Vous êtes sur la page 1sur 3
‘Single-click this logo position Single-click this position(2nd) Input secret: 5355, then press “ENT” Hand-Switch to do adjusting value mor this setting. Choose KV value at 40, do not change mA value. Press exposure. If mA test value >77, adjust mA set value(minimum re than 10) to make mA test value to 7, then press "Rec" to save If mA test value between 77 and 74, do not need adjust mA set value KV 50 mAtest 68 KV 60 mAtest 60 kV 70 mAtest 57 kV 80 mAtest 53 KV 90 mAtest 47 kV 100 mAtest 37 kv 110 mAtest 29 Choose KV value at 50, do not change mA value. Press Hand-Switch to do exposure. If mA test value >68, adjust mA set value(minimum adjusting value more than 10) to make mA test value to 68, then press "Rec” to save this setting If MA test value between 68 and 65, do not need adjust mA set value, Then do same setting at KV60, KV70, KVB0, KV90, KV100 & KV110. After all setting, please press "Back", then press "Photography mode", then you can use XM101F well