sion is allowed be installed on the device only.
2 Upgrade procedure
2.1 Preparation
2.1.1 Remove the battery of the C-MAC PM. Then connect the C-MAC PM to the PC.
Seite 2 von 6
If that is the case you have to wait for approx. 10 minutes. Then the message will disappear.
To install a new software release on the C-MAC PM connect the Pocket Monitor via the USB
cable 8403XDP with the computer.
Note
If the message ‘Check and repair (recommended)’ appears,
then this process should be initiated.
Always follow the instructions shown on the PC.
2.1.2 Unzip the C-MAC PM software
Copy the files of the latest software release for the C-MAC PM to your computer. The soft-
ware will be provided as zip file (747vXXX_update.zip). The unpacked file will create a folder
named ‘update’ on your computer.
Attention
Make sure that your installed anti-virus system on your com-
puter allows an access of C-MAC PM as an external system.
Otherwise interrupt anti-virus scan of C-MAC PM.
As soon as the C-MAC PM and the PC is connection is built up, the C-MAC PM boots up. Af-
ter a few seconds the automatic play will be opened. Choose the option open folder to dis-
play files.
If the automatic play does not appear, open the explorer and choose the drive named C-
MAC PM.
Seite 3 von 6
Content of the folder named ‘update’.
To upgrade the C-MAC PM, copy the folder named ‘update’ by drag and drop to the root of
the C-MAC PM (do not copy the ‘update’ folder in a subdirectory).
Seite 4 von 6
If the copying is completed, open the Windows Explorer, click to the drive with the right
mouse key and choose ‘eject’.
Note
Choose the ‘eject‘ function in the Windows Explorer by a click
on the right mouse button. If you eject the device on the
symbol on the task list, the update process does not start
with all windows installations.
After rejecting the drive, the upgrade process of the C-MAC PM will be started. The progress
of the upgrade will be displayed on the screen of the C-MAC PM. This process will last about
3 minutes.
Seite 5 von 6
As soon as the update process is completed, the C-MAC PM displays the message ‚please
unplug cable‘.
If the verification failed and the C-MAC Pocket Monitor does not display the correct software
version, repeat the procedure from step 2.2.
Seite 6 von 6