QP-822V
This License Agreement is made and entered into between NIHON KOHDEN CORPORATION or its overseas subsidiary
company (Hereinafter called “NK”) and User with respect to the copyright, the license to use the Data Management Software
QP-822V (Hereinafter called “Software”) and the conditions for using other relevant materials, such as the operator’s manual.
To use this Software, the User must agree to this Agreement. When the seal of the Software package is opened, it means that
the User has agreed to this Agreement.
Article 1 Definition
For the purpose of this Agreement, the following terms shall have the following meanings.
1. “Software” means any program or data file which are provided by NK solely in connection with the use of the Software,
which is used on the computer owned by the User to acquire data from the monitor network specified by NK and output
the acquired data to another network according to the protocol manual of the Software.
2. “Use of the Software” means running or use of the software on a computer.
3. “Output Data” means data output from the Software according to the protocol manual of the Software.
Article 3 Responsibility
1. User must thoroughly check the safety of the use of the Software, Output Data and the instruments. If death, personal
injury, property damage has occurred to a third party or dispute has occurred with a third party, the User shall settle such
problem at its own risk and expense and NK takes no responsibility.
2. NK does not guarantee that use of the Software does not infringe patent rights, copyrights and other rights of a third party.
3. When it becomes necessary to obtain permits or to make notifications in relation to use or application of the Software, on
the basis of the relevant national laws and regulations, User shall obtain said permits or make said notifications at its own
risk and expense.
Article 4 Prohibition
User is prohibited to do the following, regardless of whether it is free of charge or not.
1. Using or duplicating the Software for purposes other than specified in this Agreement and duplicating relevant material
2. Modifying and reverse engineering the Software
3. Giving all or part of the Software to a third party
4. Licensing the use of the Software to a third party
5. Lending or renting the Software or transacting it as secondhand goods
6. Installing the Software on a network server
Article 5 Guarantee
1. NK does not guarantee the commerciality and compatibility of the Software to the purpose of the User and has no
responsibility for guarantee and warranty for the items other than specified in this Agreement. The responsibility for
selecting and using of this Software and the outcome is on the User, not on NK.
2. NK guarantees the operation of the Software on Windows XP Professional. However, if the Software does not function due
to other software installed in the same personal computer, NK has no responsibility for the operation of the Software.
Article 7 Arbitration
When any discrepancy occurs in the interpretation of the stipulations of this agreement or when any disagreement occurs
between the User and NK in regard to matters not covered by the stipulations of this agreement, the two parties shall discuss
the matter amicably in order to reach a solution in good faith. If such an amicable settlement is not attained through such
negotiation between the parties, the matter shall be settled by arbitration in Tokyo, according to the Rules of the Japan
Commercial Arbitration Association. The decision or award of such arbitration shall be final and binding upon the parties.
LICENSEE Date:
Hospital installed:
Product Model: Company:
(Signature):
LICENSEE’s Name:
Title:
LICENSOR Date:
Engineering Operations
(Signature):
LICENSOR’s Name:
Title:
Contents
CD-ROM Handling and Storing Caution .......................................................................... i
Introduction
Overview The QP-822V Data Management Software can be installed in a PC with Windows XP
Professional. The Data Management Software allows data communication with a
Nihon Kohden hematology analyzer. You can receive the measured data from a
hematology analyzer, and edit and print the data on the PC.
The Data Management Software also supports work lists. You can create and edit a
work list on the PC or receive it from the hospital network and send it to the
hematology analyzer.
Composition
Hematology Analyzer To use the Data Management Software with the hematology analyzer, the following
Software Version software version of the hematology analyzer is required.
NOTE
The Data Management Software cannot function properly with earlier
versions.
Receiving QC data
QC is enabled by receiving QC data from the hematology analyzer.
Work list
Work list measurement is performed by sending a work list from the Data
Management Software to the hematology analyzer. Creating a work list is easy with
the work list function. ID bar code labels can also be printed. Work lists created on
other software applications can also be used.
A work list can be created automatically from an order from the LIS (Laboratory
Information System) and then sent to the hematology analyzer. Measured data can
also be sent to the LIS.
NOTE
You need software for the LIS to use the Data Management Software in
the LIS mode. Contact your Nihon Kohden distributor for the
communication specifications for LAN access.
System Configuration
System Example
Network
Host (LIS)
TM-L90
Possible to connect
up to 2 units
Data Flow between the Data Management Software and Hematology Analyzer
Worklist
USB
USB
LAN USB
Data Management Software
Main Window
Toolbar
Measured Data
Window
Select tab to display other patient information
Patient information:
Displays the patient
information. Some items
can be edited.
Scattergrams
View Screen Display the measured data for the selected date.
Display style: display style can be changed View older or newer data
Date
Installation Flowchart
PC Hardware Minimum hardware requirements to install the Data Management Software are
Requirements indicated below. Before installation, check that the PC meets the hardware
requirements.
NOTE
If the minimum hardware requirements are not met, the Data
Management Software cannot function. Even if the requirements are
met, it is possible that the Data Management Software may not function
depending on the model and settings of the PC.
• Operator’s manual
Installing the Data STEP-1: Insert the installation CD in the CD drive of the PC.
Management Software
The QP-822V Installer window appears.
Installer starts.
* If it does not start automatically, run Install.exe on CD.
Uninstalling the Data 1. From the Windows [start] menu, select [Settings] → [Control Panel].
Management Software
Connecting the Data Connect the Data Management Software PC and the hematology analyzer with a
Management Software PC USB cable.
to the Hematology
Analyzer
Installation of driver
starts automatically
when the USB cable USB terminal
is connected. USB terminal
Refer to the next
section.
USB cable
NOTE
• A USB cable is not provided with the Data Management Software.
• The position of USB terminal differs depending on the hematology
analyzer. Refer to the operator’s manual of your hematology analyzer.
Connecting Two If you connect two hematology analyzers to the Data Management Software PC, the
Hematology Analyzers to analyzer name (Unit No.) must be set on the hematology analyzers so that the Data
the Data Management Management Software can recognize the data origin. Refer to the operator’s manual
Software PC of the hematology analyzers.
When connecting only one hematology analyzer, the analyzer name does not need
to be set.
<Example>
When connecting two MEK-8222J/K hematology analyzers:
Set [UNIT 1] or [UNIT 2] for item [ANALYZER NAME] of [USB SETTING] on the
[MENU] → [SETTINGS] → [OUTPUT FORMAT] screen.
When you connect the Data Management Software PC and hematology analyzer
with the USB cable, the PC detects the hematology analyzer and starts installing the
driver.
1. When the PC detects the hematology analyzer, the following window appears.
2. Select [Search for a suitable driver for my device [recommended] and click
[Next].
After installing the driver, check that the Data Management Software starts
correctly. Refer to “Starting the Data Management Software” in Section 3.
Installing Drivers for P-touch Printer for Printing Bar Code Labels
NOTE
This procedure is for using P-touch printer for printing bar code labels.
The bar code label width for the Data When not using P-touch printer, there is no need to do this procedure.
Management Software is either 12 or 24
mm.
1. Connect the P-touch printer to the Data Management Software PC and set P-
touch printer settings on the analyzer. For details, refer to the P-touch manual.
2. In Windows File Explorer, open the CD-ROM drive which contains the QP 822V
installation CD.
4. Select [Install] from the pop-up menu. The driver for the P-touch printer is
installed.
NOTE
• Do not disconnect the USB cable between the Data Management
Software PC and the hematology analyzer when the Data Management
Software is running. It may cause the PC to shut down. If the PC shuts
down, restart it.
• Do not turn off the power or disconnect the power cord of the
hematology analyzer connected to the Data Management Software
when the Data Management Software is running. It may cause the PC
to shut down. If the PC shuts down, restart it.
• In the Windows Power properties, set standby to “Never” when the
Data Management Software is running. Otherwise it may cause the PC
to shut down. If the PC shuts down, restart it.
• Always close the Data Management Software before turning off the
connected hematology analyzer.
Starting the Data Check the following before starting the Data Management Software.
Management Software • The Data Management Software is installed in the PC.
• The hematology analyzer is connected to the PC.
• The hematology analyzer is turned on.
USB mark
When the hematology analyzer is connected to the PC, the USB mark is displayed in
the DMS-Pro screen. Two USB marks are displayed when two hematology analyzers
are connected.
NOTE
The USB mark may not be displayed if you connect the hematology
analyzer to the PC after the Data Management Software is started. In
this case, reconnect the USB cable.
Checking Items Before Check the following items before operation to assure proper operation. If you find
Operation any problem, take proper measures referring to this manual. If the problem cannot
be solved, contact your Nihon Kohden distributor.
Item Description
PC, hematology analyzer and printer cables are connected
Connection
properly.
The hematology analyzer is turned on.
The PC is turned on.
Power on
Printer and other equipment are turned on.
The Data Management Software starts normally.
There is enough free space on the hard disk.
The printing is not faint.
Basic operation
Setting items are set properly.
Date and time settings are correct.
NOTE
Check that the date and time of the PC and hematology analyzer are
correct at the start of operation. Refer to the operator’s manual of the
hematology analyzer and PC.
Checking Items After Check the following items after operation to assure normal operation for next use.
Operation
Item Description
Temporarily changed settings are changed back to the
Before shut down
previous settings.
Error confirmation Make sure no errors were found during use.
Power PC, hematology analyzer and printer are turned off.
Setting the Folder for You can receive the data measured by the hematology analyzer and save it in the
Saving Measured Data PC. Before receiving data, set the folder to save measured data.
2. Click [Other] from the [Other settings] on the sDMS_Pro. The Other settings
It is recommended to regularly change window appears.
the folder to save measured data.
Otherwise, the software will run slowly
because of the data increase in the folder.
Name the folder according to the date of
measurement.
4. Select the folder which you created in step 1 and click [OK].
5. Check that the selected folder path is displayed, then click [OK]. This folder is
now set for saving received data.
Save as CSV File: Check this option when using the received measured data
in other applications.
Save as Bitmap File: Check this option when saving the histogram and
scattergram as bitmaps.
Receiving Currently The Data Management Software receives the measured data when the hematology
Measured Data analyzer completes measuring the sample. Received data is automatically saved in
the specified folder.
Measured data
It is recommended to regularly change
the folder to save measured data.
Otherwise, the software will run slowly
because of the data increase in the folder.
Name the folder according to the date of
measurement.
NOTE
• A message is displayed when the measured data in one folder
exceeds 1000. In this case, change the folder to save data.
• Change the folder to save data when its saved data exceeds 4000. Data
cannot be saved in the received data history when the data exceeds
4000.
• A message is displayed when there is not enough free space in the
hard disk. Make sure there is enough free space in the hard disk.
Receiving Previously The Data Management Software can receive already measured data from the
Measured Data hematology analyzer and save them to the selected folder.
Measured data
Send the measured data to the Data Management Software by operation on the
hematology analyzer. Refer to the operator’s manual of your hematology analyzer
for details on how to send the data.
<Example>
On the MEK-8222J/K hematology analyzer:
Press the [TRANSFER] key on the [DETAILS] screen, or press the [TRANSFER] key
on the [DAY] screen.
NOTE
• When the measured data is sent from the DAY screen, only the
numeric data is transferred. Scattergram and histogram are not sent.
• When two hematology analyzers are connected to the Data
Management Software, do not transfer the daily data from both
hematology analyzers at the same time. Transfer daily measured data
only from one hematology analyzer at a time.
Receiving QC Data
Create a folder for QC data. The received QC data is automatically saved in this
folder.
The measured data saved in the Data Management Software can be displayed by
doing one of the following.
Displaying Measured Data Display a list of received measured data by selecting the date.
from the VIEW Window
1. Click [View] button on the toolbar.
2. Display a list of measured data by selecting the date. Select the measured data
and click it.
Switch the display of Display the page. Switch the page
scattergram or The display is “present
histogram page/total page”
Display the number of total
data by selecting the date.
Displaying Measured Data You can set search keyword(s) to display only the specific measured data from the
from the Received Data received data folder.
Folder by Using Search
Function 1. Click the [SEARCH] icon on the toolbar or select [Search] in the menu.
* Search criteria
[Search in all Database]---Search from all data
[Search in the last]---Search from the last data to the specified month
[Range of data]---Search among the specified months/years
[Date]---Search on the specified date
[Today]---Search on data for today
NAME
DEPARTMENT
To search all data in the selected period, ID
start search by not checking any criteria. DATE OF BIRTH
AGE
PATHOLOGICAL DATA
(To search for samples with a pathological flag, check this box.)
[Close] button
Close search result
If the search results have over 1000 * If other measured data screen is displayed, each selecting of data on search result
data, the [view all] button is not usable. list displays its measured data screen and data on the selected column.
Printing Search Result 1. Click [Print] button on search result list. Print setting window appears.
Display preview
3. Check the search result list to be printed. If OK, click the [Print] button to print
it.
Printing Displayed Received data from the hematology analyzer can be printed with the Data
Measured Data Management Software.
Display the measured data you want to print and click [Print] in the [File] menu.
The displayed data is printed.
Viewing Print Image of Click [Preview] in the [File] on toolbar on the DMS-Pro with the measured data
Displayed Measured Data displayed.
Work List A work list is an instruction file for measuring samples, including patient information
such as measurement type, normal range, and patient name.
Work List Measurement The work list is transferred from the Data Management Software (DMS) to the
hematology analyzer and used to control measurement.
Modes The work list has two operation modes-Bar code and No Bar Code.
Bar code
The embedded bar code reader (option) reads the bar code label on the sample tube
to identify the sample regardless of its rack position. Make bar code labels for all
samples and attach them on the sample tubes before measurement.
Bar code mode is further divided into the following 3 modes.
• Stand-alone mode Section 4
• Network Path mode Section 4
• Network communication mode Section 4
No Bar code
The sample is indentified by its rack location. Measurement is quicker because there
is no need to make and attach bar codes. However, you must be sure not to mix up
the samples.
This mode is further divided into the following 2 modes.
• Stand-alone mode Section 4
• Network Path mode Section 4
Stand-alone Mode:
This mode sends the work list made by the Data Management Software (without
external software or files) to the hematology analyzer to perform measurement.
4. Send the work list to the hematology 2. Receive the work list from the DMS
analyzer
3. Start measurement.
Bar Code Stand-alone 1. Click [Work list] icon on the menu or [Work list] icon on the toolbar.
Mode Setting Procedure
3. Select [TUBES WITH BAR CODE] on the work list operation tab. Also select
[Stand alone] on the combo box and click [OK].
2. Send the work list to the hematology 2. Receive the work list from the DMS
analyzer
3. Start measurement
3. Receive the measured data from the 6. End the measurement of all samples
hematology analyzer
No Bar Code Stand-alone 1. Click [Work list] icon on the menu or [Work list] icon on the toolbar.
Mode Setting Procedure
3. Select [TUBES WITHOUT BAR CODE] on the work list operation tab. Also
select [Stand alone] on the combo box and click [OK].
Bar Code 1. Select [TUBES WITH Bar code] on the option form.
3. Display the work list registration form. Enter the data in each box.
ID: Enter the patient ID. You can enter some of the digits and
click [Search patient] to display all patients with a similar ID.
Then select from the shorter list. ID is required for
registration.
The data are added to the work list and the work list registration form is
displayed again. To edit or to enter more data, Repeat step 3. After you finish,
click [Cancel].
No Bar Corde 1. Select [TUBES WITHOUT Bar code] on the option form.
ID: Enter the patient ID. You can enter some of the digits and
click [Search patient] to display all patients with a similar ID.
Then select from the shorter list. For work list when rack ID
is “M” without bar code, ID is required for registration. For
rack ID other than ”M”, ID is not required.
The registered work list with the items selected can be printed.
Select Print item: Click items to add them to the list on the right.
Delete print item: If you registered a wrong item, select it and press
[Delete] on the keyboard.
[Previous/Next page]
Click this to display a preview of the
previous or next page.
Setting Bar Code Print 1. Click [Options] on the work list form
Settigs
3. Set the bar code printer, the bar code label width, other settings and click [OK].
Printing a Bar Code Label Select the data to print on the bar code label.
1. Before sending the work list, put the hematology analyzer on standby to receive
a work list, and set the work list screen display as follows.
2. Click [Send] on the work list window in the Data Management Software.
If the IDs match, a confirmation box appears.
3. Select [Yes] on the dialog box to send the work list data from the Data
Management Software to the hematology analyzer.
In this mode, an other application is required to make a work list file in the specified
folder. Also bar codes must be attached to the sample tubes.
NOTE
• The work list cannot be read if the file name is wrong. The file name is
case-sensitive.
• The work list can have up to 500 data.
• Do not register data of the same ID on the work list.
• The DMS tries to transfer data via USB to the connected hematology
analyzer. If the hematology analyzer is not reading the work list,
transfer is impossible. Before reading the work list file on the Data
Management Software, change the hematology analyzer screen to the
READY or work list screen.
• On the Data Management Software, automatic transfer of the work list
is followed by automatic deleting of the transferred work list.
Measurement Flow
3. Start measurement
3. Receive the measured data from the 5. Send the measured data of one sample
hematology analyzer to the DMS
Setting the Network Path 1. Click [Work list] on the menu or [Work list] on the toolbar.
Using Bar Code
3. Select [TUBES WITH BARCODE] on the work list operation tab. Also select
[Network Path] on the combo box. The watch folder settings are displayed.
Set the folder to make the work list file and click [OK].
Setting the Network Path 1. Click [Work list] on the menu or [Work list] on the toolbar.
Using No Bar Code
3. Select [TUBES WITHOUT BAR CODE] on the work list operation tab. Also
select [Network Path] on the combo box. The watch folder settings are
displayed. Set the folder to make the work list file and click [OK].
Flow of Operation
1. Receive the work list from the server 1. Display the READY or work list
software screen screen
2. Send the work list to the hematology 2. Receive the work list from the DMS
analyzer
3. Start measurement
Setting the Network 1. Click [Work List] on the menu or [Work List] icon on the toolbar.
Communication Mode
Using Bar Code
3. Select [TUBES WITH BAR CODE] on the work list operation tab. Also select
[Network communication] on the combo box. Click [OK].
Retry interval:
Timeout time:
Note
If these settings do not match the server settings, communication is
impossible.
Select Print item: Click items to add them to the list on the right.
Delete print item: If you registered a wrong item, select it and press [Delete] on the
keyboard.
Margin: You can set the printing margins. Use the arrow buttons to
change the order of items.
Q&A
Q.1 Can a printer for bar code and a printer for measured data be connected to the
PC at the same time?
A.1 Yes.
The WA-820V printer (TM-L90 by Seiko Epson equivalent) must be connected
to the serial port, but a locally purchased printer can be connected to the USB
(or parallel) port. The P-touch printer for bar code label is connected to the
USB port.
Q.3 No printing.
A.3 Printer may be disconnected or set incorrectly.
Connect and set the printer correctly, referring to the operator’s manual of the
printer. Close the Data Management Software before changing the Windows
printer settings.
Q.4 The “ Registration code required” message is displayed when the Data
Management Software is started.
How to operate the LIS, Data Management Software PC, and Hematology Analyzer with LIS
............................................................................................................................................ 6.1
The Data Management Software has a Direct LIS Connection .................................. 6.1
Procedure in Data Communication between LIS, Data Management Software
and Hematology Analyzer ............................................................................... 6.2
When the Data Management Software is Connected to the LIS through a Gateway
Software .................................................................................................................... 6.4
Procedure in Data Communication between Gateway, Data Management
Software and Hematology Analyzer ................................................................ 6.5
Gateway Sending the Order Information to Receiving the Measured Data ...... 6.7
Quality Control .................................................................................................................... 6.9
QC Graph ...................................................................................................... 6.12
QC Limit Setting ........................................................................................... 6.14
Database ........................................................................................................................... 6.15
Deleting Data ................................................................................................ 6.16
Backing Up the Database .............................................................................. 6.16
Data Restore ................................................................................................. 6.17
Levey & Jennings Program ..................................................................................... 6.18
How to Read the L-J Chart ............................................................................ 6.18
Files Created and Managed by the Data Management Software ....................................... 6.20
The LIS and Data Management Software PC are connected to the analyzer by LAN.
The Data Management Software and analyzer are connected by USB.
The Data Management This explanation assumes that the Data Management Software PC manages the
Software has a Direct LIS measured data on the hematology analyzer.
Connection
Orders are sent from the LIS to the Data Management Software PC. The Data
Management Software combines the received orders from the LIS on one order list,
processes the order list as the work list, and sends the work list to the hematology
analyzer. The hematology analyzer receives the work list only after confirming
whether to receive it. The hematology analyzer performs measurement according to
the work list, and sends the measured data to the Data Management Software PC.
Measured data are saved on the Data Management Software PC. The user sends the
measured data from the Data Management Software PC to the LIS.
The hematology analyzer performs measurement according to the work list and
sends the measured data to the Data Management Software PC. The received
measured data is saved to the hard disk of the Data Management Software PC. The
user sends the measured data from the Data Management Software to the LIS.
Measured Data
based on the order Measured Data
1. LIS
Orders are sent from the LIS to the Data Management Software PC. Through the
LIS, you can add, delete, or edit orders on the Data Management Software.
[Add] registers the order on the order list on the Data Management Software.
[Delete] or [Edit] applies to registered orders. The LIS cannot manage work lists
that were already sent.
3. MEK-8222
Set a rack with measuring samples on the MEK-8222. Set the MEK-8222 on the
READY screen to Work list for [Parameters].
4. MEK-8222
Each time the Data Management Software receives an order, it edits the order
list and sends the work list to the MEK-8222. The MEK-8222 displays a
confirmation dialog box to receive the work list. Selecting [YES] overwrites the
work list received on a previously received work list.
NOTE
MEK-8222 can receive work list only on the READY screen or the WORK
LIST screen.
5. MEK-8222
Work list measurement starts. During measurement, The MEK-8222 reads bar
codes on the samples to acquire IDs. The analyzer uses the work list to perform
measurement in accordance with the order. After finishing measurement,
measured data are automatically sent to the Data Management Software. If the
analyzer reads an ID not on the work list or a sample with no bar code,
measurement is not performed.
7. MEK-8222
After measuring all samples, display the READY screen. The measured status
of each rack is indicated by colors as follows:
8. MEK-8222
If there are more orders from the LIS, the Data Management Software sends a
work list again. Add any new samples to the rack with not measured samples.
Repeat steps 3 to 8 until all samples are measured.
When the Data Management This explanation assumes that the Gateway Server manages the measured data
Software is Connected to the on the hematology analyzers.
LIS through a Gateway
Software Orders are sent from the LIS to the Gateway Server. The Gateway Server sends
the orders to the appropriate hematology analyzer. The hematology analyzer
receives the work list, performs measurement according to the work list, and
sends the measured data to the Data Management Software PC. The measured
result is sent to the Gateway Server. The Gateway Server manages the measured
result.
When the Gateway Software is installed in the Gateway Server, it automatically
sends and receives data to and from the analyzer. So, no Data Management
Software operation is needed.
Network
Gateway
Software
HOST(LIS)
Another
Device
Another
Device
Data Flow between Gateway, Data Management Software and Hematology Analyzer
Gateway
Software
Measured Data
based on orders Measured Data
LIS
Data Management Software Hematology Analyzer
1. LIS
Orders are sent from the LIS to the Data Management Software PC. Through the
LIS, you can add, delete, or edit orders on the DMS. [Add] registers the order
on the order list on the Data Management Software. [Delete] or [Edit] applies to
registered orders. The LIS cannot manage the work lists that were already sent.
3. MEK-8222
Set a rack with measuring samples on the MEK-8222. MEK-8222 on the READY
screen, set as follows:
Measuring Parameter: Worklist
4. MEK-8222
Each time the Data Management Software receives an order, it edits the order
list and sends the work list to the MEK-8222 .The MEK-8222 displays a
confirmation dialog box to receive the work list. Selecting [YES] overwrites the
work list received a previously received work list.
NOTE
MEK-8222 can only receive the work list only on the READY screen or
the WORK LIST screen.
5. MEK-8222
Work list measurement starts. During measurement, The MEK-8222 reads the
bar codes on the samples to acquire IDs. The analyzer uses the work list to
perform measurement. After finishing measurement, measured data are
automatically sent to the Data Management Software. If the analyzer reads an
ID not on the work list or a sample with no bar code, measurement is not
performed.
7. MEK-8222
After measuring all samples, display the READY screen. The measured status
of each rack is indicated as follows:.
BLUE: Measured normally. Remove them from the rack.
RED: Alarm during measurement. Remove them from the rack.
WHITE: Not measured. Leave these samples on the rack. These samples either
have not received any order, have no bar code, or have a wrong bar
code. Wait for the order or check the bar codes.
8. MEK-8222
If there are more orders from the LIS, the Data Management Software sends a
work list again. Add any new samples to the rack with unmeasured samples.
Repeat steps 3 to 8 until all samples are measured.
NOTE
The data on the Data Management Software is not automatically deleted
when the measured data is sent to the gateway. Periodically check and
delete data.
10. Gateway
Confirm that all measured data corresponding to the work order are sent.
1. Gateway software
Orders are sent from the Gateway software to the Data Management Software
PC. Through the LIS, you can add, delete, or edit orders on the DMS. [ADD]
registers the order on the order list on the Data Management Software. [Delete]
or [Edit] applies to registered orders. The LIS cannot manage work lists that
were already sent.
3. MEK-8222
Set a rack with measuring samples on the MEK-8222. MEK-8222 on the
READY screen, set as follows: Measuring Parameter: Work list
4. MEK-8222
Each time the Data Management Software receives an order, it edits the order
list and sends the work list to the work list to the MEK-8222. The MEK-8222
displays a confirmation dialog box to receive the work list. The MEK-8222
displays a confirmation dialog box to receive the work list.
NOTE
MEK-8222 can only receive the work list on the READY or work list
screen.
5. MEK-8222
Work list measurement starts. During measurement, The MEK-8222 reads the
bar codes on the samples to acquire IDs. The analyzer uses the work list to
perform measurement in accordance with the order. After finishingmeasurement,
measured data are automatically sent to the Data Management Software. If the
analyzer reads an ID not on the work list or a sample with no bar code,
measurement is not performed.
7. MEK-8222
After measuring all samples, display the READY screen. The measured status
of each rack is indicated as follows:
BLUE: Measured normally. Remove them from the rack.
RED: Alarm during measurement. Remove them from the rack.
WHITE: Not measured. Leave these samples on the rack. These samples
either have not received any order, have no bar code, or have a
wrong bar code. Wait for the order or check the bar codes.
8. MEK-8222
If there are more orders from the LIS, the Data Management Software sends a
work list again. Add any new samples to the rack with not measured samples.
Repeat steps 3 to 8 until all samples are measured.
Quality Control
A graph of quality control data received from the hematology analyzer can be
displayed.
2. The quality control screen is displayed. Select the hematology analyzer under
Quality control.
QC Graph
[Print chart]
Print the enlarged graph on display.
[Print page]
Print all graphs on the lower screen.
* When [Selecting mode of no print items] is checked, clicking a graph changes its
background color. The selected graph is not printed.
QC Limit Setting
Use this screen to enter QC limits when the control blood lot number is changed.
Database
Deleting Data
Deleting QC data
1. Click [Delete QC]. A confirmation dialog box appears.
2. Click [YES]. QC data is deleted.
* You can free space only by deleting data and checking [Delete after pack].
4. Click [OK]. Database backup starts. After completion, the following message
box appears.
Data Restore
Levey & Jennings Program The Levey & Jennings program counts one hematology control sample every day
and the mean and standard deviation (SD) are automatically calculated to plot L-J
chart for quality control.
L-J chart
4.90
+3 SD
+2 SD
4.80
2 SD
4.70 3 SD
The following files are created and managed by the Data Management Software.
AutoWorklistData.csv:
Used in file trigger mode. This is the work list to be automatically sent to the
hematology analyzer. The file name cannot be changed.