Académique Documents
Professionnel Documents
Culture Documents
John Shinoda
Senior Support Specialist
Waters Corporation
November 5, 2009
Business Issues
Analytical Changes
— Chromatography
— Detector considerations
— Software considerations
Test Case Study
— Budesonide method development
— Budesonide batch analysis
Leachables
— Polymer Additives
— Detector Considerations
Provide analytical results for all batches of drug substance used for
clinical, safety, stability testing; also for batches representative of
proposed commercial process
Guidance for Industry, Q3A Impurities in New Drug Substances, February 2003; \\CDS029\CDERGUID\4164fnl.doc
Draft Guidance for Industry, ANDAs: Impurities in Drug Substances, January 2005; J:!GUIDANC\6422dft.doc
©2007 Waters Corporation 5
Impurity Thresholds
Drug Substances
Threshold
Maximum
Daily Dose Reporting Identification Qualification
0.10% 0.15%
≤ 2 g/day
or or
0.05%
1.0 mg/day intake 1.0 mg/day intake
(lower) (lower)
1.419
3.246
3.700
0.006 Lovastatin Simvastatin
1.184
4.609
4.713
3.443
0.002
6.786
5.187
5.372
1.309
2.855
1.758
0.000
0.016
0.014
0.012
0.010
0.008
0.006
0.004
0.002
0.000
0.565 0.570 0.575 0.580 0.585 0.590 0.595 0.600 0.605 0.610 0.615 0.620 0.625 0.630
Minutes
Sampling Rate Points Across Peak Peak Area %RSD Peak Height %RSD
1 pt/s 2 2.436 15.515
2 pts/s 4 1.790 13.455
5 pts/s 7 0.971 3.962
10 pts/s 13 1.129 1.015
20 pts/s 25 0.603 1.156
40 pts/s 49 0.284 1.127
©2007 Waters Corporation 11
Detector Dynamic Range
3.5
Linearity of detector > linear 3 1.5% Deviation at 2.0 AU
range of impurity and parent 2.5
5.0% Deviation at 2.7 AU
Absorbance (AU)
large parent compounds 0.5 PDA - Analytical Flow Cell
PDA - High Sensitivity Flow Cell
simultaneously 0
0 10 20 30 40 50 60 70 80
3.5
3
Analytical flow cell 2.2% Deviation at 2.5 AU
2.5 5.0% Deviation at 3.0 AU
— 10mm pathlength
2 2.4% Deviation at 2.5 AU
High sensitivity cell 5.0% Deviation at 3.0 AU
1.5
— 25 mm pathlength 1
UV - Analytical Flow Cell
0.5
UV - High Sensitivity Flow Cell
0
0 10 20 30 40 50 60 70 80
% Absorber in Eluent
Prilocaine Spectrum
Prilocaine
o-Toluidine
o-Toluidine Spectrum
0.01% Impurity
Expanded View
Part I
Re-development of European Pharmacopoeial (EP) Assay for
Budesonide Using UPLC®
Goal
—Decrease raw material inventory costs by shortening the time to
accept new lots of budesonide drug substance
Analytical Needs
—Method development
o Explore column, pH, organic solvent, temperature
—Meet EP assay criteria
—MS compatible solvent system
o Single quadrupole mass detection for impurity confirmation
0.040
R-epimer S-epimer
0.035
0.030
0.025
AU
0.020
0.015
0.010
0.005
0.000
2.00 4.00 6.00 8.00 10.00 12.00
Minutes
14.00 16.00 18.00 20.00 22.00 24.00
26.00
0.005
0.000
0.005
0.000
0.010
BEH Phenyl
AU
0.005
0.000
0.010
BEH Shield
AU
0.005
0.000
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00
Minutes
0.015 30 °C
0.010
AU
0.005
0.000
0.015 40 °C
0.010
AU
0.005
0.000
0.015
60 °C
0.010
AU
0.005
0.000
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00
Minutes ©2007 Waters Corporation 21
Budesonide
Final UPLC Method
0.028 0.80
R-epimer S-epimer
Conditions
0.40 Column: ACQUITY UPLC BEH C18
Dimensions: 100 x 2.1, 1.7um
Mobile Phase A: 20mM Ammonium
0.021 0.00 formate, pH 3.2
Mobile Phase B: acetonitrile
0.00 3.00 6.00 Flow Rate: 0.6 mL/min
Isocratic: 68%A : 32%B
Injection Volume: 5.0 µL
Temperature: 40C
7 Detection: UV @ 240 nm
AU
0.007 3 4 6
2 5
0.000
0.00 0.80 1.60 2.40 3.20 4.00 4.80 5.60 6.40 7.20 8.00
Minutes 8.00
©2007 Waters Corporation 22
Budesonide EP System Suitability
Specifications
R-epimer - 4.819
S-epimer - 5.194
0.028 Peak Results
Width
Name RT % Area Area % Height Height Signal_to_Noise
(sec)
1 Peak1 0.647 0.11 9450 0.60 9007 3.450 72.755
2 Peak2 1.860 0.08 7109 0.18 2733 8.900 22.079
3 Peak3 2.018 0.21 18639 0.40 5960 11.100 48.137
Peak7 - 4.291
5 Peak5 3.249 0.19 16603 0.19 2793 12.150 22.556
6 Peak6 4.111 0.33 28804 0.39 5908 12.850 47.718
7 Peak7 4.291 0.81 70719 0.93 14019 15.500 113.234
8 R-epimer 4.819 57.91 5081008 58.80 885434 19.200 7151.854
Peak1 - 0.647
Peak3 - 2.018
Peak6 - 4.111
Peak4 - 2.598
Peak2 - 1.860
Peak5 - 3.249
Peak10 - 5.753
0.007
0.000
0.00 0.80 1.60 2.40 3.20 4.00 4.80 5.60 6.40 7.20 8.00
Minutes
Part II
Qualification of Multiple Batch Lots of Budesonide
from Different Suppliers
Goal
— To determine most consistent and cost-effective supplier of
budesonide
Analytical needs
— Quality of each manufacturer’s batch lot of budesonide need to be
evaluated
— Perform the EP related substances test on each lot
o Individual impurities
o Total impurities
o R/S epimer ratio
o % purity
— Confirmation of detected impurities ©2007 Waters Corporation 25
EP Related Substances Test
*There should be 3 standard preps for each lot of substance (12 total)
©2007 Waters Corporation 26
Required Tests
1. Individual Impurities
x < 2.5µg/mL ∑ of epimers areas in the 500 µg/mL test
solution
2. Total Impurities
x < 7.5µg/mL ∑ of epimers areas in the 500 µg/mL test
solution
3. R-epimer/S-epimer Ratio
S-epimer is 40.0% to 51% of the ∑ of epimers areas in the
500 µg/mL test solution
4. Purity
98% to 102% (as the case for most raw material qualification for
use as an authentic reference standard)
Supplier A
Supplier B
Supplier C1
Supplier C2
Supplier A
Supplier B
Supplier C1
Supplier C2
Supplier A Supplier C2
European Pharmacopoeia Supplier B Supplier C1
>99% EP grade
Related Substances Test 100.2% R&D grade
purity 98% -
Specification (no spec)
102%
Individual Impurities
(x < 2.5 µg/mL ∑ of epimers Fail Pass Fail Fail
areas)
Total Impurities
(x < 7.5µg/mL ∑ of epimers Fail Pass Fail Fail
areas)
R-epimer/S-epimer Ratio
50.49%/ 51.38%/ 58.66%/ 59.24%/
(S-epimer is 40.0% to 51%
49.51 48.62 41.34 40.76%
∑ of epimers areas)
Plastic containers
Metal container coatings
Elastomeric closures or septa
Label Adhesives
Printing inks
1 O HO OH
2 O 3
Lowilite 20
Chimassorb 81
Lowinox TBM6
OH OH
NH 2
HO
O
4 N
N 6
O
Irganox 1035 O Cl N
S
5 Erucamide
O O
Tinuvin 326
HO
OH
N
N
O
Cl N
7 8
Lowilite 27 Vitamine E
O S O
O O
Irganox PS 800 9
N N
N N
N OH OH N
10
Lowilite 36
©2007 Waters Corporation 36
Typical Sample Preparation uses
SPE
3
0.30 1 Lowilite 20
1
2 Lowinox TBM6
3 Chimasorb 81
4 Irganox 1035
5 Tinuvin 326
6 Erucamide (?)
7 Lowilite 27
8 Vitamin E
2 9 Irganox PS 800 (?)
AU
10
4
8
0.00
0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40
Minutes
PDA Timed wavelength chromatogram (0 min, 320 nm; 0.6 min, 275 nm)
Lowilite 20
Lowilite 27
Vitamin E
Lowilite 36
3
Erucamide
7
4
5
120.00
6
Irganox PS 800
8
LSU
60.00
10
9
1
0.00
0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40
Minutes
ELS Chromatogram of 40 ppm polymer additives
Calibration Plot
100000.0
Erucamide (6)
75000.0
Area
50000.0
0.0